- Up to 3.3x faster with up to 76% less helium consumption compared to conventional methods.
- Get the speed advantage of LPGC-MS with a simple column change.
- Robust, factory-coupled column kit ensures leak-free performance.
Multiresidue pesticides analysis is a cornerstone of food safety testing, and labs are generally under pressure to manage both a high volume of samples and rapid turnaround time requirements. This creates demand for faster GC-MS and GC-MS/MS methods, but typical approaches involve expensive instrumentation or “fast GC” techniques that have capacity issues (narrow-bore columns) or MS-compatibility concerns (hydrogen carrier gas). Low-pressure GC-MS (LPGC-MS) is option that can provide significant speed gains without these drawbacks, but, historically, the challenging setup has been a barrier to implementation. This LPGC-MS pesticides analysis utilizes a unique LPGC Rtx-5ms column kit (cat.# 11800 or 11802) that is comprised of a narrow restrictor column that is factory coupled to a wider Rtx-5ms analytical column . Using either of these pre-connected kits allows the speed gains and helium savings of LPGC-MS to be obtained by making a simple column change and updating the instrument method with the new column dimensions, oven ramp, and flow rates. Either Rtx-5ms kit will perform the analysis 3x faster. Choose the kit based on your needs: cat.# 11800 has an integrated transfer line that can make background stabilization faster and can handle higher flow rates. Choose cat.# 11802 if less helium usage and lower flow rates are preferred.
As shown in the LPGC-MS pesticides analysis of 209 compounds in strawberries (Figure 1), all analytes elute quickly, with deltamethrin eluting last at 8.33 minutes. This time is three-fold faster than our analysis of the same extract on a conventional 30 m, 0.25 mm ID, 0.25 µm 5-type column where deltamethrin was again the final compound and eluted at 26.34 minutes. In addition to the speed improvement, LPGC kit (cat.# 11800) used 54% less helium than the conventional method. LPGC kit cat.# 11802 used 76% less helium while remaining 3x faster (see Figure 2).
For busy food labs needing faster methods, the speed gains of LPGC-MS pesticides analysis are an effective way to increase sample throughput. This once-challenging setup is now much simpler to implement using a low-pressure GC column kit from Restek. Visit www.restek.com/lpgc to learn more about this powerful technique.

GC_FS0572
Peaks
| Peaks | tR (min) | Precursor Ion | Product Ion | Collision Energy | Confirmation Precursor Ion | Confirmation Product Ion | Collision Energy | |
|---|---|---|---|---|---|---|---|---|
| 1. | Allidochlor | 3.37 | 138 | 81 | 8 | 132 | 56 | 6 |
