{"id":43823,"date":"2020-10-27T14:30:00","date_gmt":"2020-10-27T14:30:00","guid":{"rendered":"https:\/\/discover.restek.com\/uncategorized\/gc-inlet-maintenance-resteks-quick-reference-guide\/"},"modified":"2026-09-25T20:52:09","modified_gmt":"2026-09-25T20:52:09","slug":"gc-inlet-maintenance-resteks-quick-reference-guide","status":"publish","type":"post","link":"https:\/\/discover.restek.com\/fr\/articles\/gnss2704\/gc-inlet-maintenance-resteks-quick-reference-guide","title":{"rendered":"GC Inlet Maintenance: Restek\u2019s Quick-Reference Guide"},"content":{"rendered":"\n<ul class=\"wp-block-list\">\n<li>Help ensure quality data by selecting the right GC inlet supplies.<\/li>\n\n\n\n<li>Prevent unwanted downtime with a preventative GC inlet maintenance program.<\/li>\n\n\n\n<li>Easily order the GC inlet supplies you need for your specific instrument.<\/li>\n<\/ul>\n\n\n\n<p class=\"wp-block-paragraph\">Liners, septa, inlet seals&#8230; when compared to your analytical column, your instrument, or a customer lost due to missed deadlines or inaccurate data, these small consumables carry an almost nonexistent price. And yet, the impact they can have on your analysis\u2014not to mention your livelihood\u2014is huge: breakdown of active compounds, loss in response, and other performance issues. In extreme cases, improperly selected or poorly maintained inlet supplies can mean the difference between detecting your compounds or not.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">With Restek\u2019s guidance and our comprehensive line of high-quality GC inlet supplies, performing proper GC inlet maintenance to help maximize your lab\u2019s uptime and ensure accurate, reproducible data is as easy as replacing a liner (and a few other items).<\/p>\n\n\n\n<figure class=\"wp-block-embed is-type-video is-provider-youtube wp-block-embed-youtube wp-embed-aspect-16-9 wp-has-aspect-ratio\"><div class=\"wp-block-embed__wrapper\">\n<iframe loading=\"lazy\" title=\"GC Inlet Maintenance\" width=\"720\" height=\"405\" src=\"https:\/\/www.youtube.com\/embed\/IvycxlyZDaQ?feature=oembed\" frameborder=\"0\" allow=\"accelerometer; autoplay; clipboard-write; encrypted-media; gyroscope; picture-in-picture; web-share\" referrerpolicy=\"strict-origin-when-cross-origin\" allowfullscreen><\/iframe>\n<\/div><\/figure>\n\n\n\n<h2 class=\"wp-block-heading\">Selecting the Right GC Inlet Supplies<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">Using the wrong liner or septa can have just as negative an impact on your data as a fouled inlet. So, even if you are a seasoned GC veteran, it pays to take the time to be sure that you are using the right GC inlet supplies before worrying about how or when to perform GC inlet maintenance.<\/p>\n\n\n\n<div style=\"height:30px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<h2 class=\"wp-block-heading\">Inlet Liners<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">In gas chromatography, the inlet is the first part of the instrument that your sample comes in contact with. Using the right liner will aid in sample vaporization; promote efficient, loss-free analyte transfer; and protect your column from contamination\u2014so choosing the correct GC inlet liner is critical for ensuring accurate and precise data. Liners differ in geometric configuration\/design, volume, base material, deactivation, and packing. But, choosing the right liner can be greatly simplified by looking at the type of injection. Variations in sample and method may necessitate a different liner, but the following are great initial selections for any analysis:<\/p>\n\n\n\n<ul class=\"wp-block-list\">\n<li>Split injection:\n<ul class=\"wp-block-list\">\n<li>Topaz Precision split liner with wool.<\/li>\n<\/ul>\n<\/li>\n\n\n\n<li>Splitless injection:\n<ul class=\"wp-block-list\">\n<li>Topaz single taper liner with wool.<\/li>\n<\/ul>\n<\/li>\n\n\n\n<li>Direct injection:\n<ul class=\"wp-block-list\">\n<li><em>For semivolatiles or compounds that could be affected by a tailing solvent peak:<\/em>&nbsp;Topaz Uniliner liner with the hole drilled near the bottom.<\/li>\n\n\n\n<li><em>For aqueous injections or compounds that elute away from the solvent peak:<\/em>&nbsp;Topaz Uniliner liner with the hole drilled near the top.