{"id":43635,"date":"2020-10-28T14:30:00","date_gmt":"2020-10-28T14:30:00","guid":{"rendered":"https:\/\/discover.restek.com\/uncategorized\/gc-column-installation-made-easier-with-extended-reducing-nuts-for-the-agilent-split-splitless-inlet\/"},"modified":"2026-08-11T15:18:08","modified_gmt":"2026-08-11T15:18:08","slug":"gc-column-installation-made-easier-with-extended-reducing-nuts-for-the-agilent-split-splitless-inlet","status":"publish","type":"post","link":"https:\/\/discover.restek.com\/fr\/artikel\/gnar2056\/gc-column-installation-made-easier-with-extended-reducing-nuts-for-the-agilent-split-splitless-inlet","title":{"rendered":"GC Column Installation Made Easier with Extended Reducing Nuts for the Agilent Split\/Splitless Inlet"},"content":{"rendered":"\n<p class=\"wp-block-paragraph\">Every new Agilent GC comes with an insulating cup (also known as a nut warmer cup with insulation) that is installed over the bottom of the split\/splitless inlet. The purpose of this insulating cup is to protect the bottom of the inlet from the temperature fluctuations of the GC oven during a temperature ramp. Unfortunately, this insulating cup makes it difficult to install a GC column. The column installation is typically done \u201cblind\u201d as the user cannot see the reducing nut at the bottom of the inlet. Removal of the insulating cup makes it easier to install the GC column and as such, once the insulating cup is removed, it is rarely reinstalled.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">While some analysts run their GCs without the insulating cup and get satisfactory results, unless you compare analyses with and without the insulating cup, you don\u2019t know how good \u201cgood\u201d can be. In Figure 1 we show low volatility pesticides, with one analysis using the insulating cup and with the other analysis not using the insulating cup. You can see from the difference in peak heights that you get better analyte responses with the insulating cup installed.<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><strong>Figure 1:<\/strong>\u00a0Higher analyte responses result from analysis with the insulating cup (nut warmer cup) installed, especially for lower volatility pesticides.<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n<style>.kb-image43635_32e828-f2 .kb-image-has-overlay:after{opacity:0.3;}<\/style>\n<div class=\"wp-block-kadence-image kb-image43635_32e828-f2\"><figure class=\"aligncenter size-large\"><img loading=\"lazy\" decoding=\"async\" width=\"1024\" height=\"477\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR2056-01-1024x477.jpg\" alt=\"\" class=\"kb-img wp-image-20620\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR2056-01-1024x477.jpg 1024w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR2056-01-300x140.jpg 300w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR2056-01-768x358.jpg 768w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR2056-01-1536x716.jpg 1536w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR2056-01.jpg 1800w\" sizes=\"auto, (max-width: 1024px) 100vw, 1024px\" \/><\/figure><\/div>\n\n<\/div><\/div><\/div>\n\n\n<p class=\"wp-block-paragraph\">Not only do you get better analyte response by using the insulating cup, you also see improved repeatability with it installed. Table I shows the analysis of several pesticides using a 4 mm single taper liner with wool at the bottom. The initial oven temperature was set at 90 \u00b0C and seven injections were made. As you can see, when the insulating cup is used, the percent relative standard deviation (RSD) for the peak areas of all compounds is under 6%; whereas, for the analyses when the insulating cup is not used, percent RSD ranged from the mid-twenties up to 45% RSD. For more on the benefits of using the insulating cup, visit our nut warmer cup <a href=\"https:\/\/www.restek.com\/chromablography-library\" target=\"_blank\" rel=\"noopener\">technical blogs<\/a>.