{"id":45314,"date":"2021-03-03T00:00:00","date_gmt":"2021-03-03T00:00:00","guid":{"rendered":"https:\/\/discover.restek.com\/uncategorized\/lanalisi-di-pesticidi-micotossine-e-cannabinoidi-nelle-caramelle-gommose-alla-cannabis\/"},"modified":"2026-01-28T22:06:53","modified_gmt":"2026-01-28T22:06:53","slug":"lanalisi-di-pesticidi-micotossine-e-cannabinoidi-nelle-caramelle-gommose-alla-cannabis","status":"publish","type":"post","link":"https:\/\/discover.restek.com\/it\/application-notes-it\/ffan3481-it\/lanalisi-di-pesticidi-micotossine-e-cannabinoidi-nelle-caramelle-gommose-alla-cannabis","title":{"rendered":"L\u2019analisi di pesticidi, micotossine e cannabinoidi nelle caramelle gommose alla cannabis"},"content":{"rendered":"\n<h2 class=\"wp-block-heading\">Abstract<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">Lo stato della California richiede l\u2019analisi di pesticidi, micotossine e cannabinoidi in tutti i tipi di prodotti derivati dalla cannabis [1]. Per&nbsp;questo motivo \u00e8 fondamentale sviluppare flussi di lavoro affidabili che consentano di determinare facilmente tali analiti in diverse matrici. Le caramelle gommose alla cannabis sono un prodotto alimentare molto diffuso, la cui composizione le rende una matrice altamente complessa. In questo studio illustreremo un flusso di lavoro completo, che utilizza un\u2019unica procedura di estrazione, per l\u2019analisi nelle caramelle gommose dei pesticidi, delle micotossine e dei cannabinoidi indicati dal governo californiano.<\/p>\n\n\n\n<h2 class=\"wp-block-heading\">Introduzione<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">I prodotti alimentari arricchiti di cannabis o cannabidiolo (CBD) sono sempre pi\u00f9 diffusi fra i consumatori di cannabis. In ambito analitico, \u00e8 obbligatorio eseguire analisi di potenza dei prodotti alimentari in tutti gli Stati che richiedono per legge le analisi di laboratorio. Inoltre, lo stato della California prevede anche la determinazione di contaminanti, come pesticidi e micotossine, negli alimenti e in altri tipi di prodotti derivati dalla cannabis. Data l\u2019ampia variet\u00e0 di prodotti alimentari disponibili in commercio, per ottenere dati affidabili per tutti gli analiti di interesse e in qualsiasi tipo di matrice \u00e8 necessario adottare strategie analitiche differenti. Fra i vari tipi di prodotti alimentari a base di cannabis, le caramelle gommose rappresentano una grande sfida per i laboratori di analisi;<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">si tratta di una matrice \u201cappiccicosa\u201d composta in genere da zucchero, amido, pectina e gelatina. Data la composizione del campione, le strategie di preparazione per l\u2019analisi di potenza solitamente prevedono una fase di solubilizzazione in solventi come acqua o DMSO, oppure, in alternativa, si possono utilizzare grandi volumi di solventi come il metanolo. Una volta omogenizzato il campione, i cannabinoidi vengono quantificati tramite HPLC-UV iniettando l\u2019estratto diluito direttamente oppure dopo una fase di salting-out eseguita con i sali QuEChERS. Esistono poche informazioni di dominio pubblico relative all\u2019analisi di contaminanti come pesticidi e micotossine in matrici di caramelle gommose; \u00e8 infatti disponibile un solo report che illustra i dati semi-quantitativi di 35 pesticidi utilizzando i QuEChERS [2].<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">In questo studio delineiamo un flusso di lavoro affidabile per la determinazione quantitativa nei campioni di caramelle gommose di pesticidi, micotossine e cannabinoidi presenti nell&#8217;elenco stilato dal governo californiano. La metodologia ottimizzata di preparazione prevede la solubilizzazione del campione, seguita da una fase di estrazione con acetonitrile acidificato e un passaggio di salting-out per il quale si utilizzano i sali QuEChERS per EN. Per l\u2019analisi di cannabinoidi e contaminanti compatibili con la LC \u00e8 stata condotta una diluizione semplice prima dell\u2019iniezione, mentre per i pesticidi analizzabili in GC, prima dell\u2019analisi, \u00e8 stato necessario utilizzare una miscela di sorbente dSPE contenente ammine primarie\/secondarie (PSA), nero di carbonio grafitato (GCB) e solfato di magnesio. Nel complesso, il flusso di lavoro proposto per l\u2019analisi di pesticidi, micotossine e cannabinoidi nelle caramelle gommose alla cannabis ha fornito risultati soddisfacenti in termini di linearit\u00e0, accuratezza, precisione e limiti di quantificazione (LOQ).<\/p>\n\n\n\n<h2 class=\"wp-block-heading\">Studio sperimentale<\/h2>\n\n\n\n<h3 class=\"wp-block-heading\">Preparazione del campione<\/h3>\n\n\n\n<p class=\"wp-block-paragraph\">A valle dei test preliminari sui vari parametri sperimentali \u00e8 stato implementato il seguente flusso di lavoro (Figura 1). \u00c8 stato pesato 1\u202fg di caramelle tritate in una provetta da 50 mL (cat.# 25846) e sono stati aggiunti 5 mL di acqua. I campioni sono stati poi vortexati con forza finch\u00e9 i pezzi di caramella si sono solubilizzati. La soluzione del campione \u00e8 stata poi adeguatamente fortificata con analiti target e\/o standard interni (gli standard interni sono stati utilizzati solo per l\u2019analisi dei contaminanti e non per l\u2019analisi di potenza). Dopodich\u00e9 il tutto \u00e8 stato vortexato per ulteriori 30 secondi. A seguire, sono stati aggiunti 5 mL di acetonitrile acidificato con acido acetico (1% v:v) e i&nbsp;campioni sono stati nuovamente vortexati per 30 secondi. \u00c8 stato aggiunto un pacchetto di sali di estrazione Q-sep QuEChERS per la norma europea EN\u202f15662 (cat.# 25849), poi il campione \u00e8 stato passato nuovamente al vortex e centrifugato per 5 minuti.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Per l\u2019analisi di potenza sono stati miscelati 100 \u00b5L di estratto del campione e 900 \u00b5L di 25:75 acqua:acetonitrile, mentre 2 \u00b5L sono stati analizzati in LC-UV. Per quanto riguarda i contaminanti compatibili con la LC, sono stati miscelati 750 \u00b5L di estratto del campione e 250\u202f\u00b5L di acqua, mentre 2 \u03bcL sono stati analizzati in LC-MS\/MS. Sul fronte invece dei contaminanti analizzabili in GC, sono stati trasferiti 1,9\u202fmL di estratto in una provetta dSPE Q-sep QuEChERS contenente sorbenti pre-pesati di PSA, GCB e solfato di magnesio (cat.# 26217). Dopo&nbsp;il passaggio al vortex e la centrifuga sono stati miscelati 500 \u03bcL di estratto con 500 \u03bcL di 1% di acido acetico in acetonitrile, mentre 1\u202f\u03bcL \u00e8&nbsp;stato analizzato in GC-MS\/MS.<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><strong>Figura 1:<\/strong>\u00a0Procedura di preparazione del campione per l\u2019analisi di pesticidi, micotossine e cannabinoidi nelle caramelle gommose alla cannabis.<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n<style>.kb-image45314_4e0ec1-f5 .kb-image-has-overlay:after{opacity:0.3;}<\/style>\n<div class=\"wp-block-kadence-image kb-image45314_4e0ec1-f5\"><figure class=\"aligncenter size-large\"><img loading=\"lazy\" decoding=\"async\" width=\"629\" height=\"1024\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/\/figure-article-FFAN3481-01-629x1024.jpg\" alt=\"\" class=\"kb-img wp-image-19838\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-FFAN3481-01-629x1024.jpg 629w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-FFAN3481-01-184x300.jpg 184w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-FFAN3481-01-768x1251.jpg 768w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-FFAN3481-01-943x1536.jpg 943w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-FFAN3481-01-1258x2048.jpg 1258w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-FFAN3481-01-scaled.jpg 1572w\" sizes=\"auto, (max-width: 629px) 100vw, 629px\" \/><\/figure><\/div>\n\n<\/div><\/div><\/div>\n\n\n<div style=\"height:20px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<h3 class=\"wp-block-heading\">Quantificazione<\/h3>\n\n\n\n<p class=\"wp-block-paragraph\">Ai fini delle analisi di pesticidi e micotossine, le soluzioni di calibrazione sono state preparate fortificando con analiti e standard interni le aliquote dell\u2019estratto ottenuto da campioni bianchi di caramelle gommose (\u00e8 stato ottenuto un estratto misto miscelando gli estratti di vari campioni bianchi di caramelle gommose). La Tabella I mostra il volume della soluzione di spiking con analiti target e della soluzione della miscela di standard interni aggiunto a ciascuna aliquota (volume della soluzione di calibrazione finale = 3 mL). Per costruire le curve di calibrazione dei pesticidi analizzabili in GC, 1,9 mL dei 3 mL delle soluzioni di calibrazione sono stati sottoposti a purificazione dSPE, come descritto nella sezione dedicata alla preparazione del campione. L\u2019accuratezza e la precisione del metodo sono state valutate fortificando in quadruplicato le caramelle gommose omogeneizzate a 10, 50, 100 e 500 ng\/g, dopo l\u2019aggiunta di acqua e il passaggio al vortex (Tabella II), e seguendo tutto il flusso di lavoro di preparazione del campione descritto nella sezione dedicata.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Per l\u2019analisi dei cannabinoidi sono state preparate soluzioni di calibrazione a 2, 5, 10, 20, 50, 100 e 200 ppm in 75:25 acetonitrile:acqua. \u00c8 stato valutato il recupero dei cannabinoidi fortificando campioni di 1 g di caramelle gommose solubilizzate in acqua a 0,2 e 0,5\u202fmg\/g (n=2), procedendo poi all\u2019estrazione come descritto sopra.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\"><strong>Tabella I:<\/strong>&nbsp;Preparazione dei calibratori per l\u2019analisi di pesticidi e micotossine con l\u2019utilizzo di aliquote di estratti ottenuti da campioni bianchi di caramelle gommose (volume finale di ciascuna soluzione di calibrazione&nbsp;=&nbsp;3 mL).<\/p>\n\n\n\n<figure class=\"wp-block-table\"><table class=\"has-fixed-layout\"><tbody><tr><td class=\"has-text-align-center\" data-align=\"center\"><strong>Analiti desiderati<\/strong><br><strong>Conc. nella matrice (ng\/g)<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>Conc. dell\u2019analita nell\u2019estratto finale ipotizzando un recupero del 100% dalla matrice (ng\/mL)<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>\u00b5L di soluzione di analiti target addizionati nell&#8217;estratto bianco<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>Conc. dell\u2019analita nella soluzione di spiking (ng\/mL)<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>\u00b5L di 5000 ng\/mL di miscela di standard interno aggiunti all\u2019estratto bianco<\/strong><\/td><\/tr><tr><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><td class=\"has-text-align-center\" data-align=\"center\">1<\/td><td class=\"has-text-align-center\" data-align=\"center\">30<\/td><td class=\"has-text-align-center\" data-align=\"center\">100<\/td><td class=\"has-text-align-center\" data-align=\"center\">24<\/td><\/tr><tr><td class=\"has-text-align-center\" data-align=\"center\">20<\/td><td class=\"has-text-align-center\" data-align=\"center\">4<\/td><td class=\"has-text-align-center\" data-align=\"center\">12<\/td><td class=\"has-text-align-center\" data-align=\"center\">1000<\/td><td class=\"has-text-align-center\" data-align=\"center\">24<\/td><\/tr><tr><td class=\"has-text-align-center\" data-align=\"center\">50<\/td><td class=\"has-text-align-center\" data-align=\"center\">10<\/td><td class=\"has-text-align-center\" data-align=\"center\">30<\/td><td class=\"has-text-align-center\" data-align=\"center\">1000<\/td><td class=\"has-text-align-center\" data-align=\"center\">24<\/td><\/tr><tr><td class=\"has-text-align-center\" data-align=\"center\">75<\/td><td class=\"has-text-align-center\" data-align=\"center\">15<\/td><td class=\"has-text-align-center\" data-align=\"center\">45<\/td><td class=\"has-text-align-center\" data-align=\"center\">1000<\/td><td class=\"has-text-align-center\" data-align=\"center\">24<\/td><\/tr><tr><td class=\"has-text-align-center\" data-align=\"center\">150<\/td><td class=\"has-text-align-center\" data-align=\"center\">30<\/td><td class=\"has-text-align-center\" data-align=\"center\">90<\/td><td class=\"has-text-align-center\" data-align=\"center\">1000<\/td><td class=\"has-text-align-center\" data-align=\"center\">24<\/td><\/tr><tr><td class=\"has-text-align-center\" data-align=\"center\">200<\/td><td class=\"has-text-align-center\" data-align=\"center\">40<\/td><td class=\"has-text-align-center\" data-align=\"center\">24<\/td><td class=\"has-text-align-center\" data-align=\"center\">5000<\/td><td class=\"has-text-align-center\" data-align=\"center\">24<\/td><\/tr><tr><td class=\"has-text-align-center\" data-align=\"center\">400<\/td><td class=\"has-text-align-center\" data-align=\"center\">80<\/td><td class=\"has-text-align-center\" data-align=\"center\">48<\/td><td class=\"has-text-align-center\" data-align=\"center\">5000<\/td><td class=\"has-text-align-center\" data-align=\"center\">24<\/td><\/tr><tr><td class=\"has-text-align-center\" data-align=\"center\">700<\/td><td class=\"has-text-align-center\" data-align=\"center\">140<\/td><td class=\"has-text-align-center\" data-align=\"center\">84<\/td><td class=\"has-text-align-center\" data-align=\"center\">5000<\/td><td class=\"has-text-align-center\" data-align=\"center\">24<\/td><\/tr><\/tbody><\/table><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\"><strong>Tabella II:<\/strong>&nbsp;Fortificazione di micotossine e pesticidi nelle caramelle gommose (1&nbsp;g di campione omogeneizzato con 5 mL di acqua) a diversi livelli di concentrazione..<\/p>\n\n\n\n<figure class=\"wp-block-table\"><table class=\"has-fixed-layout\"><tbody><tr><td class=\"has-text-align-center\" data-align=\"center\"><strong>Conc. nella matrice<\/strong><br><strong>(ng\/g)<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>\u00b5L di soluzione di analiti target addizionati nei campioni omogeneizzati in acqua<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>Conc. dell\u2019analita nella soluzione di spiking<\/strong><br><strong>(ng\/mL)<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>\u00b5L di 5000 ng\/mL di miscela di standard interno aggiunti a campioni omogeneizzati in acqua<\/strong><\/td><\/tr><tr><td class=\"has-text-align-center\" data-align=\"center\">10<\/td><td class=\"has-text-align-center\" data-align=\"center\">10<\/td><td class=\"has-text-align-center\" data-align=\"center\">1000<\/td><td class=\"has-text-align-center\" data-align=\"center\">40<\/td><\/tr><tr><td class=\"has-text-align-center\" data-align=\"center\">50<\/td><td class=\"has-text-align-center\" data-align=\"center\">50<\/td><td class=\"has-text-align-center\" data-align=\"center\">1000<\/td><td class=\"has-text-align-center\" data-align=\"center\">40<\/td><\/tr><tr><td class=\"has-text-align-center\" data-align=\"center\">100<\/td><td class=\"has-text-align-center\" data-align=\"center\">100<\/td><td class=\"has-text-align-center\" data-align=\"center\">1000<\/td><td class=\"has-text-align-center\" data-align=\"center\">40<\/td><\/tr><tr><td class=\"has-text-align-center\" data-align=\"center\">500<\/td><td class=\"has-text-align-center\" data-align=\"center\">100<\/td><td class=\"has-text-align-center\" data-align=\"center\">5000<\/td><td class=\"has-text-align-center\" data-align=\"center\">40<\/td><\/tr><\/tbody><\/table><\/figure>\n\n\n\n<h3 class=\"wp-block-heading\">Parametri dello strumento<\/h3>\n\n\n\n<p class=\"wp-block-paragraph\">La strumentazione e le condizioni per l&#8217;analisi di pesticidi, micotossine e cannabinoidi nelle caramelle di cannabis sono presentate nelle tabelle III, IV e V. Le transizioni ioniche per i contaminanti ammessi alla LC e alla GC sono presentate rispettivamente nelle tabelle VI e VII. I tempi di ritenzione dei cannabinoidi sono riportati nella Tabella VIII.