Articles

Roc HPLC Columns: The Reliable Cornerstone of Your LC Lab

Conventional HPLC columns built for solid performance, exceptional value, and compact storage.

17 Dec 2020

header article GNSS2332B
  • Conventional HPLC column built to be the cornerstone for your lab; pressure rated for any 400 bar HPLC system.
  • Solid and reliable—delivers the peak shape, reproducibility, ruggedness, and performance you demand.
  • Exceptional value for routine analyses and large-volume workflows—high-purity, fully porous silica backed by quality manufacturing features a variety of phases to fit most HPLC methods.
  • Ideal for a wide range of applications from food to pharma; 5 phases cover 98% of USP methods.
  • Newly designed minimal packaging for compact storage—download serial #-specific certificates of analysis at restek.com/documentation

Roc Family of HPLC Phases

 C18
figure article GNSS2332B 01
C8
figure article GNSS2332B 02
Phenyl-Hexyl
figure article GNSS2332B 03
Cyano
figure article GNSS2332B 04
Silica
figure article GNSS2332B 05
USP Phase CodeL1L7L11L10L3
Stationary Phase CategoryC18, octadecylsilaneC8, octylsilanephenylcyanobare silica
Ligand Typemonomeric C18monomeric C8phenyl-hexylcyanopropyln/a
Particle Size3 μm or 5 μm, spherical3 μm or 5 μm, spherical3 μm or 5 μm, spherical3 μm or 5 μm, spherical3 μm or 5 μm, spherical
Pore Size100 Å100 Å100 Å100 Å100 Å
Surface Area300 m2/g300 m2/g300 m2/g300 m2/g300 m2/g
Carbon Load20%12%15%8%n/a
End-Capyesyesyesyesn/a
pH Range2.5 to 8.02.5 to 8.02.0 to 8.02.5 to 8.02.5 to 8.0
Maximum Temperature80 °C80 °C80 °C80 °C80 °C
Figure 1: Over 1,000 injections and Roc HPLC columns will still offer reliable performance, regardless of which dimension your method calls for.
decorative
Figure 2: Add a Roc HPLC guard to further extend the life of your analytical column without impacting your retention, peak shape, or efficiency.
Net Chromatographic Effects of Roc C8 with Guard Column

LC_GN0565

Peaks

PeaksConc.
(mg/mL)
1.Uracil0.02
2.Toluene3.00
3.Naphthalene0.50
4.Biphenyl0.06

Conditions

ColumnRoc C8 (cat.# 9533565)
Dimensions:150 mm x 4.6 mm ID
Particle Size:5 µm
Pore Size:100 Å
Guard Column:Roc C8 10 mm, 4.0 mm ID, 5 µm (cat.# 953350210)
Temp.:30 °C
Standard/SampleCustom mix
Diluent:50:50 Acetonitrile:water
Inj. Vol.:2 µL
Mobile Phase
A:Water
B:Acetonitrile
Time (min)Flow (mL/min)%A%B
0.001.03070
8.001.03070
DetectorUV @ 254 nm
InstrumentHPLC
Figure 3: From one column to the next, one lot to the next, Roc columns provide results you can trust time and time again.

Reversed Phase

Retention Time Stability for Roc C18

LC_GN0562

Peaks

PeakstR (min)
1.p-Tolunitrile3.7
2.4-Butylbenzoic acid5.2
3.3,4-Dimethylanisole7.3
4.Mefenamic acid8.1
5.p-Xylene9.7

Conditions

ColumnRoc C18 (cat.# 9534565)
Dimensions:150 mm x 4.6 mm ID
Particle Size:5 µm
Pore Size:100 Å
Temp.:35 °C
Standard/Sample
Inj. Vol.:5 µL
Mobile Phase0.05% Formic acid in water:0.05% Formic acid in acetonitrile (40:60)
Flow:1.0 mL/min
DetectorUV/Vis @ 220 nm
Cell Temp:35 °C
InstrumentHPLC

Normal Phase

Retention Time Stability for Roc Cyano

LC_GN0563

Peaks

PeakstR (min)
1.Benzene3.5
2.Nitrobenzene4.1
3.1,3,5-Trinitrobenzene5.5
4.2,4-Dinitrotoluene7.4

