{"id":43572,"date":"2020-10-21T14:30:00","date_gmt":"2020-10-21T14:30:00","guid":{"rendered":"https:\/\/discover.restek.com\/uncategorized\/analyse-lc-de-multi-residus-de-pesticides-dans-la-farine-de-riz-brun-en-7-5-minutes\/"},"modified":"2026-01-28T19:15:27","modified_gmt":"2026-01-28T19:15:27","slug":"analyse-lc-de-multi-residus-de-pesticides-dans-la-farine-de-riz-brun-en-7-5-minutes","status":"publish","type":"post","link":"https:\/\/discover.restek.com\/fr\/articles-fr\/ffss2930-fr\/analyse-lc-de-multi-residus-de-pesticides-dans-la-farine-de-riz-brun-en-7-5-minutes","title":{"rendered":"Analyse LC de multi-r\u00e9sidus de pesticides dans la farine de riz brun en 7.5 minutes"},"content":{"rendered":"\n<ul class=\"wp-block-list\">\n<li>Analyse de 7.5 minutes avec un cycle analytique total de moins de 10 minutes.<\/li>\n\n\n\n<li>Plus de 90% des 200 r\u00e9sidus analys\u00e9s donnent des rendements compris entre 70 et 120 %.<\/li>\n\n\n\n<li>M\u00e9thode compatible avec les syst\u00e8mes HPLC et UHPLC.<\/li>\n<\/ul>\n\n\n\n<p class=\"wp-block-paragraph\">L\u2019analyse multi-r\u00e9sidus de pesticides n\u2019est pas chose ais\u00e9e, et elle est rendue encore plus difficile par les matrices d\u2019\u00e9chantillons complexes et l\u2019instrumentation LC-MS\/MS couramment utilis\u00e9es pour cette analyse. La LC-MS\/MS peut fournir une sensibilit\u00e9 remarquable et elle peut m\u00eame permettre des co\u00e9lutions chromatographiques tout en pouvant quantifier les analytes d\u2019int\u00e9r\u00eat. Mais trop de co\u00e9lutions peuvent surcharger le spectrom\u00e8tre de masse et une bonne m\u00e9thode chromatographique est donc essentielle. La m\u00e9thode chromatographique optimis\u00e9e d\u00e9crite ici permet l\u2019analyse de centaines de compos\u00e9s en moins de 7.5 minutes, garantissant un &#8220;screening&#8221; rapide et pr\u00e9cis des \u00e9chantillons et une productivit\u00e9 accrue du laboratoire.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Comme d\u00e9taill\u00e9 ci-dessous, un \u00e9chantillon de farine de riz brun a \u00e9t\u00e9 dop\u00e9 avec tous les compos\u00e9s inclus dans le Kit LC multi-r\u00e9sidus de pesticides Restek, extrait \u00e0 l\u2019aide de sels d\u2019extraction QuEChERS Q-sep, purifi\u00e9 par dSPE, et analys\u00e9 sur une colonne Raptor ARC-18 2.7 \u00b5m. L\u2019efficacit\u00e9 inh\u00e9rente \u00e0 une colonne faite de particules superficiellement poreuses de 2.7 \u00b5m offre des pics fins, ce qui permet d\u2019analyser plus de compos\u00e9s dans une unit\u00e9 de temps donn\u00e9e. La s\u00e9lectivit\u00e9 de la phase stationnaire de la colonne Raptor ARC-18 garantit que les compos\u00e9s \u00e9luent sur toute la dur\u00e9e de l\u2019analyse, ce qui \u00e9vite de surcharger le chromatogramme \u00e0 certains endroits et permet au spectrom\u00e8tre de masse de suivre plus efficacement les ions d\u2019int\u00e9r\u00eat. Notez qu\u2019il est important de r\u00e9duire au minimum le volume interne du syst\u00e8me donc utiliser des colonnes LC &#8220;narrow-bore&#8221; et \u00e9viter des longueurs de tubulures excessives aideront \u00e0 obtenir des pics plus fins et une chromatographie plus rapide et plus efficace. Cette analyse multi-r\u00e9sidus de pesticides dans la farine de riz brun utilisent des phases mobiles simples et compatibles avec une analyse MS, et un d\u00e9bit maintenant la contre-pression dans la plage de travail des instruments HPLC conventionnels. M\u00eame dans une matrice aussi complexe, tous ces avantages permettent d\u2019analyser plus de 200 pesticides et standards internes en quelques minutes.<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n<div class=\"wp-block-custom-chromatogram-article\"><div class=\"wp-block-custom-chromatogram-article\"><div class=\"chromatogram-image wide-image\"><img decoding=\"async\" src=\"https:\/\/ez.restek.com\/images\/cgram\/lc_fs0521.png\" alt=\"LC Multiresidue Pesticide Mix Extracted from Brown Rice Flour with QuEChERS Slim Pouch on Raptor ARC-18 by LC-MS\/MS\" title=\"-\"><\/div><p class=\"article-id\" style=\"text-align:center\"> LC_FS0521<\/p><div class=\"chromatogram-peaks\"><h4>Peaks<\/h4><table class=\"peaks col-lg-6 col-12 peak-50\">\n<thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 75px\">t<sub>R<\/sub> (min)<\/th><th style=\"text-align: center;width: 75px\">Precursor Ion<\/th><th style=\"text-align: center;width: 75px\">Product Ion 1<\/th><th style=\"text-align: center;width: 75px\">Product Ion 2<\/th><\/tr><\/thead>\n<tbody><tr><td class=\"num\">1.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Cyromazine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/66215-27-8\/Cyromazine\" target=\"_blank\" rel=\"noopener\">Cyromazine<\/a><\/td><td class=\"oth\">0.747<\/td><td class=\"oth\">167.1<\/td><td class=\"oth\">68.1<\/td><td class=\"oth\">43.1<\/td><\/tr>\n<tr><td class=\"num\">2.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Methamidophos\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/10265-92-6\/Methamidophos\" target=\"_blank\" rel=\"noopener\">Methamidophos<\/a><\/td><td class=\"oth\">1.318<\/td><td class=\"oth\">142.0<\/td><td class=\"oth\">94.0<\/td><td class=\"oth\">125.1<\/td><\/tr>\n<tr><td class=\"num\">3.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Formetanate\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/22259-30-9\/Formetanate\" target=\"_blank\" rel=\"noopener\">Formetanate<\/a><\/td><td class=\"oth\">1.681<\/td><td class=\"oth\">222.1<\/td><td class=\"oth\">165.0<\/td><td class=\"oth\">46.1<\/td><\/tr>\n<tr><td class=\"num\">4.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Carbofuran\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/1563-66-2\/Carbofuran\" target=\"_blank\" rel=\"noopener\">Carbofuran<\/a><\/td><td class=\"oth\">1.682<\/td><td class=\"oth\">222.2<\/td><td class=\"oth\">165.0<\/td><td class=\"oth\">123.1<\/td><\/tr>\n<tr><td class=\"num\">5.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Methabenzthiazuron\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/18691-97-9\/Methabenzthiazuron\" target=\"_blank\" rel=\"noopener\">Methabenzthiazuron<\/a><\/td><td class=\"oth\">1.686<\/td><td class=\"oth\">222.0<\/td><td class=\"oth\">165.0<\/td><td class=\"oth\">150.0<\/td><\/tr>\n<tr><td class=\"num\">6.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Acephate\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/30560-19-1\/Acephate\" target=\"_blank\" rel=\"noopener\">Acephate<\/a><\/td><td class=\"oth\">1.690<\/td><td class=\"oth\">184.0<\/td><td class=\"oth\">143.0<\/td><td class=\"oth\">49.0<\/td><\/tr>\n<tr><td class=\"num\">7.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Pymetrozine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/123312-89-0\/Pymetrozine\" target=\"_blank\" rel=\"noopener\">Pymetrozine<\/a><\/td><td class=\"oth\">1.708<\/td><td class=\"oth\">218.1<\/td><td class=\"oth\">105.0<\/td><td class=\"oth\">78.0<\/td><\/tr>\n<tr><td class=\"num\">8.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Aminocarb\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/2032-59-9\/Aminocarb\" target=\"_blank\" rel=\"noopener\">Aminocarb<\/a><\/td><td class=\"oth\">1.716<\/td><td class=\"oth\">209.1<\/td><td class=\"oth\">137.1<\/td><td class=\"oth\">152.1<\/td><\/tr>\n<tr><td class=\"num\">9.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Propamocarb\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/24579-73-5\/Propamocarb\" target=\"_blank\" rel=\"noopener\">Propamocarb<\/a><\/td><td class=\"oth\">1.735<\/td><td class=\"oth\">189.1<\/td><td class=\"oth\">102.0<\/td><td class=\"oth\">144.1<\/td><\/tr>\n<tr><td class=\"num\">10.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Omethoate\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/1113-02-6\/Omethoate\" target=\"_blank\" rel=\"noopener\">Omethoate<\/a><\/td><td class=\"oth\">1.905<\/td><td class=\"oth\">214.0<\/td><td class=\"oth\">125.0<\/td><td class=\"oth\">183.0<\/td><\/tr>\n<tr><td class=\"num\">11.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Dinotefuran\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/165252-70-0\/Dinotefuran\" target=\"_blank\" rel=\"noopener\">Dinotefuran<\/a><\/td><td class=\"oth\">1.972<\/td><td class=\"oth\">203.1<\/td><td class=\"oth\">129.1<\/td><td class=\"oth\">87.1<\/td><\/tr>\n<tr><td class=\"num\">12.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Aldicarb sulfoxide\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/1646-87-3\/Aldicarb sulfoxide\" target=\"_blank\" rel=\"noopener\">Aldicarb sulfoxide<\/a><\/td><td class=\"oth\">2.010<\/td><td class=\"oth\">207.1<\/td><td class=\"oth\">89.0<\/td><td class=\"oth\">132.1<\/td><\/tr>\n<tr><td class=\"num\">13.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Butoxycarboxim\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/34681-23-7\/Butoxycarboxim\" target=\"_blank\" rel=\"noopener\">Butoxycarboxim<\/a><\/td><td class=\"oth\">2.022<\/td><td class=\"oth\">223.0<\/td><td class=\"oth\">106.0<\/td><td class=\"oth\">166.0<\/td><\/tr>\n<tr><td class=\"num\">14.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Nitenpyram\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/120738-89-8\/Nitenpyram\" target=\"_blank\" rel=\"noopener\">Nitenpyram<\/a><\/td><td class=\"oth\">2.049<\/td><td class=\"oth\">271.0<\/td><td class=\"oth\">225.0<\/td><td class=\"oth\">56.0<\/td><\/tr>\n<tr><td class=\"num\">15.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Aldicarb sulfone\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/1646-88-4\/Aldicarb sulfone\" target=\"_blank\" rel=\"noopener\">Aldicarb sulfone<\/a><\/td><td class=\"oth\">2.062<\/td><td class=\"oth\">240.2<\/td><td class=\"oth\">148.0<\/td><td class=\"oth\">86.0<\/td><\/tr>\n<tr><td class=\"num\">16.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Carbendazim\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/10605-21-7\/Carbendazim\" target=\"_blank\" rel=\"noopener\">Carbendazim<\/a><\/td><td class=\"oth\">2.090<\/td><td class=\"oth\">192.0<\/td><td class=\"oth\">160.0<\/td><td class=\"oth\">131.9<\/td><\/tr>\n<tr><td class=\"num\">17.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Oxamyl\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/23135-22-0\/Oxamyl\" target=\"_blank\" rel=\"noopener\">Oxamyl<\/a><\/td><td class=\"oth\">2.154<\/td><td class=\"oth\">237.1<\/td><td class=\"oth\">72.0<\/td><td class=\"oth\">90.0<\/td><\/tr>\n<tr><td class=\"num\">18.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Flonicamid\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/158062-67-0\/Flonicamid\" target=\"_blank\" rel=\"noopener\">Flonicamid<\/a><\/td><td class=\"oth\">2.234<\/td><td class=\"oth\">230.1<\/td><td class=\"oth\">203.0<\/td><td class=\"oth\">148.0<\/td><\/tr>\n<tr><td class=\"num\">19.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Methomyl\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/16752-77-5\/Methomyl\" target=\"_blank\" rel=\"noopener\">Methomyl<\/a><\/td><td class=\"oth\">2.259<\/td><td class=\"oth\">163.1<\/td><td class=\"oth\">88.0<\/td><td class=\"oth\">106.1<\/td><\/tr>\n<tr><td class=\"num\">20.