Articles

Packed Column USP Phase and Support Cross-Reference Chart

26 Oct 2020

Table I:  Cross-reference a non-Restek packed column phase or support to an equivalent Restek packed column phase or support.

LIQUID PHASESUSP NomenclatureRestek EquivalentNotes
Dimethyl polysiloxane oilG1Rtx-1, Rt-2100, OV-101 
100% Dimethylpolysiloxane gum, polydimethylsiloxane, or polymethylsiloxaneG2  Rtx-1, OV-1   
50% Phenyl-50% methylpolysiloxaneG3Rt-2250, OV-17 
DEGSG4DEGSDiethylene glycol succinate polyester
3-CyanopropylpolysiloxaneG5Rt-2340 
TrifluoropropylmethylpolysiloxaneG6Rt-2401 
50% 3-Cyanopropyl-50% phenylmethylsiliconeG7Rt-2300 
MethylvinylpolysiloxaneG9Rtx-1 
Phenyldiethanolamine succinate polyesterG12Phenyldiethanolamine succinate polyester 
D-SorbitolG13SorbitolSame properties without binding.
Polyethylene glycol (avg. mol. wt. of 950 to 1050) or PEG 1000G14Carbowax 1000 (CW-1000) g/mole 
Macrogol 20000 or PEG 20MG16Carbowax 20M (CW-20M) g/mole; StabilwaxPolyethylene glycol compound (avg. mol. wt. about 15,000). A high molecular weight compound of polyethylene glycol with a diepoxide linker.
25% Phenyl-25% cyanopropyl-50% methylsilicone or p/p poli (cyano-propyl)(methyl)(phenyl)(methyl) siloxaneG19OV-225 
Polyethylene glycol (avg. mol. wt. of 380 to 420)G20Carbowax 400 (CW-400) g/mole 
Diisodecyl phthalateG24Diisodecyl phthalate (DIDP) 
Polyethylene glycol compound TPA. A high molecular weight compound of a polyethylene glycol and a diepoxide that is esterified with terephthalic acid.G25Carbowax 20 TPA (CW-20MTPA) g/mole 
25% 2-Cyanoethyl-75% methylpolysiloxaneG26XE-60 
5% Phenyl-95% methylpolysiloxaneG27Rtx-5, SE-52 
25% Phenyl-75% methylpolysiloxaneG28DC 550 
CO880 or nonylphenoxypoly(ethyleneoxy)ethanol (avg. ethyleneoxy chain length is 30); nonoxynol 30.G31IGEPAL-CO 880 
AT-1000 or a high molecular weight compound of a polyethylene glycol and a diepoxide that is esterified with nitroterephthalic acid.G35Rt-1000 
1% vinyl-5% phenylmethylpolysiloxaneG36Rtx-5, SE-54 
SOLID SUPPORTSUSPRestek EquivalentNotes
Siliceous earth flux calcined, acid washed and water washed until neutralS1ASilcoport-W, DiatoSorb-WAW, DiatoSorb-WHP 
A support prepared from crushed firebrick and calcined or burned with a clay binder above 900 °C with subsequent acid wash. It may be silanized.S1CChromosorb PAW, Chromosorb PAW DMDCS 
A support prepared from crushed firebrick and calcined or burned with a clay binder above 900 °C, not acid washed. It may be silanized.S1DChromosorb PNAW 
Untreated siliceous earthS1NSDiatoSorb-WNAW 
Styrene-divinylbenzene co-polymer having a nominal surface area of less than 50 m2 per gram and an average pore size of 0.3 to 0.4 μmS2Chromosorb 101 
Ethyl vinylbenzene-divinylbenzene co-polymer having a nominal surface area of less than 500-600 m2 per gram and an average pore size of 0.0075 μmS3HayeSep Q 
Styrene-divinylbenzene copolymer with aromatic -O and -N groups having a nominal surface area of 400 to 600 m2/g and an average pore diameter of 0.0076 μmS4Porapak R 
High molecular weight tetrafluoroethylene polymer 40 to 60 meshS5Chromosorb T 40/60 
Styrene-divinylbenzene co-polymer having a nominal surface area of less than 250 – 300 m2 per gram and an average pore size of 0.0091 μmS6HayeSep P 
Graphitized carbon having a nominal surface of 12 m2 per gramS7CarboBlack C 
Copolymer of 4-vinylpyridine styrene-divinylbenzeneS8HayeSep S 
Porous polymer based on 2,6-diphenyl-p-phenylene oxideS9Tenax TA 
Highly cross-linked co-polymer of acrylonitrile and divinylbenzeneS10HayeSep C 
Graphitized carbon having a nominal surface of 100 m2 per gram modified with small amounts of petrolatum and polyethylene glycol compoundS11CarboBlack B 80/120 3% Rt-1500 
Graphitized carbon having a nominal surface area of 100 m2 per gS12CarboBlack B 

Our packed columns are prepared with SilcoSmooth tubing and stable bonded-phase technology resulting in a column with the inertness of glass and the strength of stainless steel. Learn more about our packed columns in this technical article.

Author

  • Restek Corporation

    Restek is a leading provider of chromatography columns, accessories, and certified reference materials. Trust Restek for reliable, high-quality analytical solutions.

    View all posts
PHOT3287
Version: D