   {"id":45288,"date":"2020-10-26T14:30:00","date_gmt":"2020-10-26T14:30:00","guid":{"rendered":"https:\/\/discover.restek.com\/uncategorized\/residual-pesticides-analysis-of-botanical-ingredients-using-gas-chromatography-triple-quadrupole-mass-spectrometry\/"},"modified":"2026-01-28T17:00:44","modified_gmt":"2026-01-28T17:00:44","slug":"residual-pesticides-analysis-of-botanical-ingredients-using-gas-chromatography-triple-quadrupole-mass-spectrometry","status":"publish","type":"post","link":"https:\/\/discover.restek.com\/it\/articles-fr\/fsar3080\/residual-pesticides-analysis-of-botanical-ingredients-using-gas-chromatography-triple-quadrupole-mass-spectrometry","title":{"rendered":"Residual Pesticides Analysis of Botanical Ingredients Using Gas Chromatography Triple Quadrupole Mass Spectrometry"},"content":{"rendered":"\n<p class=\"wp-block-paragraph\"><em>Originally published by Shimadzu Corporation in collaboration with Restek Corporation.<\/em><\/p>\n\n\n\n<h2 class=\"wp-block-heading\">Introduction<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">The use of dietary supplements is increasing in the United States. These dietary supplements are made from various dried botanicals, which may contain residual pesticides from agricultural practices, and because of that, they have to be monitored to ensure their quality and prevent exposure. The ingredients are dried which requires the standard QuEChERS methods [1] to be modified to overcome this difficulty. Gas chromatography is the best technique to separate multicomponents including coextracted interferences, and because the triple quadruple mass spectrometer (GC-MS\/MS) is highly sensitive this allows analysis of trace level contamination. In this study, we analyzed over 200 compounds simultaneously using a triple quadrupole gas chromatograph mass spectrometer with the modified QuEChERS method.<\/p>\n\n\n\n<h2 class=\"wp-block-heading\">Materials and Methods<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">Pesticides standards, internal standards, quality control standards and a QuEChERS kit were obtained from Restek:<\/p>\n\n\n\n<ul class=\"wp-block-list\">\n<li>GC Multiresidue Pesticide Kit (cat.# 32562)<\/li>\n\n\n\n<li>QuEChERS Internal Standard Mix for GC-MS Analysis (cat.# 33267)<\/li>\n\n\n\n<li>SV Internal Standard Mix (cat.# 31206)<\/li>\n\n\n\n<li>Q-sep QuEChERS Extraction Kit (Original cat.# 23991 discontinued; see&nbsp;cat.# 25848)<\/li>\n<\/ul>\n\n\n\n<p class=\"wp-block-paragraph\">The total number of targets was 232 compounds (220 pesticides, 6 internal standards, and 6 quality control standards). Ginseng, which can be purchased in any store, was used as a matrix. Using this ginseng, matrix-matched calibration standards (1 to 200 ng\/mL) and fortified samples (each two 10 and 50 ng\/g) were prepared. Calibration curves were generated by an internal standard method, weighted 1\/C and the internal standard was PCB52.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">An MRM analytical method was created using the Smart Pesticides Database (Shimadzu). This database has retention indices for all registered compounds, and retention times can be predicted by running an&nbsp;<em>n<\/em>-alkane sample mixture (AART: Automatic Adjustment of Retention Time). According to estimated retention times, Smart MRM creates an optimum data acquisition time program (Figure 1).<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><strong>Figure 1:<\/strong>\u00a0Data Acquisition Time Program Created by Smart MRM<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n<style>.kb-image45288_b4ea3d-48 .kb-image-has-overlay:after{opacity:0.3;}<\/style>\n<div class=\"wp-block-kadence-image kb-image45288_b4ea3d-48\"><figure class=\"aligncenter size-medium_large\"><img loading=\"lazy\" decoding=\"async\" width=\"768\" height=\"720\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-FSAR3080-01-768x720.jpg\" alt=\"\" class=\"kb-img wp-image-20038\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-FSAR3080-01-768x720.jpg 768w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-FSAR3080-01-300x281.jpg 300w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-FSAR3080-01-1024x960.jpg 1024w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-FSAR3080-01-1536x1440.jpg 1536w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-FSAR3080-01.jpg 1800w\" sizes=\"auto, (max-width: 768px) 100vw, 768px\" \/><\/figure><\/div>\n\n<\/div><\/div><\/div>\n\n\n<p class=\"wp-block-paragraph\"><\/p>\n\n\n\n<div style=\"height:20px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<h2 class=\"wp-block-heading\">Extraction and Clean-up Procedure<\/h2>\n\n\n\n<ul class=\"wp-block-list\">\n<li>Weigh 1.0 \u00b1 0.05 g ground ginseng powder into 50 mL polypropylene centrifuge tube.<\/li>\n\n\n\n<li>Add 10 mL HPLC-grade water and vortex the tube vigorously.<\/li>\n\n\n\n<li>Add 10 mL of the ACN\/IS Extraction Solvent.<\/li>\n\n\n\n<li>Allow the tube to sit for 15 min.<\/li>\n\n\n\n<li>Add 4 g anhydrous MgSO<sub>4<\/sub>&nbsp;and 1 g NaCl.<\/li>\n\n\n\n<li>Shake the tube vigorously on a mechanical shaker for 30 min.<\/li>\n\n\n\n<li>Centrifuge the 50 mL tubes at 3000\u20134500 rpm \u00d7 5 min.<\/li>\n\n\n\n<li>Condition the GCB\/PSA (0.25 g\/0.5 g) SPE columns with ~250 mg anhydrous Na<sub>2<\/sub>SO<sub>4<\/sub>&nbsp;on top using 3 column volumes of acetone.<\/li>\n\n\n\n<li>Insert a collection rack consisting of 15 mL disposable glass centrifuge tube on a SPE vacuum manifold.<\/li>\n\n\n\n<li>Add 1.25 mL of the ACN extract.<\/li>\n\n\n\n<li>Rinse with 1 mL acetone.<\/li>\n\n\n\n<li>Elute with 12 mL of 3:1 v\/v acetone:toluene.<\/li>\n\n\n\n<li>Evaporate (50 \u00b0C) the eluent to ~100 \u03bcL gently.<\/li>\n\n\n\n<li>Add 500 \u03bcL of toluene to the Blank\/fortified samples, calibration standard solutions to matrix matched calibration standards.<\/li>\n\n\n\n<li>Add 20 \u03bcL quality control standards (12.5 \u03bcg\/mL) and ~50 mg of anhydrous MgSO<sub>4<\/sub>&nbsp;to all samples.<\/li>\n\n\n\n<li>Vortex for 5 sec.<\/li>\n\n\n\n<li>Centrifuge the tubes at 3000 g \u00d7 5 min.<\/li>\n\n\n\n<li>Transfer the toluene extract using a Pasteur pipette to ALS GC vials.<\/li>\n<\/ul>\n\n\n\n<div style=\"height:20px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<h2 class=\"wp-block-heading\">Result and Discussion<\/h2>\n\n\n\n<h3 class=\"wp-block-heading\">Matrix Matched Calibration<\/h3>\n\n\n\n<p class=\"wp-block-paragraph\">The chromatogram in Figure 2 shows a 10 ng\/mL matrix matched calibration standard. Of the 232 compounds, 230 could be detected in \u00b10.1 min of estimated retention time by AART. The remaining two compounds, 1,4-Dichlorobenzene-d4 and Naphthalene-d8 of six quality control standards, had eluted before 4 min. Although retention times were shifted, they were with identified within about \u00b10.2 min of estimated.<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><strong>Figure 2<\/strong>: MRM Chromatogram of 10 ng\/mL Matrix Matched Calibration Standard (Internal standards and quality control standards are not displayed.)<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n<div class=\"wp-block-custom-chromatogram-article\"><div class=\"wp-block-custom-chromatogram-article\"><div class=\"chromatogram-image wide-image\"><img decoding=\"async\" src=\"https:\/\/ez.restek.com\/images\/cgram\/gc_fs0568.png\" alt=\"Residual Pesticides in Botanicals by GC-MS\/MS\" title=\"-\"><\/div><p class=\"article-id\" style=\"text-align:center\"> GC_FS0568<\/p><div class=\"chromatogram-conditions\"><h4>Conditions<\/h4><div class=\"conditions-container container-fluid\"><div class=\"row\"><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Column<\/th><td>Rxi-5ms, 30 m, 0.25 mm ID, 0.25 \u00b5m (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/13423\" rel=\"noopener\">cat.# 13423<\/a>)<\/td><\/tr><tr><td><\/td><td>using Rxi guard column 5 m, 0.25 mm ID (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/10029\" rel=\"noopener\">cat.# 10029<\/a>)<\/td><\/tr><tr><td><\/td><td>with SilTite \u03bc-Union connector kit (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/23885\" rel=\"noopener\">cat.# 23885<\/a>)<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Standard\/Sample<\/th><td>GC Multiresidue pesticide kit (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/32562\" rel=\"noopener\">cat.# 32562<\/a>)<\/td><\/tr><tr><td><\/td><\/td><\/tr>\n<tr><th class=\"sub conditions_header\" scope=\"row\">Diluent:<\/th><td>Toluene<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Conc.:<\/th><td>10 ng\/mL<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Injection<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Vol.:<\/th><td>2 \u00b5L pulsed splitless<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Liner:<\/th><td>Topaz 3.5 mm ID single taper inlet liner w\/wool (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/23336\" rel=\"noopener\">cat.# 23336<\/a>)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Temp.:<\/th><td>250 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Pulse Pressure:<\/th><td>36 psi (248.2kPa)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Pulse Time:<\/th><td>1.5  min<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\" colspan=\"2\">Oven<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Oven Temp.:<\/th><td>90 &#176;C (hold 1 min) to 130 &#176;C at 30 &#176;C\/min to 330 &#176;C at 10 &#176;C\/min (hold 2 min)<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Carrier Gas<\/th><td>He, constant linear velocity<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Linear Velocity:<\/th><td>55 cm\/sec<\/td><\/tr><\/table><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Detector<\/th><td>Shimadzu GCMS TQ8040<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Transfer Line Temp.:<\/th><td>290 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Analyzer Type:<\/th><td>Quadrupole <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Source Temp.:<\/th><td>230 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Electron Energy:<\/th><td>70 eV<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Ionization Mode:<\/th><td>EI <\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Instrument<\/th><td>Shimadzu GCMS-TQ8040<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Notes<\/th><td>Matrix-matched calibration standards were prepared in ginseng. Calibration curves were generated by the internal standard method using PCB52 as the internal standard. An MRM analytical method was created using the Shimadzu pesticides database. <br \/><br \/>Although the multiresidue pesticides kit mixes are formulated to ensure maximum long-term stability and reliability as packaged, stability may become an issue when a large number of compounds with different chemical functionalities are combined together into a single mix. This should be taken into consideration for quantitative analysis.<br \/><br \/><br \/><\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Acknowledgement<\/th><td>Chromatogram provided by Shimadzu. Publication 3655-11615-10ANS (C146-E334), First Edition, December 2016. Residual Pesticides Analysis of Botanical Ingredients Using Gas Chromatography Triple Quadrupole Mass Spectrometry. Riki Kitano, Tairo Ogura, Nicole Lock, Robert Clifford, Julie Kowalski, Jack Cochran, Dan Li<\/td><\/tr><\/table><\/div><\/div><\/div><div class=\"chromatogram-related-skus\"><h4>Related SKUs<\/h4><ul><li>32562<\/li><li>13423<\/li><li>10029<\/li><li>23885<\/li><li>23336<\/li><\/ul><\/div><div class=\"chromatogram-related-searches\"><h4>Related Searches<\/h4><ul><li>QuEChERS<\/li><li>Rxi<\/li><li>botanical<\/li><li>pesticides<\/li><li>pesticides-food-gc-library<\/li><\/ul><\/div><div class=\"chromatogram-pdf-link\"><a href=\"https:\/\/ez.restek.com\/images\/cgram\/gc_fs0568.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"><svg xmlns=\"http:\/\/www.w3.org\/2000\/svg\" width=\"18\" height=\"18\" viewBox=\"0 0 18 18\"><g data-name=\"Group 2996\"><path data-name=\"Rectangle 1246\" d=\"M0 0h18v18H0z\" style=\"fill: none;\"><\/path><\/g><g data-name=\"Group 2997\"><path data-name=\"Path 729\" d=\"M13.412 11.4v2.017H5.345V11.4H4v2.017a1.349 1.349 0 0 0 1.345 1.345h8.068a1.349 1.349 0 0 0 1.345-1.345V11.4zm-.672-2.694-.948-.948-1.741 1.735V4H8.706v5.493L6.965 7.758l-.948.948 3.361 3.361z\" transform=\"translate(-.437 -.414)\" style=\"fill: rgb(13, 123, 196);\"><\/path><\/g><\/svg>Download PDF<\/a><\/div><\/div><\/div>\n<\/div><\/div><\/div>\n\n\n<p class=\"wp-block-paragraph\"><\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Calibration curves were generated from matrix matched calibration standards, then back calculation and linearity were evaluated.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Back calculation was performed by calculating the concentration of each calibration point, and if the concentration exceeds \u00b120% of theoretical value, the calibration point would be interpolated with the nearest two points. Over 93% of the compounds with concentration of 1 ng\/mL were within \u00b120% of theoretical calculations and all compounds of concentration 20 to 200 ng\/mL were within \u00b120% (Figure 3).<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><strong>Figure 3:<\/strong>\u00a0Difference Between Back Calculation and Theoretical Concentration<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n<style>.kb-image45288_982ec3-53 .kb-image-has-overlay:after{opacity:0.3;}<\/style>\n<div class=\"wp-block-kadence-image kb-image45288_982ec3-53\"><figure class=\"aligncenter size-large\"><img loading=\"lazy\" decoding=\"async\" width=\"1024\" height=\"548\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-FSAR3080-03-1024x548.jpg\" alt=\"\" class=\"kb-img wp-image-20044\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-FSAR3080-03-1024x548.jpg 1024w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-FSAR3080-03-300x161.jpg 300w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-FSAR3080-03-768x411.jpg 768w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-FSAR3080-03-1536x823.jpg 1536w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-FSAR3080-03.jpg 1800w\" sizes=\"auto, (max-width: 1024px) 100vw, 1024px\" \/><\/figure><\/div>\n\n<\/div><\/div><\/div>\n\n\n<p class=\"wp-block-paragraph\"><\/p>\n\n\n\n<p class=\"wp-block-paragraph\">This modified QuEChERS method contains dilution steps, and samples will be diluted by quarter. To quantify 10 ng\/g concentration, 2 ng\/g or lower calibration point are required. Even at low concentrations, the calibration curves show good linearity (Figure 4) and all coefficients of determination (220 pesticides) were greater than 0.99.