{"id":43697,"date":"2021-07-15T00:00:00","date_gmt":"2021-07-15T00:00:00","guid":{"rendered":"https:\/\/discover.restek.com\/uncategorized\/evaluation-de-linertie-des-injecteurs-gc-avec-des-etalons-de-reference-snap-shoot-dendrine-et-de-ddt\/"},"modified":"2026-01-28T19:02:11","modified_gmt":"2026-01-28T19:02:11","slug":"evaluation-de-linertie-des-injecteurs-gc-avec-des-etalons-de-reference-snap-shoot-dendrine-et-de-ddt","status":"publish","type":"post","link":"https:\/\/discover.restek.com\/fr\/articles-fr\/gnar3689-fr\/evaluation-de-linertie-des-injecteurs-gc-avec-des-etalons-de-reference-snap-shoot-dendrine-et-de-ddt","title":{"rendered":"\u00c9valuation de l\u2019inertie des injecteurs GC avec des \u00e9talons de r\u00e9f\u00e9rence Snap &#038; Shoot&#8221; d\u2019endrine et de DDT&#8221;"},"content":{"rendered":"\n<p>Il existe diff\u00e9rentes mani\u00e8res de d\u00e9terminer l\u2019inertie des syst\u00e8mes de chromatographie. L\u2019un des essais les plus rigoureux pour les injecteurs GC consiste \u00e0 utiliser de faibles concentrations de&nbsp; dichlorodiph\u00e9nyltrichloro\u00e9thane (DDT) et d\u2019endrine comme indicateurs d\u2019inertie. Ces pesticides figurent dans la liste des analytes d\u2019int\u00e9r\u00eat de plusieurs m\u00e9thodes d\u2019analyses environnementales, mais lorsque les instruments sont assez sensibles pour les d\u00e9tecter au niveau des ppb ou des ppt (par exemple, GC-ECD, GC-MS\/MS), ils peuvent aussi servir d&#8217;excellents indicateurs de l\u2019inertie d&#8217;un injecteur GC. Ils sont moins utiles pour les d\u00e9tecteurs moins sensibles (par exemple, GC-FID), o\u00f9 les concentrations d\u00e9tectables d\u2019endrine et de DDT sont bien plus \u00e9lev\u00e9es. Toutefois, \u00e0 des concentrations de l\u2019ordre du ppb, ces deux compos\u00e9s sont excellents pour r\u00e9v\u00e9ler l\u2019inertie r\u00e9elle des injecteurs GC. Restek propose un \u00e9talon de r\u00e9f\u00e9rence pratique, pr\u00e9par\u00e9 aux concentrations appropri\u00e9es et pr\u00eat \u00e0 l\u2019emploi.<\/p>\n\n\n\n<h2 class=\"wp-block-heading\">Pourquoi l\u2019endrine et le DDT&nbsp;?<\/h2>\n\n\n\n<p>L\u2019endrine et le DDT sont d\u2019excellents indicateurs de l\u2019inertie des injecteurs. Si l\u2019inertie du circuit d&#8217;\u00e9chantillon du GC n\u2019est pas parfaite, l\u2019endrine et le DDT subiront probablement une d\u00e9gradation dans l\u2019injecteur. Selon le type de sites actifs pr\u00e9sents dans l\u2019injecteur et leurs degr\u00e9s d\u2019activit\u00e9, la r\u00e9ponse de l\u2019endrine et du DDT va diminuer en raison de leur d\u00e9gradation. La preuve de cette d\u00e9gradation appara\u00eet normalement sur le chromatogramme sous la forme de pics distincts.<\/p>\n\n\n\n<p>Les r\u00e9actions chimiques g\u00e9n\u00e8rent des produits pr\u00e9visibles. Pour l\u2019endrine, l\u2019ouverture de l\u2019anneau \u00e9poxyde pr\u00e9sent dans l\u2019endrine g\u00e9n\u00e8re de l\u2019endrine ald\u00e9hyde et de l\u2019endrine c\u00e9tone, en particulier en pr\u00e9sence de sites actifs acides. Le DDT peut \u00eatre converti en DDE (d\u00e9shydrochloration) et en DDD (d\u00e9chloration). En cas d\u2019activit\u00e9 extr\u00eame dans l\u2019injecteur, l\u2019endrine et le DDT peuvent \u00eatre enti\u00e8rement d\u00e9grad\u00e9s en ces produits et en d\u2019autres produits qui repr\u00e9sentent d\u2019autres r\u00e9actions survenant dans l\u2019injecteur.<\/p>\n\n\n\n<p>Lorsque l\u2019injecteur contient des sites actifs pouvant r\u00e9agir avec l\u2019endrine ou le DDT, les produits de ces r\u00e9actions chimiques se forment avant la colonne GC, c\u2019est pourquoi ils apparaissent sous la forme de pics fins et souvent sym\u00e9triques. Lorsqu\u2019ils atteignent ensuite la colonne analytique, ils font l\u2019objet d\u2019une s\u00e9paration chromatographique comme tous les autres compos\u00e9s inject\u00e9s dans l\u2019instrument.<\/p>\n\n\n\n<p>Ces produits apparaissants sous la forme de pics distincts, ils peuvent \u00eatre surveill\u00e9s. Leur pr\u00e9sence permet d\u2019\u00e9valuer le degr\u00e9 de d\u00e9gradation d\u00fb \u00e0 l\u2019activit\u00e9 de l\u2019injecteur GC, cette mesure est un marqueur fort de l\u2019inertie globale d\u2019un injecteur GC.