Blog

Adding THC-O Acetate to your Potency Method

20 Jan 2022

The cannabis industry is always innovating with a constant revolving door of hot new products. In the past we’ve seen buzz around products containing delta-8-THC, delta-10-THC, and THCP. 

Recently there has been a lot of interest in tetrahydrocannabinol acetate, or THC-O acetate. THC-O acetate is a psychoactive synthetic compound produced by acetylating THC. As far as legality, there is arguably some grey area and varies by state. There are claims of THC-O acetate being stronger than delta-9-THC and many products containing THC-O acetate are becoming available on the market. As such, labs are interested in including this compound into their cannabinoid potency methods.

Herein, three methods have been developed to monitor 12 cannabinoids, including delta-10-THC and THC-O acetate by LC-UV. Since labs have different requirements, different methods were developed so the user has the flexibility to use an unbuffered or a buffered mobile phase system as well as methanol or acetonitrile in order to best fit their needs. For this analysis, a Raptor ARC-18 150 x 3 mm, 2.7 µm column was chosen to achieve desired resolution while keeping the analysis time under 12 minutes.

Figure 1. Method conditions and extracted chromatogram for the analysis of 13 cannabinoids by LC-UV using acetonitrile as an organic modifier and buffered aqueous mobile phase.

blog adding THC O acetate to your potency method 01

Figure 2. Method conditions and extracted chromatogram for the analysis of 13 cannabinoids by LC-UV using acetonitrile as an organic modifier without the use of buffer.

blog adding THC O acetate to your potency method 02

Figure 3. Method conditions and extracted chromatogram for the analysis of 13 cannabinoids by LC-UV using methanol as an organic modifier and buffered aqueous mobile phase.

blog adding THC O acetate to your potency method 01

By using one of these methods, resolution can be achieved for all compounds analyzed within a 12 minute total analysis time. 

Author

  • Jamie York, PhD

    Jamie York is a scientist in the Applications Lab at Restek Corporation in the LC Solutions department, where she works on the development of novel applications for the food, clinical, and cannabis markets. She earned her PhD in chemistry from The University of Texas at Arlington in 2019. There, she mastered many analytical techniques including gas chromatography–vacuum ultraviolet; gas chromatography–mass spectrometry; matrix-assisted laser desorption/ionization; and liquid chromatography–mass spectrometry with a focus on food and environmental research. Jamie continued her post-doctoral work at The University of Texas at Arlington with a focus on the analysis of mammalian cell culture media by LC-MS/MS.

    View all posts
GNBL5462