{"id":43479,"date":"2020-10-27T00:00:00","date_gmt":"2020-10-27T00:00:00","guid":{"rendered":"https:\/\/discover.restek.com\/uncategorized\/method-505-analysis-of-organohalide-pesticides-and-commercial-pcbs-in-water-by-microextraction-and-gc\/"},"modified":"2026-01-02T18:25:38","modified_gmt":"2026-01-02T18:25:38","slug":"method-505-analysis-of-organohalide-pesticides-and-commercial-pcbs-in-water-by-microextraction-and-gc","status":"publish","type":"post","link":"https:\/\/discover.restek.com\/fr\/articles-ja\/evar4504\/method-505-analysis-of-organohalide-pesticides-and-commercial-pcbs-in-water-by-microextraction-and-gc","title":{"rendered":"Method 505: Analysis of Organohalide Pesticides and Commercial PCBs in Water by Microextraction and GC"},"content":{"rendered":"\n<p class=\"wp-block-paragraph\">With the advent of modern agriculture and its vast selection of chemical pest control measures, the farming community has made significant increases in productivity and efficiency. Crop yield per acre is extremely high, due in part to the role of pesticides and herbicides in mitigating the devastating effects of many plant and insect pests. However, the use of these chemicals can have drawbacks, including surface and ground water contamination. EPA Method 505 is used for organohalide pesticide analysis in drinking and ground water. The&nbsp;Rtx-CLPesticides and Rtx-CLPesticides2 column set&nbsp;works well for the analysis of organohalide compounds because the columns are highly selective for analytes that contain electronegative substituents. The optimized parallel dual-column method shown here provides fast analysis times, which improves sample throughput.<\/p>\n\n\n\n<figure class=\"wp-block-table\"><table class=\"has-fixed-layout\"><tbody><tr><th class=\"has-text-align-left\" data-align=\"left\">Vendor<\/th><th class=\"has-text-align-left\" data-align=\"left\">Column Pair<\/th><th class=\"has-text-align-left\" data-align=\"left\">Analysis Time (min)<\/th><th class=\"has-text-align-left\" data-align=\"left\">Coelutions<\/th><th class=\"has-text-align-left\" data-align=\"left\">Restek Advantage<\/th><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Restek<\/td><td class=\"has-text-align-left\" data-align=\"left\"><a href=\"https:\/\/www.restek.com\/p\/11120\" target=\"_blank\" rel=\"noopener\">Rtx-CLPesticides<\/a>&nbsp;\/<br><a href=\"https:\/\/www.restek.com\/p\/11321\" target=\"_blank\" rel=\"noopener\">Rtx-CLPesticides2<\/a><\/td><td class=\"has-text-align-left\" data-align=\"left\">18 \/ 18.5<\/td><td class=\"has-text-align-left\" data-align=\"left\">1 \/ 1<\/td><td class=\"has-text-align-left\" data-align=\"left\" rowspan=\"3\">All compounds resolved on at least one column.<\/td><\/tr><tr><th class=\"has-text-align-left\" data-align=\"left\">Competitor A<\/th><td class=\"has-text-align-left\" data-align=\"left\">column 1 \/ column 2<\/td><td class=\"has-text-align-left\" data-align=\"left\">14 \/ 14<\/td><td class=\"has-text-align-left\" data-align=\"left\">0 \/ 1<\/td><\/tr><tr><th class=\"has-text-align-left\" data-align=\"left\">Competitor B<\/th><td class=\"has-text-align-left\" data-align=\"left\">column 1 \/ column 2<\/td><td class=\"has-text-align-left\" data-align=\"left\">35 \/ 36<\/td><td class=\"has-text-align-left\" data-align=\"left\">1 \/ 2<\/td><\/tr><\/tbody><\/table><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\">Comparison based on published competitor data. All columns tested were 0.32 mm ID.<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><strong>Organohalide pesticide analysis (Method 505) on Rtx-CLPesticides and Rtx-CLPesticides2 columns.<\/strong><\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n<div class=\"wp-block-custom-chromatogram-article\"><div class=\"wp-block-custom-chromatogram-article\"><div class=\"chromatogram-image regular-image\"><img decoding=\"async\" src=\"https:\/\/ez.restek.com\/images\/cgram\/gc_ev01004.png\" alt=\"Pesticides of U.S. EPA Method 505 on Rtx-CLPesticides and Rtx-CLPesticides2\" title=\"-\"><\/div><p class=\"article-id\" style=\"text-align:center\"> GC_EV01004<\/p><div class=\"chromatogram-peaks\"><h4>Peaks<\/h4><table class=\"peaks col-lg-6 col-12 peak-50\">\n<thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><\/tr><\/thead>\n<tbody><tr><td class=\"num\">1.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Hexachlorocyclopentadiene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/77-47-4\/Hexachlorocyclopentadiene\" target=\"_blank\" rel=\"noopener\">Hexachlorocyclopentadiene<\/a><\/td><\/tr>\n<tr><td class=\"num\">2.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Hexachlorobenzene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/118-74-1\/Hexachlorobenzene\" target=\"_blank\" rel=\"noopener\">Hexachlorobenzene<\/a><\/td><\/tr>\n<tr><td class=\"num\">3.