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\"><strong>Tabella III:<\/strong>&nbsp;Condizioni LC-MS\/MS (pesticidi e micotossine).<\/p>\n\n\n\n<figure class=\"wp-block-table\"><table class=\"has-fixed-layout\"><tbody><tr><td><strong>Colonna<\/strong><\/td><td colspan=\"4\">Raptor ARC-18 2.7 \u00b5m, 100 mm x 2.1 mm (cat.# 9314A12)<\/td><\/tr><tr><td><strong>Precolonna<\/strong><\/td><td colspan=\"4\">Cartuccia per precolonna Raptor ARC-18 EXP 2,7 \u00b5m, 5 x 2,1 mm (cat.# 9314A0252)<\/td><\/tr><tr><td><strong>Fase mobile A<\/strong><\/td><td colspan=\"4\">Acqua, formato di ammonio 2 mM, acido formico 0,1%<\/td><\/tr><tr><td><strong>Fase mobile B<\/strong><\/td><td colspan=\"4\">Metanolo, formato di ammonio 2 mM, acido formico 0,1%<\/td><\/tr><tr><td rowspan=\"5\"><strong>Scansione temporale<\/strong><\/td><td>Tempo (min.)<\/td><td>%B<\/td><td>Tempo (min.)<\/td><td>%B<\/td><\/tr><tr><td>0<\/td><td>5<\/td><td>10.5<\/td><td>100<\/td><\/tr><tr><td>1.5<\/td><td>65<\/td><td>10.6<\/td><td>5<\/td><\/tr><tr><td>8.5<\/td><td>95<\/td><td>12.0<\/td><td>5<\/td><\/tr><tr><td>9.5<\/td><td>100<\/td><td>&#8211;<\/td><td>&#8211;<\/td><\/tr><tr><td><strong>Flusso<\/strong><\/td><td colspan=\"4\">0.5 mL\/min<\/td><\/tr><tr><td><strong>Temp. colonna<\/strong><\/td><td colspan=\"4\">40 \u00b0C<\/td><\/tr><tr><td><strong>Temp. autocampionatore<\/strong><\/td><td colspan=\"4\">10 \u00b0C<\/td><\/tr><tr><td><strong>Volume di iniezione<\/strong><\/td><td colspan=\"4\">2 \u03bcL<\/td><\/tr><tr><td><strong>Strumento<\/strong><\/td><td colspan=\"4\">Shimadzu LCMS-8060<\/td><\/tr><\/tbody><\/table><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\"><strong>Tabella IV:<\/strong>&nbsp;Condizioni GC-MS\/MS (pesticidi).<\/p>\n\n\n\n<figure class=\"wp-block-table\"><table class=\"has-fixed-layout\"><tbody><tr><td><strong>Strumento<\/strong><\/td><td>Thermo Trace 1310-TSQ 8000<\/td><\/tr><tr><td><strong>Colonna:<\/strong><\/td><td>Rxi-5ms, 30 m x 0.25 mm ID x 0.25 \u00b5m, (cat.# 13423)<\/td><\/tr><tr><td><strong>Modalit\u00e0 di iniezione<\/strong><\/td><td>Splitless<\/td><\/tr><tr><td><strong>Vol. iniezione<\/strong><\/td><td>1 \u00b5L<\/td><\/tr><tr><td><strong>Liner<\/strong><\/td><td>Topaz a cono singolo con ID 4,0 mm con lana (cat.# 23447)<\/td><\/tr><tr><td><strong>Temp. iniez.<\/strong><\/td><td>250 \u00b0C<\/td><\/tr><tr><td><strong>Flusso split<\/strong><\/td><td>14.0 mL\/min<\/td><\/tr><tr><td><strong>Tempo splitless<\/strong><\/td><td>0.50 min<\/td><\/tr><tr><td><strong>Flusso di spurgo<\/strong><\/td><td>5 mL\/min<\/td><\/tr><tr><td><strong>Forno<\/strong><\/td><td>90&nbsp;\u00b0C (mantenimento 1 min.) fino a 310&nbsp;\u00b0C a 25&nbsp;\u00b0C\/min (mantenimento 10 min.)<\/td><\/tr><tr><td><strong>Carrier Gas<\/strong><\/td><td>He, flusso costante<\/td><\/tr><tr><td><strong>Velocit\u00e0 flusso<\/strong><\/td><td>1.40 mL\/min<\/td><\/tr><tr><td><strong>Detector<\/strong><\/td><td>MS\/MS<\/td><\/tr><tr><td><strong>Tipo di metodo<\/strong><\/td><td>Acquisizione &#8211; a tempo<\/td><\/tr><tr><td><strong>Modalit\u00e0 di ionizzazione<\/strong><\/td><td>EI<\/td><\/tr><tr><td><strong>Temp. linea di trasferimento<\/strong><\/td><td>290 \u00b0C<\/td><\/tr><tr><td><strong>Temp. sorgente<\/strong><\/td><td>330 \u00b0C<\/td><\/tr><\/tbody><\/table><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\"><strong>Tabella V:<\/strong>&nbsp;Condizioni HPLC-UV (analisi di potenza con un metodo a risparmio di solvente [3]).<\/p>\n\n\n\n<figure class=\"wp-block-table\"><table class=\"has-fixed-layout\"><tbody><tr><td><strong>Strumento<\/strong><\/td><td>Waters ACQUITY<\/td><\/tr><tr><td><strong>Colonna:<\/strong><\/td><td>Raptor ARC-18 2.7 \u00b5m, 150 mm x 2.1 mm (cat.# 9314A62)<\/td><\/tr><tr><td><strong>Precolonna<\/strong><\/td><td>Cartuccia per precolonna Raptor ARC-18 EXP 2,7\u202f\u00b5m, 5 x 2,1 mm (cat.# 9314A0252)<\/td><\/tr><tr><td><strong>Vol. iniezione<\/strong><\/td><td>2 \u00b5L<\/td><\/tr><tr><td><strong>Fase mobile A<\/strong><\/td><td>Acqua, formato di ammonio 5 mM,<br>acido formico 0,1%<\/td><\/tr><tr><td><strong>Fase mobile B<\/strong><\/td><td>Acetonitrile, acido formico 0,1%<\/td><\/tr><tr><td><strong>Gradiente<\/strong><\/td><td>Isocratico, 75% B<\/td><\/tr><tr><td><strong>Flusso<\/strong><\/td><td>0.4 mL\/min<\/td><\/tr><tr><td><strong>Temp. Colonna<\/strong><\/td><td>30 \u00b0C<\/td><\/tr><tr><td><strong>Autosampler Temp.<\/strong><\/td><td>10 \u00b0C<\/td><\/tr><tr><td><strong>Lunghezza d\u2019onda<\/strong><\/td><td>228 nm<\/td><\/tr><\/tbody><\/table><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\"><strong>Tabella VI:<\/strong>&nbsp;Transizioni LC-MS\/MS<\/p>\n\n\n\n<figure class=\"wp-block-table\"><table class=\"has-fixed-layout\"><tbody><tr><td><strong>Nome<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>Tempo di ritenzione (min.)<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>Ione precursore<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>Ione prodotto 1<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>Product Ion 2<\/strong><\/td><\/tr><tr><td>Daminozide-D6<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.7<\/td><td class=\"has-text-align-center\" data-align=\"center\">167.0<\/td><td class=\"has-text-align-center\" data-align=\"center\">149.3<\/td><td class=\"has-text-align-center\" data-align=\"center\">49.3<\/td><\/tr><tr><td>Daminozide<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.7<\/td><td class=\"has-text-align-center\" data-align=\"center\">161.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">44.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">143.2<\/td><\/tr><tr><td>Acephate<\/td><td class=\"has-text-align-center\" data-align=\"center\">1.7<\/td><td class=\"has-text-align-center\" data-align=\"center\">184.0<\/td><td class=\"has-text-align-center\" data-align=\"center\">143.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">95.1<\/td><\/tr><tr><td>Oxamyl<\/td><td class=\"has-text-align-center\" data-align=\"center\">2.0<\/td><td class=\"has-text-align-center\" data-align=\"center\">237.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">72.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">90.1<\/td><\/tr><tr><td>Flonicamid<\/td><td class=\"has-text-align-center\" data-align=\"center\">2.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">230.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">203.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">174.1<\/td><\/tr><tr><td>Methomyl<\/td><td class=\"has-text-align-center\" data-align=\"center\">2.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">163.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">88.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">106.1<\/td><\/tr><tr><td>Thiamethoxam<\/td><td class=\"has-text-align-center\" data-align=\"center\">2.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">292.0<\/td><td class=\"has-text-align-center\" data-align=\"center\">211.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">181.1<\/td><\/tr><tr><td>Imidacloprid<\/td><td class=\"has-text-align-center\" data-align=\"center\">2.3<\/td><td class=\"has-text-align-center\" data-align=\"center\">256.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">209.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">175.1<\/td><\/tr><tr><td>Mevinphos<\/td><td class=\"has-text-align-center\" data-align=\"center\">2.4<\/td><td class=\"has-text-align-center\" data-align=\"center\">225.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">127.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">193.2<\/td><\/tr><tr><td>Acetamiprid<\/td><td class=\"has-text-align-center\" data-align=\"center\">2.4<\/td><td class=\"has-text-align-center\" data-align=\"center\">223.0<\/td><td class=\"has-text-align-center\" data-align=\"center\">126.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">56.1<\/td><\/tr><tr><td>Dimethoathe-D6<\/td><td class=\"has-text-align-center\" data-align=\"center\">2.4<\/td><td class=\"has-text-align-center\" data-align=\"center\">236.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">205.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">&#8211;<\/td><\/tr><tr><td>Dimethoate<\/td><td class=\"has-text-align-center\" data-align=\"center\">2.4<\/td><td class=\"has-text-align-center\" data-align=\"center\">230.0<\/td><td class=\"has-text-align-center\" data-align=\"center\">199.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">125.1<\/td><\/tr><tr><td>Thiacloprid<\/td><td class=\"has-text-align-center\" data-align=\"center\">2.5<\/td><td class=\"has-text-align-center\" data-align=\"center\">253.0<\/td><td class=\"has-text-align-center\" data-align=\"center\">126.0<\/td><td class=\"has-text-align-center\" data-align=\"center\">90.1<\/td><\/tr><tr><td>Aflatoxin G2<\/td><td class=\"has-text-align-center\" data-align=\"center\">2.5<\/td><td class=\"has-text-align-center\" data-align=\"center\">331.2<\/td><td class=\"has-text-align-center\" data-align=\"center\">189.3<\/td><td class=\"has-text-align-center\" data-align=\"center\">115.2<\/td><\/tr><tr><td>Aflatoxin G1<\/td><td class=\"has-text-align-center\" data-align=\"center\">2.5<\/td><td class=\"has-text-align-center\" data-align=\"center\">329.2<\/td><td class=\"has-text-align-center\" data-align=\"center\">243.2<\/td><td class=\"has-text-align-center\" data-align=\"center\">215.3<\/td><\/tr><tr><td>Aldicarb<\/td><td class=\"has-text-align-center\" data-align=\"center\">2.6<\/td><td class=\"has-text-align-center\" data-align=\"center\">116.0<\/td><td class=\"has-text-align-center\" data-align=\"center\">89.2<\/td><td class=\"has-text-align-center\" data-align=\"center\">70.2<\/td><\/tr><tr><td>Aflatoxin B2<\/td><td class=\"has-text-align-center\" data-align=\"center\">2.6<\/td><td class=\"has-text-align-center\" data-align=\"center\">315.3<\/td><td class=\"has-text-align-center\" data-align=\"center\">287.2<\/td><td class=\"has-text-align-center\" data-align=\"center\">243.3<\/td><\/tr><tr><td>Dichlorvos<\/td><td class=\"has-text-align-center\" data-align=\"center\">2.7<\/td><td class=\"has-text-align-center\" data-align=\"center\">220.9<\/td><td class=\"has-text-align-center\" data-align=\"center\">109.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">79.2<\/td><\/tr><tr><td>Dichlorvos-D6<\/td><td class=\"has-text-align-center\" data-align=\"center\">2.7<\/td><td class=\"has-text-align-center\" data-align=\"center\">227.0<\/td><td class=\"has-text-align-center\" data-align=\"center\">115.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">&#8211;<\/td><\/tr><tr><td>Aflatoxin B1<\/td><td class=\"has-text-align-center\" data-align=\"center\">2.7<\/td><td class=\"has-text-align-center\" data-align=\"center\">313.2<\/td><td class=\"has-text-align-center\" data-align=\"center\">241.2<\/td><td class=\"has-text-align-center\" data-align=\"center\">128.2<\/td><\/tr><tr><td>Imazalil<\/td><td class=\"has-text-align-center\" data-align=\"center\">2.7<\/td><td class=\"has-text-align-center\" data-align=\"center\">297.0<\/td><td class=\"has-text-align-center\" data-align=\"center\">159.0<\/td><td class=\"has-text-align-center\" data-align=\"center\">201.0<\/td><\/tr><tr><td>Carbofuran<\/td><td class=\"has-text-align-center\" data-align=\"center\">2.7<\/td><td class=\"has-text-align-center\" data-align=\"center\">222.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">123.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">165.2<\/td><\/tr><tr><td>Propoxur<\/td><td class=\"has-text-align-center\" data-align=\"center\">2.7<\/td><td class=\"has-text-align-center\" data-align=\"center\">210.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">111.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">93.1<\/td><\/tr><tr><td>Carbaryl-D7<\/td><td class=\"has-text-align-center\" data-align=\"center\">2.8<\/td><td class=\"has-text-align-center\" data-align=\"center\">209.2<\/td><td class=\"has-text-align-center\" data-align=\"center\">152.2<\/td><td class=\"has-text-align-center\" data-align=\"center\">&#8211;<\/td><\/tr><tr><td>Carbaryl<\/td><td class=\"has-text-align-center\" data-align=\"center\">2.8<\/td><td class=\"has-text-align-center\" data-align=\"center\">202.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">145.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">127.1<\/td><\/tr><tr><td>Diuron-D6<\/td><td class=\"has-text-align-center\" data-align=\"center\">3.0<\/td><td class=\"has-text-align-center\" data-align=\"center\">239.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">78.2<\/td><td class=\"has-text-align-center\" data-align=\"center\">&#8211;<\/td><\/tr><tr><td>Atrazine-D5<\/td><td class=\"has-text-align-center\" data-align=\"center\">3.0<\/td><td class=\"has-text-align-center\" data-align=\"center\">221.2<\/td><td class=\"has-text-align-center\" data-align=\"center\">179.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">&#8211;<\/td><\/tr><tr><td>Naled<\/td><td class=\"has-text-align-center\" data-align=\"center\">3.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">397.8<\/td><td class=\"has-text-align-center\" data-align=\"center\">127.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">109.1<\/td><\/tr><tr><td>Metalaxyl<\/td><td class=\"has-text-align-center\" data-align=\"center\">3.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">280.2<\/td><td class=\"has-text-align-center\" data-align=\"center\">220.2<\/td><td class=\"has-text-align-center\" data-align=\"center\">192.2<\/td><\/tr><tr><td>Spiroxamine<\/td><td class=\"has-text-align-center\" data-align=\"center\">3.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">298.3<\/td><td class=\"has-text-align-center\" data-align=\"center\">144.2<\/td><td class=\"has-text-align-center\" data-align=\"center\">100.2<\/td><\/tr><tr><td>Chlorantraniliprole<\/td><td class=\"has-text-align-center\" data-align=\"center\">3.2<\/td><td class=\"has-text-align-center\" data-align=\"center\">483.9<\/td><td class=\"has-text-align-center\" data-align=\"center\">452.9<\/td><td class=\"has-text-align-center\" data-align=\"center\">285.9<\/td><\/tr><tr><td>Phosmet<\/td><td class=\"has-text-align-center\" data-align=\"center\">3.2<\/td><td class=\"has-text-align-center\" data-align=\"center\">318.0<\/td><td class=\"has-text-align-center\" data-align=\"center\">160.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">77.2<\/td><\/tr><tr><td>Azoxystrobin<\/td><td class=\"has-text-align-center\" data-align=\"center\">3.3<\/td><td class=\"has-text-align-center\" data-align=\"center\">404.0<\/td><td class=\"has-text-align-center\" data-align=\"center\">372.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">344.1<\/td><\/tr><tr><td>Linuron-D6<\/td><td class=\"has-text-align-center\" data-align=\"center\">3.3<\/td><td class=\"has-text-align-center\" data-align=\"center\">255.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">160.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">&#8211;<\/td><\/tr><tr><td>Fludioxonil*<\/td><td class=\"has-text-align-center\" data-align=\"center\">3.4<\/td><td class=\"has-text-align-center\" data-align=\"center\">247.0<\/td><td class=\"has-text-align-center\" data-align=\"center\">180.0<\/td><td class=\"has-text-align-center\" data-align=\"center\">126.0<\/td><\/tr><tr><td>Methiocarb<\/td><td class=\"has-text-align-center\" data-align=\"center\">3.4<\/td><td class=\"has-text-align-center\" data-align=\"center\">226.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">169.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">121.1<\/td><\/tr><tr><td>Dimethomorph<\/td><td class=\"has-text-align-center\" data-align=\"center\">3.5<\/td><td class=\"has-text-align-center\" data-align=\"center\">388.2<\/td><td class=\"has-text-align-center\" data-align=\"center\">301.2<\/td><td class=\"has-text-align-center\" data-align=\"center\">165.3<\/td><\/tr><tr><td>Boscalid<\/td><td class=\"has-text-align-center\" data-align=\"center\">3.5<\/td><td class=\"has-text-align-center\" data-align=\"center\">342.9<\/td><td class=\"has-text-align-center\" data-align=\"center\">307.