{"id":76446,"date":"2020-03-16T00:00:00","date_gmt":"2020-03-16T00:00:00","guid":{"rendered":"https:\/\/discover.restek.com\/?p=76446"},"modified":"2025-12-26T19:47:16","modified_gmt":"2025-12-26T19:47:16","slug":"over-tightening-your-fittings-has-consequences","status":"publish","type":"post","link":"https:\/\/discover.restek.com\/fr\/blogs\/gnbl5523\/over-tightening-your-fittings-has-consequences","title":{"rendered":"Over tightening your fittings has consequences."},"content":{"rendered":"\n<p class=\"wp-block-paragraph\">If you use the correct fittings and seals in the gas chromatography system, and follow some simple steps, there is absolutely no need to over tighten any fitting to achieve a leak free seal.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">For almost all fittings that use seals in a GC system, whether they are ferrules, septa or o-rings, the tightening process should be the same:<\/p>\n\n\n\n<ol class=\"wp-block-list\">\n<li>Make sure the place where you are applying the fitting is cooled down and the gas flow is turned off<\/li>\n\n\n\n<li>Select the correct size fitting and seal for the application<\/li>\n\n\n\n<li>Correctly fit the fitting and seal together according to the manufacturer\u2019s guidelines<\/li>\n\n\n\n<li>Finger-tighten the fitting on to the receiving part<\/li>\n\n\n\n<li>Using the appropriate tool (wrench or fingers) tighten the fitting and extra\u00a0<strong><u>\u00bc turn<\/u><\/strong>.<\/li>\n\n\n\n<li>Establish a gas flow and pressurize the section of the GC system.<\/li>\n\n\n\n<li>Use a Restek electronic leak detector (28500) to check for leaks around the fitting<\/li>\n\n\n\n<li>If you do detect a leak further tighten the fitting in\u00a0<strong><u>1\/8 turn<\/u><\/strong>\u00a0increments until the leak ceases.<\/li>\n<\/ol>\n\n\n\n<p class=\"wp-block-paragraph\"><em>Note: If you continue to notice a leak and you have done&nbsp;<u>one complete revolution<\/u>&nbsp;of the fitting&nbsp;<strong>STOP<\/strong>.&nbsp; There is something wrong.&nbsp; Rather than damaging the fitting, or the receiving part, remove the fitting, get a new one and a new seal and start again.<\/em><\/p>\n\n\n\n<p class=\"wp-block-paragraph\">\u00a0<strong>Common fittings and seal, and the consequences of over tightening<\/strong><\/p>\n\n\n\n<p class=\"wp-block-paragraph\"><em>Septa<\/em><\/p>\n\n\n\n<p class=\"wp-block-paragraph\">This is the easiest seal to get leak free.&nbsp; You should not need a wrench; minimal finger tightening is usually sufficient.&nbsp; If you over tighten you compress the septum.&nbsp; This puts the septum under stress, and so when the septum is penetrated by a needle it is more easily ripped apart leading to coring and leaks.&nbsp; Coring can result in particles of septum in the liner which can adversely affect chromatography results.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">You should leak check with and electronic leak detector around the septum fitting every day.&nbsp; It is one of the hardest working seals, and the easiest and cheapest to change.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\"><em>Reducing Nut<\/em><\/p>\n\n\n\n<p class=\"wp-block-paragraph\">This is the fitting at the bottom of the inlet where the column nut attaches.&nbsp; Over tightening this nut can damage the threads on both the nut and the inlet fitting.&nbsp; We have also seen the inlet bent by over tightening to such an extent that liners could not fit down inside.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">For GCs like Agilent, Thermo Trace 1300\/1310 and Perkin Elmer 590\/690 the simple way to avoid over tightening, or the use of excessive force, is to use a dual vespel ring inlet seal.&nbsp; The dual vespel ring allows for a soft but effective seal, and enables you to follow the tightening procedure above.<\/p>\n\n\n<style>.kb-image76446_07252d-83 .kb-image-has-overlay:after{opacity:0.3;}<\/style>\n<figure class=\"wp-block-kadence-image kb-image76446_07252d-83 size-full\"><img loading=\"lazy\" decoding=\"async\" width=\"200\" height=\"200\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-over-tightening-your-fittings-has-consequences-01.jpg\" alt=\"\" class=\"kb-img wp-image-78434\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-over-tightening-your-fittings-has-consequences-01.jpg 200w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-over-tightening-your-fittings-has-consequences-01-150x150.jpg 150w\" sizes=\"auto, (max-width: 200px) 100vw, 200px\" \/><figcaption>Dual Vespel Inlet Seals come in several different surface finishes.<br><\/figcaption><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\">It should be noted that when moderately high inlet temperatures are employed the dual vespel ring can shrink slightly, which can cause a leak, but that is easily remedied by a slight additional tightening after a couple of heat cycles.&nbsp; With good, frequent leak monitoring with a Restek electronic leak detector this would not cause any issues.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\"><em>Column Nuts<\/em><\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Over tightening column nuts has many consequences.&nbsp; First you could crush the column, whether that is a fused silica or MXT column.&nbsp; This in turn can lead to leaks and\/or flow restrictions, and in some cases the column may break off in the fitting.