Articoli

Fast Screening and Confirmation of Gamma-Hydroxybutyrate (GHB) in Urine

26 Oct 2020

feature CFAR3233

Maximize your analytical options with this versatile GHB extraction method. No derivatization means faster sample preparation. Extracts are amenable to both liquid injection GC-FID and headspace GC-MS methods.

Gamma-hydroxybutyrate (GHB) and its precursor, gamma-butyrolactone (GBL), are controlled substances associated with drug-facilitated sexual assault. Criminal cases often hinge on lab results, which can include screening urine samples and then quantifying GHB using GC/MS. In its native state, GHB is extremely difficult to chromatograph and must be analyzed as a trimethylsilyl derivative or converted to GBL. The headspace (HS) procedure described here (adapted from an FBI Chemistry Unit method) eliminates time-consuming derivatization [1]. This procedure reduces sample preparation time and minimizes both column contamination from derivatization reagents and contamination from sample matrix caused by liquid injections.

Eliminate Derivatization and Reduce System Contamination

Samples were spiked in urine and extracted according to the procedure detailed in Table I using alpha-methylene-gamma-butyrolactone (AMGB) as an internal standard. GHB is converted to GBL with sulfuric acid, eliminating the need for derivatization (Figure 1). Note the unconverted sample shows comparable levels of GBL and AMGB whereas GBL levels in the converted sample are significantly higher due to the conversion of GHB to GBL.

Table I:  Extraction procedure for GHB and GBL.

  1. Label two screw-top test tubes per specimen. One for total GHB, the other for GBL only.
  2. Add 1 mL of sample (urine) to each tube.
  3. Add 50 µL of AMGB (internal standard) to each tube.
  4. Add 150 µL concentrated sulfuric acid only to tubes used for analysis of total GHB.
  5. Vortex all tubes and allow them to sit 5 minutes.
  6. Add 5 mL methylene chloride to each tube. Shake 10 minutes to extract.
  7. Centrifuge samples at 3000 rpm for 5 minutes.
  8. Transfer bottom (methylene chloride) layer to a clean test tube for drying.
  9. Concentrate samples to ~100 µL at 30 °C under nitrogen.
  10. For headspace analysis, inject 15 µL of sample into a capped headspace vial. Or, for liquid injection, transfer extract to a limited volume insert.

Reliably Screen Samples Using Existing Blood Alcohol Testing Set-Up

Headspace injections (using the total vaporization technique) of the final urine extracts were screened by GC-FID using an Rtx-BAC1 column in a blood alcohol headspace GC system. This system is commonly used in clinical/forensic labs, eliminating the need for additional equipment. Excellent linear response was obtained from both unconverted (r2 = 0.9992, 10-100 µg/mL 4-point curve) and converted GHB in matrix (r2 = 0.9910, 20-200 µg/mL 4-point curve) with AMBG at 50 μg/mL.

Fast, Definitive Confirmation Analysis by Headspace GC/MS

Positive screening results were quickly confirmed on an Rtx-5MS column by headspace GC-MS; several quantification and qualifier ions were identified for each compound (GBL: 42, 56, 86; AMBG: 40, 68, 98). Again, excellent linearity was achieved (Figure 2) and analysis time was less than 7 minutes (Figure 3).

In summary, the versatile extraction and headspace method shown here saves lab time and minimizes contamination by eliminating the need for derivatization and by reducing matrix effects. Rapid screening is accomplished on commonly used blood alcohol GC columns, allowing labs to reduce costs by using existing equipment. Confirmation testing using the Rtx-5MS column, provides the definitive results needed in court with a fast analysis time of less than 7 minutes.