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Thiabendazole\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/148-79-8\/Thiabendazole\" target=\"_blank\" rel=\"noopener\">Thiabendazole<\/a><\/td><td class=\"oth\">2.284<\/td><td class=\"oth\">201.8<\/td><td class=\"oth\">92.2<\/td><td class=\"oth\">104.0<\/td><\/tr>\n<tr><td class=\"num\">21.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Thiamethoxam\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/153719-23-4\/Thiamethoxam\" target=\"_blank\" rel=\"noopener\">Thiamethoxam<\/a><\/td><td class=\"oth\">2.289<\/td><td class=\"oth\">292.1<\/td><td class=\"oth\">211.1<\/td><td class=\"oth\">181.1<\/td><\/tr>\n<tr><td class=\"num\">22.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Fuberidazole\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/3878-19-1\/Fuberidazole\" target=\"_blank\" rel=\"noopener\">Fuberidazole<\/a><\/td><td class=\"oth\">2.368<\/td><td class=\"oth\">185.0<\/td><td class=\"oth\">157.1<\/td><td class=\"oth\">103.0<\/td><\/tr>\n<tr><td class=\"num\">23.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Monocrotophos\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/6923-22-4\/Monocrotophos\" target=\"_blank\" rel=\"noopener\">Monocrotophos<\/a><\/td><td class=\"oth\">2.375<\/td><td class=\"oth\">224.1<\/td><td class=\"oth\">127.0<\/td><td class=\"oth\">193.1<\/td><\/tr>\n<tr><td class=\"num\">24.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Mexacarbate\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/315-18-4\/Mexacarbate\" target=\"_blank\" rel=\"noopener\">Mexacarbate<\/a><\/td><td class=\"oth\">2.393<\/td><td class=\"oth\">223.2<\/td><td class=\"oth\">166.1<\/td><td class=\"oth\">151.1<\/td><\/tr>\n<tr><td class=\"num\">25.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Dioxacarb\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/6988-21-2\/Dioxacarb\" target=\"_blank\" rel=\"noopener\">Dioxacarb<\/a><\/td><td class=\"oth\">2.397<\/td><td class=\"oth\">224.1<\/td><td class=\"oth\">167.0<\/td><td class=\"oth\">77.1<\/td><\/tr>\n<tr><td class=\"num\">26.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Dicrotophos\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/141-66-2\/Dicrotophos\" target=\"_blank\" rel=\"noopener\">Dicrotophos<\/a><\/td><td class=\"oth\">2.526<\/td><td class=\"oth\">238.1<\/td><td class=\"oth\">112.1<\/td><td class=\"oth\">72.0<\/td><\/tr>\n<tr><td class=\"num\">27.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Imidacloprid\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/138261-41-3\/Imidacloprid\" target=\"_blank\" rel=\"noopener\">Imidacloprid<\/a><\/td><td class=\"oth\">2.536<\/td><td class=\"oth\">256.1<\/td><td class=\"oth\">175.0<\/td><td class=\"oth\">209.0<\/td><\/tr>\n<tr><td class=\"num\">28.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Clothianidin\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/210880-92-5\/Clothianidin\" target=\"_blank\" rel=\"noopener\">Clothianidin<\/a><\/td><td class=\"oth\">2.545<\/td><td class=\"oth\">250.0<\/td><td class=\"oth\">169.0<\/td><td class=\"oth\">113.0<\/td><\/tr>\n<tr><td class=\"num\">29.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Dimethoate-d6 (IS)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/1219794-81-6\/Dimethoate-d6\" target=\"_blank\" rel=\"noopener\">Dimethoate-d6 (IS)<\/a><\/td><td class=\"oth\">2.652<\/td><td class=\"oth\">236.2<\/td><td class=\"oth\">205.2<\/td><td class=\"oth\">131.2<\/td><\/tr>\n<tr><td class=\"num\">30.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Fenuron\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/101-42-8\/Fenuron\" target=\"_blank\" rel=\"noopener\">Fenuron<\/a><\/td><td class=\"oth\">2.662<\/td><td class=\"oth\">165.0<\/td><td class=\"oth\">72.1<\/td><td class=\"oth\">46.1<\/td><\/tr>\n<tr><td class=\"num\">31.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Trichlorfon\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/52-68-6\/Trichlorfon\" target=\"_blank\" rel=\"noopener\">Trichlorfon<\/a><\/td><td class=\"oth\">2.667<\/td><td class=\"oth\">257.0<\/td><td class=\"oth\">108.9<\/td><td class=\"oth\">220.8<\/td><\/tr>\n<tr><td class=\"num\">32.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Dimethoate\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/60-51-5\/Dimethoate\" target=\"_blank\" rel=\"noopener\">Dimethoate<\/a><\/td><td class=\"oth\">2.668<\/td><td class=\"oth\">230.0<\/td><td class=\"oth\">199.0<\/td><td class=\"oth\">125.0<\/td><\/tr>\n<tr><td class=\"num\">33.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 3-Hydroxycarbofuran\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/16655-82-6\/3-Hydroxycarbofuran\" target=\"_blank\" rel=\"noopener\">3-Hydroxycarbofuran<\/a><\/td><td class=\"oth\">2.679<\/td><td class=\"oth\">238.1<\/td><td class=\"oth\">181.0<\/td><td class=\"oth\">163.0<\/td><\/tr>\n<tr><td class=\"num\">34.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Vamidothion\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/2275-23-2\/Vamidothion\" target=\"_blank\" rel=\"noopener\">Vamidothion<\/a><\/td><td class=\"oth\">2.695<\/td><td class=\"oth\">288.0<\/td><td class=\"oth\">146.0<\/td><td class=\"oth\">118.0<\/td><\/tr>\n<tr><td class=\"num\">35.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Mevinphos\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/7786-34-7\/Mevinphos\" target=\"_blank\" rel=\"noopener\">Mevinphos<\/a><\/td><td class=\"oth\">2.715<\/td><td class=\"oth\">225.1<\/td><td class=\"oth\">127.0<\/td><td class=\"oth\">127.0<\/td><\/tr>\n<tr><td class=\"num\">36.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Acetamiprid\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/135410-20-7\/Acetamiprid\" target=\"_blank\" rel=\"noopener\">Acetamiprid<\/a><\/td><td class=\"oth\">2.724<\/td><td class=\"oth\">223.0<\/td><td class=\"oth\">126.0<\/td><td class=\"oth\">56.1<\/td><\/tr>\n<tr><td class=\"num\">37.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Propoxur\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/114-26-1\/Propoxur\" target=\"_blank\" rel=\"noopener\">Propoxur<\/a><\/td><td class=\"oth\">2.777<\/td><td class=\"oth\">210.0<\/td><td class=\"oth\">111.0<\/td><td class=\"oth\">168.0<\/td><\/tr>\n<tr><td class=\"num\">38.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Ethirimol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/23947-60-6\/Ethirimol\" target=\"_blank\" rel=\"noopener\">Ethirimol<\/a><\/td><td class=\"oth\">2.780<\/td><td class=\"oth\">210.2<\/td><td class=\"oth\">140.2<\/td><td class=\"oth\">98.2<\/td><\/tr>\n<tr><td class=\"num\">39.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Cymoxanil\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/57966-95-7\/Cymoxanil\" target=\"_blank\" rel=\"noopener\">Cymoxanil<\/a><\/td><td class=\"oth\">2.805<\/td><td class=\"oth\">199.1<\/td><td class=\"oth\">128.2<\/td><td class=\"oth\">111.2<\/td><\/tr>\n<tr><td class=\"num\">40.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Thiacloprid\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/111988-49-9\/Thiacloprid\" target=\"_blank\" rel=\"noopener\">Thiacloprid<\/a><\/td><td class=\"oth\">2.865<\/td><td class=\"oth\">253.0<\/td><td class=\"oth\">126.0<\/td><td class=\"oth\">99.0<\/td><\/tr>\n<tr><td class=\"num\">41.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Pirimicarb\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/23103-98-2\/Pirimicarb\" target=\"_blank\" rel=\"noopener\">Pirimicarb<\/a><\/td><td class=\"oth\">2.896<\/td><td class=\"oth\">239.2<\/td><td class=\"oth\">72.0<\/td><td class=\"oth\">182.1<\/td><\/tr>\n<tr><td class=\"num\">42.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Aldicarb\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/116-06-3\/Aldicarb\" target=\"_blank\" rel=\"noopener\">Aldicarb<\/a><\/td><td class=\"oth\">3.022<\/td><td class=\"oth\">208.2<\/td><td class=\"oth\">116.2<\/td><td class=\"oth\">89.2<\/td><\/tr>\n<tr><td class=\"num\">43.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Tricyclazole\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/41814-78-2\/Tricyclazole\" target=\"_blank\" rel=\"noopener\">Tricyclazole<\/a><\/td><td class=\"oth\">3.066<\/td><td class=\"oth\">190.0<\/td><td class=\"oth\">163.0<\/td><td class=\"oth\">136.0<\/td><\/tr>\n<tr><td class=\"num\">44.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Oxadixyl\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/77732-09-3\/Oxadixyl\" target=\"_blank\" rel=\"noopener\">Oxadixyl<\/a><\/td><td class=\"oth\">3.076<\/td><td class=\"oth\">279.1<\/td><td class=\"oth\">219.1<\/td><td class=\"oth\">132.1<\/td><\/tr>\n<tr><td class=\"num\">45.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Carbetamide\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/16118-49-3\/Carbetamide\" target=\"_blank\" rel=\"noopener\">Carbetamide<\/a><\/td><td class=\"oth\">3.085<\/td><td class=\"oth\">237.1<\/td><td class=\"oth\">192.0<\/td><td class=\"oth\">118.1<\/td><\/tr>\n<tr><td class=\"num\">46.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Simetryn\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/1014-70-6\/Simetryn\" target=\"_blank\" rel=\"noopener\">Simetryn<\/a><\/td><td class=\"oth\">3.122<\/td><td class=\"oth\">214.1<\/td><td class=\"oth\">68.0<\/td><td class=\"oth\">96.0<\/td><\/tr>\n<tr><td class=\"num\">47.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Thiophanate-methyl\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/23564-05-8\/Thiophanate-methyl\" target=\"_blank\" rel=\"noopener\">Thiophanate-methyl<\/a><\/td><td class=\"oth\">3.146<\/td><td class=\"oth\">343.0<\/td><td class=\"oth\">151.0<\/td><td class=\"oth\">311.1<\/td><\/tr>\n<tr><td class=\"num\">48.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Thidiazuron\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/51707-55-2\/Thidiazuron\" target=\"_blank\" rel=\"noopener\">Thidiazuron<\/a><\/td><td class=\"oth\">3.150<\/td><td class=\"oth\">221.1<\/td><td class=\"oth\">102.1<\/td><td class=\"oth\">127.9<\/td><\/tr>\n<tr><td class=\"num\">49.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Bendiocarb\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/22781-23-3\/Bendiocarb\" target=\"_blank\" rel=\"noopener\">Bendiocarb<\/a><\/td><td class=\"oth\">3.212<\/td><td class=\"oth\">224.1<\/td><td class=\"oth\">167.1<\/td><td class=\"oth\">109.0<\/td><\/tr>\n<tr><td class=\"num\">50.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Imazalil\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/35554-44-0\/Imazalil\" target=\"_blank\" rel=\"noopener\">Imazalil<\/a><\/td><td class=\"oth\">3.225<\/td><td class=\"oth\">297.1<\/td><td class=\"oth\">159.0<\/td><td class=\"oth\">41.1<\/td><\/tr>\n<tr><td class=\"num\">51.