<\/p>\n\n\n\n<p>La m\u00e9thode US EPA 508.1 est un exemple de m\u00e9thode fond\u00e9e sur ce comportement. Comme d\u2019autres m\u00e9thodes, elle propose un protocole de qualification des instruments avant leur utilisation. Ce protocole permet de v\u00e9rifier que l\u2019injecteur est suffisamment inerte pour la recherche de pesticides dans les \u00e9chantillons. Mais cette technique n\u2019est pas seulement utile aux analystes qui utilisent cette m\u00e9thode sp\u00e9cifique.<\/p>\n\n\n\n<h2 class=\"wp-block-heading\">V\u00e9rification de l\u2019inertie du circuit analytique avec la solution-\u00e9talon pour le contr\u00f4le de la d\u00e9gradation 508.1<\/h2>\n\n\n\n<p>La\u00a0<a href=\"https:\/\/www.nemi.gov\/methods\/method_summary\/4802\/\" target=\"_blank\" rel=\"noopener\">m\u00e9thode US EPA 508.1<\/a>\u00a0combine l\u2019extraction liquide-solide sur disque et la chromatographie en phase gazeuse (GC) sur colonne capillaire avec injection flash, elle utilise aussi un d\u00e9tecteur \u00e0 capture d&#8217;\u00e9lectrons (ECD). Elle est extr\u00eamement sensible aux compos\u00e9s halog\u00e9n\u00e9s comme l\u2019endrine et le DDT. Cette m\u00e9thode permet d\u2019analyser de faibles concentrations (pg dans la colonne) de pesticides halog\u00e9n\u00e9s dans l\u2019eau potable. La d\u00e9gradation (d\u00e9composition) des pesticides est v\u00e9rifi\u00e9e avant l\u2019analyse et au moins toutes les 12 heures pendant une m\u00eame s\u00e9quence d\u2019\u00e9chantillons. Cette v\u00e9rification implique d\u2019injecter un \u00e9talon d\u2019endrine et de DDT, puis de calculer le &#8220;pourcentage de d\u00e9gradation&#8221;. Restek propose un \u00e9talon de r\u00e9f\u00e9rence qui permet de tester l\u2019inertie de l\u2019injecteur, la\u00a0solution-\u00e9talon de contr\u00f4le de la d\u00e9gradation 508.1. Cet \u00e9talon s\u2019utilise sans dilution suppl\u00e9mentaire car il contient d\u00e9j\u00e0 les niveaux de concentration de souhait\u00e9s pour l\u2019endrine et le DDT.<\/p>\n\n\n\n<p>Mode de calcul des d\u00e9gradations :<\/p>\n\n\n<div class=\"wp-block-image\">\n<figure class=\"aligncenter size-large\"><img loading=\"lazy\" decoding=\"async\" width=\"1024\" height=\"84\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/\/equation-article-gnar3689-01a-FR-1024x84.jpg\" alt=\"\" class=\"wp-image-18378\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/equation-article-gnar3689-01a-FR-1024x84.jpg 1024w, https:\/\/discover.restek.com\/wp-content\/uploads\/equation-article-gnar3689-01a-FR-300x25.jpg 300w, https:\/\/discover.restek.com\/wp-content\/uploads\/equation-article-gnar3689-01a-FR-768x63.jpg 768w, https:\/\/discover.restek.com\/wp-content\/uploads\/equation-article-gnar3689-01a-FR.jpg 1236w\" sizes=\"auto, (max-width: 1024px) 100vw, 1024px\" \/><\/figure>\n<\/div>\n\n\n<div style=\"height:20px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n<div class=\"wp-block-image\">\n<figure class=\"aligncenter size-full\"><img loading=\"lazy\" decoding=\"async\" width=\"1021\" height=\"76\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/equation-article-gnar3689-02a-FR.jpg\" alt=\"\" class=\"wp-image-18396\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/equation-article-gnar3689-02a-FR.jpg 1021w, https:\/\/discover.restek.com\/wp-content\/uploads\/equation-article-gnar3689-02a-FR-300x22.jpg 300w, https:\/\/discover.restek.com\/wp-content\/uploads\/equation-article-gnar3689-02a-FR-768x57.jpg 768w\" sizes=\"auto, (max-width: 1021px) 100vw, 1021px\" \/><\/figure>\n<\/div>\n\n\n<p>Dans le cas de la m\u00e9thode US EPA 508.1, si la d\u00e9gradation du DDT ou de l\u2019endrine d\u00e9passe 20&nbsp;%, un entretien de l\u2019injecteur et de la colonne est n\u00e9cessaire pour restaurer les performances. C\u2019est une bonne r\u00e8gle de base pour quiconque souhaite utiliser cet \u00e9talon de r\u00e9f\u00e9rence pour \u00e9valuer l\u2019inertie de l&#8217;injecteur GC. Le mod\u00e9lisateur de chromatogramme Pro&nbsp;<em>EZ<\/em>GC de Restek peut aider \u00e0 d\u00e9terminer les temps d\u2019\u00e9lution de ces composants pour une m\u00e9thode ou une colonne Restek particuli\u00e8re. L\u2019endrine et le DDT doivent \u00eatre s\u00e9par\u00e9s des compos\u00e9s r\u00e9sultant de leur d\u00e9gradation pour garantir l\u2019exactitude des calculs (voir Figure 1).