FAQ

Sometimes when my original method contains a gradient there is a mismatch between the number of steps in my “Original” and “Translation” methods. Why does this happen and when should I expect it to happen?

30 Jul 2025

This occurs because of a significant change to either column diameter or dwell volume (e.g., when transferring a method from one LC system to another LC system). In some instances, these differences can effectively act as a hold time at the initial %B. Depending on the scenario, an isocratic hold may be needed or not needed in the translated method for equivalence.

  • Scenarios when an initial isocratic hold will be added to the translated method:
    • The dwell volume of the original method is greater than the dwell volume of the translated method.
    • Going from a narrow bore (ID) column to a wide bore (ID) column.
    • Going from a shorter (length) column to a longer (length) column with the ID of the columns changing.
    • Any combination of the above.
  • Scenarios when an initial isocratic hold is not needed in the translated method:
    • The dwell volume of the translated method is greater than the dwell volume of the original method.
    • Going from a wide bore (ID) column to a narrow bore (ID) column.
    • Going from a long (length) column to a shorter (length) column with the ID of the columns changing.
    • Any combination of the above.

Autori

  • Restek Corporation

    Restek is a leading provider of chromatography columns, accessories, and certified reference materials. Trust Restek for reliable, high-quality analytical solutions.

    View Articles
GNOT5139