| 2. | Dichlobenil | 3.63 | 171 | 100 | 25 | 136 | 100.1 | 10 |
| 3. | Biphenyl | 3.75 | 154.1 | 115 | 25 | 152.1 | 126.1 | 23 |
| 4. | Mevinphos | 3.88 | 192 | 127.1 | 10 | 127.1 | 95 | 16 |
| 5. | 3,4-Dichloroaniline | 3.92 | 163 | 90 | 16 | 161 | 126.1 | 8 |
| 6. | Pebulate | 4.01 | 203 | 128 | 8 | 128 | 57 | 8 |
| 7. | Etridiazole | 4.02 | 211 | 139.9 | 22 | 182.8 | 139.9 | 16 |
| 8. | <a class="cmpd_link" title="View compound information for N-(2,4-Dimethylphenyl) formamide” title=”View compound information for N-(2,4-Dimethylphenyl) formamide” href=”https://ez.restek.com/compound/view/en/60397-77-5/N-(2,4-Dimethylphenyl) formamide”>N-(2,4-Dimethylphenyl) formamide | 4.10 | 121 | 106 | 8 | 120 | 77 | 15 |
| 9. | Tetrahydrophthalimide | 4.10 | 151.1 | 80.1 | 6 | 151.1 | 122 | 10 |
| 10. | Methacrifos | 4.19 | 240 | 180 | 10 | 125 | 79 | 6 |
| 11. | Chloroneb | 4.24 | 206 | 191 | 10 | 191 | 113 | 13 |
| 12. | 2-Phenylphenol | 4.31 | 170.2 | 141.2 | 22 | 141 | 115.1 | 12 |
| 13. | Pentachlorobenzene | 4.33 | 250 | 142 | 30 | 250 | 179 | 30 |
| 14. | Propachlor | 4.60 | 176 | 57 | 8 | 120 | 77 | 19 |
| 15. | Tecnazene | 4.60 | 260.9 | 202.9 | 13 | 214.9 | 178.9 | 8 |
| 16. | Diphenylamine | 4.63 | 169.1 | 167.1 | 25 | 169.1 | 168.1 | 12 |
| 17. | Cycloate | 4.67 | 215 | 154 | 6 | 154 | 83 | 8 |
| 18. | 2,3,5,6-Tetrachloroaniline | 4.69 | 231 | 122 | 30 | 231 | 160 | 22 |
| 19. | Chlorpropham | 4.70 | 213.1 | 171.1 | 8 | 171 | 127.1 | 8 |
| 20. | Ethalfluralin | 4.75 | 316.1 | 276.1 | 10 | 276.1 | 202.1 | 15 |
| 21. | Trifluralin | 4.78 | 306.1 | 206.2 | 12 | 306.1 | 264.1 | 8 |
| 22. | Benfluralin | 4.80 | 292 | 160.1 | 20 | 292 | 264.1 | 8 |
| 23. | Sulfotepp | 4.85 | 322 | 146 | 23 | 202 | 146 | 10 |
| 24. | Diallate 1 | 4.89 | 234.1 | 150 | 18 | 234.1 | 192.1 | 12 |
| 25. | Phorate | 4.90 | 260.1 | 75 | 8 | 121.1 | 65 | 10 |
| 26. | Diallate 2 | 4.95 | 234.1 | 150 | 18 | 234.1 | 192.1 | 12 |
| 27. | α-BHC | 5.00 | 219 | 183 | 8 | 181 | 145 | 15 |
| 28. | Hexachlorobenzene | 5.02 | 283.8 | 213.9 | 30 | 248.9 | 213.9 | 14 |
| 29. | Dicloran | 5.03 | 206 | 147.9 | 18 | 176 | 148 | 10 |
| 30. | Pentachloroanisole | 5.04 | 279.9 | 236.9 | 23 | 266.9 | 238.9 | 10 |
| 31. | Atrazine | 5.05 | 215 | 173.1 | 8 | 200.1 | 122.1 | 10 |
| 32. | Clomazone | 5.10 | 204 | 107 | 18 | 125 | 89 | 13 |
| 33. | β-BHC | 5.12 | 219 | 183 | 8 | 181 | 145 | 15 |
| 34. | Profluralin | 5.14 | 330.1 | 69 | 25 | 318.1 | 199 | 17 |
| 35. | Terbuthylazine | 5.16 | 229.1 | 173.1 | 8 | 172.8 | 137.9 | 8 |
| 36. | Terbufos | 5.17 | 231 | 129 | 23 | 231 | 175 | 12 |
| 37. | γ-BHC | 5.16 | 219 | 183 | 8 | 181 | 145 | 15 |
| 38. | Propyzamide | 5.18 | 173 | 109 | 27 | 173 | 145 | 13 |
| 39. | Diazinon | 5.19 | 304 | 179.1 | 12 | 137.1 | 84.1 | 12 |