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Simazine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/122-34-9\/Simazine\" target=\"_blank\" rel=\"noopener\">Simazine<\/a><\/td><\/tr>\n<tr><td class=\"num\">4.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Atrazine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/1912-24-9\/Atrazine\" target=\"_blank\" rel=\"noopener\">Atrazine<\/a><\/td><\/tr>\n<tr><td class=\"num\">5.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for \u03b3-BHC\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/58-89-9\/\u03b3-BHC\" target=\"_blank\" rel=\"noopener\">\u03b3-BHC<\/a><\/td><\/tr>\n<tr><td class=\"num\">6.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Heptachlor\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/76-44-8\/Heptachlor\" target=\"_blank\" rel=\"noopener\">Heptachlor<\/a><\/td><\/tr>\n<tr><td class=\"num\">7.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Alachlor\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/15972-60-8\/Alachlor\" target=\"_blank\" rel=\"noopener\">Alachlor<\/a><\/td><\/tr>\n<tr><td class=\"num\">8.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Aldrin\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/309-00-2\/Aldrin\" target=\"_blank\" rel=\"noopener\">Aldrin<\/a><\/td><\/tr>\n<\/tbody><\/table>\n<table class=\"peaks col-lg-6 col-12 peak-50\"><thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><\/tr><\/thead><tbody>\n<tr><td class=\"num\">9.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Heptachlor epoxide\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/1024-57-3\/Heptachlor epoxide\" target=\"_blank\" rel=\"noopener\">Heptachlor epoxide<\/a><\/td><\/tr>\n<tr><td class=\"num\">10.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for trans-Chlordane\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/5103-74-2\/trans-Chlordane\" target=\"_blank\" rel=\"noopener\">trans-Chlordane<\/a><\/td><\/tr>\n<tr><td class=\"num\">11.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for trans-Nonachlor\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/39765-80-5\/trans-Nonachlor\" target=\"_blank\" rel=\"noopener\">trans-Nonachlor<\/a><\/td><\/tr>\n<tr><td class=\"num\">12.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for cis-Chlordane\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/5103-71-9\/cis-Chlordane\" target=\"_blank\" rel=\"noopener\">cis-Chlordane<\/a><\/td><\/tr>\n<tr><td class=\"num\">13.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Dieldrin\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/60-57-1\/Dieldrin\" target=\"_blank\" rel=\"noopener\">Dieldrin<\/a><\/td><\/tr>\n<tr><td class=\"num\">14.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Endrin\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/72-20-8\/Endrin\" target=\"_blank\" rel=\"noopener\">Endrin<\/a><\/td><\/tr>\n<tr><td class=\"num\">15.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for cis-Nonachlor\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/5103-73-1\/cis-Nonachlor\" target=\"_blank\" rel=\"noopener\">cis-Nonachlor<\/a><\/td><\/tr>\n<tr><td class=\"num\">16.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Methoxychlor\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/72-43-5\/Methoxychlor\" target=\"_blank\" rel=\"noopener\">Methoxychlor<\/a><\/td><\/tr>\n<\/tbody><\/table><\/div><div class=\"chromatogram-conditions\"><h4>Conditions<\/h4><div class=\"conditions-container container-fluid\"><div class=\"row\"><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Columns<\/th><td>Rtx-CLPesticides 30 m, 0.32 mm ID, 0.32 \u00b5m (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/11141\" rel=\"noopener\">cat.# 11141<\/a>)<tr><td><\/td><td>and Rtx-CLPesticides2 30 m, 0.32 mm ID, 0.25 \u00b5m (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/11324\" rel=\"noopener\">cat.# 11324<\/a>)<tr><td><\/td><td>using Rxi guard column 5 m, 0.32 mm ID (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/10039\" rel=\"noopener\">cat.# 10039<\/a>)<tr><td><\/td><td>with deactivated universal \u201cY\u201d Press-Tight connector (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/20405-261\" rel=\"noopener\">cat.# 20405-261<\/a>)<tr><td><\/td><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Standard\/Sample<\/th><td><\/td><\/tr><tr><td><\/td><td>200 ng\/mL 505 organohalide pesticide mix (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/32024\" rel=\"noopener\">cat.# 32024<\/a>)<\/td><\/tr><tr><td><\/td><td>4.2 \u00b5g\/mL simazine (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/32236\" rel=\"noopener\">cat.# 32236<\/a>)<\/td><\/tr><tr><td><\/td><td>4.2 \u00b5g\/mL atrazine (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/32208\" rel=\"noopener\">cat.