{"id":65643,"date":"2025-11-18T16:29:58","date_gmt":"2025-11-18T16:29:58","guid":{"rendered":"https:\/\/discover.restek.com\/?p=65643"},"modified":"2026-08-10T18:49:56","modified_gmt":"2026-08-10T18:49:56","slug":"alcoholic-beverage-analysis-by-gc","status":"publish","type":"post","link":"https:\/\/discover.restek.com\/fr\/artikel\/ffar4494\/alcoholic-beverage-analysis-by-gc","title":{"rendered":"Alcoholic Beverage Analysis by GC"},"content":{"rendered":"\n<h2 class=\"wp-block-heading\">Introduction<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">The volatile component profiles of alcoholic beverages consist of a wide range of compounds, including acids, alcohols, aldehydes, and other trace-level flavor compounds. Analysts trained in the sensory evaluation of distilled liquors, wines, or beers tell us no two products are exactly alike. The unique sensory properties of different types and brands of distilled liquor products often are due to minor differences among the volatile components present. By using instrumental methods to qualitatively or quantitatively evaluate these differences, in addition to sensory techniques, quality assurance analysts can obtain a wealth of information about their products.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Gas chromatography (GC) is not only a powerful tool in the analysis of alcoholic beverages, but it is also a somewhat easy technique to use for the following reasons. Minimal sample preparation is required because the samples are analyzed in their original liquid state with alcohol or alcohol\/water as the matrix. The flavor compounds tend to be volatile in nature, which fulfills one of the main requirements of GC. General detectors, such as the flame ionization detector (FID), or more information-rich detectors, such as the mass selective detector (MSD), can be used. Additionally, the ability to automate the analysis makes GC a very practical tool in a QA\/QC environment. In this article, we will discuss how GC can be used to monitor the volatile components of alcoholic beverages.<\/p>\n\n\n\n<h2 class=\"wp-block-heading\">Analysis of Alcohols and Aldehydes in Alcoholic Beverages<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">Alcoholic beverages contain a wide range of volatile compounds, including alcohols and short-chain aldehydes. GC can be used to analyze these compounds without requiring sample extraction. For example, AOAC International has published methods for the analysis of fusel alcohols (also known as fusel oils), methanol, ethanol, and higher order alcohols by GC [1].<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Since capillary columns offer efficient separations, capillary GC is especially useful in the analysis of structurally similar compounds, such as fusel alcohols. However, selecting a GC column stationary phase with appropriate selectivity is essential for obtaining good peak shapes and excellent resolution of a range of alcohols. The two predominant fusel alcohols, 2-methyl-1-butanol (isoamyl alcohol) and 3-methyl-1-butanol (active amyl alcohol), are critical target compounds that must be resolved. Also, since alcoholic beverage analysis involves liquid samples that are predominantly water\/ethanol-based, water, acetaldehyde, methanol, and ethanol must also be separated to avoid signal suppression during sample analysis. Acetic acid presents an additional challenge because it can exhibit poor peak shapes on some stationary phases.<\/p>\n\n\n\n<figure class=\"wp-block-table\"><table class=\"has-fixed-layout\"><tbody><tr><td><strong>Tech Tip<\/strong><br>Use a gas-tight syringe to maximize syringe lifetime with ethanol samples.<\/td><\/tr><\/tbody><\/table><\/figure>\n\n\n\n<h2 class=\"wp-block-heading\">Choosing a GC Column with the Right Selectivity<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">To determine which GC column is best overall to use for alcoholic beverage analysis, three stationary phases with different polarity were evaluated: Rtx-502.2 (low-polarity); Rtx-VMS (unique polarity for solvents and volatiles); and Rtx-1301 (mid-polarity). Particular attention was paid to peak shapes and the resolution of specific compounds.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">A standard containing many of the target compounds commonly tested for in alcoholic beverage analysis was prepared at 100 \u00b5g\/mL in ethanol. Initial screening was conducted by FID. MSD analysis was then performed with a full scan method (10-400 m\/z) to ensure that water was adequately resolved.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">All three column phases yielded good separation of water, acetaldehyde, methanol, and ethanol. The Rtx-1301 column chromatogram shown in Figure 1 is representative of a typical separation of these compounds.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\"><strong>Figure 1:<\/strong> Typical Water\/Acetaldehyde Separation on an Rtx-1301 Column Using GC-MS.<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article\"><div class=\"wp-block-custom-chromatogram-article\"><div class=\"chromatogram-image wide-image\"><img decoding=\"async\" src=\"https:\/\/ez.restek.com\/images\/cgram\/gc_ff1389.png\" alt=\"Rtx-1301 Water, Acetaldehyde, and Methanol Separation\" title=\"-\"><\/div><p class=\"article-id\" style=\"text-align:center\"> GC_FF1389<\/p><div class=\"chromatogram-peaks\"><h4>Peaks<\/h4><table class=\"peaks col-lg-4 col-12\">\n<thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 50px\">t<sub>R<\/sub> (min)<\/th><th style=\"text-align: center;width: 50px\">Conc.<br \/>(\u00b5g\/mL)<\/th><\/tr><\/thead>\n<tbody><tr><td class=\"num\">1.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Water\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/7732-18-5\/Water\" target=\"_blank\" rel=\"noopener\">Water<\/a><\/td><td class=\"oth\">4.312<\/td><td class=\"oth\"><\/td><\/tr>\n<tr><td class=\"num\">2.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Acetaldehyde\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/75-07-0\/Acetaldehyde\" target=\"_blank\" rel=\"noopener\">Acetaldehyde<\/a><\/td><td class=\"oth\">4.483<\/td><td class=\"oth\">100<\/td><\/tr>\n<tr><td class=\"num\">3.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Methanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/67-56-1\/Methanol\" target=\"_blank\" rel=\"noopener\">Methanol<\/a><\/td><td class=\"oth\">4.699<\/td><td class=\"oth\">100<\/td><\/tr>\n<tr><td class=\"num\">4.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Ethanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/64-17-5\/Ethanol\" target=\"_blank\" rel=\"noopener\">Ethanol<\/a><\/td><td class=\"oth\">5.871<\/td><td class=\"oth\">100<\/td><\/tr>\n<\/tbody><\/table><\/div><div class=\"chromatogram-conditions\"><h4>Conditions<\/h4><div class=\"conditions-container container-fluid\"><div class=\"row\"><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Column<\/th><td>Rtx-1301, 60 m, 0.25 mm ID, 1.00 \u00b5m (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/16056\" rel=\"noopener\">cat.# 16056<\/a>)<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Standard\/Sample<\/th><td>Custom acetaldehyde, methanol, and ethanol standard in water<\/td><\/tr>\n<tr><th class=\"sub conditions_header\" scope=\"row\">Diluent:<\/th><td>Water<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Conc.:<\/th><td>100 \u00b5g\/mL<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Injection<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Vol.:<\/th><td>1 \u00b5L split (split ratio 100:1)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Liner:<\/th><td>Topaz Precision inlet liner with wool, 4.0 mm x 6.3 x 78.5 (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/23305\" rel=\"noopener\">cat.# 23305<\/a>)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Temp.:<\/th><td>250 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Split Vent Flow Rate:<\/th><td>141.4 mL\/min<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\" colspan=\"2\">Oven<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Oven Temp.:<\/th><td>35 &#176;C (hold 2.45 min) to 220 &#176;C at 7 &#176;C\/min (hold 5 min)<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Carrier Gas<\/th><td>He, constant flow<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Flow Rate:<\/th><td>1.4 mL\/min<tr><th class=\"sub conditions_header\" scope=\"row\">Linear Velocity:<\/th><td>30.261 cm\/sec @ 35 \u00b0C<\/td><\/tr><\/table><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Detector<\/th><td>MS<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Mode:<\/th><td>Scan<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Scan Program:<\/th><td><\/td><\/tr><tr><td style=\"text-align:left\" colspan=\"2\"><table class=\"conditions_sim\"><thead><tr><th style=\"padding: 0 5px\">Group<\/th><th style=\"padding: 0 5px\">Start Time<br \/>(min)<\/th><th style=\"padding: 0 5px\">Scan Range<br \/>(amu)<\/th><th style=\"padding: 0 5px\">Scan Rate<br \/>(scans\/sec)<\/th><\/tr><\/thead><tbody><tr><td>1<\/td><td>0<\/td><td>10-400<\/td><td>6.84<\/td><\/tr><\/tbody><\/table><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Transfer Line Temp.:<\/th><td>280 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Analyzer Type:<\/th><td>Quadrupole <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Source Temp.:<\/th><td>230 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Quad Temp.:<\/th><td>150 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Electron Energy:<\/th><td>70 eV<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Solvent Delay Time:<\/th><td>0.0  min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Tune Type:<\/th><td>PFTBA <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Ionization Mode:<\/th><td>EI <\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Instrument<\/th><td>Agilent 7890A GC &amp; 5975C MSD<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Notes<\/th><td>The <i>EZ<\/i>GC method translator was used to establish the MS conditions.<\/td><\/tr><\/table><\/div><\/div><\/div><div class=\"chromatogram-related-skus\"><h4>Related SKUs<\/h4><ul><li>16056<\/li><li>23305<\/li><\/ul><\/div><div class=\"chromatogram-pdf-link\"><a href=\"https:\/\/ez.restek.com\/images\/cgram\/gc_ff1389.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"><svg xmlns=\"http:\/\/www.w3.org\/2000\/svg\" width=\"18\" height=\"18\" viewBox=\"0 0 18 18\"><g data-name=\"Group 2996\"><path data-name=\"Rectangle 1246\" d=\"M0 0h18v18H0z\" style=\"fill: none;\"><\/path><\/g><g data-name=\"Group 2997\"><path data-name=\"Path 729\" d=\"M13.412 11.4v2.017H5.345V11.4H4v2.017a1.349 1.349 0 0 0 1.345 1.345h8.068a1.349 1.349 0 0 0 1.345-1.345V11.4zm-.672-2.694-.948-.948-1.741 1.735V4H8.706v5.493L6.965 7.758l-.948.948 3.361 3.361z\" transform=\"translate(-.437 -.414)\" style=\"fill: rgb(13, 123, 196);\"><\/path><\/g><\/svg>Download PDF<\/a><\/div><\/div><\/div>\n\n\n<figure class=\"wp-block-table\">\n<table style=\"background-color:#E6DEDB; border: none;\">\n<tbody>\n<tr style=\"border: none;\">\n<td style=\"width=75%; border: none;\"><strong>Did You&nbsp;Know?<\/strong><br>Fusel alcohols are higher order (i.e., secondary or tertiary) alcohols, traces of which usually are present in all beers. They are produced through a pathway very similar to the pathway for ethanol, the preferred alcohol for beer. Fusel alcohols contribute a hot, spicy, solvent-like flavor and an alcohol &#8220;burn.&#8221; Small amounts of these components can be desirable in a strong ale or barley wine, but they can be offensive and therefore are unwanted in a Pilsner or other lager. In addition to their influence on flavor, they usually cause low carbonation and poor head retention in bottle-conditioned beers because they are deadly to yeast. Higher fermentation temperatures can produce excessively rapid yeast growth and yeast mutations, which, in turn, stimulate the formation of these components.