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">140.1<\/td><\/tr><tr><td>Paclobutrazol<\/td><td class=\"has-text-align-center\" data-align=\"center\">3.6<\/td><td class=\"has-text-align-center\" data-align=\"center\">294.3<\/td><td class=\"has-text-align-center\" data-align=\"center\">70.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">125.1<\/td><\/tr><tr><td>Malathion<\/td><td class=\"has-text-align-center\" data-align=\"center\">3.6<\/td><td class=\"has-text-align-center\" data-align=\"center\">331.0<\/td><td class=\"has-text-align-center\" data-align=\"center\">127.2<\/td><td class=\"has-text-align-center\" data-align=\"center\">285.2<\/td><\/tr><tr><td>Myclobutanil<\/td><td class=\"has-text-align-center\" data-align=\"center\">3.7<\/td><td class=\"has-text-align-center\" data-align=\"center\">289.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">70.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">125.1<\/td><\/tr><tr><td>Bifenazate<\/td><td class=\"has-text-align-center\" data-align=\"center\">3.7<\/td><td class=\"has-text-align-center\" data-align=\"center\">301.0<\/td><td class=\"has-text-align-center\" data-align=\"center\">198.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">170.2<\/td><\/tr><tr><td>Ochratoxin A<\/td><td class=\"has-text-align-center\" data-align=\"center\">3.8<\/td><td class=\"has-text-align-center\" data-align=\"center\">404.2<\/td><td class=\"has-text-align-center\" data-align=\"center\">239.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">358.3<\/td><\/tr><tr><td>Fenhexamid<\/td><td class=\"has-text-align-center\" data-align=\"center\">3.9<\/td><td class=\"has-text-align-center\" data-align=\"center\">302.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">97.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">55.2<\/td><\/tr><tr><td>Spirotetramat<\/td><td class=\"has-text-align-center\" data-align=\"center\">4.0<\/td><td class=\"has-text-align-center\" data-align=\"center\">374.2<\/td><td class=\"has-text-align-center\" data-align=\"center\">302.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">216.1<\/td><\/tr><tr><td>Ethoprophos<\/td><td class=\"has-text-align-center\" data-align=\"center\">4.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">243.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">131.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">97.1<\/td><\/tr><tr><td>Fipronil*<\/td><td class=\"has-text-align-center\" data-align=\"center\">4.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">436.8<\/td><td class=\"has-text-align-center\" data-align=\"center\">331.8<\/td><td class=\"has-text-align-center\" data-align=\"center\">251.9<\/td><\/tr><tr><td>Fenoxycarb<\/td><td class=\"has-text-align-center\" data-align=\"center\">4.2<\/td><td class=\"has-text-align-center\" data-align=\"center\">302.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">88.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">116.1<\/td><\/tr><tr><td>Kresoxim methyl<\/td><td class=\"has-text-align-center\" data-align=\"center\">4.4<\/td><td class=\"has-text-align-center\" data-align=\"center\">314.2<\/td><td class=\"has-text-align-center\" data-align=\"center\">267.2<\/td><td class=\"has-text-align-center\" data-align=\"center\">222.2<\/td><\/tr><tr><td>Tebuconazole<\/td><td class=\"has-text-align-center\" data-align=\"center\">4.4<\/td><td class=\"has-text-align-center\" data-align=\"center\">308.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">70.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">125.1<\/td><\/tr><tr><td>Diazinon-D10<\/td><td class=\"has-text-align-center\" data-align=\"center\">4.6<\/td><td class=\"has-text-align-center\" data-align=\"center\">315.2<\/td><td class=\"has-text-align-center\" data-align=\"center\">170.2<\/td><td class=\"has-text-align-center\" data-align=\"center\">&#8211;<\/td><\/tr><tr><td>Spinosad (spinosyn A)<\/td><td class=\"has-text-align-center\" data-align=\"center\">4.6<\/td><td class=\"has-text-align-center\" data-align=\"center\">732.4<\/td><td class=\"has-text-align-center\" data-align=\"center\">142.2<\/td><td class=\"has-text-align-center\" data-align=\"center\">98.1<\/td><\/tr><tr><td>Diazinon<\/td><td class=\"has-text-align-center\" data-align=\"center\">4.6<\/td><td class=\"has-text-align-center\" data-align=\"center\">305.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">169.2<\/td><td class=\"has-text-align-center\" data-align=\"center\">153.2<\/td><\/tr><tr><td>Coumaphos<\/td><td class=\"has-text-align-center\" data-align=\"center\">4.7<\/td><td class=\"has-text-align-center\" data-align=\"center\">363.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">227.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">307.1<\/td><\/tr><tr><td>Pyridaben<\/td><td class=\"has-text-align-center\" data-align=\"center\">4.7<\/td><td class=\"has-text-align-center\" data-align=\"center\">365.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">309.2<\/td><td class=\"has-text-align-center\" data-align=\"center\">147.2<\/td><\/tr><tr><td>Propiconazole<\/td><td class=\"has-text-align-center\" data-align=\"center\">4.7<\/td><td class=\"has-text-align-center\" data-align=\"center\">342.0<\/td><td class=\"has-text-align-center\" data-align=\"center\">159.0<\/td><td class=\"has-text-align-center\" data-align=\"center\">69.2<\/td><\/tr><tr><td>Clofentezine<\/td><td class=\"has-text-align-center\" data-align=\"center\">4.8<\/td><td class=\"has-text-align-center\" data-align=\"center\">303.0<\/td><td class=\"has-text-align-center\" data-align=\"center\">138.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">102.1<\/td><\/tr><tr><td>Spinosad (spinosyn D)<\/td><td class=\"has-text-align-center\" data-align=\"center\">5.0<\/td><td class=\"has-text-align-center\" data-align=\"center\">746.5<\/td><td class=\"has-text-align-center\" data-align=\"center\">142.3<\/td><td class=\"has-text-align-center\" data-align=\"center\">98.4<\/td><\/tr><tr><td>Spinetoram (spinosyn J)<\/td><td class=\"has-text-align-center\" data-align=\"center\">5.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">748.5<\/td><td class=\"has-text-align-center\" data-align=\"center\">142.3<\/td><td class=\"has-text-align-center\" data-align=\"center\">98.3<\/td><\/tr><tr><td>Trifloxystrobin<\/td><td class=\"has-text-align-center\" data-align=\"center\">5.3<\/td><td class=\"has-text-align-center\" data-align=\"center\">409.2<\/td><td class=\"has-text-align-center\" data-align=\"center\">186.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">145.1<\/td><\/tr><tr><td>Prallethrin<\/td><td class=\"has-text-align-center\" data-align=\"center\">5.3<\/td><td class=\"has-text-align-center\" data-align=\"center\">301.2<\/td><td class=\"has-text-align-center\" data-align=\"center\">123.2<\/td><td class=\"has-text-align-center\" data-align=\"center\">105.2<\/td><\/tr><tr><td>Pyrethrin II<\/td><td class=\"has-text-align-center\" data-align=\"center\">5.5<\/td><td class=\"has-text-align-center\" data-align=\"center\">373.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">161.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">133.2<\/td><\/tr><tr><td>Spinetoram (spinosyn L)<\/td><td class=\"has-text-align-center\" data-align=\"center\">5.6<\/td><td class=\"has-text-align-center\" data-align=\"center\">760.5<\/td><td class=\"has-text-align-center\" data-align=\"center\">142.2<\/td><td class=\"has-text-align-center\" data-align=\"center\">98.1<\/td><\/tr><tr><td>Piperonyl butoxide<\/td><td class=\"has-text-align-center\" data-align=\"center\">6.0<\/td><td class=\"has-text-align-center\" data-align=\"center\">356.3<\/td><td class=\"has-text-align-center\" data-align=\"center\">177.2<\/td><td class=\"has-text-align-center\" data-align=\"center\">119.2<\/td><\/tr><tr><td>Chlorpyrifos<\/td><td class=\"has-text-align-center\" data-align=\"center\">6.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">349.9<\/td><td class=\"has-text-align-center\" data-align=\"center\">198.0<\/td><td class=\"has-text-align-center\" data-align=\"center\">97.1<\/td><\/tr><tr><td>Hexythiazox<\/td><td class=\"has-text-align-center\" data-align=\"center\">6.2<\/td><td class=\"has-text-align-center\" data-align=\"center\">353.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">228.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">168.1<\/td><\/tr><tr><td>Etoxazole<\/td><td class=\"has-text-align-center\" data-align=\"center\">6.6<\/td><td class=\"has-text-align-center\" data-align=\"center\">360.2<\/td><td class=\"has-text-align-center\" data-align=\"center\">141.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">304.2<\/td><\/tr><tr><td>Spiromesifen<\/td><td class=\"has-text-align-center\" data-align=\"center\">6.7<\/td><td class=\"has-text-align-center\" data-align=\"center\">273.2<\/td><td class=\"has-text-align-center\" data-align=\"center\">255.2<\/td><td class=\"has-text-align-center\" data-align=\"center\">187.2<\/td><\/tr><tr><td>Pyrethrin I<\/td><td class=\"has-text-align-center\" data-align=\"center\">6.9<\/td><td class=\"has-text-align-center\" data-align=\"center\">329.2<\/td><td class=\"has-text-align-center\" data-align=\"center\">161.2<\/td><td class=\"has-text-align-center\" data-align=\"center\">105.2<\/td><\/tr><tr><td>Cyfluthrin (qualifier)<\/td><td class=\"has-text-align-center\" data-align=\"center\">6.9<\/td><td class=\"has-text-align-center\" data-align=\"center\">453.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">193.2<\/td><td class=\"has-text-align-center\" data-align=\"center\">&#8211;<\/td><\/tr><tr><td>Cyfluthrin<\/td><td class=\"has-text-align-center\" data-align=\"center\">6.9<\/td><td class=\"has-text-align-center\" data-align=\"center\">451.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">191.2<\/td><td class=\"has-text-align-center\" data-align=\"center\">&#8211;<\/td><\/tr><tr><td>Cypermethrin<\/td><td class=\"has-text-align-center\" data-align=\"center\">7.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">433.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">191.0<\/td><td class=\"has-text-align-center\" data-align=\"center\">416.0<\/td><\/tr><tr><td>(E)-Fenpyroximate<\/td><td class=\"has-text-align-center\" data-align=\"center\">7.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">422.2<\/td><td class=\"has-text-align-center\" data-align=\"center\">366.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">138.1<\/td><\/tr><tr><td>Permethrin-<em>trans<\/em><\/td><td class=\"has-text-align-center\" data-align=\"center\">7.6<\/td><td class=\"has-text-align-center\" data-align=\"center\">408.3<\/td><td class=\"has-text-align-center\" data-align=\"center\">183.2<\/td><td class=\"has-text-align-center\" data-align=\"center\">355.1<\/td><\/tr><tr><td>Permethrin-<em>cis<\/em><\/td><td class=\"has-text-align-center\" data-align=\"center\">7.9<\/td><td class=\"has-text-align-center\" data-align=\"center\">408.3<\/td><td class=\"has-text-align-center\" data-align=\"center\">183.2<\/td><td class=\"has-text-align-center\" data-align=\"center\">355.1<\/td><\/tr><tr><td>Avermectin B1a<\/td><td class=\"has-text-align-center\" data-align=\"center\">7.9<\/td><td class=\"has-text-align-center\" data-align=\"center\">890.5<\/td><td class=\"has-text-align-center\" data-align=\"center\">305.4<\/td><td class=\"has-text-align-center\" data-align=\"center\">567.4<\/td><\/tr><tr><td>Etofenprox<\/td><td class=\"has-text-align-center\" data-align=\"center\">8.0<\/td><td class=\"has-text-align-center\" data-align=\"center\">394.3<\/td><td class=\"has-text-align-center\" data-align=\"center\">177.2<\/td><td class=\"has-text-align-center\" data-align=\"center\">359.3<\/td><\/tr><tr><td>Bifenthrin<\/td><td class=\"has-text-align-center\" data-align=\"center\">8.2<\/td><td class=\"has-text-align-center\" data-align=\"center\">440.0<\/td><td class=\"has-text-align-center\" data-align=\"center\">181.2<\/td><td class=\"has-text-align-center\" data-align=\"center\">166.2<\/td><\/tr><tr><td>Acequinocyl precursor ion 1<\/td><td class=\"has-text-align-center\" data-align=\"center\">9.4<\/td><td class=\"has-text-align-center\" data-align=\"center\">402.3<\/td><td class=\"has-text-align-center\" data-align=\"center\">343.2<\/td><td class=\"has-text-align-center\" data-align=\"center\">189.0<\/td><\/tr><tr><td>Acequinocyl precursor ion 2<\/td><td class=\"has-text-align-center\" data-align=\"center\">9.4<\/td><td class=\"has-text-align-center\" data-align=\"center\">386.0<\/td><td class=\"has-text-align-center\" data-align=\"center\">344.2<\/td><td class=\"has-text-align-center\" data-align=\"center\">189.1<\/td><\/tr><\/tbody><\/table><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\">* Analizzati in modalit\u00e0 negativa.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\"><strong>Tabella VII:<\/strong>&nbsp;Transizioni GC-MS\/MS.<\/p>\n\n\n\n<figure class=\"wp-block-table\"><table class=\"has-fixed-layout\"><tbody><tr><td><strong>Nome<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>Tempo di ritenzione (min.)<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>Polarit\u00e0 ionica<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>Ione precursore<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>Ione prodotto<\/strong><\/td><\/tr><tr><td>Atrazine-D5 (IS) (Quan)<\/td><td class=\"has-text-align-center\" data-align=\"center\">6.82<\/td><td class=\"has-text-align-center\" data-align=\"center\">Positive<\/td><td class=\"has-text-align-center\" data-align=\"center\">220.0<\/td><td class=\"has-text-align-center\" data-align=\"center\">58.0<\/td><\/tr><tr><td>Atrazine-D5 (IS) (Qual)<\/td><td class=\"has-text-align-center\" data-align=\"center\">6.82<\/td><td class=\"has-text-align-center\" data-align=\"center\">Positive<\/td><td class=\"has-text-align-center\" data-align=\"center\">205.0<\/td><td class=\"has-text-align-center\" data-align=\"center\">127.0<\/td><\/tr><tr><td>Diazinon-D10 (Quan)<\/td><td class=\"has-text-align-center\" data-align=\"center\">7.01<\/td><td class=\"has-text-align-center\" data-align=\"center\">Positive<\/td><td class=\"has-text-align-center\" data-align=\"center\">183.0<\/td><td class=\"has-text-align-center\" data-align=\"center\">139.0<\/td><\/tr><tr><td>Diazinon-D10 (Qual)<\/td><td class=\"has-text-align-center\" data-align=\"center\">7.01<\/td><td class=\"has-text-align-center\" data-align=\"center\">Positive<\/td><td class=\"has-text-align-center\" data-align=\"center\">183.0<\/td><td class=\"has-text-align-center\" data-align=\"center\">168.0<\/td><\/tr><tr><td>Quintozene (PCNB) (Quan)<\/td><td class=\"has-text-align-center\" data-align=\"center\">7.03<\/td><td class=\"has-text-align-center\" data-align=\"center\">Positive<\/td><td class=\"has-text-align-center\" data-align=\"center\">294.9<\/td><td class=\"has-text-align-center\" data-align=\"center\">236.9<\/td><\/tr><tr><td>Quintozene (PCNB) (Qual)<\/td><td class=\"has-text-align-center\" data-align=\"center\">7.03<\/td><td class=\"has-text-align-center\" data-align=\"center\">Positive<\/td><td class=\"has-text-align-center\" data-align=\"center\">236.8<\/td><td class=\"has-text-align-center\" data-align=\"center\">118.9<\/td><\/tr><tr><td>Methyl parathion (Quan)<\/td><td class=\"has-text-align-center\" data-align=\"center\">7.50<\/td><td class=\"has-text-align-center\" data-align=\"center\">Positive<\/td><td class=\"has-text-align-center\" data-align=\"center\">263.0<\/td><td class=\"has-text-align-center\" data-align=\"center\">109.0<\/td><\/tr><tr><td>Methyl parathion (Qual)<\/td><td class=\"has-text-align-center\" data-align=\"center\">7.50<\/td><td class=\"has-text-align-center\" data-align=\"center\">Positive<\/td><td class=\"has-text-align-center\" data-align=\"center\">263.0<\/td><td class=\"has-text-align-center\" data-align=\"center\">79.0<\/td><\/tr><tr><td>Captan (Quan)<\/td><td class=\"has-text-align-center\" data-align=\"center\">8.37<\/td><td class=\"has-text-align-center\" data-align=\"center\">Positive<\/td><td class=\"has-text-align-center\" data-align=\"center\">184.0<\/td><td class=\"has-text-align-center\" data-align=\"center\">149.1<\/td><\/tr><tr><td>Captan (Qual)<\/td><td class=\"has-text-align-center\" data-align=\"center\">8.37<\/td><td class=\"has-text-align-center\" data-align=\"center\">Positive<\/td><td class=\"has-text-align-center\" data-align=\"center\">184.0<\/td><td class=\"has-text-align-center\" data-align=\"center\">134.1<\/td><\/tr><tr><td><em>trans<\/em>-Chlordane (Quan)<\/td><td class=\"has-text-align-center\" data-align=\"center\">8.41<\/td><td class=\"has-text-align-center\" data-align=\"center\">Positive<\/td><td class=\"has-text-align-center\" data-align=\"center\">271.9<\/td><td class=\"has-text-align-center\" data-align=\"center\">237.0<\/td><\/tr><tr><td><em>trans<\/em>-Chlordane (Qual)<\/td><td class=\"has-text-align-center\" data-align=\"center\">8.41<\/td><td class=\"has-text-align-center\" data-align=\"center\">Positive<\/td><td class=\"has-text-align-center\" data-align=\"center\">372.9<\/td><td class=\"has-text-align-center\" data-align=\"center\">265.9<\/td><\/tr><tr><td><em>cis<\/em>-Chlordane (Quan)<\/td><td class=\"has-text-align-center\" data-align=\"center\">8.53<\/td><td class=\"has-text-align-center\" data-align=\"center\">Positive<\/td><td class=\"has-text-align-center\" data-align=\"center\">372.9<\/td><td class=\"has-text-align-center\" data-align=\"center\">265.9<\/td><\/tr><tr><td><em>cis<\/em>-Chlordane (Qual)<\/td><td class=\"has-text-align-center\" data-align=\"center\">8.53<\/td><td class=\"has-text-align-center\" data-align=\"center\">Positive<\/td><td class=\"has-text-align-center\" data-align=\"center\">271.9<\/td><td class=\"has-text-align-center\" data-align=\"center\">237.0<\/td><\/tr><tr><td>Chlorfenapyr (Quan)<\/td><td class=\"has-text-align-center\" data-align=\"center\">8.80<\/td><td class=\"has-text-align-center\" data-align=\"center\">Positive<\/td><td class=\"has-text-align-center\" data-align=\"center\">247.