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Prometon\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/1610-18-0\/Prometon\" target=\"_blank\" rel=\"noopener\">Prometon<\/a><\/td><td class=\"oth\">3.227<\/td><td class=\"oth\">226.1<\/td><td class=\"oth\">142.1<\/td><td class=\"oth\">184.0<\/td><\/tr>\n<tr><td class=\"num\">52.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Metribuzin\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/21087-64-9\/Metribuzin\" target=\"_blank\" rel=\"noopener\">Metribuzin<\/a><\/td><td class=\"oth\">3.230<\/td><td class=\"oth\">215.1<\/td><td class=\"oth\">187.1<\/td><td class=\"oth\">131.0<\/td><\/tr>\n<tr><td class=\"num\">53.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Secbumeton\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/26259-45-0\/Secbumeton\" target=\"_blank\" rel=\"noopener\">Secbumeton<\/a><\/td><td class=\"oth\">3.232<\/td><td class=\"oth\">226.2<\/td><td class=\"oth\">170.2<\/td><td class=\"oth\">100.2<\/td><\/tr>\n<tr><td class=\"num\">54.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Terbumeton\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/33693-04-8\/Terbumeton\" target=\"_blank\" rel=\"noopener\">Terbumeton<\/a><\/td><td class=\"oth\">3.235<\/td><td class=\"oth\">226.1<\/td><td class=\"oth\">114.1<\/td><td class=\"oth\">170.1<\/td><\/tr>\n<tr><td class=\"num\">55.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Sulfentrazone*\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/122836-35-5\/Sulfentrazone\" target=\"_blank\" rel=\"noopener\">Sulfentrazone*<\/a><\/td><td class=\"oth\">3.263<\/td><td class=\"oth\">387.0<\/td><td class=\"oth\">309.0<\/td><td class=\"oth\">201.0<\/td><\/tr>\n<tr><td class=\"num\">56.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Pyracarbolid\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/24691-76-7\/Pyracarbolid\" target=\"_blank\" rel=\"noopener\">Pyracarbolid<\/a><\/td><td class=\"oth\">3.279<\/td><td class=\"oth\">218.1<\/td><td class=\"oth\">125.1<\/td><td class=\"oth\">97.0<\/td><\/tr>\n<tr><td class=\"num\">57.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Tebuthiuron\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/34014-18-1\/Tebuthiuron\" target=\"_blank\" rel=\"noopener\">Tebuthiuron<\/a><\/td><td class=\"oth\">3.310<\/td><td class=\"oth\">229.1<\/td><td class=\"oth\">172.1<\/td><td class=\"oth\">116.0<\/td><\/tr>\n<tr><td class=\"num\">58.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Carbaryl\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/63-25-2\/Carbaryl\" target=\"_blank\" rel=\"noopener\">Carbaryl<\/a><\/td><td class=\"oth\">3.316<\/td><td class=\"oth\">202.1<\/td><td class=\"oth\">145.0<\/td><td class=\"oth\">127.0<\/td><\/tr>\n<tr><td class=\"num\">59.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Carboxin\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/5234-68-4\/Carboxin\" target=\"_blank\" rel=\"noopener\">Carboxin<\/a><\/td><td class=\"oth\">3.326<\/td><td class=\"oth\">236.1<\/td><td class=\"oth\">143.0<\/td><td class=\"oth\">87.1<\/td><\/tr>\n<tr><td class=\"num\">60.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Monolinuron\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/1746-81-2\/Monolinuron\" target=\"_blank\" rel=\"noopener\">Monolinuron<\/a><\/td><td class=\"oth\">3.373<\/td><td class=\"oth\">215.1<\/td><td class=\"oth\">126.0<\/td><td class=\"oth\">148.0<\/td><\/tr>\n<tr><td class=\"num\">61.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Fluometuron\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/2164-17-2\/Fluometuron\" target=\"_blank\" rel=\"noopener\">Fluometuron<\/a><\/td><td class=\"oth\">3.381<\/td><td class=\"oth\">233.1<\/td><td class=\"oth\">72.1<\/td><td class=\"oth\">46.1<\/td><\/tr>\n<tr><td class=\"num\">62.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Diuron\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/330-54-1\/Diuron\" target=\"_blank\" rel=\"noopener\">Diuron<\/a><\/td><td class=\"oth\">3.382<\/td><td class=\"oth\">233.1<\/td><td class=\"oth\">72.1<\/td><td class=\"oth\">46.0<\/td><\/tr>\n<tr><td class=\"num\">63.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Ethiofencarb\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/29973-13-5\/Ethiofencarb\" target=\"_blank\" rel=\"noopener\">Ethiofencarb<\/a><\/td><td class=\"oth\">3.406<\/td><td class=\"oth\">226.0<\/td><td class=\"oth\">107.0<\/td><td class=\"oth\">164.0<\/td><\/tr>\n<tr><td class=\"num\">64.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Ametryn\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/834-12-8\/Ametryn\" target=\"_blank\" rel=\"noopener\">Ametryn<\/a><\/td><td class=\"oth\">3.447<\/td><td class=\"oth\">228.1<\/td><td class=\"oth\">68.1<\/td><td class=\"oth\">186.1<\/td><\/tr>\n<tr><td class=\"num\">65.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Chlorotoluron\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/15545-48-9\/Chlorotoluron\" target=\"_blank\" rel=\"noopener\">Chlorotoluron<\/a><\/td><td class=\"oth\">3.450<\/td><td class=\"oth\">213.0<\/td><td class=\"oth\">72.1<\/td><td class=\"oth\">46.1<\/td><\/tr>\n<tr><td class=\"num\">66.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Metobromuron\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/3060-89-7\/Metobromuron\" target=\"_blank\" rel=\"noopener\">Metobromuron<\/a><\/td><td class=\"oth\">3.483<\/td><td class=\"oth\">259.0<\/td><td class=\"oth\">170.0<\/td><td class=\"oth\">148.1<\/td><\/tr>\n<tr><td class=\"num\">67.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Propham\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/122-42-9\/Propham\" target=\"_blank\" rel=\"noopener\">Propham<\/a><\/td><td class=\"oth\">3.500<\/td><td class=\"oth\">180.1<\/td><td class=\"oth\">120.2<\/td><td class=\"oth\">77.1<\/td><\/tr>\n<tr><td class=\"num\">68.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Flutriafol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/76674-21-0\/Flutriafol\" target=\"_blank\" rel=\"noopener\">Flutriafol<\/a><\/td><td class=\"oth\">3.508<\/td><td class=\"oth\">302.0<\/td><td class=\"oth\">70.1<\/td><td class=\"oth\">123.0<\/td><\/tr>\n<tr><td class=\"num\">69.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Methoprotryne\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/841-06-5\/Methoprotryne\" target=\"_blank\" rel=\"noopener\">Methoprotryne<\/a><\/td><td class=\"oth\">3.532<\/td><td class=\"oth\">272.1<\/td><td class=\"oth\">170.0<\/td><td class=\"oth\">198.1<\/td><\/tr>\n<tr><td class=\"num\">70.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Forchlorfenuron\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/68157-60-8\/Forchlorfenuron\" target=\"_blank\" rel=\"noopener\">Forchlorfenuron<\/a><\/td><td class=\"oth\">3.536<\/td><td class=\"oth\">248.1<\/td><td class=\"oth\">129.0<\/td><td class=\"oth\">93.0<\/td><\/tr>\n<tr><td class=\"num\">71.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Isoprocarb\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/2631-40-5\/Isoprocarb\" target=\"_blank\" rel=\"noopener\">Isoprocarb<\/a><\/td><td class=\"oth\">3.537<\/td><td class=\"oth\">194.1<\/td><td class=\"oth\">95.1<\/td><td class=\"oth\">137.1<\/td><\/tr>\n<tr><td class=\"num\">72.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Fenpropimorph\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/67564-91-4\/Fenpropimorph\" target=\"_blank\" rel=\"noopener\">Fenpropimorph<\/a><\/td><td class=\"oth\">3.561<\/td><td class=\"oth\">304.2<\/td><td class=\"oth\">147.1<\/td><td class=\"oth\">117.0<\/td><\/tr>\n<tr><td class=\"num\">73.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Atrazine-d5 (SS)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/163165-75-1\/Atrazine-d5\" target=\"_blank\" rel=\"noopener\">Atrazine-d5 (SS)<\/a><\/td><td class=\"oth\">3.575<\/td><td class=\"oth\">221.2<\/td><td class=\"oth\">179.2<\/td><td class=\"oth\">101.3<\/td><\/tr>\n<tr><td class=\"num\">74.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Isocarbophos\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/24353-61-5\/Isocarbophos\" target=\"_blank\" rel=\"noopener\">Isocarbophos<\/a><\/td><td class=\"oth\">3.606<\/td><td class=\"oth\">291.1<\/td><td class=\"oth\">231.1<\/td><td class=\"oth\">121.1<\/td><\/tr>\n<tr><td class=\"num\">75.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Isoproturon\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/34123-59-6\/Isoproturon\" target=\"_blank\" rel=\"noopener\">Isoproturon<\/a><\/td><td class=\"oth\">3.612<\/td><td class=\"oth\">207.1<\/td><td class=\"oth\">72.1<\/td><td class=\"oth\">46.1<\/td><\/tr>\n<tr><td class=\"num\">76.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Desmedipham\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/13684-56-5\/Desmedipham\" target=\"_blank\" rel=\"noopener\">Desmedipham<\/a><\/td><td class=\"oth\">3.620<\/td><td class=\"oth\">318.0<\/td><td class=\"oth\">154.1<\/td><td class=\"oth\">301.0<\/td><\/tr>\n<tr><td class=\"num\">77.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Phenmedipham\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/13684-63-4\/Phenmedipham\" target=\"_blank\" rel=\"noopener\">Phenmedipham<\/a><\/td><td class=\"oth\">3.623<\/td><td class=\"oth\">318.1<\/td><td class=\"oth\">108.1<\/td><td class=\"oth\">93.1<\/td><\/tr>\n<tr><td class=\"num\">78.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Spiroxamine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/118134-30-8\/Spiroxamine\" target=\"_blank\" rel=\"noopener\">Spiroxamine<\/a><\/td><td class=\"oth\">3.626<\/td><td class=\"oth\">298.2<\/td><td class=\"oth\">144.2<\/td><td class=\"oth\">58.2<\/td><\/tr>\n<tr><td class=\"num\">79.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Pyrimethanil\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/53112-28-0\/Pyrimethanil\" target=\"_blank\" rel=\"noopener\">Pyrimethanil<\/a><\/td><td class=\"oth\">3.659<\/td><td class=\"oth\">200.0<\/td><td class=\"oth\">107.0<\/td><td class=\"oth\">82.0<\/td><\/tr>\n<tr><td class=\"num\">80.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Metalaxyl\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/57837-19-1\/Metalaxyl\" target=\"_blank\" rel=\"noopener\">Metalaxyl<\/a><\/td><td class=\"oth\">3.691<\/td><td class=\"oth\">280.1<\/td><td class=\"oth\">220.1<\/td><td class=\"oth\">192.1<\/td><\/tr>\n<tr><td class=\"num\">81.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Chlorantraniliprole\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/500008-45-7\/Chlorantraniliprole\" target=\"_blank\" rel=\"noopener\">Chlorantraniliprole<\/a><\/td><td class=\"oth\">3.730<\/td><td class=\"oth\">484.0<\/td><td class=\"oth\">452.8<\/td><td class=\"oth\">285.8<\/td><\/tr>\n<tr><td class=\"num\">82.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Cycluron\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/2163-69-1\/Cycluron\" target=\"_blank\" rel=\"noopener\">Cycluron<\/a><\/td><td class=\"oth\">3.785<\/td><td class=\"oth\">199.1<\/td><td class=\"oth\">89.1<\/td><td class=\"oth\">46.0<\/td><\/tr>\n<tr><td class=\"num\">83.