<\/td>\n<td style=\"width:25%; border: none;\"><img loading=\"lazy\" decoding=\"async\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/teaser-article-FFAR4494-01.jpg\" alt=\"picture of beer bottles\" width=\"250\" height=\"165\" title=\"-\"><\/td><\/tr><\/tbody><\/table><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\">The first phase evaluated for resolution and peak shape was the Rtx-VMS phase in a 30 m x 0.25 mm ID x 1.40 \u00b5m column format (cat.# <a href=\"https:\/\/restek.com\/p\/19915\" data-type=\"link\" data-id=\"https:\/\/restek.com\/p\/19915\" target=\"_blank\" rel=\"noopener\">19915<\/a>). While the separation of water, acetaldehyde, and methanol was adequate (similar to the Rtx-1301 column), the Rtx-VMS column does not resolve 2-methyl-1-butanol (isoamyl alcohol) and 3-methyl-1-butanol (active amyl alcohol) as shown in Figure 2. Even with the oven set to 35 \u00b0C and held isothermally, the amyl alcohols coelute regardless of the Rtx-VMS column dimensions.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\"><strong>Figure 2:<\/strong> Common alcoholic beverage target compounds 2-methyl-1-butanol and 3-methyl-1-butanol are not resolved on an Rtx-VMS column (peaks 12 and 13).<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article\"><div class=\"wp-block-custom-chromatogram-article\"><div class=\"chromatogram-image wide-image\"><img decoding=\"async\" src=\"https:\/\/ez.restek.com\/images\/cgram\/gc_ff1385.png\" alt=\"Custom Beverage Analysis Mix on Rtx-VMS\" title=\"-\"><\/div><p class=\"article-id\" style=\"text-align:center\"> GC_FF1385<\/p><div class=\"chromatogram-peaks\"><h4>Peaks<\/h4><table class=\"peaks col-lg-6 col-12 peak-50\">\n<thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 75px\">t<sub>R<\/sub> (min)<\/th><th style=\"text-align: center;width: 75px\">Conc.<br \/>(\u00b5g\/mL)<\/th><\/tr><\/thead>\n<tbody><tr><td class=\"num\">1.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Acetaldehyde\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/75-07-0\/Acetaldehyde\" target=\"_blank\" rel=\"noopener\">Acetaldehyde<\/a><\/td><td class=\"oth\">2.607<\/td><td class=\"oth\">100<\/td><\/tr>\n<tr><td class=\"num\">2.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Methyl alcohol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/67-56-1\/Methyl alcohol\" target=\"_blank\" rel=\"noopener\">Methyl alcohol<\/a><\/td><td class=\"oth\">2.685<\/td><td class=\"oth\">100<\/td><\/tr>\n<tr><td class=\"num\">3.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Ethanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/64-17-5\/Ethanol\" target=\"_blank\" rel=\"noopener\">Ethanol<\/a><\/td><td class=\"oth\">3.917<\/td><td class=\"oth\"><\/td><\/tr>\n<tr><td class=\"num\">4.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Acetone\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/67-64-1\/Acetone\" target=\"_blank\" rel=\"noopener\">Acetone<\/a><\/td><td class=\"oth\">4.521<\/td><td class=\"oth\">100<\/td><\/tr>\n<tr><td class=\"num\">5.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Ethyl formate\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/109-94-4\/Ethyl formate\" target=\"_blank\" rel=\"noopener\">Ethyl formate<\/a><\/td><td class=\"oth\">4.627<\/td><td class=\"oth\">100<\/td><\/tr>\n<tr><td class=\"num\">6.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 1-Propanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/71-23-8\/1-Propanol\" target=\"_blank\" rel=\"noopener\">1-Propanol<\/a><\/td><td class=\"oth\">6.040<\/td><td class=\"oth\">100<\/td><\/tr>\n<tr><td class=\"num\">7.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Ethyl acetate\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/141-78-6\/Ethyl acetate\" target=\"_blank\" rel=\"noopener\">Ethyl acetate<\/a><\/td><td class=\"oth\">6.874<\/td><td class=\"oth\">100<\/td><\/tr>\n<tr><td class=\"num\">8.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2-Butanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/78-92-2\/2-Butanol\" target=\"_blank\" rel=\"noopener\">2-Butanol<\/a><\/td><td class=\"oth\">6.959<\/td><td class=\"oth\">100<\/td><\/tr>\n<\/tbody><\/table>\n<table class=\"peaks col-lg-6 col-12 peak-50\"><thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 75px\">t<sub>R<\/sub> (min)<\/th><th style=\"text-align: center;width: 75px\">Conc.<br \/>(\u00b5g\/mL)<\/th><\/tr><\/thead><tbody>\n<tr><td class=\"num\">9.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 1-Propanol, 2-methyl-\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/78-83-1\/1-Propanol, 2-methyl-\" target=\"_blank\" rel=\"noopener\">1-Propanol, 2-methyl-<\/a><\/td><td class=\"oth\">7.799<\/td><td class=\"oth\">100<\/td><\/tr>\n<tr><td class=\"num\">10.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Acetic acid\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/64-19-7\/Acetic acid\" target=\"_blank\" rel=\"noopener\">Acetic acid<\/a><\/td><td class=\"oth\">8.497<\/td><td class=\"oth\">100<\/td><\/tr>\n<tr><td class=\"num\">11.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Ethane, 1,1-diethoxy-\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/105-57-7\/Ethane, 1,1-diethoxy-\" target=\"_blank\" rel=\"noopener\">Ethane, 1,1-diethoxy-<\/a><\/td><td class=\"oth\">9.112<\/td><td class=\"oth\">100<\/td><\/tr>\n<tr><td class=\"num\">12.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 3-Methyl-1-butanol \" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/123-51-3\/3-Methyl-1-butanol \" target=\"_blank\" rel=\"noopener\">3-Methyl-1-butanol <\/a><\/td><td class=\"oth\">10.614<\/td><td class=\"oth\">100<\/td><\/tr>\n<tr><td class=\"num\">13.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2-Methyl-1-butanol \" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/137-32-6\/2-Methyl-1-butanol \" target=\"_blank\" rel=\"noopener\">2-Methyl-1-butanol <\/a><\/td><td class=\"oth\">10.632<\/td><td class=\"oth\">100<\/td><\/tr>\n<tr><td class=\"num\">14.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 1-Pentanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/71-41-0\/1-Pentanol\" target=\"_blank\" rel=\"noopener\">1-Pentanol<\/a><\/td><td class=\"oth\">11.328<\/td><td class=\"oth\">100<\/td><\/tr>\n<tr><td class=\"num\">15.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Propylene glycol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/57-55-6\/Propylene glycol\" target=\"_blank\" rel=\"noopener\">Propylene glycol<\/a><\/td><td class=\"oth\">12.375<\/td><td class=\"oth\">100<\/td><\/tr>\n<tr><td class=\"num\">16.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Isoamyl acetate\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/123-92-2\/Isoamyl acetate\" target=\"_blank\" rel=\"noopener\">Isoamyl acetate<\/a><\/td><td class=\"oth\">13.478<\/td><td class=\"oth\">100<\/td><\/tr>\n<tr><td class=\"num\">17.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Furfural\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/98-01-1\/Furfural\" target=\"_blank\" rel=\"noopener\">Furfural<\/a><\/td><td class=\"oth\">13.839<\/td><td class=\"oth\">100<\/td><\/tr>\n<\/tbody><\/table><div class=\"peaknotes\">Final on-column concentration is 1 ppm due to 100:1 split.  Ethanol is the diluent solvent.<\/div><\/div><div class=\"chromatogram-conditions\"><h4>Conditions<\/h4><div class=\"conditions-container container-fluid\"><div class=\"row\"><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Column<\/th><td>Rtx-VMS, 30 m, 0.25 mm ID, 1.40 \u00b5m (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/19915\" rel=\"noopener\">cat.# 19915<\/a>)<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Standard\/Sample<\/th><td>Custom beverage analysis mix<\/td><\/tr>\n<tr><th class=\"sub conditions_header\" scope=\"row\">Diluent:<\/th><td>Ethanol<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Conc.:<\/th><td>100 \u00b5g\/mL Final on-column concentration is 1 ppm.<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Injection<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Vol.:<\/th><td>1 \u00b5L split (split ratio 100:1)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Liner:<\/th><td>Topaz, precision inlet liner, 4.0 mm x 6.3 x 78.5, for Agilent GCs, w\/quartz wool (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/23305\" rel=\"noopener\">cat.# 23305<\/a>)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Temp.:<\/th><td>250 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Split Vent Flow Rate:<\/th><td>144.4 mL\/min<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\" colspan=\"2\">Oven<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Oven Temp.:<\/th><td>35 &#176;C (hold 3 min) to 220 &#176;C at 8 &#176;C\/min (hold 5 min)<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Carrier Gas<\/th><td>He, constant flow<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Flow Rate:<\/th><td>1.4 mL\/min<tr><th class=\"sub conditions_header\" scope=\"row\">Linear Velocity:<\/th><td>31.993 cm\/sec @ 35 \u00b0C<\/td><\/tr><\/table><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Detector<\/th><td>FID @ 250 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Make-up Gas Flow Rate:<\/th><td>25 mL\/min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Make-up Gas Type:<\/th><td>N<sub>2<\/sub> <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Hydrogen flow:<\/th><td>35 mL\/min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Air flow:<\/th><td>400 mL\/min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Data Rate:<\/th><td>50 Hz<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Instrument<\/th><td>Agilent 7890A GC<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Sample Preparation<\/th><td>Single component compounds combined to make final custom mix in ethanol.<\/td><\/tr><\/table><\/div><\/div><\/div><div class=\"chromatogram-related-skus\"><h4>Related SKUs<\/h4><ul><li>19915<\/li><li>23305<\/li><\/ul><\/div><div class=\"chromatogram-pdf-link\"><a href=\"https:\/\/ez.restek.com\/images\/cgram\/gc_ff1385.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"><svg xmlns=\"http:\/\/www.w3.org\/2000\/svg\" width=\"18\" height=\"18\" viewBox=\"0 0 18 18\"><g data-name=\"Group 2996\"><path data-name=\"Rectangle 1246\" d=\"M0 0h18v18H0z\" style=\"fill: none;\"><\/path><\/g><g data-name=\"Group 2997\"><path data-name=\"Path 729\" d=\"M13.412 11.4v2.017H5.345V11.4H4v2.017a1.349 1.349 0 0 0 1.345 1.345h8.068a1.349 1.349 0 0 0 1.345-1.345V11.4zm-.672-2.694-.948-.948-1.741 1.735V4H8.706v5.493L6.965 7.758l-.948.948 3.361 3.361z\" transform=\"translate(-.437 -.414)\" style=\"fill: rgb(13, 123, 196);\"><\/path><\/g><\/svg>Download PDF<\/a><\/div><\/div><\/div>\n\n\n<p class=\"wp-block-paragraph\">The Rtx-502.2 phase in a 60 m x 0.25 mm ID x 1.40 \u00b5m column format (cat.# <a href=\"https:\/\/restek.com\/p\/10916\" target=\"_blank\" rel=\"noopener\">10916<\/a>) offered excellent resolution and an acceptable analysis time for all target compounds. However, the acetic acid peak shape shows fronting, which indicates poor solubility of acetic acid in the low-polarity stationary phase (Figure 3).<\/p>\n\n\n\n<p class=\"wp-block-paragraph\"><strong>Figure 3:<\/strong> Acetic acid exhibits poor peak shape on an Rtx-502.2 column, indicating poor solubility in the stationary phase (peak 7).<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article\"><div class=\"wp-block-custom-chromatogram-article\"><div class=\"chromatogram-image wide-image\"><img decoding=\"async\" src=\"https:\/\/ez.restek.com\/images\/cgram\/gc_ff1387.png\" alt=\"Custom Beverage Analysis Mix on Rtx-502.2\" title=\"-\"><\/div><p class=\"article-id\" style=\"text-align:center\"> GC_FF1387<\/p><div class=\"chromatogram-peaks\"><h4>Peaks<\/h4><table class=\"peaks col-lg-6 col-12 peak-50\">\n<thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 75px\">t<sub>R<\/sub> (min)<\/th><th style=\"text-align: center;width: 75px\">Conc.<br \/>(\u00b5g\/mL)<\/th><\/tr><\/thead>\n<tbody><tr><td class=\"num\">1.