{"id":43700,"date":"2022-09-15T00:00:00","date_gmt":"2022-09-15T00:00:00","guid":{"rendered":"https:\/\/discover.restek.com\/uncategorized\/gc-ms-troubleshooting-avoid-poor-performance-from-slow-acquisition-speeds\/"},"modified":"2026-03-17T17:03:09","modified_gmt":"2026-03-17T17:03:09","slug":"gc-ms-troubleshooting-avoid-poor-performance-from-slow-acquisition-speeds","status":"publish","type":"post","link":"https:\/\/discover.restek.com\/it\/articles\/gnar3935\/gc-ms-troubleshooting-avoid-poor-performance-from-slow-acquisition-speeds","title":{"rendered":"GC-MS Troubleshooting: Avoid Poor Performance from Slow Acquisition Speeds"},"content":{"rendered":"\n<h2 class=\"wp-block-heading\">Abstract<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">Data acquisition rate is an essential component to achieving accurate and precise chromatographic results. A data acquisition rate that is too slow of a speed often leads to loss of sensitivity, precision, and accuracy while an acquisition speed that is too fast may cause excessive noise and decreased sensitivity. In this article, we discuss how to select the ideal MS data acquisition rate to achieve optimal integration and reliably calculate peak areas.<\/p>\n\n\n\n<h2 class=\"wp-block-heading\">Introduction<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">The foundation for quantitative chromatography is the relationship between peak area and compound concentration.&nbsp; The ability to define the shape of the peak is the cornerstone for reliably calculating peak area, but there are method parameters that can hamper our ability to get a well-defined peak, undermining our ability to achieve accurate and precise results.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">To avoid poorly defined peaks, it is important to make sure your mass spectrometer\u2019s data acquisition rate is matched to chromatographic performance. By comparing the width of the chromatographic peak with the MS data acquisition rate, you can verify that a sufficient amount of data is being collected to accurately define the shape of the chromatographic peak.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">In addition to the shape of the peak itself, data acquisition rates can affect how well the mass spectrometer is able to accurately determine the spectral composition of a given peak, which can affect your ability to properly identify a compound.&nbsp;<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">In this article, we\u2019ll explore examples of deformed peak shapes and skewed mass spectra caused by improperly set data acquisition rates and discuss how to correct data acquisition issues.<\/p>\n\n\n\n<h2 class=\"wp-block-heading\">How are Peak Shape and MS Data Acquisition Rate Related?<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">Since most mass spectrometers coupled to chromatographs rely on quadrupole mass analyzers (single or triple quad), we\u2019ll use a single-quad MS for this example.<\/p>\n\n\n\n<h3 class=\"wp-block-heading\"><em>A Data Acquisition Rate that is Too Slow<\/em><\/h3>\n\n\n\n<p class=\"wp-block-paragraph\">Let\u2019s consider a data acquisition rate that is too slow as our first example.&nbsp; In this case, the MS is reporting data twice every second, or at 2 Hz (Figure 1).<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><strong>Figure 1:<\/strong>\u00a0Points Recorded over Time at a Data Acquisition Rate of 2 Hz<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n<style>.kb-image43700_fd1bc6-16 .kb-image-has-overlay:after{opacity:0.3;}<\/style>\n<div class=\"wp-block-kadence-image kb-image43700_fd1bc6-16\"><figure class=\"aligncenter size-medium_large\"><img loading=\"lazy\" decoding=\"async\" width=\"768\" height=\"214\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR3935-01-768x214.jpg\" alt=\"\" class=\"kb-img wp-image-21460\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR3935-01-768x214.jpg 768w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR3935-01-300x84.jpg 300w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR3935-01-1024x286.jpg 1024w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR3935-01-1536x428.jpg 1536w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR3935-01.jpg 1800w\" sizes=\"auto, (max-width: 768px) 100vw, 768px\" \/><\/figure><\/div>\n\n<\/div><\/div><\/div>\n\n\n<p class=\"wp-block-paragraph\">If a 2-second-wide chromatographic peak enters the mass spectrometer, a 2 Hz acquisition speed will only record a few data points across its width.