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\"><strong>Table I:<\/strong>&nbsp;&nbsp;More consistent results were achieved when analyzing pesticides with the insulating cup installed compared to analysis without it in place. (n = 7)<\/p>\n\n\n\n<figure class=\"wp-block-table\"><table class=\"has-fixed-layout\"><tbody><tr><td>&nbsp;<\/td><td colspan=\"2\"><strong>Percent RSD<\/strong><\/td><\/tr><tr><td><strong>Pesticide<\/strong><\/td><td><strong>Without Insulating Cup<\/strong><\/td><td><strong>With Insulating Cup<\/strong><\/td><\/tr><tr><td>4,4\u2019-DDT<\/td><td>24<\/td><td>4.0<\/td><\/tr><tr><td>Bifenthrin<\/td><td>24<\/td><td>3.9<\/td><\/tr><tr><td>Fenpropathrin<\/td><td>26<\/td><td>4.2<\/td><\/tr><tr><td><em>cis<\/em>-Permethrin<\/td><td>31<\/td><td>4.3<\/td><\/tr><tr><td><em>trans<\/em>-Permethrin<\/td><td>34<\/td><td>4.3<\/td><\/tr><tr><td>Deltamethrin<\/td><td>45<\/td><td>5.6<\/td><\/tr><\/tbody><\/table><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\">As previously stated, the main reason that a user would remove the insulating cup would be to facilitate the installation of the GC column since, with the insulating cup installed, the user would have trouble seeing the bottom of the inlet. Using a\u00a0Restek extended reducing nut\u00a0instead of a standard reducing nut makes column installation easier, as the threaded opening of the inlet where you install the GC column is more visible beneath the bottom of the insulating cup. Note that while the extended reducing nut simplifies column installation by improving visibility, it does not change the column installation distance (i.e., the length from the top of the ferrule to the end of the column).<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><strong>Figure 2:<\/strong>\u00a0An extended reducing nut is visible below the insulating cup, making column installation much simpler.<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n<div class=\"wp-block-image\">\n<figure class=\"aligncenter size-large\"><img loading=\"lazy\" decoding=\"async\" width=\"1024\" height=\"284\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR2056-02-1024x284.jpg\" alt=\"\" class=\"wp-image-20626\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR2056-02-1024x284.jpg 1024w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR2056-02-300x83.jpg 300w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR2056-02-768x213.jpg 768w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR2056-02-1536x427.jpg 1536w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR2056-02.jpg 1800w\" sizes=\"auto, (max-width: 1024px) 100vw, 1024px\" \/><\/figure>\n<\/div>\n<\/div><\/div><\/div>\n\n\n<p class=\"wp-block-paragraph\"><\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Restek&#8217;s extended reducing nuts, which are used with\u00a0extended capillary column nuts, perform similarly to the original standard capillary nut. To demonstrate this, a hydrocarbon test mix containing\u00a0<em>n<\/em>-alkanes covering a wide range of volatilities was analyzed. Multiple analyses on the extended reducing nut were compared to multiple analyses on the standard reducing nut. The ratio of peak area for C40 versus C8 was used to determine if any discrimination was seen between these two styles of reducing nuts. The average ratio of the C40 to C8 areas for the standard capillary nut was 1.01. The average ratio of the C40 to C8 areas for the extended capillary nut was 1.04. There was no difference between the performances of these two capillary nuts. Figure 3 shows an overlay of two chromatograms, one showing this analysis on a standard reducing nut and one showing this analysis on the extended reducing nut.<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><strong>Figure 3:<\/strong>\u00a0Overlay of a Standard Reducing Nut vs. Restek Extended Reducing Inlet Nut while running FL TRPH Standard.<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n<div class=\"wp-block-custom-chromatogram-article\"><div class=\"wp-block-custom-chromatogram-article\"><div class=\"chromatogram-image wide-image\"><img decoding=\"async\" src=\"https:\/\/ez.restek.com\/images\/cgram\/gc_gn1187.png\" alt=\"Comparison: Agilent Inlet Reducing Nut vs. Restek Extended 8 mm Reducing Nut\" title=\"-\"><\/div><p class=\"article-id\" style=\"text-align:center\"> GC_GN1187<\/p><div class=\"chromatogram-peaks\"><h4>Peaks<\/h4><table class=\"peaks col-lg-6 col-12 peak-50\">\n<thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 50px\">t<sub>R<\/sub> (min)<\/th><\/tr><\/thead>\n<tbody><tr><td class=\"num\">1.