Articoli

PFAS Analysis by CDC 6304.09 for Serum

03 Oct 2025

feature CFAR4291
  • Sample Matrix: serum
  • Sample Preparation: online solid-phase extraction
  • Detection Technique: LC-MS/MS

Online Solid Phase Extraction-High Performance Liquid Chromatography-Turbo Ion Spray-Tandem Mass Spectrometry (online SPE-HPLC-TISMS/MS)

Method CDC 6304.09 is used for the analysis of 15 PFAS in serum. It combines online solid-phase extraction sample preparation with turbo ion spray LC-MS/MS analysis. Notably, PFOS is analyzed separately as n-perfluorooctane sulfonate (n-PFOS) and the sum of perfluoromethylheptane sulfonate isomers (sm-PFOS). Similarly, PFOA is analyzed as separately as n-perfluorooctanoate (n-PFOA) and the sum of branched perfluorooctanoate isomers (sb-PFOA). The full analyte list and recommended products for PFAS analysis serum by CDC 6304.09 are provided below. In addition, you can register and directly compare analyte lists and matrices across PFAS analysis methods here or visit www.restek.com/PFAS to explore our technical resources for PFAS analysis.

Analyte List for CDC 6304.09 for PFAS Analysis

PFAS acronyms, compound names, and CAS numbers vary across methods and guidelines, and they also depend on compound form (e.g., neutral vs. salt). The following list uses standardized terminology to allow easy comparison across the methods and guidelines in our comprehensive guide. Please refer to CDC 6304.09 for method-specific nomenclature.

AcronymCompound NameCAS#
NEtFOSAAN-ethyl perfluorooctanesulfonamidoacetic acid2991-50-6
NMeFOSAAN-methyl perfluorooctanesulfonamidoacetic acid 2355-31-9
PFDAPerfluorodecanoic acid335-76-2
PFHpAPerfluoroheptanoic acid375-85-9
PFHxSPerfluoro-1-hexanesulfonic acid355-46-4
PFHxAPerfluorohexanoic acid307-24-4
PFNAPerfluorononanoic acid375-95-1
PFOSPerfluorooctanesulfonic acid1763-23-1
PFOAPerfluorooctanoic acid335-67-1
PFUnAPerfluoroundecanoic acid2058-94-8
HFPO-DA, GenXHexafluoropropylene oxide dimer acid13252-13-6
PFHpSPerfluoro-1-heptanesulfonic acid375-92-8
FOSAPerfluoro-1-octanesulfonamide754-91-6
9Cl-PF3ONS9-Chlorohexadecafluoro-3-oxanonane-1-sulfonic acid (F-53B Major)756426-58-1
ADONA4,8-Dioxa-3H-perfluorononanoic acid919005-14-4

Recommended Products for CDC 6304.09 for PFAS Analysis

Analytical LC Columns:


Colonna HPLC Raptor C18, 2,7 µm, 50 x 2,1 mm
Colonna LC Force C18, 1,8 µm, 50 x 2,1 mm

Delay Columns:


PFAS Delay Column, 5 µm, 50 x 2,1 mm, colonna HPLC

PFAS delay column: 5 µm, 50 x 3.0 mm (cat.# 580959 [custom product])

Sample Prep:


Cartucce SPE Resprep S-DVB, 500mg/6ml, 30-pk.

Note that this CDC method is based on automated online SPE using C8 or C18 SPE materials. The product listed here is recommended for off-line SPE.

Vacuum Manifolds:


Vacuum Manifold Resprep QR-24
Vacuum Manifold Resprep QR-12

Vacuum Pumps:


Pompa da vuoto in PTFE Chemker 300, 20 l/min, CA 110 V, 60 Hz
Pompa da vuoto in PTFE Chemker 300, 18 l/min, CA 220 V, 50 Hz

Note that this CDC method is based on automated online SPE using C8 or C18 SPE materials. The products listed here are recommended for off-line SPE.

General Lab Supplies:


Vial a volume ridotto in polipropilene con filettatura, 9 mm, 700 µL, 12 x 32 mm, 100 pz.
Vial a volume ridotto in polipropilene con filettatura, 9 mm, 700 µL, 12 x 32 mm, 1000 pz.
Tappi in polietilene da 2,0 mL con top solido di 9 mm, filettatura, membrana spessore 10 mm, trasparente, 100 pz.
Tappi in polietilene da 2,0 mL con top solido di 9 mm, filettatura, membrana spessore 10 mm, trasparente, 1000 pz.
Provette da centrifuga vuote da 50 ml, Polipropilene con tappo, 50 provette, 50-conf.
Provette da centrifuga vuote da 50 ml, Polipropilene con tappo, 50 provette, 500-conf.
Siringa di Plastica Norm-Ject, Luer Lock Tip, graduazione fino a 12 mL, 10 mL, Conf. da 100
Siringa di Plastica Norm-Ject, Luer Lock Tip, graduazione fino a 6 mL, 5 mL, Conf. da 100

Discover More Solutions for PFAS Analysis!

Start Saving
Time Today

Develop PFAS Methods in Minutes with our Pro EZLC Chromatogram Modeler

Start Saving Time Today

Develop PFAS Methods in Minutes with our Pro EZLC Chromatogram Modeler

Tags

Author

  • Restek Corporation

    Restek is a leading provider of chromatography columns, accessories, and certified reference materials. Trust Restek for reliable, high-quality analytical solutions.

    View all posts
CFAR4291
Version: B