&nbsp; As shown in Figure 2, this results in a jagged-looking peak that cannot capture the complete peak area.&nbsp; At best, this results in some degree of a loss in sensitivity since the available peak area is not fully captured due to the slow acquisition speed.&nbsp; Quantitation accuracy may suffer as well, but since the instrument was likely calibrated under the same low acquisition speed conditions, it may not be off as much as if the instrument had been calibrated using an ideal data acquisition rate.<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><strong>Figure 2:<\/strong>\u00a0A data acquisition rate that is too slow does not capture the true peak shape.<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n<style>.kb-image43700_43902c-8c .kb-image-has-overlay:after{opacity:0.3;}<\/style>\n<div class=\"wp-block-kadence-image kb-image43700_43902c-8c\"><figure class=\"aligncenter size-medium_large\"><img loading=\"lazy\" decoding=\"async\" width=\"768\" height=\"477\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR3935-02-768x477.jpg\" alt=\"\" class=\"kb-img wp-image-21466\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR3935-02-768x477.jpg 768w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR3935-02-300x186.jpg 300w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR3935-02-1024x635.jpg 1024w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR3935-02-1536x953.jpg 1536w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR3935-02.jpg 1800w\" sizes=\"auto, (max-width: 768px) 100vw, 768px\" \/><\/figure><\/div>\n\n<\/div><\/div><\/div>\n\n\n<p class=\"wp-block-paragraph\">While calibrating under the slow acquisition rate mitigates some of the accuracy error, it does not account for the degradation in precision that will likely also occur.&nbsp; Even a small shift in retention time from analysis to analysis can affect how much of the peak is recorded with a slow acquisition rate (Figure 3).<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><strong>Figure 3:<\/strong>\u00a0Varying Retention Times and a Slow Data Acquisition Rate<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n<style>.kb-image43700_92150e-b1 .kb-image-has-overlay:after{opacity:0.3;}<\/style>\n<div class=\"wp-block-kadence-image kb-image43700_92150e-b1\"><figure class=\"aligncenter size-medium_large\"><img loading=\"lazy\" decoding=\"async\" width=\"768\" height=\"409\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR3935-03-768x409.jpg\" alt=\"\" class=\"kb-img wp-image-21472\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR3935-03-768x409.jpg 768w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR3935-03-300x160.jpg 300w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR3935-03-1024x545.jpg 1024w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR3935-03-1536x817.jpg 1536w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR3935-03.jpg 1800w\" sizes=\"auto, (max-width: 768px) 100vw, 768px\" \/><\/figure><\/div>\n\n<\/div><\/div><\/div>\n\n\n<p class=\"wp-block-paragraph\">By doubling the data acquisition speed to 4 Hz, we can see that the true shape of the chromatographic peak is better defined, allowing for better accuracy, precision, and sensitivity (Figure 4).<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><strong>Figure 4:<\/strong>\u00a0Peak Area after Doubling the Data Acquisition Rate<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n<style>.kb-image43700_726458-f0 .kb-image-has-overlay:after{opacity:0.3;}<\/style>\n<div class=\"wp-block-kadence-image kb-image43700_726458-f0\"><figure class=\"aligncenter size-medium_large\"><img loading=\"lazy\" decoding=\"async\" width=\"768\" height=\"411\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR3935-04-768x411.jpg\" alt=\"\" class=\"kb-img wp-image-21478\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR3935-04-768x411.jpg 768w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR3935-04-300x161.jpg 300w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR3935-04-1024x548.jpg 1024w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR3935-04-1536x823.jpg 1536w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR3935-04.jpg 1800w\" sizes=\"auto, (max-width: 768px) 100vw, 768px\" \/><\/figure><\/div>\n\n<\/div><\/div><\/div>\n\n\n<div style=\"height:30px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<h3 class=\"wp-block-heading\"><em>A Data Acquisition Speed that is Too Fast<\/em><\/h3>\n\n\n\n<p class=\"wp-block-paragraph\">If acquiring data too slowly leads to a poorly defined peak shape that reduces sensitivity, precision, and, in some cases, accuracy, it would be easy to think that collecting data as fast as the instrument will allow is a good solution.