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">227.1<\/td><\/tr><tr><td>Chlorfenapyr (Qual)<\/td><td class=\"has-text-align-center\" data-align=\"center\">8.80<\/td><td class=\"has-text-align-center\" data-align=\"center\">Positive<\/td><td class=\"has-text-align-center\" data-align=\"center\">59.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">31.1<\/td><\/tr><tr><td>Cyfluthrin (Quan)<\/td><td class=\"has-text-align-center\" data-align=\"center\">10.61<\/td><td class=\"has-text-align-center\" data-align=\"center\">Positive<\/td><td class=\"has-text-align-center\" data-align=\"center\">226.0<\/td><td class=\"has-text-align-center\" data-align=\"center\">206.0<\/td><\/tr><tr><td>Cyfluthrin (Qual)<\/td><td class=\"has-text-align-center\" data-align=\"center\">10.61<\/td><td class=\"has-text-align-center\" data-align=\"center\">Positive<\/td><td class=\"has-text-align-center\" data-align=\"center\">163.0<\/td><td class=\"has-text-align-center\" data-align=\"center\">127.0<\/td><\/tr><tr><td>Cypermethrin (Quan)<\/td><td class=\"has-text-align-center\" data-align=\"center\">10.87<\/td><td class=\"has-text-align-center\" data-align=\"center\">Positive<\/td><td class=\"has-text-align-center\" data-align=\"center\">163.0<\/td><td class=\"has-text-align-center\" data-align=\"center\">127.1<\/td><\/tr><tr><td>Cypermethrin (Qual)<\/td><td class=\"has-text-align-center\" data-align=\"center\">10.87<\/td><td class=\"has-text-align-center\" data-align=\"center\">Positive<\/td><td class=\"has-text-align-center\" data-align=\"center\">181.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">152.1<\/td><\/tr><\/tbody><\/table><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\"><strong>Tabella VIII:<\/strong>&nbsp;Tempi di ritenzione dei cannabinoidi..<\/p>\n\n\n\n<figure class=\"wp-block-table\"><table class=\"has-fixed-layout\"><tbody><tr><td><strong>Composto<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>Tempo di ritenzione (min.)<\/strong><\/td><\/tr><tr><td>Cannabidiolic acid (CBDA)<\/td><td class=\"has-text-align-center\" data-align=\"center\">2.142<\/td><\/tr><tr><td>Cannabigerol (CBG)<\/td><td class=\"has-text-align-center\" data-align=\"center\">2.405<\/td><\/tr><tr><td>Cannabidiol (CBD)<\/td><td class=\"has-text-align-center\" data-align=\"center\">2.535<\/td><\/tr><tr><td>Cannabinol (CBN)<\/td><td class=\"has-text-align-center\" data-align=\"center\">3.776<\/td><\/tr><tr><td>Delta-9-tetrahydrocannabinol (Delta-9-THC)<\/td><td class=\"has-text-align-center\" data-align=\"center\">4.753<\/td><\/tr><tr><td>Tetrahydrocannabinolic acid (THCA)<\/td><td class=\"has-text-align-center\" data-align=\"center\">6.279<\/td><\/tr><\/tbody><\/table><\/figure>\n\n\n\n<h2 class=\"wp-block-heading\">Risultati e discussione<\/h2>\n\n\n\n<h3 class=\"wp-block-heading\">Ottimizzazione del metodo<\/h3>\n\n\n\n<p class=\"wp-block-paragraph\">La variet\u00e0 delle matrici di cannabis e le caratteristiche chimiche differenti degli analiti target impongono l\u2019adozione di strategie diversificate per ottenere risultati accurati. L\u2019analisi di pesticidi, micotossine e cannabinoidi nelle caramelle gommose alla cannabis, come previsto dallo stato della California, richiede condizioni sperimentali totalmente diverse rispetto a quelle adottate per i brownie nel nostro precedente articolo tecnico [4]. Innanzitutto, abbiamo constatato che trattare le caramelle gommose polverizzate (tritate con ghiaccio secco e un robot da cucina) pu\u00f2 sollevare problematiche complesse perch\u00e9 la matrice, una volta a temperatura ambiente, diventa molto appiccicosa. Per questo motivo si \u00e8 rivelata un\u2019alternativa decisamente migliore tagliare le caramelle gommose in piccoli pezzi, in modo da riuscire a manipolarle pi\u00f9 facilmente quando si pesano le quantit\u00e0 necessarie di campione.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">La sfida successiva \u00e8 stata quella di individuare l\u2019approccio migliore per dissolvere la matrice prima dell\u2019estrazione e ottenere cos\u00ec risultati affidabili. Poich\u00e9 il campione \u00e8 stato tritato in piccoli pezzi di dimensione casuale, avere una matrice omogenea \u00e8 stato fondamentale per garantire una riproducibilit\u00e0 del metodo soddisfacente. In primo luogo, abbiamo valutato l\u2019utilizzo di solventi come acetonitrile e metanolo per dissolvere i pezzi delle caramelle gommose ed estrarre gli analiti di interesse tramite una semplice estrazione di solvente. Tuttavia, abbiamo riscontrato grandi difficolt\u00e0 nel solubilizzare la matrice in queste condizioni, quindi abbiamo valutato anche l\u2019utilizzo di DMSO. Nonostante tutti i pezzi di caramelle gommose siano stati dissolti facilmente in questo solvente, il daminozide non \u00e8 stato rilevato in nessuno degli estratti perch\u00e9 la piccola percentuale di DMSO rimasta nei campioni iniettati ha portato alla soppressione della ionizzazione. Dopo questi test, abbiamo concluso che il modo migliore per ottenere un campione omogeneo fosse idratare il campione con 5 mL di acqua e poi effettuare un vigoroso passaggio al vortex. Dopodich\u00e9, per estrarre tutti i contaminanti dalla matrice sono stati aggiunti al campione dissolto 5 mL di acetonitrile acidificato a 1% con acido acetico.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Ai fini della separazione dello strato organico da quello acquoso sono stati confrontati tre differenti sali di estrazione Q-sep QuEChERS: AOAC (cat.# 25851), senza tampone (cat.# 25847) ed EN (cat.# 25849). I sali EN hanno ottenuto le prestazioni migliori, mostrando recuperi superiori all\u201983% per tutti i composti a esclusione del daminozide e del suo analogo deuterato, che invece hanno raggiunto un recupero del 25%. Per quanto riguarda la fase di purificazione, \u00e8 stato valutato l\u2019effetto di quattro diverse miscele di sorbenti dSPE Q-sep QuEChERS (cat.#&nbsp;26215,&nbsp;26216,&nbsp;26217,&nbsp;26242) in tutti i pesticidi analizzabili in LC-MS. \u00c8 stato confermato che tutte le miscele contenenti 25&nbsp;mg di PSA (cat.# &nbsp;26215,&nbsp;26216,&nbsp;26217) hanno provocato perdite significative di daminozide e di ocratossina A, con un recupero pari solo al 40% del quantitativo iniziale presente nella soluzione. La miscela dSPE contenente MgSO4 e C18 (cat.#&nbsp;26242) non ha causato perdite significative di pesticidi, ma abbiamo deciso di valutare l\u2019idoneit\u00e0, per la quantificazione di analiti compatibili con la LC, dell\u2019utilizzo dell\u2019estratto organico senza alcuna ulteriore fase di purificazione.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Per valutare la fattibilit\u00e0 di un\u2019analisi diretta degli estratti sono stati condotti esperimenti volti a studiare gli effetti assoluti della matrice negli estratti di caramelle senza alcuna fase di purificazione, utilizzando la metodologia proposta da Matuszweski et al. [5]. Gli estratti ottenuti dai campioni bianchi sono stati fortificati alle concentrazioni finali di 5, 15 e 50 ppb e tramite LC-MS\/MS sono state confrontate le loro risposte con quelle del solvente puro, fortificato agli stessi livelli di concentrazione. A 5 ppb, 12 pesticidi hanno mostrato effetti di matrice superiori al 120%, mentre a 15 ppb e 50 ppb solo il daminozide \u00e8 risultato potenziato in modo significativo (Tabella IX). Considerando questo e lo scarso recupero di daminozide, utilizzare il daminozide-d6 come standard interno \u00e8 stato fondamentale per riuscire a ottenere dati affidabili.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Per quanto riguarda i recuperi dei pesticidi analizzabili in GC, i dati relativi alla valutazione di tre miscele di sorbenti dSPE (cat.#&nbsp;26215,&nbsp;26216, e&nbsp;26217) hanno mostrato che, mettendo a confronto le risposte dell\u2019estratto purificato e quelle dell\u2019estratto originale, in tutti i casi i recuperi sono stati superiori al 96%. Dato l\u2019alto contenuto di zucchero e pigmenti negli estratti di caramelle gommose, per la procedura finale per la preparazione del campione \u00e8 stato scelto il sorbente dSPE contenente PSA, GCB e solfato di magnesio (cat.#&nbsp;26217).<\/p>\n\n\n\n<h3 class=\"wp-block-heading\">Verifica del metodo<\/h3>\n\n\n\n<p class=\"wp-block-paragraph\">La Tabella X mostra i risultati quanto a limiti di quantificazione, linearit\u00e0, accuratezza e precisione relativi a pesticidi e micotossine presenti nella lista stilata dal governo della California, determinati nella matrice di caramelle gommose. Per tutti i contaminanti analizzati in LC-MS\/MS, le curve di calibrazione sono state tracciate utilizzando i rapporti delle risposte degli analiti e degli standard interni e un fattore di ponderazione di 1\/x. In riferimento agli analiti analizzabili in GC, solo la curva di calibrazione del PCNB \u00e8 stata elaborata con il rapporto analita\/standard interno, avendo come standard interno il diazinone-d10. La quantificazione per i restanti composti compatibili con la GC \u00e8 stata effettuata con curve di calibrazione esterne (area vs. concentrazione fortificata) perch\u00e9 in questo modo si sono ottenuti risultati migliori rispetto al procedimento con gli standard interni. Sono stati ottenuti valori di RSD inferiori al 24% per tutti gli analiti e a tutti i livelli di concentrazione testati. I valori di accuratezza riscontrati sono stati fra il 75 e il 118%, mentre i coefficienti di determinazione (R2) erano tutti superiori a 0,99.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Infine, i risultati delle analisi dei cannabinoidi hanno dimostrato che l\u2019estratto ottenuto per la determinazione dei contaminanti \u00e8 adatto anche per l\u2019analisi di potenza. La Tabella XI mostra i dati relativi alle curve di calibrazione preparate in solvente per la quantificazione di ciascun cannabinoide. Come illustrato nella Tabella XII, i campioni di caramelle gommose fortificati con sei cannabinoidi a 0,2\u202fmg\/g hanno mostrato recuperi fra il 99% e il 107%, mentre i campioni fortificati a 0,5 mg\/g hanno avuto recuperi fra il 99% e il 106%; le Figure da 2 a 4 riportano i rispettivi cromatogrammi.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\"><strong>Tabella IX:<\/strong>&nbsp;Effetti assoluti della matrice (ME) per pesticidi e micotossine nelle caramelle gommose alla cannabis.<\/p>\n\n\n\n<figure class=\"wp-block-table\"><table class=\"has-fixed-layout\"><tbody><tr><td>&nbsp;<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>ME at 5 ppb (%)<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>RSD<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>ME at 15 ppb (%)<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>RSD<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>ME at 50 ppb (%)<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>RSD<\/strong><\/td><\/tr><tr><td>Daminozide<\/td><td class=\"has-text-align-center\" data-align=\"center\">216<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><td class=\"has-text-align-center\" data-align=\"center\">251<\/td><td class=\"has-text-align-center\" data-align=\"center\">7<\/td><td class=\"has-text-align-center\" data-align=\"center\">185<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><\/tr><tr><td>Acephate<\/td><td class=\"has-text-align-center\" data-align=\"center\">95<\/td><td class=\"has-text-align-center\" data-align=\"center\">6<\/td><td class=\"has-text-align-center\" data-align=\"center\">85<\/td><td class=\"has-text-align-center\" data-align=\"center\">1<\/td><td class=\"has-text-align-center\" data-align=\"center\">89<\/td><td class=\"has-text-align-center\" data-align=\"center\">6<\/td><\/tr><tr><td>Oxamyl<\/td><td class=\"has-text-align-center\" data-align=\"center\">105<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><td class=\"has-text-align-center\" data-align=\"center\">96<\/td><td class=\"has-text-align-center\" data-align=\"center\">2<\/td><td class=\"has-text-align-center\" data-align=\"center\">98<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><\/tr><tr><td>Flonicamid<\/td><td class=\"has-text-align-center\" data-align=\"center\">91<\/td><td class=\"has-text-align-center\" data-align=\"center\">32<\/td><td class=\"has-text-align-center\" data-align=\"center\">90<\/td><td class=\"has-text-align-center\" data-align=\"center\">22<\/td><td class=\"has-text-align-center\" data-align=\"center\">97<\/td><td class=\"has-text-align-center\" data-align=\"center\">17<\/td><\/tr><tr><td>Methomyl<\/td><td class=\"has-text-align-center\" data-align=\"center\">104<\/td><td class=\"has-text-align-center\" data-align=\"center\">4<\/td><td class=\"has-text-align-center\" data-align=\"center\">92<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><td class=\"has-text-align-center\" data-align=\"center\">99<\/td><td class=\"has-text-align-center\" data-align=\"center\">4<\/td><\/tr><tr><td>Thiamethoxam<\/td><td class=\"has-text-align-center\" data-align=\"center\">105<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><td class=\"has-text-align-center\" data-align=\"center\">92<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><td class=\"has-text-align-center\" data-align=\"center\">96<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><\/tr><tr><td>Imidacloprid<\/td><td class=\"has-text-align-center\" data-align=\"center\">114<\/td><td class=\"has-text-align-center\" data-align=\"center\">9<\/td><td class=\"has-text-align-center\" data-align=\"center\">87<\/td><td class=\"has-text-align-center\" data-align=\"center\">9<\/td><td