| 40. | Fonofos | 5.19 | 246.1 | 137.1 | 8 | 137.1 | 109 | 6 |
| 41. | Quintozene | 5.19 | 294.8 | 237 | 15 | 236.8 | 119 | 22 |
| 42. | Fluchloralin | 5.21 | 306.1 | 159.7 | 20 | 306.1 | 264 | 8 |
| 43. | Pentachlorobenzonitrile | 5.21 | 275 | 205 | 30 | 273 | 238 | 17 |
| 44. | Pyrimethanil | 5.21 | 198.1 | 118.1 | 32 | 198.1 | 183 | 16 |
| 45. | Tefluthrin | 5.24 | 177 | 127.1 | 15 | 177 | 137.1 | 15 |
| 46. | Disulfoton | 5.25 | 153.1 | 97 | 12 | 88.1 | 59.9 | 6 |
| 47. | Terbacil | 5.26 | 161 | 144 | 14 | 160 | 117 | 8 |
| 48. | Isazophos | 5.29 | 172.1 | 130 | 10 | 161.1 | 119 | 8 |
| 49. | δ-BHC | 5.31 | 219 | 183 | 8 | 181 | 145 | 15 |
| 50. | Triallate | 5.32 | 270 | 186 | 18 | 268 | 184 | 20 |
| 51. | Chlorothalonil | 5.34 | 266 | 170 | 23 | 264 | 168 | 23 |
| 52. | Anthracene (IS) | 5.35 | 178 | 152 | 20 | |||
| 53. | Endosulfan ether | 5.42 | 240.9 | 206 | 13 | 238.9 | 204 | 13 |
| 54. | Propanil | 5.44 | 219 | 163 | 8 | 217 | 161 | 8 |
| 55. | Pentachloroaniline | 5.45 | 265 | 194 | 22 | 263 | 192 | 20 |
| 56. | Dimethachlor | 5.46 | 197 | 148 | 8 | 134 | 105 | 13 |
| 57. | Acetochlor | 5.48 | 223 | 132 | 20 | 174 | 146 | 12 |
| 58. | Transfluthrin | 5.48 | 163 | 143 | 13 | 127.1 | 91.1 | 8 |
| 59. | Vinclozolin | 5.50 | 285 | 212 | 12 | 212 | 172 | 12 |
| 60. | Chlorpyrifos methyl | 5.52 | 286 | 93 | 22 | 286 | 207.9 | 12 |
| 61. | Methyl parathion | 5.52 | 263 | 109 | 10 | 263 | 136.2 | 8 |
| 62. | Tolclofos-methyl | 5.54 | 267 | 252 | 10 | 265 | 250 | 10 |
| 63. | Alachlor | 5.55 | 188 | 160 | 10 | 146 | 118 | 8 |
| 64. | Metalaxyl | 5.56 | 220.1 | 160.2 | 10 | 160.1 | 145.1 | 10 |
| 65. | Propisochlor | 5.56 | 223 | 132 | 18 | 162 | 120 | 13 |
| 66. | Fenchlorphos | 5.59 | 287 | 272 | 11 | 285 | 240 | 23 |
| 67. | Heptachlor | 5.59 | 272 | 237 | 13 | 100 | 65 | 12 |
| 68. | Prodiamine | 5.63 | 321 | 279 | 6 | 279 | 203 | 8 |
| 69. | Pirimiphos methyl | 5.65 | 305.1 | 180.1 | 8 | 290.1 | 125 | 22 |
| 70. | Fenitrothion | 5.66 | 277 | 109 | 16 | 260 | 125 | 12 |
| 71. | Linuron | 5.69 | 248 | 61 | 8 | 187 | 124 | 21 |
| 72. | Malathion | 5.69 | 173 | 99 | 13 | 127 | 99 | 6 |
| 73. | Dichlofluanid | 5.72 | 224 | 123 | 12 | 123 | 77.1 | 16 |
| 74. | Pentachlorothioanisole | 5.73 | 296 | 246 | 32 | 296 | 263 | 12 |
| 75. | Metolachlor | 5.76 | 238 | 162 | 10 | 162 | 133 | 13 |
| 76. | Chlorpyrifos | 5.77 | 314 | 257.9 | 12 | 196.9 | 168.9 | 14 |
| 77. | Fenthion | 5.77 | 278 | 109 | 18 | 278 | 169 | 17 |
| 78. | Parathion | 5.78 | 291.2 | 109.1 | 10 | 109 | 81 | 8 |
| 79. | Anthraquinone | 5.79 | 208.1 | 180.1 | 10 | 180.1 | 152.1 | 12 |
| 80. | Triadimefon | 5.79 | 208 | 110.9 | 24 | 208 | 180.8 | 10 |
| 81. | Aldrin | 5.80 | 298 | 263 | 8 | 263 | 193 | 31 |
