Articoli

Understanding Uncertainty on Your Restek Reference Standard Certificate of Analysis

20 Jan 2021

Restek’s Certificates of Analysis, or CofAs, offer comprehensive information to help you understand the composition, proper storage condi­tions, and method of shipment for your reference standard. Some customers also require the use of an “uncertainty” value for their analysis, which we provide on our CofAs.

In November 2022, we began rolling out exciting upgrades to our reference standard packaging, labeling, and certificates of analysis. Our manufacturing processes; QC methods; ISO accreditations; certified reference material (CRM) designations; ampuls; and the composition of the standards themselves were not changed in any way. However, as part of these upgrades, the appearance and information contained on our certificates of analysis has changed.

We formulate our reference standards for optimal compound stability, which can result in long shelf lives for many of our standards. As such, it’s likely you have reference standards that predate this upgrade and, therefore, have a previous version of our CofA. Additionally, some standards ordered from Restek today through 2023 may also have a previous version of our CofA due to the long shelf life of in-stock products.

This article is separated into two sections, with each section specific to either previous versions or upgraded versions of our certificates of analysis:

  • If your standard was manufactured on or before 21 November 2022, it uses the previous version. Click here to jump to the lower section of this article that focuses on these specific certificates.
  • If your standard was manufactured after 21 November 2022, it uses the upgraded version. You may continue reading below.

Understanding Uncertainty on Your Restek CofA

For standards manufactured after 21 November 2022.

In reference standards, “uncertainty” is a quantitative measurement that determines the quality of a standard. It’s used to provide a margin of error that could potentially apply to the product. It’s important to note, however, that unless your analysis results require the presentation of uncertainty, it may not be necessary for you to use the uncertainty value provided on your CofA.

Expanded Uncertainty

On your CofA, there is a section within the table labelled “Certified Values,” and within that is “Expanded Uncertainty” (Figure 1). These values are labeled “certified” because the standard has been manufactured with its uncertainties calculated and presented in adherence to our ISO accreditations. If your analysis requires the presentation of uncertainty, the “Expanded Uncertainty” is the value you would use.

Figure 1: An excerpt from the certificate of analysis of our Cannabinoids Acids 7 multicomponent mix (cat.# 34144). In this example, the “Expanded Uncertainty” value of the first eluting compound, CBDVA, is +/- 25.2543 μg/mL.

figure article GNAR3428A 01a

The expanded uncertainty value on your CofA is calculated using four uncertainty components: gravimetric, homogeneity, storage, and shipping.

Gravimetric assesses all uncertainty from the physical making of the standard. It includes balance uncertainty, volumetric and pipette uncertainty, and purity uncertainty.

Homogeneity assesses if all ampuls within the batch are the same. For example, if we place 1 mL of solution into 50 ampuls, would each ampul have the same concentration of every compound?

Storage Stability assesses if the standard concentration is the same from the first day it was made up to the expiration date on the label.

Shipping Stability assesses the conditions the standard experienced during shipping.

Figure 2:

figure article GNAR3428A 02

Figure 2 illustrates how the four components are used to determine the expanded uncertainty value. Simplified, the calculation can be thought of this way:

Expanded uncertainty = gravimetric uncertainty + homogeneity uncertainty + storage uncertainty + shipping uncertainty

The actual calculation we use for the expanded uncertainty value displayed on our CofAs is made using this formula:

equation article GNAR3428A 01

Where “k” is a coverage factor of 2, which gives a level of confidence of approximately 95%.

Note: While we display this formula on our CofAs, you do not need to apply or use this formula itself. It is only displayed to show how the expanded uncertainty value was calculated.

Storage and Shipping Conditions

Your CofA also includes information about your standard’s storage and shipping conditions.

Storage

The storage field refers to the temperature that the product should be stored at once you receive it to guarantee its uncertainty, concentration, and expiration. This does not mean the ampul must remain at or below this temperature during shipment from Restek to you.

Shipping

The shipping field refers to the packaging conditions that the product can be shipped in and still maintain stability. Most—approximately 98%—of our reference standards will remain stable when shipped under “Ambient” conditions. The remaining 2% of our reference standards require “On ice” packaging.