<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><strong>Figure 4:<\/strong>\u00a0Calibration Curves of Representative Six Compounds<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n<style>.kb-image45288_dacf43-8a .kb-image-has-overlay:after{opacity:0.3;}<\/style>\n<div class=\"wp-block-kadence-image kb-image45288_dacf43-8a\"><figure class=\"aligncenter size-large\"><img loading=\"lazy\" decoding=\"async\" width=\"1024\" height=\"605\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-FSAR3080-04-1024x605.jpg\" alt=\"\" class=\"kb-img wp-image-20050\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-FSAR3080-04-1024x605.jpg 1024w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-FSAR3080-04-300x177.jpg 300w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-FSAR3080-04-768x454.jpg 768w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-FSAR3080-04-1536x908.jpg 1536w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-FSAR3080-04.jpg 1800w\" sizes=\"auto, (max-width: 1024px) 100vw, 1024px\" \/><\/figure><\/div>\n\n<\/div><\/div><\/div>\n\n\n<p class=\"wp-block-paragraph\"><\/p>\n\n\n\n<div style=\"height:40px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<h2 class=\"wp-block-heading\">Recovery of Fortified Samples<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">Each two 10 ng\/g and 50 ng\/g fortified samples were prepared (10 ng\/g-1, 10 ng\/g-2, 50 ng\/g-1, 50 ng\/g-2) and these recovery rates were evaluated from the average of three successive data points for each sample.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Of the compounds, 85% showed good recovery within the range of 70 to 120% on 10 ng\/g-1 and 50 ng\/g-2. As mentioned previously, fortified samples were diluted to 2.5 ng\/mL and 12.5 ng\/mL. Since calibration curves showed good linearity at low concentration and modified QuEChERS method suppressed interference, good recovery results were achieved. (Some compounds were not quantified correctly because the matrix for the calibration standards originally contained them. Y-intercept were lifted up and this shift might cause incorrect quantification, especially at low concentration. From the standard addition method, 16 pesticides were detected with more than 10 ng\/g in matrix.)<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">In this study, recovery results were rechecked and combined with qualitative information, the relative ion ratio. Ion ratios between the target and reference were compared to that of the 100 ng\/g standard and evaluated according to SANCO\/12571\/2013 [2]). This mentions that the use of relative ratio \u00b130% as a criterion is recommended.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Table I shows recovery and relative ion ratio for all compounds and Figure 5 shows the combination map of recovery and relative ion ratio, which was generated from this table. Of the compounds, 76% in the 10 ng\/g-1 were within \u00b130% on relative ion ratio with good recovery of between 70 to 120%. For 50 ng\/g-2, 83% of the compounds were within a \u00b130% relative ion ratio. And here, compounds which showed poor recovery and\/or over \u00b130% relative ion ratio were examined.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\"><strong>Table I:<\/strong>&nbsp;Recovery and Relative Ion Ratio; Relative ion ratios were calculated by those of 100 ng\/mL standard solution.<\/p>\n\n\n<figure class=\"wp-block-table align left table table-bordered\">\n<table>\n<tbody>\n<tr>\n<td rowspan=\"2\"><strong>ID<\/strong><\/td>\n<td style=\"text-align: left;\" rowspan=\"2\"><strong>Compound Name<\/strong><\/td>\n<td class=\"has-text-align-left\" colspan=\"4\" data-align=\"left\"><strong>Transitions<\/strong><\/td>\n<td class=\"has-text-align-left\" colspan=\"4\" data-align=\"left\"><strong>Recovery<\/strong><br role=\"presentation\" data-uw-rm-sr=\"\" \/><strong>(Average of n = 3)<\/strong><\/td>\n<td class=\"has-text-align-left\" colspan=\"4\" data-align=\"left\"><strong>Relative Ion Ratio<\/strong><br role=\"presentation\" data-uw-rm-sr=\"\" \/><strong>(Average of n = 3)<\/strong><\/td>\n<\/tr>\n<tr>\n<td style=\"text-align: left;\"><strong>Target<\/strong><\/td>\n<td style=\"text-align: left;\"><strong>CE<\/strong><\/td>\n<td class=\"has-text-align-left\" data-align=\"left\"><strong>Reference<\/strong><\/td>\n<td class=\"has-text-align-left\" data-align=\"left\"><strong>CE<\/strong><\/td>\n<td class=\"has-text-align-left\" data-align=\"left\"><strong>10 ng\/g-1<\/strong><\/td>\n<td><strong>10 ng\/g-2<\/strong><\/td>\n<td><strong>50 ng\/g-1<\/strong><\/td>\n<td><strong>50 ng\/g-2<\/strong><\/td>\n<td><strong>10 ng\/g-1<\/strong><\/td>\n<td><strong>10 ng\/g-2<\/strong><\/td>\n<td><strong>50 ng\/g-1<\/strong><\/td>\n<td><strong>50 ng\/g-2<\/strong><\/td>\n<\/tr>\n<tr>\n<td>1<\/td>\n<td>2,3,5,6-Tetrachloroaniline<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">228.9 &gt; 158.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">18<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">230.9 &gt; 158.0<\/td>\n<td>22<\/td>\n<td>72.5<\/td>\n<td>67.0<\/td>\n<td>61.4<\/td>\n<td>68.8<\/td>\n<td>93.6<\/td>\n<td>97.2<\/td>\n<td>98.9<\/td>\n<td>101.1<\/td>\n<\/tr>\n<tr>\n<td>2<\/td>\n<td>2,4&#8242;-Methoxychlor<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">227.1 &gt; 121.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">16<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">228.1 &gt; 122.1<\/td>\n<td>16<\/td>\n<td>83.0<\/td>\n<td>101.1<\/td>\n<td>86.5<\/td>\n<td>80.2<\/td>\n<td>94.7<\/td>\n<td>105.0<\/td>\n<td>104.6<\/td>\n<td>106.7<\/td>\n<\/tr>\n<tr>\n<td>3<\/td>\n<td>2-Phenylphenol<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">170.1 &gt; 141.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">24<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">170.1 &gt; 115.1<\/td>\n<td>28<\/td>\n<td>72.7<\/td>\n<td>72.3<\/td>\n<td>63.2<\/td>\n<td>69.0<\/td>\n<td>113.1<\/td>\n<td>112.9<\/td>\n<td>99.1<\/td>\n<td>102.1<\/td>\n<\/tr>\n<tr>\n<td>4<\/td>\n<td>3,4-Dichloroaniline<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">161.0 &gt; 99.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">22<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">161.0 &gt; 126.0<\/td>\n<td>14<\/td>\n<td>74.5<\/td>\n<td>66.2<\/td>\n<td>56.4<\/td>\n<td>64.4<\/td>\n<td>103.8<\/td>\n<td>105.5<\/td>\n<td>105.8<\/td>\n<td>106.0<\/td>\n<\/tr>\n<tr>\n<td>5<\/td>\n<td>4,4&#8242;-Dichlorobenzophenone<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">139.0 &gt; 111.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">14<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">139.0 &gt; 75.0<\/td>\n<td>26<\/td>\n<td>77.4<\/td>\n<td>83.2<\/td>\n<td>73.5<\/td>\n<td>73.8<\/td>\n<td>97.5<\/td>\n<td>104.7<\/td>\n<td>102.0<\/td>\n<td>100.3<\/td>\n<\/tr>\n<tr>\n<td>6<\/td>\n<td>4,4&#8242;-methoxychlor olefin<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">308.0 &gt; 238.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">16<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">310.0 &gt; 238.1<\/td>\n<td>20<\/td>\n<td>84.4<\/td>\n<td>94.8<\/td>\n<td>85.7<\/td>\n<td>83.0<\/td>\n<td>99.4<\/td>\n<td>101.0<\/td>\n<td>100.0<\/td>\n<td>98.4<\/td>\n<\/tr>\n<tr>\n<td>7<\/td>\n<td>Acequinocyl deg.<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">342.2 &gt; 188.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">14<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">342.2 &gt; 160.1<\/td>\n<td>22<\/td>\n<td>105.3<\/td>\n<td>269.8<\/td>\n<td>201.1<\/td>\n<td>115.2<\/td>\n<td>61.2<\/td>\n<td>62.8<\/td>\n<td>90.7<\/td>\n<td>64.0<\/td>\n<\/tr>\n<tr>\n<td>8<\/td>\n<td>Acetochlor<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">223.1 &gt; 132.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">22<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">223.1 &gt; 147.1<\/td>\n<td>10<\/td>\n<td>77.5<\/td>\n<td>88.3<\/td>\n<td>77.4<\/td>\n<td>77.1<\/td>\n<td>109.0<\/td>\n<td>104.1<\/td>\n<td>102.8<\/td>\n<td>106.2<\/td>\n<\/tr>\n<tr>\n<td>9<\/td>\n<td>Acrinathrin<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">289.1 &gt; 93.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">14<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">181.1 &gt; 152.1<\/td>\n<td>26<\/td>\n<td>102.5<\/td>\n<td>121.4<\/td>\n<td>91.0<\/td>\n<td>82.4<\/td>\n<td>87.0<\/td>\n<td>94.7<\/td>\n<td>92.3<\/td>\n<td>95.1<\/td>\n<\/tr>\n<tr>\n<td>10<\/td>\n<td>Alachlor<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">188.1 &gt; 160.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">10<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">188.1 &gt; 132.1<\/td>\n<td>18<\/td>\n<td>74.1<\/td>\n<td>87.2<\/td>\n<td>77.8<\/td>\n<td>74.1<\/td>\n<td>114.7<\/td>\n<td>107.2<\/td>\n<td>97.3<\/td>\n<td>103.5<\/td>\n<\/tr>\n<tr>\n<td>11<\/td>\n<td>Aldrin<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">262.9 &gt; 191.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">34<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">262.9 &gt; 193.0<\/td>\n<td>28<\/td>\n<td>86.0<\/td>\n<td>66.9<\/td>\n<td>65.2<\/td>\n<td>74.3<\/td>\n<td>79.8<\/td>\n<td>101.1<\/td>\n<td>104.6<\/td>\n<td>93.2<\/td>\n<\/tr>\n<tr>\n<td>12<\/td>\n<td>Allidochlor<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">132.1 &gt; 56.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">8<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">132.1 &gt; 49.0<\/td>\n<td>24<\/td>\n<td>71.8<\/td>\n<td>64.5<\/td>\n<td>59.0<\/td>\n<td>65.9<\/td>\n<td>123.7<\/td>\n<td>118.4<\/td>\n<td>113.4<\/td>\n<td>117.2<\/td>\n<\/tr>\n<tr>\n<td>13<\/td>\n<td>Anthraquinone<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">208.1 &gt; 180.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">10<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">208.1 &gt; 152.1<\/td>\n<td>22<\/td>\n<td>0.0<\/td>\n<td>126.7<\/td>\n<td>47.4<\/td>\n<td>48.0<\/td>\n<td>97.0<\/td>\n<td>83.6<\/td>\n<td>90.4<\/td>\n<td>94.0<\/td>\n<\/tr>\n<tr>\n<td>14<\/td>\n<td>Atrazine<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">200.1 &gt; 104.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">18<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">200.1 &gt; 122.1<\/td>\n<td>8<\/td>\n<td>77.8<\/td>\n<td>94.0<\/td>\n<td>83.3<\/td>\n<td>78.6<\/td>\n<td>105.3<\/td>\n<td>110.1 \u00a0<\/td>\n<td>92.5<\/td>\n<td>100.6<\/td>\n<\/tr>\n<tr>\n<td>15<\/td>\n<td>Azinphos-ethyl<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">160.1 &gt; 132.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">4<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">160.1 &gt; 77.0<\/td>\n<td>18<\/td>\n<td>91.4<\/td>\n<td>115.5<\/td>\n<td>94.5<\/td>\n<td>86.5<\/td>\n<td>102.1<\/td>\n<td>94.8<\/td>\n<td>100.0<\/td>\n<td>98.7<\/td>\n<\/tr>\n<tr>\n<td>16<\/td>\n<td>Azinphos-methyl<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">160.1 &gt; 132.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">6<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">160.1 &gt; 77.0<\/td>\n<td>20<\/td>\n<td>84.1<\/td>\n<td>124.7<\/td>\n<td>93.0<\/td>\n<td>83.3<\/td>\n<td>89.6<\/td>\n<td>103.3<\/td>\n<td>98.2<\/td>\n<td>100.0<\/td>\n<\/tr>\n<tr>\n<td>17<\/td>\n<td>Benfluralin<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">292.1 &gt; 264.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">8<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">292.1 &gt; 160.0<\/td>\n<td>22<\/td>\n<td>78.4<\/td>\n<td>77.0<\/td>\n<td>64.5<\/td>\n<td>70.7<\/td>\n<td>97.7<\/td>\n<td>89.1<\/td>\n<td>96.8<\/td>\n<td>95.3<\/td>\n<\/tr>\n<tr>\n<td>18<\/td>\n<td>BHC,\u00a0alpha-<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">180.9 &gt; 144.9<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">16<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">218.9 &gt; 182.9<\/td>\n<td>8<\/td>\n<td>62.5<\/td>\n<td>50.4<\/td>\n<td>58.2<\/td>\n<td>66.1<\/td>\n<td>99.0<\/td>\n<td>104.0<\/td>\n<td>99.2<\/td>\n<td>104.3<\/td>\n<\/tr>\n<tr>\n<td>19<\/td>\n<td>BHC,\u00a0beta-<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">180.9 &gt; 144.9<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">16<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">218.9 &gt; 182.9<\/td>\n<td>8<\/td>\n<td>69.6<\/td>\n<td>87.6<\/td>\n<td>76.8<\/td>\n<td>71.8<\/td>\n<td>102.2<\/td>\n<td>102.9<\/td>\n<td>98.9<\/td>\n<td>104.1<\/td>\n<\/tr>\n<tr>\n<td>20<\/td>\n<td>BHC,\u00a0delta-<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">180.9 &gt; 144.9<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">16<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">218.9 &gt; 182.9<\/td>\n<td>8<\/td>\n<td>28.6<\/td>\n<td>87.6<\/td>\n<td>69.0<\/td>\n<td>67.8<\/td>\n<td>102.5<\/td>\n<td>104.3<\/td>\n<td>106.4<\/td>\n<td>104.2<\/td>\n<\/tr>\n<tr>\n<td>21<\/td>\n<td>BHC,\u00a0gamma-<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">180.9 &gt; 144.9<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">16<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">218.9 &gt; 182.9<\/td>\n<td>8<\/td>\n<td>64.1<\/td>\n<td>73.3<\/td>\n<td>62.0<\/td>\n<td>69.8<\/td>\n<td>94.8<\/td>\n<td>93.9<\/td>\n<td>103.1<\/td>\n<td>102.8<\/td>\n<\/tr>\n<tr>\n<td>22<\/td>\n<td>Bifenthrin<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">181.1 &gt; 166.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">12<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">181.1 &gt; 179.1<\/td>\n<td>12<\/td>\n<td>84.6<\/td>\n<td>99.2<\/td>\n<td>89.3<\/td>\n<td>81.4<\/td>\n<td>97.0<\/td>\n<td>104.4<\/td>\n<td>116.7<\/td>\n<td>105.9<\/td>\n<\/tr>\n<tr>\n<td>23<\/td>\n<td>Bioallethrin<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">123.1 &gt; 81.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">10<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">136.1 &gt; 93.1<\/td>\n<td>14<\/td>\n<td>81.7<\/td>\n<td>100.7<\/td>\n<td>83.6<\/td>\n<td>71.9<\/td>\n<td>463.4<\/td>\n<td>563.8<\/td>\n<td>200.4<\/td>\n<td>228.1<\/td>\n<\/tr>\n<tr>\n<td>24<\/td>\n<td>Biphenyl<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">154.1 &gt; 128.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">22<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">154.1 &gt; 115.1<\/td>\n<td>24<\/td>\n<td>90.0<\/td>\n<td>67.9<\/td>\n<td>56.2<\/td>\n<td>65.1<\/td>\n<td>105.5<\/td>\n<td>105.1<\/td>\n<td>106.3<\/td>\n<td>106.0<\/td>\n<\/tr>\n<tr>\n<td>25<\/td>\n<td>Bromfenvinfos-methyl<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">294.9 &gt; 109.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">16<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">296.9 &gt; 109.0<\/td>\n<td>16<\/td>\n<td>85.5<\/td>\n<td>98.5<\/td>\n<td>84.2<\/td>\n<td>77.6<\/td>\n<td>103.5<\/td>\n<td>92.2<\/td>\n<td>100.9<\/td>\n<td>102.7<\/td>\n<\/tr>\n<tr>\n<td>26<\/td>\n<td>Bromfenvinphos<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">266.9 &gt; 159.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">14<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">268.9 &gt; 161.0<\/td>\n<td>16<\/td>\n<td>78.<\/td>\n<td>97.8<\/td>\n<td>88.2<\/td>\n<td>79.2<\/td>\n<td>102.1<\/td>\n<td>94.7<\/td>\n<td>100.9<\/td>\n<td>102.3<\/td>\n<\/tr>\n<tr>\n<td>27<\/td>\n<td>Bromophos<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">330.9 &gt; 315.9<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">14<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">328.9 &gt; 313.9<\/td>\n<td>18<\/td>\n<td>70.9<\/td>\n<td>86.2<\/td>\n<td>77.2<\/td>\n<td>76.4<\/td>\n<td>104.6<\/td>\n<td>98.6<\/td>\n<td>97.2<\/td>\n<td>98.6<\/td>\n<\/tr>\n<tr>\n<td>28<\/td>\n<td>Bromophos-ethyl<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">358.9 &gt; 302.9<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">16<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">302.9 &gt; 284.9<\/td>\n<td>18<\/td>\n<td>78.0<\/td>\n<td>86.9<\/td>\n<td>76.5<\/td>\n<td>75.8<\/td>\n<td>91.9<\/td>\n<td>89.6<\/td>\n<td>98.5<\/td>\n<td>98.4<\/td>\n<\/tr>\n<tr>\n<td>29<\/td>\n<td>Bromopropylate<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">340.9 &gt; 182.9<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">18<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">340.9 &gt; 184.9<\/td>\n<td>20<\/td>\n<td>85.8<\/td>\n<td>104.5<\/td>\n<td>94.3<\/td>\n<td>85.6<\/td>\n<td>101.4<\/td>\n<td>100.9<\/td>\n<td>101.0<\/td>\n<td>99.2<\/td>\n<\/tr>\n<tr>\n<td>30<\/td>\n<td>Bupirimate<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">273.1 &gt; 108.