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Acetaldehyde\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/75-07-0\/Acetaldehyde\" target=\"_blank\" rel=\"noopener\">Acetaldehyde<\/a><\/td><td class=\"oth\">5.785<\/td><td class=\"oth\">100<\/td><\/tr>\n<tr><td class=\"num\">2.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Methanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/67-56-1\/Methanol\" target=\"_blank\" rel=\"noopener\">Methanol<\/a><\/td><td class=\"oth\">5.508<\/td><td class=\"oth\">100<\/td><\/tr>\n<tr><td class=\"num\">3.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Ethanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/64-17-5\/Ethanol\" target=\"_blank\" rel=\"noopener\">Ethanol<\/a><\/td><td class=\"oth\">6.971<\/td><td class=\"oth\">100<\/td><\/tr>\n<tr><td class=\"num\">4.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Acetone\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/67-64-1\/Acetone\" target=\"_blank\" rel=\"noopener\">Acetone<\/a><\/td><td class=\"oth\">8.103<\/td><td class=\"oth\">100<\/td><\/tr>\n<tr><td class=\"num\">5.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Ethyl formate\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/109-94-4\/Ethyl formate\" target=\"_blank\" rel=\"noopener\">Ethyl formate<\/a><\/td><td class=\"oth\">8.523<\/td><td class=\"oth\">100<\/td><\/tr>\n<tr><td class=\"num\">6.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 1-Propanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/71-23-8\/1-Propanol\" target=\"_blank\" rel=\"noopener\">1-Propanol<\/a><\/td><td class=\"oth\">9.559<\/td><td class=\"oth\">100<\/td><\/tr>\n<tr><td class=\"num\">7.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Acetic acid\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/64-19-7\/Acetic acid\" target=\"_blank\" rel=\"noopener\">Acetic acid<\/a><\/td><td class=\"oth\">10.244<\/td><td class=\"oth\">100<\/td><\/tr>\n<tr><td class=\"num\">8.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2-Butanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/78-92-2\/2-Butanol\" target=\"_blank\" rel=\"noopener\">2-Butanol<\/a><\/td><td class=\"oth\">10.750<\/td><td class=\"oth\">100<\/td><\/tr>\n<\/tbody><\/table>\n<table class=\"peaks col-lg-6 col-12 peak-50\"><thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 75px\">t<sub>R<\/sub> (min)<\/th><th style=\"text-align: center;width: 75px\">Conc.<br \/>(\u00b5g\/mL)<\/th><\/tr><\/thead><tbody>\n<tr><td class=\"num\">9.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Ethyl acetate\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/141-78-6\/Ethyl acetate\" target=\"_blank\" rel=\"noopener\">Ethyl acetate<\/a><\/td><td class=\"oth\">11.318<\/td><td class=\"oth\">100<\/td><\/tr>\n<tr><td class=\"num\">10.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Isobutanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/78-83-1\/Isobutanol\" target=\"_blank\" rel=\"noopener\">Isobutanol<\/a><\/td><td class=\"oth\">11.544<\/td><td class=\"oth\">100<\/td><\/tr>\n<tr><td class=\"num\">11.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Acetal\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/105-57-7\/Acetal\" target=\"_blank\" rel=\"noopener\">Acetal<\/a><\/td><td class=\"oth\">14.260<\/td><td class=\"oth\">100<\/td><\/tr>\n<tr><td class=\"num\">12.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 3-Methyl-1-butanol \" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/123-51-3\/3-Methyl-1-butanol \" target=\"_blank\" rel=\"noopener\">3-Methyl-1-butanol <\/a><\/td><td class=\"oth\">14.730<\/td><td class=\"oth\">100<\/td><\/tr>\n<tr><td class=\"num\">13.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2-Methyl-1-butanol \" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/137-32-6\/2-Methyl-1-butanol \" target=\"_blank\" rel=\"noopener\">2-Methyl-1-butanol <\/a><\/td><td class=\"oth\">14.850<\/td><td class=\"oth\">100<\/td><\/tr>\n<tr><td class=\"num\">14.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Propylene glycol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/57-55-6\/Propylene glycol\" target=\"_blank\" rel=\"noopener\">Propylene glycol<\/a><\/td><td class=\"oth\">15.670<\/td><td class=\"oth\">100<\/td><\/tr>\n<tr><td class=\"num\">15.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for n-Amyl alcohol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/71-41-0\/n-Amyl alcohol\" target=\"_blank\" rel=\"noopener\">n-Amyl alcohol<\/a><\/td><td class=\"oth\">16.286<\/td><td class=\"oth\">100<\/td><\/tr>\n<tr><td class=\"num\">16.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Isoamyl acetate\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/123-92-2\/Isoamyl acetate\" target=\"_blank\" rel=\"noopener\">Isoamyl acetate<\/a><\/td><td class=\"oth\">18.511<\/td><td class=\"oth\">100<\/td><\/tr>\n<tr><td class=\"num\">17.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Furfural\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/98-01-1\/Furfural\" target=\"_blank\" rel=\"noopener\">Furfural<\/a><\/td><td class=\"oth\">18.708<\/td><td class=\"oth\">100<\/td><\/tr>\n<\/tbody><\/table><\/div><div class=\"chromatogram-conditions\"><h4>Conditions<\/h4><div class=\"conditions-container container-fluid\"><div class=\"row\"><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Column<\/th><td>Rtx-502.2, 60 m, 0.25 mm ID, 1.4 \u00b5m (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/10916\" rel=\"noopener\">cat.# 10916<\/a>)<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Standard\/Sample<\/th><td>Custom beverage analysis<\/td><\/tr>\n<tr><th class=\"sub conditions_header\" scope=\"row\">Diluent:<\/th><td>Ethanol<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Conc.:<\/th><td>100 \u00b5g\/mL 100:1 split<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Injection<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Vol.:<\/th><td>1.0 \u00b5L split (split ratio 100:1)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Liner:<\/th><td>Topaz precision split liner (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/23305\" rel=\"noopener\">cat.# 23305<\/a>)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Temp.:<\/th><td>250 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Split Vent Flow Rate:<\/th><td>140 mL\/min<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\" colspan=\"2\">Oven<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Oven Temp.:<\/th><td>35 &#176;C (hold 3 min) to 220 &#176;C at 8 &#176;C\/min (hold 5 min)<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Carrier Gas<\/th><td>He, constant flow<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Flow Rate:<\/th><td>1.4 mL\/min<tr><th class=\"sub conditions_header\" scope=\"row\">Linear Velocity:<\/th><td>25.503 cm\/sec @ 35 \u00b0C<\/td><\/tr><\/table><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Detector<\/th><td>FID @ 250 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Make-up Gas Flow Rate:<\/th><td>25 mL\/min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Make-up Gas Type:<\/th><td>N<sub>2<\/sub> <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Hydrogen flow:<\/th><td>35 mL\/min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Air flow:<\/th><td>400 mL\/min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Data Rate:<\/th><td>50 Hz<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Instrument<\/th><td>Agilent 7890A GC<\/td><\/tr><\/table><\/div><\/div><\/div><div class=\"chromatogram-related-skus\"><h4>Related SKUs<\/h4><ul><li>10916<\/li><li>23305<\/li><\/ul><\/div><div class=\"chromatogram-pdf-link\"><a href=\"https:\/\/ez.restek.com\/images\/cgram\/gc_ff1387.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"><svg xmlns=\"http:\/\/www.w3.org\/2000\/svg\" width=\"18\" height=\"18\" viewBox=\"0 0 18 18\"><g data-name=\"Group 2996\"><path data-name=\"Rectangle 1246\" d=\"M0 0h18v18H0z\" style=\"fill: none;\"><\/path><\/g><g data-name=\"Group 2997\"><path data-name=\"Path 729\" d=\"M13.412 11.4v2.017H5.345V11.4H4v2.017a1.349 1.349 0 0 0 1.345 1.345h8.068a1.349 1.349 0 0 0 1.345-1.345V11.4zm-.672-2.694-.948-.948-1.741 1.735V4H8.706v5.493L6.965 7.758l-.948.948 3.361 3.361z\" transform=\"translate(-.437 -.414)\" style=\"fill: rgb(13, 123, 196);\"><\/path><\/g><\/svg>Download PDF<\/a><\/div><\/div><\/div>\n\n\n<p class=\"wp-block-paragraph\">The last column evaluated was an Rtx-1301 phase in a 60 m x 0.25 mm ID x 1.00 \u00b5m format (cat.# <a href=\"https:\/\/restek.com\/p\/16056\" target=\"_blank\" rel=\"noopener\">16056<\/a>). This column offered the best overall chromatography, particularly in terms of peak shape and resolution of the critical pairs. Importantly, the 2-methyl-1-butanol and 3-methyl-1-butanol peaks are adequately resolved. In addition, by using the higher polarity Rtx-1301 stationary phase (a cyanopropylphenyl-type polysiloxane phase) compared to the lower polarity Rtx-502.2 (a phenylmethyl-type polysiloxane phase), the acetic acid peak shape was much improved (Figure 4). This occurs because acetic acid is more soluble in the Rtx-1301 phase. Because the Rtx-1301 column produced the best overall chromatography in less than 20 minutes, it is the recommended column for GC analysis of alcoholic beverages.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\"><strong>Figure 4:<\/strong> The Rtx-1301 column shows the best overall chromatography for targeted alcoholic beverage compounds.<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article\"><div class=\"wp-block-custom-chromatogram-article\"><div class=\"chromatogram-image wide-image\"><img decoding=\"async\" src=\"https:\/\/ez.restek.com\/images\/cgram\/gc_ff1388.png\" alt=\"Custom Beverage Analysis Mix on Rtx-1301\" title=\"-\"><\/div><p class=\"article-id\" style=\"text-align:center\"> GC_FF1388<\/p><div class=\"chromatogram-peaks\"><h4>Peaks<\/h4><table class=\"peaks col-lg-6 col-12 peak-50\">\n<thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 75px\">t<sub>R<\/sub> (min)<\/th><th style=\"text-align: center;width: 75px\">Conc.<br \/>(\u00b5g\/mL)<\/th><\/tr><\/thead>\n<tbody><tr><td class=\"num\">1.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Acetaldehyde\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/75-07-0\/Acetaldehyde\" target=\"_blank\" rel=\"noopener\">Acetaldehyde<\/a><\/td><td class=\"oth\">5.230<\/td><td class=\"oth\">100<\/td><\/tr>\n<tr><td class=\"num\">2.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Methyl alcohol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/67-56-1\/Methyl alcohol\" target=\"_blank\" rel=\"noopener\">Methyl alcohol<\/a><\/td><td class=\"oth\">5.431<\/td><td class=\"oth\">100<\/td><\/tr>\n<tr><td class=\"num\">3.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Ethanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/64-17-5\/Ethanol\" target=\"_blank\" rel=\"noopener\">Ethanol<\/a><\/td><td class=\"oth\">6.911<\/td><td class=\"oth\">100<\/td><\/tr>\n<tr><td class=\"num\">4.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Acetone\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/67-64-1\/Acetone\" target=\"_blank\" rel=\"noopener\">Acetone<\/a><\/td><td class=\"oth\">7.476<\/td><td class=\"oth\">100<\/td><\/tr>\n<tr><td class=\"num\">5.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Ethyl formate\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/109-94-4\/Ethyl formate\" target=\"_blank\" rel=\"noopener\">Ethyl formate<\/a><\/td><td class=\"oth\">7.674<\/td><td class=\"oth\">100<\/td><\/tr>\n<tr><td class=\"num\">6.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 1-Propanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/71-23-8\/1-Propanol\" target=\"_blank\" rel=\"noopener\">1-Propanol<\/a><\/td><td class=\"oth\">9.517<\/td><td class=\"oth\">100<\/td><\/tr>\n<tr><td class=\"num\">7.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Ethyl acetate\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/141-78-6\/Ethyl acetate\" target=\"_blank\" rel=\"noopener\">Ethyl acetate<\/a><\/td><td class=\"oth\">10.400<\/td><td class=\"oth\">100<\/td><\/tr>\n<tr><td class=\"num\">8.