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Prometryn\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/7287-19-6\/Prometryn\" target=\"_blank\" rel=\"noopener\">Prometryn<\/a><\/td><td class=\"oth\">3.830<\/td><td class=\"oth\">242.1<\/td><td class=\"oth\">158.0<\/td><td class=\"oth\">200.2<\/td><\/tr>\n<tr><td class=\"num\">84.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Linuron-d6 (SS)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/1219804-76-8\/Linuron-d6\" target=\"_blank\" rel=\"noopener\">Linuron-d6 (SS)<\/a><\/td><td class=\"oth\">3.843<\/td><td class=\"oth\">255.1<\/td><td class=\"oth\">160.1<\/td><td class=\"oth\">185.2<\/td><\/tr>\n<tr><td class=\"num\">85.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Linuron\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/330-55-2\/Linuron\" target=\"_blank\" rel=\"noopener\">Linuron<\/a><\/td><td class=\"oth\">3.854<\/td><td class=\"oth\">249.0<\/td><td class=\"oth\">160.0<\/td><td class=\"oth\">182.0<\/td><\/tr>\n<tr><td class=\"num\">86.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Thiofanox\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/39196-18-4\/Thiofanox\" target=\"_blank\" rel=\"noopener\">Thiofanox<\/a><\/td><td class=\"oth\">3.876<\/td><td class=\"oth\">242.1<\/td><td class=\"oth\">57.2<\/td><td class=\"oth\">184.0<\/td><\/tr>\n<tr><td class=\"num\">87.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Fenobucarb\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/3766-81-2\/Fenobucarb\" target=\"_blank\" rel=\"noopener\">Fenobucarb<\/a><\/td><td class=\"oth\">3.894<\/td><td class=\"oth\">208.0<\/td><td class=\"oth\">95.1<\/td><td class=\"oth\">152.1<\/td><\/tr>\n<tr><td class=\"num\">88.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Azoxystrobin\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/131860-33-8\/Azoxystrobin\" target=\"_blank\" rel=\"noopener\">Azoxystrobin<\/a><\/td><td class=\"oth\">3.915<\/td><td class=\"oth\">404.2<\/td><td class=\"oth\">329.0<\/td><td class=\"oth\">344.0<\/td><\/tr>\n<tr><td class=\"num\">89.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Diethofencarb\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/87130-20-9\/Diethofencarb\" target=\"_blank\" rel=\"noopener\">Diethofencarb<\/a><\/td><td class=\"oth\">3.924<\/td><td class=\"oth\">268.1<\/td><td class=\"oth\">226.1<\/td><td class=\"oth\">124.0<\/td><\/tr>\n<tr><td class=\"num\">90.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Ethofumesate\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/26225-79-6\/Ethofumesate\" target=\"_blank\" rel=\"noopener\">Ethofumesate<\/a><\/td><td class=\"oth\">3.925<\/td><td class=\"oth\">304.1<\/td><td class=\"oth\">121.1<\/td><td class=\"oth\">241.0<\/td><\/tr>\n<tr><td class=\"num\">91.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Ethiprole\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/181587-01-9\/Ethiprole\" target=\"_blank\" rel=\"noopener\">Ethiprole<\/a><\/td><td class=\"oth\">3.932<\/td><td class=\"oth\">397.0<\/td><td class=\"oth\">350.9<\/td><td class=\"oth\">255.0<\/td><\/tr>\n<tr><td class=\"num\">92.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Fludioxonil*\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/131341-86-1\/Fludioxonil\" target=\"_blank\" rel=\"noopener\">Fludioxonil*<\/a><\/td><td class=\"oth\">3.935<\/td><td class=\"oth\">247.1<\/td><td class=\"oth\">180.1<\/td><td class=\"oth\">126.1<\/td><\/tr>\n<tr><td class=\"num\">93.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Methiocarb\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/2032-65-7\/Methiocarb\" target=\"_blank\" rel=\"noopener\">Methiocarb<\/a><\/td><td class=\"oth\">3.958<\/td><td class=\"oth\">226.1<\/td><td class=\"oth\">169.1<\/td><td class=\"oth\">121.1<\/td><\/tr>\n<tr><td class=\"num\">94.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Siduron\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/1982-49-6\/Siduron\" target=\"_blank\" rel=\"noopener\">Siduron<\/a><\/td><td class=\"oth\">3.967<\/td><td class=\"oth\">233.3<\/td><td class=\"oth\">94.0<\/td><td class=\"oth\">137.1<\/td><\/tr>\n<tr><td class=\"num\">95.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Halofenozide\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/112226-61-6\/Halofenozide\" target=\"_blank\" rel=\"noopener\">Halofenozide<\/a><\/td><td class=\"oth\">3.974<\/td><td class=\"oth\">331.1<\/td><td class=\"oth\">104.9<\/td><td class=\"oth\">275.0<\/td><\/tr>\n<tr><td class=\"num\">96.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Fenamidone\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/161326-34-7\/Fenamidone\" target=\"_blank\" rel=\"noopener\">Fenamidone<\/a><\/td><td class=\"oth\">3.993<\/td><td class=\"oth\">312.0<\/td><td class=\"oth\">92.1<\/td><td class=\"oth\">236.1<\/td><\/tr>\n<tr><td class=\"num\">97.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Furalaxyl\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/57646-30-7\/Furalaxyl\" target=\"_blank\" rel=\"noopener\">Furalaxyl<\/a><\/td><td class=\"oth\">3.995<\/td><td class=\"oth\">302.1<\/td><td class=\"oth\">242.1<\/td><td class=\"oth\">95.0<\/td><\/tr>\n<tr><td class=\"num\">98.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Acibenzolar-S-methyl\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/135158-54-2\/Acibenzolar-S-methyl\" target=\"_blank\" rel=\"noopener\">Acibenzolar-S-methyl<\/a><\/td><td class=\"oth\">4.023<\/td><td class=\"oth\">210.9<\/td><td class=\"oth\">136.1<\/td><td class=\"oth\">91.1<\/td><\/tr>\n<tr><td class=\"num\">99.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Nuarimol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/63284-71-9\/Nuarimol\" target=\"_blank\" rel=\"noopener\">Nuarimol<\/a><\/td><td class=\"oth\">4.030<\/td><td class=\"oth\">315.1<\/td><td class=\"oth\">252.0<\/td><td class=\"oth\">207.0<\/td><\/tr>\n<tr><td class=\"num\">100.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Dimethomorph\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/110488-70-5\/Dimethomorph\" target=\"_blank\" rel=\"noopener\">Dimethomorph<\/a><\/td><td class=\"oth\">4.044<\/td><td class=\"oth\">388.2<\/td><td class=\"oth\">300.9<\/td><td class=\"oth\">165.1<\/td><\/tr>\n<tr><td class=\"num\">101.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Boscalid\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/188425-85-6\/Boscalid\" target=\"_blank\" rel=\"noopener\">Boscalid<\/a><\/td><td class=\"oth\">4.050<\/td><td class=\"oth\">343.0<\/td><td class=\"oth\">306.8<\/td><td class=\"oth\">140.0<\/td><\/tr>\n<tr><td class=\"num\">102.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Mandipropamid\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/374726-62-2\/Mandipropamid\" target=\"_blank\" rel=\"noopener\">Mandipropamid<\/a><\/td><td class=\"oth\">4.067<\/td><td class=\"oth\">412.1<\/td><td class=\"oth\">328.1<\/td><td class=\"oth\">125.0<\/td><\/tr>\n<tr><td class=\"num\">103.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Flutolanil\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/66332-96-5\/Flutolanil\" target=\"_blank\" rel=\"noopener\">Flutolanil<\/a><\/td><td class=\"oth\">4.079<\/td><td class=\"oth\">324.1<\/td><td class=\"oth\">262.0<\/td><td class=\"oth\">242.0<\/td><\/tr>\n<\/tbody><\/table>\n<table class=\"peaks col-lg-6 col-12 peak-50\"><thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 75px\">t<sub>R<\/sub> (min)<\/th><th style=\"text-align: center;width: 75px\">Precursor Ion<\/th><th style=\"text-align: center;width: 75px\">Product Ion 1<\/th><th style=\"text-align: center;width: 75px\">Product Ion 2<\/th><\/tr><\/thead><tbody>\n<tr><td class=\"num\">104.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Promecarb\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/2631-37-0\/Promecarb\" target=\"_blank\" rel=\"noopener\">Promecarb<\/a><\/td><td class=\"oth\">4.104<\/td><td class=\"oth\">208.1<\/td><td class=\"oth\">109.0<\/td><td class=\"oth\">151.0<\/td><\/tr>\n<tr><td class=\"num\">105.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Paclobutrazol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/76738-62-0\/Paclobutrazol\" target=\"_blank\" rel=\"noopener\">Paclobutrazol<\/a><\/td><td class=\"oth\">4.119<\/td><td class=\"oth\">294.0<\/td><td class=\"oth\">70.1<\/td><td class=\"oth\">125.0<\/td><\/tr>\n<tr><td class=\"num\">106.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Triadimenol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/55219-65-3\/Triadimenol\" target=\"_blank\" rel=\"noopener\">Triadimenol<\/a><\/td><td class=\"oth\">4.119<\/td><td class=\"oth\">296.1<\/td><td class=\"oth\">70.1<\/td><td class=\"oth\">43.1<\/td><\/tr>\n<tr><td class=\"num\">107.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Mepronil\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/55814-41-0\/Mepronil\" target=\"_blank\" rel=\"noopener\">Mepronil<\/a><\/td><td class=\"oth\">4.161<\/td><td class=\"oth\">270.1<\/td><td class=\"oth\">91.1<\/td><td class=\"oth\">119.0<\/td><\/tr>\n<tr><td class=\"num\">108.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Cyproconazole\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/94361-06-5\/Cyproconazole\" target=\"_blank\" rel=\"noopener\">Cyproconazole<\/a><\/td><td class=\"oth\">4.175<\/td><td class=\"oth\">292.2<\/td><td class=\"oth\">70.1<\/td><td class=\"oth\">125.0<\/td><\/tr>\n<tr><td class=\"num\">109.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Myclobutanil\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/88671-89-0\/Myclobutanil\" target=\"_blank\" rel=\"noopener\">Myclobutanil<\/a><\/td><td class=\"oth\">4.195<\/td><td class=\"oth\">289.2<\/td><td class=\"oth\">70.1<\/td><td class=\"oth\">125.1<\/td><\/tr>\n<tr><td class=\"num\">110.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Chloroxuron\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/1982-47-4\/Chloroxuron\" target=\"_blank\" rel=\"noopener\">Chloroxuron<\/a><\/td><td class=\"oth\">4.225<\/td><td class=\"oth\">291.1<\/td><td class=\"oth\">72.1<\/td><td class=\"oth\">46.1<\/td><\/tr>\n<tr><td class=\"num\">111.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Bupirimate\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/41483-43-6\/Bupirimate\" target=\"_blank\" rel=\"noopener\">Bupirimate<\/a><\/td><td class=\"oth\">4.226<\/td><td class=\"oth\">317.2<\/td><td class=\"oth\">166.0<\/td><td class=\"oth\">237.1<\/td><\/tr>\n<tr><td class=\"num\">112.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Methoxyfenozide\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/161050-58-4\/Methoxyfenozide\" target=\"_blank\" rel=\"noopener\">Methoxyfenozide<\/a><\/td><td class=\"oth\">4.230<\/td><td class=\"oth\">369.2<\/td><td class=\"oth\">149.0<\/td><td class=\"oth\">313.1<\/td><\/tr>\n<tr><td class=\"num\">113.