<\/li>\n<\/ul>\n<\/li>\n\n\n\n<li>Gas samples via sample loop injection:\n<ul class=\"wp-block-list\">\n<li>Topaz straight liner with a 1.0 mm ID.<\/li>\n<\/ul>\n<\/li>\n\n\n\n<li>PTV injection:\n<ul class=\"wp-block-list\">\n<li>Instrument-specific Topaz liner with a small inner diameter and at least one baffle or dimple.<\/li>\n<\/ul>\n<\/li>\n<\/ul>\n\n\n\n<p class=\"wp-block-paragraph\">For more information about liner selection, visit&nbsp;<a href=\"https:\/\/www.restek.com\/articles\/how-to-choose-a-GC-inlet-liner\" data-type=\"link\" data-id=\"https:\/\/www.restek.com\/articles\/how-to-choose-a-GC-inlet-liner\" target=\"_blank\" rel=\"noopener\">www.restek.com\/liner-choice<\/a><\/p>\n\n\n\n<div style=\"height:30px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<h2 class=\"wp-block-heading\">Septa<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">Choosing the right septum for your inlet is even easier. Restek offers several options that are designed to avoid septum bleed and coring, which can damage your column and your data.<\/p>\n\n\n\n<ul class=\"wp-block-list\">\n<li>For temperatures up to 350 \u00b0C*:\n<ul class=\"wp-block-list\">\n<li>Thermolite Plus Septa.<\/li>\n<\/ul>\n<\/li>\n\n\n\n<li>For temperatures up to 400 \u00b0C*:\n<ul class=\"wp-block-list\">\n<li>Premium Non-Stick BTO Septa.<\/li>\n<\/ul>\n<\/li>\n\n\n\n<li>When coring is an issue (can create unwanted activity):\n<ul class=\"wp-block-list\">\n<li>Thermolite Plus or Premium Non-Stick BTO Septa with CenterGuide dimple.<\/li>\n\n\n\n<li>Merlin Microseal Septa.<\/li>\n<\/ul>\n<\/li>\n<\/ul>\n\n\n\n<p class=\"wp-block-paragraph\">For more information choosing the right septum for your analysis, as well as a handy size chart and septum choice guidelines, visit&nbsp;<a href=\"https:\/\/www.restek.com\/articles\/preventing-GC-septum-problems\" target=\"_blank\" rel=\"noopener\">www.restek.com\/septum-choice<\/a>.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">*For 17 mm inlets, the maximum temperature for Thermolite Plus septa is 300 \u00b0C and the maximum temperature for BTO septa is 330 \u00b0C. For all injectors, minimum recommended operating temperature for BTO septa is 250 \u00b0C.<\/p>\n\n\n\n<div style=\"height:30px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<h2 class=\"wp-block-heading\">No More Burned Fingers!<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">Instead of reaching for a hot inlet liner or septa, reach for The Claw (cat.# 26261). This finger-saving tool lets you safely and cleanly remove liners, O-rings, and other hot objects from your inlet! Our inlet liner removal tool (cat.# 20181) is another option for easy and burn-free liner removal.<\/p>\n\n\n\n<div style=\"height:30px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<h2 class=\"wp-block-heading\">Inlet Seals<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">Choosing the right inlet seal is more a matter of convenience and preference than data integrity.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">As a general rule, however, especially for sensitive or splitless analyses, you want to avoid bare stainless steel and, instead, use highly inert gold-plated seals to reduce breakdown and adsorption of active compounds.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Restek offers several models of gold-plated inlet seals with washerless installation, including the customer favorite Dual Vespel Ring inlet seal. Its patented design features two soft Vespel rings\u2014embedded in both the top and bottom surfaces\u2014to help ensure a low-torque, leak-tight seal, even after repeated temperature cycles.<\/p>\n\n\n\n<div style=\"height:30px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<h2 class=\"wp-block-heading\">Which One to Choose&#8230;<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">Do you need additional assistance with selecting the right GC inlet supplies or with GC inlet maintenance? Contact&nbsp;<a href=\"https:\/\/www.restek.com\/contact-us\" data-type=\"link\" data-id=\"https:\/\/www.restek.com\/contact-us\" target=\"_blank\" rel=\"noopener\">Restek Technical Service<\/a>&nbsp;or your&nbsp;<a href=\"https:\/\/www.restek.com\/find-sales-rep\" data-type=\"link\" data-id=\"https:\/\/www.restek.com\/find-sales-rep\" target=\"_blank\" rel=\"noopener\">local Restek representative<\/a>.