Insert the GC column through the 1/16” adaptor sealing nut and a standard 1/16” graphite or graphite-Vespel capillary ferrule. Make a clean cut across the GC column. Install the column into the adaptor 30-32 mm from the bottom of the 1/16” nut, which places the column end a few millimeters above the bottom of the inlet seal, the perfect installation distance! Don’t forget to leak check with a Restek Electronic Leak Detector (cat.# 28500) after tightening the adaptor nut to seal the GC column to the inlet and re-pressurizing the inlet.

Seat the column sealing nut so that
the bottom is 30-32 mm from
the end of the column.

Insert the column into the adaptor fitting.

Tighten the sealing nut onto the adaptor fitting.

Perfect installation distance for column in adaptor fitting.