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2-Butanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/78-92-2\/2-Butanol\" target=\"_blank\" rel=\"noopener\">2-Butanol<\/a><\/td><td class=\"oth\">10.620<\/td><td class=\"oth\">100<\/td><\/tr>\n<\/tbody><\/table>\n<table class=\"peaks col-lg-6 col-12 peak-50\"><thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 75px\">t<sub>R<\/sub> (min)<\/th><th style=\"text-align: center;width: 75px\">Conc.<br \/>(\u00b5g\/mL)<\/th><\/tr><\/thead><tbody>\n<tr><td class=\"num\">9.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Isobutanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/78-83-1\/Isobutanol\" target=\"_blank\" rel=\"noopener\">Isobutanol<\/a><\/td><td class=\"oth\">11.590<\/td><td class=\"oth\">100<\/td><\/tr>\n<tr><td class=\"num\">10.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Acetic acid\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/64-19-7\/Acetic acid\" target=\"_blank\" rel=\"noopener\">Acetic acid<\/a><\/td><td class=\"oth\">11.753<\/td><td class=\"oth\">100<\/td><\/tr>\n<tr><td class=\"num\">11.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Acetal\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/105-57-7\/Acetal\" target=\"_blank\" rel=\"noopener\">Acetal<\/a><\/td><td class=\"oth\">13.517<\/td><td class=\"oth\">100<\/td><\/tr>\n<tr><td class=\"num\">12.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 3-Methyl-1-butanol \" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/123-51-3\/3-Methyl-1-butanol \" target=\"_blank\" rel=\"noopener\">3-Methyl-1-butanol <\/a><\/td><td class=\"oth\">14.884<\/td><td class=\"oth\">100<\/td><\/tr>\n<tr><td class=\"num\">13.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2-Methyl-1-butanol \" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/137-32-6\/2-Methyl-1-butanol \" target=\"_blank\" rel=\"noopener\">2-Methyl-1-butanol <\/a><\/td><td class=\"oth\">14.975<\/td><td class=\"oth\">100<\/td><\/tr>\n<tr><td class=\"num\">14.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 1-Pentanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/71-41-0\/1-Pentanol\" target=\"_blank\" rel=\"noopener\">1-Pentanol<\/a><\/td><td class=\"oth\">15.802<\/td><td class=\"oth\">100<\/td><\/tr>\n<tr><td class=\"num\">15.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Propylene glycol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/57-55-6\/Propylene glycol\" target=\"_blank\" rel=\"noopener\">Propylene glycol<\/a><\/td><td class=\"oth\">16.490<\/td><td class=\"oth\">100<\/td><\/tr>\n<tr><td class=\"num\">16.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Isoamyl acetate\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/123-92-2\/Isoamyl acetate\" target=\"_blank\" rel=\"noopener\">Isoamyl acetate<\/a><\/td><td class=\"oth\">18.246<\/td><td class=\"oth\">100<\/td><\/tr>\n<tr><td class=\"num\">17.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Furfural\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/98-01-1\/Furfural\" target=\"_blank\" rel=\"noopener\">Furfural<\/a><\/td><td class=\"oth\">18.363<\/td><td class=\"oth\">100<\/td><\/tr>\n<\/tbody><\/table><\/div><div class=\"chromatogram-conditions\"><h4>Conditions<\/h4><div class=\"conditions-container container-fluid\"><div class=\"row\"><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Column<\/th><td>Rtx-1301, 60 m, 0.25 mm ID, 1.0 \u00b5m (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/16056\" rel=\"noopener\">cat.# 16056<\/a>)<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Standard\/Sample<\/th><td>Custom beverage analysis standard<\/td><\/tr>\n<tr><th class=\"sub conditions_header\" scope=\"row\">Diluent:<\/th><td>Ethanol<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Conc.:<\/th><td>100 \u00b5g\/mL<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Injection<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Vol.:<\/th><td>1 \u00b5L split (split ratio 100:1)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Liner:<\/th><td>Topaz precision split liner (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/23305\" rel=\"noopener\">cat.# 23305<\/a>)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Temp.:<\/th><td>250 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Split Vent Flow Rate:<\/th><td>140 mL\/min<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\" colspan=\"2\">Oven<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Oven Temp.:<\/th><td>35 &#176;C (hold 3 min) to 220 &#176;C at 7 &#176;C\/min (hold 5 min)<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Carrier Gas<\/th><td>He, constant flow<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Flow Rate:<\/th><td>1.4 mL\/min<tr><th class=\"sub conditions_header\" scope=\"row\">Linear Velocity:<\/th><td>25.457 cm\/sec @ 35 \u00b0C<\/td><\/tr><\/table><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Detector<\/th><td>FID @ 250 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Make-up Gas Flow Rate:<\/th><td>25 mL\/min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Make-up Gas Type:<\/th><td>N<sub>2<\/sub> <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Hydrogen flow:<\/th><td>35 mL\/min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Air flow:<\/th><td>400 mL\/min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Data Rate:<\/th><td>50 Hz<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Instrument<\/th><td>Agilent 7890A GC<\/td><\/tr><\/table><\/div><\/div><\/div><div class=\"chromatogram-related-skus\"><h4>Related SKUs<\/h4><ul><li>16056<\/li><li>23305<\/li><\/ul><\/div><div class=\"chromatogram-pdf-link\"><a href=\"https:\/\/ez.restek.com\/images\/cgram\/gc_ff1388.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"><svg xmlns=\"http:\/\/www.w3.org\/2000\/svg\" width=\"18\" height=\"18\" viewBox=\"0 0 18 18\"><g data-name=\"Group 2996\"><path data-name=\"Rectangle 1246\" d=\"M0 0h18v18H0z\" style=\"fill: none;\"><\/path><\/g><g data-name=\"Group 2997\"><path data-name=\"Path 729\" d=\"M13.412 11.4v2.017H5.345V11.4H4v2.017a1.349 1.349 0 0 0 1.345 1.345h8.068a1.349 1.349 0 0 0 1.345-1.345V11.4zm-.672-2.694-.948-.948-1.741 1.735V4H8.706v5.493L6.965 7.758l-.948.948 3.361 3.361z\" transform=\"translate(-.437 -.414)\" style=\"fill: rgb(13, 123, 196);\"><\/path><\/g><\/svg>Download PDF<\/a><\/div><\/div><\/div>\n\n\n<h2 class=\"wp-block-heading\">Alcoholic Beverage Sample Analysis<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">To confirm the performance of the Rtx-1301 column for targeted alcoholic beverage compounds, real-world alcohol samples were analyzed. Data was collected with both an FID and an MS detector with the method initially developed by GC-FID and then transferred to GC-MS. To ease this transition, Restek\u2019s <em>EZ<\/em>GC method translator was used. The translation provided by this tool proved to be an excellent starting point for the MS method\u2014it was generated in seconds and needed only minor optimization and adjustment of the temperature program to ensure that isobutanol and acetic acid were adequately resolved. For example, Figures 5 and 6 show the same chromatography for rye whiskey analysis by GC-FID and GC-MS, respectively; note the excellent resolution between the isobutanol and acetic acid as well as the separation of the 2-methyl-1-butanol and 3-methyl-1-butanol peaks. The acetic acid peak shape is satisfactory, and the analysis time is less than 15 minutes.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\"><strong>Figure 5:<\/strong> Rye whiskey by GC-FID on an Rtx-1301 column exhibits excellent resolution, satisfactory acetic acid peak shape, and a fast &lt;15-minute analysis time.<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article\"><div class=\"wp-block-custom-chromatogram-article\"><div class=\"chromatogram-image wide-image\"><img decoding=\"async\" src=\"https:\/\/ez.restek.com\/images\/cgram\/gc_ff1395.png\" alt=\"Rye Whiskey on Rtx-1301\" title=\"-\"><\/div><p class=\"article-id\" style=\"text-align:center\"> GC_FF1395<\/p><div class=\"chromatogram-peaks\"><h4>Peaks<\/h4><table class=\"peaks col-lg-4 col-12\">\n<thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 50px\">t<sub>R<\/sub> (min)<\/th><\/tr><\/thead>\n<tbody><tr><td class=\"num\">1.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Acetaldehyde\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/75-07-0\/Acetaldehyde\" target=\"_blank\" rel=\"noopener\">Acetaldehyde<\/a><\/td><td class=\"oth\">5.405<\/td><\/tr>\n<tr><td class=\"num\">2.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Methanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/67-56-1\/Methanol\" target=\"_blank\" rel=\"noopener\">Methanol<\/a><\/td><td class=\"oth\">5.631<\/td><\/tr>\n<tr><td class=\"num\">3.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Ethanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/64-17-5\/Ethanol\" target=\"_blank\" rel=\"noopener\">Ethanol<\/a><\/td><td class=\"oth\">7.032<\/td><\/tr>\n<tr><td class=\"num\">4.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 1-Propanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/71-23-8\/1-Propanol\" target=\"_blank\" rel=\"noopener\">1-Propanol<\/a><\/td><td class=\"oth\">9.489<\/td><\/tr>\n<tr><td class=\"num\">5.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Ethyl acetate\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/141-78-6\/Ethyl acetate\" target=\"_blank\" rel=\"noopener\">Ethyl acetate<\/a><\/td><td class=\"oth\">10.318<\/td><\/tr>\n<tr><td class=\"num\">6.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Isobutanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/78-83-1\/Isobutanol\" target=\"_blank\" rel=\"noopener\">Isobutanol<\/a><\/td><td class=\"oth\">11.385<\/td><\/tr>\n<tr><td class=\"num\">7.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Acetic acid\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/64-19-7\/Acetic acid\" target=\"_blank\" rel=\"noopener\">Acetic acid<\/a><\/td><td class=\"oth\">11.472<\/td><\/tr>\n<tr><td class=\"num\">8.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Acetal\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/105-57-7\/Acetal\" target=\"_blank\" rel=\"noopener\">Acetal<\/a><\/td><td class=\"oth\">13.137<\/td><\/tr>\n<tr><td class=\"num\">9.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 3-Methyl-1-butanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/123-51-3\/3-Methyl-1-butanol\" target=\"_blank\" rel=\"noopener\">3-Methyl-1-butanol<\/a><\/td><td class=\"oth\">14.334<\/td><\/tr>\n<tr><td class=\"num\">10.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2-Methyl-1-butanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/137-32-6\/2-Methyl-1-butanol\" target=\"_blank\" rel=\"noopener\">2-Methyl-1-butanol<\/a><\/td><td class=\"oth\">14.419<\/td><\/tr>\n<\/tbody><\/table><\/div><div class=\"chromatogram-conditions\"><h4>Conditions<\/h4><div class=\"conditions-container container-fluid\"><div class=\"row\"><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Column<\/th><td>Rtx-1301, 60 m, 0.25 mm ID, 1.0 \u00b5m (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/16056\" rel=\"noopener\">cat.# 16056<\/a>)<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Standard\/Sample<\/th><td>Rye whiskey (neat)<\/td><\/tr>\n<tr><th class=\"sub conditions_header\" scope=\"row\">Diluent:<\/th><td>n\/a<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Injection<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Vol.:<\/th><td>1 \u00b5L split (split ratio 100:1)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Liner:<\/th><td>Topaz precision inlet liner, 4.0 mm x 6.3 x 78.5, with wool (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/23305\" rel=\"noopener\">cat.