Articoli

PFAS Analysis by U.S. EPA Method 537.1 for Drinking Water

03 Oct 2025

feature EVAR4284
  • Sample Matrix: drinking water
  • Sample Preparation: S-DVB solid-phase extraction
  • Detection Technique: LC-MS/MS

Determination of Selected Per- and Polyfluorinated Alkyl Substances in Drinking Water by Solid Phase Extraction and Liquid Chromatography/Tandem Mass Spectrometry (LC-MS/MS)

U.S. EPA Method 537.1 is used for the analysis of 18 PFAS in drinking water. It combines solid-phase extraction sample preparation using styrene-divinylbenzene cartridges with LC-MS/MS analysis on a C18 column. The method allows flexibility in some aspects, such as evaporation technique, separation technique, mobile phase composition, LC column and conditions, and detector conditions, but changes to sample collection, preservation, and extraction and quality control requirements are prohibited. The full analyte list and recommended products for PFAS analysis in drinking water by Method 537.1 are provided below. In addition, you can register and directly compare analyte lists and matrices across PFAS analysis methods here or visit www.restek.com/PFAS to explore our technical resources for PFAS analysis.

Analyte List for U.S. EPA Method 537.1 for PFAS Analysis

PFAS acronyms, compound names, and CAS numbers vary across methods and guidelines, and they also depend on compound form (e.g., neutral vs. salt). The following list uses standardized terminology to allow easy comparison across the methods and guidelines in our comprehensive guide. Please refer to U.S. EPA Method 537.1 for method-specific nomenclature.

AcronymCompound NameCAS#
NEtFOSAAN-ethyl perfluorooctanesulfonamidoacetic acid2991-50-6
NMeFOSAAN-methyl perfluorooctanesulfonamidoacetic acid2355-31-9
PFBSPerfluoro-1-butanesulfonic acid375-73-5
PFDAPerfluorodecanoic acid335-76-2
PFDoAPerfluorododecanoic acid307-55-1
PFHpAPerfluoroheptanoic acid375-85-9
PFHxSPerfluoro-1-hexanesulfonic acid355-46-4
PFHxAPerfluorohexanoic acid307-24-4
PFNAPerfluorononanoic acid375-95-1
PFOSPerfluorooctanesulfonic acid1763-23-1
PFOAPerfluorooctanoic acid335-67-1
PFTeDAPerfluorotetradecanoic acid376-06-7
PFTrDAPerfluorotridecanoic acid72629-94-8
PFUnAPerfluoroundecanoic acid2058-94-8
HFPO-DA, GenXHexafluoropropylene oxide dimer acid13252-13-6
9Cl-PF3ONS9-Chlorohexadecafluoro-3-oxanonane-1-sulfonic acid (F-53B Major)756426-58-1
11Cl-PF3OUdS11-Chloroeicosafluoro-3-oxaundecane-1-sulfonic acid (F-53B Minor)763051-92-9
ADONA4,8-Dioxa-3H-perfluorononanoic acid919005-14-4

Recommended Products for U.S. EPA Method 537.1 for PFAS Analysis

Analytical LC Columns:


Colonna HPLC Raptor C18, 2,7 µm, 50 x 2,1 mm
Colonna LC Force C18, 1,8 µm, 50 x 2,1 mm

Delay Columns:


PFAS Delay Column, 5 µm, 50 x 2,1 mm, colonna HPLC

PFAS delay column: 5 µm, 50 x 3.0 mm (cat.# 580959 [custom product])

Guard Columns:


Filtro precolonna UHPLC UltraShield, frit 0,2 µm, cad.
Holder EXP a connessione diretta per cartucce per precolonne EXP, comprende raccordo e ferrule
Cartuccia per precolonna Raptor C18 EXP, 2,7 µm, 5 x 2,1 mm, 3 pz.

Standards:


Standard di calibrazione PFAS EPA 537.1, 2 μg/mL, metanolo (1 mM KOH), 1 mL/ampolla

Sample Prep:


Cartucce SPE Resprep S-DVB, 500mg/6ml, 30-pk.

Sample Reservoirs:


Provette vuote SPE, serbatoio per campioni, 75 mL, polipropilene, 12 pz.
Connettore SPE, 1, 3, 6, 10 o 15 mL, polipropilene, 15 pz.

Vacuum Manifolds:


Vacuum Manifold Resprep QR-24
Vacuum Manifold Resprep QR-12

Vacuum Pumps:


Pompa da vuoto in PTFE Chemker 300, 20 l/min, CA 110 V, 60 Hz
Pompa da vuoto in PTFE Chemker 300, 18 l/min, CA 220 V, 50 Hz

General Lab Supplies:


Vial a volume ridotto in polipropilene con filettatura, 9 mm, 700 µL, 12 x 32 mm, 100 pz.
Vial a volume ridotto in polipropilene con filettatura, 9 mm, 700 µL, 12 x 32 mm, 1000 pz.
Tappi in polietilene da 2,0 mL con top solido di 9 mm, filettatura, membrana spessore 10 mm, trasparente, 100 pz.
Tappi in polietilene da 2,0 mL con top solido di 9 mm, filettatura, membrana spessore 10 mm, trasparente, 1000 pz.
Provette da centrifuga vuote da 50 ml, Polipropilene con tappo, 50 provette, 50-conf.
Provette da centrifuga vuote da 50 ml, Polipropilene con tappo, 50 provette, 500-conf.
Siringa di Plastica Norm-Ject, Luer Lock Tip, graduazione fino a 12 mL, 10 mL, Conf. da 100
Siringa di Plastica Norm-Ject, Luer Lock Tip, graduazione fino a 6 mL, 5 mL, Conf. da 100

Discover More Solutions for PFAS Analysis!

EZLC modeler PFAS Placeholder Image

Author

  • Restek Corporation

    Restek is a leading provider of chromatography columns, accessories, and certified reference materials. Trust Restek for reliable, high-quality analytical solutions.

    View all posts
EVAR4284
Version: A