class=\"has-text-align-center\" data-align=\"center\">100<\/td><td class=\"has-text-align-center\" data-align=\"center\">9<\/td><\/tr><tr><td>Mevinphos<\/td><td class=\"has-text-align-center\" data-align=\"center\">102<\/td><td class=\"has-text-align-center\" data-align=\"center\">6<\/td><td class=\"has-text-align-center\" data-align=\"center\">93<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><td class=\"has-text-align-center\" data-align=\"center\">93<\/td><td class=\"has-text-align-center\" data-align=\"center\">6<\/td><\/tr><tr><td>Acetamiprid<\/td><td class=\"has-text-align-center\" data-align=\"center\">100<\/td><td class=\"has-text-align-center\" data-align=\"center\">2<\/td><td class=\"has-text-align-center\" data-align=\"center\">88<\/td><td class=\"has-text-align-center\" data-align=\"center\">0<\/td><td class=\"has-text-align-center\" data-align=\"center\">91<\/td><td class=\"has-text-align-center\" data-align=\"center\">2<\/td><\/tr><tr><td>Dimethoate<\/td><td class=\"has-text-align-center\" data-align=\"center\">97<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><td class=\"has-text-align-center\" data-align=\"center\">92<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><td class=\"has-text-align-center\" data-align=\"center\">93<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><\/tr><tr><td>Thiacloprid<\/td><td class=\"has-text-align-center\" data-align=\"center\">109<\/td><td class=\"has-text-align-center\" data-align=\"center\">7<\/td><td class=\"has-text-align-center\" data-align=\"center\">93<\/td><td class=\"has-text-align-center\" data-align=\"center\">1<\/td><td class=\"has-text-align-center\" data-align=\"center\">93<\/td><td class=\"has-text-align-center\" data-align=\"center\">7<\/td><\/tr><tr><td>Aflatoxin G2<\/td><td class=\"has-text-align-center\" data-align=\"center\">102<\/td><td class=\"has-text-align-center\" data-align=\"center\">6<\/td><td class=\"has-text-align-center\" data-align=\"center\">86<\/td><td class=\"has-text-align-center\" data-align=\"center\">0<\/td><td class=\"has-text-align-center\" data-align=\"center\">93<\/td><td class=\"has-text-align-center\" data-align=\"center\">6<\/td><\/tr><tr><td>Aflatoxin G1<\/td><td class=\"has-text-align-center\" data-align=\"center\">106<\/td><td class=\"has-text-align-center\" data-align=\"center\">7<\/td><td class=\"has-text-align-center\" data-align=\"center\">93<\/td><td class=\"has-text-align-center\" data-align=\"center\">1<\/td><td class=\"has-text-align-center\" data-align=\"center\">91<\/td><td class=\"has-text-align-center\" data-align=\"center\">7<\/td><\/tr><tr><td>Aldicarb<\/td><td class=\"has-text-align-center\" data-align=\"center\">81<\/td><td class=\"has-text-align-center\" data-align=\"center\">22<\/td><td class=\"has-text-align-center\" data-align=\"center\">85<\/td><td class=\"has-text-align-center\" data-align=\"center\">14<\/td><td class=\"has-text-align-center\" data-align=\"center\">90<\/td><td class=\"has-text-align-center\" data-align=\"center\">22<\/td><\/tr><tr><td>Aflatoxin B2<\/td><td class=\"has-text-align-center\" data-align=\"center\">113<\/td><td class=\"has-text-align-center\" data-align=\"center\">11<\/td><td class=\"has-text-align-center\" data-align=\"center\">79<\/td><td class=\"has-text-align-center\" data-align=\"center\">9<\/td><td class=\"has-text-align-center\" data-align=\"center\">95<\/td><td class=\"has-text-align-center\" data-align=\"center\">11<\/td><\/tr><tr><td>Dichlorvos<\/td><td class=\"has-text-align-center\" data-align=\"center\">126<\/td><td class=\"has-text-align-center\" data-align=\"center\">15<\/td><td class=\"has-text-align-center\" data-align=\"center\">90<\/td><td class=\"has-text-align-center\" data-align=\"center\">6<\/td><td class=\"has-text-align-center\" data-align=\"center\">92<\/td><td class=\"has-text-align-center\" data-align=\"center\">15<\/td><\/tr><tr><td>Aflatoxin B1<\/td><td class=\"has-text-align-center\" data-align=\"center\">104<\/td><td class=\"has-text-align-center\" data-align=\"center\">9<\/td><td class=\"has-text-align-center\" data-align=\"center\">93<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><td class=\"has-text-align-center\" data-align=\"center\">95<\/td><td class=\"has-text-align-center\" data-align=\"center\">9<\/td><\/tr><tr><td>Imazalil<\/td><td class=\"has-text-align-center\" data-align=\"center\">92<\/td><td class=\"has-text-align-center\" data-align=\"center\">1<\/td><td class=\"has-text-align-center\" data-align=\"center\">96<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><td class=\"has-text-align-center\" data-align=\"center\">98<\/td><td class=\"has-text-align-center\" data-align=\"center\">1<\/td><\/tr><tr><td>Carbofuran<\/td><td class=\"has-text-align-center\" data-align=\"center\">106<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><td class=\"has-text-align-center\" data-align=\"center\">98<\/td><td class=\"has-text-align-center\" data-align=\"center\">1<\/td><td class=\"has-text-align-center\" data-align=\"center\">102<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><\/tr><tr><td>Propoxur<\/td><td class=\"has-text-align-center\" data-align=\"center\">104<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><td class=\"has-text-align-center\" data-align=\"center\">96<\/td><td class=\"has-text-align-center\" data-align=\"center\">4<\/td><td class=\"has-text-align-center\" data-align=\"center\">96<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><\/tr><tr><td>Carbaryl<\/td><td class=\"has-text-align-center\" data-align=\"center\">103<\/td><td class=\"has-text-align-center\" data-align=\"center\">4<\/td><td class=\"has-text-align-center\" data-align=\"center\">99<\/td><td class=\"has-text-align-center\" data-align=\"center\">8<\/td><td class=\"has-text-align-center\" data-align=\"center\">93<\/td><td class=\"has-text-align-center\" data-align=\"center\">4<\/td><\/tr><tr><td>Naled<\/td><td class=\"has-text-align-center\" data-align=\"center\">101<\/td><td class=\"has-text-align-center\" data-align=\"center\">4<\/td><td class=\"has-text-align-center\" data-align=\"center\">94<\/td><td class=\"has-text-align-center\" data-align=\"center\">2<\/td><td class=\"has-text-align-center\" data-align=\"center\">84<\/td><td class=\"has-text-align-center\" data-align=\"center\">4<\/td><\/tr><tr><td>Metalaxyl<\/td><td class=\"has-text-align-center\" data-align=\"center\">105<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><td class=\"has-text-align-center\" data-align=\"center\">97<\/td><td class=\"has-text-align-center\" data-align=\"center\">2<\/td><td class=\"has-text-align-center\" data-align=\"center\">97<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><\/tr><tr><td>Spiroxamine<\/td><td class=\"has-text-align-center\" data-align=\"center\">105<\/td><td class=\"has-text-align-center\" data-align=\"center\">4<\/td><td class=\"has-text-align-center\" data-align=\"center\">94<\/td><td class=\"has-text-align-center\" data-align=\"center\">1<\/td><td class=\"has-text-align-center\" data-align=\"center\">97<\/td><td class=\"has-text-align-center\" data-align=\"center\">4<\/td><\/tr><tr><td>Chlorantraniliprole<\/td><td class=\"has-text-align-center\" data-align=\"center\">114<\/td><td class=\"has-text-align-center\" data-align=\"center\">6<\/td><td class=\"has-text-align-center\" data-align=\"center\">88<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><td class=\"has-text-align-center\" data-align=\"center\">100<\/td><td class=\"has-text-align-center\" data-align=\"center\">6<\/td><\/tr><tr><td>Phosmet<\/td><td class=\"has-text-align-center\" data-align=\"center\">108<\/td><td class=\"has-text-align-center\" data-align=\"center\">10<\/td><td class=\"has-text-align-center\" data-align=\"center\">100<\/td><td class=\"has-text-align-center\" data-align=\"center\">6<\/td><td class=\"has-text-align-center\" data-align=\"center\">92<\/td><td class=\"has-text-align-center\" data-align=\"center\">10<\/td><\/tr><tr><td>Azoxystrobin<\/td><td class=\"has-text-align-center\" data-align=\"center\">104<\/td><td class=\"has-text-align-center\" data-align=\"center\">1<\/td><td class=\"has-text-align-center\" data-align=\"center\">94<\/td><td class=\"has-text-align-center\" data-align=\"center\">1<\/td><td class=\"has-text-align-center\" data-align=\"center\">94<\/td><td class=\"has-text-align-center\" data-align=\"center\">1<\/td><\/tr><tr><td>Fludioxonil<\/td><td class=\"has-text-align-center\" data-align=\"center\">101<\/td><td class=\"has-text-align-center\" data-align=\"center\">7<\/td><td class=\"has-text-align-center\" data-align=\"center\">90<\/td><td class=\"has-text-align-center\" data-align=\"center\">11<\/td><td class=\"has-text-align-center\" data-align=\"center\">94<\/td><td class=\"has-text-align-center\" data-align=\"center\">7<\/td><\/tr><tr><td>Methiocarb<\/td><td class=\"has-text-align-center\" data-align=\"center\">102<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><td class=\"has-text-align-center\" data-align=\"center\">94<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><td class=\"has-text-align-center\" data-align=\"center\">95<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><\/tr><tr><td>Dimethomorph<\/td><td class=\"has-text-align-center\" data-align=\"center\">108<\/td><td class=\"has-text-align-center\" data-align=\"center\">15<\/td><td class=\"has-text-align-center\" data-align=\"center\">93<\/td><td class=\"has-text-align-center\" data-align=\"center\">0<\/td><td class=\"has-text-align-center\" data-align=\"center\">93<\/td><td class=\"has-text-align-center\" data-align=\"center\">15<\/td><\/tr><tr><td>Boscalid<\/td><td class=\"has-text-align-center\" data-align=\"center\">119<\/td><td class=\"has-text-align-center\" data-align=\"center\">11<\/td><td class=\"has-text-align-center\" data-align=\"center\">94<\/td><td class=\"has-text-align-center\" data-align=\"center\">4<\/td><td class=\"has-text-align-center\" data-align=\"center\">85<\/td><td class=\"has-text-align-center\" data-align=\"center\">11<\/td><\/tr><tr><td>Paclobutrazol<\/td><td class=\"has-text-align-center\" data-align=\"center\">110<\/td><td class=\"has-text-align-center\" data-align=\"center\">10<\/td><td class=\"has-text-align-center\" data-align=\"center\">90<\/td><td class=\"has-text-align-center\" data-align=\"center\">2<\/td><td class=\"has-text-align-center\" data-align=\"center\">92<\/td><td class=\"has-text-align-center\" data-align=\"center\">10<\/td><\/tr><tr><td>Malathion<\/td><td class=\"has-text-align-center\" data-align=\"center\">101<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><td class=\"has-text-align-center\" data-align=\"center\">92<\/td><td class=\"has-text-align-center\" data-align=\"center\">4<\/td><td class=\"has-text-align-center\" data-align=\"center\">95<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><\/tr><tr><td>Myclobutanil<\/td><td class=\"has-text-align-center\" data-align=\"center\">92<\/td><td class=\"has-text-align-center\" data-align=\"center\">11<\/td><td class=\"has-text-align-center\" data-align=\"center\">93<\/td><td class=\"has-text-align-center\" data-align=\"center\">0<\/td><td class=\"has-text-align-center\" data-align=\"center\">95<\/td><td class=\"has-text-align-center\" data-align=\"center\">11<\/td><\/tr><tr><td>Bifenazate<\/td><td class=\"has-text-align-center\" data-align=\"center\">105<\/td><td class=\"has-text-align-center\" data-align=\"center\">2<\/td><td class=\"has-text-align-center\" data-align=\"center\">98<\/td><td class=\"has-text-align-center\" data-align=\"center\">4<\/td><td class=\"has-text-align-center\" data-align=\"center\">99<\/td><td class=\"has-text-align-center\" data-align=\"center\">2<\/td><\/tr><tr><td>Ochratoxin A<\/td><td class=\"has-text-align-center\" data-align=\"center\">111<\/td><td class=\"has-text-align-center\" data-align=\"center\">10<\/td><td class=\"has-text-align-center\" data-align=\"center\">111<\/td><td class=\"has-text-align-center\" data-align=\"center\">0<\/td><td class=\"has-text-align-center\" data-align=\"center\">98<\/td><td class=\"has-text-align-center\" data-align=\"center\">10<\/td><\/tr><tr><td>Fenhexamid<\/td><td class=\"has-text-align-center\" data-align=\"center\">135<\/td><td class=\"has-text-align-center\" data-align=\"center\">6<\/td><td class=\"has-text-align-center\" data-align=\"center\">102<\/td><td class=\"has-text-align-center\" data-align=\"center\">0<\/td><td class=\"has-text-align-center\" data-align=\"center\">91<\/td><td class=\"has-text-align-center\" data-align=\"center\">6<\/td><\/tr><tr><td>Spirotetramat<\/td><td class=\"has-text-align-center\" data-align=\"center\">107<\/td><td class=\"has-text-align-center\" data-align=\"center\">7<\/td><td class=\"has-text-align-center\" data-align=\"center\">87<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><td class=\"has-text-align-center\" data-align=\"center\">95<\/td><td class=\"has-text-align-center\" data-align=\"center\">7<\/td><\/tr><tr><td>Ethoprophos<\/td><td class=\"has-text-align-center\" data-align=\"center\">104<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><td class=\"has-text-align-center\" data-align=\"center\">97<\/td><td class=\"has-text-align-center\" data-align=\"center\">2<\/td><td class=\"has-text-align-center\" data-align=\"center\">96<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><\/tr><tr><td>Fipronil<\/td><td class=\"has-text-align-center\" data-align=\"center\">99<\/td><td class=\"has-text-align-center\" data-align=\"center\">12<\/td><td class=\"has-text-align-center\" data-align=\"center\">93<\/td><td class=\"has-text-align-center\" data-align=\"center\">9<\/td><td class=\"has-text-align-center\" data-align=\"center\">96<\/td><td class=\"has-text-align-center\" data-align=\"center\">12<\/td><\/tr><tr><td>Fenoxycarb<\/td><td class=\"has-text-align-center\" data-align=\"center\">108<\/td><td class=\"has-text-align-center\" data-align=\"center\">9<\/td><td class=\"has-text-align-center\" data-align=\"center\">97<\/td><td class=\"has-text-align-center\" data-align=\"center\">1<\/td><td class=\"has-text-align-center\" data-align=\"center\">93<\/td><td class=\"has-text-align-center\" data-align=\"center\">9<\/td><\/tr><tr><td>Kresoxim methyl<\/td><td class=\"has-text-align-center\" data-align=\"center\">93<\/td><td class=\"has-text-align-center\" data-align=\"center\">21<\/td><td class=\"has-text-align-center\" data-align=\"center\">108<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><td class=\"has-text-align-center\" data-align=\"center\">90<\/td><td class=\"has-text-align-center\" data-align=\"center\">21<\/td><\/tr><tr><td>Tebuconazole<\/td><td class=\"has-text-align-center\" data-align=\"center\">102<\/td><td class=\"has-text-align-center\" data-align=\"center\">6<\/td><td class=\"has-text-align-center\" data-align=\"center\">95<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><td class=\"has-text-align-center\" data-align=\"center\">97<\/td><td class=\"has-text-align-center\" data-align=\"center\">6<\/td><\/tr><tr><td>Spinosyn A<\/td><td class=\"has-text-align-center\" data-align=\"center\">100<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><td class=\"has-text-align-center\" data-align=\"center\">98<\/td><td class=\"has-text-align-center\" data-align=\"center\">2<\/td><td class=\"has-text-align-center\" data-align=\"center\">95<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><\/tr><tr><td>Diazinon<\/td><td class=\"has-text-align-center\" data-align=\"center\">107<\/td><td class=\"has-text-align-center\" data-align=\"center\">2<\/td><td class=\"has-text-align-center\" data-align=\"center\">95<\/td><td class=\"has-text-align-center\" data-align=\"center\">1<\/td><td class=\"has-text-align-center\" data-align=\"center\">96<\/td><td class=\"has-text-align-center\" data-align=\"center\">2<\/td><\/tr><tr><td>Coumaphos<\/td><td