Figure 1: GHB can be converted to GBL for analysis, saving time by eliminating derivatization.
Comparison of Converted and Unconverted GHB on Rtx-BAC1

GC_CF01039

Peaks

Peaks
1.GBL
2.AMGB (IS)

Conditions

ColumnRtx-BAC1, 30 m, 0.32 mm ID, 1.80 µm (cat.# 18003)
Standard/Sample
Conc.:50 µg/mL GHB, GBL, and AMGB (IS) in urine
A: Unconverted
B: Converted with sulfuric acid
Injection
Inj. Vol.:1000 µL headspace-syringe split (split ratio 10:1)
Liner:1 mm Split
Inj. Temp.:200 °C
Headspace-Syringe
Sample Temp.:100 °C
Sample Equil. Time:10 min
Oven
Oven Temp.:50 °C (hold 3 min) to 150 °C at 20 °C/min (hold 7 min)
Carrier GasHe, constant flow
Flow Rate:1.0 mL/min
DetectorFID @ 240 °C
Make-up Gas Flow Rate:40 mL/min
NotesLiner cat.# 20972 was used to produce this chromatogram, but it has since been discontinued. We recommend cat.# 20973 [or cat.# 23333] as an alternative. For assistance choosing a replacement for this application, contact Restek Technical Service or your local Restek representative.
Figure 2: GHB (Analyzed as GBL) Confirmation Method Calibration Curve for Headspace GC-MS Analysis (10-300 µg/mL in Urine)
figure article CFAR3233 01
Figure 3: Confirmation Headspace GC-MS Analysis of 300 µg/mL Converted GHB (Analyzed as GBL) Standard in Urine
GHB (Analyzed as GBL) in Urine on Rtx-5MS

GC_CF01038

Peaks

Peaks
1.GBL
2.AMGB (IS)

Conditions

ColumnRtx-5MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 12623)
Standard/Sample300 µg/mL GHB, 50 µg/mL AMGB in urine, converted with sulfuric acid
Injection
Inj. Vol.:1000 µL headspace-syringe split (split ratio 10:1)
Liner:1 mm Split
Inj. Temp.:200 °C
Headspace-Syringe
Sample Temp.:100 °C
Sample Equil. Time:10 min
Oven
Oven Temp.:50 °C (hold 3 min) to 150 °C at 20 °C/min (hold 7 min)
Carrier GasHe, constant flow
Flow Rate:1.0 mL/min
DetectorMS
Mode:Scan
Transfer Line Temp.:280 °C
Ionization Mode:EI
Scan Range:20-150 amu
NotesLiner cat.# 20972 was used to produce this chromatogram, but it has since been discontinued. We recommend cat.# 20973 [or cat.# 23333] as an alternative. For assistance choosing a replacement for this application, contact Restek Technical Service or your local Restek representative.

References

  1. M.A. LeBeau, M.A. Montgomery, M.L Miller, S. G. Burmeister, Analysis of biofluids for gamma-hydroxybutyrate (GHB) and gamma-butyrolactone (GBL) by headspace GC-FID and GC-MS, J. Anal. Toxicol. 24 (6) (2000) 421-428. https://doi.org/10.1093/jat/24.6.421

Prodotti Menzionati


Colonna capillare GC Rtx-BAC1, 30 m, ID 0,32 mm, 1,80 µm
Colonna capillare GC Rtx-5ms, 30 m, ID 0,25 mm, 0,25 µm

Autori

  • Amanda Rigdon is an analytical chemist with more than 21 years of experience in chromatography and mass spectrometry. Throughout her career, she has worked in a variety of analytical fields, including clinical forensics, food safety, environmental analysis, and cannabis testing. Her experience spans method development, validation, troubleshooting, and laboratory operations using both GC and LC methodologies. Amanda also served as Chief Scientific Officer of an ISO/IEC 17025-accredited cannabis testing laboratory, where she gained extensive experience in quality systems, regulatory compliance, and laboratory management. She is currently a Principal Scientist in the GC Applications Group at Restek, where she enjoys helping customers solve analytical challenges and sharing practical chromatography knowledge through technical support, presentations, and educational content.

    View Articles
  • Kristi Sellers
CFAR3233
Versione: A