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">16<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">273.1 &gt; 193.1<\/td>\n<td>8<\/td>\n<td>94.4<\/td>\n<td>102.8<\/td>\n<td>92.2<\/td>\n<td>87.0<\/td>\n<td>82.0<\/td>\n<td>98.3<\/td>\n<td>98.1<\/td>\n<td>92.8<\/td>\n<\/tr>\n<tr>\n<td>31<\/td>\n<td>Captafol<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">79.0 &gt; 77.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">14<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">79.0 &gt; 51.0<\/td>\n<td>20<\/td>\n<td>N.D.<\/td>\n<td>N.D.<\/td>\n<td>N.D.<\/td>\n<td>N.D.<\/td>\n<td>N.D.<\/td>\n<td>N.D.<\/td>\n<td>N.D.<\/td>\n<td>N.D.<\/td>\n<\/tr>\n<tr>\n<td>32<\/td>\n<td>Captan<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">149.1 &gt; 105.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">4<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">149.1 &gt; 79.1<\/td>\n<td>14<\/td>\n<td>90.7<\/td>\n<td>95.3<\/td>\n<td>83.4<\/td>\n<td>75.3<\/td>\n<td>232.6<\/td>\n<td>240.2<\/td>\n<td>118.7<\/td>\n<td>135.5<\/td>\n<\/tr>\n<tr>\n<td>33<\/td>\n<td>Carbophenothion<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">341.9 &gt; 157.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">14<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">341.9 &gt; 199.0<\/td>\n<td>8<\/td>\n<td>81.9<\/td>\n<td>97.6<\/td>\n<td>84.2<\/td>\n<td>79.2<\/td>\n<td>58.6<\/td>\n<td>66.3<\/td>\n<td>80.4<\/td>\n<td>91.6<\/td>\n<\/tr>\n<tr>\n<td>34<\/td>\n<td>Carfentrazone-ethyl<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">340.1 &gt; 312.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">14<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">312.1 &gt; 151.1<\/td>\n<td>24<\/td>\n<td>85.3<\/td>\n<td>100.6<\/td>\n<td>94.0<\/td>\n<td>87.3<\/td>\n<td>81.5<\/td>\n<td>81.8<\/td>\n<td>101.9<\/td>\n<td>94.7<\/td>\n<\/tr>\n<tr>\n<td>35<\/td>\n<td>Chlorbenside<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">125.0 &gt; 99.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">18<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">127.0 &gt; 89.0<\/td>\n<td>18<\/td>\n<td>80.3<\/td>\n<td>83.4<\/td>\n<td>73.5<\/td>\n<td>71.8<\/td>\n<td>86.9<\/td>\n<td>93.5<\/td>\n<td>96.8<\/td>\n<td>99.3<\/td>\n<\/tr>\n<tr>\n<td>36<\/td>\n<td>Chlordane,\u00a0cis-<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">374.8 &gt; 265.9<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">26<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">372.8 &gt; 265.9<\/td>\n<td>22<\/td>\n<td>76.4<\/td>\n<td>84.2<\/td>\n<td>76.6<\/td>\n<td>77.1<\/td>\n<td>94.8<\/td>\n<td>99.2<\/td>\n<td>105.3<\/td>\n<td>101.1<\/td>\n<\/tr>\n<tr>\n<td>37<\/td>\n<td>Chlordane,\u00a0trans-<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">374.8 &gt; 265.9<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">26<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">372.8 &gt; 265.9<\/td>\n<td>22<\/td>\n<td>76.6<\/td>\n<td>77.3<\/td>\n<td>75.9<\/td>\n<td>72.9<\/td>\n<td>96.3<\/td>\n<td>106.5<\/td>\n<td>103.2<\/td>\n<td>102.7<\/td>\n<\/tr>\n<tr>\n<td>38<\/td>\n<td>Chlorfenapyr<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">247.1 &gt; 227.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">16<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">247.1 &gt; 200.0<\/td>\n<td>24<\/td>\n<td>84.3<\/td>\n<td>114.9<\/td>\n<td>87.1<\/td>\n<td>85.0<\/td>\n<td>51.3<\/td>\n<td>71.3<\/td>\n<td>112.2<\/td>\n<td>98.6<\/td>\n<\/tr>\n<tr>\n<td>39<\/td>\n<td>Chlorfenson<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">175.0 &gt; 111.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">12<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">301.9 &gt; 175.0<\/td>\n<td>8<\/td>\n<td>77.0<\/td>\n<td>89.1<\/td>\n<td>81.4<\/td>\n<td>76.4<\/td>\n<td>99.5<\/td>\n<td>100.6<\/td>\n<td>100.5<\/td>\n<td>98.0<\/td>\n<\/tr>\n<tr>\n<td>40<\/td>\n<td>Chlorfenvinphos, (E)-<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">323.0 &gt; 267.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">16<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">267.0 &gt; 159.0<\/td>\n<td>18<\/td>\n<td>110.3<\/td>\n<td>114.8<\/td>\n<td>74.7<\/td>\n<td>72.9<\/td>\n<td>270.4<\/td>\n<td>484.8<\/td>\n<td>103.9<\/td>\n<td>99.9<\/td>\n<\/tr>\n<tr>\n<td>41<\/td>\n<td>Chlorfenvinphos, (Z)-<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">323.0 &gt; 267.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">16<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">267.0 &gt; 159.0<\/td>\n<td>18<\/td>\n<td>85.5<\/td>\n<td>96.6<\/td>\n<td>82.7<\/td>\n<td>78.5<\/td>\n<td>105.8<\/td>\n<td>101.5<\/td>\n<td>104.8<\/td>\n<td>101.2<\/td>\n<\/tr>\n<tr>\n<td>42<\/td>\n<td>Chlorobenzilate<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">251.0 &gt; 139.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">14<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">139.0 &gt; 75.0<\/td>\n<td>26<\/td>\n<td>87.5<\/td>\n<td>106.9<\/td>\n<td>88.9<\/td>\n<td>80.3<\/td>\n<td>98.3<\/td>\n<td>92.0<\/td>\n<td>100.1<\/td>\n<td>101.1<\/td>\n<\/tr>\n<tr>\n<td>43<\/td>\n<td>Chloroneb<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">206.0 &gt; 141.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">20<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">193.0 &gt; 113.0<\/td>\n<td>18<\/td>\n<td>65.4<\/td>\n<td>59.8<\/td>\n<td>59.8<\/td>\n<td>68.6<\/td>\n<td>111.1<\/td>\n<td>110.8<\/td>\n<td>107.9<\/td>\n<td>105.9<\/td>\n<\/tr>\n<tr>\n<td>44<\/td>\n<td>Chlorothalonil<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">263.9 &gt; 168.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">24<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">263.9 &gt; 228.8<\/td>\n<td>18<\/td>\n<td>N.D.<\/td>\n<td>N.D.<\/td>\n<td>5.2<\/td>\n<td>N.D.<\/td>\n<td>N.D.<\/td>\n<td>N.D.<\/td>\n<td>82.3<\/td>\n<td>N.D.<\/td>\n<\/tr>\n<tr>\n<td>45<\/td>\n<td>Chlorpropham<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">213.1 &gt; 171.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">6<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">127.1 &gt; 92.0<\/td>\n<td>18<\/td>\n<td>77.7<\/td>\n<td>81.7<\/td>\n<td>72.5<\/td>\n<td>73.3<\/td>\n<td>99.6<\/td>\n<td>101.7<\/td>\n<td>99.0<\/td>\n<td>101.4<\/td>\n<\/tr>\n<tr>\n<td>46<\/td>\n<td>Chlorpyrifos<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">313.9 &gt; 257.9<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">14<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">313.9 &gt; 285.9<\/td>\n<td>8<\/td>\n<td>71.2<\/td>\n<td>77.8<\/td>\n<td>74.9<\/td>\n<td>74.2<\/td>\n<td>100.5<\/td>\n<td>93.8<\/td>\n<td>93.6<\/td>\n<td>95.7<\/td>\n<\/tr>\n<tr>\n<td>47<\/td>\n<td>Chlorpyrifos-methyl<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">285.9 &gt; 93.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">22<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">287.9 &gt; 93.0<\/td>\n<td>22<\/td>\n<td>77.0<\/td>\n<td>89.3<\/td>\n<td>73.1<\/td>\n<td>75.3<\/td>\n<td>92.9<\/td>\n<td>91.7<\/td>\n<td>102.6<\/td>\n<td>98.0<\/td>\n<\/tr>\n<tr>\n<td>48<\/td>\n<td>Chlorthal-dimethyl<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">298.9 &gt; 220.9<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">24<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">300.9 &gt; 222.9<\/td>\n<td>26<\/td>\n<td>78.8<\/td>\n<td>81.9<\/td>\n<td>77.3<\/td>\n<td>73.7<\/td>\n<td>90.7<\/td>\n<td>95.7<\/td>\n<td>96.1<\/td>\n<td>99.9<\/td>\n<\/tr>\n<tr>\n<td>49<\/td>\n<td>Chlorthiophos-1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">256.9 &gt; 239.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">14<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">256.9 &gt; 193.0<\/td>\n<td>22<\/td>\n<td>91.4<\/td>\n<td>108.5<\/td>\n<td>84.2<\/td>\n<td>83.5<\/td>\n<td>8.4<\/td>\n<td>49.4<\/td>\n<td>77.2<\/td>\n<td>68.6<\/td>\n<\/tr>\n<tr>\n<td>50<\/td>\n<td>Chlorthiophos-2<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">324.9 &gt; 268.9<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">14<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">268.9 &gt; 205.0<\/td>\n<td>18<\/td>\n<td>76.1<\/td>\n<td>96.8<\/td>\n<td>84.6<\/td>\n<td>80.0<\/td>\n<td>73.8<\/td>\n<td>79.3<\/td>\n<td>95.8<\/td>\n<td>101.5<\/td>\n<\/tr>\n<tr>\n<td>51<\/td>\n<td>Chlorthiophos-3<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">324.9 &gt; 268.9<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">14<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">268.9 &gt; 205.0<\/td>\n<td>18<\/td>\n<td>78.4<\/td>\n<td>100.2<\/td>\n<td>86.8<\/td>\n<td>77.8<\/td>\n<td>92.6<\/td>\n<td>85.9<\/td>\n<td>100.8<\/td>\n<td>97.8<\/td>\n<\/tr>\n<tr>\n<td>52<\/td>\n<td>Chlozolinate<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">258.9 &gt; 188.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">14<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">330.9 &gt; 258.9<\/td>\n<td>6<\/td>\n<td>75.6<\/td>\n<td>82.2<\/td>\n<td>75.9<\/td>\n<td>76.1<\/td>\n<td>91.7<\/td>\n<td>98.3<\/td>\n<td>103.8<\/td>\n<td>104.3<\/td>\n<\/tr>\n<tr>\n<td>53<\/td>\n<td>Clomazone<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">204.1 &gt; 107.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">20<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">204.1 &gt; 78.0<\/td>\n<td>26<\/td>\n<td>72.7<\/td>\n<td>72.2<\/td>\n<td>70.9<\/td>\n<td>74.9<\/td>\n<td>100.7<\/td>\n<td>99.6<\/td>\n<td>101.8<\/td>\n<td>98.6<\/td>\n<\/tr>\n<tr>\n<td>54<\/td>\n<td>Coumaphos<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">362.0 &gt; 109.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">16<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">362.0 &gt; 226.0<\/td>\n<td>14<\/td>\n<td>90.1<\/td>\n<td>110.4<\/td>\n<td>98.2<\/td>\n<td>89.7<\/td>\n<td>84.4<\/td>\n<td>84.7<\/td>\n<td>94.5<\/td>\n<td>94.9<\/td>\n<\/tr>\n<tr>\n<td>55<\/td>\n<td>Cycloate<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">154.2 &gt; 72.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">6<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">215.1 &gt; 154.2<\/td>\n<td>4<\/td>\n<td>69.8<\/td>\n<td>66.7<\/td>\n<td>61.2<\/td>\n<td>69.5<\/td>\n<td>89.3<\/td>\n<td>89.9<\/td>\n<td>93.3<\/td>\n<td>94.0<\/td>\n<\/tr>\n<tr>\n<td>56<\/td>\n<td>Cyfluthrin-1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">226.1 &gt; 206.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">14<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">163.1 &gt; 127.1<\/td>\n<td>6<\/td>\n<td>92.8<\/td>\n<td>111.7<\/td>\n<td>98.2<\/td>\n<td>92.1<\/td>\n<td>105.2<\/td>\n<td>118.8<\/td>\n<td>100.9<\/td>\n<td>96.1<\/td>\n<\/tr>\n<tr>\n<td>57<\/td>\n<td>Cyfluthrin-2<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">226.1 &gt; 206.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">14<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">163.1 &gt; 127.1<\/td>\n<td>6<\/td>\n<td>92.5<\/td>\n<td>113.1<\/td>\n<td>91.9<\/td>\n<td>90.2<\/td>\n<td>124.6<\/td>\n<td>121.5<\/td>\n<td>105.1<\/td>\n<td>97.6<\/td>\n<\/tr>\n<tr>\n<td>58<\/td>\n<td>Cyfluthrin-3<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">226.1 &gt; 206.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">14<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">163.1 &gt; 127.1<\/td>\n<td>6<\/td>\n<td>81.4<\/td>\n<td>96.8<\/td>\n<td>92.5<\/td>\n<td>81.7<\/td>\n<td>117.8<\/td>\n<td>160.7<\/td>\n<td>124.5<\/td>\n<td>128.0<\/td>\n<\/tr>\n<tr>\n<td>59<\/td>\n<td>Cyfluthrin-4<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">226.1 &gt; 206.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">14<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">163.1 &gt; 127.1<\/td>\n<td>6<\/td>\n<td>76.5<\/td>\n<td>95.7<\/td>\n<td>106.8<\/td>\n<td>88.7<\/td>\n<td>167.1<\/td>\n<td>159.4<\/td>\n<td>121.8<\/td>\n<td>126.4<\/td>\n<\/tr>\n<tr>\n<td>60<\/td>\n<td>Cyhalothrin,\u00a0lambda-<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">208.1 &gt; 181.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">8<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">197.1 &gt; 141.0<\/td>\n<td>12<\/td>\n<td>88.4<\/td>\n<td>108.0<\/td>\n<td>92.1<\/td>\n<td>85.5<\/td>\n<td>99.2<\/td>\n<td>113.2<\/td>\n<td>99.5<\/td>\n<td>96.9<\/td>\n<\/tr>\n<tr>\n<td>61<\/td>\n<td>Cypermethrin-1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">163.1 &gt; 127.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">6<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">163.1 &gt; 109.1<\/td>\n<td>22<\/td>\n<td>96.0<\/td>\n<td>113.8<\/td>\n<td>94.6<\/td>\n<td>89.9<\/td>\n<td>108.8<\/td>\n<td>93.7<\/td>\n<td>117.4<\/td>\n<td>111.3<\/td>\n<\/tr>\n<tr>\n<td>62<\/td>\n<td>Cypermethrin-2<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">163.1 &gt; 127.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">6<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">163.1 &gt; 109.1<\/td>\n<td>22<\/td>\n<td>89.6<\/td>\n<td>119.7<\/td>\n<td>98.0<\/td>\n<td>88.7<\/td>\n<td>99.2<\/td>\n<td>113.1<\/td>\n<td>94.8<\/td>\n<td>100.1<\/td>\n<\/tr>\n<tr>\n<td>63<\/td>\n<td>Cypermethrin-3<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">163.1 &gt; 127.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">6<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">163.1 &gt; 109.1<\/td>\n<td>22<\/td>\n<td>75.9<\/td>\n<td>124.9<\/td>\n<td>103.1<\/td>\n<td>96.9<\/td>\n<td>126.1<\/td>\n<td>114.1<\/td>\n<td>96.6<\/td>\n<td>104.1<\/td>\n<\/tr>\n<tr>\n<td>64<\/td>\n<td>Cypermethrin-4<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">163.1 &gt; 127.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">6<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">163.1 &gt; 109.1<\/td>\n<td>22<\/td>\n<td>76.6<\/td>\n<td>100.1<\/td>\n<td>84.6<\/td>\n<td>81.5<\/td>\n<td>115.3<\/td>\n<td>119.9<\/td>\n<td>124.9<\/td>\n<td>120.1<\/td>\n<\/tr>\n<tr>\n<td>65<\/td>\n<td>Cyprodinil<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">224.1 &gt; 197.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">22<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">224.1 &gt; 131.1<\/td>\n<td>14<\/td>\n<td>82.0<\/td>\n<td>88.9<\/td>\n<td>80.0<\/td>\n<td>71.2<\/td>\n<td>138.1<\/td>\n<td>117.5<\/td>\n<td>100.5<\/td>\n<td>112.6<\/td>\n<\/tr>\n<tr>\n<td>66<\/td>\n<td>DDD,\u00a0o,p&#8217;-<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">235.0 &gt; 165.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">24<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">235.0 &gt; 199.0<\/td>\n<td>16<\/td>\n<td>83.6<\/td>\n<td>96.0<\/td>\n<td>79.8<\/td>\n<td>75.3<\/td>\n<td>87.0<\/td>\n<td>86.6<\/td>\n<td>100.1<\/td>\n<td>99.1<\/td>\n<\/tr>\n<tr>\n<td>67<\/td>\n<td>DDD,\u00a0p,p&#8217;-<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">235.0 &gt; 165.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">24<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">235.0 &gt; 199.0<\/td>\n<td>16<\/td>\n<td>80.2<\/td>\n<td>94.6<\/td>\n<td>83.6<\/td>\n<td>78.2<\/td>\n<td>104.4<\/td>\n<td>98.8<\/td>\n<td>102.6<\/td>\n<td>104.3<\/td>\n<\/tr>\n<tr>\n<td>68<\/td>\n<td>DDE,\u00a0o,p&#8217;-<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">246.0 &gt; 176.