Articoli

Rapid LPGC-MS Analysis of Arylamines Increases Throughput and Reduces Helium Use

Featured Application: Arylamines on LPGC Rxi-35Sil MS

03 Oct 2025

feature GNFA4235
  • Analyze arylamines 3.3x faster than conventional GC-MS.
  • Cut costs by reducing helium consumption 81%.
  • Pre-connected LPGC column kit prevents leaks, simplifies installation, and allows large-volume injection.

Azo dyes are synthetically produced colorants that are commonly used in the food, cosmetic, pharmaceutical, and textile industries. They are present in a wide variety of consumer products, and some can produce carcinogenic or mutagenic arylamines as breakdown products. Due to concerns about human health and exposure, effective methods are needed for the analysis of arylamines. Testing is usually performed by conventional GC-MS or LC-MS, but long analysis times limit sample throughput. 

A much faster alternative, low-pressure GC-MS (LPGC-MS) is a paired-column technique that can bring significant speed gains to labs analyzing arylamines. This approach uses the MS vacuum to lower pressure within the analytical column, thus increasing speed. For example, in the analysis of arylamines (22 of which are regulated by EU REACH legislation [1]) shown in Figure 1, the run time was 3.3x faster than the conventional method, and it also used 81% less helium, which provides a substantial cost savings. 

LPGC-MS only works because the inlet pressure is maintained by pairing the analytical column to a narrow restrictor column. However, due to the difference in tubing diameters, it can be difficult to make a robust connection manually, which has limited adoption of the technique. LPGC column kits from Restek solve this problem because they are factory connected and individually leak checked, which ensures a leak-free connection and simplifies setup. Visit www.restek.com/lpgc to learn more about this powerful technique.

References

  1. Regulation (EC) No. 1907/2006 of the European Parliament and of the Council of 18 December 2006 Concerning the Registration, Evaluation, Authorization and Restriction of Chemicals (REACH). https://osha.europa.eu/en/legislation/directives/regulation-ec-no-1907-2006-of-the-european-parliament-and-of-the-council.
Figure 1: Arylamines on LPGC Rxi-35Sil MS Compared to Conventional GC-MS Analysis
Comparison of Conventional and LPGC-MS Analysis of REACH Carcinogenic Arylamines

GC_GN1244

Peaks

PeakstR (30 m)tR (LPGC)Conc. (ppm)Quant Ion
1.o-Toluidine3.2171.2072107
2.o-Anisidine4.421.5892123
3.4-Chloroaniline4.8321.712127
4.p-Cresidine5.5191.922137
5.2,4,5-Trimethylaniline5.5821.9352135
6.3-Chloro-o-toluidine5.9022.0262141
7.4-Chloro-o-toluidine6.0252.0612141
8.2,4-Diaminotoluene7.6372.5232122
9.2,4-Diaminoanisole8.892.8722138
10.2-Naphthylamine9.7733.1122143
11.2-Aminobiphenyl9.8923.1482169
12.2-Amino-4-nitrotoluene10.5373.3252152
PeakstR (30 m)tR (LPGC)Conc. (ppm)Quant Ion
13.4-Aminobiphenyl12.1543.7672169
14.p-Aminoazobenzene15.954.8052197
15.4,4′-Oxydianiline16.6134.992200
16.4,4′-Diaminodiphenylmethane16.7145.0162198
17.Benzidine16.7875.0342184
18.o-Aminoazotoluene17.535.242225
19.3,3′-Dimethyl-4,4′-diaminodiphenylmethane18.235.4362226
20.3,3′-Dimethylbenzidine18.4835.5042212
21.4,4’-Thiodianiline19.4395.7652216
22.3,3′-Dichlorobenzidine19.7955.8642252
23.4,4′-Methylenebis(2-chloroaniline)19.7955.8642266
24.3,3′-Dimethoxybenzidine19.9055.8982244

Conditions

ColumnSee notes
Standard/Sample
AccuStandard carcinogenic aryl amine mix (AE-00049-R1-10X)
AccuStandard 2,4-diaminoanisole (ALR-070S-R2)
Diluent:Ethyl acetate
Conc.:20 µg/mL
Injection
Inj. Vol.:1 µL split (split ratio 10:1)
Liner:Topaz, splitless, single taper inlet liner, 4.0 mm x 6.5 x 78.5 (cat.# 23303)
Inj. Temp.:280 °C
Carrier GasHe
DetectorMS
Mode:Scan
Transfer Line Temp.:300 °C
Analyzer Type:Quadrupole
Source Temp.:230 °C
Quad Temp.:150 °C
Electron Energy:70 eV
Tune Type:PFTBA
Ionization Mode:EI
InstrumentAgilent 7890B GC & 5977A MSD
Sample PreparationThe standards were diluted with ethyl acetate to 20 ppm; analyzed in a 2 mL, short-cap, screw-thread vial (cat.# 21143); and capped with a short-cap, screw-vial closure (cat.# 24495).
Notes Conventional (30 m) Analysis:
Column: Rxi-35Sil MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13823)
Temp. program: 200 °C (hold 0.5 min) to 320 °C at 9.5 °C/min (hold 5 min)
Flow: 2.0 mL/min
Scan start time: 2 min
Scan range: 30-300 amu
Scan rate: 10 scans/sec

LPGC-MS Analysis:
Column: LPGC Rxi-35Sil MS column kit, includes 10 m x 0.32 mm ID x 0.25 μm Rxi-35Sil MS analytical column and 5 m x 0.15 mm ID Rxi restrictor factory connected via SilTite connector (cat.# 11806)
Temp. program: 100 °C (hold 0.5 min) to 300 °C at 35 °C/min (hold 5 min)
Flow: 0.9 mL/min
Scan start time: 1 min
Scan range: 35-300 amu
Scan rate: 9.7 scans/sec

Compound list based on EU legislation of Registration, Evaluation, Authorisation and Restriction of Chemicals (REACH) Appendix 8 (https://reachonline.eu/reach/en/appendix-8.html). The 3-chloro-o-toluidine and 2-aminibiphenyl compounds are not part of the list.

Products Mentioned


Kit colonna LPGC Rxi-35Sil MS
Kit inserti per forno GC Accelerator, per GC Agilent 5890, 6890, 7890 e 8890
Topaz Inlet Liner a Cono Singolo, 4,0 mm x 6,5 x 78,5 per GC Agilent, con Lana di Quarzo, Deattivazione Premium, 5 pz.
Vial con tappo corto con indicatore graduato, filettatura 9-425, 2,0 mL, 9 mm, 12 x 32 (solo vial), ambrato, 1000 pz.
Tappi corti a vite, polipropilene, filettatura, setti in PTFE/silicone/PTFE, neri, preassemblati, 2,0 mL, 9 mm, 100 pz.

Author

  • Restek Corporation

    Restek is a leading provider of chromatography columns, accessories, and certified reference materials. Trust Restek for reliable, high-quality analytical solutions.

    View all posts
GNFA4235