# 23305<\/a>)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Temp.:<\/th><td>250 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Split Vent Flow Rate:<\/th><td>144.4 mL\/min<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\" colspan=\"2\">Oven<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Oven Temp.:<\/th><td>35 &#176;C (hold 3 min) to 220 &#176;C at 8 &#176;C\/min (hold 5 min)<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Carrier Gas<\/th><td>He, constant flow<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Flow Rate:<\/th><td>1.4 mL\/min<tr><th class=\"sub conditions_header\" scope=\"row\">Linear Velocity:<\/th><td>25.457 cm\/sec @ 35 \u00b0C<\/td><\/tr><\/table><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Detector<\/th><td>FID @ 250 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Make-up Gas Flow Rate:<\/th><td>25 mL\/min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Make-up Gas Type:<\/th><td>N<sub>2<\/sub> <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Hydrogen flow:<\/th><td>35 mL\/min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Air flow:<\/th><td>400 mL\/min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Data Rate:<\/th><td>50 Hz<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Instrument<\/th><td>Agilent 7890A GC<\/td><\/tr><\/table><\/div><\/div><\/div><div class=\"chromatogram-related-skus\"><h4>Related SKUs<\/h4><ul><li>16056<\/li><li>23305<\/li><\/ul><\/div><div class=\"chromatogram-pdf-link\"><a href=\"https:\/\/ez.restek.com\/images\/cgram\/gc_ff1395.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"><svg xmlns=\"http:\/\/www.w3.org\/2000\/svg\" width=\"18\" height=\"18\" viewBox=\"0 0 18 18\"><g data-name=\"Group 2996\"><path data-name=\"Rectangle 1246\" d=\"M0 0h18v18H0z\" style=\"fill: none;\"><\/path><\/g><g data-name=\"Group 2997\"><path data-name=\"Path 729\" d=\"M13.412 11.4v2.017H5.345V11.4H4v2.017a1.349 1.349 0 0 0 1.345 1.345h8.068a1.349 1.349 0 0 0 1.345-1.345V11.4zm-.672-2.694-.948-.948-1.741 1.735V4H8.706v5.493L6.965 7.758l-.948.948 3.361 3.361z\" transform=\"translate(-.437 -.414)\" style=\"fill: rgb(13, 123, 196);\"><\/path><\/g><\/svg>Download PDF<\/a><\/div><\/div><\/div>\n\n\n<p class=\"wp-block-paragraph\"><strong>Figure 6:<\/strong> GC-MS analysis of rye whiskey shows the same chromatography as the GC-FID analysis in Figure 5.<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article\"><div class=\"wp-block-custom-chromatogram-article\"><div class=\"chromatogram-image wide-image\"><img decoding=\"async\" src=\"https:\/\/ez.restek.com\/images\/cgram\/gc_ff1396.png\" alt=\"Rye Whiskey on Rtx-1301\" title=\"-\"><\/div><p class=\"article-id\" style=\"text-align:center\"> GC_FF1396<\/p><div class=\"chromatogram-peaks\"><h4>Peaks<\/h4><table class=\"peaks col-lg-4 col-12\">\n<thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 50px\">t<sub>R<\/sub> (min)<\/th><\/tr><\/thead>\n<tbody><tr><td class=\"num\">1.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Acetaldehyde\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/75-07-0\/Acetaldehyde\" target=\"_blank\" rel=\"noopener\">Acetaldehyde<\/a><\/td><td class=\"oth\">4.493<\/td><\/tr>\n<tr><td class=\"num\">2.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Methanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/67-56-1\/Methanol\" target=\"_blank\" rel=\"noopener\">Methanol<\/a><\/td><td class=\"oth\">4.684<\/td><\/tr>\n<tr><td class=\"num\">3.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Ethanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/64-17-5\/Ethanol\" target=\"_blank\" rel=\"noopener\">Ethanol<\/a><\/td><td class=\"oth\">5.988<\/td><\/tr>\n<tr><td class=\"num\">4.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 1-Propanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/71-23-8\/1-Propanol\" target=\"_blank\" rel=\"noopener\">1-Propanol<\/a><\/td><td class=\"oth\">8.392<\/td><\/tr>\n<tr><td class=\"num\">5.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Ethyl acetate\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/141-78-6\/Ethyl acetate\" target=\"_blank\" rel=\"noopener\">Ethyl acetate<\/a><\/td><td class=\"oth\">9.234<\/td><\/tr>\n<tr><td class=\"num\">6.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Isobutanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/78-83-1\/Isobutanol\" target=\"_blank\" rel=\"noopener\">Isobutanol<\/a><\/td><td class=\"oth\">10.373<\/td><\/tr>\n<tr><td class=\"num\">7.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Acetic acid\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/64-19-7\/Acetic acid\" target=\"_blank\" rel=\"noopener\">Acetic acid<\/a><\/td><td class=\"oth\">10.521<\/td><\/tr>\n<tr><td class=\"num\">8.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Acetal\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/105-57-7\/Acetal\" target=\"_blank\" rel=\"noopener\">Acetal<\/a><\/td><td class=\"oth\">12.265<\/td><\/tr>\n<tr><td class=\"num\">9.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 3-Methyl-1-butanol \" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/123-51-3\/3-Methyl-1-butanol \" target=\"_blank\" rel=\"noopener\">3-Methyl-1-butanol <\/a><\/td><td class=\"oth\">13.595<\/td><\/tr>\n<tr><td class=\"num\">10.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2-Methyl-1-butanol \" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/137-32-6\/2-Methyl-1-butanol \" target=\"_blank\" rel=\"noopener\">2-Methyl-1-butanol <\/a><\/td><td class=\"oth\">13.667<\/td><\/tr>\n<\/tbody><\/table><\/div><div class=\"chromatogram-conditions\"><h4>Conditions<\/h4><div class=\"conditions-container container-fluid\"><div class=\"row\"><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Column<\/th><td>Rtx-1301, 60 m, 0.25 mm ID, 1.0 \u00b5m (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/16056\" rel=\"noopener\">cat.# 16056<\/a>)<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Standard\/Sample<\/th><td>Rye whiskey<\/td><\/tr>\n<tr><th class=\"sub conditions_header\" scope=\"row\">Diluent:<\/th><td>n\/a<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Injection<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Vol.:<\/th><td>1 \u00b5L split (split ratio 100:1)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Liner:<\/th><td>Topaz precision inlet liner w\/wool, 4.0 mm x 6.3 x 78.5 (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/23305\" rel=\"noopener\">cat.# 23305<\/a>)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Temp.:<\/th><td>250 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Split Vent Flow Rate:<\/th><td>140 mL\/min<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\" colspan=\"2\">Oven<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Oven Temp.:<\/th><td>35 &#176;C (hold 2.45 min) to 220 &#176;C at 7 &#176;C\/min (hold 5 min)<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Carrier Gas<\/th><td>He, constant flow<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Flow Rate:<\/th><td>1.4 mL\/min<tr><th class=\"sub conditions_header\" scope=\"row\">Linear Velocity:<\/th><td>30.261 cm\/sec @ 35 \u00b0C<\/td><\/tr><\/table><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Detector<\/th><td>MS<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Mode:<\/th><td>Scan<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Transfer Line Temp.:<\/th><td>280 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Analyzer Type:<\/th><td>Quadrupole <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Source Temp.:<\/th><td>230 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Quad Temp.:<\/th><td>150 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Electron Energy:<\/th><td>70 eV<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Solvent Delay Time:<\/th><td>0.0  min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Tune Type:<\/th><td>PFTBA <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Ionization Mode:<\/th><td>EI <\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Instrument<\/th><td>Agilent 7890A GC &amp; 5975C MSD<\/td><\/tr><\/table><\/div><\/div><\/div><div class=\"chromatogram-related-skus\"><h4>Related SKUs<\/h4><ul><li>16056<\/li><li>23305<\/li><\/ul><\/div><div class=\"chromatogram-pdf-link\"><a href=\"https:\/\/ez.restek.com\/images\/cgram\/gc_ff1396.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"><svg xmlns=\"http:\/\/www.w3.org\/2000\/svg\" width=\"18\" height=\"18\" viewBox=\"0 0 18 18\"><g data-name=\"Group 2996\"><path data-name=\"Rectangle 1246\" d=\"M0 0h18v18H0z\" style=\"fill: none;\"><\/path><\/g><g data-name=\"Group 2997\"><path data-name=\"Path 729\" d=\"M13.412 11.4v2.017H5.345V11.4H4v2.017a1.349 1.349 0 0 0 1.345 1.345h8.068a1.349 1.349 0 0 0 1.345-1.345V11.4zm-.672-2.694-.948-.948-1.741 1.735V4H8.706v5.493L6.965 7.758l-.948.948 3.361 3.361z\" transform=\"translate(-.437 -.414)\" style=\"fill: rgb(13, 123, 196);\"><\/path><\/g><\/svg>Download PDF<\/a><\/div><\/div><\/div>\n\n\n<p class=\"wp-block-paragraph\">Additives are commonly encountered when analyzing liquors. In fact, up to 2% of any spirit can be unreported and not disclosed on the label, so unlabeled additives are frequently present. For example, propylene glycol and glycerin are sometimes added to liquor. Rum may have propylene glycol added to give the product a mouthfeel similar to an aged rum. Similarly, glycerin may be added for smoothness. Some of these compounds appeared in the range of liquors that were analyzed in Figures 7-11. Note that for these real-world samples, the Rtx-1301 column again provided excellent resolution of critical compounds, good peak shape for acetic acid, and fast analysis times, making it an outstanding column choice for alcoholic beverage analysis by GC.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\"><strong>Figure 7:<\/strong> Brandy analysis using an Rtx-1301 column and GC-FID shows excellent resolution.<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article\"><div class=\"wp-block-custom-chromatogram-article\"><div class=\"chromatogram-image wide-image\"><img decoding=\"async\" src=\"https:\/\/ez.restek.com\/images\/cgram\/gc_ff1391.png\" alt=\"Brandy on Rtx-1301\" title=\"-\"><\/div><p class=\"article-id\" style=\"text-align:center\"> GC_FF1391<\/p><div class=\"chromatogram-peaks\"><h4>Peaks<\/h4><table class=\"peaks col-lg-4 col-12\">\n<thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 50px\">t<sub>R<\/sub> (min)<\/th><\/tr><\/thead>\n<tbody><tr><td class=\"num\">1.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Acetaldehyde\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/75-07-0\/Acetaldehyde\" target=\"_blank\" rel=\"noopener\">Acetaldehyde<\/a><\/td><td class=\"oth\">5.449<\/td><\/tr>\n<tr><td class=\"num\">2.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Methanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/67-56-1\/Methanol\" target=\"_blank\" rel=\"noopener\">Methanol<\/a><\/td><td class=\"oth\">5.634<\/td><\/tr>\n<tr><td class=\"num\">3.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Ethanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/64-17-5\/Ethanol\" target=\"_blank\" rel=\"noopener\">Ethanol<\/a><\/td><td class=\"oth\">7.028<\/td><\/tr>\n<tr><td class=\"num\">4.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 1-Propanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/71-23-8\/1-Propanol\" target=\"_blank\" rel=\"noopener\">1-Propanol<\/a><\/td><td class=\"oth\">9.495<\/td><\/tr>\n<tr><td class=\"num\">5.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Ethyl acetate\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/141-78-6\/Ethyl acetate\" target=\"_blank\" rel=\"noopener\">Ethyl acetate<\/a><\/td><td class=\"oth\">10.321<\/td><\/tr>\n<tr><td class=\"num\">6.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Isobutanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/78-83-1\/Isobutanol\" target=\"_blank\" rel=\"noopener\">Isobutanol<\/a><\/td><td class=\"oth\">11.387<\/td><\/tr>\n<tr><td class=\"num\">7.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Acetal\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/105-57-7\/Acetal\" target=\"_blank\" rel=\"noopener\">Acetal<\/a><\/td><td class=\"oth\">13.137<\/td><\/tr>\n<tr><td class=\"num\">8.