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">30<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">248.0 &gt; 176.0<\/td>\n<td>28<\/td>\n<td>73.7<\/td>\n<td>83.1<\/td>\n<td>75.7<\/td>\n<td>71.7<\/td>\n<td>101.7<\/td>\n<td>97.9<\/td>\n<td>99.0<\/td>\n<td>100.3<\/td>\n<\/tr>\n<tr>\n<td>69<\/td>\n<td>DDE,\u00a0p,p&#8217;-<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">246.0 &gt; 176.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">30<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">317.9 &gt; 248.0<\/td>\n<td>24<\/td>\n<td>76.9<\/td>\n<td>99.7<\/td>\n<td>76.2<\/td>\n<td>73.5<\/td>\n<td>101.1<\/td>\n<td>93.8<\/td>\n<td>98.8<\/td>\n<td>98.3<\/td>\n<\/tr>\n<tr>\n<td>70<\/td>\n<td>DDT,\u00a0o,p&#8217;-<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">235.0 &gt; 165.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">24<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">235.0 &gt; 199.0<\/td>\n<td>16<\/td>\n<td>79.0<\/td>\n<td>89.1<\/td>\n<td>78.0<\/td>\n<td>74.3<\/td>\n<td>93.9<\/td>\n<td>97.6<\/td>\n<td>100.1<\/td>\n<td>101.2<\/td>\n<\/tr>\n<tr>\n<td>71<\/td>\n<td>DDT,\u00a0p,p&#8217;-<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">235.0 &gt; 165.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">24<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">235.0 &gt; 199.0<\/td>\n<td>16<\/td>\n<td>75.2<\/td>\n<td>95.1<\/td>\n<td>80.7<\/td>\n<td>75.8<\/td>\n<td>96.8<\/td>\n<td>95.0<\/td>\n<td>102.9<\/td>\n<td>97.7<\/td>\n<\/tr>\n<tr>\n<td>72<\/td>\n<td>Deltamethrin<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">252.9 &gt; 93.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">20<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">252.9 &gt; 171.9<\/td>\n<td>8<\/td>\n<td>83.8<\/td>\n<td>109.4<\/td>\n<td>92.2<\/td>\n<td>85.3<\/td>\n<td>99.9<\/td>\n<td>99.3<\/td>\n<td>102.5<\/td>\n<td>99.7<\/td>\n<\/tr>\n<tr>\n<td>73<\/td>\n<td>Di-allate-1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">234.1 &gt; 150.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">20<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">234.1 &gt; 192.1<\/td>\n<td>14<\/td>\n<td>75.3<\/td>\n<td>66.5<\/td>\n<td>63.3<\/td>\n<td>70.7<\/td>\n<td>93.8<\/td>\n<td>93.0<\/td>\n<td>100.9<\/td>\n<td>103.3<\/td>\n<\/tr>\n<tr>\n<td>74<\/td>\n<td>Di-allate-2<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">234.1 &gt; 150.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">20<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">234.1 &gt; 192.1<\/td>\n<td>14<\/td>\n<td>72.4<\/td>\n<td>65.1<\/td>\n<td>63.0<\/td>\n<td>70.5<\/td>\n<td>88.9<\/td>\n<td>107.7<\/td>\n<td>100.0<\/td>\n<td>94.1<\/td>\n<\/tr>\n<tr>\n<td>75<\/td>\n<td>Diazinon<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">304.1 &gt; 179.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">10<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">304.1 &gt; 162.1<\/td>\n<td>8<\/td>\n<td>76.4<\/td>\n<td>70.3<\/td>\n<td>69.5<\/td>\n<td>74.4<\/td>\n<td>62.3<\/td>\n<td>80.6<\/td>\n<td>97.9<\/td>\n<td>86.4<\/td>\n<\/tr>\n<tr>\n<td>76<\/td>\n<td>Dichlobenil<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">170.9 &gt; 100.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">24<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">170.9 &gt; 136.0<\/td>\n<td>14<\/td>\n<td>71.2<\/td>\n<td>65.3<\/td>\n<td>58.2<\/td>\n<td>65.4<\/td>\n<td>96.5<\/td>\n<td>97.4<\/td>\n<td>96.5<\/td>\n<td>98.2<\/td>\n<\/tr>\n<tr>\n<td>77<\/td>\n<td>Dichlofluanid<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">223.9 &gt; 123.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">8<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">223.9 &gt; 77.0<\/td>\n<td>28<\/td>\n<td>65.0<\/td>\n<td>71.1<\/td>\n<td>57.8<\/td>\n<td>56.9<\/td>\n<td>75.4<\/td>\n<td>87.4<\/td>\n<td>94.4<\/td>\n<td>101.7<\/td>\n<\/tr>\n<tr>\n<td>78<\/td>\n<td>Dicloran<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">206.0 &gt; 176.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">10<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">206.0 &gt; 124.0<\/td>\n<td>24<\/td>\n<td>78.1<\/td>\n<td>76.6<\/td>\n<td>69.7<\/td>\n<td>77.0<\/td>\n<td>79.0<\/td>\n<td>87.7<\/td>\n<td>100.6<\/td>\n<td>91.0<\/td>\n<\/tr>\n<tr>\n<td>79<\/td>\n<td>Dieldrin<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">276.9 &gt; 241.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">8<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">262.9 &gt; 193.0<\/td>\n<td>34<\/td>\n<td>80.2<\/td>\n<td>80.3<\/td>\n<td>89.0<\/td>\n<td>85.0<\/td>\n<td>154.2<\/td>\n<td>185.9<\/td>\n<td>102.7<\/td>\n<td>91.4<\/td>\n<\/tr>\n<tr>\n<td>80<\/td>\n<td>Dimethachlor<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">197.1 &gt; 148.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">10<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">199.1 &gt; 148.1<\/td>\n<td>10<\/td>\n<td>78.4<\/td>\n<td>90.6<\/td>\n<td>77.9<\/td>\n<td>76.8<\/td>\n<td>99.6<\/td>\n<td>89.1<\/td>\n<td>100.2<\/td>\n<td>101.6<\/td>\n<\/tr>\n<tr>\n<td>81<\/td>\n<td>Diphenamid<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">239.1 &gt; 167.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">8<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">239.1 &gt; 72.0<\/td>\n<td>16<\/td>\n<td>94.7<\/td>\n<td>96.0<\/td>\n<td>88.3<\/td>\n<td>79.8<\/td>\n<td>126.1<\/td>\n<td>123.6<\/td>\n<td>112.2<\/td>\n<td>115.5<\/td>\n<\/tr>\n<tr>\n<td>82<\/td>\n<td>Diphenylamine<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">169.1 &gt; 66.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">24<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">169.1 &gt; 77.0<\/td>\n<td>28<\/td>\n<td>78.2<\/td>\n<td>74.6<\/td>\n<td>65.6<\/td>\n<td>71.1<\/td>\n<td>85.3<\/td>\n<td>99.9<\/td>\n<td>98.3<\/td>\n<td>100.6<\/td>\n<\/tr>\n<tr>\n<td>83<\/td>\n<td>Disulfoton<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">186.0 &gt; 153.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">6<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">186.0 &gt; 97.0<\/td>\n<td>16<\/td>\n<td>71.4<\/td>\n<td>66.2<\/td>\n<td>64.6<\/td>\n<td>78.9<\/td>\n<td>142.4<\/td>\n<td>124.1<\/td>\n<td>107.9<\/td>\n<td>85.8<\/td>\n<\/tr>\n<tr>\n<td>84<\/td>\n<td>Edifenphos<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">173.0 &gt; 109.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">10<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">310.0 &gt; 173.0<\/td>\n<td>14<\/td>\n<td>82.6<\/td>\n<td>103.7<\/td>\n<td>91.0<\/td>\n<td>83.6<\/td>\n<td>97.9<\/td>\n<td>94.6<\/td>\n<td>100.3<\/td>\n<td>98.2<\/td>\n<\/tr>\n<tr>\n<td>85<\/td>\n<td>Endosulfan ether<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">240.9 &gt; 205.9<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">16<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">238.9 &gt; 203.9<\/td>\n<td>16<\/td>\n<td>62.7<\/td>\n<td>64.4<\/td>\n<td>67.2<\/td>\n<td>69.5<\/td>\n<td>126.7<\/td>\n<td>95.3<\/td>\n<td>104.8<\/td>\n<td>101.2<\/td>\n<\/tr>\n<tr>\n<td>86<\/td>\n<td>Endosulfan sulfate<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">271.8 &gt; 236.9<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">18<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">386.8 &gt; 252.9<\/td>\n<td>16<\/td>\n<td>83.8<\/td>\n<td>88.4<\/td>\n<td>87.0<\/td>\n<td>85.1<\/td>\n<td>42.5<\/td>\n<td>55.6<\/td>\n<td>83.3<\/td>\n<td>85.8<\/td>\n<\/tr>\n<tr>\n<td>87<\/td>\n<td>Endosulfan,\u00a0alpha-<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">194.9 &gt; 160.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">8<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">194.9 &gt; 125.0<\/td>\n<td>24<\/td>\n<td>74.0<\/td>\n<td>80.0<\/td>\n<td>75.6<\/td>\n<td>80.7<\/td>\n<td>77.6<\/td>\n<td>76.9<\/td>\n<td>92.3<\/td>\n<td>83.4<\/td>\n<\/tr>\n<tr>\n<td>88<\/td>\n<td>Endosulfan,\u00a0beta-<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">194.9 &gt; 160.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">8<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">194.9 &gt; 125.0<\/td>\n<td>24<\/td>\n<td>83.1<\/td>\n<td>99.4<\/td>\n<td>81.6<\/td>\n<td>81.7<\/td>\n<td>102.1<\/td>\n<td>67.3<\/td>\n<td>93.2<\/td>\n<td>100.2<\/td>\n<\/tr>\n<tr>\n<td>89<\/td>\n<td>Endrin<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">262.9 &gt; 193.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">28<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">262.9 &gt; 228.0<\/td>\n<td>22<\/td>\n<td>84.1<\/td>\n<td>89.9<\/td>\n<td>77.2<\/td>\n<td>75.3<\/td>\n<td>50.2<\/td>\n<td>49.4<\/td>\n<td>85.3<\/td>\n<td>88.2<\/td>\n<\/tr>\n<tr>\n<td>90<\/td>\n<td>Endrin aldehyde<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">249.8 &gt; 214.9<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">26<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">344.9 &gt; 244.9<\/td>\n<td>16<\/td>\n<td>N.D.<\/td>\n<td>N.D.<\/td>\n<td>N.D.<\/td>\n<td>N.D.<\/td>\n<td>N.D.<\/td>\n<td>N.D.<\/td>\n<td>N.D.<\/td>\n<td>N.D.<\/td>\n<\/tr>\n<tr>\n<td>91<\/td>\n<td>Endrin ketone<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">316.9 &gt; 244.9<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">20<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">314.9 &gt; 242.9<\/td>\n<td>18<\/td>\n<td>73.4<\/td>\n<td>91.5<\/td>\n<td>84.1<\/td>\n<td>88.0<\/td>\n<td>108.9<\/td>\n<td>95.4<\/td>\n<td>107.6<\/td>\n<td>101.9<\/td>\n<\/tr>\n<tr>\n<td>92<\/td>\n<td>EPN<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">169.1 &gt; 140.9<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">8<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">169.1 &gt; 77.0<\/td>\n<td>22<\/td>\n<td>90.3<\/td>\n<td>107.4<\/td>\n<td>89.2<\/td>\n<td>82.3<\/td>\n<td>94.3<\/td>\n<td>102.9<\/td>\n<td>100.2<\/td>\n<td>98.8<\/td>\n<\/tr>\n<tr>\n<td>93<\/td>\n<td>Ethalfluralin<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">276.0 &gt; 202.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">18<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">316.1 &gt; 276.0<\/td>\n<td>10<\/td>\n<td>73.3<\/td>\n<td>67.7<\/td>\n<td>63.0<\/td>\n<td>67.1<\/td>\n<td>95.2<\/td>\n<td>96.4<\/td>\n<td>100.9<\/td>\n<td>100.6<\/td>\n<\/tr>\n<tr>\n<td>94<\/td>\n<td>Ethion<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">230.9 &gt; 129.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">24<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">230.9 &gt; 174.9<\/td>\n<td>14<\/td>\n<td>81.4<\/td>\n<td>97.1<\/td>\n<td>84.2<\/td>\n<td>77.9<\/td>\n<td>118.3<\/td>\n<td>108.6<\/td>\n<td>101.8<\/td>\n<td>102.6<\/td>\n<\/tr>\n<tr>\n<td>95<\/td>\n<td>Etofenprox<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">163.1 &gt; 135.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">10<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">163.1 &gt; 107.1<\/td>\n<td>18<\/td>\n<td>87.1<\/td>\n<td>101.1<\/td>\n<td>95.6<\/td>\n<td>87.3<\/td>\n<td>97.4<\/td>\n<td>103.1<\/td>\n<td>98.8<\/td>\n<td>97.8<\/td>\n<\/tr>\n<tr>\n<td>96<\/td>\n<td>Etridiazole<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">210.9 &gt; 182.9<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">10<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">210.9 &gt; 139.9<\/td>\n<td>22<\/td>\n<td>70.0<\/td>\n<td>60.1<\/td>\n<td>58.0<\/td>\n<td>66.9<\/td>\n<td>88.2<\/td>\n<td>102.0<\/td>\n<td>102.0<\/td>\n<td>100.4<\/td>\n<\/tr>\n<tr>\n<td>97<\/td>\n<td>Fenamiphos<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">303.1 &gt; 195.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">8<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">288.1 &gt; 260.1<\/td>\n<td>6<\/td>\n<td>88.7<\/td>\n<td>104.6<\/td>\n<td>99.3<\/td>\n<td>87.1<\/td>\n<td>120.9<\/td>\n<td>106.0<\/td>\n<td>96.8<\/td>\n<td>100.1<\/td>\n<\/tr>\n<tr>\n<td>98<\/td>\n<td>Fenarimol<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">251.0 &gt; 139.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">14<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">330.0 &gt; 139.0<\/td>\n<td>8<\/td>\n<td>83.9<\/td>\n<td>106.7<\/td>\n<td>96.5<\/td>\n<td>89.4<\/td>\n<td>97.4<\/td>\n<td>92.6<\/td>\n<td>99.3<\/td>\n<td>96.0<\/td>\n<\/tr>\n<tr>\n<td>99<\/td>\n<td>Fenchlorphos<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">284.9 &gt; 269.9<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">16<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">286.9 &gt; 271.9<\/td>\n<td>18<\/td>\n<td>80.5<\/td>\n<td>81.4<\/td>\n<td>71.5<\/td>\n<td>74.8<\/td>\n<td>89.8<\/td>\n<td>83.8<\/td>\n<td>99.1<\/td>\n<td>97.2<\/td>\n<\/tr>\n<tr>\n<td>100<\/td>\n<td>Fenitrothion<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">277.0 &gt; 260.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">6<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">260.0 &gt; 125.1<\/td>\n<td>12<\/td>\n<td>78.0<\/td>\n<td>89.3<\/td>\n<td>79.3<\/td>\n<td>75.8<\/td>\n<td>104.9<\/td>\n<td>99.2<\/td>\n<td>111.4<\/td>\n<td>109.6<\/td>\n<\/tr>\n<tr>\n<td>101<\/td>\n<td>Fenpropathrin<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">265.1 &gt; 210.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">12<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">265.1 &gt; 89.0<\/td>\n<td>28<\/td>\n<td>90.3<\/td>\n<td>102.8<\/td>\n<td>96.8<\/td>\n<td>88.2<\/td>\n<td>68.9<\/td>\n<td>75.9<\/td>\n<td>90.9<\/td>\n<td>94.4<\/td>\n<\/tr>\n<tr>\n<td>102<\/td>\n<td>Fenson<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">141.0 &gt; 77.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">16<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">267.9 &gt; 141.0<\/td>\n<td>6<\/td>\n<td>78.3<\/td>\n<td>88.0<\/td>\n<td>77.9<\/td>\n<td>74.5<\/td>\n<td>91.3<\/td>\n<td>92.7<\/td>\n<td>100.8<\/td>\n<td>96.8<\/td>\n<\/tr>\n<tr>\n<td>103<\/td>\n<td>Fenthion<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">278.0 &gt; 169.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">14<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">278.0 &gt; 125.0<\/td>\n<td>20<\/td>\n<td>80.4<\/td>\n<td>81.9<\/td>\n<td>78.2<\/td>\n<td>76.0<\/td>\n<td>93.9<\/td>\n<td>96.0<\/td>\n<td>102.2<\/td>\n<td>104.1<\/td>\n<\/tr>\n<tr>\n<td>104<\/td>\n<td>Fenvalerate-1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">225.1 &gt; 147.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">10<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">419.1 &gt; 225.1<\/td>\n<td>6<\/td>\n<td>82.2<\/td>\n<td>100.7<\/td>\n<td>94.6<\/td>\n<td>88.9<\/td>\n<td>95.5<\/td>\n<td>92.4<\/td>\n<td>103.2<\/td>\n<td>99.3<\/td>\n<\/tr>\n<tr>\n<td>105<\/td>\n<td>Fenvalerate-2<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">225.1 &gt; 147.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">10<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">419.1 &gt; 225.1<\/td>\n<td>6<\/td>\n<td>75.7<\/td>\n<td>106.7<\/td>\n<td>93.1<\/td>\n<td>86.5<\/td>\n<td>73.5<\/td>\n<td>70.0<\/td>\n<td>86.2<\/td>\n<td>89.7<\/td>\n<\/tr>\n<tr>\n<td>106<\/td>\n<td>Fipronil<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">366.9 &gt; 212.9<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">30<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">368.9 &gt; 214.9<\/td>\n<td>30<\/td>\n<td>92.8<\/td>\n<td>113.2<\/td>\n<td>99.5<\/td>\n<td>82.8<\/td>\n<td>93.8<\/td>\n<td>96.2<\/td>\n<td>88.0<\/td>\n<td>96.7<\/td>\n<\/tr>\n<tr>\n<td>107<\/td>\n<td>Fluazifop-P-butyl<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">282.1 &gt; 91.