&nbsp; An infinite number of points across the peaks would perfectly define the peak shape, right?&nbsp; In reality, though, there is a \u201ctoo fast\u201d just as there is a \u201ctoo slow\u201d when it comes to data acquisition rate.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">To explain how a mass spectrometer data acquisition rate can be too fast, we must first understand how data is collected.&nbsp; A quadrupole mass spectrometer often has a balance to strike when acquiring data.&nbsp; As the MS scans the mass range, it does not simply collect a single data point at a given m\/z ratio value.&nbsp; Rather, it collects multiple \u201csamples\u201d and then averages them together.&nbsp; And, as is typical with averages, the greater the sample size, the better a representation the average will be, leading to an improved signal to noise ratio. Collecting all of those samples takes time, though.&nbsp; If you set an overall data acquisition rate that is too fast, you may end up acquiring more points across the chromatographic curve, but the data will be noisier, and the signal to noise will be reduced considerably.&nbsp; The result is a peak that no longer looks \u201cblocky\u201d but one that looks jagged.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Poor peak shape caused by a data acquisition rate that is too fast can pose a challenge to the processing software\u2019s auto-integration, resulting in the need to do more manual data review.&nbsp; Additionally, a scan rate that is too fast can result in significantly lower abundances of the ions detected, which leads to a decrease in sensitivity.<\/p>\n\n\n\n<h2 class=\"wp-block-heading\">What MS Data Acquisition Speed is \u201cJust Right\u201d?<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">How do you avoid the pitfalls of a nonoptimal acquisition rate and select conditions that are \u201cjust right?\u201d&nbsp; Thankfully, there is a helpful rule of thumb.&nbsp; In general, aim to have 10-20 points across the width of the chromatographic peak (8-10 at minimum) to ensure an accurate peak shape for quantitation.&nbsp; Exactly how to set your specific instrument to achieve this goal will depend on the mass spectrometer itself as well as the chromatography.&nbsp; For instance, a peak that is 2 seconds wide will require a data acquisition rate that is twice as fast than a peak that is 4 seconds wide. Tailoring MS conditions as much as possible to match the chromatographic performance is ideal.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Figure 5 illustrates how chromatographic peak shape and overall intensity changes with different scan speeds, highlighting the value of making sure your scan speed is tailored to your chromatography.<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><strong>Figure 5:<\/strong>\u00a0The effects of scan rate on peak shape.<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n<div class=\"wp-block-custom-chromatogram-article\"><div class=\"wp-block-custom-chromatogram-article\"><div class=\"chromatogram-image wide-image\"><img decoding=\"async\" src=\"https:\/\/ez.restek.com\/images\/cgram\/gc_gn1216.png\" alt=\"Impact of MS Scan Rate on Peak Shape and Area\" title=\"-\"><\/div><p class=\"article-id\" style=\"text-align:center\"> GC_GN1216<\/p><div class=\"chromatogram-peaks\"><h4>Peaks<\/h4><table class=\"peaks col-lg-4 col-12\">\n<thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><\/tr><\/thead>\n<tbody><tr><td class=\"num\">1.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Benzo[b]fluoranthene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/205-99-2\/Benzo[b]fluoranthene\" target=\"_blank\" rel=\"noopener\">Benzo[b]fluoranthene<\/a><\/td><\/tr>\n<tr><td class=\"num\">2.