class=\"has-text-align-center\" data-align=\"center\">114<\/td><td class=\"has-text-align-center\" data-align=\"center\">7<\/td><td class=\"has-text-align-center\" data-align=\"center\">94<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><td class=\"has-text-align-center\" data-align=\"center\">97<\/td><td class=\"has-text-align-center\" data-align=\"center\">7<\/td><\/tr><tr><td>Pyridaben<\/td><td class=\"has-text-align-center\" data-align=\"center\">139<\/td><td class=\"has-text-align-center\" data-align=\"center\">18<\/td><td class=\"has-text-align-center\" data-align=\"center\">103<\/td><td class=\"has-text-align-center\" data-align=\"center\">15<\/td><td class=\"has-text-align-center\" data-align=\"center\">94<\/td><td class=\"has-text-align-center\" data-align=\"center\">18<\/td><\/tr><tr><td>Propiconazole<\/td><td class=\"has-text-align-center\" data-align=\"center\">106<\/td><td class=\"has-text-align-center\" data-align=\"center\">2<\/td><td class=\"has-text-align-center\" data-align=\"center\">97<\/td><td class=\"has-text-align-center\" data-align=\"center\">1<\/td><td class=\"has-text-align-center\" data-align=\"center\">95<\/td><td class=\"has-text-align-center\" data-align=\"center\">2<\/td><\/tr><tr><td>Clofentezine<\/td><td class=\"has-text-align-center\" data-align=\"center\">126<\/td><td class=\"has-text-align-center\" data-align=\"center\">14<\/td><td class=\"has-text-align-center\" data-align=\"center\">87<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><td class=\"has-text-align-center\" data-align=\"center\">92<\/td><td class=\"has-text-align-center\" data-align=\"center\">14<\/td><\/tr><tr><td>Spinosyn D<\/td><td class=\"has-text-align-center\" data-align=\"center\">104<\/td><td class=\"has-text-align-center\" data-align=\"center\">8<\/td><td class=\"has-text-align-center\" data-align=\"center\">93<\/td><td class=\"has-text-align-center\" data-align=\"center\">8<\/td><td class=\"has-text-align-center\" data-align=\"center\">97<\/td><td class=\"has-text-align-center\" data-align=\"center\">8<\/td><\/tr><tr><td>Spinosyn J<\/td><td class=\"has-text-align-center\" data-align=\"center\">110<\/td><td class=\"has-text-align-center\" data-align=\"center\">6<\/td><td class=\"has-text-align-center\" data-align=\"center\">92<\/td><td class=\"has-text-align-center\" data-align=\"center\">6<\/td><td class=\"has-text-align-center\" data-align=\"center\">101<\/td><td class=\"has-text-align-center\" data-align=\"center\">6<\/td><\/tr><tr><td>Trifloxystrobin<\/td><td class=\"has-text-align-center\" data-align=\"center\">103<\/td><td class=\"has-text-align-center\" data-align=\"center\">2<\/td><td class=\"has-text-align-center\" data-align=\"center\">95<\/td><td class=\"has-text-align-center\" data-align=\"center\">2<\/td><td class=\"has-text-align-center\" data-align=\"center\">101<\/td><td class=\"has-text-align-center\" data-align=\"center\">2<\/td><\/tr><tr><td>Prallethrin<\/td><td class=\"has-text-align-center\" data-align=\"center\">101<\/td><td class=\"has-text-align-center\" data-align=\"center\">17<\/td><td class=\"has-text-align-center\" data-align=\"center\">103<\/td><td class=\"has-text-align-center\" data-align=\"center\">0<\/td><td class=\"has-text-align-center\" data-align=\"center\">98<\/td><td class=\"has-text-align-center\" data-align=\"center\">17<\/td><\/tr><tr><td>Pyrethrin II<\/td><td class=\"has-text-align-center\" data-align=\"center\">92<\/td><td class=\"has-text-align-center\" data-align=\"center\">23<\/td><td class=\"has-text-align-center\" data-align=\"center\">82<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><td class=\"has-text-align-center\" data-align=\"center\">101<\/td><td class=\"has-text-align-center\" data-align=\"center\">23<\/td><\/tr><tr><td>Spinosyn L<\/td><td class=\"has-text-align-center\" data-align=\"center\">106<\/td><td class=\"has-text-align-center\" data-align=\"center\">6<\/td><td class=\"has-text-align-center\" data-align=\"center\">95<\/td><td class=\"has-text-align-center\" data-align=\"center\">0<\/td><td class=\"has-text-align-center\" data-align=\"center\">98<\/td><td class=\"has-text-align-center\" data-align=\"center\">6<\/td><\/tr><tr><td>Piperonyl butoxide<\/td><td class=\"has-text-align-center\" data-align=\"center\">103<\/td><td class=\"has-text-align-center\" data-align=\"center\">2<\/td><td class=\"has-text-align-center\" data-align=\"center\">97<\/td><td class=\"has-text-align-center\" data-align=\"center\">0<\/td><td class=\"has-text-align-center\" data-align=\"center\">96<\/td><td class=\"has-text-align-center\" data-align=\"center\">2<\/td><\/tr><tr><td>Chlorpyrifos<\/td><td class=\"has-text-align-center\" data-align=\"center\">105<\/td><td class=\"has-text-align-center\" data-align=\"center\">10<\/td><td class=\"has-text-align-center\" data-align=\"center\">93<\/td><td class=\"has-text-align-center\" data-align=\"center\">0<\/td><td class=\"has-text-align-center\" data-align=\"center\">91<\/td><td class=\"has-text-align-center\" data-align=\"center\">10<\/td><\/tr><tr><td>Hexythiazox<\/td><td class=\"has-text-align-center\" data-align=\"center\">121<\/td><td class=\"has-text-align-center\" data-align=\"center\">15<\/td><td class=\"has-text-align-center\" data-align=\"center\">96<\/td><td class=\"has-text-align-center\" data-align=\"center\">0<\/td><td class=\"has-text-align-center\" data-align=\"center\">86<\/td><td class=\"has-text-align-center\" data-align=\"center\">15<\/td><\/tr><tr><td>Etoxazole<\/td><td class=\"has-text-align-center\" data-align=\"center\">102<\/td><td class=\"has-text-align-center\" data-align=\"center\">1<\/td><td class=\"has-text-align-center\" data-align=\"center\">95<\/td><td class=\"has-text-align-center\" data-align=\"center\">1<\/td><td class=\"has-text-align-center\" data-align=\"center\">95<\/td><td class=\"has-text-align-center\" data-align=\"center\">1<\/td><\/tr><tr><td>Spiromesifen<\/td><td class=\"has-text-align-center\" data-align=\"center\">112<\/td><td class=\"has-text-align-center\" data-align=\"center\">6<\/td><td class=\"has-text-align-center\" data-align=\"center\">95<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><td class=\"has-text-align-center\" data-align=\"center\">100<\/td><td class=\"has-text-align-center\" data-align=\"center\">6<\/td><\/tr><tr><td>Pyrethrin I<\/td><td class=\"has-text-align-center\" data-align=\"center\">131<\/td><td class=\"has-text-align-center\" data-align=\"center\">13<\/td><td class=\"has-text-align-center\" data-align=\"center\">97<\/td><td class=\"has-text-align-center\" data-align=\"center\">4<\/td><td class=\"has-text-align-center\" data-align=\"center\">95<\/td><td class=\"has-text-align-center\" data-align=\"center\">13<\/td><\/tr><tr><td>Cyfluthrin<\/td><td class=\"has-text-align-center\" data-align=\"center\">&#8211;<\/td><td class=\"has-text-align-center\" data-align=\"center\">&#8211;<\/td><td class=\"has-text-align-center\" data-align=\"center\">&#8211;<\/td><td class=\"has-text-align-center\" data-align=\"center\">&#8211;<\/td><td class=\"has-text-align-center\" data-align=\"center\">80<\/td><td class=\"has-text-align-center\" data-align=\"center\">8<\/td><\/tr><tr><td>Cypermethrin<\/td><td class=\"has-text-align-center\" data-align=\"center\">&#8211;<\/td><td class=\"has-text-align-center\" data-align=\"center\">&#8211;<\/td><td class=\"has-text-align-center\" data-align=\"center\">95<\/td><td class=\"has-text-align-center\" data-align=\"center\">16<\/td><td class=\"has-text-align-center\" data-align=\"center\">96<\/td><td class=\"has-text-align-center\" data-align=\"center\">16<\/td><\/tr><tr><td>(E)-Fenpyroximate<\/td><td class=\"has-text-align-center\" data-align=\"center\">107<\/td><td class=\"has-text-align-center\" data-align=\"center\">6<\/td><td class=\"has-text-align-center\" data-align=\"center\">95<\/td><td class=\"has-text-align-center\" data-align=\"center\">4<\/td><td class=\"has-text-align-center\" data-align=\"center\">94<\/td><td class=\"has-text-align-center\" data-align=\"center\">6<\/td><\/tr><tr><td>Permethrin-<em>trans<\/em><\/td><td class=\"has-text-align-center\" data-align=\"center\">136<\/td><td class=\"has-text-align-center\" data-align=\"center\">22<\/td><td class=\"has-text-align-center\" data-align=\"center\">103<\/td><td class=\"has-text-align-center\" data-align=\"center\">4<\/td><td class=\"has-text-align-center\" data-align=\"center\">94<\/td><td class=\"has-text-align-center\" data-align=\"center\">22<\/td><\/tr><tr><td>Permethrin-<em>cis<\/em><\/td><td class=\"has-text-align-center\" data-align=\"center\">113<\/td><td class=\"has-text-align-center\" data-align=\"center\">12<\/td><td class=\"has-text-align-center\" data-align=\"center\">98<\/td><td class=\"has-text-align-center\" data-align=\"center\">8<\/td><td class=\"has-text-align-center\" data-align=\"center\">101<\/td><td class=\"has-text-align-center\" data-align=\"center\">12<\/td><\/tr><tr><td>Avermectin B1a<\/td><td class=\"has-text-align-center\" data-align=\"center\">122<\/td><td class=\"has-text-align-center\" data-align=\"center\">17<\/td><td class=\"has-text-align-center\" data-align=\"center\">96<\/td><td class=\"has-text-align-center\" data-align=\"center\">0<\/td><td class=\"has-text-align-center\" data-align=\"center\">87<\/td><td class=\"has-text-align-center\" data-align=\"center\">17<\/td><\/tr><tr><td>Etofenprox<\/td><td class=\"has-text-align-center\" data-align=\"center\">110<\/td><td class=\"has-text-align-center\" data-align=\"center\">4<\/td><td class=\"has-text-align-center\" data-align=\"center\">100<\/td><td class=\"has-text-align-center\" data-align=\"center\">2<\/td><td class=\"has-text-align-center\" data-align=\"center\">101<\/td><td class=\"has-text-align-center\" data-align=\"center\">4<\/td><\/tr><tr><td>Bifenthrin<\/td><td class=\"has-text-align-center\" data-align=\"center\">135<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><td class=\"has-text-align-center\" data-align=\"center\">98<\/td><td class=\"has-text-align-center\" data-align=\"center\">6<\/td><td class=\"has-text-align-center\" data-align=\"center\">85<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><\/tr><tr><td>Acequinocyl<\/td><td class=\"has-text-align-center\" data-align=\"center\">129<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><td class=\"has-text-align-center\" data-align=\"center\">95<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><td class=\"has-text-align-center\" data-align=\"center\">78<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><\/tr><\/tbody><\/table><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\"><strong>Tabella X:<\/strong>&nbsp;Limiti di quantificazione (LOQ), linearit\u00e0, accuratezza e precisione per pesticidi e micotossine nelle caramelle gommose alla cannabis.<\/p>\n\n\n\n<figure class=\"wp-block-table\"><table class=\"has-fixed-layout\"><tbody><tr><td class=\"has-text-align-left\" data-align=\"left\" rowspan=\"2\"><strong>Contaminante<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\" rowspan=\"2\"><strong>Livello di azione (ng\/g)<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\" rowspan=\"2\"><strong>LOQ (ng\/g)<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\" rowspan=\"2\"><strong>R<sup>2<\/sup><\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\" colspan=\"2\"><strong>10 ng\/g (n=4)<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\" colspan=\"2\"><strong>50 ng\/g (n=4)<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\" colspan=\"2\"><strong>100 ng\/g (n=4)<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\" colspan=\"2\"><strong>500 ng\/g (n=4)<\/strong><\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\"><strong>Accuratezza (%)<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>Precisione (RSD)<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>Accuratezza (%)<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>Precisione (RSD)<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>Accuratezza (%)<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>Precisione (RSD)<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>Accuratezza (%)<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>Precisione (RSD)<\/strong><\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Daminozide*<\/td><td class=\"has-text-align-center\" data-align=\"center\">&lt;LOD<\/td><td class=\"has-text-align-center\" data-align=\"center\">20<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.9999<\/td><td class=\"has-text-align-center\" data-align=\"center\">&#8211;<\/td><td class=\"has-text-align-center\" data-align=\"center\">&#8211;<\/td><td class=\"has-text-align-center\" data-align=\"center\">114<\/td><td class=\"has-text-align-center\" data-align=\"center\">6<\/td><td class=\"has-text-align-center\" data-align=\"center\">114<\/td><td class=\"has-text-align-center\" data-align=\"center\">4<\/td><td class=\"has-text-align-center\" data-align=\"center\">116<\/td><td class=\"has-text-align-center\" data-align=\"center\">4<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Acephate<\/td><td class=\"has-text-align-center\" data-align=\"center\">5000<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.9996<\/td><td class=\"has-text-align-center\" data-align=\"center\">100<\/td><td class=\"has-text-align-center\" data-align=\"center\">4<\/td><td class=\"has-text-align-center\" data-align=\"center\">93<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><td class=\"has-text-align-center\" data-align=\"center\">90<\/td><td class=\"has-text-align-center\" data-align=\"center\">2<\/td><td class=\"has-text-align-center\" data-align=\"center\">88<\/td><td class=\"has-text-align-center\" data-align=\"center\">2<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Oxamyl<\/td><td class=\"has-text-align-center\" data-align=\"center\">200<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.999<\/td><td class=\"has-text-align-center\" data-align=\"center\">108<\/td><td class=\"has-text-align-center\" data-align=\"center\">1<\/td><td class=\"has-text-align-center\" data-align=\"center\">102<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><td class=\"has-text-align-center\" data-align=\"center\">105<\/td><td class=\"has-text-align-center\" data-align=\"center\">4<\/td><td class=\"has-text-align-center\" data-align=\"center\">99<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Flonicamid<\/td><td class=\"has-text-align-center\" data-align=\"center\">2000<\/td><td class=\"has-text-align-center\" data-align=\"center\">50<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.999<\/td><td class=\"has-text-align-center\" data-align=\"center\">&#8211;<\/td><td class=\"has-text-align-center\" data-align=\"center\">&#8211;<\/td><td class=\"has-text-align-center\" data-align=\"center\">117<\/td><td class=\"has-text-align-center\" data-align=\"center\">6<\/td><td class=\"has-text-align-center\" data-align=\"center\">105<\/td><td class=\"has-text-align-center\" data-align=\"center\">11<\/td><td class=\"has-text-align-center\" data-align=\"center\">101<\/td><td class=\"has-text-align-center\" data-align=\"center\">4<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Methomyl<\/td><td class=\"has-text-align-center\" data-align=\"center\">100<\/td><td class=\"has-text-align-center\" data-align=\"center\">20<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.9989<\/td><td class=\"has-text-align-center\" data-align=\"center\">&#8211;<\/td><td class=\"has-text-align-center\" data-align=\"center\">&#8211;<\/td><td class=\"has-text-align-center\" data-align=\"center\">102<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><td class=\"has-text-align-center\" data-align=\"center\">103<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><td class=\"has-text-align-center\" data-align=\"center\">100<\/td><td