Blog

BAC Analysis Using Hydrogen Carrier Gas: Get the Same Results at a Lower Cost!

07 Oct 2012

blog bac analysis using hydrogen carrier gas get the same results at a lower cost 01

Who doesn’t want the same thing they’re getting now at a lower cost, especially if historically that cost has only been rising? That question got me thinking about the 2012 SOFT meeting where I presented a talk about Restek’s new BAC Plus columns. A large portion of the talk focused on improving headspace chromatography of blood alcohol compounds by using a fast linear velocity and split injection technique. As you can see in Figure 1, going from a 5:1 split injection to a 20:1 split drastically improves peak shape without sacrificing much in the way of sensitivity. Note that these chromatograms were generated on the BAC Plus 2 column only – the same thing happens on the BAC Plus 1, I promise.

However, a 20:1 split injection on a 0.32mm ID column operating at 80cm/sec uses a 150mL/min of helium. A 0.53mm column operated in this manner uses a whopping 275mL/min helium. Wow! That’s a lot of helium! Unfortunately, since helium is becoming increasingly scarce, implementing this method change in your lab may be prohibitively costly or just plain impossible, depending on your helium supply. Fortunately, there’s a really easy way to get around this problem. There have been several other blog posts from my colleagues discussing the use of alternative carrier gases over the past few months, so that got me thinking: why not use a hydrogen carrier for blood alcohol analyses?

Usually, changing carrier gases for FID analyses doesn’t pose much of a problem – especially for isothermal FID analyses. In fact, the QC tests for the BAC Plus columns are run using hydrogen as a carrier gas. So I gave it a try and ended up with beautiful chromatography using exactly the same conditions that I was using for helium. Figure 2 shows my BAC run with hydrogen.

diagram

I used nitrogen as the pressurization gas for my headspace instrument. For obvious safety reasons, hydrogen shouldn’t be used as the vial pressurization gas. By performing a split injection, I’ve avoided any potential problems arising from injecting a large slug of nitrogen into the carrier gas stream.

It’s every chromatographer’s dream to have their analysis work on the first try, and it appears that switching to hydrogen worked that way. To date, I haven’t done any quantitative work with it, but it definitely seems like a viable solution. I would love to hear comments from labs who have tried this approach.

Prodotti Menzionati


Colonna capillare GC Rtx- BAC PLUS 1, 30 m, ID 0,32 mm, 1,8 µm
Colonna capillare GC Rtx- BAC PLUS 2, 30 m, ID 0,32 mm, 0,6 µm
Topaz Inlet Liner Dritto, 2,0 mm x 6,5 x 78,5, per GC Agilent, Deattivazione Premium, 5 pz.

Autori

  • Amanda Rigdon is an analytical chemist with more than 21 years of experience in chromatography and mass spectrometry. Throughout her career, she has worked in a variety of analytical fields, including clinical forensics, food safety, environmental analysis, and cannabis testing. Her experience spans method development, validation, troubleshooting, and laboratory operations using both GC and LC methodologies. Amanda also served as Chief Scientific Officer of an ISO/IEC 17025-accredited cannabis testing laboratory, where she gained extensive experience in quality systems, regulatory compliance, and laboratory management. She is currently a Principal Scientist in the GC Applications Group at Restek, where she enjoys helping customers solve analytical challenges and sharing practical chromatography knowledge through technical support, presentations, and educational content.

    Guarda tutti i post
GNBL4588