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Triadimefon\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/43121-43-3\/Triadimefon\" target=\"_blank\" rel=\"noopener\">Triadimefon<\/a><\/td><td class=\"oth\">4.287<\/td><td class=\"oth\">294.0<\/td><td class=\"oth\">69.0<\/td><td class=\"oth\">197.0<\/td><\/tr>\n<tr><td class=\"num\">114.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Bifenazate\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/149877-41-8\/Bifenazate\" target=\"_blank\" rel=\"noopener\">Bifenazate<\/a><\/td><td class=\"oth\">4.290<\/td><td class=\"oth\">301.1<\/td><td class=\"oth\">198.0<\/td><td class=\"oth\">170.1<\/td><\/tr>\n<tr><td class=\"num\">115.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Cyprodinil\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/121552-61-2\/Cyprodinil\" target=\"_blank\" rel=\"noopener\">Cyprodinil<\/a><\/td><td class=\"oth\">4.313<\/td><td class=\"oth\">226.1<\/td><td class=\"oth\">93.0<\/td><td class=\"oth\">77.0<\/td><\/tr>\n<tr><td class=\"num\">116.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Mefenacet\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/73250-68-7\/Mefenacet\" target=\"_blank\" rel=\"noopener\">Mefenacet<\/a><\/td><td class=\"oth\">4.325<\/td><td class=\"oth\">299.1<\/td><td class=\"oth\">148.1<\/td><td class=\"oth\">120.1<\/td><\/tr>\n<tr><td class=\"num\">117.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Mepanipyrim\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/110235-47-7\/Mepanipyrim\" target=\"_blank\" rel=\"noopener\">Mepanipyrim<\/a><\/td><td class=\"oth\">4.334<\/td><td class=\"oth\">224.1<\/td><td class=\"oth\">77.1<\/td><td class=\"oth\">106.1<\/td><\/tr>\n<tr><td class=\"num\">118.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Fluquinconazole\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/136426-54-5\/Fluquinconazole\" target=\"_blank\" rel=\"noopener\">Fluquinconazole<\/a><\/td><td class=\"oth\">4.346<\/td><td class=\"oth\">376.0<\/td><td class=\"oth\">348.9<\/td><td class=\"oth\">307.0<\/td><\/tr>\n<tr><td class=\"num\">119.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Fenhexamid\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/126833-17-8\/Fenhexamid\" target=\"_blank\" rel=\"noopener\">Fenhexamid<\/a><\/td><td class=\"oth\">4.383<\/td><td class=\"oth\">302.1<\/td><td class=\"oth\">97.1<\/td><td class=\"oth\">55.1<\/td><\/tr>\n<tr><td class=\"num\">120.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Fluoxastrobin\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/361377-29-9\/Fluoxastrobin\" target=\"_blank\" rel=\"noopener\">Fluoxastrobin<\/a><\/td><td class=\"oth\">4.385<\/td><td class=\"oth\">459.2<\/td><td class=\"oth\">427.0<\/td><td class=\"oth\">188.1<\/td><\/tr>\n<tr><td class=\"num\">121.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Tetraconazole\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/112281-77-3\/Tetraconazole\" target=\"_blank\" rel=\"noopener\">Tetraconazole<\/a><\/td><td class=\"oth\">4.387<\/td><td class=\"oth\">372.0<\/td><td class=\"oth\">159.0<\/td><td class=\"oth\">70.0<\/td><\/tr>\n<tr><td class=\"num\">122.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Bromuconazole isomer 1\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/116255-48-2\/Bromuconazole isomer 1\" target=\"_blank\" rel=\"noopener\">Bromuconazole isomer 1<\/a><\/td><td class=\"oth\">4.392<\/td><td class=\"oth\">376.0<\/td><td class=\"oth\">159.0<\/td><td class=\"oth\">159.0<\/td><\/tr>\n<tr><td class=\"num\">123.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Butafenacil\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/134605-64-4\/Butafenacil\" target=\"_blank\" rel=\"noopener\">Butafenacil<\/a><\/td><td class=\"oth\">4.411<\/td><td class=\"oth\">492.1<\/td><td class=\"oth\">330.9<\/td><td class=\"oth\">179.9<\/td><\/tr>\n<tr><td class=\"num\">124.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Iprovalicarb\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/140923-17-7\/Iprovalicarb\" target=\"_blank\" rel=\"noopener\">Iprovalicarb<\/a><\/td><td class=\"oth\">4.412<\/td><td class=\"oth\">321.2<\/td><td class=\"oth\">119.0<\/td><td class=\"oth\">91.1<\/td><\/tr>\n<tr><td class=\"num\">125.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Flufenacet\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/142459-58-3\/Flufenacet\" target=\"_blank\" rel=\"noopener\">Flufenacet<\/a><\/td><td class=\"oth\">4.427<\/td><td class=\"oth\">364.1<\/td><td class=\"oth\">152.0<\/td><td class=\"oth\">194.1<\/td><\/tr>\n<tr><td class=\"num\">126.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Triticonazole\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/131983-72-7\/Triticonazole\" target=\"_blank\" rel=\"noopener\">Triticonazole<\/a><\/td><td class=\"oth\">4.429<\/td><td class=\"oth\">318.0<\/td><td class=\"oth\">70.0<\/td><td class=\"oth\">125.0<\/td><\/tr>\n<tr><td class=\"num\">127.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Cyazofamid\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/120116-88-3\/Cyazofamid\" target=\"_blank\" rel=\"noopener\">Cyazofamid<\/a><\/td><td class=\"oth\">4.441<\/td><td class=\"oth\">325.1<\/td><td class=\"oth\">108.0<\/td><td class=\"oth\">44.0<\/td><\/tr>\n<tr><td class=\"num\">128.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Diflubenzuron\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/35367-38-5\/Diflubenzuron\" target=\"_blank\" rel=\"noopener\">Diflubenzuron<\/a><\/td><td class=\"oth\">4.464<\/td><td class=\"oth\">311.0<\/td><td class=\"oth\">158.0<\/td><td class=\"oth\">141.0<\/td><\/tr>\n<tr><td class=\"num\">129.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Spirotetramat\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/203313-25-1\/Spirotetramat\" target=\"_blank\" rel=\"noopener\">Spirotetramat<\/a><\/td><td class=\"oth\">4.478<\/td><td class=\"oth\">374.1<\/td><td class=\"oth\">330.1<\/td><td class=\"oth\">302.0<\/td><\/tr>\n<tr><td class=\"num\">130.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Fenarimol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/60168-88-9\/Fenarimol\" target=\"_blank\" rel=\"noopener\">Fenarimol<\/a><\/td><td class=\"oth\">4.510<\/td><td class=\"oth\">331.0<\/td><td class=\"oth\">267.9<\/td><td class=\"oth\">189.1<\/td><\/tr>\n<tr><td class=\"num\">131.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Fenbuconazole\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/114369-43-6\/Fenbuconazole\" target=\"_blank\" rel=\"noopener\">Fenbuconazole<\/a><\/td><td class=\"oth\">4.516<\/td><td class=\"oth\">337.0<\/td><td class=\"oth\">125.0<\/td><td class=\"oth\">70.1<\/td><\/tr>\n<tr><td class=\"num\">132.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Epoxiconazole\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/133855-98-8\/Epoxiconazole\" target=\"_blank\" rel=\"noopener\">Epoxiconazole<\/a><\/td><td class=\"oth\">4.529<\/td><td class=\"oth\">330.0<\/td><td class=\"oth\">121.1<\/td><td class=\"oth\">101.0<\/td><\/tr>\n<tr><td class=\"num\">133.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Etaconazole\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/60207-93-4\/Etaconazole\" target=\"_blank\" rel=\"noopener\">Etaconazole<\/a><\/td><td class=\"oth\">4.552<\/td><td class=\"oth\">328.1<\/td><td class=\"oth\">159.0<\/td><td class=\"oth\">55.0<\/td><\/tr>\n<tr><td class=\"num\">134.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Fipronil*\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/120068-37-3\/Fipronil\" target=\"_blank\" rel=\"noopener\">Fipronil*<\/a><\/td><td class=\"oth\">4.571<\/td><td class=\"oth\">436.9<\/td><td class=\"oth\">332.0<\/td><td class=\"oth\">252.0<\/td><\/tr>\n<tr><td class=\"num\">135.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Neburon\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/555-37-3\/Neburon\" target=\"_blank\" rel=\"noopener\">Neburon<\/a><\/td><td class=\"oth\">4.635<\/td><td class=\"oth\">275.0<\/td><td class=\"oth\">57.1<\/td><td class=\"oth\">88.1<\/td><\/tr>\n<tr><td class=\"num\">136.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Flusilazole\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/85509-19-9\/Flusilazole\" target=\"_blank\" rel=\"noopener\">Flusilazole<\/a><\/td><td class=\"oth\">4.646<\/td><td class=\"oth\">316.1<\/td><td class=\"oth\">247.1<\/td><td class=\"oth\">165.1<\/td><\/tr>\n<tr><td class=\"num\">137.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Fenoxycarb\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/72490-01-8\/Fenoxycarb\" target=\"_blank\" rel=\"noopener\">Fenoxycarb<\/a><\/td><td class=\"oth\">4.653<\/td><td class=\"oth\">302.2<\/td><td class=\"oth\">88.1<\/td><td class=\"oth\">116.1<\/td><\/tr>\n<tr><td class=\"num\">138.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Picoxystrobin\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/117428-22-5\/Picoxystrobin\" target=\"_blank\" rel=\"noopener\">Picoxystrobin<\/a><\/td><td class=\"oth\">4.674<\/td><td class=\"oth\">368.0<\/td><td class=\"oth\">145.0<\/td><td class=\"oth\">205.0<\/td><\/tr>\n<tr><td class=\"num\">139.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Rotenone\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/83-79-4\/Rotenone\" target=\"_blank\" rel=\"noopener\">Rotenone<\/a><\/td><td class=\"oth\">4.683<\/td><td class=\"oth\">395.1<\/td><td class=\"oth\">213.1<\/td><td class=\"oth\">192.1<\/td><\/tr>\n<tr><td class=\"num\">140.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Tebufenozide\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/112410-23-8\/Tebufenozide\" target=\"_blank\" rel=\"noopener\">Tebufenozide<\/a><\/td><td class=\"oth\">4.719<\/td><td class=\"oth\">353.2<\/td><td class=\"oth\">133.1<\/td><td class=\"oth\">105.0<\/td><\/tr>\n<tr><td class=\"num\">141.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Dimoxystrobin\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/149961-52-4\/Dimoxystrobin\" target=\"_blank\" rel=\"noopener\">Dimoxystrobin<\/a><\/td><td class=\"oth\">4.738<\/td><td class=\"oth\">327.1<\/td><td class=\"oth\">205.1<\/td><td class=\"oth\">116.1<\/td><\/tr>\n<tr><td class=\"num\">142.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Bromuconazole isomer 2\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/116255-48-2\/Bromuconazole isomer 2\" target=\"_blank\" rel=\"noopener\">Bromuconazole isomer 2<\/a><\/td><td class=\"oth\">4.740<\/td><td class=\"oth\">378.0<\/td><td class=\"oth\">159.0<\/td><td class=\"oth\">159.0<\/td><\/tr>\n<tr><td class=\"num\">143.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Flubendiamide\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/272451-65-7\/Flubendiamide\" target=\"_blank\" rel=\"noopener\">Flubendiamide<\/a><\/td><td class=\"oth\">4.745<\/td><td class=\"oth\">700.0<\/td><td class=\"oth\">408.0<\/td><td class=\"oth\">274.0<\/td><\/tr>\n<tr><td class=\"num\">144.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Diclobutrazol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/75736-33-3\/Diclobutrazol\" target=\"_blank\" rel=\"noopener\">Diclobutrazol<\/a><\/td><td class=\"oth\">4.764<\/td><td class=\"oth\">328.0<\/td><td class=\"oth\">70.1<\/td><td class=\"oth\">159.0<\/td><\/tr>\n<tr><td class=\"num\">145.