class=\"has-text-align-center\" data-align=\"center\">1<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Thiamethoxam<\/td><td class=\"has-text-align-center\" data-align=\"center\">4500<\/td><td class=\"has-text-align-center\" data-align=\"center\">10<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.9988<\/td><td class=\"has-text-align-center\" data-align=\"center\">112<\/td><td class=\"has-text-align-center\" data-align=\"center\">23<\/td><td class=\"has-text-align-center\" data-align=\"center\">108<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><td class=\"has-text-align-center\" data-align=\"center\">108<\/td><td class=\"has-text-align-center\" data-align=\"center\">6<\/td><td class=\"has-text-align-center\" data-align=\"center\">100<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Imidacloprid<\/td><td class=\"has-text-align-center\" data-align=\"center\">3000<\/td><td class=\"has-text-align-center\" data-align=\"center\">10<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.9985<\/td><td class=\"has-text-align-center\" data-align=\"center\">105<\/td><td class=\"has-text-align-center\" data-align=\"center\">17<\/td><td class=\"has-text-align-center\" data-align=\"center\">107<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><td class=\"has-text-align-center\" data-align=\"center\">109<\/td><td class=\"has-text-align-center\" data-align=\"center\">4<\/td><td class=\"has-text-align-center\" data-align=\"center\">102<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Mevinphos (I and II)*<\/td><td class=\"has-text-align-center\" data-align=\"center\">&lt;LOD<\/td><td class=\"has-text-align-center\" data-align=\"center\">20<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.9981<\/td><td class=\"has-text-align-center\" data-align=\"center\">&#8211;<\/td><td class=\"has-text-align-center\" data-align=\"center\">&#8211;<\/td><td class=\"has-text-align-center\" data-align=\"center\">101<\/td><td class=\"has-text-align-center\" data-align=\"center\">4<\/td><td class=\"has-text-align-center\" data-align=\"center\">104<\/td><td class=\"has-text-align-center\" data-align=\"center\">7<\/td><td class=\"has-text-align-center\" data-align=\"center\">101<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Acetamiprid<\/td><td class=\"has-text-align-center\" data-align=\"center\">5000<\/td><td class=\"has-text-align-center\" data-align=\"center\">10<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.9968<\/td><td class=\"has-text-align-center\" data-align=\"center\">100<\/td><td class=\"has-text-align-center\" data-align=\"center\">8<\/td><td class=\"has-text-align-center\" data-align=\"center\">108<\/td><td class=\"has-text-align-center\" data-align=\"center\">6<\/td><td class=\"has-text-align-center\" data-align=\"center\">109<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><td class=\"has-text-align-center\" data-align=\"center\">105<\/td><td class=\"has-text-align-center\" data-align=\"center\">1<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Dimethoate*<\/td><td class=\"has-text-align-center\" data-align=\"center\">&lt;LOD<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.9994<\/td><td class=\"has-text-align-center\" data-align=\"center\">109<\/td><td class=\"has-text-align-center\" data-align=\"center\">15<\/td><td class=\"has-text-align-center\" data-align=\"center\">104<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><td class=\"has-text-align-center\" data-align=\"center\">101<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><td class=\"has-text-align-center\" data-align=\"center\">99<\/td><td class=\"has-text-align-center\" data-align=\"center\">2<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Thiacloprid*<\/td><td class=\"has-text-align-center\" data-align=\"center\">&lt;LOD<\/td><td class=\"has-text-align-center\" data-align=\"center\">20<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.9981<\/td><td class=\"has-text-align-center\" data-align=\"center\">&#8211;<\/td><td class=\"has-text-align-center\" data-align=\"center\">&#8211;<\/td><td class=\"has-text-align-center\" data-align=\"center\">100<\/td><td class=\"has-text-align-center\" data-align=\"center\">1<\/td><td class=\"has-text-align-center\" data-align=\"center\">103<\/td><td class=\"has-text-align-center\" data-align=\"center\">4<\/td><td class=\"has-text-align-center\" data-align=\"center\">103<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Aflatoxin G2<\/td><td class=\"has-text-align-center\" data-align=\"center\">20<sup>#<\/sup><\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.9957<\/td><td class=\"has-text-align-center\" data-align=\"center\">112<\/td><td class=\"has-text-align-center\" data-align=\"center\">17<\/td><td class=\"has-text-align-center\" data-align=\"center\">101<\/td><td class=\"has-text-align-center\" data-align=\"center\">6<\/td><td class=\"has-text-align-center\" data-align=\"center\">97<\/td><td class=\"has-text-align-center\" data-align=\"center\">2<\/td><td class=\"has-text-align-center\" data-align=\"center\">&#8211;<\/td><td class=\"has-text-align-center\" data-align=\"center\">&#8211;<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Aflatoxin G1<\/td><td class=\"has-text-align-center\" data-align=\"center\">20<sup>#<\/sup><\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.9984<\/td><td class=\"has-text-align-center\" data-align=\"center\">114<\/td><td class=\"has-text-align-center\" data-align=\"center\">9<\/td><td class=\"has-text-align-center\" data-align=\"center\">98<\/td><td class=\"has-text-align-center\" data-align=\"center\">1<\/td><td class=\"has-text-align-center\" data-align=\"center\">100<\/td><td class=\"has-text-align-center\" data-align=\"center\">4<\/td><td class=\"has-text-align-center\" data-align=\"center\">&#8211;<\/td><td class=\"has-text-align-center\" data-align=\"center\">&#8211;<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Aldicarb*<\/td><td class=\"has-text-align-center\" data-align=\"center\">&lt;LOD<\/td><td class=\"has-text-align-center\" data-align=\"center\">20<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.9971<\/td><td class=\"has-text-align-center\" data-align=\"center\">&#8211;<\/td><td class=\"has-text-align-center\" data-align=\"center\">&#8211;<\/td><td class=\"has-text-align-center\" data-align=\"center\">91<\/td><td class=\"has-text-align-center\" data-align=\"center\">17<\/td><td class=\"has-text-align-center\" data-align=\"center\">104<\/td><td class=\"has-text-align-center\" data-align=\"center\">8<\/td><td class=\"has-text-align-center\" data-align=\"center\">97<\/td><td class=\"has-text-align-center\" data-align=\"center\">4<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Aflatoxin B2<\/td><td class=\"has-text-align-center\" data-align=\"center\">20<sup>#<\/sup><\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.9973<\/td><td class=\"has-text-align-center\" data-align=\"center\">97<\/td><td class=\"has-text-align-center\" data-align=\"center\">23<\/td><td class=\"has-text-align-center\" data-align=\"center\">107<\/td><td class=\"has-text-align-center\" data-align=\"center\">6<\/td><td class=\"has-text-align-center\" data-align=\"center\">94<\/td><td class=\"has-text-align-center\" data-align=\"center\">7<\/td><td class=\"has-text-align-center\" data-align=\"center\">&#8211;<\/td><td class=\"has-text-align-center\" data-align=\"center\">&#8211;<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Dichlorvos*<\/td><td class=\"has-text-align-center\" data-align=\"center\">&lt;LOD<\/td><td class=\"has-text-align-center\" data-align=\"center\">10<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.9984<\/td><td class=\"has-text-align-center\" data-align=\"center\">98<\/td><td class=\"has-text-align-center\" data-align=\"center\">17<\/td><td class=\"has-text-align-center\" data-align=\"center\">103<\/td><td class=\"has-text-align-center\" data-align=\"center\">4<\/td><td class=\"has-text-align-center\" data-align=\"center\">97<\/td><td class=\"has-text-align-center\" data-align=\"center\">18<\/td><td class=\"has-text-align-center\" data-align=\"center\">106<\/td><td class=\"has-text-align-center\" data-align=\"center\">4<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Aflatoxin B1<\/td><td class=\"has-text-align-center\" data-align=\"center\">20<sup>#<\/sup><\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.9978<\/td><td class=\"has-text-align-center\" data-align=\"center\">113<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><td class=\"has-text-align-center\" data-align=\"center\">101<\/td><td class=\"has-text-align-center\" data-align=\"center\">6<\/td><td class=\"has-text-align-center\" data-align=\"center\">96<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><td class=\"has-text-align-center\" data-align=\"center\">&#8211;<\/td><td class=\"has-text-align-center\" data-align=\"center\">&#8211;<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Imazalil*<\/td><td class=\"has-text-align-center\" data-align=\"center\">&lt;LOD<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.9977<\/td><td class=\"has-text-align-center\" data-align=\"center\">97<\/td><td class=\"has-text-align-center\" data-align=\"center\">19<\/td><td class=\"has-text-align-center\" data-align=\"center\">109<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><td class=\"has-text-align-center\" data-align=\"center\">107<\/td><td class=\"has-text-align-center\" data-align=\"center\">4<\/td><td class=\"has-text-align-center\" data-align=\"center\">105<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Carbofuran*<\/td><td class=\"has-text-align-center\" data-align=\"center\">&lt;LOD<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.9973<\/td><td class=\"has-text-align-center\" data-align=\"center\">93<\/td><td class=\"has-text-align-center\" data-align=\"center\">7<\/td><td class=\"has-text-align-center\" data-align=\"center\">108<\/td><td class=\"has-text-align-center\" data-align=\"center\">1<\/td><td class=\"has-text-align-center\" data-align=\"center\">109<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><td class=\"has-text-align-center\" data-align=\"center\">99<\/td><td class=\"has-text-align-center\" data-align=\"center\">4<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Propoxur*<\/td><td class=\"has-text-align-center\" data-align=\"center\">&lt;LOD<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.9977<\/td><td class=\"has-text-align-center\" data-align=\"center\">108<\/td><td class=\"has-text-align-center\" data-align=\"center\">6<\/td><td class=\"has-text-align-center\" data-align=\"center\">108<\/td><td class=\"has-text-align-center\" data-align=\"center\">2<\/td><td class=\"has-text-align-center\" data-align=\"center\">107<\/td><td class=\"has-text-align-center\" data-align=\"center\">4<\/td><td class=\"has-text-align-center\" data-align=\"center\">102<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Carbaryl<\/td><td class=\"has-text-align-center\" data-align=\"center\">500<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.9988<\/td><td class=\"has-text-align-center\" data-align=\"center\">95<\/td><td class=\"has-text-align-center\" data-align=\"center\">14<\/td><td class=\"has-text-align-center\" data-align=\"center\">109<\/td><td class=\"has-text-align-center\" data-align=\"center\">6<\/td><td class=\"has-text-align-center\" data-align=\"center\">108<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><td class=\"has-text-align-center\" data-align=\"center\">102<\/td><td class=\"has-text-align-center\" data-align=\"center\">2<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Naled<\/td><td class=\"has-text-align-center\" data-align=\"center\">500<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.9968<\/td><td class=\"has-text-align-center\" data-align=\"center\">98<\/td><td class=\"has-text-align-center\" data-align=\"center\">4<\/td><td class=\"has-text-align-center\" data-align=\"center\">112<\/td><td class=\"has-text-align-center\" data-align=\"center\">8<\/td><td class=\"has-text-align-center\" data-align=\"center\">110<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><td class=\"has-text-align-center\" data-align=\"center\">101<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Metalaxyl<\/td><td class=\"has-text-align-center\" data-align=\"center\">15,000<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.9988<\/td><td class=\"has-text-align-center\" data-align=\"center\">101<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><td class=\"has-text-align-center\" data-align=\"center\">105<\/td><td class=\"has-text-align-center\" data-align=\"center\">4<\/td><td class=\"has-text-align-center\" data-align=\"center\">106<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><td class=\"has-text-align-center\" data-align=\"center\">99<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Spiroxamine*<\/td><td class=\"has-text-align-center\" data-align=\"center\">&lt;LOD<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.9977<\/td><td class=\"has-text-align-center\" data-align=\"center\">104<\/td><td class=\"has-text-align-center\" data-align=\"center\">6<\/td><td class=\"has-text-align-center\" data-align=\"center\">106<\/td><td class=\"has-text-align-center\" data-align=\"center\">2<\/td><td class=\"has-text-align-center\" data-align=\"center\">105<\/td><td class=\"has-text-align-center\" data-align=\"center\">2<\/td><td class=\"has-text-align-center\" data-align=\"center\">101<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Chlorantraniliprole<\/td><td class=\"has-text-align-center\" data-align=\"center\">40,000<\/td><td class=\"has-text-align-center\" data-align=\"center\">20<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.9971<\/td><td class=\"has-text-align-center\" data-align=\"center\">&#8211;<\/td><td class=\"has-text-align-center\" data-align=\"center\">&#8211;<\/td><td class=\"has-text-align-center\" data-align=\"center\">93<\/td><td class=\"has-text-align-center\" data-align=\"center\">7<\/td><td class=\"has-text-align-center\" data-align=\"center\">104<\/td><td class=\"has-text-align-center\" data-align=\"center\">6<\/td><td class=\"has-text-align-center\" data-align=\"center\">104<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Phosmet<\/td><td class=\"has-text-align-center\" data-align=\"center\">200<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.9992<\/td><td class=\"has-text-align-center\" data-align=\"center\">109<\/td><td class=\"has-text-align-center\" data-align=\"center\">14<\/td><td class=\"has-text-align-center\" data-align=\"center\">105<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><td class=\"has-text-align-center\" data-align=\"center\">104<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><td class=\"has-text-align-center\" data-align=\"center\">100<\/td><td class=\"has-text-align-center\" data-align=\"center\">4<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Azoxystrobin<\/td><td class=\"has-text-align-center\" data-align=\"center\">40,000<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.9992<\/td><td class=\"has-text-align-center\" data-align=\"center\">100<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><td class=\"has-text-align-center\" data-align=\"center\">104<\/td><td class=\"has-text-align-center\" data-align=\"center\">1<\/td><td class=\"has-text-align-center\" data-align=\"center\">105<\/td><td class=\"has-text-align-center\" data-align=\"center\">4<\/td><td class=\"has-text-align-center\" data-align=\"center\">102<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Fludioxonil<\/td><td class=\"has-text-align-center\" data-align=\"center\">30,000<\/td><td class=\"has-text-align-center\" data-align=\"center\">20<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.9949<\/td><td class=\"has-text-align-center\" data-align=\"center\">&#8211;<\/td><td