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">18<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">383.1 &gt; 282.1<\/td>\n<td>14<\/td>\n<td>77.4<\/td>\n<td>94.0<\/td>\n<td>89.0<\/td>\n<td>80.6<\/td>\n<td>101.9<\/td>\n<td>96.4<\/td>\n<td>99.6<\/td>\n<td>100.5<\/td>\n<\/tr>\n<tr>\n<td>108<\/td>\n<td>Fluchloralin<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">306.0 &gt; 264.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">8<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">326.0 &gt; 63.0<\/td>\n<td>16<\/td>\n<td>76.4<\/td>\n<td>78.4<\/td>\n<td>76.9<\/td>\n<td>75.1<\/td>\n<td>95.8<\/td>\n<td>96.0<\/td>\n<td>90.4<\/td>\n<td>96.5<\/td>\n<\/tr>\n<tr>\n<td>109<\/td>\n<td>Flucythrinate-1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">157.1 &gt; 107.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">12<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">199.1 &gt; 107.1<\/td>\n<td>22<\/td>\n<td>86.9<\/td>\n<td>111.9<\/td>\n<td>95.6<\/td>\n<td>87.3<\/td>\n<td>119.8<\/td>\n<td>120.4<\/td>\n<td>107.1<\/td>\n<td>105.2<\/td>\n<\/tr>\n<tr>\n<td>110<\/td>\n<td>Flucythrinate-2<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">157.1 &gt; 107.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">12<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">199.1 &gt; 107.1<\/td>\n<td>22<\/td>\n<td>89.4<\/td>\n<td>111.0<\/td>\n<td>96.0<\/td>\n<td>86.6<\/td>\n<td>105.1<\/td>\n<td>108.4<\/td>\n<td>102.6<\/td>\n<td>104.6<\/td>\n<\/tr>\n<tr>\n<td>111<\/td>\n<td>Fludioxonil<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">248.0 &gt; 127.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">26<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">248.0 &gt; 154.0<\/td>\n<td>20<\/td>\n<td>94.1<\/td>\n<td>110.3<\/td>\n<td>96.5<\/td>\n<td>86.1<\/td>\n<td>91.9<\/td>\n<td>101.1<\/td>\n<td>98.1<\/td>\n<td>96.0<\/td>\n<\/tr>\n<tr>\n<td>112<\/td>\n<td>Fluquinconazole<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">340.0 &gt; 298.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">20<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">340.0 &gt; 313.0<\/td>\n<td>14<\/td>\n<td>85.2<\/td>\n<td>107.0<\/td>\n<td>98.9<\/td>\n<td>88.9<\/td>\n<td>97.7<\/td>\n<td>98.5<\/td>\n<td>101.4<\/td>\n<td>101.1<\/td>\n<\/tr>\n<tr>\n<td>113<\/td>\n<td>Fluridone<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">328.1 &gt; 259.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">24<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">328.1 &gt; 127.0<\/td>\n<td>24<\/td>\n<td>91.5<\/td>\n<td>117.7<\/td>\n<td>100.3<\/td>\n<td>91.4<\/td>\n<td>80.1<\/td>\n<td>77.3<\/td>\n<td>93.4<\/td>\n<td>95.7<\/td>\n<\/tr>\n<tr>\n<td>114<\/td>\n<td>Flusilazole<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">233.1 &gt; 165.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">14<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">233.1 &gt; 152.1<\/td>\n<td>14<\/td>\n<td>78.1<\/td>\n<td>93.3<\/td>\n<td>84.9<\/td>\n<td>83.2<\/td>\n<td>105.9<\/td>\n<td>91.5<\/td>\n<td>94.8<\/td>\n<td>95.4<\/td>\n<\/tr>\n<tr>\n<td>115<\/td>\n<td>Flutolanil<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">173.0 &gt; 95.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">26<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">281.1 &gt; 173.0<\/td>\n<td>12<\/td>\n<td>84.1<\/td>\n<td>108.0<\/td>\n<td>93.4<\/td>\n<td>84.7<\/td>\n<td>104.1<\/td>\n<td>101.1<\/td>\n<td>100.0<\/td>\n<td>99.8<\/td>\n<\/tr>\n<tr>\n<td>116<\/td>\n<td>Flutriafol<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">219.1 &gt; 123.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">14<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">219.1 &gt; 95.0<\/td>\n<td>28<\/td>\n<td>82.9<\/td>\n<td>109.6<\/td>\n<td>89.8<\/td>\n<td>84.2<\/td>\n<td>120.1<\/td>\n<td>115.2<\/td>\n<td>103.9<\/td>\n<td>103.9<\/td>\n<\/tr>\n<tr>\n<td>117<\/td>\n<td>Fluvalinate-1,\u00a0tau-<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">250.1 &gt; 55.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">18<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">250.1 &gt; 200.1<\/td>\n<td>16<\/td>\n<td>86.7<\/td>\n<td>112.8<\/td>\n<td>91.9<\/td>\n<td>85.5<\/td>\n<td>79.4<\/td>\n<td>79.9<\/td>\n<td>90.8<\/td>\n<td>90.4<\/td>\n<\/tr>\n<tr>\n<td>118<\/td>\n<td>Fluvalinate-2,\u00a0tau-<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">250.1 &gt; 55.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">18<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">250.1 &gt; 200.1<\/td>\n<td>16<\/td>\n<td>90.4<\/td>\n<td>112.0<\/td>\n<td>89.1<\/td>\n<td>83.9<\/td>\n<td>89.6<\/td>\n<td>95.5<\/td>\n<td>102.5<\/td>\n<td>104.7<\/td>\n<\/tr>\n<tr>\n<td>119<\/td>\n<td>Folpet<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">259.9 &gt; 130.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">14<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">261.9 &gt; 130.0<\/td>\n<td>18<\/td>\n<td>65.2<\/td>\n<td>82.9<\/td>\n<td>72.8<\/td>\n<td>67.0<\/td>\n<td>97.2<\/td>\n<td>89.1<\/td>\n<td>99.1<\/td>\n<td>98.3<\/td>\n<\/tr>\n<tr>\n<td>120<\/td>\n<td>Fonofos<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">246.0 &gt; 137.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">6<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">246.0 &gt; 109.1<\/td>\n<td>18<\/td>\n<td>76.6<\/td>\n<td>74.7<\/td>\n<td>65.8<\/td>\n<td>71.4<\/td>\n<td>106.8<\/td>\n<td>97.6<\/td>\n<td>102.1<\/td>\n<td>101.2<\/td>\n<\/tr>\n<tr>\n<td>121<\/td>\n<td>Heptachlor<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">271.8 &gt; 236.9<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">20<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">273.8 &gt; 238.9<\/td>\n<td>16<\/td>\n<td>68.6<\/td>\n<td>65.4<\/td>\n<td>65.0<\/td>\n<td>68.6<\/td>\n<td>92.3<\/td>\n<td>102.0<\/td>\n<td>104.0<\/td>\n<td>101.2<\/td>\n<\/tr>\n<tr>\n<td>122<\/td>\n<td>Heptachlor-exo-epoxide<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">352.8 &gt; 262.9<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">14<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">352.8 &gt; 316.9<\/td>\n<td>10<\/td>\n<td>81.1<\/td>\n<td>80.4<\/td>\n<td>71.8<\/td>\n<td>79.1<\/td>\n<td>33.7<\/td>\n<td>39.0<\/td>\n<td>88.6<\/td>\n<td>71.5<\/td>\n<\/tr>\n<tr>\n<td>123<\/td>\n<td>Hexachlorobenzene<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">283.8 &gt; 248.8<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">24<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">283.8 &gt; 213.8<\/td>\n<td>28<\/td>\n<td>45.7<\/td>\n<td>28.8<\/td>\n<td>48.1<\/td>\n<td>62.3<\/td>\n<td>101.4<\/td>\n<td>100.7<\/td>\n<td>102.7<\/td>\n<td>99.3<\/td>\n<\/tr>\n<tr>\n<td>124<\/td>\n<td>Hexazinone<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">171.1 &gt; 71.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">16<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">171.1 &gt; 85.0<\/td>\n<td>16<\/td>\n<td>93.2<\/td>\n<td>112.9<\/td>\n<td>96.1<\/td>\n<td>88.2<\/td>\n<td>105.3<\/td>\n<td>101.3<\/td>\n<td>104.0<\/td>\n<td>105.2<\/td>\n<\/tr>\n<tr>\n<td>125<\/td>\n<td>Iodofenphos<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">376.9 &gt; 361.8<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">22<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">376.9 &gt; 331.8<\/td>\n<td>32<\/td>\n<td>80.0<\/td>\n<td>86.8<\/td>\n<td>75.8<\/td>\n<td>74.2<\/td>\n<td>83.8<\/td>\n<td>102.3<\/td>\n<td>97.3<\/td>\n<td>97.1<\/td>\n<\/tr>\n<tr>\n<td>126<\/td>\n<td>Iprodione<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">314.0 &gt; 245.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">12<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">314.0 &gt; 56.0<\/td>\n<td>22<\/td>\n<td>110.5<\/td>\n<td>156.0<\/td>\n<td>102.0<\/td>\n<td>89.9<\/td>\n<td>91.0<\/td>\n<td>99.9<\/td>\n<td>102.2<\/td>\n<td>107.0<\/td>\n<\/tr>\n<tr>\n<td>127<\/td>\n<td>Isazofos<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">257.0 &gt; 162.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">8<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">257.0 &gt; 119.0<\/td>\n<td>18<\/td>\n<td>80.0<\/td>\n<td>87.4<\/td>\n<td>74.4<\/td>\n<td>74.7<\/td>\n<td>109.5<\/td>\n<td>107.1<\/td>\n<td>96.0<\/td>\n<td>102.6<\/td>\n<\/tr>\n<tr>\n<td>128<\/td>\n<td>Isodrin<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">192.9 &gt; 157.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">20<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">262.9 &gt; 192.9<\/td>\n<td>28<\/td>\n<td>76.4<\/td>\n<td>73.7<\/td>\n<td>67.4<\/td>\n<td>71.2<\/td>\n<td>88.2<\/td>\n<td>99.0<\/td>\n<td>110.1<\/td>\n<td>98.6<\/td>\n<\/tr>\n<tr>\n<td>129<\/td>\n<td>Isopropalin<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">280.1 &gt; 238.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">8<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">280.1 &gt; 133.1<\/td>\n<td>18<\/td>\n<td>76.5<\/td>\n<td>88.1<\/td>\n<td>73.6<\/td>\n<td>71.4<\/td>\n<td>89.8<\/td>\n<td>82.7<\/td>\n<td>97.9<\/td>\n<td>106.1<\/td>\n<\/tr>\n<tr>\n<td>130<\/td>\n<td>Lenacil<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">153.1 &gt; 136.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">14<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">153.1 &gt; 82.1<\/td>\n<td>16<\/td>\n<td>85.1<\/td>\n<td>103.2<\/td>\n<td>97.5<\/td>\n<td>85.7<\/td>\n<td>109.0<\/td>\n<td>108.9<\/td>\n<td>124.1<\/td>\n<td>111.8<\/td>\n<\/tr>\n<tr>\n<td>131<\/td>\n<td>Leptophos<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">376.9 &gt; 361.9<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">24<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">374.9 &gt; 359.9<\/td>\n<td>24<\/td>\n<td>86.8<\/td>\n<td>102.6<\/td>\n<td>89.0<\/td>\n<td>82.6<\/td>\n<td>95.4<\/td>\n<td>99.8<\/td>\n<td>101.9<\/td>\n<td>103.0<\/td>\n<\/tr>\n<tr>\n<td>132<\/td>\n<td>Linuron<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">248.0 &gt; 61.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">16<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">250.0 &gt; 61.0<\/td>\n<td>16<\/td>\n<td>68.3<\/td>\n<td>84.7<\/td>\n<td>77.7<\/td>\n<td>77.9<\/td>\n<td>98.3<\/td>\n<td>83.1<\/td>\n<td>105.8<\/td>\n<td>108.6<\/td>\n<\/tr>\n<tr>\n<td>133<\/td>\n<td>Malathion<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">173.1 &gt; 99.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">14<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">158.1 &gt; 125.0<\/td>\n<td>10<\/td>\n<td>79.9<\/td>\n<td>87.3<\/td>\n<td>77.6<\/td>\n<td>75.3<\/td>\n<td>95.2<\/td>\n<td>97.4<\/td>\n<td>101.3<\/td>\n<td>100.9<\/td>\n<\/tr>\n<tr>\n<td>134<\/td>\n<td>Metalaxyl<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">249.2 &gt; 190.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">8<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">249.2 &gt; 146.1<\/td>\n<td>22<\/td>\n<td>83.5<\/td>\n<td>88.7<\/td>\n<td>89.5<\/td>\n<td>81.1<\/td>\n<td>103.0<\/td>\n<td>122.4<\/td>\n<td>104.0<\/td>\n<td>100.7<\/td>\n<\/tr>\n<tr>\n<td>135<\/td>\n<td>Metazachlor<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">209.1 &gt; 132.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">18<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">211.1 &gt; 132.1<\/td>\n<td>20<\/td>\n<td>83.3<\/td>\n<td>93.2<\/td>\n<td>83.8<\/td>\n<td>80.1<\/td>\n<td>97.8<\/td>\n<td>100.1<\/td>\n<td>101.6<\/td>\n<td>104.8<\/td>\n<\/tr>\n<tr>\n<td>136<\/td>\n<td>Methacrifos<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">208.0 &gt; 180.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">8<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">240.0 &gt; 208.0<\/td>\n<td>4<\/td>\n<td>75.6<\/td>\n<td>68.7<\/td>\n<td>62.9<\/td>\n<td>67.2<\/td>\n<td>93.3<\/td>\n<td>89.0<\/td>\n<td>95.3<\/td>\n<td>101.8<\/td>\n<\/tr>\n<tr>\n<td>137<\/td>\n<td>Methoxychlor<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">227.1 &gt; 169.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">24<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">227.1 &gt; 212.1<\/td>\n<td>14<\/td>\n<td>82.8<\/td>\n<td>101.8<\/td>\n<td>90.8<\/td>\n<td>83.5<\/td>\n<td>109.0<\/td>\n<td>108.5<\/td>\n<td>103.5<\/td>\n<td>103.2<\/td>\n<\/tr>\n<tr>\n<td>138<\/td>\n<td>Metolachlor<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">238.1 &gt; 162.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">12<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">238.1 &gt; 133.1<\/td>\n<td>26<\/td>\n<td>77.0<\/td>\n<td>89.0<\/td>\n<td>77.9<\/td>\n<td>75.4<\/td>\n<td>104.8<\/td>\n<td>96.6<\/td>\n<td>101.4<\/td>\n<td>102.3<\/td>\n<\/tr>\n<tr>\n<td>139<\/td>\n<td>Mevinphos-1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">192.0 &gt; 127.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">12<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">127.0 &gt; 95.0<\/td>\n<td>18<\/td>\n<td>75.7<\/td>\n<td>70.7<\/td>\n<td>64.1<\/td>\n<td>71.4<\/td>\n<td>95.8<\/td>\n<td>104.8<\/td>\n<td>103.6<\/td>\n<td>102.2<\/td>\n<\/tr>\n<tr>\n<td>140<\/td>\n<td>MGK 264-1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">164.1 &gt; 93.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">10<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">164.1 &gt; 80.0<\/td>\n<td>24<\/td>\n<td>112.8<\/td>\n<td>102.9<\/td>\n<td>86.2<\/td>\n<td>81.3<\/td>\n<td>92.5<\/td>\n<td>104.4<\/td>\n<td>108.1<\/td>\n<td>104.7<\/td>\n<\/tr>\n<tr>\n<td>141<\/td>\n<td>MGK 264-2<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">164.1 &gt; 98.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">12<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">164.1 &gt; 67.0<\/td>\n<td>8<\/td>\n<td>74.8<\/td>\n<td>89.7<\/td>\n<td>79.3<\/td>\n<td>78.2<\/td>\n<td>116.0<\/td>\n<td>100.8<\/td>\n<td>93.7<\/td>\n<td>96.0<\/td>\n<\/tr>\n<tr>\n<td>142<\/td>\n<td>Mirex<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">271.8 &gt; 236.8<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">18<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">273.8 &gt; 238.8<\/td>\n<td>18<\/td>\n<td>71.4<\/td>\n<td>80.4<\/td>\n<td>73.6<\/td>\n<td>71.2<\/td>\n<td>99.9<\/td>\n<td>97.3<\/td>\n<td>100.7<\/td>\n<td>99.6<\/td>\n<\/tr>\n<tr>\n<td>143<\/td>\n<td>Myclobutanil<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">179.1 &gt; 125.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">14<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">179.1 &gt; 152.0<\/td>\n<td>8<\/td>\n<td>83.7<\/td>\n<td>104.4<\/td>\n<td>90.8<\/td>\n<td>84.1<\/td>\n<td>122.4<\/td>\n<td>127.5<\/td>\n<td>112.5<\/td>\n<td>108.8<\/td>\n<\/tr>\n<tr>\n<td>144<\/td>\n<td>N-(2,4-dimethylphenyl) formamide<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">149.1 &gt; 106.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">16<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">149.1 &gt; 121.1<\/td>\n<td>6<\/td>\n<td>86.6<\/td>\n<td>88.5<\/td>\n<td>73.0<\/td>\n<td>71.4<\/td>\n<td>334.2<\/td>\n<td>367.6<\/td>\n<td>157.5<\/td>\n<td>169.6<\/td>\n<\/tr>\n<tr>\n<td>145<\/td>\n<td>Nitralin<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">316.1 &gt; 274.