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Octane\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/111-65-9\/Octane\" target=\"_blank\" rel=\"noopener\">Octane<\/a><\/td><td class=\"oth\">1.23<\/td><\/tr>\n<tr><td class=\"num\">2.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Decane\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/124-18-5\/Decane\" target=\"_blank\" rel=\"noopener\">Decane<\/a><\/td><td class=\"oth\">2.70<\/td><\/tr>\n<tr><td class=\"num\">3.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Dodecane\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/112-40-3\/Dodecane\" target=\"_blank\" rel=\"noopener\">Dodecane<\/a><\/td><td class=\"oth\">4.18<\/td><\/tr>\n<tr><td class=\"num\">4.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Tetradecane\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/629-59-4\/Tetradecane\" target=\"_blank\" rel=\"noopener\">Tetradecane<\/a><\/td><td class=\"oth\">5.48<\/td><\/tr>\n<tr><td class=\"num\">5.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Hexadecane\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/544-76-3\/Hexadecane\" target=\"_blank\" rel=\"noopener\">Hexadecane<\/a><\/td><td class=\"oth\">6.64<\/td><\/tr>\n<tr><td class=\"num\">6.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Octadecane\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/593-45-3\/Octadecane\" target=\"_blank\" rel=\"noopener\">Octadecane<\/a><\/td><td class=\"oth\">7.68<\/td><\/tr>\n<tr><td class=\"num\">7.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Eicosane\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/112-95-8\/Eicosane\" target=\"_blank\" rel=\"noopener\">Eicosane<\/a><\/td><td class=\"oth\">8.63<\/td><\/tr>\n<tr><td class=\"num\">8.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Docosane\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/629-97-0\/Docosane\" target=\"_blank\" rel=\"noopener\">Docosane<\/a><\/td><td class=\"oth\">9.50<\/td><\/tr>\n<\/tbody><\/table>\n<table class=\"peaks col-lg-6 col-12 peak-50\"><thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 50px\">t<sub>R<\/sub> (min)<\/th><\/tr><\/thead><tbody>\n<tr><td class=\"num\">9.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for n-Tetracosane\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/646-31-1\/n-Tetracosane\" target=\"_blank\" rel=\"noopener\">n-Tetracosane<\/a><\/td><td class=\"oth\">10.30<\/td><\/tr>\n<tr><td class=\"num\">10.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for n-Hexacosane\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/630-01-3\/n-Hexacosane\" target=\"_blank\" rel=\"noopener\">n-Hexacosane<\/a><\/td><td class=\"oth\">11.04<\/td><\/tr>\n<tr><td class=\"num\">11.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for n-Octacosane\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/630-02-4\/n-Octacosane\" target=\"_blank\" rel=\"noopener\">n-Octacosane<\/a><\/td><td class=\"oth\">11.73<\/td><\/tr>\n<tr><td class=\"num\">12.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for n-Triacontane\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/638-68-6\/n-Triacontane\" target=\"_blank\" rel=\"noopener\">n-Triacontane<\/a><\/td><td class=\"oth\">12.38<\/td><\/tr>\n<tr><td class=\"num\">13.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for n-Dotriacontane\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/544-85-4\/n-Dotriacontane\" target=\"_blank\" rel=\"noopener\">n-Dotriacontane<\/a><\/td><td class=\"oth\">12.98<\/td><\/tr>\n<tr><td class=\"num\">14.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for n-Tetratriacontane\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/14167-59-0\/n-Tetratriacontane\" target=\"_blank\" rel=\"noopener\">n-Tetratriacontane<\/a><\/td><td class=\"oth\">13.55<\/td><\/tr>\n<tr><td class=\"num\">15.