| 82. | Chlorthal-dimethyl | 5.81 | 331.8 | 301 | 8 | 298.9 | 220.9 | 24 |
| 83. | 4,4′-Dichlorobenzophenone | 5.81 | 139 | 75 | 27 | 111 | 75.1 | 12 |
| 84. | Pirimiphos ethyl | 5.85 | 318.2 | 166.1 | 12 | 304.1 | 168.1 | 12 |
| 85. | Fenson | 5.86 | 268 | 77.1 | 18 | 141 | 77.1 | 8 |
| 86. | Cyprodinil | 5.87 | 225.2 | 209.8 | 12 | 224.1 | 208.1 | 20 |
| 87. | Diphenamid | 5.88 | 239 | 167 | 8 | 167 | 115 | 40 |
| 88. | Isopropalin | 5.89 | 280.1 | 118.1 | 25 | 280.1 | 238.1 | 8 |
| 89. | MGK 264 1 | 5.89 | 164.1 | 80 | 24 | 164.1 | 93.1 | 12 |
| 90. | Bromophos-methyl | 5.90 | 331 | 316 | 13 | 329 | 314 | 13 |
| 91. | Chlozolinate | 5.95 | 331 | 259.1 | 8 | 186 | 145 | 16 |
| 92. | Pendimethalin | 5.95 | 281.1 | 252.2 | 8 | 252.1 | 162.1 | 10 |
| 93. | Bioallethrin | 5.96 | 136.1 | 93 | 11 | 123.2 | 81.1 | 8 |
| 94. | Fipronil | 5.97 | 367 | 213 | 30 | 369 | 215 | 30 |
| 95. | MGK 264 2 | 5.97 | 164 | 67 | 10 | 164.1 | 80 | 24 |
| 96. | Isodrin | 5.97 | 260.9 | 190.9 | 28 | 193 | 123 | 30 |
| 97. | Metazachlor | 5.97 | 209 | 132 | 15 | 133 | 117 | 25 |
| 98. | Penconazole | 5.97 | 248.1 | 192 | 16 | 159 | 89 | 30 |
| 99. | Chlorfenvinphos 1 | 6.04 | 323 | 267 | 10 | 267 | 159 | 15 |
| 100. | Tolylfluanid | 6.05 | 240 | 137 | 10 | 238 | 137 | 10 |
| 101. | Bromfenvinphos-methyl | 6.06 | 295 | 109 | 15 | 109 | 79 | 5 |
| 102. | Heptachlor epoxide | 6.07 | 353 | 263 | 13 | 263 | 193 | 29 |
| 103. | Quinalphos | 6.07 | 298 | 156.1 | 8 | 146.1 | 118.1 | 10 |
| 104. | Triadimenol | 6.07 | 168.1 | 70.1 | 10 | 128 | 65.1 | 22 |
| 105. | Triflumizole | 6.09 | 205.9 | 179 | 14 | 179 | 143.8 | 14 |
| 106. | Captan | 6.10 | 149 | 105 | 6 | 106.8 | 79 | 12 |
| 107. | Procymidone | 6.11 | 283 | 67.1 | 28 | 283 | 96 | 10 |
| 108. | Folpet | 6.14 | 259.9 | 130 | 16 | 103.9 | 76 | 10 |
| 109. | Chlorbenside | 6.15 | 268 | 125.1 | 8 | 125 | 89.1 | 17 |
| 110. | Bromophos-ethyl | 6.17 | 358.9 | 303 | 17 | 330.9 | 303 | 8 |
| 111. | trans-Chlordane | 6.19 | 375 | 266 | 18 | 272 | 237 | 12 |
| 112. | Paclobutrazol | 6.19 | 237.8 | 127 | 10 | 236.1 | 125.1 | 12 |
| 113. | 2,4′-DDE | 6.19 | 316 | 246 | 15 | 246 | 176 | 30 |
| 114. | Tetrachlorvinphos | 6.20 | 333 | 109 | 17 | 331 | 109 | 20 |
| 115. | Fenamiphos | 6.25 | 303.1 | 195.1 | 8 | 217 | 202 | 9 |
| 116. | Flutolanil | 6.26 | 281 | 173 | 10 | 173 | 145 | 14 |
| 117. | Prothiofos | 6.26 | 309 | 239 | 15 | 267 | 221 | 16 |
| 118. | cis-Chlordane | 6.27 | 375 | 266 | 18 | 272 | 237 | 12 |
| 119. | Bromfenvinphos | 6.27 | 325 | 269 | 10 | 323 | 267 | 10 |
| 120. | Fludioxonil | 6.27 | 248 | 127 | 30 | 153.8 | 127 | 8 |
| 121. | Flutriafol | 6.27 | 219.1 | 123 | 14 | 123.1 | 95 | 12 |
| 122. | Iodofenphos | 6.27 | 377 | 332 | 30 | 377 | 362 | 15 |