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Benzo[k]fluoranthene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/207-08-9\/Benzo[k]fluoranthene\" target=\"_blank\" rel=\"noopener\">Benzo[k]fluoranthene<\/a><\/td><\/tr>\n<\/tbody><\/table><\/div><div class=\"chromatogram-conditions\"><h4>Conditions<\/h4><div class=\"conditions-container container-fluid\"><div class=\"row\"><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Column<\/th><td><\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Injection<\/th><td>split<\/td><\/tr><\/table><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Detector<\/th><td>MS<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Mode:<\/th><td>Scan<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Instrument<\/th><td>Agilent 7890B GC &amp; 5977A MSD<\/td><\/tr><\/table><\/div><\/div><\/div><div class=\"chromatogram-pdf-link\"><a href=\"https:\/\/ez.restek.com\/images\/cgram\/gc_gn1216.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"><svg xmlns=\"http:\/\/www.w3.org\/2000\/svg\" width=\"18\" height=\"18\" viewBox=\"0 0 18 18\"><g data-name=\"Group 2996\"><path data-name=\"Rectangle 1246\" d=\"M0 0h18v18H0z\" style=\"fill: none;\"><\/path><\/g><g data-name=\"Group 2997\"><path data-name=\"Path 729\" d=\"M13.412 11.4v2.017H5.345V11.4H4v2.017a1.349 1.349 0 0 0 1.345 1.345h8.068a1.349 1.349 0 0 0 1.345-1.345V11.4zm-.672-2.694-.948-.948-1.741 1.735V4H8.706v5.493L6.965 7.758l-.948.948 3.361 3.361z\" transform=\"translate(-.437 -.414)\" style=\"fill: rgb(13, 123, 196);\"><\/path><\/g><\/svg>Download PDF<\/a><\/div><\/div><\/div>\n<\/div><\/div><\/div>\n\n\n<p class=\"wp-block-paragraph\">Also, check to see if your instrument allows you to adjust the data acquisition parameters during the course of an analysis for additional optimization. Scanning a narrower, lower range of ions earlier in the chromatographic analysis when lower molecular weight compounds are likely to elute can provide a faster acquisition rate when the peaks are also likely to be the narrowest.<\/p>\n\n\n\n<div style=\"height:30px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n        <div class=\"cpb\">\n            <h3 class=\"cpb-heading\">Prodotti Menzionati<\/h3>\n            <hr class=\"cpb-heading-underline\" \/>\n            <div class=\"cpb-list\">\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/27194\">                                N\u00b0 Catalogo 27194                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Kit di pulizia per spettrometro di massa GC-MS con utensile rotante<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/27194\">Esplora Prodotto<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/23074\">                                N\u00b0 Catalogo 23074                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Moltiplicatore di elettroni ETP, per GC-MS Agilent 5973 e 5975 (non adatto per rivelatore a triplo asse 5975C) (incluso supporto per installazione iniziale)<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/23074\">Esplora Prodotto<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/24819\">                                N\u00b0 Catalogo 24819                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Olio grezzo per pompa Inland 45, 1 litro<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/24819\">Esplora Prodotto<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/22685\">                                N\u00b0 Catalogo 22685                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Polvere di pulizia della sorgente ionica: polvere di allumina per la pulizia delle parti degli spettrometri di massa, 1\u00b0kg<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/22685\">Esplora Prodotto<\/a>\n                                                    <\/div>\n                    <\/div>\n                            <\/div>\n        <\/div>\n        \n\n\n<div style=\"height:100px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n","protected":false},"excerpt":{"rendered":"<p>In this article, we discuss how to choose the optimal MS data acquisition rate for your system by considering the peak width. We also show the consequences when we haven\u2019t optimized the data acquisition rate.