# 32208<\/a>)<\/td><\/tr>\n<tr><th class=\"sub conditions_header\" scope=\"row\">Diluent:<\/th><td>Methanol<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Injection<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Vol.:<\/th><td>2 \u00b5L splitless (hold 0.75  min)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Liner:<\/th><td>Cyclo double taper (4 mm) (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/20896\" rel=\"noopener\">cat.# 20896<\/a>)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Temp.:<\/th><td>250 \u00b0C<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\" colspan=\"2\">Oven<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Oven Temp.:<\/th><td>90 &#176;C (hold 1 min) to 310 &#176;C at 10 &#176;C\/min (hold 5 min)<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Carrier Gas<\/th><td>He, constant flow<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Linear Velocity:<\/th><td>40 cm\/sec<\/td><\/tr><\/table><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Detector<\/th><td>Micro-ECD @ 325 \u00b0C<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Instrument<\/th><td>Agilent\/HP6890  GC<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Notes<\/th><td>This chromatogram was obtained using an Agilent micro-ECD. To obtain comparable results, you will need to employ a micro-ECD in addition to confirmational dual columns connected to a 5-meter guard column using a \u201cY\u201d Press-Tight connector.<\/td><\/tr><\/table><\/div><\/div><\/div><div class=\"chromatogram-related-skus\"><h4>Related SKUs<\/h4><ul><li>32024<\/li><li>32236<\/li><li>32208<\/li><li>11141<\/li><li>11324<\/li><li>10039<\/li><li>20405-261<\/li><li>20896<\/li><\/ul><\/div><div class=\"chromatogram-related-searches\"><h4>Related Searches<\/h4><ul><li>BHC<\/li><li>Chlordane<\/li><li>Drinking Water<\/li><li>EPA 505<\/li><li>EPA Method 505<\/li><li>Organochlorine pesticides<\/li><li>Pesticides<\/li><li>Triazine Herbicides<\/li><\/ul><\/div><div class=\"chromatogram-pdf-link\"><a href=\"https:\/\/ez.restek.com\/images\/cgram\/gc_ev01004.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"><svg xmlns=\"http:\/\/www.w3.org\/2000\/svg\" width=\"18\" height=\"18\" viewBox=\"0 0 18 18\"><g data-name=\"Group 2996\"><path data-name=\"Rectangle 1246\" d=\"M0 0h18v18H0z\" style=\"fill: none;\"><\/path><\/g><g data-name=\"Group 2997\"><path data-name=\"Path 729\" d=\"M13.412 11.4v2.017H5.345V11.4H4v2.017a1.349 1.349 0 0 0 1.345 1.345h8.068a1.349 1.349 0 0 0 1.345-1.345V11.4zm-.672-2.694-.948-.948-1.741 1.735V4H8.706v5.493L6.965 7.758l-.948.948 3.361 3.361z\" transform=\"translate(-.437 -.414)\" style=\"fill: rgb(13, 123, 196);\"><\/path><\/g><\/svg>Download PDF<\/a><\/div><\/div><\/div>\n<\/div><\/div><\/div>\n\n\n<p class=\"wp-block-paragraph\"><\/p>\n\n\n\n<div style=\"height:50px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n        <div class=\"cpb\">\n            <h3 class=\"cpb-heading\">Produits mentionn\u00e9s<\/h3>\n            <hr class=\"cpb-heading-underline\" \/>\n            <div class=\"cpb-list\">\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/11196\">                                R\u00e9f. 11196                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Kit colonne capillaire Rtx-CLPesticides, DI 0.32\u00a0mm, 0.32 \u00b5m<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/11196\">Voir le produit<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/11120\">                                R\u00e9f. 11120                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Colonne capillaire Rtx-CLPesticides, L 15\u00a0m, DI 0.25 mm, 0.25 \u00b5m<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/11120\">Voir le produit<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/11321\">                                R\u00e9f. 11321                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Colonne capillaire Rtx-CLPesticides2, L 15 m, DI 0.32 mm, 0.25 \u00b5m<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/11321\">Voir le produit<\/a>\n                                                    <\/div>\n                    <\/div>\n                            <\/div>\n        <\/div>\n        \n\n\n<div style=\"height:100px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n","protected":false},"excerpt":{"rendered":"<p>Rtx-CLPesticides and Rtx-CLPesticides2 columns are ideal for multiple environmental ECD methods. Here we demonstrate the effectiveness of this column pair for organohalide pesticides by EPA Method 505.