Stereochemistry refers to the three-dimensional arrangement of atoms within a molecule. The stereochemistry of a molecule can have significant implications on both its chemical and physical properties. It can affect the ability of the molecule to undergo a chemical reaction, impact its solubility, or even influence the biological activity of the compound.
There are two different types of stereoisomers—enantiomers and diastereomers. Molecules that are enantiomers will be nonsuperimposable mirror images of each other, meaning all chiral centers of the molecules will be inverted, but the structures will otherwise be identical. Because of this, a pair of enantiomers will have the same physical properties (melting points, boiling points, solubility, etc.). In a chiral environment, however, enantiomers will behave uniquely and can have extremely different effects in the human body. Diastereomers are nonsuperimposable, non-mirror images of each other. In diastereomers, the configuration of at least one chiral center within the structures will differ. Unlike enantiomers, diastereomers will have different physical properties.
Because enantiomers share the same physical properties, differentiating between them using HPLC can be challenging. Separating a pair of enantiomers by HPLC typically requires a chiral stationary phase, which are notoriously expensive, fragile, and lack broad utility. In some cases, enantiomers may be differentiated with reversed-phase HPLC columns if a derivatization step is performed prior to analysis. Diastereomers, on the other hand, can be more easily separated using traditional HPLC stationary phases without any additional steps, such as derivatization.
9(S)-hexahydrocannabinol (9(S)-HHC) and 9(R)-hexahydrocannabinol (9(R)-HHC) are both cannabinoids that occur naturally in trace amounts in the cannabis plant. 9(S)-HHC and 9(R)-HHC are epimers, which are a specific kind of diastereomer. Epimers will differ in their stereochemical configuration at only one specific chiral center on the molecule. While their chemical formulas will be identical, this difference in configuration at one chiral center will lead to the epimers having different physiochemical properties. These HHC compounds have recently piqued the interest of toxicology testing laboratories as they have started to appear in various products as a “legal high” alternative to THC. HHC is found at much higher concentrations in these products than would occur naturally in the cannabis plant, indicating they are being synthetically manufactured.
In the human body, the HHC compounds will follow a similar metabolic pathway to THC, forming 9(S)-HHC-COOH and 9(R)-HHC-COOH as metabolites. Recently, I was working on a method that incorporated 9(S)-HHC, 9(R)-HHC, and their metabolites into a comprehensive LC-MS/MS method for the analysis of cannabinoids in whole blood specimens. The instrument conditions, analyte list, and retention times are shown below in tables I and II and figure 1.
Table I: Instrument Conditions for Panel of Cannabinoids in Whole Blood by LC-MS/MS
| Analytical Column | Raptor FluoroPhenyl, 100 x 3 mm, 2.7 µm | ||
| Mobile Phase A | Water, 0.1% formic acid | ||
| Mobile Phase B | Methanol, 0.1% formic acid | ||
| Column Temperature | 30 °C | ||
| Flow Rate | 0.8 mL/min | ||
| Gradient | Time (min) | %A | %B |
| 0.00 | 34 | 66 | |
| 5.50 | 34 | 66 | |
| 5.60 | 29 | 71 | |
| 10.50 | 29 | 71 | |
| 11.00 | 15 | 85 | |
| 13.00 | 0 | 100 | |
| 14.00 | 0 | 100 | |
| 14.10 | 34 | 66 | |
| 16.00 | 34 | 66 | |
Figure 1. Example Chromatogram (100 ng/mL for all analytes)

Table II: Peak ID and Retention Times
| Peak # | Analyte | RT (min) | Peak # | Analyte | RT (min) |
| 1 | 11-OH-Δ8-THC | 4.81 | 8 | Δ8-THC | 10.02 |
| 2 | 11-OH-Δ9-THC | 5.24 | 9 | 9(S)-HHC | 10.14 |
| 3 | Δ8-THC-COOH | 5.27 | 10 | Δ9-THC | 10.45 |
| 4 | 9(R)-HHC-COOH | 5.44 | 11 | 9(R)-HHC | 10.84 |
| 5 | Δ9-THC-COOH | 6.15 | 12 | (6aR, 9R)-Δ10-THC | 11.63 |
| 6 | 9(S)-HHC-COOH | 6.36 | 13 | Δ9-THCP | 12.11 |
| 7 | CBDP | 9.15 | 14 | Δ9-THC-O-Acetate | 12.40 |
When separating isomers, metabolites with the same configuration will often follow the same elution pattern as the parent compounds. This is observed with the carboxy and hydroxy metabolites of Δ8-THC and Δ9-THC. Δ8- and Δ9-THC are positional isomers, differing in the position of a single double bond in their structures. I expected the HHC-COOH compounds to follow the same pattern as the parent molecules, with 9(S)-HHC-COOH eluting before 9(R)-HHC-COOH. I was surprised to find that the HHC-COOH metabolites eluted in the opposite order of the parent HHC compounds. After verifying the retention times of all four compounds were correct, I took a look at the analyte structures to see if this difference could be explained (Figures 2 and 3).
Figure 2*: 9(R)-HHC [left] and 9(S)-HHC [right]


Figure 3*: 9(R)-HHC-COOH [left] and 9(S)-HHC-COOH [right]


Looking at the stereochemistry of these compounds, 9(R)-HHC has an equatorial methyl group while 9(S)-HHC has an axial methyl group, making 9(R)-HHC more nonpolar and causing it to elute after 9(S)-HHC. Based on this logic, we might also expect 9(R)-HHC-COOH to be more non-polar than 9(S)-HHC-COOH. However, the stereocenter on the HHC-COOH compounds is a part of a carboxylic acid group, which has an impact on how these compounds interact with the stationary phase. Groups with axial orientations, like these 9(S) compounds, will have significantly more steric hinderance than equatorial groups. While the axial group on 9(S)-HHC is a small methyl group that will not contribute much to steric hindrance, the axial group on 9(S)-HHC-COOH is a much larger carboxylic acid group that will exhibit much greater steric hindrance. This increased steric hindrance causes 9(S)-HHC-COOH to travel slower through the stationary phase, ultimately eluting later than 9(R)-HHC-COOH. This difference in stereochemistry explains why the 9(S) and 9(R) carboxy metabolites do not follow the same elution pattern as the parent compounds.
The next time you find yourself questioning the elution order of a set of isomers, don’t forget to consider stereochemistry!
*Figure 2 and Figure 3 images can be found on the Cayman Chemicals website at https://www.caymanchem.com/



