Articoli

Multi-Mycotoxin Method for Alternaria Toxins, Ergot Alkaloid Epimers, and Other Major Mycotoxins in Food

Featured Application: Mycotoxins on Raptor Biphenyl

17 Aug 2022

feature FSFA3904
  • Increase lab productivity by analyzing emerging mycotoxins using the same method as priority regulated mycotoxins.
  • No separate run under high pH conditions—Raptor Biphenyl column allows quantitative analysis of all groups under MS-friendly, acidic conditions.
  • Simple, universal sample prep saves time and does not require further purification procedures.

As more mycotoxins come under regulatory purview, new methods are needed to help food safety labs continue to operate efficiently. Comprehensive, multi-mycotoxin methods are an attractive alternative to separate methods for different analyte lists, but they can be difficult to develop due to the wide range of chemical characteristics among mycotoxin classes. In particular, the Alternaria toxins and ergot alkaloids create additional challenges for method developers. These emerging mycotoxin food contaminants are unique in that, when analyzed on a C18 column, high pH conditions must be used to obtain acceptable peak shape for the Alternaria toxins and adequate separation of the ergot alkaloid epimers. Use of high pH conditions is stressful for LC columns and not suitable for analyzing other classes of mycotoxins, so another approach is required for a truly comprehensive method.

Here, we show the simultaneous analysis of Alternaria toxins, ergot alkaloid epimers, and other major mycotoxins using a simple sample preparation procedure and a Raptor Biphenyl column run under acidic conditions. Under these column- and MS-friendly conditions, accurate quantitative results were obtained in a fast, 11-minute total cycle time run for 37 regulated and emerging mycotoxins (including the Alternaria toxins and all six essential ergot alkaloids and their epimers). In a separate study, method performance was verified in wheat baby cereal, peanut, tomato puree, and blended flour to demonstrate the applicability of this method to a wide range of food types [1]. Quantitation was performed using matrix-matched standards because isotopically labeled standards were not available for most mycotoxins. Use of this Raptor Biphenyl-based, multi-mycotoxin method provides food safety labs with a means of significantly improving productivity compared to using separate methods for different mycotoxin panels.

Figure 1: Increase productivity using one multi-mycotoxin method for concurrent analysis of 37 major mycotoxins, including Alternaria toxins and ergot alkaloid epimers (blended flour sample).
Alternaria Toxins, Ergot Alkaloid Epimers, and Other Major Mycotoxins in Fortified Flour Sample on Raptor Biphenyl by LC-MS/MS

LC_FS0550

Peaks

PeakstR (min)Precursor IonProduct Ion 1Product Ion 2
1.Nivalenol 0.92295.1137.191.0
2.Patulin1.03155.099.081.0
3.Ergometrine1.27326.2223.2208.1
4.Deoxynivalenol1.30297.2231.0249.0
5.Ergometrinine1.83326.2223.2208.1
6.Fusarenon-X1.98355.1137.1247.1
7.15-Acetyldeoxynivalenol3.14339.2137.1321.2
8.3-Acetyldeoxynivalenol3.21339.2213.1231.1
9.Tenuazonic acid4.22198.1125.0153.1
10.Altenuene4.70293.2257.1275.2
11.Alternariol5.30259.0185.1130.0
12.Citrinin5.43251.2233.1205.1
13.Ergosine5.47548.4208.1223.2
14.Fumonisin B15.63722.5352.3334.2
15.Ergosinine5.67548.4208.1223.2
16.Diacetoxyscirpenol5.73384.2247.1307.2
17.Ergotamine5.90582.4223.2268.2
18.Ergocornine6.03562.4268.2223.2
PeakstR (min)Precursor IonProduct Ion 1Product Ion 2
19.Ergotaminine6.13582.4223.2268.2
20.HT-26.20447.2345.1285.1
21.Fumonisin B36.32706.4336.2318.3
22.Ergocryptine6.32576.4268.2223.2
23.Ergocristine6.56610.4223.2592.4
24.Fumonisin B26.68706.4336.2318.3
25.Tentoxin6.70415.2312.2302.2
26.α-Zearalenol6.96303.1285.1175.0
27.Ergocorninine7.07562.4268.2223.2
28.T-27.14489.2387.1245.1
29.Aflatoxin G27.15331.2189.0313.0
30.Ergocryptinine7.31576.4268.2223.2
31.Ergocristinine7.53610.4223.2592.4
32.Aflatoxin G17.62329.1199.7243.0
33.Zearalenone7.65319.2283.1187.0
34.Alternariol monomethylether7.69273.0199.1128.0
35.Aflatoxin B27.81315.1287.0259.0
36.Aflatoxin B18.20313.2241.1284.9
37.Ochratoxin A8.31404.1239.0358.0