Conditions

ColumnRoc Cyano (cat.# 953631E)
Dimensions:100 mm x 3.0 mm ID
Particle Size:3 µm
Pore Size:100 Å
Temp.:40 °C
Standard/Sample
Inj. Vol.:1 µL
Mobile PhaseWater:acetonitrile (80:20)
Flow:0.6 mL/min
DetectorUV/Vis @ 254 nm
Cell Temp:40 °C
InstrumentHPLC

Roc HPLC Columns from Restek Perform Better Than or Equal To Columns from Other Vendors

ColumnsAnalytetRAreaHeightUSP TailingAsymmetry (10%)k’Plate NumberResolution
RocEthylparaben3.2521,407,609281,3581.1761.1481.1829,73812.5
Ethinyl Estradiol5.4101,869,684234,3461.0751.0642.63110,264
Vendor AEthylparaben3.0601,424,621215,6981.9071.6081.0136,73210.1
Ethinyl Estradiol4.8891,860,659223,8791.3111.2172.2178,575
Vendor BEthylparaben2.8731,426,068319,8501.1431.1240.8669,7069.9
Ethinyl Estradiol4.3071,867,049287,4141.0641.0571.79710,031
Vendor CEthylparaben2.7631,413,099315,0801.2621.2170.9328,64311.9
Ethinyl Estradiol4.5801,868,724270,6621.1431.1222.2039,700
Vendor DEthylparaben3.1572,939,187357,5411.1590.9651.0844,47812.2
Ethinyl Estradiol5.8161,771,785182,0561.3011.2322.8398,954
Vendor EEthylparaben3.8731,498,446244,6421.1721.1341.2219,34813.7
Ethinyl Estradiol6.8391,781,453174,9551.0801.0632.9219,963
Figure 4: Count on Roc HPLC columns from Restek for outstanding chromatographic performance.

Roc HPLC Column

Ethinyl Estradiol on Roc C18 - USP Method

LC_EX0518

Peaks

PeakstR (min)Conc.
(µg/mL)
1.Ethylparaben3.25250
2.Ethinyl estradiol5.410200

Vendor A

Ethinyl Estradiol on Vendor A C18 - USP Method

LC_EX0521

Peaks

PeakstR (min)Conc.
(µg/mL)
1.Ethylparaben3.06050
2.Ethinyl estradiol4.889200

Vendor B

Ethinyl Estradiol on Vendor B C18 - USP Method

LC_EX0520

Peaks

PeakstR (min)Conc.
(µg/mL)
1.Ethylparaben2.87350
2.Ethinyl estradiol4.307200

Vendor C

Ethinyl Estradiol on Vendor C C18 - USP Method

LC_EX0519

Peaks

PeakstR (min)Conc.
(µg/mL)
1.Ethylparaben2.76350
2.Ethinyl estradiol4.580200

Vendor D

Ethinyl Estradiol on Vendor D C18 - USP Method

LC_EX0522

Peaks

PeakstR (min)Conc.
(µg/mL)
1.Ethylparaben3.15750
2.Ethinyl estradiol5.816200

Vendor E

Ethinyl Estradiol on Vendor E C18 - USP Method

LC_EX0523

Peaks

PeakstR (min)Conc.
(µg/mL)
1.Ethylparaben3.87350
2.Ethinyl estradiol6.839200

Chromatographic conditions for Table I and Figure 4
All columns are C18 stationary phase (5 µm, 150 x 4.6 mm). Conditions are as follows. Mobile phase: water:acetonitrile (1:1); Flow rate: 1 mL/min; Injection volume: 25 µL; Instrument: HPLC with UV detector, 280 nm; Sample: 200 µg/mL ethinyl estradiol and 50 µg/mL ethylparaben in water:acetonitrile (1:1).

Get more of what you want, and less of what you don’t, with Roc HPLC columns from Restek.

Author

  • Restek Corporation

    Restek is a leading provider of chromatography columns, accessories, and certified reference materials. Trust Restek for reliable, high-quality analytical solutions.

    View all posts
GNSS2332
Version: B