If your shipment was in transit for more than seven days and you have reason to believe it was exposed to extreme temperature conditions, contact Restek Technical Service at www.restek.com/contact-us for recommendations.

We hope the information contained in this article has helped you to better understand uncertainty, storage, and shipping conditions for Restek’s reference standards. If you have any questions, we recommend reviewing the Reference Standard FAQ (GNOT3397-UNV) or contacting Restek’s Technical Service at www.restek.com/contact-us

For standards manufactured before 21 November 2022:

Understanding Your Restek Reference Standard Certificate of Analysis: Uncertainty, Storage, and Shipping Conditions

Restek’s Certificates of Analysis, or CofAs, offer comprehensive information to help you understand the composition, proper storage conditions, and method of shipment for your reference standard. Some customers also require “uncertainty” values for their analyses, so we include “certified values” that include “expanded uncertainty.” In this article, we’ll explain these terms, discuss when you might need to apply uncertainty values, and clarify why we provide them. We’ll also detail the differences between storage conditions and shipping conditions along with what else you need to keep in mind to ensure the long-term integrity of your standard.

Uncertainty

In reference standards, “uncertainty” is a quantitative measurement that determines the quality of a standard. It’s used to provide a margin of error that could potentially apply to the product.

It’s important to note, however, that unless your analysis results require the presentation of uncertainties, it may not be necessary for you to use the uncertainty values provided on your CofA.

Certified Values

On your certificate of analysis, there is a “Certified Values” table. These values are labeled “certified” because the standard has been manufactured with its uncertainties calculated and presented in adherence to our ISO accreditations. Certified values include the standard’s gravimetric concentration as well as its expanded uncertainty (Figure 3).

Figure 3: An excerpt from the certificate of analysis of our LC Multiresidue Pesticide Standard #5 (cat.# 31976). In this example, we will focus on the certified values of its first eluting compound, thiabendazole.

figure article GNAR3428A 03

Gravimetric Concentration

This value represents the actual concentration of the compound in the ampul.

For our above compound, the gravimetric concentration is 100.0 µg/mL.

Most of our customers use only gravimetric concentration for their analyses. If the results of your analysis do not require the use or presentation of uncertainties, you can skip the Expanded Uncertainty sections below and continue on to the Homogeneity, Storage, and Shipping sections.

Expanded Uncertainty

Expanded Uncertainty takes a closer look at the gravimetric concentration and provides a detailed breakdown of several uncertainties that your ampul may be subject to. These uncertainty values will be needed if the results of your analysis require their presentation.

There are three expanded uncertainty components that we provide on our CofAs:

  • Gravimetric
  • Unstressed
  • Stressed

Gravimetric Expanded Uncertainty

This value contains all uncertainty from the physical making of the standard. It includes balance uncertainty, volumetric and pipette uncertainty, and purity uncertainty.

For our above compound, the gravimetric uncertainty is +/- 0.7091 µg/mL.

Unstressed Expanded Uncertainty

This value represents combined uncertainties present within your reference standard, including transit from Restek to your lab under standard shipping conditions (explained below).

unstressed expanded uncertainty = gravimetric uncertainty + homogeneity uncertainty + storage uncertainty

For our above compound, the unstressed expanded uncertainty is +/- 4.6114 µg/mL.

Stressed Expanded Uncertainty

This value (also referred to as “combined stressed uncertainty”) represents the uncertainties present within your reference standard, including transit from Restek to your lab under nonstandard shipping conditions (explained below).

stressed expanded uncertainty = gravimetric uncertainty + homogeneity uncertainty + storage uncertainty + shipping uncertainty

For our above compound, the stressed uncertainty is +/- 4.7285 µg/mL.

To further explain, we also display a formula on the CofA that shows the calculation we used to determine the stressed expanded uncertainty:

equation article GNAR3428A 01

Where “k” is a coverage factor of 2, which gives a level of confidence of approximately 95%.

Note: You do not need to apply or use this formula itself—we already provide these uncertainties values on our CofAs. We only display the formula to show how the combined stressed value was calculated.

Homogeneity, Storage, and Shipping

Beyond the factors contributing to gravimetric expanded uncertainty, unstressed and stressed expanded uncertainties also include these additional factors:

Homogeneity – are all ampuls from the batch the same? For example, if we place 1 mL of solution into 50 ampuls, would each ampul have the same concentration of every compound?