<\/p>\n\n\n\n<div style=\"height:30px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<h2 class=\"wp-block-heading\">Preventative GC Inlet Maintenance<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">Only after you have confirmed that you are using the right GC inlet supplies is it time to think about GC inlet maintenance. Even within a single industry like environmental testing\u2014let alone across separate industries such as petrochemical, food safety, and clinical diagnostics\u2014there are simply too many variables in samples, injection techniques, instruments, detectors, etc. to be able to recommend a single general schedule to all analysts for replacing all liners, septa, and inlet seals. Your GC inlet maintenance schedule could be twice a day if you are running particularly dirty samples, or once a year if you are using headspace GC. You need to determine your preventative maintenance (PM) schedule on an analysis-by-analysis basis.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">When establishing a PM schedule specific to an analysis, the key term is \u201cpreventative.\u201d You are trying to&nbsp;<em><strong>prevent<\/strong><\/em>&nbsp;unwanted downtime by replacing your inlet supplies&nbsp;<strong><em>before<\/em><\/strong>&nbsp;you have a problem. This way, you can plan for the downtime and choose a time that is not disruptive to your lab. You can also avoid lengthy troubleshooting sessions and missed deadlines.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Simply keep track of how long it generally takes after maintenance to begin seeing data problems commonly associated with the inlet, and then schedule your PM for that analysis and instrument to occur&nbsp;<em><strong>before<\/strong><\/em>&nbsp;data problems tend to arise.<\/p>\n\n\n\n<figure class=\"wp-block-embed is-type-video is-provider-youtube wp-block-embed-youtube wp-embed-aspect-16-9 wp-has-aspect-ratio\"><div class=\"wp-block-embed__wrapper\">\n<iframe loading=\"lazy\" title=\"Replacing a GC Inlet Liner\" width=\"720\" height=\"405\" src=\"https:\/\/www.youtube.com\/embed\/e2sGYmg1B-o?feature=oembed\" frameborder=\"0\" allow=\"accelerometer; autoplay; clipboard-write; encrypted-media; gyroscope; picture-in-picture; web-share\" referrerpolicy=\"strict-origin-when-cross-origin\" allowfullscreen><\/iframe>\n<\/div><\/figure>\n\n\n\n<div style=\"height:30px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<h3 class=\"wp-block-heading\">Common Data Problems that can Alert You to the Need for GC Inlet Maintenance<\/h3>\n\n\n\n<ul class=\"wp-block-list\">\n<li>Loss of signal \/ loss of response:\n<ul class=\"wp-block-list\">\n<li>Inlet activity<\/li>\n\n\n\n<li>Inlet discrimination<\/li>\n\n\n\n<li>Inlet leaks (e.g., seals around septa, O-rings, ferrules, etc.)<\/li>\n<\/ul>\n<\/li>\n\n\n\n<li>Extraneous signal:\n<ul class=\"wp-block-list\">\n<li>Contamination\n<ul class=\"wp-block-list\">\n<li>O-ring bleed<\/li>\n\n\n\n<li>Septum particles in the liner<\/li>\n\n\n\n<li>Matrix buildup in the liner (off-gassing)<\/li>\n<\/ul>\n<\/li>\n\n\n\n<li>Inlet activity\n<ul class=\"wp-block-list\">\n<li>Analyte breakdown creating new compounds (extra peaks)<\/li>\n<\/ul>\n<\/li>\n<\/ul>\n<\/li>\n\n\n\n<li>Distorted signal \/ deteriorating peak shape (e.g., peak tailing):\n<ul class=\"wp-block-list\">\n<li>Inlet activity<\/li>\n\n\n\n<li>Inlet leaks (e.g., seals around septa, O-rings, ferrules, etc.)<\/li>\n<\/ul>\n<\/li>\n<\/ul>\n\n\n\n<p class=\"wp-block-paragraph\">Once you\u2019ve set your preventative maintenance schedule, you will also want to plan ahead for the components you will be replacing and order your GC inlet supplies in advance. Restek\u2019s chromatography experts recommend replacing septa, liners, O-rings, and inlet seals at the same time when you perform GC inlet maintenance. After all, you could unknowingly have issues with multiple components, and the cost of each of these items is minimal compared to having to take your instrument offline more often than needed.<\/p>\n\n\n\n<div style=\"height:30px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<h2 class=\"wp-block-heading\">Leak-Free is the Way to Be<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">A high-quality leak detector\u2014like the Restek Electronic Leak Detector&nbsp;(cat.