Conditions

ColumnRaptor Biphenyl (cat.# 9309A12)
Dimensions:100 mm x 2.1 mm ID
Particle Size:2.7 µm
Pore Size:90 Å
Guard Column:Raptor Biphenyl EXP guard column cartridge 5 mm, 2.1 mm ID, 2.7 µm (cat.# 9309A0252)
Temp.:60 °C
Standard/Sample
Aflatoxins standard (cat.# 34121)
Ochratoxin A standard (cat.# 34122)
Diluent:50:50 Water:methanol
Conc.:6.25 ng/mL final concentration after sample preparation
Inj. Vol.:5 µL
Mobile Phase
A:Water, 0.05% formic acid
B:Methanol, 0.05% formic acid
Time (min)Flow (mL/min)%A%B
0.000.47525
5.000.45050
9.000.40100
9.010.47525
11.00.47525
DetectorMS/MS
Ion Mode:ESI+
Mode:MRM
InstrumentUHPLC
Sample PreparationA blended flour was prepared by mixing white rice flour (75%); brown rice flour (5%); millet flour (5%); oat flour (5%); all-purpose wheat flour (5%); and all-purpose, gluten-free flour (5%). Two grams of the flour sample were weighed into a 50-mL polypropylene centrifuge tube (cat.# 25846) and fortified at 50 µg/kg for all analytes with a stock standard solution. After sitting at room temperature for 10 minutes, 16 mL of extraction solution (80:20 acetonitrile:water) containing 0.5% formic acid were added, and the tube was stirred to create a homogenous suspension. The extraction was carried out by shaking horizontally on a digital pulse mixer (Glas-Col LLC, Terre Haute, IN) at 800 rpm for 20 minutes. After centrifuging for 5 minutes at 4000 rpm, 1 mL of extract was evaporated to dryness at 45 °C under a gentle stream of nitrogen. The dried extract was reconstituted with 1 mL of 50:50 water:methanol solution, and a 0.4 mL aliquot was transferred to and filtered using a Thomson SINGLE StEP filter vial with a 0.2 µm PTFE filter cat.# 25874. Five µL of the filtered solution was injected for the LC-MS/MS analysis.

NotesThe chromatogram shows peaks with the MS transition of product ion 1.

Note that method development work demonstrated that whenever a new column is installed it must be rinsed and maintained under mobile phase overnight to ensure an acceptable peak shape for tenuazonic acid.

Want even better performance when analyzing metal-sensitive compounds? Check out Inert LC columns at www.restek.com/inert.

References

  1. H. Liang, J. York, J. Konschnik, H. Majer, J. Steimling, Comprehensive mycotoxin analysis: simultaneous determination of Alternaria toxins, ergot alkaloid epimers, and other major mycotoxins in various food matrices by LC-MS/MS, Restek Corporation, 2022.

Learn more at www.restek.com/inert

teaser product lc columns inert lc 02

Autori

  • Restek Corporation

    Restek is a leading provider of chromatography columns, accessories, and certified reference materials. Trust Restek for reliable, high-quality analytical solutions.

    Guarda tutti i post
FSFA3904
Versione: A