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for n-Hexatriacontane\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/630-06-8\/n-Hexatriacontane\" target=\"_blank\" rel=\"noopener\">n-Hexatriacontane<\/a><\/td><td class=\"oth\">14.09<\/td><\/tr>\n<tr><td class=\"num\">16.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for n-Octatriacontane\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/7194-85-6\/n-Octatriacontane\" target=\"_blank\" rel=\"noopener\">n-Octatriacontane<\/a><\/td><td class=\"oth\">14.60<\/td><\/tr>\n<tr><td class=\"num\">17.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for n-Tetracontane\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/4181-95-7\/n-Tetracontane\" target=\"_blank\" rel=\"noopener\">n-Tetracontane<\/a><\/td><td class=\"oth\">15.08<\/td><\/tr>\n<\/tbody><\/table><\/div><div class=\"chromatogram-conditions\"><h4>Conditions<\/h4><div class=\"conditions-container container-fluid\"><div class=\"row\"><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Column<\/th><td>Rxi-1HT, 15 m, 0.32 mm ID, 0.10 \u00b5m (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/13953\" rel=\"noopener\">cat.# 13953<\/a>)<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Standard\/Sample<\/th><td>Florida TRPH standard  (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/31266\" rel=\"noopener\">cat.# 31266<\/a>)<\/td><\/tr>\n<tr><th class=\"sub conditions_header\" scope=\"row\">Diluent:<\/th><td>Hexane<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Conc.:<\/th><td>50 ppm 1 ng on column<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Injection<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Vol.:<\/th><td>1 \u00b5L split (split ratio 50:1)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Liner:<\/th><td>Premium 4.0 mm ID Precision inlet liner w\/wool (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/23305\" rel=\"noopener\">cat.# 23305<\/a>)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Temp.:<\/th><td>300 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Split Vent Flow Rate:<\/th><td>114 mL\/min<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\" colspan=\"2\">Oven<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Oven Temp.:<\/th><td>40 &#176;C (hold 1.0 min) to 400 &#176;C at 20 &#176;C\/min<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Carrier Gas<\/th><td>He, constant flow<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Flow Rate:<\/th><td>2.3 mL\/min<tr><th class=\"sub conditions_header\" scope=\"row\">Linear Velocity:<\/th><td>55 cm\/sec @ 40 \u00b0C<\/td><\/tr><\/table><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Detector<\/th><td>FID @ 400 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Make-up Gas Flow Rate:<\/th><td>45 mL\/min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Make-up Gas Type:<\/th><td>N<sub>2<\/sub> <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Hydrogen flow:<\/th><td>40 mL\/min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Air flow:<\/th><td>450 mL\/min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Data Rate:<\/th><td>50 Hz<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Instrument<\/th><td>Agilent\/HP6890  GC<\/td><\/tr><\/table><\/div><\/div><\/div><div class=\"chromatogram-related-skus\"><h4>Related SKUs<\/h4><ul><li>31266<\/li><li>13953<\/li><li>23305<\/li><\/ul><\/div><div class=\"chromatogram-related-searches\"><h4>Related Searches<\/h4><ul><li>extended<\/li><li>inlet<\/li><li>knurled<\/li><li>nut<\/li><\/ul><\/div><div class=\"chromatogram-pdf-link\"><a href=\"https:\/\/ez.restek.com\/images\/cgram\/gc_gn1187.