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Carfentrazone ethyl\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/128639-02-1\/Carfentrazone ethyl\" target=\"_blank\" rel=\"noopener\">Carfentrazone ethyl<\/a><\/td><td class=\"oth\">4.773<\/td><td class=\"oth\">429.1<\/td><td class=\"oth\">412.1<\/td><td class=\"oth\">346.1<\/td><\/tr>\n<tr><td class=\"num\">146.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Kresoxim-methyl\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/143390-89-0\/Kresoxim-methyl\" target=\"_blank\" rel=\"noopener\">Kresoxim-methyl<\/a><\/td><td class=\"oth\">4.795<\/td><td class=\"oth\">314.0<\/td><td class=\"oth\">267.1<\/td><td class=\"oth\">116.0<\/td><\/tr>\n<tr><td class=\"num\">147.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Tebuconazole\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/107534-96-3\/Tebuconazole\" target=\"_blank\" rel=\"noopener\">Tebuconazole<\/a><\/td><td class=\"oth\">4.827<\/td><td class=\"oth\">308.0<\/td><td class=\"oth\">70.1<\/td><td class=\"oth\">124.9<\/td><\/tr>\n<tr><td class=\"num\">148.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Spinosad A\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/168316-95-8\/Spinosad A\" target=\"_blank\" rel=\"noopener\">Spinosad A<\/a><\/td><td class=\"oth\">4.840<\/td><td class=\"oth\">732.5<\/td><td class=\"oth\">142.1<\/td><td class=\"oth\">98.1<\/td><\/tr>\n<tr><td class=\"num\">149.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Penconazole\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/66246-88-6\/Penconazole\" target=\"_blank\" rel=\"noopener\">Penconazole<\/a><\/td><td class=\"oth\">4.853<\/td><td class=\"oth\">284.1<\/td><td class=\"oth\">70.1<\/td><td class=\"oth\">159.0<\/td><\/tr>\n<tr><td class=\"num\">150.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Prothioconazole\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/178928-70-6\/Prothioconazole\" target=\"_blank\" rel=\"noopener\">Prothioconazole<\/a><\/td><td class=\"oth\">4.857<\/td><td class=\"oth\">344.0<\/td><td class=\"oth\">189.0<\/td><td class=\"oth\">154.0<\/td><\/tr>\n<tr><td class=\"num\">151.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Zoxamide\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/156052-68-5\/Zoxamide\" target=\"_blank\" rel=\"noopener\">Zoxamide<\/a><\/td><td class=\"oth\">4.893<\/td><td class=\"oth\">336.1<\/td><td class=\"oth\">187.0<\/td><td class=\"oth\">159.0<\/td><\/tr>\n<tr><td class=\"num\">152.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Famoxadone\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/131807-57-3\/Famoxadone\" target=\"_blank\" rel=\"noopener\">Famoxadone<\/a><\/td><td class=\"oth\">4.933<\/td><td class=\"oth\">392.2<\/td><td class=\"oth\">331.1<\/td><td class=\"oth\">238.0<\/td><\/tr>\n<tr><td class=\"num\">153.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Diazinon-diethyl-d10 (IS)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/100155-47-3\/Diazinon-diethyl-d10\" target=\"_blank\" rel=\"noopener\">Diazinon-diethyl-d10 (IS)<\/a><\/td><td class=\"oth\">4.961<\/td><td class=\"oth\">315.2<\/td><td class=\"oth\">170.2<\/td><td class=\"oth\">154.3<\/td><\/tr>\n<tr><td class=\"num\">154.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Hydramethylnon\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/67485-29-4\/Hydramethylnon\" target=\"_blank\" rel=\"noopener\">Hydramethylnon<\/a><\/td><td class=\"oth\">4.979<\/td><td class=\"oth\">495.2<\/td><td class=\"oth\">323.2<\/td><td class=\"oth\">45.2<\/td><\/tr>\n<tr><td class=\"num\">155.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Triflumuron\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/64628-44-0\/Triflumuron\" target=\"_blank\" rel=\"noopener\">Triflumuron<\/a><\/td><td class=\"oth\">4.986<\/td><td class=\"oth\">359.1<\/td><td class=\"oth\">156.0<\/td><td class=\"oth\">111.0<\/td><\/tr>\n<tr><td class=\"num\">156.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Prochloraz\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/67747-09-5\/Prochloraz\" target=\"_blank\" rel=\"noopener\">Prochloraz<\/a><\/td><td class=\"oth\">4.986<\/td><td class=\"oth\">376.0<\/td><td class=\"oth\">308.0<\/td><td class=\"oth\">70.1<\/td><\/tr>\n<tr><td class=\"num\">157.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Hexaconazole\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/79983-71-4\/Hexaconazole\" target=\"_blank\" rel=\"noopener\">Hexaconazole<\/a><\/td><td class=\"oth\">4.993<\/td><td class=\"oth\">314.1<\/td><td class=\"oth\">70.0<\/td><td class=\"oth\">158.8<\/td><\/tr>\n<tr><td class=\"num\">158.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Benalaxyl\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/71626-11-4\/Benalaxyl\" target=\"_blank\" rel=\"noopener\">Benalaxyl<\/a><\/td><td class=\"oth\">5.021<\/td><td class=\"oth\">326.2<\/td><td class=\"oth\">148.0<\/td><td class=\"oth\">208.1<\/td><\/tr>\n<tr><td class=\"num\">159.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Metconazole\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/125116-23-6\/Metconazole\" target=\"_blank\" rel=\"noopener\">Metconazole<\/a><\/td><td class=\"oth\">5.033<\/td><td class=\"oth\">320.1<\/td><td class=\"oth\">70.1<\/td><td class=\"oth\">125.0<\/td><\/tr>\n<tr><td class=\"num\">160.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Propiconazole\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/60207-90-1\/Propiconazole\" target=\"_blank\" rel=\"noopener\">Propiconazole<\/a><\/td><td class=\"oth\">5.054<\/td><td class=\"oth\">342.0<\/td><td class=\"oth\">159.0<\/td><td class=\"oth\">69.0<\/td><\/tr>\n<tr><td class=\"num\">161.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Pyraclostrobin\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/175013-18-0\/Pyraclostrobin\" target=\"_blank\" rel=\"noopener\">Pyraclostrobin<\/a><\/td><td class=\"oth\">5.076<\/td><td class=\"oth\">388.0<\/td><td class=\"oth\">194.0<\/td><td class=\"oth\">163.1<\/td><\/tr>\n<tr><td class=\"num\">162.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Clofentezine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/74115-24-5\/Clofentezine\" target=\"_blank\" rel=\"noopener\">Clofentezine<\/a><\/td><td class=\"oth\">5.080<\/td><td class=\"oth\">303.0<\/td><td class=\"oth\">138.0<\/td><td class=\"oth\">102.0<\/td><\/tr>\n<tr><td class=\"num\">163.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Bitertanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/55179-31-2\/Bitertanol\" target=\"_blank\" rel=\"noopener\">Bitertanol<\/a><\/td><td class=\"oth\">5.120<\/td><td class=\"oth\">338.2<\/td><td class=\"oth\">269.2<\/td><td class=\"oth\">70.1<\/td><\/tr>\n<tr><td class=\"num\">164.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Spinosad D\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/168316-95-8\/Spinosad D\" target=\"_blank\" rel=\"noopener\">Spinosad D<\/a><\/td><td class=\"oth\">5.142<\/td><td class=\"oth\">746.5<\/td><td class=\"oth\">142.0<\/td><td class=\"oth\">98.0<\/td><\/tr>\n<tr><td class=\"num\">165.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Benzoximate\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/29104-30-1\/Benzoximate\" target=\"_blank\" rel=\"noopener\">Benzoximate<\/a><\/td><td class=\"oth\">5.171<\/td><td class=\"oth\">364.0<\/td><td class=\"oth\">199.1<\/td><td class=\"oth\">105.0<\/td><\/tr>\n<tr><td class=\"num\">166.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Diniconazole\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/83657-24-3\/Diniconazole\" target=\"_blank\" rel=\"noopener\">Diniconazole<\/a><\/td><td class=\"oth\">5.184<\/td><td class=\"oth\">326.1<\/td><td class=\"oth\">70.0<\/td><td class=\"oth\">158.9<\/td><\/tr>\n<tr><td class=\"num\">167.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Thiobencarb\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/28249-77-6\/Thiobencarb\" target=\"_blank\" rel=\"noopener\">Thiobencarb<\/a><\/td><td class=\"oth\">5.197<\/td><td class=\"oth\">258.0<\/td><td class=\"oth\">125.0<\/td><td class=\"oth\">89.1<\/td><\/tr>\n<tr><td class=\"num\">168.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Pencycuron\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/66063-05-6\/Pencycuron\" target=\"_blank\" rel=\"noopener\">Pencycuron<\/a><\/td><td class=\"oth\">5.229<\/td><td class=\"oth\">329.1<\/td><td class=\"oth\">125.0<\/td><td class=\"oth\">89.0<\/td><\/tr>\n<tr><td class=\"num\">169.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Spinetoram\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/935545-74-7\/Spinetoram\" target=\"_blank\" rel=\"noopener\">Spinetoram<\/a><\/td><td class=\"oth\">5.234<\/td><td class=\"oth\">748.5<\/td><td class=\"oth\">142.1<\/td><td class=\"oth\">98.2<\/td><\/tr>\n<tr><td class=\"num\">170.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Hexaflumuron*\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/86479-06-3\/Hexaflumuron\" target=\"_blank\" rel=\"noopener\">Hexaflumuron*<\/a><\/td><td class=\"oth\">5.351<\/td><td class=\"oth\">459.0<\/td><td class=\"oth\">439.0<\/td><td class=\"oth\">175.0<\/td><\/tr>\n<tr><td class=\"num\">171.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Triflumizole\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/68694-11-1\/Triflumizole\" target=\"_blank\" rel=\"noopener\">Triflumizole<\/a><\/td><td class=\"oth\">5.366<\/td><td class=\"oth\">346.1<\/td><td class=\"oth\">278.0<\/td><td class=\"oth\">73.1<\/td><\/tr>\n<tr><td class=\"num\">172.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Indoxacarb\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/173584-44-6\/Indoxacarb\" target=\"_blank\" rel=\"noopener\">Indoxacarb<\/a><\/td><td class=\"oth\">5.375<\/td><td class=\"oth\">528.1<\/td><td class=\"oth\">149.9<\/td><td class=\"oth\">203.0<\/td><\/tr>\n<tr><td class=\"num\">173.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Difenoconazole\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/119446-68-3\/Difenoconazole\" target=\"_blank\" rel=\"noopener\">Difenoconazole<\/a><\/td><td class=\"oth\">5.380<\/td><td class=\"oth\">406.1<\/td><td class=\"oth\">251.1<\/td><td class=\"oth\">152.0<\/td><\/tr>\n<tr><td class=\"num\">174.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Ipconazole\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/125225-28-7\/Ipconazole\" target=\"_blank\" rel=\"noopener\">Ipconazole<\/a><\/td><td class=\"oth\">5.426<\/td><td class=\"oth\">334.2<\/td><td class=\"oth\">70.0<\/td><td class=\"oth\">124.9<\/td><\/tr>\n<tr><td class=\"num\">175.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Trifloxystrobin\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/141517-21-7\/Trifloxystrobin\" target=\"_blank\" rel=\"noopener\">Trifloxystrobin<\/a><\/td><td class=\"oth\">5.426<\/td><td class=\"oth\">409.0<\/td><td class=\"oth\">186.0<\/td><td class=\"oth\">206.0<\/td><\/tr>\n<tr><td class=\"num\">176.