<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><strong>Figure 1\u00a0<\/strong>: Temps d&#8217;\u00e9lution affich\u00e9s dans le logiciel Pro\u00a0<em>EZ<\/em>GC de Restek.<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n<div class=\"wp-block-image\">\n<figure class=\"aligncenter size-large\"><img loading=\"lazy\" decoding=\"async\" width=\"1024\" height=\"659\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/\/figure-article-GNAR3689-01-1024x659.jpg\" alt=\"\" class=\"wp-image-21354\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR3689-01-1024x659.jpg 1024w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR3689-01-300x193.jpg 300w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR3689-01-768x495.jpg 768w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR3689-01-1536x989.jpg 1536w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR3689-01.jpg 1800w\" sizes=\"auto, (max-width: 1024px) 100vw, 1024px\" \/><\/figure>\n<\/div>\n<\/div><\/div><\/div>\n\n\n<div style=\"height:20px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<h2 class=\"wp-block-heading\">Conclusion<\/h2>\n\n\n\n<p>L\u2019endrine et le DDT ont \u00e9t\u00e9 choisis sp\u00e9cifiquement pour la m\u00e9thode US EPA 508.1 car ils sont sensibles aux sites actifs pr\u00e9sents \u00e0 l\u2019int\u00e9rieur des injecteurs et qu\u2019ils exigent une forte inertie pour traverser l\u2019injecteur sans d\u00e9gradation. Les performances pour ces pesticides ont une implication qui d\u00e9passe l\u2019inertie vis-\u00e0-vis du DDT et de l\u2019endrine. De mauvaises performances (forte d\u00e9gradation) pour ces pesticides pr\u00e9disent des probl\u00e8mes semblables pour d\u2019autres analytes sensibles. Les laboratoires qui ne travaillent pas dans le domaine environnemental peuvent aussi v\u00e9rifier l\u2019inertie du circuit d\u2019\u00e9chantillon au moyen de ce test. Les m\u00e9thodes de d\u00e9sactivation des inserts comme les inserts Topaz de Restek utilisent des compositions chimiques complexes pour emp\u00eacher les interactions et la d\u00e9gradation des analytes dans l\u2019injecteur. Chaque lot d\u2019inserts Topaz est test\u00e9 avec le DDT et l\u2019endrine pour d\u00e9montrer leur inertie et pour prouver qu\u2019il convient aux applications les plus exigeantes. La solution-\u00e9talon de contr\u00f4le de la d\u00e9gradation 508.1 de Restek permet d&#8217;\u00e9valuer facilement l\u2019inertie de l\u2019injecteur du GC.<\/p>\n\n\n\n<p>Pour toute question concernant l\u2019utilisation de l\u2019endrine et du DDT comme indicateurs de l\u2019inertie des injecteurs ou pour obtenir de l\u2019aide pour la pr\u00e9paration des \u00e9chantillons ou pour l\u2019analyse, contactez&nbsp;<a href=\"https:\/\/www.restek.com\/global\/fr\/contact-us\" target=\"_blank\" rel=\"noopener\">notre service technique<\/a>.<\/p>\n\n\n\n<div style=\"height:50px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n        <div class=\"cpb\">\n            <h3 class=\"cpb-heading\">Products Mentioned<\/h3>\n            <hr class=\"cpb-heading-underline\" \/>\n            <div class=\"cpb-list\">\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/32093\">                                Catalog No. 32093                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Solution \u00e9talon de contr\u00f4le de la d\u00e9gradation en GC EPA 508.1 n\u00b01, 100 \u00b5g\/ml, ac\u00e9tate d\u2019\u00e9thyle, ampoule de 1ml<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/32093\">Voir le produit<\/a>\n                                                    <\/div>\n                    <\/div>\n                            <\/div>\n        <\/div>\n        \n\n\n<div style=\"height:100px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n","protected":false},"excerpt":{"rendered":"<p>Il existe diff\u00e9rentes mani\u00e8res de d\u00e9terminer l\u2019inertie des syst\u00e8mes de chromatographie. L\u2019un des essais les plus rigoureux pour les injecteurs GC consiste \u00e0 utiliser de faibles concentrations de dichlorodiph\u00e9nyltrichloro\u00e9thane (DDT) et d\u2019endrine comme indicateurs d\u2019inertie.