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"><svg xmlns=\"http:\/\/www.w3.org\/2000\/svg\" width=\"18\" height=\"18\" viewBox=\"0 0 18 18\"><g data-name=\"Group 2996\"><path data-name=\"Rectangle 1246\" d=\"M0 0h18v18H0z\" style=\"fill: none;\"><\/path><\/g><g data-name=\"Group 2997\"><path data-name=\"Path 729\" d=\"M13.412 11.4v2.017H5.345V11.4H4v2.017a1.349 1.349 0 0 0 1.345 1.345h8.068a1.349 1.349 0 0 0 1.345-1.345V11.4zm-.672-2.694-.948-.948-1.741 1.735V4H8.706v5.493L6.965 7.758l-.948.948 3.361 3.361z\" transform=\"translate(-.437 -.414)\" style=\"fill: rgb(13, 123, 196);\"><\/path><\/g><\/svg>Download PDF<\/a><\/div><\/div><\/div>\n<\/div><\/div><\/div>\n\n\n<p class=\"wp-block-paragraph\">Based on the data shown here, keeping the insulating cup installed on your GC can improve both peak response and analytical reproducibility. In order to reap the benefits of using the insulating cup while keeping column installation easy, we recommend replacing standard reducing nuts with extended reducing nuts for the Agilent split\/splitless GC inlet. This simple change improves visibility of the nut during column installation and does not affect chromatographic performance.<\/p>\n\n\n\n<div style=\"height:50px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n        <div class=\"cpb\">\n            <h3 class=\"cpb-heading\">Produits mentionn\u00e9s<\/h3>\n            <hr class=\"cpb-heading-underline\" \/>\n            <div class=\"cpb-list\">\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/22635\">                                R\u00e9f. 22635                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Ecrou r\u00e9ducteur rallong\u00e9. Pour GC Agilent.<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/22635\">Voir le produit<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/22636\">                                R\u00e9f. 22636                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">\u00c9crou rallong\u00e9 pour colonne capillaire pour ferrule standard 1\/16&quot;, pour GC Agilent<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/22636\">Voir le produit<\/a>\n                                                    <\/div>\n                    <\/div>\n                            <\/div>\n        <\/div>\n        \n\n\n<div style=\"height:100px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n","protected":false},"excerpt":{"rendered":"<p>In order to make GC column installation easier when using an insulating cup in your Agilent split\/splitless GC inlet, standard reducing nuts should be replaced with extended reducing nuts. This simple change improves visibility of the nut during column installation and allows you to reap the benefits of using the insulating cup.<\/p>\n","protected":false},"author":12,"featured_media":90030,"comment_status":"closed","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"_acf_changed":false,"_kad_blocks_custom_css":"","_kad_blocks_head_custom_js":"","_kad_blocks_body_custom_js":"","_kad_blocks_footer_custom_js":"","_kadence_starter_templates_imported_post":false,"_kad_post_transparent":"","_kad_post_title":"","_kad_post_layout":"","_kad_post_sidebar_id":"","_kad_post_content_style":"","_kad_post_vertical_padding":"","_kad_post_feature":"","_kad_post_feature_position":"","_kad_post_header":false,"_kad_post_footer":false,"footnotes":"","_ppma_block_editor_authors":""},"categories":[787],"tags":[],"industries-application":[],"post-badge":[],"resource-type":[563],"product-library":[2480,2492,2494],"resource-technique":[2324],"ppma_author":[431,590,446],"class_list":["post-43635","post","type-post","status-publish","format-standard","has-post-thumbnail","hentry","category-artikel","resource-type-column-installation-and-care","product-library-gaschromatographie-produkte","product-library-gc-saulen","product-library-inlet-zubehor","resource-technique-gaschromatographie-gc"],"acf":[],"taxonomy_info":{"category":[{"value":787,"label":"Artikel"}],"resource-type":[{"value":563,"label":"Column Installation and Care"}],"product-library":[{"value":2480,"label":"Gaschromatographie Produkte"},{"value":2492,"label":"GC S\u00e4ulen"},{"value":2494,"label":"Inlet Zubeh\u00f6r"}],"resource-technique":[{"value":2324,"label":"Gaschromatographie (GC)"}]},"featured_image_src_large":["https:\/\/discover.restek.com\/wp-content\/uploads\/feature-gnar2056-1024x536.jpg",1024,536,true],"author_info":{"display_name":"Scott Adams","author_link":"https:\/\/discover.restek.com\/fr\/author\/scott-adams\/"},"comment_info":0,"category_info":[{"term_id":787,"name":"Artikel","slug":"artikel","term_group":0,"term_taxonomy_id":787,"taxonomy":"category","description":"","parent":0,"count":510,"filter":"raw","cat_ID":787,"category_count":510,"category_description":"","cat_name":"Artikel","category_nicename":"artikel","category_parent":0}],"tag_info":false,"authors":[{"term_id":431,"user_id":12,"is_guest":0,"slug":"scott-adams","display_name":"Scott