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 3-Methyl-1-butanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/123-51-3\/3-Methyl-1-butanol\" target=\"_blank\" rel=\"noopener\">3-Methyl-1-butanol<\/a><\/td><td class=\"oth\">14.333<\/td><\/tr>\n<tr><td class=\"num\">9.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2-Methyl-1-butanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/137-32-6\/2-Methyl-1-butanol\" target=\"_blank\" rel=\"noopener\">2-Methyl-1-butanol<\/a><\/td><td class=\"oth\">14.420<\/td><\/tr>\n<\/tbody><\/table><\/div><div class=\"chromatogram-conditions\"><h4>Conditions<\/h4><div class=\"conditions-container container-fluid\"><div class=\"row\"><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Column<\/th><td>Rtx-1301, 60 m, 0.25 mm ID, 1.0 \u00b5m (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/16056\" rel=\"noopener\">cat.# 16056<\/a>)<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Standard\/Sample<\/th><td>Brandy (neat)<\/td><\/tr>\n<tr><th class=\"sub conditions_header\" scope=\"row\">Diluent:<\/th><td>n\/a<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Injection<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Vol.:<\/th><td>1 \u00b5L split (split ratio 100:1)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Liner:<\/th><td>Topaz precision inlet liner, 4.0 mm x 6.3 x 78.5, with wool (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/23305\" rel=\"noopener\">cat.# 23305<\/a>)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Temp.:<\/th><td>250 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Split Vent Flow Rate:<\/th><td>144.4 mL\/min<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\" colspan=\"2\">Oven<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Oven Temp.:<\/th><td>35 &#176;C (hold 3 min) to 220 &#176;C at 8 &#176;C\/min (hold 5 min)<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Carrier Gas<\/th><td>He, constant flow<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Flow Rate:<\/th><td>1.4 mL\/min<tr><th class=\"sub conditions_header\" scope=\"row\">Linear Velocity:<\/th><td>25.457 cm\/sec @ 35 \u00b0C<\/td><\/tr><\/table><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Detector<\/th><td>FID @ 250 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Make-up Gas Flow Rate:<\/th><td>25 mL\/min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Make-up Gas Type:<\/th><td>N<sub>2<\/sub> <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Hydrogen flow:<\/th><td>35 mL\/min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Air flow:<\/th><td>400 mL\/min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Data Rate:<\/th><td>50 Hz<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Instrument<\/th><td>Agilent 7890A GC<\/td><\/tr><\/table><\/div><\/div><\/div><div class=\"chromatogram-related-skus\"><h4>Related SKUs<\/h4><ul><li>16056<\/li><li>23305<\/li><\/ul><\/div><div class=\"chromatogram-pdf-link\"><a href=\"https:\/\/ez.restek.com\/images\/cgram\/gc_ff1391.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"><svg xmlns=\"http:\/\/www.w3.org\/2000\/svg\" width=\"18\" height=\"18\" viewBox=\"0 0 18 18\"><g data-name=\"Group 2996\"><path data-name=\"Rectangle 1246\" d=\"M0 0h18v18H0z\" style=\"fill: none;\"><\/path><\/g><g data-name=\"Group 2997\"><path data-name=\"Path 729\" d=\"M13.412 11.4v2.017H5.345V11.4H4v2.017a1.349 1.349 0 0 0 1.345 1.345h8.068a1.349 1.349 0 0 0 1.345-1.345V11.4zm-.672-2.694-.948-.948-1.741 1.735V4H8.706v5.493L6.965 7.758l-.948.948 3.361 3.361z\" transform=\"translate(-.437 -.414)\" style=\"fill: rgb(13, 123, 196);\"><\/path><\/g><\/svg>Download PDF<\/a><\/div><\/div><\/div>\n\n\n<p class=\"wp-block-paragraph\"><strong>Figure 8:<\/strong> Sherry analysis using an Rtx-1301 column and GC-FID not only shows excellent resolution but also exhibits good peak shape for the additive propylene glycol.<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article\"><div class=\"wp-block-custom-chromatogram-article\"><div class=\"chromatogram-image wide-image\"><img decoding=\"async\" src=\"https:\/\/ez.restek.com\/images\/cgram\/gc_ff1392.png\" alt=\"Sherry on Rtx-1301\" title=\"-\"><\/div><p class=\"article-id\" style=\"text-align:center\"> GC_FF1392<\/p><div class=\"chromatogram-peaks\"><h4>Peaks<\/h4><table class=\"peaks col-lg-4 col-12\">\n<thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 50px\">t<sub>R<\/sub> (min)<\/th><\/tr><\/thead>\n<tbody><tr><td class=\"num\">1.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Acetaldehyde\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/75-07-0\/Acetaldehyde\" target=\"_blank\" rel=\"noopener\">Acetaldehyde<\/a><\/td><td class=\"oth\">5.405<\/td><\/tr>\n<tr><td class=\"num\">2.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Methanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/67-56-1\/Methanol\" target=\"_blank\" rel=\"noopener\">Methanol<\/a><\/td><td class=\"oth\">5.636<\/td><\/tr>\n<tr><td class=\"num\">3.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Ethanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/64-17-5\/Ethanol\" target=\"_blank\" rel=\"noopener\">Ethanol<\/a><\/td><td class=\"oth\">6.963<\/td><\/tr>\n<tr><td class=\"num\">4.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 1-Propanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/71-23-8\/1-Propanol\" target=\"_blank\" rel=\"noopener\">1-Propanol<\/a><\/td><td class=\"oth\">9.490<\/td><\/tr>\n<tr><td class=\"num\">5.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Ethyl acetate\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/141-78-6\/Ethyl acetate\" target=\"_blank\" rel=\"noopener\">Ethyl acetate<\/a><\/td><td class=\"oth\">10.319<\/td><\/tr>\n<tr><td class=\"num\">6.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Isobutanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/78-83-1\/Isobutanol\" target=\"_blank\" rel=\"noopener\">Isobutanol<\/a><\/td><td class=\"oth\">11.385<\/td><\/tr>\n<tr><td class=\"num\">7.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 3-Methyl-1-butanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/123-51-3\/3-Methyl-1-butanol\" target=\"_blank\" rel=\"noopener\">3-Methyl-1-butanol<\/a><\/td><td class=\"oth\">14.333<\/td><\/tr>\n<tr><td class=\"num\">8.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2-Methyl-1-butanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/137-32-6\/2-Methyl-1-butanol\" target=\"_blank\" rel=\"noopener\">2-Methyl-1-butanol<\/a><\/td><td class=\"oth\">14.419<\/td><\/tr>\n<tr><td class=\"num\">9.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Propylene glycol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/57-55-6\/Propylene glycol\" target=\"_blank\" rel=\"noopener\">Propylene glycol<\/a><\/td><td class=\"oth\">15.752<\/td><\/tr>\n<tr><td class=\"num\">10.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Furfural\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/98-01-1\/Furfural\" target=\"_blank\" rel=\"noopener\">Furfural<\/a><\/td><td class=\"oth\">17.486<\/td><\/tr>\n<\/tbody><\/table><\/div><div class=\"chromatogram-conditions\"><h4>Conditions<\/h4><div class=\"conditions-container container-fluid\"><div class=\"row\"><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Column<\/th><td>Rtx-1301, 60 m, 0.25 mm ID, 1.0 \u00b5m (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/16056\" rel=\"noopener\">cat.# 16056<\/a>)<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Standard\/Sample<\/th><td>Sherry (neat)<\/td><\/tr>\n<tr><th class=\"sub conditions_header\" scope=\"row\">Diluent:<\/th><td>n\/a<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Injection<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Vol.:<\/th><td>1 \u00b5L split (split ratio 100:1)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Liner:<\/th><td>Topaz precision inlet liner, 4.0 mm x 6.3 x 78.5, with wool (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/23305\" rel=\"noopener\">cat.# 23305<\/a>)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Temp.:<\/th><td>250 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Split Vent Flow Rate:<\/th><td>144.4 mL\/min<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\" colspan=\"2\">Oven<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Oven Temp.:<\/th><td>35 &#176;C (hold 3 min) to 220 &#176;C at 8 &#176;C\/min (hold 5 min)<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Carrier Gas<\/th><td>He, constant flow<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Flow Rate:<\/th><td>1.4 mL\/min<tr><th class=\"sub conditions_header\" scope=\"row\">Linear Velocity:<\/th><td>25.457 cm\/sec @ 35 \u00b0C<\/td><\/tr><\/table><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Detector<\/th><td>FID @ 250 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Make-up Gas Flow Rate:<\/th><td>25 mL\/min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Make-up Gas Type:<\/th><td>N<sub>2<\/sub> <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Hydrogen flow:<\/th><td>35 mL\/min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Air flow:<\/th><td>400 mL\/min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Data Rate:<\/th><td>50 Hz<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Instrument<\/th><td>Agilent 7890A GC<\/td><\/tr><\/table><\/div><\/div><\/div><div class=\"chromatogram-related-skus\"><h4>Related SKUs<\/h4><ul><li>16056<\/li><li>23305<\/li><\/ul><\/div><div class=\"chromatogram-pdf-link\"><a href=\"https:\/\/ez.restek.com\/images\/cgram\/gc_ff1392.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"><svg xmlns=\"http:\/\/www.w3.org\/2000\/svg\" width=\"18\" height=\"18\" viewBox=\"0 0 18 18\"><g data-name=\"Group 2996\"><path data-name=\"Rectangle 1246\" d=\"M0 0h18v18H0z\" style=\"fill: none;\"><\/path><\/g><g data-name=\"Group 2997\"><path data-name=\"Path 729\" d=\"M13.412 11.4v2.017H5.345V11.4H4v2.017a1.349 1.349 0 0 0 1.345 1.345h8.068a1.349 1.349 0 0 0 1.345-1.345V11.4zm-.672-2.694-.948-.948-1.741 1.735V4H8.706v5.493L6.965 7.758l-.948.948 3.361 3.361z\" transform=\"translate(-.437 -.414)\" style=\"fill: rgb(13, 123, 196);\"><\/path><\/g><\/svg>Download PDF<\/a><\/div><\/div><\/div>\n\n\n<p class=\"wp-block-paragraph\"><strong>Figure 9:<\/strong> Irish whiskey analysis using an Rtx-1301 column and GC-FID shows good resolution and a &lt;15-minute analysis time.<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article\"><div class=\"wp-block-custom-chromatogram-article\"><div class=\"chromatogram-image wide-image\"><img decoding=\"async\" src=\"https:\/\/ez.restek.com\/images\/cgram\/gc_ff1393.png\" alt=\"Irish Whiskey on Rtx-1301\" title=\"-\"><\/div><p class=\"article-id\" style=\"text-align:center\"> GC_FF1393<\/p><div class=\"chromatogram-peaks\"><h4>Peaks<\/h4><table class=\"peaks col-lg-4 col-12\">\n<thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 50px\">t<sub>R<\/sub> (min)<\/th><\/tr><\/thead>\n<tbody><tr><td class=\"num\">1.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Acetaldehyde\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/75-07-0\/Acetaldehyde\" target=\"_blank\" rel=\"noopener\">Acetaldehyde<\/a><\/td><td class=\"oth\">5.410<\/td><\/tr>\n<tr><td class=\"num\">2.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Methanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/67-56-1\/Methanol\" target=\"_blank\" rel=\"noopener\">Methanol<\/a><\/td><td class=\"oth\">5.638<\/td><\/tr>\n<tr><td class=\"num\">3.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Ethanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/64-17-5\/Ethanol\" target=\"_blank\" rel=\"noopener\">Ethanol<\/a><\/td><td class=\"oth\">7.029<\/td><\/tr>\n<tr><td class=\"num\">4.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 1-Propanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/71-23-8\/1-Propanol\" target=\"_blank\" rel=\"noopener\">1-Propanol<\/a><\/td><td class=\"oth\">9.492<\/td><\/tr>\n<tr><td class=\"num\">5.