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">8<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">274.0 &gt; 169.0<\/td>\n<td>12<\/td>\n<td>101.1<\/td>\n<td>116.0<\/td>\n<td>95.4<\/td>\n<td>87.2<\/td>\n<td>98.7<\/td>\n<td>104.0<\/td>\n<td>106.2<\/td>\n<td>97.4<\/td>\n<\/tr>\n<tr>\n<td>146<\/td>\n<td>Nitrofen<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">202.0 &gt; 139.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">24<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">282.9 &gt; 253.0<\/td>\n<td>12<\/td>\n<td>83.4<\/td>\n<td>91.9<\/td>\n<td>85.9<\/td>\n<td>77.0<\/td>\n<td>89.0<\/td>\n<td>95.2<\/td>\n<td>99.6<\/td>\n<td>104.3<\/td>\n<\/tr>\n<tr>\n<td>147<\/td>\n<td>Nonachlor,\u00a0cis-<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">406.8 &gt; 299.9<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">24<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">406.8 &gt; 334.9<\/td>\n<td>16<\/td>\n<td>80.0<\/td>\n<td>91.3<\/td>\n<td>79.8<\/td>\n<td>77.0<\/td>\n<td>59.7<\/td>\n<td>48.5<\/td>\n<td>84.3<\/td>\n<td>87.0<\/td>\n<\/tr>\n<tr>\n<td>148<\/td>\n<td>Nonachlor,\u00a0trans-<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">406.8 &gt; 299.9<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">24<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">406.8 &gt; 334.9<\/td>\n<td>16<\/td>\n<td>74.8<\/td>\n<td>86.0<\/td>\n<td>80.8<\/td>\n<td>78.2<\/td>\n<td>58.4<\/td>\n<td>72.4<\/td>\n<td>82.3<\/td>\n<td>84.6<\/td>\n<\/tr>\n<tr>\n<td>149<\/td>\n<td>Norflurazon<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">303.0 &gt; 145.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">22<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">145.0 &gt; 95.0<\/td>\n<td>18<\/td>\n<td>94.7<\/td>\n<td>111.0<\/td>\n<td>98.7<\/td>\n<td>85.0<\/td>\n<td>92.7<\/td>\n<td>95.4<\/td>\n<td>99.2<\/td>\n<td>100.8<\/td>\n<\/tr>\n<tr>\n<td>150<\/td>\n<td>Oxadiazon<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">258.0 &gt; 175.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">8<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">302.0 &gt; 175.0<\/td>\n<td>14<\/td>\n<td>78.8<\/td>\n<td>88.7<\/td>\n<td>84.7<\/td>\n<td>76.9<\/td>\n<td>95.1<\/td>\n<td>101.4<\/td>\n<td>100.4<\/td>\n<td>103.2<\/td>\n<\/tr>\n<tr>\n<td>151<\/td>\n<td>Oxyfluorfen<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">361.0 &gt; 300.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">14<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">361.0 &gt; 317.0<\/td>\n<td>6<\/td>\n<td>101.4<\/td>\n<td>106.2<\/td>\n<td>90.9<\/td>\n<td>80.4<\/td>\n<td>91.2<\/td>\n<td>97.7<\/td>\n<td>101.4<\/td>\n<td>102.1<\/td>\n<\/tr>\n<tr>\n<td>152<\/td>\n<td>Paclobutrazol<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">236.1 &gt; 125.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">14<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">236.1 &gt; 167.0<\/td>\n<td>10<\/td>\n<td>92.2<\/td>\n<td>116.7<\/td>\n<td>92.8<\/td>\n<td>86.1<\/td>\n<td>98.0<\/td>\n<td>86.2<\/td>\n<td>100.9<\/td>\n<td>100.1<\/td>\n<\/tr>\n<tr>\n<td>153<\/td>\n<td>Parathion<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">291.1 &gt; 137.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">6<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">291.1 &gt; 81.0<\/td>\n<td>24<\/td>\n<td>99.8<\/td>\n<td>96.8<\/td>\n<td>80.7<\/td>\n<td>76.0<\/td>\n<td>100.1<\/td>\n<td>115.2<\/td>\n<td>104.9<\/td>\n<td>108.1<\/td>\n<\/tr>\n<tr>\n<td>154<\/td>\n<td>Parathion-methyl<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">263.0 &gt; 109.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">14<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">263.0 &gt; 246.0<\/td>\n<td>6<\/td>\n<td>85.5<\/td>\n<td>91.6<\/td>\n<td>74.2<\/td>\n<td>77.4<\/td>\n<td>71.7<\/td>\n<td>69.8<\/td>\n<td>96.2<\/td>\n<td>93.7<\/td>\n<\/tr>\n<tr>\n<td>155<\/td>\n<td>Pebulate<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">161.1 &gt; 128.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">6<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">128.1 &gt; 57.0<\/td>\n<td>6<\/td>\n<td>64.1<\/td>\n<td>59.1<\/td>\n<td>56.9<\/td>\n<td>66.7<\/td>\n<td>115.3<\/td>\n<td>111.1<\/td>\n<td>101.3<\/td>\n<td>101.7<\/td>\n<\/tr>\n<tr>\n<td>156<\/td>\n<td>Penconazole<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">248.1 &gt; 157.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">26<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">159.1 &gt; 123.1<\/td>\n<td>22<\/td>\n<td>90.9<\/td>\n<td>92.6<\/td>\n<td>75.3<\/td>\n<td>75.4<\/td>\n<td>90.0<\/td>\n<td>94.4<\/td>\n<td>97.6<\/td>\n<td>97.5<\/td>\n<\/tr>\n<tr>\n<td>157<\/td>\n<td>Pendimethalin<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">252.1 &gt; 162.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">10<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">252.1 &gt; 191.1<\/td>\n<td>8<\/td>\n<td>84.0<\/td>\n<td>84.2<\/td>\n<td>73.9<\/td>\n<td>73.8<\/td>\n<td>89.5<\/td>\n<td>98.9<\/td>\n<td>102.2<\/td>\n<td>104.0<\/td>\n<\/tr>\n<tr>\n<td>158<\/td>\n<td>Pentachloroaniline<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">262.9 &gt; 191.9<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">22<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">264.9 &gt; 193.9<\/td>\n<td>18<\/td>\n<td>23.6<\/td>\n<td>83.6<\/td>\n<td>52.0<\/td>\n<td>72.4<\/td>\n<td>99.4<\/td>\n<td>99.0<\/td>\n<td>101.1<\/td>\n<td>98.8<\/td>\n<\/tr>\n<tr>\n<td>159<\/td>\n<td>Pentachloroanisole<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">279.9 &gt; 236.8<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">26<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">279.9 &gt; 264.8<\/td>\n<td>12<\/td>\n<td>69.0<\/td>\n<td>65.8<\/td>\n<td>60.2<\/td>\n<td>67.9<\/td>\n<td>102.8<\/td>\n<td>93.3<\/td>\n<td>101.7<\/td>\n<td>101.4<\/td>\n<\/tr>\n<tr>\n<td>160<\/td>\n<td>Pentachlorobenzene<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">249.9 &gt; 214.9<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">18<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">249.9 &gt; 176.9<\/td>\n<td>26<\/td>\n<td>59.9<\/td>\n<td>27.0<\/td>\n<td>47.6<\/td>\n<td>62.0<\/td>\n<td>104.3<\/td>\n<td>113.2<\/td>\n<td>108.4<\/td>\n<td>104.3<\/td>\n<\/tr>\n<tr>\n<td>161<\/td>\n<td>Pentachlorobenzonitrile<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">274.8 &gt; 239.8<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">18<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">272.8 &gt; 202.9<\/td>\n<td>30<\/td>\n<td>69.3<\/td>\n<td>65.6<\/td>\n<td>63.2<\/td>\n<td>68.7<\/td>\n<td>90.1<\/td>\n<td>88.2<\/td>\n<td>96.3<\/td>\n<td>95.7<\/td>\n<\/tr>\n<tr>\n<td>162<\/td>\n<td>Pentachlorothioanisole<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">295.8 &gt; 262.9<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">14<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">295.8 &gt; 245.8<\/td>\n<td>30<\/td>\n<td>55.8<\/td>\n<td>75.7<\/td>\n<td>62.7<\/td>\n<td>69.7<\/td>\n<td>92.4<\/td>\n<td>96.3<\/td>\n<td>94.9<\/td>\n<td>93.5<\/td>\n<\/tr>\n<tr>\n<td>163<\/td>\n<td>Permethrine,\u00a0cis-<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">183.1 &gt; 153.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">14<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">183.1 &gt; 168.1<\/td>\n<td>14<\/td>\n<td>86.6<\/td>\n<td>112.1<\/td>\n<td>96.8<\/td>\n<td>87.5<\/td>\n<td>101.7<\/td>\n<td>108.3<\/td>\n<td>99.5<\/td>\n<td>100.2<\/td>\n<\/tr>\n<tr>\n<td>164<\/td>\n<td>Permethrine,\u00a0trans-<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">183.1 &gt; 153.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">14<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">183.1 &gt; 168.1<\/td>\n<td>14<\/td>\n<td>96.0<\/td>\n<td>131.5<\/td>\n<td>97.0<\/td>\n<td>88.4<\/td>\n<td>100.4<\/td>\n<td>98.3<\/td>\n<td>103.6<\/td>\n<td>102.4<\/td>\n<\/tr>\n<tr>\n<td>165<\/td>\n<td>Perthane<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">223.2 &gt; 167.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">14<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">223.2 &gt; 193.1<\/td>\n<td>28<\/td>\n<td>85.3<\/td>\n<td>92.7<\/td>\n<td>81.9<\/td>\n<td>78.1<\/td>\n<td>97.1<\/td>\n<td>106.8<\/td>\n<td>102.3<\/td>\n<td>100.3<\/td>\n<\/tr>\n<tr>\n<td>166<\/td>\n<td>Phenothrin-1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">183.1 &gt; 153.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">14<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">183.1 &gt; 168.1<\/td>\n<td>14<\/td>\n<td>N.D.<\/td>\n<td>N.D.<\/td>\n<td>N.D.<\/td>\n<td>84.8<\/td>\n<td>N.D.<\/td>\n<td>N.D.<\/td>\n<td>111.5<\/td>\n<td>92.4<\/td>\n<\/tr>\n<tr>\n<td>167<\/td>\n<td>Phenothrin-2<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">183.1 &gt; 153.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">14<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">183.1 &gt; 168.1<\/td>\n<td>14<\/td>\n<td>100.7<\/td>\n<td>113.7<\/td>\n<td>95.7<\/td>\n<td>86.7<\/td>\n<td>101.4<\/td>\n<td>112.1<\/td>\n<td>105.1<\/td>\n<td>100.1<\/td>\n<\/tr>\n<tr>\n<td>168<\/td>\n<td>Phorate<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">260.0 &gt; 75.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">8<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">231.0 &gt; 129.0<\/td>\n<td>24<\/td>\n<td>74.3<\/td>\n<td>63.8<\/td>\n<td>61.9<\/td>\n<td>69.7<\/td>\n<td>92.7<\/td>\n<td>116.3<\/td>\n<td>99.8<\/td>\n<td>101.6<\/td>\n<\/tr>\n<tr>\n<td>169<\/td>\n<td>Phosalone<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">182.0 &gt; 102.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">14<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">182.0 &gt; 111.0<\/td>\n<td>14<\/td>\n<td>86.3<\/td>\n<td>101.6<\/td>\n<td>93.9<\/td>\n<td>84.4<\/td>\n<td>117.0<\/td>\n<td>123.8<\/td>\n<td>101.2<\/td>\n<td>104.5<\/td>\n<\/tr>\n<tr>\n<td>170<\/td>\n<td>Phosmet<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">160.0 &gt; 77.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">24<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">160.0 &gt; 105.0<\/td>\n<td>18<\/td>\n<td>86.7<\/td>\n<td>105.4<\/td>\n<td>92.6<\/td>\n<td>83.0<\/td>\n<td>100.8<\/td>\n<td>100.1<\/td>\n<td>103.5<\/td>\n<td>101.2<\/td>\n<\/tr>\n<tr>\n<td>171<\/td>\n<td>Piperonyl butoxide<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">176.1 &gt; 131.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">12<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">176.1 &gt; 117.1<\/td>\n<td>20<\/td>\n<td>84.1<\/td>\n<td>112.8<\/td>\n<td>92.1<\/td>\n<td>85.9<\/td>\n<td>104.9<\/td>\n<td>118.6<\/td>\n<td>102.2<\/td>\n<td>102.3<\/td>\n<\/tr>\n<tr>\n<td>172<\/td>\n<td>Pirimiphos ethyl<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">304.1 &gt; 168.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">12<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">318.1 &gt; 166.1<\/td>\n<td>12<\/td>\n<td>78.1<\/td>\n<td>94.3<\/td>\n<td>79.1<\/td>\n<td>73.8<\/td>\n<td>83.5<\/td>\n<td>84.4<\/td>\n<td>97.4<\/td>\n<td>105.8<\/td>\n<\/tr>\n<tr>\n<td>173<\/td>\n<td>Pirimiphos-methyl<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">290.1 &gt; 125.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">22<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">290.1 &gt; 233.1<\/td>\n<td>12<\/td>\n<td>80.4<\/td>\n<td>88.9<\/td>\n<td>78.5<\/td>\n<td>76.1<\/td>\n<td>93.3<\/td>\n<td>92.5<\/td>\n<td>101.2<\/td>\n<td>99.4<\/td>\n<\/tr>\n<tr>\n<td>174<\/td>\n<td>Pretilachlor<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">262.1 &gt; 202.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">10<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">238.1 &gt; 162.1<\/td>\n<td>10<\/td>\n<td>77.0<\/td>\n<td>95.2<\/td>\n<td>86.0<\/td>\n<td>79.7<\/td>\n<td>109.6<\/td>\n<td>79.5<\/td>\n<td>93.4<\/td>\n<td>102.5<\/td>\n<\/tr>\n<tr>\n<td>175<\/td>\n<td>Prochloraz<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">180.1 &gt; 138.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">12<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">180.1 &gt; 69.0<\/td>\n<td>20<\/td>\n<td>68.1<\/td>\n<td>110.1<\/td>\n<td>85.4<\/td>\n<td>81.5<\/td>\n<td>106.2<\/td>\n<td>102.1<\/td>\n<td>91.1<\/td>\n<td>91.2<\/td>\n<\/tr>\n<tr>\n<td>176<\/td>\n<td>Procymidone<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">283.0 &gt; 96.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">10<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">285.0 &gt; 96.0<\/td>\n<td>10<\/td>\n<td>12.4<\/td>\n<td>140.1<\/td>\n<td>78.8<\/td>\n<td>79.4<\/td>\n<td>103.0<\/td>\n<td>101.7<\/td>\n<td>104.7<\/td>\n<td>101.8<\/td>\n<\/tr>\n<tr>\n<td>177<\/td>\n<td>Prodiamine<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">321.1 &gt; 279.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">6<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">321.1 &gt; 203.1<\/td>\n<td>10<\/td>\n<td>86.1<\/td>\n<td>94.7<\/td>\n<td>81.6<\/td>\n<td>78.3<\/td>\n<td>88.5<\/td>\n<td>88.3<\/td>\n<td>93.6<\/td>\n<td>99.0<\/td>\n<\/tr>\n<tr>\n<td>178<\/td>\n<td>Profenofos<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">338.9 &gt; 268.9<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">18<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">336.9 &gt; 266.9<\/td>\n<td>14<\/td>\n<td>87.1<\/td>\n<td>93.5<\/td>\n<td>90.0<\/td>\n<td>85.6<\/td>\n<td>107.6<\/td>\n<td>107.6<\/td>\n<td>93.3<\/td>\n<td>91.1<\/td>\n<\/tr>\n<tr>\n<td>179<\/td>\n<td>Profluralin<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">318.1 &gt; 199.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">16<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">318.1 &gt; 55.0<\/td>\n<td>22<\/td>\n<td>66.3<\/td>\n<td>64.9<\/td>\n<td>65.7<\/td>\n<td>74.8<\/td>\n<td>105.2<\/td>\n<td>95.9<\/td>\n<td>97.6<\/td>\n<td>91.7<\/td>\n<\/tr>\n<tr>\n<td>180<\/td>\n<td>Propachlor<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">176.1 &gt; 57.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">8<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">176.1 &gt; 77.0<\/td>\n<td>24<\/td>\n<td>74.9<\/td>\n<td>76.4<\/td>\n<td>67.3<\/td>\n<td>71.4<\/td>\n<td>114.9<\/td>\n<td>106.5<\/td>\n<td>103.8<\/td>\n<td>102.7<\/td>\n<\/tr>\n<tr>\n<td>181<\/td>\n<td>Propanil<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">217.0 &gt; 161.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">10<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">160.9 &gt; 126.0<\/td>\n<td>18<\/td>\n<td>93.5<\/td>\n<td>105.3<\/td>\n<td>91.4<\/td>\n<td>78.7<\/td>\n<td>100.7<\/td>\n<td>106.9<\/td>\n<td>95.8<\/td>\n<td>104.6<\/td>\n<\/tr>\n<tr>\n<td>182<\/td>\n<td>Propargite<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">173.1 &gt; 135.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">16<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">173.1 &gt; 107.1<\/td>\n<td>24<\/td>\n<td>88.0<\/td>\n<td>98.7<\/td>\n<td>90.8<\/td>\n<td>85.5<\/td>\n<td>120.0<\/td>\n<td>122.8<\/td>\n<td>74.1<\/td>\n<td>66.1<\/td>\n<\/tr>\n<tr>\n<td>183<\/td>\n<td>Propisochlor<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">223.1 &gt; 132.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">20<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">223.1 &gt; 147.1<\/td>\n<td>8<\/td>\n<td>83.7<\/td>\n<td>90.8<\/td>\n<td>79.4<\/td>\n<td>77.9<\/td>\n<td>91.0<\/td>\n<td>100.1<\/td>\n<td>98.4<\/td>\n<td>101.0<\/td>\n<\/tr>\n<tr>\n<td>184<\/td>\n<td>Propyzamide<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">172.9 &gt; 109.