Storage – is your standard concentration the same from the first day it was made up to the expiration date on the label?

Shipping (applies only to stressed uncertainty) – this is an assessment of the worst-case scenario the standard experienced during shipping. Essentially, we’re asking the question: what is the hottest temperature the ampul experienced and how long did this last? For this, we are evaluating whether the integrity of the standard is the same before and after shipment.

Figure 4 below shows the relationships between gravimetric concentration as well as the expanded uncertainty values for gravimetric uncertainty, unstressed uncertainty, and stressed uncertainty.

Figure 4:

figure article GNAR3428A 04

Storage Conditions vs. Shipping Conditions

To determine the uncertainty of your standard, it’s helpful to understand the difference between the storage conditions and shipping conditions listed on your CofA (Figure 5).

Figure 5: The CofA for our LC Multiresidue Pesticide Standard #5 (cat.# 31976) has fields for both the storage and shipping conditions.

figure article GNAR3428A 05

Storage

The storage field refers to the temperature that the product should be stored at once you receive it to guarantee its uncertainty, concentration, and expiration. This does not mean the ampul must remain at or below this temperature during shipment from Restek to you.

Shipping

The shipping field refers to the packaging conditions that the product can be shipped in and still maintain stability. Most—approximately 98%—of all our reference standards will remain stable when shipped under “ambient” conditions. The remaining 2% of our reference standards require “on ice” packaging. These are the only two options for this field.

Our reference standards have been tested under sustained temperature extremes for specific time intervals—extremes that most ampuls will never experience during transit (Table I).

Table I: Included on every CofA, this table shows the label conditions, standard shipping conditions, and nonstandard shipping conditions of our standards.

Label ConditionsStandard ConditionsNonstandard Conditions
25 °C Nominal (Room Temperature)< 60 °C≥ 60 °C up to 7 days
10 °C or colder (Refrigerate)< 40 °C≥ 40 °C up to 7 days
0 °C or colder (Freezer) -20 °C or colder (Deep Freezer)< 25 °C≥ 25 °C up to 7 days

Standard Shipping Conditions

If you receive your standard under standard shipping conditions, you should apply the Unstressed Uncertainty Value to your standard.

Our example reference standard has a label condition of 10 °C or colder. Under standard temperature conditions, the standard may experience up to 40 °C temperatures with no adverse effects expected.

Nonstandard Shipping Conditions

If you receive your standard under nonstandard shipping conditions, you should apply the Stressed Uncertainty Value to your standard.

Our example reference standard has a label condition of 10 °C or colder. Under nonstandard temperature conditions, the standard may experience temperatures equal to or greater than 40 °C temperatures for up to 7 days with no adverse effects expected.

Standard vs. Nonstandard Shipping Conditions

In our experience, it’s highly unlikely for a standard to experience extreme temperatures that are sustained throughout transit. Therefore, it’s highly unlikely that you will need to apply the Nonstandard Shipping Conditions when calculating the uncertainty of your standard.

If your shipment was in transit for more than seven days and you have reason to believe it exceeded the nonstandard temperature conditions, contact Restek Technical Service at www.restek.com/contact-us for recommendations.

We hope the information contained in this article has helped you to better understand uncertainty, storage, and shipping conditions for your Restek certified reference standards. If you have any questions, we recommend reviewing the Reference Standard FAQ or contacting Restek’s Technical Service at www.restek.com/contact-us

Find our full selection of reference standards here.

Need Help with Reference Standards?

From method selection to product recommendations, our reference standards experts are here to help you navigate complex workflows and achieve accurate, reliable results.

Prodotti Menzionati


Standard di pesticidi per analisi multiresiduale LC n 5, composti organici azotati, 100 g/mL, acetonitrile, 1 mL/ampolla
Standard 7 Cannabinoidi Acidi, 1000 μg/mL, Acetonitrile con 1% DIPEA e 0.05% Acido Ascorbico, 1 mL/ampollla

Autori

  • Restek Corporation

    Restek is a leading provider of chromatography columns, accessories, and certified reference materials. Trust Restek for reliable, high-quality analytical solutions.

    View Articles
GNAR3428
Versione: A