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Ethyl acetate\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/141-78-6\/Ethyl acetate\" target=\"_blank\" rel=\"noopener\">Ethyl acetate<\/a><\/td><td class=\"oth\">10.322<\/td><\/tr>\n<tr><td class=\"num\">6.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Isobutanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/78-83-1\/Isobutanol\" target=\"_blank\" rel=\"noopener\">Isobutanol<\/a><\/td><td class=\"oth\">11.388<\/td><\/tr>\n<tr><td class=\"num\">7.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Acetal\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/105-57-7\/Acetal\" target=\"_blank\" rel=\"noopener\">Acetal<\/a><\/td><td class=\"oth\">13.137<\/td><\/tr>\n<tr><td class=\"num\">8.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 3-Methyl-1-butanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/123-51-3\/3-Methyl-1-butanol\" target=\"_blank\" rel=\"noopener\">3-Methyl-1-butanol<\/a><\/td><td class=\"oth\">14.334<\/td><\/tr>\n<tr><td class=\"num\">9.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2-Methyl-1-butanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/137-32-6\/2-Methyl-1-butanol\" target=\"_blank\" rel=\"noopener\">2-Methyl-1-butanol<\/a><\/td><td class=\"oth\">14.420<\/td><\/tr>\n<\/tbody><\/table><\/div><div class=\"chromatogram-conditions\"><h4>Conditions<\/h4><div class=\"conditions-container container-fluid\"><div class=\"row\"><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Column<\/th><td>Rtx-1301, 60 m, 0.25 mm ID, 1.0 \u00b5m (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/16056\" rel=\"noopener\">cat.# 16056<\/a>)<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Standard\/Sample<\/th><td>Irish whiskey<\/td><\/tr>\n<tr><th class=\"sub conditions_header\" scope=\"row\">Diluent:<\/th><td>n\/a<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Injection<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Vol.:<\/th><td>1 \u00b5L split (split ratio 100:1)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Liner:<\/th><td>Topaz precision split liner, 4.0 mm x 6.3 x 78.5, with wool (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/23305\" rel=\"noopener\">cat.# 23305<\/a>)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Temp.:<\/th><td>250 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Split Vent Flow Rate:<\/th><td>144.4 mL\/min<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\" colspan=\"2\">Oven<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Oven Temp.:<\/th><td>35 &#176;C (hold 3 min) to 220 &#176;C at 8 &#176;C\/min (hold 5 min)<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Carrier Gas<\/th><td>He, constant flow<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Flow Rate:<\/th><td>1.4 mL\/min<tr><th class=\"sub conditions_header\" scope=\"row\">Linear Velocity:<\/th><td>25.457 cm\/sec @ 35 \u00b0C<\/td><\/tr><\/table><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Detector<\/th><td>FID @ 250 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Make-up Gas Flow Rate:<\/th><td>25 mL\/min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Make-up Gas Type:<\/th><td>N<sub>2<\/sub> <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Hydrogen flow:<\/th><td>35 mL\/min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Air flow:<\/th><td>400 mL\/min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Data Rate:<\/th><td>50 Hz<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Instrument<\/th><td>Agilent 7890A GC<\/td><\/tr><\/table><\/div><\/div><\/div><div class=\"chromatogram-related-skus\"><h4>Related SKUs<\/h4><ul><li>16056<\/li><li>23305<\/li><\/ul><\/div><div class=\"chromatogram-pdf-link\"><a href=\"https:\/\/ez.restek.com\/images\/cgram\/gc_ff1393.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"><svg xmlns=\"http:\/\/www.w3.org\/2000\/svg\" width=\"18\" height=\"18\" viewBox=\"0 0 18 18\"><g data-name=\"Group 2996\"><path data-name=\"Rectangle 1246\" d=\"M0 0h18v18H0z\" style=\"fill: none;\"><\/path><\/g><g data-name=\"Group 2997\"><path data-name=\"Path 729\" d=\"M13.412 11.4v2.017H5.345V11.4H4v2.017a1.349 1.349 0 0 0 1.345 1.345h8.068a1.349 1.349 0 0 0 1.345-1.345V11.4zm-.672-2.694-.948-.948-1.741 1.735V4H8.706v5.493L6.965 7.758l-.948.948 3.361 3.361z\" transform=\"translate(-.437 -.414)\" style=\"fill: rgb(13, 123, 196);\"><\/path><\/g><\/svg>Download PDF<\/a><\/div><\/div><\/div>\n\n\n<p class=\"wp-block-paragraph\"><strong>Figure 10:<\/strong> Marsala analysis using an Rtx-1301 column and GC-FID shows good resolution.<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article\"><div class=\"wp-block-custom-chromatogram-article\"><div class=\"chromatogram-image wide-image\"><img decoding=\"async\" src=\"https:\/\/ez.restek.com\/images\/cgram\/gc_ff1394.png\" alt=\"Marsala on Rtx-1301\" title=\"-\"><\/div><p class=\"article-id\" style=\"text-align:center\"> GC_FF1394<\/p><div class=\"chromatogram-peaks\"><h4>Peaks<\/h4><table class=\"peaks col-lg-6 col-12 peak-50\">\n<thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 50px\">t<sub>R<\/sub> (min)<\/th><\/tr><\/thead>\n<tbody><tr><td class=\"num\">1.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Acetaldehyde\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/75-07-0\/Acetaldehyde\" target=\"_blank\" rel=\"noopener\">Acetaldehyde<\/a><\/td><td class=\"oth\">5.404<\/td><\/tr>\n<tr><td class=\"num\">2.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Methanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/67-56-1\/Methanol\" target=\"_blank\" rel=\"noopener\">Methanol<\/a><\/td><td class=\"oth\">5.632<\/td><\/tr>\n<tr><td class=\"num\">3.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Ethanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/64-17-5\/Ethanol\" target=\"_blank\" rel=\"noopener\">Ethanol<\/a><\/td><td class=\"oth\">6.956<\/td><\/tr>\n<tr><td class=\"num\">4.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Ethyl formate\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/109-94-4\/Ethyl formate\" target=\"_blank\" rel=\"noopener\">Ethyl formate<\/a><\/td><td class=\"oth\">7.779<\/td><\/tr>\n<tr><td class=\"num\">5.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 1-Propanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/71-23-8\/1-Propanol\" target=\"_blank\" rel=\"noopener\">1-Propanol<\/a><\/td><td class=\"oth\">9.485<\/td><\/tr>\n<tr><td class=\"num\">6.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Ethyl acetate\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/141-78-6\/Ethyl acetate\" target=\"_blank\" rel=\"noopener\">Ethyl acetate<\/a><\/td><td class=\"oth\">10.318<\/td><\/tr>\n<\/tbody><\/table>\n<table class=\"peaks col-lg-6 col-12 peak-50\"><thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 50px\">t<sub>R<\/sub> (min)<\/th><\/tr><\/thead><tbody>\n<tr><td class=\"num\">7.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for sec-Butanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/78-92-2\/sec-Butanol\" target=\"_blank\" rel=\"noopener\">sec-Butanol<\/a><\/td><td class=\"oth\">10.507<\/td><\/tr>\n<tr><td class=\"num\">8.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Isobutanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/78-83-1\/Isobutanol\" target=\"_blank\" rel=\"noopener\">Isobutanol<\/a><\/td><td class=\"oth\">11.384<\/td><\/tr>\n<tr><td class=\"num\">9.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Acetic acid\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/64-19-7\/Acetic acid\" target=\"_blank\" rel=\"noopener\">Acetic acid<\/a><\/td><td class=\"oth\">11.479<\/td><\/tr>\n<tr><td class=\"num\">10.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Acetal\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/105-57-7\/Acetal\" target=\"_blank\" rel=\"noopener\">Acetal<\/a><\/td><td class=\"oth\">13.143<\/td><\/tr>\n<tr><td class=\"num\">11.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 3-Methyl-1-butanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/123-51-3\/3-Methyl-1-butanol\" target=\"_blank\" rel=\"noopener\">3-Methyl-1-butanol<\/a><\/td><td class=\"oth\">14.333<\/td><\/tr>\n<tr><td class=\"num\">12.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2-Methyl-1-butanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/137-32-6\/2-Methyl-1-butanol\" target=\"_blank\" rel=\"noopener\">2-Methyl-1-butanol<\/a><\/td><td class=\"oth\">14.419<\/td><\/tr>\n<tr><td class=\"num\">13.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Furfural\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/98-01-1\/Furfural\" target=\"_blank\" rel=\"noopener\">Furfural<\/a><\/td><td class=\"oth\">17.491<\/td><\/tr>\n<\/tbody><\/table><\/div><div class=\"chromatogram-conditions\"><h4>Conditions<\/h4><div class=\"conditions-container container-fluid\"><div class=\"row\"><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Column<\/th><td>Rtx-1301, 60 m, 0.25 mm ID, 1.0 \u00b5m (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/16056\" rel=\"noopener\">cat.# 16056<\/a>)<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Standard\/Sample<\/th><td>Marsala (neat)<\/td><\/tr>\n<tr><th class=\"sub conditions_header\" scope=\"row\">Diluent:<\/th><td>n\/a<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Injection<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Vol.:<\/th><td>1 \u00b5L split (split ratio 100:1)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Liner:<\/th><td>Topaz precision inlet liner, 4.0 mm x 6.3 x 78.5, with wool (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/23305\" rel=\"noopener\">cat.# 23305<\/a>)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Temp.:<\/th><td>250 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Split Vent Flow Rate:<\/th><td>144.4 mL\/min<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\" colspan=\"2\">Oven<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Oven Temp.:<\/th><td>35 &#176;C (hold 3 min) to 220 &#176;C at 8 &#176;C\/min (hold 5 min)<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Carrier Gas<\/th><td>He, constant flow<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Flow Rate:<\/th><td>1.4 mL\/min<tr><th class=\"sub conditions_header\" scope=\"row\">Linear Velocity:<\/th><td>25.457 cm\/sec @ 35 \u00b0C<\/td><\/tr><\/table><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Detector<\/th><td>FID @ 250 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Make-up Gas Flow Rate:<\/th><td>25 mL\/min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Make-up Gas Type:<\/th><td>N<sub>2<\/sub> <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Hydrogen flow:<\/th><td>35 mL\/min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Air flow:<\/th><td>400 mL\/min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Data Rate:<\/th><td>50 Hz<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Instrument<\/th><td>Agilent 7890A GC<\/td><\/tr><\/table><\/div><\/div><\/div><div class=\"chromatogram-related-skus\"><h4>Related SKUs<\/h4><ul><li>16056<\/li><li>23305<\/li><\/ul><\/div><div class=\"chromatogram-pdf-link\"><a href=\"https:\/\/ez.restek.com\/images\/cgram\/gc_ff1394.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"><svg xmlns=\"http:\/\/www.w3.org\/2000\/svg\" width=\"18\" height=\"18\" viewBox=\"0 0 18 18\"><g data-name=\"Group 2996\"><path data-name=\"Rectangle 1246\" d=\"M0 0h18v18H0z\" style=\"fill: none;\"><\/path><\/g><g data-name=\"Group 2997\"><path data-name=\"Path 729\" d=\"M13.412 11.4v2.017H5.345V11.4H4v2.017a1.349 1.349 0 0 0 1.345 1.345h8.068a1.349 1.349 0 0 0 1.345-1.345V11.4zm-.672-2.694-.948-.948-1.741 1.735V4H8.706v5.493L6.965 7.758l-.948.948 3.361 3.361z\" transform=\"translate(-.437 -.414)\" style=\"fill: rgb(13, 123, 196);\"><\/path><\/g><\/svg>Download PDF<\/a><\/div><\/div><\/div>\n\n\n<p class=\"wp-block-paragraph\"><strong>Figure 11:<\/strong> Rum analysis using an Rtx-1301 column and GC-FID shows good resolution and good peak shape for the additive propylene glycol.