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">26<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">172.9 &gt; 74.0<\/td>\n<td>28<\/td>\n<td>83.0<\/td>\n<td>91.4<\/td>\n<td>77.1<\/td>\n<td>76.5<\/td>\n<td>98.7<\/td>\n<td>100.0<\/td>\n<td>108.7<\/td>\n<td>106.3<\/td>\n<\/tr>\n<tr>\n<td>185<\/td>\n<td>Prothiofos<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">266.9 &gt; 238.9<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">10<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">309.0 &gt; 238.9<\/td>\n<td>14<\/td>\n<td>74.3<\/td>\n<td>88.9<\/td>\n<td>78.0<\/td>\n<td>73.9<\/td>\n<td>102.4<\/td>\n<td>94.4<\/td>\n<td>102.7<\/td>\n<td>101.2<\/td>\n<\/tr>\n<tr>\n<td>186<\/td>\n<td>Pyraclofos<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">194.0 &gt; 138.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">22<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">360.1 &gt; 194.0<\/td>\n<td>14<\/td>\n<td>91.6<\/td>\n<td>111.4<\/td>\n<td>97.0<\/td>\n<td>87.0<\/td>\n<td>84.1<\/td>\n<td>89.9<\/td>\n<td>99.3<\/td>\n<td>100.1<\/td>\n<\/tr>\n<tr>\n<td>187<\/td>\n<td>Pyrazophos<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">221.1 &gt; 193.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">12<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">221.1 &gt; 149.1<\/td>\n<td>14<\/td>\n<td>88.7<\/td>\n<td>111.9<\/td>\n<td>97.3<\/td>\n<td>86.1<\/td>\n<td>101.7<\/td>\n<td>96.0<\/td>\n<td>99.9<\/td>\n<td>103.7<\/td>\n<\/tr>\n<tr>\n<td>188<\/td>\n<td>Pyridaben<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">147.1 &gt; 117.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">22<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">147.1 &gt; 132.1<\/td>\n<td>14<\/td>\n<td>87.5<\/td>\n<td>106.4<\/td>\n<td>92.5<\/td>\n<td>84.1<\/td>\n<td>99.2<\/td>\n<td>105.1<\/td>\n<td>102.0<\/td>\n<td>102.5<\/td>\n<\/tr>\n<tr>\n<td>189<\/td>\n<td>Pyridaphenthion<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">340.0 &gt; 199.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">8<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">199.1 &gt; 92.0<\/td>\n<td>16<\/td>\n<td>100.2<\/td>\n<td>120.0<\/td>\n<td>95.0<\/td>\n<td>89.3<\/td>\n<td>114.1<\/td>\n<td>129.2<\/td>\n<td>109.5<\/td>\n<td>107.5<\/td>\n<\/tr>\n<tr>\n<td>190<\/td>\n<td>Pyrimethanil<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">198.1 &gt; 118.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">28<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">198.1 &gt; 158.1<\/td>\n<td>18<\/td>\n<td>74.4<\/td>\n<td>82.5<\/td>\n<td>75.3<\/td>\n<td>73.2<\/td>\n<td>95.2<\/td>\n<td>99.1<\/td>\n<td>96.4<\/td>\n<td>98.5<\/td>\n<\/tr>\n<tr>\n<td>191<\/td>\n<td>Pyriproxyfen<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">136.1 &gt; 96.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">14<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">226.1 &gt; 186.1<\/td>\n<td>14<\/td>\n<td>81.3<\/td>\n<td>91.9<\/td>\n<td>92.8<\/td>\n<td>84.4<\/td>\n<td>59.7<\/td>\n<td>78.1<\/td>\n<td>88.1<\/td>\n<td>84.8<\/td>\n<\/tr>\n<tr>\n<td>192<\/td>\n<td>Quinalphos<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">146.1 &gt; 118.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">10<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">146.1 &gt; 91.0<\/td>\n<td>24<\/td>\n<td>71.7<\/td>\n<td>86.6<\/td>\n<td>79.6<\/td>\n<td>72.9<\/td>\n<td>204.6<\/td>\n<td>155.7<\/td>\n<td>122.4<\/td>\n<td>117.8<\/td>\n<\/tr>\n<tr>\n<td>193<\/td>\n<td>Quintozene<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">294.8 &gt; 236.8<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">16<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">264.8 &gt; 236.8<\/td>\n<td>10<\/td>\n<td>61.7<\/td>\n<td>0.0<\/td>\n<td>18.5<\/td>\n<td>71.1<\/td>\n<td>100.1<\/td>\n<td>100.7<\/td>\n<td>101.5<\/td>\n<td>100.3<\/td>\n<\/tr>\n<tr>\n<td>194<\/td>\n<td>Resmethrin-1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">171.1 &gt; 128.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">12<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">171.1 &gt; 143.1<\/td>\n<td>6<\/td>\n<td>N.D.<\/td>\n<td>N.D.<\/td>\n<td>N.D.<\/td>\n<td>N.D.<\/td>\n<td>N.D.<\/td>\n<td>N.D.<\/td>\n<td>N.D.<\/td>\n<td>N.D.<\/td>\n<\/tr>\n<tr>\n<td>195<\/td>\n<td>Resmethrin-2<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">171.1 &gt; 143.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">6<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">171.1 &gt; 128.1<\/td>\n<td>14<\/td>\n<td>85.8<\/td>\n<td>96.4<\/td>\n<td>85.8<\/td>\n<td>78.1<\/td>\n<td>86.0<\/td>\n<td>97.9<\/td>\n<td>95.6<\/td>\n<td>100.0<\/td>\n<\/tr>\n<tr>\n<td>196<\/td>\n<td>Sulfotep<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">322.0 &gt; 294.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">4<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">322.0 &gt; 202.0<\/td>\n<td>10<\/td>\n<td>79.1<\/td>\n<td>71.8<\/td>\n<td>65.5<\/td>\n<td>71.8<\/td>\n<td>94.6<\/td>\n<td>98.9<\/td>\n<td>98.0<\/td>\n<td>95.0<\/td>\n<\/tr>\n<tr>\n<td>197<\/td>\n<td>Sulprofos<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">322.0 &gt; 156.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">8<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">156.0 &gt; 108.0<\/td>\n<td>28<\/td>\n<td>80.7<\/td>\n<td>97.3<\/td>\n<td>86.0<\/td>\n<td>83.2<\/td>\n<td>101.1<\/td>\n<td>95.9<\/td>\n<td>97.3<\/td>\n<td>95.1<\/td>\n<\/tr>\n<tr>\n<td>198<\/td>\n<td>Tebuconazole<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">250.1 &gt; 125.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">22<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">250.1 &gt; 153.1<\/td>\n<td>12<\/td>\n<td>76.8<\/td>\n<td>82.9<\/td>\n<td>68.1<\/td>\n<td>71.6<\/td>\n<td>112.5<\/td>\n<td>96.6<\/td>\n<td>98.9<\/td>\n<td>98.1<\/td>\n<\/tr>\n<tr>\n<td>199<\/td>\n<td>Tebufenpyrad<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">333.1 &gt; 171.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">20<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">333.1 &gt; 276.1<\/td>\n<td>8<\/td>\n<td>87.7<\/td>\n<td>105.2<\/td>\n<td>96.7<\/td>\n<td>84.7<\/td>\n<td>97.6<\/td>\n<td>97.3<\/td>\n<td>94.6<\/td>\n<td>95.6<\/td>\n<\/tr>\n<tr>\n<td>200<\/td>\n<td>Tecnazene<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">260.9 &gt; 202.9<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">14<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">202.9 &gt; 85.0<\/td>\n<td>24<\/td>\n<td>78.8<\/td>\n<td>59.9<\/td>\n<td>59.7<\/td>\n<td>72.4<\/td>\n<td>93.9<\/td>\n<td>99.1<\/td>\n<td>96.8<\/td>\n<td>95.8<\/td>\n<\/tr>\n<tr>\n<td>201<\/td>\n<td>Tefluthrin<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">177.0 &gt; 127.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">16<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">177.0 &gt; 137.1<\/td>\n<td>16<\/td>\n<td>79.0<\/td>\n<td>81.2<\/td>\n<td>67.5<\/td>\n<td>72.5<\/td>\n<td>92.6<\/td>\n<td>94.9<\/td>\n<td>101.0<\/td>\n<td>98.3<\/td>\n<\/tr>\n<tr>\n<td>202<\/td>\n<td>Terbacil<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">161.0 &gt; 88.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">20<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">117.0 &gt; 76.0<\/td>\n<td>8<\/td>\n<td>97.5<\/td>\n<td>111.6<\/td>\n<td>86.5<\/td>\n<td>82.0<\/td>\n<td>81.8<\/td>\n<td>77.9<\/td>\n<td>92.5<\/td>\n<td>96.1<\/td>\n<\/tr>\n<tr>\n<td>203<\/td>\n<td>Terbufos<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">231.0 &gt; 128.9<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">26<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">231.0 &gt; 174.9<\/td>\n<td>14<\/td>\n<td>78.8<\/td>\n<td>80.6<\/td>\n<td>65.2<\/td>\n<td>70.3<\/td>\n<td>111.6<\/td>\n<td>100.1<\/td>\n<td>102.8<\/td>\n<td>104.7<\/td>\n<\/tr>\n<tr>\n<td>204<\/td>\n<td>Terbuthylazine<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">229.1 &gt; 173.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">6<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">214.1 &gt; 71.0<\/td>\n<td>16<\/td>\n<td>94.9<\/td>\n<td>91.0<\/td>\n<td>77.6<\/td>\n<td>77.1<\/td>\n<td>82.6<\/td>\n<td>91.6<\/td>\n<td>99.7<\/td>\n<td>103.3<\/td>\n<\/tr>\n<tr>\n<td>205<\/td>\n<td>Tetrachlorvinphos<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">328.9 &gt; 109.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">20<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">330.9 &gt; 109.0<\/td>\n<td>22<\/td>\n<td>87.4<\/td>\n<td>102.5<\/td>\n<td>87.3<\/td>\n<td>81.0<\/td>\n<td>96.0<\/td>\n<td>93.6<\/td>\n<td>99.9<\/td>\n<td>96.6<\/td>\n<\/tr>\n<tr>\n<td>206<\/td>\n<td>Tetradifon<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">355.9 &gt; 159.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">18<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">355.9 &gt; 228.9<\/td>\n<td>12<\/td>\n<td>84.1<\/td>\n<td>96.6<\/td>\n<td>93.6<\/td>\n<td>88.5<\/td>\n<td>87.9<\/td>\n<td>103.1<\/td>\n<td>102.9<\/td>\n<td>96.7<\/td>\n<\/tr>\n<tr>\n<td>207<\/td>\n<td>Tetramethrin-1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">164.1 &gt; 107.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">14<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">164.1 &gt; 77.0<\/td>\n<td>22<\/td>\n<td>N.D.<\/td>\n<td>N.D.<\/td>\n<td>100.5<\/td>\n<td>94.8<\/td>\n<td>N.D.<\/td>\n<td>N.D.<\/td>\n<td>106.5<\/td>\n<td>114.1<\/td>\n<\/tr>\n<tr>\n<td>208<\/td>\n<td>Tetramethrin-2<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">164.1 &gt; 107.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">14<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">164.1 &gt; 77.0<\/td>\n<td>22<\/td>\n<td>103.2<\/td>\n<td>129.6<\/td>\n<td>98.4<\/td>\n<td>88.4<\/td>\n<td>115.2<\/td>\n<td>117.0<\/td>\n<td>107.9<\/td>\n<td>113.8<\/td>\n<\/tr>\n<tr>\n<td>209<\/td>\n<td>THPI<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">151.1 &gt; 79.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">18<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">151.1 &gt; 77.0<\/td>\n<td>28<\/td>\n<td>79.4<\/td>\n<td>85.0<\/td>\n<td>81.5<\/td>\n<td>78.2<\/td>\n<td>104.7<\/td>\n<td>104.1<\/td>\n<td>103.6<\/td>\n<td>108.0<\/td>\n<\/tr>\n<tr>\n<td>210<\/td>\n<td>Tolclofos-methyl<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">264.9 &gt; 93.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">24<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">264.9 &gt; 219.9<\/td>\n<td>22<\/td>\n<td>72.8<\/td>\n<td>78.6<\/td>\n<td>72.6<\/td>\n<td>73.8<\/td>\n<td>98.8<\/td>\n<td>100.9<\/td>\n<td>102.3<\/td>\n<td>103.6<\/td>\n<\/tr>\n<tr>\n<td>211<\/td>\n<td>Tolylfluanid<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">238.0 &gt; 137.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">14<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">181.1 &gt; 138.1<\/td>\n<td>10<\/td>\n<td>66.8<\/td>\n<td>79.8<\/td>\n<td>65.9<\/td>\n<td>64.1<\/td>\n<td>118.7<\/td>\n<td>117.3<\/td>\n<td>114.3<\/td>\n<td>107.4<\/td>\n<\/tr>\n<tr>\n<td>212<\/td>\n<td>Transfluthrin<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">163.1 &gt; 127.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">6<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">163.1 &gt; 143.1<\/td>\n<td>16<\/td>\n<td>84.2<\/td>\n<td>85.7<\/td>\n<td>77.1<\/td>\n<td>77.6<\/td>\n<td>106.3<\/td>\n<td>115.0<\/td>\n<td>97.7<\/td>\n<td>99.0<\/td>\n<\/tr>\n<tr>\n<td>213<\/td>\n<td>Triadimefon<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">208.1 &gt; 111.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">22<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">208.1 &gt; 127.0<\/td>\n<td>14<\/td>\n<td>88.4<\/td>\n<td>98.2<\/td>\n<td>86.7<\/td>\n<td>80.5<\/td>\n<td>99.5<\/td>\n<td>103.9<\/td>\n<td>97.9<\/td>\n<td>104.7<\/td>\n<\/tr>\n<tr>\n<td>214<\/td>\n<td>Triadimenol<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">168.1 &gt; 70.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">10<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">128.1 &gt; 65.0<\/td>\n<td>22<\/td>\n<td>N.D.<\/td>\n<td>N.D.<\/td>\n<td>101.2<\/td>\n<td>94.2<\/td>\n<td>N.D.<\/td>\n<td>N.D.<\/td>\n<td>80.9<\/td>\n<td>85.6<\/td>\n<\/tr>\n<tr>\n<td>215<\/td>\n<td>Tri-allate<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">268.1 &gt; 184.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">20<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">270.1 &gt; 186.0<\/td>\n<td>20<\/td>\n<td>80.0<\/td>\n<td>75.6<\/td>\n<td>68.6<\/td>\n<td>74.6<\/td>\n<td>102.4<\/td>\n<td>85.9<\/td>\n<td>94.6<\/td>\n<td>98.1<\/td>\n<\/tr>\n<tr>\n<td>216<\/td>\n<td>Triazophos<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">257.0 &gt; 162.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">8<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">257.0 &gt; 134.0<\/td>\n<td>22<\/td>\n<td>89.7<\/td>\n<td>112.4<\/td>\n<td>94.8<\/td>\n<td>86.4<\/td>\n<td>95.9<\/td>\n<td>80.8<\/td>\n<td>90.3<\/td>\n<td>88.2<\/td>\n<\/tr>\n<tr>\n<td>217<\/td>\n<td>Tricyclazole<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">189.0 &gt; 161.9<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">12<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">189.0 &gt; 135.0<\/td>\n<td>18<\/td>\n<td>91.5<\/td>\n<td>95.7<\/td>\n<td>84.5<\/td>\n<td>81.8<\/td>\n<td>105.9<\/td>\n<td>118.7<\/td>\n<td>105.2<\/td>\n<td>97.3<\/td>\n<\/tr>\n<tr>\n<td>218<\/td>\n<td>Triflumizole<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">278.1 &gt; 73.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">6<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">206.1 &gt; 186.1<\/td>\n<td>8<\/td>\n<td>87.2<\/td>\n<td>88.2<\/td>\n<td>71.5<\/td>\n<td>76.8<\/td>\n<td>83.2<\/td>\n<td>110.9<\/td>\n<td>97.0<\/td>\n<td>99.2<\/td>\n<\/tr>\n<tr>\n<td>219<\/td>\n<td>Trifluralin<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">306.1 &gt; 264.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">8<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">306.1 &gt; 160.1<\/td>\n<td>22<\/td>\n<td>79.9<\/td>\n<td>73.6<\/td>\n<td>64.4<\/td>\n<td>72.5<\/td>\n<td>92.7<\/td>\n<td>100.3<\/td>\n<td>100.8<\/td>\n<td>101.1<\/td>\n<\/tr>\n<tr>\n<td>220<\/td>\n<td>Vinclozolin<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">285.0 &gt; 212.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">12<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">212.0 &gt; 172.0<\/td>\n<td>16<\/td>\n<td>86.3<\/td>\n<td>92.8<\/td>\n<td>79.9<\/td>\n<td>80.7<\/td>\n<td>84.3<\/td>\n<td>96.4<\/td>\n<td>104.2<\/td>\n<td>103.3<\/td>\n<\/tr>\n<tr>\n<td>QC-1<\/td>\n<td>1,4-Dichlorobenzene-d4<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">150.0 &gt; 78.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">24<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">115.1 &gt; 78.0<\/td>\n<td>12<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<\/tr>\n<tr>\n<td>QC-2<\/td>\n<td>Acenaphthene-d10<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">164.0 &gt; 160.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">30<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">164.0 &gt; 134.0<\/td>\n<td>38<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<\/tr>\n<tr>\n<td>QC-3<\/td>\n<td>Chrysene-d12<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">240.0 &gt; 236.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">30<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">240.0 &gt; 212.0<\/td>\n<td>24<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<\/tr>\n<tr>\n<td>QC-4<\/td>\n<td>Naphthalene-d8<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">136.0 &gt; 84.