<\/p>\n","protected":false},"author":55,"featured_media":89916,"comment_status":"closed","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"_acf_changed":false,"_kad_blocks_custom_css":"","_kad_blocks_head_custom_js":"","_kad_blocks_body_custom_js":"","_kad_blocks_footer_custom_js":"","_kadence_starter_templates_imported_post":false,"_kad_post_transparent":"","_kad_post_title":"","_kad_post_layout":"","_kad_post_sidebar_id":"","_kad_post_content_style":"","_kad_post_vertical_padding":"","_kad_post_feature":"","_kad_post_feature_position":"","_kad_post_header":false,"_kad_post_footer":false,"footnotes":"","_ppma_block_editor_authors":""},"categories":[7],"tags":[],"industries-application":[],"post-badge":[],"resource-type":[569],"product-library":[2388,2372],"resource-technique":[2294,2301],"ppma_author":[584,610],"class_list":["post-43700","post","type-post","status-publish","format-standard","has-post-thumbnail","hentry","category-articles","resource-type-troubleshooting","product-library-detector-supplies","product-library-gas-chromatography-products","resource-technique-gas-chromatography-gc","resource-technique-mass-spectrometry-ms"],"acf":[],"taxonomy_info":{"category":[{"value":7,"label":"Articles"}],"resource-type":[{"value":569,"label":"Troubleshooting"}],"product-library":[{"value":2388,"label":"Detector Supplies"},{"value":2372,"label":"Gas Chromatography Products"}],"resource-technique":[{"value":2294,"label":"Gas Chromatography (GC)"},{"value":2301,"label":"Mass Spectrometry (MS)"}]},"featured_image_src_large":["https:\/\/discover.restek.com\/wp-content\/uploads\/feature-GNAR3935-1024x536.jpg",1024,536,true],"author_info":{"display_name":"Scott Grossman","author_link":"https:\/\/discover.restek.com\/it\/author\/scott-grossman\/"},"comment_info":0,"category_info":[{"term_id":7,"name":"Articles","slug":"articles","term_group":0,"term_taxonomy_id":7,"taxonomy":"category","description":"","parent":0,"count":510,"filter":"raw","cat_ID":7,"category_count":510,"category_description":"","cat_name":"Articles","category_nicename":"articles","category_parent":0}],"tag_info":false,"authors":[{"term_id":584,"user_id":0,"is_guest":1,"slug":"chris-rattray","display_name":"Chris Rattray","avatar_url":"https:\/\/secure.gravatar.com\/avatar\/?s=96&d=mm&r=g","author_category":"1","first_name":"Chris","last_name":"Rattray","user_url":"","job_title":"","description":""},{"term_id":610,"user_id":55,"is_guest":0,"slug":"scott-grossman","display_name":"Scott Grossman","avatar_url":{"url":"https:\/\/discover.restek.com\/wp-content\/uploads\/people-grossman-scott.jpg","url2x":"https:\/\/discover.restek.com\/wp-content\/uploads\/people-grossman-scott.jpg"},"author_category":"1","first_name":"Scott","last_name":"Grossman","user_url":"https:\/\/www.linkedin.com\/in\/scott-grossman-57163970\/","job_title":"Field Sales Representative I","description":""}],"_links":{"self":[{"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/posts\/43700","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/users\/55"}],"replies":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/comments?post=43700"}],"version-history":[{"count":6,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/posts\/43700\/revisions"}],"predecessor-version":[{"id":79487,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/posts\/43700\/revisions\/79487"}],"wp:featuredmedia":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/media\/89916"}],"wp:attachment":[{"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/media?parent=43700"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/categories?post=43700"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/tags?post=43700"},{"taxonomy":"industries-application","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/industries-application?post=43700"},{"taxonomy":"post-badge","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/post-badge?post=43700"},{"taxonomy":"resource-type","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/resource-type?post=43700"},{"taxonomy":"product-library","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/product-library?post=43700"},{"taxonomy":"resource-technique","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/resource-technique?post=43700"},{"taxonomy":"author","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/ppma_author?post=43700"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}