| 123. | Pretilachlor | 6.27 | 262 | 202 | 8 | 162 | 132 | 18 |
| 124. | Profenofos | 6.27 | 339 | 188 | 30 | 337 | 267 | 13 |
| 125. | Chlorfenvinphos 2 | 6.27 | 323 | 267 | 10 | 267 | 159 | 15 |
| 126. | Endosulfan I | 6.28 | 241 | 206 | 10 | 195 | 159 | 6 |
| 127. | trans-Nonachlor | 6.28 | 409 | 263 | 25 | 409 | 300 | 23 |
| 128. | Chlorfenson | 6.29 | 302 | 175 | 8 | 175 | 111 | 8 |
| 129. | Oxadiazon | 6.33 | 258 | 175 | 8 | 175 | 112 | 13 |
| 130. | Oxyfluorfen | 6.34 | 300 | 223 | 15 | 252.1 | 146.1 | 33 |
| 131. | Myclobutanil | 6.38 | 179 | 125 | 14 | 179 | 152 | 8 |
| 132. | Bupirimate | 6.39 | 316.2 | 208.2 | 8 | 273.1 | 193.2 | 8 |
| 133. | Flusilazole | 6.39 | 233.1 | 164.9 | 18 | 206.1 | 137 | 20 |
| 134. | 4,4′-DDE | 6.40 | 316 | 246 | 15 | 246 | 176 | 30 |
| 135. | Tricyclazole | 6.41 | 189 | 135.2 | 18 | 189 | 162 | 12 |
| 136. | 2,4′-DDD | 6.42 | 235 | 165.1 | 21 | 235 | 199.1 | 16 |
| 137. | Dieldrin | 6.42 | 277 | 241 | 8 | 263 | 193 | 34 |
| 138. | Fluazifop-P-butyl | 6.42 | 383.1 | 268.2 | 8 | 383.1 | 282.2 | 12 |
| 139. | Chlorfenapyr | 6.46 | 247 | 227.1 | 12 | 137 | 75 | 28 |
| 140. | Ethylan | 6.46 | 223 | 167 | 12 | 223 | 193 | 28 |
| 141. | Nitrofen | 6.49 | 283 | 202 | 10 | 202 | 139.1 | 20 |
| 142. | Chlorthiophos 1 | 6.50 | 256.9 | 165 | 26 | 256.9 | 239 | 12 |
| 143. | Chlorobenzilate | 6.52 | 251 | 139 | 14 | 139 | 111 | 12 |
| 144. | Endrin | 6.55 | 263 | 193 | 30 | 245 | 173 | 25 |
| 145. | Ethion | 6.58 | 231 | 129 | 24 | 153 | 97 | 10 |
| 146. | 4,4′-DDD | 6.59 | 235 | 165.1 | 21 | 235 | 199.1 | 16 |
| 147. | Chlorthiophos 2 | 6.61 | 325 | 269 | 14 | 297 | 269 | 8 |
| 148. | 2,4′-DDT | 6.62 | 235 | 165.1 | 21 | 235 | 199.1 | 16 |
| 149. | Triazophos | 6.65 | 256.9 | 161.7 | 8 | 161 | 134.1 | 8 |
| 150. | cis-Nonachlor | 6.65 | 409 | 263 | 25 | 409 | 300 | 23 |
| 151. | Sulprofos | 6.66 | 322 | 156 | 8 | 156 | 141 | 13 |
| 152. | Carfentrazone ethyl | 6.67 | 340.2 | 312.2 | 10 | 330 | 310.1 | 8 |
| 153. | Endrin aldehyde | 6.68 | 344.9 | 243.1 | 17 | 344.9 | 244.9 | 14 |
| 154. | 4,4′-Methoxychlor olefin | 6.70 | 308 | 223.1 | 30 | 238.1 | 152.1 | 35 |
| 155. | Carbophenothion | 6.70 | 342 | 157 | 10 | 125 | 97 | 6 |
| 156. | Norflurazon | 6.74 | 303 | 145 | 17 | 145 | 95 | 18 |
| 157. | Edifenphos | 6.75 | 310 | 109 | 29 | 173 | 109 | 8 |
| 158. | Lenacil | 6.76 | 153 | 135.7 | 14 | 135.8 | 52.9 | 18 |
| 159. | 4,4′-DDT | 6.61 | 235 | 165.1 | 21 | 235 | 199.1 | 16 |
| 160. | Endosulfan sulfate | 6.81 | 272 | 237 | 10 | 241 | 206 | 8 |
| 161. | Hexazinone | 6.83 | 171.1 | 71.1 | 16 | 128 | 83 | 10 |
| 162. | Piperonyl butoxide | 6.83 | 176.1 | 103 | 26 | 176.1 | 131.1 | 14 |
| 163. | Propargite | 6.83 | 201.1 | 81.1 | 10 | 135.1 | 107 | 14 |