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">22<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">136.0 &gt; 82.0<\/td>\n<td>28<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<\/tr>\n<tr>\n<td>QC-5<\/td>\n<td>Perylene-d12<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">264.0 &gt; 263.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">34<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">264.0 &gt; 262.0<\/td>\n<td>24<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<\/tr>\n<tr>\n<td>QC-6<\/td>\n<td>Phenanthrene-d10<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">188.0 &gt; 160.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">24<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">187.0 &gt; 159.0<\/td>\n<td>18<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<\/tr>\n<tr>\n<td>IS-1<\/td>\n<td>2,2&#8242;,5-Trichlorobiphenyl<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">255.9 &gt; 186.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">26<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">257.9 &gt; 186.0<\/td>\n<td>26<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<\/tr>\n<tr>\n<td>IS-2<\/td>\n<td>2,4,4&#8242;-Trichlorobiphenyl<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">255.9 &gt; 186.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">26<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">257.9 &gt; 186.0<\/td>\n<td>26<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<\/tr>\n<tr>\n<td>IS-3<\/td>\n<td>2,2&#8242;,5,5&#8242;-Tetrachlorobiphenyl<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">257.0 &gt; 222.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">12<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">292.0 &gt; 220.0<\/td>\n<td>26<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<\/tr>\n<tr>\n<td>IS-4<\/td>\n<td>Triphenylmethane<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">244.1 &gt; 167.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">16<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">244.1 &gt; 165.1<\/td>\n<td>26<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<\/tr>\n<tr>\n<td>IS-5<\/td>\n<td>Triphenylphosphate<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">215.1 &gt; 168.1<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">16<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">325.1 &gt; 169.1<\/td>\n<td>20<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<\/tr>\n<tr>\n<td>IS-6<\/td>\n<td>Tris(1,3-dichloroisopropyl)phosphate<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">379.0 &gt; 159.0<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">12<\/td>\n<td class=\"has-text-align-left\" data-align=\"left\">381.0 &gt; 159.0<\/td>\n<td>12<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<td>\u2014<\/td>\n<\/tr>\n<\/tbody>\n<\/table>\n<\/figure>\n\n\n<div style=\"height:20px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<p class=\"wp-block-paragraph\">Compounds outside the red box (Figure 5) were classified to four groups.<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><strong>Figure 5:<\/strong>\u00a0Combination Map Between Recovery and Ion Ratio (Average of n = 3 for Each Fortified Sample) Red box shows the area of 70\u2013120% recovery and \u00b130% relative ion ratio.<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n<style>.kb-image45288_8232fb-0d .kb-image-has-overlay:after{opacity:0.3;}<\/style>\n<div class=\"wp-block-kadence-image kb-image45288_8232fb-0d\"><figure class=\"aligncenter size-large\"><img loading=\"lazy\" decoding=\"async\" width=\"1024\" height=\"630\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-FSAR3080-05-1024x630.jpg\" alt=\"\" class=\"kb-img wp-image-20056\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-FSAR3080-05-1024x630.jpg 1024w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-FSAR3080-05-300x185.jpg 300w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-FSAR3080-05-768x473.jpg 768w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-FSAR3080-05-1536x945.jpg 1536w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-FSAR3080-05.jpg 1800w\" sizes=\"auto, (max-width: 1024px) 100vw, 1024px\" \/><\/figure><\/div>\n\n<\/div><\/div><\/div>\n\n\n<p class=\"wp-block-paragraph\"><\/p>\n\n\n\n<div style=\"height:20px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<p class=\"wp-block-paragraph\">Group A showed low recovery; this group consisted mainly of compounds which have a low boiling point. They may have been lost in the evaporation step. Group B showed high relative ion ratios and this was caused by interference from matrix. Group C showed high recovery; this group consisted of 10 ng\/g fortified sample. Some of these were in matrix originally and quantified incorrectly. Others caused by their transitions which had low response and low stability. Group D showed low relative ion ratio. It was necessary to set higher response transitions. By modifying some procedures and parameters, positions of these compounds may improve.<\/p>\n\n\n\n<div style=\"height:20px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<h2 class=\"wp-block-heading\">Conclusion<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">This study shows that the modified QuEChERS method combined with GC-MS\/MS achieved consistent pesticides monitoring in botanical ingredients.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Although dried sample could make a heavy and difficult matrix, the modified QuEChERS method, SPE column cleanup, and toluene dilution steps suppressed interference from matrix. The GC-MS\/MS detected very low amounts of pesticides even though the sample was diluted. This analytical method takes only 30 minutes in total run time and covers over 200 pesticides. It provides a high throughput solution in laboratories doing this type of analysis.<\/p>\n\n\n\n<div style=\"height:20px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<h2 class=\"wp-block-heading\">References<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">[1] M. Anastassiades, S. J. Lehotay, D. \u0160tajnbaher, F. J. Schenck, Fast and Easy Multiresidue Method Employing Acetonitrile Extraction\/Partitioning and \u201cDispersive Solid-Phase Extraction\u201d for the Determination of Pesticide Residues in Produce,&nbsp;<em>J. AOAC Int<\/em>.,&nbsp;<strong>86<\/strong>&nbsp;(2003) 412\u2013431<br>[2] European Commission, Health &amp; Consumer Protection Directorate-General, Guidance document on analytical quality control and validation procedures for pesticide residues analysis in food and feed, SANCO\/12571\/2013<\/p>\n\n\n\n<div style=\"height:50px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n        <div class=\"cpb\">\n            <h3 class=\"cpb-heading\">Prodotti Menzionati<\/h3>\n            <hr class=\"cpb-heading-underline\" \/>\n            <div class=\"cpb-list\">\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/32562\">                                N\u00b0 Catalogo 32562                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Kit di standard per l\u2019analisi multiresiduale GC dei pesticidi<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/32562\">Esplora Prodotto<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/33267\">                                N\u00b0 Catalogo 33267                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">\ufeffMiscela di standard interno di QuEChERS per analisi GC\/MS\ufeff, in acetonitrile, 5 mL\/ampolla<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/33267\">Esplora Prodotto<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/31206\">                                N\u00b0 Catalogo 31206                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Miscela di standard interno SV, 2000 \u00b5g\/mL, diclorometano, 1 mL\/ampolla<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/31206\">Esplora Prodotto<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/25848\">                                N\u00b0 Catalogo 25848                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Kit di estrazione \ufeffQ-sep QuEChERS (originale), 4 g MgSO4, 1 g NaCl con provetta da centrifuga da 50 mL, 50 pacchetti e 50 provette<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/25848\">Esplora Prodotto<\/a>\n                                                    <\/div>\n                    <\/div>\n                            <\/div>\n        <\/div>\n        \n\n\n<div style=\"height:100px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n","protected":false},"excerpt":{"rendered":"<p>Dietary supplements, which are consumed worldwide, are made from various botanical ingredients. To be safe from pesticides exposure, residual pesticides must be monitored by chromatographic instrumentation. Issues arise, however, due to the fact these botanical samples are dried and cause large interferences in the chromatography. This study shows that the modified QuEChERS method combined with GC-MS\/MS achieves consistent pesticides monitoring in botanical ingredients.<\/p>\n","protected":false},"author":5,"featured_media":6979,"comment_status":"closed","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"_acf_changed":false,"_kad_blocks_custom_css":"","_kad_blocks_head_custom_js":"","_kad_blocks_body_custom_js":"","_kad_blocks_footer_custom_js":"","_kadence_starter_templates_imported_post":false,"_kad_post_transparent":"","_kad_post_title":"","_kad_post_layout":"","_kad_post_sidebar_id":"","_kad_post_content_style":"","_kad_post_vertical_padding":"","_kad_post_feature":"","_kad_post_feature_position":"","_kad_post_header":false,"_kad_post_footer":false,"footnotes":"","_ppma_block_editor_authors":""},"categories":[462],"tags":[],"industries-application":[2214,2216],"post-badge":[],"resource-type":[],"product-library":[2444,2456],"resource-technique":[2316,2320,2362,2367,2321],"ppma_author":[738,754,726,739,594,590,587],"class_list":["post-45288","post","type-post","status-publish","format-standard","has-post-thumbnail","hentry","category-articles-fr","industries-application-industrie-agroalimentaire","industries-application-securite-alimentaire","product-library-produits-pour-la-chromatographie-en-phase-gazeuse","product-library-colonnes-gc","resource-technique-chromatographie-en-phase-gazeuse-gc","resource-technique-spectrometrie-de-masse-ms","resource-technique-ms-ms-fr","resource-technique-quechers-fr","resource-technique-extraction-en-phase-solide-spe"],"acf":[],"taxonomy_info":{"category":[{"value":462,"label":"Articles"}],"industries-application":[{"value":2214,"label":"Industrie Agroalimentaire"},{"value":2216,"label":"S\u00e9curit\u00e9 alimentaire"}],"product-library":[{"value":2444,"label":"Produits pour la chromatographie en phase gazeuse"},{"value":2456,"label":"Colonnes GC"}],"resource-technique":[{"value":2316,"label":"Chromatographie en phase gazeuse (GC)"},{"value":2320,"label":"Spectrom\u00e9trie de masse (MS)"},{"value":2362,"label":"MS\/MS"},{"value":2367,"label":"QuEChERS"},{"value":2321,"label":"Extraction en phase solide (SPE)"}]},"featured_image_src_large":["https:\/\/discover.restek.com\/wp-content\/uploads\/feature-FSAR3080-1024x536.jpg",1024,536,true],"author_info":{"display_name":"Brenda Seprish","author_link":"https:\/\/discover.restek.com\/it\/author\/brenda-coganrestek-com\/"},"comment_info":0,"category_info":[{"term_id":462,"name":"Articles","slug":"articles-fr","term_group":0,"term_taxonomy_id":462,"taxonomy":"category","description":"","parent":0,"count":511,"filter":"raw","cat_ID":462,"category_count":511,"category_description":"","cat_name":"Articles","category_nicename":"articles-fr","category_parent":0}],"tag_info":false,"authors":[{"term_id":738,"user_id":0,"is_guest":1,"slug":"riki-kitano","display_name":"Riki Kitano","avatar_url":"https:\/\/secure.gravatar.com\/avatar\/?s=96&d=mm&r=g","author_category":"1","first_name":"Riki","last_name":"Kitano","user_url":"","job_title":"","description":"Shimadzu Scientific Instruments"},{"term_id":754,"user_id":0,"is_guest":1,"slug":"tairo-ogura","display_name":"Tairo Ogura","avatar_url":"https:\/\/secure.gravatar.com\/avatar\/?s=96&d=mm&r=g","author_category":"1","first_name":"Tairo","last_name":"Ogura","user_url":"","job_title":"","description":"Shimadzu Scientific Instruments"},{"term_id":726,"user_id":0,"is_guest":1,"slug":"nicole-lock","display_name":"Nicole Lock","avatar_url":"https:\/\/secure.gravatar.com\/avatar\/?s=96&d=mm&r=g","author_category":"1","first_name":"Nicole","last_name":"Lock","user_url":"","job_title":"","description":"Shimadzu Scientific Instruments"},{"term_id":739,"user_id":0,"is_guest":1,"slug":"robert-clifford","display_name":"Robert Clifford","avatar_url":"https:\/\/secure.gravatar.com\/avatar\/?s=96&d=mm&r=g","author_category":"1","first_name":"Robert","last_name":"Clifford","user_url":"","job_title":"","description":"Shimadzu Scientific Instruments"},{"term_id":594,"user_id":0,"is_guest":1,"slug":"julie-kowalski","display_name":"Julie Kowalski","avatar_url":"https:\/\/secure.gravatar.com\/avatar\/?s=96&d=mm&r=g","author_category":"1","first_name":"Julie","last_name":"Kowalski","user_url":"","job_title":"","description":""},{"term_id":590,"user_id":0,"is_guest":1,"slug":"jack-cochran","display_name":"Jack Cochran","avatar_url":"https:\/\/secure.gravatar.com\/avatar\/?s=96&d=mm&r=g","author_category":"","first_name":"","last_name":"","user_url":"","job_title":"","description":""},{"term_id":587,"user_id":0,"is_guest":1,"slug":"dan-li","display_name":"Dan Li","avatar_url":"https:\/\/secure.gravatar.com\/avatar\/?s=96&d=mm&r=g","author_category":"1","first_name":"Dan","last_name":"Li","user_url":"","job_title":"","description":""}],"_links":{"self":[{"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/posts\/45288","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/users\/5"}],"replies":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/comments?post=45288"}],"version-history":[{"count":18,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/posts\/45288\/revisions"}],"predecessor-version":[{"id":85579,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/posts\/45288\/revisions\/85579"}],"wp:featuredmedia":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/media\/6979"}],"wp:attachment":[{"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/media?parent=45288"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/categories?post=45288"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/tags?post=45288"},{"taxonomy":"industries-application","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/industries-application?post=45288"},{"taxonomy":"post-badge","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/post-badge?post=45288"},{"taxonomy":"resource-type","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/resource-type?post=45288"},{"taxonomy":"product-library","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/product-library?post=45288"},{"taxonomy":"resource-technique","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/resource-technique?post=45288"},{"taxonomy":"author","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/ppma_author?post=45288"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}