# 28500)\u2014is a must for every GC lab. Every time you install a new column or perform GC inlet maintenance, be sure to leak-check your GC system before you condition or run an analysis. Minimizing the chances of damaging your column or detector will be well worth the time.<\/p>\n\n\n\n<div style=\"height:30px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<h2 class=\"wp-block-heading\">Peak Performance Ensured by Proper GC Inlet Supplies, Maintained Properly<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">Choosing the right GC inlet supplies\u2014Restek Topaz liners, septa, inlet seals\u2014and implementing a rigorous and timely preventative GC inlet maintenance schedule can help ensure optimum performance, maximum reproducibility, and minimized downtime for your lab.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">The inlet supplies you need to get started are right here, and you can also find them at&nbsp;www.restek.com\/GCacc<\/p>\n\n\n\n<div style=\"height:50px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n        <div class=\"cpb\">\n            <h3 class=\"cpb-heading\">Produits mentionn\u00e9s<\/h3>\n            <hr class=\"cpb-heading-underline\" \/>\n            <div class=\"cpb-list\">\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/26261\">                                R\u00e9f. 26261                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Pince sp\u00e9ciale GC<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/26261\">Voir le produit<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/20181\">                                R\u00e9f. 20181                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Extracteur d&#039;insert d&#039;injection, lot de 3<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/20181\">Voir le produit<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/28500\">                                R\u00e9f. 28500                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">D\u00e9tecteur \u00e9lectronique de fuites de gaz Restek<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/28500\">Voir le produit<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/23303\">                                R\u00e9f. 23303                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Insert Topaz Gooseneck, 4.0\u00a0mm x 6.5 x 78.5, pour GC Agilent, avec laine de quartz, d\u00e9sactivation Premium, lot de 5<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/23303\">Voir le produit<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/23305\">                                R\u00e9f. 23305                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Insert Topaz Precision, 4.0\u00a0mm x 6.3 x 78.5, pour GC Agilent, avec laine de quartz, d\u00e9sactivation Premium, lot de 5<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/23305\">Voir le produit<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/22186\">                                R\u00e9f. 22186                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Kit MLE pour l\u2019installation de colonnes, pour GC Agilent<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/22186\">Voir le produit<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/22181\">                                R\u00e9f. 22181                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Kit d&#039;entretien pour injecteur, pour GC Agilent<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/22181\">Voir le produit<\/a>\n                                                    <\/div>\n                    <\/div>\n                            <\/div>\n        <\/div>\n        \n\n\n<div style=\"height:100px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n","protected":false},"excerpt":{"rendered":"<p>With Resteks guidance and our comprehensive line of high-quality GC inlet supplies, performing proper GC inlet maintenance to help maximize your labs uptime and ensure accurate, reproducible data is as easy as replacing a liner (and a few other items).