Articoli

PFAS Analysis by ISO 23702-1:2023 for Leather

03 Oct 2025

feature GNAR4301
  • Sample Matrix: leather
  • Sample Preparation: methanol extraction
  • Detection Technique: LC-MS/MS

Determination of the nonvolatile compound content by extraction method using liquid chromatography/ tandem mass spectrometry detector (LC-MS/MS)

ISO 23702-1:2023 is used for the analysis of 42 PFAS in leather. It is not available from Restek but may be purchased from other sources. The full analyte list and recommended products for ISO 23702-1:2023 are provided below. In addition, you can register and directly compare analyte lists and matrices across PFAS analysis methods here or visit www.restek.com/PFAS to explore our technical resources for PFAS analysis.

Analyte List for ISO 23702-1:2023 for PFAS Analysis

PFAS acronyms, compound names, and CAS numbers vary across methods and guidelines, and they also depend on compound form (e.g., neutral vs. salt). The following list uses standardized terminology to allow easy comparison across the methods and guidelines in our comprehensive guide. Please refer to ISO 23702-1:2023 for method-specific nomenclature.

AcronymPFAS NameCAS
PFBSPerfluoro-1-butanesulfonic acid375-73-5
PFDAPerfluorodecanoic acid335-76-2
PFDoAPerfluorododecanoic acid307-55-1
PFHpA*Perfluoroheptanoic acid375-85-9
PFHxSPerfluoro-1-hexanesulfonic acid355-46-4
PFHxAPerfluorohexanoic acid307-24-4
PFNAPerfluorononanoic acid375-95-1
PFOSPerfluorooctanesulfonic acid1763-23-1
PFOSFHeptadecafluorooctanesulfonyl fluoride307-35-7
PFOAPerfluorooctanoic acid335-67-1
Me-PFOAMethyl perfluorooctanoate376-27-2
Et-PFOAEthyl perfluorooctanoate3108-24-5
PF-3,7-DMOAPerfluoro(3,7-dimethyloctanoic acid)172155-07-6
PFTeDAPerfluorotetradecanoic acid376-06-7
PFTrDAPerfluorotridecanoic acid72629-94-8
PFUnAPerfluoroundecanoic acid2058-94-8
HFPO-DA, GenXHexafluoropropylene oxide dimer acid13252-13-6
PFBA*Perfluorobutanoic acid375-22-4
PFPeA*Perfluoropentanoic acid2706-90-3
PFDS*Perfluoro-1-decanesulfonic acid335-77-3
PFHpS*Perfluoro-1-heptanesulfonic acid375-92-8
FOSAPerfluoro-1-octanesulfonamide754-91-6
4:2 FTS*1H,1H,2H,2H-Perfluorohexane sulfonic acid757124-72-4
6:2 FTS*1H,1H,2H,2H-Perfluorooctane sulfonic acid27619-97-2
8:2 FTS1H,1H,2H,2H-Perfluorodecane sulfonic acid39108-34-4
10:2 FTS*1H,1H,2H,2H-perfluorododecane sulfonate (10:2)120226-60-0
NMeFOSAN-Methylperfluorooctanesulfonamide31506-32-8
NEtFOSAN-Ethylperfluorooctanesulfonamide4151-50-2
NMeFOSE2-(N-methylperfluoro-1-octanesulfonamido)-ethanol24448-09-7
NEtFOSE2-(N-ethylperfluoro-1-octanesulfonamido)-ethanol1691-99-2
8:3 FTCA2H,2H,3H,3H-Heptadecafluoroundecanoic acid34598-33-9
10:2 FTOH1H,1H,2H,2H-Perfluorododecan-1-ol or 2-(Perfluorodecyl)ethanol865-86-1
8:2 FTOH1H,1H,2H,2H-Perfluoro-1-decanol678-39-7
6:2 FTOH*3,3,4,4,5,5,6,6,7,7,8,8,8-Tridecafluorooctan-1-ol or 3,3,4,4,5,5,6,6,7,7,8,8,8-Tridecafluoro-1-octanol647-42-7
4:2 FTOH*1H,1H,2H,2H-Perfluorohexan-1-ol or 1-Hexanol2043-47-2
6:2 FTA*1H,1H,2H,2H-Perfluorooctyl acrylate17527-29-6
8:2 FTA1H,1H,2H,2H-Perfluorodecyl acrylate27905-45-9
10:2 FTA1H,1H,2H,2H-Perfluorododecyl acrylate17741-60-5
6:2 FTMA*1H,1H,2H,2H-heptadecafluorooctylmethacrylate2144-53-8
8:2 FTMA1H,1H,2H,2H-heptadecafluorodecylmethacrylate1996-88-9
7HPFHpA*7H-Dodecafluoroheptanoic acid1546-95-8
1-Decanaminium, N-decyl-N, N dimethyl-0,1,1,2,2,3,3,4,4,5,5,6,6,7,7,8,8,8-heptadecafluoro-1-octanesulfonate251099-16-8