Adams","avatar_url":{"url":"https:\/\/discover.restek.com\/wp-content\/uploads\/people-adams-scott.jpg","url2x":"https:\/\/discover.restek.com\/wp-content\/uploads\/people-adams-scott.jpg"},"author_category":"1","first_name":"Scott","last_name":"Adams","user_url":"https:\/\/www.linkedin.com\/in\/scottadamsenterprises\/","job_title":"Director of Product Management","description":"Equipped with a BS from The Pennsylvania State University and two years' experience as an environmental chemist at Lancaster Labs, Scott spent 13 years with Alltech Associates\/WR Grace, where he worked in research &amp; development, technical support, applications, product management, and e-business. Most recently before joining Restek, he also worked at Thermo Scientific as an applications specialist. Currently, Scott is Restek's GC business manager"},{"term_id":590,"user_id":0,"is_guest":1,"slug":"jack-cochran","display_name":"Jack Cochran","avatar_url":"https:\/\/secure.gravatar.com\/avatar\/?s=96&d=mm&r=g","author_category":"","first_name":"","last_name":"","user_url":"","job_title":"","description":""},{"term_id":446,"user_id":26,"is_guest":0,"slug":"corby-hilliard","display_name":"Corby Hilliard","avatar_url":{"url":"https:\/\/discover.restek.com\/wp-content\/uploads\/people-Hilliard-Corby.jpg","url2x":"https:\/\/discover.restek.com\/wp-content\/uploads\/people-Hilliard-Corby.jpg"},"author_category":"1","first_name":"Corby","last_name":"Hilliard","user_url":"https:\/\/www.linkedin.com\/in\/corby-hilliard-69b1ba275\/","job_title":"Advanced Scientist, GC Applied","description":"Corby Hilliard started his Restek career in the Quality Assurance department where he spent his first seven years as a QA analyst and worked his way up to a senior QA analyst. He then moved on to the Innovations department in 2009 as a GC Solutions Advanced Scientist. His experience is predominately in gas chromatography (GC) using various detectors and mass spectrometry as well as troubleshooting and method development. His primary work for Restek is generating product application data, using new and existing products in food safety, petrochemical, environmental, and the GC accessories line. He also is involved with quality and R&amp;D as well as new product development."}],"_links":{"self":[{"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/posts\/43635","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/users\/12"}],"replies":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/comments?post=43635"}],"version-history":[{"count":5,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/posts\/43635\/revisions"}],"predecessor-version":[{"id":72230,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/posts\/43635\/revisions\/72230"}],"wp:featuredmedia":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/media\/90030"}],"wp:attachment":[{"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/media?parent=43635"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/categories?post=43635"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/tags?post=43635"},{"taxonomy":"industries-application","embeddable":true,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/industries-application?post=43635"},{"taxonomy":"post-badge","embeddable":true,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/post-badge?post=43635"},{"taxonomy":"resource-type","embeddable":true,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/resource-type?post=43635"},{"taxonomy":"product-library","embeddable":true,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/product-library?post=43635"},{"taxonomy":"resource-technique","embeddable":true,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/resource-technique?post=43635"},{"taxonomy":"author","embeddable":true,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/ppma_author?post=43635"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}