| 164. | Resmethrin | 6.83 | 143.1 | 128 | 10 | 123.2 | 81.2 | 8 |
| 165. | Tebuconazole | 6.85 | 250.1 | 125.2 | 22 | 125 | 89 | 16 |
| 166. | Nitralin | 6.95 | 316.1 | 274 | 8 | 274 | 216.2 | 6 |
| 167. | Captafol | 6.99 | 150.8 | 78.9 | 18 | 79 | 77.1 | 12 |
| 168. | TPP (IS) | 6.99 | 326 | 169 | 30 | 326 | 215 | 20 |
| 169. | Iprodione | 7.01 | 314 | 245 | 14 | 187 | 124.1 | 24 |
| 170. | Bifenthrin | 7.05 | 181.1 | 166.1 | 10 | 165.1 | 115.1 | 30 |
| 171. | Fenpropathrin | 7.05 | 265 | 210 | 8 | 181 | 152 | 24 |
| 172. | Pyridaphenthion | 7.05 | 340.1 | 199.1 | 8 | 199.1 | 77.1 | 26 |
| 173. | Tetramethrin | 7.06 | 164 | 77 | 25 | 164 | 107 | 12 |
| 174. | Bromopropylate | 7.08 | 340.8 | 183 | 15 | 183 | 154.9 | 12 |
| 175. | Azinphos-ethyl | 7.80 | 160 | 77 | 18 | 132 | 77 | 12 |
| 176. | EPN | 7.11 | 169 | 77 | 22 | 169 | 141 | 8 |
| 177. | Phosmet | 7.11 | 160 | 77 | 24 | 160 | 133 | 10 |
| 178. | Methoxychlor | 7.12 | 227 | 141 | 33 | 227 | 169 | 25 |
| 179. | Tebufenpyrad | 7.12 | 333 | 276 | 8 | 276 | 171 | 10 |
| 180. | Endrin ketone | 7.13 | 317 | 101 | 21 | 315 | 279 | 8 |
| 181. | Phenothrin | 7.19 | 183.1 | 168.1 | 12 | 123.2 | 81.1 | 8 |
| 182. | Tetradifon | 7.25 | 227 | 199 | 10 | 159 | 131 | 10 |
| 183. | Pyriproxyfen | 7.28 | 226.1 | 186.1 | 16 | 136.1 | 78 | 20 |
| 184. | Phosalone | 7.29 | 367 | 111 | 36 | 182 | 111 | 14 |
| 185. | Leptophos | 7.31 | 377 | 269 | 37 | 171 | 77 | 18 |
| 186. | λ-Cyhalothrin | 7.31 | 197.1 | 141.1 | 10 | 197 | 91 | 26 |
| 187. | Acrinathrin | 7.32 | 289.1 | 93 | 10 | 289 | 91 | 24 |
| 188. | Pyrazophos | 7.41 | 265.1 | 210.1 | 10 | 221.1 | 193.1 | 10 |
| 189. | Mirex | 7.44 | 274 | 239 | 15 | 272 | 237 | 15 |
| 190. | Fenarimol | 7.46 | 219 | 107.1 | 12 | 139 | 111 | 16 |
| 191. | Azinphos-methyl | 7.46 | 160 | 77 | 18 | 132 | 77 | 12 |
| 192. | Pyraclofos | 7.47 | 360 | 139 | 14 | 194 | 138 | 18 |
| 193. | cis-Permethrin | 7.54 | 183.1 | 153.1 | 12 | 163 | 127.1 | 6 |
| 194. | <a class="cmpd_link" title="View compound information for trans-Permethrin” title=”View compound information for trans-Permethrin” href=”https://ez.restek.com/compound/view/en/61949-77-7/trans-Permethrin”>trans-Permethrin | 7.58 | 183.1 | 153.1 | 12 | 163 | 127.1 | 6 |
| 195. | Pyridaben | 7.62 | 147 | 117 | 20 | 147 | 119 | 8 |
| 196. | Coumaphos | 7.66 | 362 | 109 | 17 | 210 | 182 | 10 |
| 197. | Fluquinconazole | 7.66 | 340 | 108 | 42 | 340 | 298 | 14 |
| 198. | Prochloraz | 7.67 | 308 | 70 | 13 | 180 | 138 | 12 |
| 199. | Cyfluthrin | 7.73 | 226.1 | 199 | 8 | 226 | 151 | 30 |
| 200. | Cypermethrin | 7.83 | 181.1 | 127 | 28 | 181.1 | 152.1 | 20 |
| 201. | Flucythrinate 1 | 7.86 | 199.1 | 157.1 | 8 | 157.1 | 107.1 | 13 |
| 202. | Acequinocyl | 7.89 | 342.4 | 188.1 | 14 | 188 | 132 | 10 |