*Optional compound

Recommended Products for ISO 23702-1:2023 for PFAS Analysis

Analytical LC Columns:


Colonna HPLC Raptor C18, 2,7 µm, 50 x 2,1 mm
Colonna LC Force C18, 1,8 µm, 50 x 2,1 mm

Delay Columns:


PFAS Delay Column, 5 µm, 50 x 2,1 mm, colonna HPLC

PFAS delay column: 5 µm, 50 x 3.0 mm (cat.# 580959 [custom product])

Guard Columns:


Filtro precolonna UHPLC UltraShield, frit 0,2 µm, cad.
Holder EXP a connessione diretta per cartucce per precolonne EXP, comprende raccordo e ferrule
Cartuccia per precolonna Raptor C18 EXP, 2,7 µm, 5 x 2,1 mm, 3 pz.

Sample Prep:


Filtro siringa da 25 mm, 0,22 µm, polipropilene, blu, Luer-Lock, 100 pz.

Vacuum Manifolds:


Vacuum Manifold Resprep QR-24
Vacuum Manifold Resprep QR-12

Vacuum Pumps:


Pompa da vuoto in PTFE Chemker 300, 20 l/min, CA 110 V, 60 Hz
Pompa da vuoto in PTFE Chemker 300, 18 l/min, CA 220 V, 50 Hz

General Lab Supplies:


Vial a volume ridotto in polipropilene con filettatura, 9 mm, 700 µL, 12 x 32 mm, 100 pz.
Vial a volume ridotto in polipropilene con filettatura, 9 mm, 700 µL, 12 x 32 mm, 1000 pz.
Tappi in polietilene da 2,0 mL con top solido di 9 mm, filettatura, membrana spessore 10 mm, trasparente, 100 pz.
Tappi in polietilene da 2,0 mL con top solido di 9 mm, filettatura, membrana spessore 10 mm, trasparente, 1000 pz.
Provette da centrifuga vuote da 50 ml, Polipropilene con tappo, 50 provette, 50-conf.
Provette da centrifuga vuote da 50 ml, Polipropilene con tappo, 50 provette, 500-conf.
Siringa di Plastica Norm-Ject, Luer Lock Tip, graduazione fino a 12 mL, 10 mL, Conf. da 100
Siringa di Plastica Norm-Ject, Luer Lock Tip, graduazione fino a 6 mL, 5 mL, Conf. da 100

Discover More Solutions for PFAS Analysis!

Start Saving
Time Today

Develop PFAS Methods in Minutes with our Pro EZLC Chromatogram Modeler

Start Saving Time Today

Develop PFAS Methods in Minutes with our Pro EZLC Chromatogram Modeler

Tag

Autori

  • Restek Corporation

    Restek is a leading provider of chromatography columns, accessories, and certified reference materials. Trust Restek for reliable, high-quality analytical solutions.

    View Articles
GNAR4301
Versione: B