<\/p>\n","protected":false},"author":11,"featured_media":0,"comment_status":"closed","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"_acf_changed":false,"_kad_blocks_custom_css":"","_kad_blocks_head_custom_js":"","_kad_blocks_body_custom_js":"","_kad_blocks_footer_custom_js":"","_kadence_starter_templates_imported_post":false,"_kad_post_transparent":"","_kad_post_title":"","_kad_post_layout":"","_kad_post_sidebar_id":"","_kad_post_content_style":"","_kad_post_vertical_padding":"","_kad_post_feature":"","_kad_post_feature_position":"","_kad_post_header":false,"_kad_post_footer":false,"footnotes":""},"categories":[462],"tags":[],"industries-application":[],"post-badge":[],"resource-type":[],"product-library":[2443,2446,2452,2458,2444],"resource-technique":[2316],"ppma_author":[414],"class_list":["post-43697","post","type-post","status-publish","format-standard","hentry","category-articles-fr","product-library-echantillonnage-dair-et-de-gaz","product-library-etalons-de-reference","product-library-etalons-gazeux-accessoires","product-library-pieces-de-rechange-pour-injecteur","product-library-produits-pour-la-chromatographie-en-phase-gazeuse","resource-technique-chromatographie-en-phase-gazeuse-gc"],"acf":[],"taxonomy_info":{"category":[{"value":462,"label":"Articles"}],"product-library":[{"value":2443,"label":"\u00c9chantillonnage d'air et de gaz"},{"value":2446,"label":"\u00c9talons de R\u00e9f\u00e9rence"},{"value":2452,"label":"\u00c9talons gazeux et accessoires"},{"value":2458,"label":"Pi\u00e8ces de rechange pour injecteur"},{"value":2444,"label":"Produits pour la chromatographie en phase gazeuse"}],"resource-technique":[{"value":2316,"label":"Chromatographie en phase gazeuse (GC)"}]},"featured_image_src_large":false,"author_info":{"display_name":"Restek Corporation","author_link":"https:\/\/discover.restek.com\/fr\/author\/restek-corporation\/"},"comment_info":0,"category_info":[{"term_id":462,"name":"Articles","slug":"articles-fr","term_group":0,"term_taxonomy_id":462,"taxonomy":"category","description":"","parent":0,"count":466,"filter":"raw","cat_ID":462,"category_count":466,"category_description":"","cat_name":"Articles","category_nicename":"articles-fr","category_parent":0}],"tag_info":false,"authors":[{"term_id":414,"user_id":11,"is_guest":0,"slug":"restek-corporation","display_name":"Restek Corporation","avatar_url":{"url":"https:\/\/discover.restek.com\/wp-content\/uploads\/Restek_Favicon_300x300.jpg","url2x":"https:\/\/discover.restek.com\/wp-content\/uploads\/Restek_Favicon_300x300.jpg"},"0":null,"1":"","2":"","3":"","4":"","5":"","6":"","7":"","8":""}],"_links":{"self":[{"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/posts\/43697","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/users\/11"}],"replies":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/comments?post=43697"}],"version-history":[{"count":8,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/posts\/43697\/revisions"}],"predecessor-version":[{"id":85645,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/posts\/43697\/revisions\/85645"}],"wp:attachment":[{"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/media?parent=43697"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/categories?post=43697"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/tags?post=43697"},{"taxonomy":"industries-application","embeddable":true,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/industries-application?post=43697"},{"taxonomy":"post-badge","embeddable":true,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/post-badge?post=43697"},{"taxonomy":"resource-type","embeddable":true,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/resource-type?post=43697"},{"taxonomy":"product-library","embeddable":true,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/product-library?post=43697"},{"taxonomy":"resource-technique","embeddable":true,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/resource-technique?post=43697"},{"taxonomy":"author","embeddable":true,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/ppma_author?post=43697"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}