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Novaluron*\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/116714-46-6\/Novaluron\" target=\"_blank\" rel=\"noopener\">Novaluron*<\/a><\/td><td class=\"oth\">5.455<\/td><td class=\"oth\">491.0<\/td><td class=\"oth\">471.0<\/td><td class=\"oth\">304.9<\/td><\/tr>\n<tr><td class=\"num\">177.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Emamectin-benzoate B1b\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/155569-91-8\/Emamectin-benzoate B1b\" target=\"_blank\" rel=\"noopener\">Emamectin-benzoate B1b<\/a><\/td><td class=\"oth\">5.463<\/td><td class=\"oth\">872.2<\/td><td class=\"oth\">158.2<\/td><td class=\"oth\">82.1<\/td><\/tr>\n<tr><td class=\"num\">178.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Clethodim\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/99129-21-2\/Clethodim\" target=\"_blank\" rel=\"noopener\">Clethodim<\/a><\/td><td class=\"oth\">5.565<\/td><td class=\"oth\">360.1<\/td><td class=\"oth\">164.1<\/td><td class=\"oth\">268.1<\/td><\/tr>\n<tr><td class=\"num\">179.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Teflubenzuron\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/83121-18-0\/Teflubenzuron\" target=\"_blank\" rel=\"noopener\">Teflubenzuron<\/a><\/td><td class=\"oth\">5.614<\/td><td class=\"oth\">380.9<\/td><td class=\"oth\">140.9<\/td><td class=\"oth\">158.0<\/td><\/tr>\n<tr><td class=\"num\">180.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Alanycarb\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/83130-01-2\/Alanycarb\" target=\"_blank\" rel=\"noopener\">Alanycarb<\/a><\/td><td class=\"oth\">5.636<\/td><td class=\"oth\">400.1<\/td><td class=\"oth\">91.0<\/td><td class=\"oth\">238.0<\/td><\/tr>\n<tr><td class=\"num\">181.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Buprofezin\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/69327-76-0\/Buprofezin\" target=\"_blank\" rel=\"noopener\">Buprofezin<\/a><\/td><td class=\"oth\">5.642<\/td><td class=\"oth\">306.2<\/td><td class=\"oth\">106.1<\/td><td class=\"oth\">86.0<\/td><\/tr>\n<tr><td class=\"num\">182.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Emamectin-benzoate B1a\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/155569-91-8\/Emamectin-benzoate B1a\" target=\"_blank\" rel=\"noopener\">Emamectin-benzoate B1a<\/a><\/td><td class=\"oth\">5.644<\/td><td class=\"oth\">886.5<\/td><td class=\"oth\">158.0<\/td><td class=\"oth\">82.1<\/td><\/tr>\n<tr><td class=\"num\">183.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Benfuracarb\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/82560-54-1\/Benfuracarb\" target=\"_blank\" rel=\"noopener\">Benfuracarb<\/a><\/td><td class=\"oth\">5.695<\/td><td class=\"oth\">411.2<\/td><td class=\"oth\">194.9<\/td><td class=\"oth\">251.9<\/td><\/tr>\n<tr><td class=\"num\">184.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Fluazinam*\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/79622-59-6\/Fluazinam\" target=\"_blank\" rel=\"noopener\">Fluazinam*<\/a><\/td><td class=\"oth\">5.726<\/td><td class=\"oth\">463.0<\/td><td class=\"oth\">415.9<\/td><td class=\"oth\">397.9<\/td><\/tr>\n<tr><td class=\"num\">185.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Metaflumizone*\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/139968-49-3\/Metaflumizone\" target=\"_blank\" rel=\"noopener\">Metaflumizone*<\/a><\/td><td class=\"oth\">5.728<\/td><td class=\"oth\">505.1<\/td><td class=\"oth\">302.1<\/td><td class=\"oth\">328.0<\/td><\/tr>\n<tr><td class=\"num\">186.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Furathiocarb\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/65907-30-4\/Furathiocarb\" target=\"_blank\" rel=\"noopener\">Furathiocarb<\/a><\/td><td class=\"oth\">5.735<\/td><td class=\"oth\">383.1<\/td><td class=\"oth\">195.0<\/td><td class=\"oth\">252.1<\/td><\/tr>\n<tr><td class=\"num\">187.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Temephos\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/3383-96-8\/Temephos\" target=\"_blank\" rel=\"noopener\">Temephos<\/a><\/td><td class=\"oth\">5.757<\/td><td class=\"oth\">467.1<\/td><td class=\"oth\">125.0<\/td><td class=\"oth\">418.9<\/td><\/tr>\n<tr><td class=\"num\">188.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Lufenuron\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/103055-07-8\/Lufenuron\" target=\"_blank\" rel=\"noopener\">Lufenuron<\/a><\/td><td class=\"oth\">5.767<\/td><td class=\"oth\">511.2<\/td><td class=\"oth\">141.0<\/td><td class=\"oth\">158.0<\/td><\/tr>\n<tr><td class=\"num\">189.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Tebufenpyrad\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/119168-77-3\/Tebufenpyrad\" target=\"_blank\" rel=\"noopener\">Tebufenpyrad<\/a><\/td><td class=\"oth\">5.778<\/td><td class=\"oth\">334.0<\/td><td class=\"oth\">117.0<\/td><td class=\"oth\">145.0<\/td><\/tr>\n<tr><td class=\"num\">190.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Pyriproxyfen\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/95737-68-1\/Pyriproxyfen\" target=\"_blank\" rel=\"noopener\">Pyriproxyfen<\/a><\/td><td class=\"oth\">5.843<\/td><td class=\"oth\">322.0<\/td><td class=\"oth\">96.0<\/td><td class=\"oth\">227.1<\/td><\/tr>\n<tr><td class=\"num\">191.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Piperonyl butoxide\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/51-03-6\/Piperonyl butoxide\" target=\"_blank\" rel=\"noopener\">Piperonyl butoxide<\/a><\/td><td class=\"oth\">5.864<\/td><td class=\"oth\">356.2<\/td><td class=\"oth\">177.0<\/td><td class=\"oth\">119.1<\/td><\/tr>\n<tr><td class=\"num\">192.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Hexythiazox\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/78587-05-0\/Hexythiazox\" target=\"_blank\" rel=\"noopener\">Hexythiazox<\/a><\/td><td class=\"oth\">5.924<\/td><td class=\"oth\">353.2<\/td><td class=\"oth\">228.0<\/td><td class=\"oth\">168.1<\/td><\/tr>\n<tr><td class=\"num\">193.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Quinoxyfen\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/124495-18-7\/Quinoxyfen\" target=\"_blank\" rel=\"noopener\">Quinoxyfen<\/a><\/td><td class=\"oth\">5.935<\/td><td class=\"oth\">308.1<\/td><td class=\"oth\">197.0<\/td><td class=\"oth\">162.0<\/td><\/tr>\n<tr><td class=\"num\">194.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Flufenoxuron\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/101463-69-8\/Flufenoxuron\" target=\"_blank\" rel=\"noopener\">Flufenoxuron<\/a><\/td><td class=\"oth\">5.984<\/td><td class=\"oth\">489.1<\/td><td class=\"oth\">158.0<\/td><td class=\"oth\">141.0<\/td><\/tr>\n<tr><td class=\"num\">195.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Propargite\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/2312-35-8\/Propargite\" target=\"_blank\" rel=\"noopener\">Propargite<\/a><\/td><td class=\"oth\">6.088<\/td><td class=\"oth\">368.1<\/td><td class=\"oth\">231.1<\/td><td class=\"oth\">175.0<\/td><\/tr>\n<tr><td class=\"num\">196.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Etoxazole\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/153233-91-1\/Etoxazole\" target=\"_blank\" rel=\"noopener\">Etoxazole<\/a><\/td><td class=\"oth\">6.099<\/td><td class=\"oth\">360.2<\/td><td class=\"oth\">141.0<\/td><td class=\"oth\">57.1<\/td><\/tr>\n<tr><td class=\"num\">197.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Spiromesifen\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/283594-90-1\/Spiromesifen\" target=\"_blank\" rel=\"noopener\">Spiromesifen<\/a><\/td><td class=\"oth\">6.129<\/td><td class=\"oth\">371.3<\/td><td class=\"oth\">273.2<\/td><td class=\"oth\">255.2<\/td><\/tr>\n<tr><td class=\"num\">198.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Chlorfluazuron\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/71422-67-8\/Chlorfluazuron\" target=\"_blank\" rel=\"noopener\">Chlorfluazuron<\/a><\/td><td class=\"oth\">6.165<\/td><td class=\"oth\">539.8<\/td><td class=\"oth\">382.9<\/td><td class=\"oth\">158.0<\/td><\/tr>\n<tr><td class=\"num\">199.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Spirodiclofen\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/148477-71-8\/Spirodiclofen\" target=\"_blank\" rel=\"noopener\">Spirodiclofen<\/a><\/td><td class=\"oth\">6.269<\/td><td class=\"oth\">428.3<\/td><td class=\"oth\">71.1<\/td><td class=\"oth\">313.1<\/td><\/tr>\n<tr><td class=\"num\">200.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for (E)-Fenpyroximate\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/111812-58-9\/(E)-Fenpyroximate\" target=\"_blank\" rel=\"noopener\">(E)-Fenpyroximate<\/a><\/td><td class=\"oth\">6.283<\/td><td class=\"oth\">422.1<\/td><td class=\"oth\">366.1<\/td><td class=\"oth\">138.1<\/td><\/tr>\n<tr><td class=\"num\">201.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Pyridaben\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/96489-71-3\/Pyridaben\" target=\"_blank\" rel=\"noopener\">Pyridaben<\/a><\/td><td class=\"oth\">6.442<\/td><td class=\"oth\">365.2<\/td><td class=\"oth\">147.0<\/td><td class=\"oth\">309.1<\/td><\/tr>\n<tr><td class=\"num\">202.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Eprinomectin\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/123997-26-2\/Eprinomectin\" target=\"_blank\" rel=\"noopener\">Eprinomectin<\/a><\/td><td class=\"oth\">6.466<\/td><td class=\"oth\">914.6<\/td><td class=\"oth\">186.1<\/td><td class=\"oth\">154.1<\/td><\/tr>\n<tr><td class=\"num\">203.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Abamectin B1a\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/71751-41-2\/Abamectin B1a\" target=\"_blank\" rel=\"noopener\">Abamectin B1a<\/a><\/td><td class=\"oth\">6.544<\/td><td class=\"oth\">890.6<\/td><td class=\"oth\">305.4<\/td><td class=\"oth\">567.2<\/td><\/tr>\n<tr><td class=\"num\">204.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Fenazaquin\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/120928-09-8\/Fenazaquin\" target=\"_blank\" rel=\"noopener\">Fenazaquin<\/a><\/td><td class=\"oth\">6.559<\/td><td class=\"oth\">307.1<\/td><td class=\"oth\">161.0<\/td><td class=\"oth\">131.1<\/td><\/tr>\n<tr><td class=\"num\">205.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Doramectin\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/117704-25-3\/Doramectin\" target=\"_blank\" rel=\"noopener\">Doramectin<\/a><\/td><td class=\"oth\">6.733<\/td><td class=\"oth\">916.6<\/td><td class=\"oth\">331.2<\/td><td class=\"oth\">593.4<\/td><\/tr>\n<tr><td class=\"num\">206.