&nbsp; Second, the ferrule may crack which can lead to leaks.&nbsp; Lastly the ferrule may deform so much that it may extrude into the fittings where it can block the flow or lead to leaks, and getting stuck.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\"><em>Standard connections<\/em><\/p>\n\n\n\n<p class=\"wp-block-paragraph\">These are like Swagelok or Parker fitting that are brass of stainless steel, and sometimes use a graphite ferrule.&nbsp; Over tightening these fittings can easily lead to stripping the threads, especially in brass fittings.&nbsp; We have seen where over tightening has led to the crushing of copper tubing, and this restricting flow, and in extreme cases cracking the tubing.&nbsp; If you use graphite ferrules in these fittings, as before the graphite can be extruded into the fitting which can cause leaks.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Never mix material.&nbsp; If you have a steel fitting, don\u2019t use a brass nut, and vice versa.&nbsp; This can lead to easier thread stripping.&nbsp; Also as a rule of thumb use brass fitting with copper piping and stainless steel fittings with stainless steel tubing.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\"><em>Screw top vials<\/em><\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Don\u2019t over tighten screw top vials.&nbsp; This can pinch the septum resulting in a poor fit and potential for loss of sample.&nbsp; We have seen it where an over tightened cap will pinch the septum so much that when a syringe tries to penetrate the septum it pushes it right into the vial.&nbsp; It is almost impossible to leak check this seal.&nbsp; So you have two choices: a gentle touch, or switch to crimp top vials.<\/p>\n\n\n\n<div style=\"height:100px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n","protected":false},"excerpt":{"rendered":"<p>If you use the correct fittings and seals in the gas chromatography system, and follow some simple steps, there is absolutely no need to over tighten any fitting to achieve a leak free seal. For almost all fittings that use&#8230;<\/p>\n","protected":false},"author":3,"featured_media":0,"comment_status":"closed","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"_acf_changed":false,"_kad_blocks_custom_css":"","_kad_blocks_head_custom_js":"","_kad_blocks_body_custom_js":"","_kad_blocks_footer_custom_js":"","_kadence_starter_templates_imported_post":false,"_kad_post_transparent":"","_kad_post_title":"","_kad_post_layout":"","_kad_post_sidebar_id":"","_kad_post_content_style":"","_kad_post_vertical_padding":"","_kad_post_feature":"","_kad_post_feature_position":"","_kad_post_header":false,"_kad_post_footer":false,"footnotes":"","_ppma_block_editor_authors":""},"categories":[9],"tags":[],"industries-application":[],"post-badge":[],"resource-type":[],"product-library":[],"resource-technique":[],"ppma_author":[599],"class_list":["post-76446","post","type-post","status-publish","format-standard","hentry","category-blogs"],"acf":[],"taxonomy_info":{"category":[{"value":9,"label":"Blogs"}]},"featured_image_src_large":false,"author_info":{"display_name":"Audrey Crater","author_link":"https:\/\/discover.restek.com\/fr\/author\/audrey-craterrestek-com\/"},"comment_info":0,"category_info":[{"term_id":9,"name":"Blogs","slug":"blogs","term_group":0,"term_taxonomy_id":9,"taxonomy":"category","description":"","parent":0,"count":458,"filter":"raw","cat_ID":9,"category_count":458,"category_description":"","cat_name":"Blogs","category_nicename":"blogs","category_parent":0}],"tag_info":false,"authors":[{"term_id":599,"user_id":0,"is_guest":1,"slug":"mark-badger","display_name":"Mark Badger","avatar_url":"https:\/\/secure.gravatar.com\/avatar\/?s=96&d=mm&r=g","author_category":"1","first_name":"Mark","last_name":"Badger","user_url":"","job_title":"","description":""}],"_links":{"self":[{"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/posts\/76446","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/users\/3"}],"replies":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/comments?post=76446"}],"version-history":[{"count":1,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/posts\/76446\/revisions"}],"predecessor-version":[{"id":78441,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/posts\/76446\/revisions\/78441"}],"wp:attachment":[{"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/media?parent=76446"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/categories?post=76446"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/tags?post=76446"},{"taxonomy":"industries-application","embeddable":true,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/industries-application?post=76446"},{"taxonomy":"post-badge","embeddable":true,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/post-badge?post=76446"},{"taxonomy":"resource-type","embeddable":true,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/resource-type?post=76446"},{"taxonomy":"product-library","embeddable":true,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/product-library?post=76446"},{"taxonomy":"resource-technique","embeddable":true,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/resource-technique?post=76446"},{"taxonomy":"author","embeddable":true,"href":"https:\/\/discover.restek.com\/fr\/wp-json\/wp\/v2\/ppma_author?post=76446"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}