<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article\"><div class=\"wp-block-custom-chromatogram-article\"><div class=\"chromatogram-image wide-image\"><img decoding=\"async\" src=\"https:\/\/ez.restek.com\/images\/cgram\/gc_ff1390.png\" alt=\"Rum on Rtx-1301\" title=\"-\"><\/div><p class=\"article-id\" style=\"text-align:center\"> GC_FF1390<\/p><div class=\"chromatogram-peaks\"><h4>Peaks<\/h4><table class=\"peaks col-lg-4 col-12\">\n<thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 50px\">t<sub>R<\/sub> (min)<\/th><\/tr><\/thead>\n<tbody><tr><td class=\"num\">1.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Acetaldehyde\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/75-07-0\/Acetaldehyde\" target=\"_blank\" rel=\"noopener\">Acetaldehyde<\/a><\/td><td class=\"oth\">5.410<\/td><\/tr>\n<tr><td class=\"num\">2.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Methanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/67-56-1\/Methanol\" target=\"_blank\" rel=\"noopener\">Methanol<\/a><\/td><td class=\"oth\">5.634<\/td><\/tr>\n<tr><td class=\"num\">3.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Ethanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/64-17-5\/Ethanol\" target=\"_blank\" rel=\"noopener\">Ethanol<\/a><\/td><td class=\"oth\">7.026<\/td><\/tr>\n<tr><td class=\"num\">4.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 1-Propanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/71-23-8\/1-Propanol\" target=\"_blank\" rel=\"noopener\">1-Propanol<\/a><\/td><td class=\"oth\">9.492<\/td><\/tr>\n<tr><td class=\"num\">5.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Ethyl acetate\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/141-78-6\/Ethyl acetate\" target=\"_blank\" rel=\"noopener\">Ethyl acetate<\/a><\/td><td class=\"oth\">10.321<\/td><\/tr>\n<tr><td class=\"num\">6.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Isobutanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/78-83-1\/Isobutanol\" target=\"_blank\" rel=\"noopener\">Isobutanol<\/a><\/td><td class=\"oth\">11.387<\/td><\/tr>\n<tr><td class=\"num\">7.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Acetal\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/105-57-7\/Acetal\" target=\"_blank\" rel=\"noopener\">Acetal<\/a><\/td><td class=\"oth\">13.138<\/td><\/tr>\n<tr><td class=\"num\">8.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 3-Methyl-1-butanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/123-51-3\/3-Methyl-1-butanol\" target=\"_blank\" rel=\"noopener\">3-Methyl-1-butanol<\/a><\/td><td class=\"oth\">14.334<\/td><\/tr>\n<tr><td class=\"num\">9.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2-Methyl-1-butanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/137-32-6\/2-Methyl-1-butanol\" target=\"_blank\" rel=\"noopener\">2-Methyl-1-butanol<\/a><\/td><td class=\"oth\">14.420<\/td><\/tr>\n<tr><td class=\"num\">10.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Propylene glycol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/57-55-6\/Propylene glycol\" target=\"_blank\" rel=\"noopener\">Propylene glycol<\/a><\/td><td class=\"oth\">15.749<\/td><\/tr>\n<\/tbody><\/table><\/div><div class=\"chromatogram-conditions\"><h4>Conditions<\/h4><div class=\"conditions-container container-fluid\"><div class=\"row\"><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Column<\/th><td>Rtx-1301, 60 m, 0.25 mm ID, 1.0 \u00b5m (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/16056\" rel=\"noopener\">cat.# 16056<\/a>)<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Standard\/Sample<\/th><td>Rum (neat)<\/td><\/tr>\n<tr><th class=\"sub conditions_header\" scope=\"row\">Diluent:<\/th><td>n\/a<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Injection<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Vol.:<\/th><td>1 \u00b5L split (split ratio 100:1)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Liner:<\/th><td>Topaz, precision inlet liner, 4.0 mm x 6.3 x 78.5, with wool (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/23305\" rel=\"noopener\">cat.# 23305<\/a>)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Temp.:<\/th><td>250 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Split Vent Flow Rate:<\/th><td>144.4 mL\/min<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\" colspan=\"2\">Oven<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Oven Temp.:<\/th><td>35 &#176;C (hold 3 min) to 220 &#176;C at 8 &#176;C\/min (hold 5 min)<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Carrier Gas<\/th><td>He, constant flow<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Flow Rate:<\/th><td>1.4 mL\/min<tr><th class=\"sub conditions_header\" scope=\"row\">Linear Velocity:<\/th><td>25.457 cm\/sec @ 35 \u00b0C<\/td><\/tr><\/table><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Detector<\/th><td>FID @ 250 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Make-up Gas Flow Rate:<\/th><td>25 mL\/min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Make-up Gas Type:<\/th><td>N<sub>2<\/sub> <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Hydrogen flow:<\/th><td>35 mL\/min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Air flow:<\/th><td>400 mL\/min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Data Rate:<\/th><td>50 Hz<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Instrument<\/th><td>Agilent 7890A GC<\/td><\/tr><\/table><\/div><\/div><\/div><div class=\"chromatogram-related-skus\"><h4>Related SKUs<\/h4><ul><li>16056<\/li><li>23305<\/li><\/ul><\/div><div class=\"chromatogram-pdf-link\"><a href=\"https:\/\/ez.restek.com\/images\/cgram\/gc_ff1390.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"><svg xmlns=\"http:\/\/www.w3.org\/2000\/svg\" width=\"18\" height=\"18\" viewBox=\"0 0 18 18\"><g data-name=\"Group 2996\"><path data-name=\"Rectangle 1246\" d=\"M0 0h18v18H0z\" style=\"fill: none;\"><\/path><\/g><g data-name=\"Group 2997\"><path data-name=\"Path 729\" d=\"M13.412 11.4v2.017H5.345V11.4H4v2.017a1.349 1.349 0 0 0 1.345 1.345h8.068a1.349 1.349 0 0 0 1.345-1.345V11.4zm-.672-2.694-.948-.948-1.741 1.735V4H8.706v5.493L6.965 7.758l-.948.948 3.361 3.361z\" transform=\"translate(-.437 -.414)\" style=\"fill: rgb(13, 123, 196);\"><\/path><\/g><\/svg>Download PDF<\/a><\/div><\/div><\/div>\n\n\n<h2 class=\"wp-block-heading\">Meet the Experts<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\"><\/p>\n\n\n\n<figure class=\"wp-block-table\"><table><tbody><tr><td><img loading=\"lazy\" decoding=\"async\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/people-Hilliard-Corby.jpg\" alt=\"photo of corby\" width=\"350\" height=\"350\" title=\"-\"><\/td><td><strong>Corby Hilliard<\/strong><br>As an advanced scientist in Restek\u2019s GC lab with over 20 years of experience, Corby\u2019s focus is on generating methodology and application data predominately for the food, petrochemical, and environmental industries. He is also involved with R&amp;D and new product development.<\/td><\/tr><tr><td><img loading=\"lazy\" decoding=\"async\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/people-King-Carolyn.jpg\" alt=\"photo of carolyn\" width=\"350\" height=\"350\" title=\"-\"><\/td><td><strong>Carolyn King<\/strong><br>Carolyn is currently a technical service specialist at Restek. She began her career in 2012 in the QC lab conducting testing to ensure product quality before joining the technical service group in 2015. Carolyn holds a BS in animal sciences and nutritional sciences from The Pennsylvania State University.<\/td><\/tr><\/tbody><\/table><\/figure>\n\n\n\n<h2 class=\"wp-block-heading\">References<\/h2>\n\n\n\n<ol class=\"wp-block-list\">\n<li>AOAC International, AOAC Official methods of analysis, 22nd edition, Edited by George W. Latimer, Jr., 3 vols., Oxford University Press, Oxford, UK, 2023.<\/li>\n<\/ol>\n","protected":false},"excerpt":{"rendered":"<p>Testing of several GC column phases for alcoholic beverage analysis demonstrated that the Rtx-1301 phase has optimal selectivity, producing excellent peak shape and resolution for key target analytes.<\/p>\n","protected":false},"author":11,"featured_media":6889,"comment_status":"closed","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"_acf_changed":false,"_kad_blocks_custom_css":"","_kad_blocks_head_custom_js":"","_kad_blocks_body_custom_js":"","_kad_blocks_footer_custom_js":"","_kadence_starter_templates_imported_post":false,"_kad_post_transparent":"","_kad_post_title":"","_kad_post_layout":"","_kad_post_sidebar_id":"","_kad_post_content_style":"","_kad_post_vertical_padding":"","_kad_post_feature":"","_kad_post_feature_position":"","_kad_post_header":false,"_kad_post_footer":false,"footnotes":"","_ppma_block_editor_authors":""},"categories":[787],"tags":[],"industries-application":[],"post-badge":[],"resource-type":[],"product-library":[],"resource-technique":[],"ppma_author":[414],"class_list":["post-65643","post","type-post","status-publish","format-standard","has-post-thumbnail","hentry","category-artikel"],"acf":[],"taxonomy_info":{"category":[{"value":787,"label":"Artikel"}]},"featured_image_src_large":["https:\/\/discover.restek.com\/wp-content\/uploads\/feature-FFAR4494-1024x536.jpg",1024,536,true],"author_info":{"display_name":"Restek Corporation","author_link":"https:\/\/discover.restek.com\/fr\/author\/restek-corporation\/"},"comment_info":0,"category_info":[{"term_id":787,"name":"Artikel","slug":"artikel","term_group":0,"term_taxonomy_id":787,"taxonomy":"category","description":"","parent":0,"count":510,"filter":"raw","cat_ID":787,"category_count":510,"category_description":"","cat_name":"Artikel","category_nicename":"artikel","category_parent":0}],"tag_info":false,"authors":[{"term_id":414,"user_id":11,"is_guest":0,"slug":"restek-corporation","display_name":"Restek Corporation","avatar_url":{"url":"https:\/\/discover.restek.com\/wp-content\/uploads\/Restek_Favicon_300x300.jpg","url2x":"https:\/\/discover.restek.com\/wp-content\/uploads\/Restek_Favicon_300x300.jpg"},"author_category":"1","first_name":"Restek","last_name":"Corporation","user_url":"https:\/\/www.linkedin.com\/company\/restek-corporation\/posts\/?feedView=all","job_title":"Chromatography Products and Solutions","description":"Restek is a leading provider of chromatography columns, accessories, and certified reference materials. Trust Restek for reliable, high-quality analytical solutions."}],"_links":{"self":[{"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/posts\/65643","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/users\/11"}],"replies":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/comments?post=65643"}],"version-history":[{"count":11,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/posts\/65643\/revisions"}],"predecessor-version":[{"id":78786,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/posts\/65643\/revisions\/78786"}],"wp:featuredmedia":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/media\/6889"}],"wp:attachment":[{"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/media?parent=65643"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/categories?post=65643"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/tags?post=65643"},{"taxonomy":"industries-application","embeddable":true,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/industries-application?post=65643"},{"taxonomy":"post-badge","embeddable":true,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/post-badge?post=65643"},{"taxonomy":"resource-type","embeddable":true,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/resource-type?post=65643"},{"taxonomy":"product-library","embeddable":true,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/product-library?post=65643"},{"taxonomy":"resource-technique","embeddable":true,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/resource-technique?post=65643"},{"taxonomy":"author","embeddable":true,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/ppma_author?post=65643"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}