| 203. | Etofenprox | 7.90 | 163.1 | 107 | 18 | 163.1 | 135.1 | 10 |
| 204. | Flucythrinate 2 | 7.91 | 199.1 | 157.1 | 8 | 157.1 | 107.1 | 13 |
| 205. | Fluridone | 8.03 | 328.1 | 127.1 | 40 | 328.1 | 259 | 26 |
| 206. | Fenvalerate | 8.10 | 169.1 | 127.1 | 10 | 167.1 | 125 | 8 |
| 207. | Tau-fluvalinate | 8.15 | 252 | 55 | 17 | 250.1 | 55 | 23 |
| 208. | Fenvalerate | 8.16 | 169.1 | 127.1 | 10 | 167.1 | 125 | 8 |
| 209. | Deltamethrin | 8.33 | 252.9 | 93.1 | 19 | 181.1 | 152 | 20 |
Conditions
| Column | LPGC Rtx-5ms column kit, includes 15 m x 0.53 mm ID x 1.00 μm analytical column w/1 m x 0.53 mm ID integrated transfer line and 5 m x 0.18 mm ID Hydroguard restrictor factory connected via SilTite connector (cat.# 11800) |
|---|---|
| Standard/Sample | |
| GC multiresidue pesticide kit (cat.# 32562) | |
| Triphenylphosphate (cat.# 33258) | |
| Anthracene | |
| Diluent: | Acetonitrile |
| Conc.: | 1 µg/mL |
| Injection | |
| Inj. Vol.: | 1 µL splitless (hold 0.5 min) |
| Liner: | Topaz 4.0 mm ID single taper inlet liner w/ wool (cat.# 23447) |
| Inj. Temp.: | 250 °C |
| Oven | |
| Oven Temp.: | 80 °C (hold 1 min) to 320 °C at 35 °C/min (hold 5 min) |
| Carrier Gas | He, constant flow |
| Flow Rate: | 2 mL/min |
| Detector | TSQ 8000 | |
|---|---|---|
| SIM Program: | 35-550 m/z | |
| Transfer Line Temp.: | 290 °C | |
| Analyzer Type: | Quadrupole | |
| Source Temp.: | 325 °C | |
| Solvent Delay Time: | 2 min | |
| Tune Type: | PFTBA | |
| Ionization Mode: | EI | |
| Instrument | Thermo Scientific TSQ 8000 Triple Quadrupole GC-MS | |
| Sample Preparation | Sample preparation: 10 g of homogenized strawberries were fortified with the internal standards at 20 ppb and then extracted with 10 mL of acetonitrile and QuEChERS EN 15662 salts (cat.# 25850). After centrifugation, 1 mL of supernatant was added to a 2 mL dSPE vial containing magnesium sulfate and PSA (cat.# 26124) for cleanup. The cleaned extract was spiked with GC multiresidue pesticide mix at 1 ppm. | |

GC_FS0591
Peaks
| Peaks | tR (30m) | tR (LPGC) | Conc. (ppm) | |
|---|---|---|---|---|
| 1. | Methamidophos | 8.16 | 3.488 | 6 |
| 2. | Dichlorvos | 8.52 | 3.582 | 6 |
| 3. | Mevinphos | 11.51 | 4.317 | 6 |
| 4. | Acephate | 11.62 | 4.356 | 6 |
| 5. | o-Phenylphenol | 12.82 | 4.711 | 6 |
| 6. | Omethoate | 13.98 | 4.979 | 6 |
| 7. | Dimethoate | 15.92 | 5.453 | 6 |
| 8. | gamma-BHC | 16.43 | 5.615 | 6 |
| 9. | Diazinon | 16.88 | 5.615 | 6 |
| 10. | Chlorothalonil | 17.18 | 5.778 | 6 |
| 11. | Vinclozolin | 18.07 | 5.95 | 6 |
| 12. | Carbaryl | 18.21 | 6.023 | 6 |
| 13. | Metalaxyl | 18.39 | 6.023 | 6 |
| Peaks | tR (30m) | tR (LPGC) | Conc. (ppm) | |
|---|---|---|---|---|
| 14. | Pirimiphos methyl | 18.80 | 6.101 | 6 |
| 15. | Methiocarb | 18.76 | 6.118 | 6 |