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Ivermectin\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/70288-86-7\/Ivermectin\" target=\"_blank\" rel=\"noopener\">Ivermectin<\/a><\/td><td class=\"oth\">6.893<\/td><td class=\"oth\">892.6<\/td><td class=\"oth\">551.4<\/td><td class=\"oth\">569.4<\/td><\/tr>\n<\/tbody><\/table><div class=\"peaknotes\">*Detected in ESI- mode<\/div><\/div><div class=\"chromatogram-conditions\"><h4>Conditions<\/h4><div class=\"conditions-container container-fluid\"><div class=\"row\"><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Column<\/th><td>Raptor ARC-18  (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/9314A12\" rel=\"noopener\">cat.# 9314A12<\/a>)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Dimensions:<\/th><td>100 mm x 2.1 mm ID<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Particle Size:<\/th><td>2.7 \u00b5m<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Pore Size:<\/th><td>90 \u00c5<\/td><\/tr><tr><td><\/td><tr><th class=\"sub conditions_header\" scope=\"row\">Guard Column:<\/th><td>Raptor ARC-18 EXP guard column cartridge 5 mm, 2.1 mm ID, 2.7 \u00b5m (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/9314A0252\" rel=\"noopener\">cat.# 9314A0252<\/a>)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Temp.:<\/th><td>50 \u00b0C<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Standard\/Sample<\/th><td>LC multiresidue pesticide kit (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/31971\" rel=\"noopener\">cat.# 31971<\/a>)<\/td><\/tr>\n<tr><th class=\"sub conditions_header\" scope=\"row\">Diluent:<\/th><td>Water:acetonitrile 9:1<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Conc.:<\/th><td>1  ng\/mL see notes for details<\/td><\/tr><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Vol.:<\/th><td>5 \u00b5L <\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Mobile Phase<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">A:<\/th><td>0.2% Formic acid, 2 mM ammonium formate in water <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">B:<\/th><td>0.2% Formic acid, 2 mM ammonium formate in methanol <\/td><\/tr><tr><td><\/td><td><table class=\"cgram_ramp\"><thead><tr><th>Time (min)<\/th><th>Flow (mL\/min)<\/th><th>%A<\/th><th>%B<\/th><\/tr><\/thead><tbody><tr><td>0.00<\/td><td>0.4<\/td><td>95<\/td><td>5<\/td><\/tr><tr><td>2.00<\/td><td>0.4<\/td><td>40<\/td><td>60<\/td><\/tr><tr><td>4.00<\/td><td>0.4<\/td><td>25<\/td><td>75<\/td><\/tr><tr><td>6.00<\/td><td>0.4<\/td><td>0<\/td><td>100<\/td><\/tr><tr><td>7.50<\/td><td>0.4<\/td><td>0<\/td><td>100<\/td><\/tr><tr><td>7.51<\/td><td>0.4<\/td><td>95<\/td><td>5<\/td><\/tr><tr><td>9.50<\/td><td>0.4<\/td><td>95<\/td><td>5<\/td><\/tr><\/tbody><\/table><\/td><\/tr><\/table><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Detector<\/th><td>MS\/MS<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Ion Mode:<\/th><td>ESI+\/ESI- <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Mode:<\/th><td>Scheduled MRM <\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Instrument<\/th><td>UHPLC<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Sample Preparation<\/th><td>Brown rice flour (10 g), purchased from a local market, was mixed with 10 mL water. After mixing for 10 min, the brown rice flour was fortified at 10 ng\/g with all residues from the LC multiresidue pesticide kit (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/31971\" rel=\"noopener\">cat.# 31971<\/a>), which had been previously combined into a single mixture that day. The fortified brown rice flour suspension was then mixed for 30 min to ensure uniform dispersion of the added pesticide residues in the brown rice flour suspension. Extraction was performed using 10 mL of acetonitrile and original unbuffered QuEChERS salts (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/25848\" rel=\"noopener\">cat.# 25848<\/a>). After centrifugation, 1 mL supernatant was added to a 2 mL dSPE vial containing magnesium sulfate, PSA, and C18 (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/26125\" rel=\"noopener\">cat.# 26125<\/a>) for cleanup.  After centrifugation, 100 \u03bcL of the 10 ng\/mL extract was diluted with 900 \u03bcL water and injected onto the UHPLC.  <\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Notes<\/th><td>Although the ten LC multiresidue pesticides kit mixes are formulated to ensure maximum long-term stability and reliability as packaged, stability may become an issue when a large number of compounds with different chemical functionalities are combined together into a single mix. This should be taken into consideration for quantitative analysis. <br \/><br \/>Want even better performance when analyzing metal-sensitive compounds? Check out Inert LC columns at <a href=\"https:\/\/www.restek.com\/inert\" target=\"_blank\" rel=\"noopener\">www.restek.com\/inert<\/a>.<\/td><\/tr><\/table><\/div><\/div><\/div><div class=\"chromatogram-related-skus\"><h4>Related SKUs<\/h4><ul><li>31971<\/li><li>9314A0252<\/li><li>9314A12<\/li><li>31971<\/li><li>25848<\/li><li>26125<\/li><\/ul><\/div><div class=\"chromatogram-related-searches\"><h4>Related Searches<\/h4><ul><li>Brown<\/li><li>Flour<\/li><li>Pesticides<\/li><li>QuEChERS<\/li><li>Rice<\/li><li>Slim Pouch<\/li><li>dSPE<\/li><\/ul><\/div><div class=\"chromatogram-pdf-link\"><a href=\"https:\/\/ez.restek.com\/images\/cgram\/lc_fs0521.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"><svg xmlns=\"http:\/\/www.w3.org\/2000\/svg\" width=\"18\" height=\"18\" viewBox=\"0 0 18 18\"><g data-name=\"Group 2996\"><path data-name=\"Rectangle 1246\" d=\"M0 0h18v18H0z\" style=\"fill: none;\"><\/path><\/g><g data-name=\"Group 2997\"><path data-name=\"Path 729\" d=\"M13.412 11.4v2.017H5.345V11.4H4v2.017a1.349 1.349 0 0 0 1.345 1.345h8.068a1.349 1.349 0 0 0 1.345-1.345V11.4zm-.672-2.694-.948-.948-1.741 1.735V4H8.706v5.493L6.965 7.758l-.948.948 3.361 3.361z\" transform=\"translate(-.437 -.414)\" style=\"fill: rgb(13, 123, 196);\"><\/path><\/g><\/svg>Download PDF<\/a><\/div><\/div><\/div>\n<\/div><\/div><\/div>\n\n\n<div style=\"height:100px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n","protected":false},"excerpt":{"rendered":"<p>L\u2019analyse multi-r\u00e9sidus de pesticides n\u2019est pas chose ais\u00e9e, et elle est rendue encore plus difficile par les matrices d\u2019\u00e9chantillons complexes et l\u2019instrumentation LC-MS\/MS. La m\u00e9thode chromatographique optimis\u00e9e d\u00e9crite ici permet l\u2019analyse de centaines de compos\u00e9s en moins de 7.5 minutes, avec un cycle analytique total de moins de 10 minutes, garantissant un screening&#8221; rapide et pr\u00e9cis des \u00e9chantillons et une productivit\u00e9 accrue du laboratoire.&#8221;<\/p>\n","protected":false},"author":11,"featured_media":70545,"comment_status":"closed","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"_acf_changed":false,"_kad_blocks_custom_css":"","_kad_blocks_head_custom_js":"","_kad_blocks_body_custom_js":"","_kad_blocks_footer_custom_js":"","_kadence_starter_templates_imported_post":false,"_kad_post_transparent":"","_kad_post_title":"","_kad_post_layout":"","_kad_post_sidebar_id":"","_kad_post_content_style":"","_kad_post_vertical_padding":"","_kad_post_feature":"","_kad_post_feature_position":"","_kad_post_header":false,"_kad_post_footer":false,"footnotes":"","_ppma_block_editor_authors":""},"categories":[462],"tags":[],"industries-application":[2214,2216],"post-badge":[],"resource-type":[],"product-library":[2446],"resource-technique":[2318,2362,2367],"ppma_author":[414],"class_list":["post-43572","post","type-post","status-publish","format-standard","has-post-thumbnail","hentry","category-articles-fr","industries-application-industrie-agroalimentaire","industries-application-securite-alimentaire","product-library-etalons-de-reference","resource-technique-chromatographie-en-phase-liquide","resource-technique-ms-ms-fr","resource-technique-quechers-fr"],"acf":[],"taxonomy_info":{"category":[{"value":462,"label":"Articles"}],"industries-application":[{"value":2214,"label":"Industrie Agroalimentaire"},{"value":2216,"label":"S\u00e9curit\u00e9 alimentaire"}],"product-library":[{"value":2446,"label":"\u00c9talons de R\u00e9f\u00e9rence"}],"resource-technique":[{"value":2318,"label":"Chromatographie en phase liquide"},{"value":2362,"label":"MS\/MS"},{"value":2367,"label":"QuEChERS"}]},"featured_image_src_large":["https:\/\/discover.restek.com\/wp-content\/uploads\/feature-FFSS2930-1024x536.jpg",1024,536,true],"author_info":{"display_name":"Restek Corporation","author_link":"https:\/\/discover.restek.com\/fr\/author\/restek-corporation\/"},"comment_info":0,"category_info":[{"term_id":462,"name":"Articles","slug":"articles-fr","term_group":0,"term_taxonomy_id":462,"taxonomy":"category","description":"","parent":0,"count":507,"filter":"raw","cat_ID":462,"category_count":507,"category_description":"","cat_name":"Articles","category_nicename":"articles-fr","category_parent":0}],"tag_info":false,"authors":[{"term_id":414,"user_id":11,"is_guest":0,"slug":"restek-corporation","display_name":"Restek Corporation","avatar_url":{"url":"https:\/\/discover.restek.com\/wp-content\/uploads\/Restek_Favicon_300x300.jpg","url2x":"https:\/\/discover.restek.com\/wp-content\/uploads\/Restek_Favicon_300x300.jpg"},"avatar_alt":"","avatar_decorative":"","author_category":"1","first_name":"Restek","last_name":"Corporation","user_url":"https:\/\/www.linkedin.com\/company\/restek-corporation\/posts\/?feedView=all","job_title":"Chromatography Products and Solutions","description":"Restek is a leading provider of chromatography columns, accessories, and certified reference materials. Trust Restek for reliable, high-quality analytical solutions."}],"_links":{"self":[{"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/posts\/43572","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/users\/11"}],"replies":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/comments?post=43572"}],"version-history":[{"count":4,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/posts\/43572\/revisions"}],"predecessor-version":[{"id":85655,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/posts\/43572\/revisions\/85655"}],"wp:featuredmedia":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/media\/70545"}],"wp:attachment":[{"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/media?parent=43572"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/categories?post=43572"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/tags?post=43572"},{"taxonomy":"industries-application","embeddable":true,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/industries-application?post=43572"},{"taxonomy":"post-badge","embeddable":true,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/post-badge?post=43572"},{"taxonomy":"resource-type","embeddable":true,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/resource-type?post=43572"},{"taxonomy":"product-library","embeddable":true,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/product-library?post=43572"},{"taxonomy":"resource-technique","embeddable":true,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/resource-technique?post=43572"},{"taxonomy":"author","embeddable":true,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/ppma_author?post=43572"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}