The following table lists the recommended orifice diameters, associated flow rate ranges, and canister volumes and sampling times:

Canister Volume/Sampling Time OrificeSiltek-TreatedStainless Steel
400 cc1 Liter3 Liter6 Liter15 LiterFlowSize cat.#cat.#
8 hour24 hour48 hour125 hour—0.5–2 mL/min0.0008″2421724216
2 hour4 hour12 hour24 hour60 hour2–4 mL/min0.0012″2416024165
1 hour2 hour6 hour12 hour30 hour4–8 mL/min0.0016″2416124166
—1 hour4 hour8 hour20 hour8–15 mL/min0.0020″2416224167
——2 hour3 hour8 hour15–30 mL/min0.0030″24163—
——1 hour1.5 hour4 hour30–80 mL/min0.0060″2416424169
———0.5 hour1 hour80–340 mL/min0.0090″22101—