<\/p>\n","protected":false},"author":11,"featured_media":6751,"comment_status":"closed","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"_acf_changed":false,"_kad_blocks_custom_css":"","_kad_blocks_head_custom_js":"","_kad_blocks_body_custom_js":"","_kad_blocks_footer_custom_js":"","_kadence_starter_templates_imported_post":false,"_kad_post_transparent":"","_kad_post_title":"","_kad_post_layout":"","_kad_post_sidebar_id":"","_kad_post_content_style":"","_kad_post_vertical_padding":"","_kad_post_feature":"","_kad_post_feature_position":"","_kad_post_header":false,"_kad_post_footer":false,"footnotes":"","_ppma_block_editor_authors":""},"categories":[789],"tags":[],"industries-application":[2287,2271,2751],"post-badge":[],"resource-type":[],"product-library":[2552,2564,2579,2554],"resource-technique":[2340],"ppma_author":[2814],"class_list":["post-43479","post","type-post","status-publish","format-standard","has-post-thumbnail","hentry","category-articles-ja","industries-application-drinking-water-wastewater-testing-ja","industries-application-environmental-ja","industries-application-semivolatiles","product-library-gas-chromatography-products-ja","product-library-gc-columns-ja","product-library-liquid-reference-standards-ja","product-library-reference-standard-products-ja","resource-technique-gas-chromatography-gc-ja"],"acf":[],"taxonomy_info":{"category":[{"value":789,"label":"\u6280\u8853\u8cc7\u6599"}],"industries-application":[{"value":2287,"label":"\u98f2\u6599\u6c34\uff5c\u975e\u98f2\u6599\u6c34\uff5c\u6392\u6c34 \u5206\u6790"},{"value":2271,"label":"\u74b0\u5883 \u5206\u6790"},{"value":2751,"label":"Semivolatiles"}],"product-library":[{"value":2552,"label":"\u30ac\u30b9\u30af\u30ed\u30de\u30c8\u30b0\u30e9\u30d5\u30a3\u30fc"},{"value":2564,"label":"GC \u30ab\u30e9\u30e0"},{"value":2579,"label":"\u6a19\u6e96\u6eb6\u6db2"},{"value":2554,"label":"\u6a19\u6e96\u7269\u8cea"}],"resource-technique":[{"value":2340,"label":"\u30ac\u30b9\u30af\u30ed\u30de\u30c8\u30b0\u30e9\u30d5\u30a3\u30fc"}]},"featured_image_src_large":["https:\/\/discover.restek.com\/wp-content\/uploads\/feature-EVSS1013-1024x536.jpg",1024,536,true],"author_info":{"display_name":"Restek Corporation","author_link":"https:\/\/discover.restek.com\/fr\/author\/restek-corporation\/"},"comment_info":0,"category_info":[{"term_id":789,"name":"\u6280\u8853\u8cc7\u6599","slug":"articles-ja","term_group":0,"term_taxonomy_id":789,"taxonomy":"category","description":"","parent":0,"count":511,"filter":"raw","cat_ID":789,"category_count":511,"category_description":"","cat_name":"\u6280\u8853\u8cc7\u6599","category_nicename":"articles-ja","category_parent":0}],"tag_info":false,"authors":[{"term_id":414,"user_id":11,"is_guest":0,"slug":"restek-corporation","display_name":"Restek Corporation","avatar_url":{"url":"https:\/\/discover.restek.com\/wp-content\/uploads\/Restek_Favicon_300x300.jpg","url2x":"https:\/\/discover.restek.com\/wp-content\/uploads\/Restek_Favicon_300x300.jpg"},"author_category":"1","first_name":"Restek","last_name":"Corporation","user_url":"https:\/\/www.linkedin.com\/company\/restek-corporation\/posts\/?feedView=all","job_title":"Chromatography Products and Solutions","description":"Restek is a leading provider of chromatography columns, accessories, and certified reference materials. Trust Restek for reliable, high-quality analytical solutions."}],"_links":{"self":[{"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/posts\/43479","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/users\/11"}],"replies":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/comments?post=43479"}],"version-history":[{"count":7,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/posts\/43479\/revisions"}],"predecessor-version":[{"id":79168,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/posts\/43479\/revisions\/79168"}],"wp:featuredmedia":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/media\/6751"}],"wp:attachment":[{"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/media?parent=43479"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/categories?post=43479"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/tags?post=43479"},{"taxonomy":"industries-application","embeddable":true,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/industries-application?post=43479"},{"taxonomy":"post-badge","embeddable":true,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/post-badge?post=43479"},{"taxonomy":"resource-type","embeddable":true,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/resource-type?post=43479"},{"taxonomy":"product-library","embeddable":true,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/product-library?post=43479"},{"taxonomy":"resource-technique","embeddable":true,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/resource-technique?post=43479"},{"taxonomy":"author","embeddable":true,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/ppma_author?post=43479"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}