| 16. | Malathion | 19.02 | 6.149 | 6 |
| 17. | Dichlofluanid | 18.96 | 6.194 | 6 |
| 18. | Fenthion | 19.23 | 6.238 | 6 |
| 19. | Chlorpyrifos | 19.27 | 6.238 | 6 |
| 20. | Cyprodinil | 19.89 | 6.4 | 6 |
| 21. | Thiabendazole | 20.18 | 6.522 | 6 |
| 22. | Captan | 20.29 | 6.545 | 6 |
| 23. | Folpet | 20.44 | 6.584 | 6 |
| 24. | Imazalil | 21.34 | 6.767 | 6 |
| 25. | Myclobutanil | 21.68 | 6.838 | 6 |
| 26. | Endrin | 22.09 | 7.033 | 6 |
| Peaks | tR (30m) | tR (LPGC) | Conc. (ppm) | |
|---|---|---|---|---|
| 27. | 4,4′-DDT | 23.34 | 7.29 | 6 |
| 28. | Fenhexamid | 23.34 | 7.317 | 6 |
| 29. | Endosulfan sulfate | 23.29 | 7.317 | 6 |
| 30. | Propargite | 23.73 | 7.328 | 6 |
| 31. | Iprodione | 24.25 | 7.469 | 6 |
| 32. | Bifenthrin | 24.50 | 7.511 | 6 |
| 33. | Fenpropathrin | 24.67 | 7.572 | 6 |
| 34. | Dicofol | 25.31 | 7.647 | 6 |
| 35. | Phosalone | 25.34 | 7.802 | 6 |
| 36. | Azinphos-methyl | 26.63 | 7.837 | 6 |
| 37. | cis-Permethrin | 26.78 | 8.114 | 6 |
| 38. | trans-Permethrin | 26.93 | 8.152 | 6 |
| 39. | Coumaphos | 27.22 | 8.235 | 6 |
| 40. | Deltamethrin | 29.74 | 9.081 | 6 |
Conditions
| Column | See notes |
|---|---|
| Standard/Sample | QuEChERS performance standards kit (cat.# 31152) |
| Diluent: | Acetonitrile |
| Conc.: | 6 µg/mL |
| Injection | |
| Inj. Vol.: | 1 µL splitless (hold 0.5 min) |
| Liner: | Topaz, single taper inlet liner, 4.0 mm x 6.5 x 78.5 (cat.# 23303) |
| Inj. Temp.: | 250 °C |
| Carrier Gas | He |
| Detector | MS |
|---|---|
| Mode: | Scan |
| Transfer Line Temp.: | 280 °C |
| Analyzer Type: | Quadrupole |
| Source Temp.: | 230 °C |
| Quad Temp.: | 150 °C |
| Electron Energy: | 70 eV |
| Tune Type: | PFTBA |
| Ionization Mode: | EI |
| Instrument | Agilent 7890A GC & 5975C MSD |
| Sample Preparation | The standards were combined and diluted with acetonitrile to 6 ppm; analyzed in a 2 mL, short-cap, screw-thread vial (cat.# 21143); and capped with a short-cap, screw-vial closure (cat.# 24495). |
| Notes | Conventional (30 m) Analysis: Column: Rtx-5ms, 30 m, 0.25 mm ID, 0.25 µm (cat.# 12623) Temp. program: 70 °C (hold 1 min) to 330 °C at 8 °C/min (hold 6.5 min) Flow: 1.4 mL/min Scan start time: 6.5 min Scan range: 45-550 amu Scan rate: 5.5 scans/sec LPGC-MS Analysis: Column: LPGC Rtx-5ms column kit, includes 10 m x 0.32 mm ID x 1.00 µm Rtx-5ms analytical column w/5 m x 0.15 mm ID Hydroguard restrictor factory connected via SilTite connector (cat.# 11802) Temp. program: 70 °C (hold 1 min) to 320 °C at 35 °C/min (hold 5 min) Flow: 1.1 mL/min Scan start time: 2 min Scan range: 45-550 amu Scan rate: 5.5 scans/sec The injections were performed on different instruments under different head pressures, resulting in different analyte responses. |