<\/p>\n","protected":false},"author":11,"featured_media":7405,"comment_status":"closed","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"_acf_changed":false,"_kad_blocks_custom_css":"","_kad_blocks_head_custom_js":"","_kad_blocks_body_custom_js":"","_kad_blocks_footer_custom_js":"","_kadence_starter_templates_imported_post":false,"_kad_post_transparent":"","_kad_post_title":"","_kad_post_layout":"","_kad_post_sidebar_id":"","_kad_post_content_style":"","_kad_post_vertical_padding":"","_kad_post_feature":"","_kad_post_feature_position":"","_kad_post_header":false,"_kad_post_footer":false,"footnotes":"","_ppma_block_editor_authors":""},"categories":[7],"tags":[2794],"industries-application":[],"post-badge":[],"resource-type":[],"product-library":[2372,2386],"resource-technique":[2294],"ppma_author":[414],"class_list":["post-43823","post","type-post","status-publish","format-standard","has-post-thumbnail","hentry","category-articles","tag-topaz-liners","product-library-gas-chromatography-products","product-library-inlet-supplies","resource-technique-gas-chromatography-gc"],"acf":[],"taxonomy_info":{"category":[{"value":7,"label":"Articles"}],"post_tag":[{"value":2794,"label":"Topaz Liners"}],"product-library":[{"value":2372,"label":"Gas Chromatography Products"},{"value":2386,"label":"Inlet Supplies"}],"resource-technique":[{"value":2294,"label":"Gas Chromatography (GC)"}]},"featured_image_src_large":["https:\/\/discover.restek.com\/wp-content\/uploads\/feature-GNSS2704-1024x536.jpg",1024,536,true],"author_info":{"display_name":"Restek Corporation","author_link":"https:\/\/discover.restek.com\/fr\/author\/restek-corporation\/"},"comment_info":0,"category_info":[{"term_id":7,"name":"Articles","slug":"articles","term_group":0,"term_taxonomy_id":7,"taxonomy":"category","description":"","parent":0,"count":506,"filter":"raw","cat_ID":7,"category_count":506,"category_description":"","cat_name":"Articles","category_nicename":"articles","category_parent":0}],"tag_info":[{"term_id":2794,"name":"Topaz Liners","slug":"topaz-liners","term_group":0,"term_taxonomy_id":2794,"taxonomy":"post_tag","description":"Explore Topaz Liner Resources","parent":0,"count":28,"filter":"raw"}],"authors":[{"term_id":414,"user_id":11,"is_guest":0,"slug":"restek-corporation","display_name":"Restek Corporation","avatar_url":{"url":"https:\/\/discover.restek.com\/wp-content\/uploads\/Restek_Favicon_300x300.jpg","url2x":"https:\/\/discover.restek.com\/wp-content\/uploads\/Restek_Favicon_300x300.jpg"},"avatar_alt":"","avatar_decorative":"","author_category":"1","first_name":"Restek","last_name":"Corporation","user_url":"https:\/\/www.linkedin.com\/company\/restek-corporation\/posts\/?feedView=all","job_title":"Chromatography Products and Solutions","description":"Restek is a leading provider of chromatography columns, accessories, and certified reference materials. Trust Restek for reliable, high-quality analytical solutions."}],"_links":{"self":[{"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/posts\/43823","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/users\/11"}],"replies":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/comments?post=43823"}],"version-history":[{"count":9,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/posts\/43823\/revisions"}],"predecessor-version":[{"id":104532,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/posts\/43823\/revisions\/104532"}],"wp:featuredmedia":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/media\/7405"}],"wp:attachment":[{"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/media?parent=43823"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/categories?post=43823"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/tags?post=43823"},{"taxonomy":"industries-application","embeddable":true,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/industries-application?post=43823"},{"taxonomy":"post-badge","embeddable":true,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/post-badge?post=43823"},{"taxonomy":"resource-type","embeddable":true,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/resource-type?post=43823"},{"taxonomy":"product-library","embeddable":true,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/product-library?post=43823"},{"taxonomy":"resource-technique","embeddable":true,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/resource-technique?post=43823"},{"taxonomy":"author","embeddable":true,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/ppma_author?post=43823"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}