- Sample Matrix: serum
- Sample Preparation: online solid-phase extraction
- Detection Technique: LC-MS/MS
Online Solid Phase Extraction-High Performance Liquid Chromatography-Turbo Ion Spray-Tandem Mass Spectrometry (online SPE-HPLC-TISMS/MS)
Method CDC 6304.09 is used for the analysis of 15 PFAS in serum. It combines online solid-phase extraction sample preparation with turbo ion spray LC-MS/MS analysis. Notably, PFOS is analyzed separately as n-perfluorooctane sulfonate (n-PFOS) and the sum of perfluoromethylheptane sulfonate isomers (sm-PFOS). Similarly, PFOA is analyzed as separately as n-perfluorooctanoate (n-PFOA) and the sum of branched perfluorooctanoate isomers (sb-PFOA). The full analyte list and recommended products for PFAS analysis serum by CDC 6304.09 are provided below. In addition, you can register and directly compare analyte lists and matrices across PFAS analysis methods here or visit www.restek.com/PFAS to explore our technical resources for PFAS analysis.
Analyte List for CDC 6304.09 for PFAS Analysis
PFAS acronyms, compound names, and CAS numbers vary across methods and guidelines, and they also depend on compound form (e.g., neutral vs. salt). The following list uses standardized terminology to allow easy comparison across the methods and guidelines in our comprehensive guide. Please refer to CDC 6304.09 for method-specific nomenclature.
| Acronym | Compound Name | CAS# |
| NEtFOSAA | N-ethyl perfluorooctanesulfonamidoacetic acid | 2991-50-6 |
| NMeFOSAA | N-methyl perfluorooctanesulfonamidoacetic acid | 2355-31-9 |
| PFDA | Perfluorodecanoic acid | 335-76-2 |
| PFHpA | Perfluoroheptanoic acid | 375-85-9 |
| PFHxS | Perfluoro-1-hexanesulfonic acid | 355-46-4 |
| PFHxA | Perfluorohexanoic acid | 307-24-4 |
| PFNA | Perfluorononanoic acid | 375-95-1 |
| PFOS | Perfluorooctanesulfonic acid | 1763-23-1 |
| PFOA | Perfluorooctanoic acid | 335-67-1 |
| PFUnA | Perfluoroundecanoic acid | 2058-94-8 |
| HFPO-DA, GenX | Hexafluoropropylene oxide dimer acid | 13252-13-6 |
| PFHpS | Perfluoro-1-heptanesulfonic acid | 375-92-8 |
| FOSA | Perfluoro-1-octanesulfonamide | 754-91-6 |
| 9Cl-PF3ONS | 9-Chlorohexadecafluoro-3-oxanonane-1-sulfonic acid (F-53B Major) | 756426-58-1 |
| ADONA | 4,8-Dioxa-3H-perfluorononanoic acid | 919005-14-4 |
Recommended Products for CDC 6304.09 for PFAS Analysis
Analytical LC Columns:
Delay Columns:
PFAS delay column: 5 µm, 50 x 3.0 mm (cat.# 580959 [custom product])
Note that this CDC method is based on automated online SPE using C8 or C18 SPE materials. The product listed here is recommended for off-line SPE.
Vacuum Manifolds:
Vacuum Pumps:
Note that this CDC method is based on automated online SPE using C8 or C18 SPE materials. The products listed here are recommended for off-line SPE.
General Lab Supplies:
Discover More Solutions for PFAS Analysis!
Start Saving
Time Today
Develop PFAS Methods in Minutes with our Pro EZLC Chromatogram Modeler
Start Saving Time Today
Develop PFAS Methods in Minutes with our Pro EZLC Chromatogram Modeler


