Articoli

Using Guard Columns and Retention Gaps in GC (Part 1)

22 Oct 2020

Guard columns and retention gaps are used widely in gas chromatography (GC). Many users have difficulty understanding the difference between these two products, even though there is a significant difference in application. Retention gaps mainly are used for focusing the sample components when introducing a large (liquid) sample directly onto the column. Guard columns are used to protect the analytical column from contamination. When using a retention gap system, the retention gap will also act as a guard column, but its primary function is to create a focusing effect.

Guard columns and retention gaps both must be coupled to the analytical column, and this connection introduces a potential point of risk. A new approach is to integrate the retention gap directly into the analytical column. By applying a “segment” coating technology the stationary phase can be deposited in a certain part of the column allowing a deactivated section at the beginning. Column coupling is not required and maintenance is greatly simplified. In Part 1 of this article we will explore retention gaps and build a foundation for a comparison to guard columns. In Part 2, we will review guard columns and discuss the new segment coating technology.

Use of retention gaps

In today’s laboratory, GC methods must be simple, fast, and low detection limits are required. Besides that, sufficient precision must also be obtained. It all starts by introducing the sample in the smallest possible injection band and making the band migrate through the capillary with minimal loss of the target components. With on-column injection a liquid sample is directly introduced into the capillary column as a liquid while the capillary column is kept at a temperature 10-15°C below the boiling point of the solvent. During this process the sample components are spread in an unreproducible way over the first 20-100cm of capillary while the solvent is evaporating. Parameters like injection speed, carrier gas flow, temperature of solvent and column, type of solvent, and pressure all will affect the injection band width. Additionally, when nonbonded stationary phases are used, the direct contact with liquids will result in a distortion of the stationary phase film and very short column lifetime. The majority of today’s stationary phases, like the Rtx and Rxi phases, are immobilized by cross- and surface bonding techniques.

Figure 1: Retention gaps are used to focus components in a tight band at the beginning of the analytical column.

decorativea) Sample introduction: liquid film of solvent and sample are deposited in the first length of capillary.
b) Oven temperature is increased (temp. program run): solvent and target compounds are vaporized and travel unretained through retention gap.
c) When target compounds come in contact with the stationary phase, they are refocused on the analytical column, resulting in a narrow initial band width.

For proper application of the on-column injection technique, the use of retention gaps is essential [1,2]. The retention gap consists of a 1-3m length of deactivated capillary that is positioned in front of the analytical column. All the processes described will still take place, but now the components are distributed over the retention gap. When the oven temperature is increased the sample components will start to move (there is very little retention…that’s why it’s called a retention “gap”). When reaching the analytical column, the components will focus in the stationary phase resulting in a narrowing of injection band width (Figure 1). As these retention gaps are mainly used for on-column injection, the inside diameter is usually 0.32mm up to 0.53mm since the needle of an on-column syringe must be able to enter the retention gap. For coupling the retention gaps to the analytical column, we need generally coupling devices that can deal with different diameter capillary tubing.

Retention gaps and splitless injection

While on-column injection minimizes discrimination and provides the best quantitative data, especially for thermolabile components, it can be challenging to perform. Many laboratories will choose a splitless method for ease of use. For splitless injection we generally do not require a retention gap. The sample is injected in a hot injection port, evaporated, and transported with a carrier gas flow of approximately 1mL/min into the capillary. The amount of solvent vapor that enters the column per unit time is much smaller than with on-column injection. Although with splitless injection the oven temperature is also 10-15°C below the boiling point of the solvent, there is little chance of the solvent condensing. The high concentration of solvent entering the capillary column will cause a strong focusing effect for the components, generating a narrow injection band. If, in splitless injection, a method is used where the initial (injection) oven temperature is much lower than the boiling point of the solvent, the risk of solvent condensation (forming a liquid plug) will increase. This can cause unwanted broadening of the injection band. Coupling a retention gap will also fix this problem.

Wettability of the retention gap

An important factor for good performance is the wettability of the retention gap surface. It is critical that the solvent spread evenly over the surface. This means that nonpolar solvents (hexane, methylene chloride, isooctane, benzene) require non/intermediate deactivated retention gaps and more polar solvents (methanol) will require polar deactivated retention gaps. If the polarity of the retention gap and solvent do not match, the solvent will form droplets inside the capillary. The carrier gas will “push” this droplet along the retention gap into the analytical column. The result is a broadened injection and possibly even peak splitting.

Retention gaps for large volume injection

Instead of injection of 1µl on a 1-2m retention gap, one can also inject much larger amounts on much longer retention gaps. Here we talk about large volume injection technique where retention gaps of 8-10m are used. Such retention gaps can be loaded with 100-200µl of sample. Injection must be slow to allow the solvent to evaporate while passing through the retention gap. With large volume injection, detection limits can be reduced a factor of 100. The technique requires some skill to optimize all the injection parameters. Additionally, the large volume retention gaps do pollute relatively quickly due to the large amounts of sample introduced.

Guard columns and retention gaps are useful tools to the practicing chemist and it is important to understand the difference between them. In Part 2 of this article we will review guard columns and discuss a new segment coating technology that allows retention gaps and guard columns to be built directly into the analytical column tubing. This new technology eliminates column coupling, substantially reducing analytical problems related to leaks and dead volume.

Read Part 2: Using Guard Columns and Retention Gaps in GC

References

  1. Grob, K., Journal of Chromatography 237:15 (1982).
  2. Hinshaw J., LC/GC Europe 17(9): 460–466 (2004).

Autori

  • Jaap de Zeeuw

    Jaap is a world-renowned chromatographer with over 40 years of experience, including 15 years with Restek and 27 years with Varian/Chrompack prior to joining Restek. Throughout his career, Jaap has focused the chromatographic challenges of industrial analysis. For his 1979 graduation from The Institute for Higher Education, where he specialized in chemistry, Jaap authored a paper titled "The Challenge of Coating Flexible Fused Silica Capillary Columns" and has since distinguished himself as an authority on the subject. Jaap has been directly involved with the creation of numerous chemically bonded columns, including the first bonded PEG column and the stabilized PLOT columns widely used in the petrochemical arena. He has also helped develop new techniques, such as fast GC-MS using vacuum GC technology (low-pressure GC or LPGC). Over his innovative career, he has filed several patents for his work. Based out of the Netherlands, Jaap is extensively published and regularly travels internationally to share his knowledge. After retiring from Restek, Jaap founded CreaVisions, where he works as a GC Consultant and teaches master classes on key GC topics as well as on creativity in science.

    Guarda tutti i post
GNAR3272