class=\"has-text-align-center\" data-align=\"center\">&#8211;<\/td><td class=\"has-text-align-center\" data-align=\"center\">109<\/td><td class=\"has-text-align-center\" data-align=\"center\">13<\/td><td class=\"has-text-align-center\" data-align=\"center\">97<\/td><td class=\"has-text-align-center\" data-align=\"center\">6<\/td><td class=\"has-text-align-center\" data-align=\"center\">101<\/td><td class=\"has-text-align-center\" data-align=\"center\">6<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Methiocarb*<\/td><td class=\"has-text-align-center\" data-align=\"center\">&lt;LOD<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.9988<\/td><td class=\"has-text-align-center\" data-align=\"center\">116<\/td><td class=\"has-text-align-center\" data-align=\"center\">16<\/td><td class=\"has-text-align-center\" data-align=\"center\">105<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><td class=\"has-text-align-center\" data-align=\"center\">107<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><td class=\"has-text-align-center\" data-align=\"center\">101<\/td><td class=\"has-text-align-center\" data-align=\"center\">4<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Dimethomorph (I and II)<\/td><td class=\"has-text-align-center\" data-align=\"center\">20,000<\/td><td class=\"has-text-align-center\" data-align=\"center\">10<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.999<\/td><td class=\"has-text-align-center\" data-align=\"center\">75<\/td><td class=\"has-text-align-center\" data-align=\"center\">14<\/td><td class=\"has-text-align-center\" data-align=\"center\">101<\/td><td class=\"has-text-align-center\" data-align=\"center\">6<\/td><td class=\"has-text-align-center\" data-align=\"center\">93<\/td><td class=\"has-text-align-center\" data-align=\"center\">8<\/td><td class=\"has-text-align-center\" data-align=\"center\">101<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Boscalid<\/td><td class=\"has-text-align-center\" data-align=\"center\">10,000<\/td><td class=\"has-text-align-center\" data-align=\"center\">10<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.9964<\/td><td class=\"has-text-align-center\" data-align=\"center\">108<\/td><td class=\"has-text-align-center\" data-align=\"center\">15<\/td><td class=\"has-text-align-center\" data-align=\"center\">102<\/td><td class=\"has-text-align-center\" data-align=\"center\">12<\/td><td class=\"has-text-align-center\" data-align=\"center\">102<\/td><td class=\"has-text-align-center\" data-align=\"center\">2<\/td><td class=\"has-text-align-center\" data-align=\"center\">103<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Paclobutrazol*<\/td><td class=\"has-text-align-center\" data-align=\"center\">&lt;LOD<\/td><td class=\"has-text-align-center\" data-align=\"center\">10<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.9979<\/td><td class=\"has-text-align-center\" data-align=\"center\">99<\/td><td class=\"has-text-align-center\" data-align=\"center\">9<\/td><td class=\"has-text-align-center\" data-align=\"center\">106<\/td><td class=\"has-text-align-center\" data-align=\"center\">1<\/td><td class=\"has-text-align-center\" data-align=\"center\">108<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><td class=\"has-text-align-center\" data-align=\"center\">100<\/td><td class=\"has-text-align-center\" data-align=\"center\">4<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Malathion<\/td><td class=\"has-text-align-center\" data-align=\"center\">5000<\/td><td class=\"has-text-align-center\" data-align=\"center\">10<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.9989<\/td><td class=\"has-text-align-center\" data-align=\"center\">117<\/td><td class=\"has-text-align-center\" data-align=\"center\">12<\/td><td class=\"has-text-align-center\" data-align=\"center\">112<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><td class=\"has-text-align-center\" data-align=\"center\">106<\/td><td class=\"has-text-align-center\" data-align=\"center\">2<\/td><td class=\"has-text-align-center\" data-align=\"center\">101<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Myclobutanil<\/td><td class=\"has-text-align-center\" data-align=\"center\">9000<\/td><td class=\"has-text-align-center\" data-align=\"center\">10<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.9986<\/td><td class=\"has-text-align-center\" data-align=\"center\">102<\/td><td class=\"has-text-align-center\" data-align=\"center\">22<\/td><td class=\"has-text-align-center\" data-align=\"center\">102<\/td><td class=\"has-text-align-center\" data-align=\"center\">6<\/td><td class=\"has-text-align-center\" data-align=\"center\">104<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><td class=\"has-text-align-center\" data-align=\"center\">101<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Bifenazate<\/td><td class=\"has-text-align-center\" data-align=\"center\">5000<\/td><td class=\"has-text-align-center\" data-align=\"center\">10<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.9994<\/td><td class=\"has-text-align-center\" data-align=\"center\">118<\/td><td class=\"has-text-align-center\" data-align=\"center\">20<\/td><td class=\"has-text-align-center\" data-align=\"center\">110<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><td class=\"has-text-align-center\" data-align=\"center\">104<\/td><td class=\"has-text-align-center\" data-align=\"center\">10<\/td><td class=\"has-text-align-center\" data-align=\"center\">102<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Ochratoxin A<\/td><td class=\"has-text-align-center\" data-align=\"center\">20<\/td><td class=\"has-text-align-center\" data-align=\"center\">10<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.9957<\/td><td class=\"has-text-align-center\" data-align=\"center\">99<\/td><td class=\"has-text-align-center\" data-align=\"center\">8<\/td><td class=\"has-text-align-center\" data-align=\"center\">99<\/td><td class=\"has-text-align-center\" data-align=\"center\">24<\/td><td class=\"has-text-align-center\" data-align=\"center\">104<\/td><td class=\"has-text-align-center\" data-align=\"center\">8<\/td><td class=\"has-text-align-center\" data-align=\"center\">&#8211;<\/td><td class=\"has-text-align-center\" data-align=\"center\">&#8211;<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Fenhexamid<\/td><td class=\"has-text-align-center\" data-align=\"center\">10,000<\/td><td class=\"has-text-align-center\" data-align=\"center\">10<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.9969<\/td><td class=\"has-text-align-center\" data-align=\"center\">96<\/td><td class=\"has-text-align-center\" data-align=\"center\">21<\/td><td class=\"has-text-align-center\" data-align=\"center\">109<\/td><td class=\"has-text-align-center\" data-align=\"center\">4<\/td><td class=\"has-text-align-center\" data-align=\"center\">106<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><td class=\"has-text-align-center\" data-align=\"center\">107<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Spirotetramat<\/td><td class=\"has-text-align-center\" data-align=\"center\">13,000<\/td><td class=\"has-text-align-center\" data-align=\"center\">10<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.9987<\/td><td class=\"has-text-align-center\" data-align=\"center\">83<\/td><td class=\"has-text-align-center\" data-align=\"center\">18<\/td><td class=\"has-text-align-center\" data-align=\"center\">107<\/td><td class=\"has-text-align-center\" data-align=\"center\">6<\/td><td class=\"has-text-align-center\" data-align=\"center\">106<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><td class=\"has-text-align-center\" data-align=\"center\">105<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Ethoprophos*<\/td><td class=\"has-text-align-center\" data-align=\"center\">&lt;LOD<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.9985<\/td><td class=\"has-text-align-center\" data-align=\"center\">101<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><td class=\"has-text-align-center\" data-align=\"center\">106<\/td><td class=\"has-text-align-center\" data-align=\"center\">4<\/td><td class=\"has-text-align-center\" data-align=\"center\">104<\/td><td class=\"has-text-align-center\" data-align=\"center\">1<\/td><td class=\"has-text-align-center\" data-align=\"center\">102<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Fipronil*<\/td><td class=\"has-text-align-center\" data-align=\"center\">&lt;LOD<\/td><td class=\"has-text-align-center\" data-align=\"center\">20<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.998<\/td><td class=\"has-text-align-center\" data-align=\"center\">&#8211;<\/td><td class=\"has-text-align-center\" data-align=\"center\">&#8211;<\/td><td class=\"has-text-align-center\" data-align=\"center\">96<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><td class=\"has-text-align-center\" data-align=\"center\">104<\/td><td class=\"has-text-align-center\" data-align=\"center\">8<\/td><td class=\"has-text-align-center\" data-align=\"center\">103<\/td><td class=\"has-text-align-center\" data-align=\"center\">4<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Fenoxycarb*<\/td><td class=\"has-text-align-center\" data-align=\"center\">&lt;LOD<\/td><td class=\"has-text-align-center\" data-align=\"center\">10<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.9967<\/td><td class=\"has-text-align-center\" data-align=\"center\">115<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><td class=\"has-text-align-center\" data-align=\"center\">106<\/td><td class=\"has-text-align-center\" data-align=\"center\">4<\/td><td class=\"has-text-align-center\" data-align=\"center\">107<\/td><td class=\"has-text-align-center\" data-align=\"center\">2<\/td><td class=\"has-text-align-center\" data-align=\"center\">102<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Kresoxym-methyl<\/td><td class=\"has-text-align-center\" data-align=\"center\">1000<\/td><td class=\"has-text-align-center\" data-align=\"center\">10<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.9993<\/td><td class=\"has-text-align-center\" data-align=\"center\">112<\/td><td class=\"has-text-align-center\" data-align=\"center\">20<\/td><td class=\"has-text-align-center\" data-align=\"center\">104<\/td><td class=\"has-text-align-center\" data-align=\"center\">10<\/td><td class=\"has-text-align-center\" data-align=\"center\">102<\/td><td class=\"has-text-align-center\" data-align=\"center\">4<\/td><td class=\"has-text-align-center\" data-align=\"center\">103<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Tebuconazole<\/td><td class=\"has-text-align-center\" data-align=\"center\">2000<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.999<\/td><td class=\"has-text-align-center\" data-align=\"center\">100<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><td class=\"has-text-align-center\" data-align=\"center\">110<\/td><td class=\"has-text-align-center\" data-align=\"center\">1<\/td><td class=\"has-text-align-center\" data-align=\"center\">105<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><td class=\"has-text-align-center\" data-align=\"center\">101<\/td><td class=\"has-text-align-center\" data-align=\"center\">4<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Spinosad- spinosyn A (71 %)<sup>a<\/sup><\/td><td class=\"has-text-align-center\" data-align=\"center\">3000<sup>\u00a5<\/sup><\/td><td class=\"has-text-align-center\" data-align=\"center\">7.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.9988<\/td><td class=\"has-text-align-center\" data-align=\"center\">117<\/td><td class=\"has-text-align-center\" data-align=\"center\">2<\/td><td class=\"has-text-align-center\" data-align=\"center\">110<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><td class=\"has-text-align-center\" data-align=\"center\">108<\/td><td class=\"has-text-align-center\" data-align=\"center\">2<\/td><td class=\"has-text-align-center\" data-align=\"center\">102<\/td><td class=\"has-text-align-center\" data-align=\"center\">1<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Diazinon<\/td><td class=\"has-text-align-center\" data-align=\"center\">200<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.9997<\/td><td class=\"has-text-align-center\" data-align=\"center\">104<\/td><td class=\"has-text-align-center\" data-align=\"center\">1<\/td><td class=\"has-text-align-center\" data-align=\"center\">102<\/td><td class=\"has-text-align-center\" data-align=\"center\">1<\/td><td class=\"has-text-align-center\" data-align=\"center\">104<\/td><td class=\"has-text-align-center\" data-align=\"center\">2<\/td><td class=\"has-text-align-center\" data-align=\"center\">101<\/td><td class=\"has-text-align-center\" data-align=\"center\">1<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Coumaphos*<\/td><td class=\"has-text-align-center\" data-align=\"center\">&lt;LOD<\/td><td class=\"has-text-align-center\" data-align=\"center\">10<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.9991<\/td><td class=\"has-text-align-center\" data-align=\"center\">107<\/td><td class=\"has-text-align-center\" data-align=\"center\">11<\/td><td class=\"has-text-align-center\" data-align=\"center\">109<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><td class=\"has-text-align-center\" data-align=\"center\">108<\/td><td class=\"has-text-align-center\" data-align=\"center\">2<\/td><td class=\"has-text-align-center\" data-align=\"center\">103<\/td><td class=\"has-text-align-center\" data-align=\"center\">2<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Pyridaben<\/td><td class=\"has-text-align-center\" data-align=\"center\">3000<\/td><td class=\"has-text-align-center\" data-align=\"center\">50<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.9994<\/td><td class=\"has-text-align-center\" data-align=\"center\">&#8211;<\/td><td class=\"has-text-align-center\" data-align=\"center\">&#8211;<\/td><td class=\"has-text-align-center\" data-align=\"center\">93<\/td><td class=\"has-text-align-center\" data-align=\"center\">12<\/td><td class=\"has-text-align-center\" data-align=\"center\">113<\/td><td class=\"has-text-align-center\" data-align=\"center\">6<\/td><td class=\"has-text-align-center\" data-align=\"center\">103<\/td><td class=\"has-text-align-center\" data-align=\"center\">1<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Propiconazole<\/td><td class=\"has-text-align-center\" data-align=\"center\">20,000<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.9991<\/td><td class=\"has-text-align-center\" data-align=\"center\">96<\/td><td class=\"has-text-align-center\" data-align=\"center\">6<\/td><td class=\"has-text-align-center\" data-align=\"center\">108<\/td><td class=\"has-text-align-center\" data-align=\"center\">4<\/td><td class=\"has-text-align-center\" data-align=\"center\">104<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><td class=\"has-text-align-center\" data-align=\"center\">103<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Clofentezine<\/td><td class=\"has-text-align-center\" data-align=\"center\">500<\/td><td class=\"has-text-align-center\" data-align=\"center\">20<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.9978<\/td><td class=\"has-text-align-center\" data-align=\"center\">&#8211;<\/td><td class=\"has-text-align-center\" data-align=\"center\">&#8211;<\/td><td class=\"has-text-align-center\" data-align=\"center\">102<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><td class=\"has-text-align-center\" data-align=\"center\">105<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><td class=\"has-text-align-center\" data-align=\"center\">106<\/td><td class=\"has-text-align-center\" data-align=\"center\">4<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Spinosad &#8211; spinosyn D (29%)<sup>b<\/sup><\/td><td class=\"has-text-align-center\" data-align=\"center\">3000<sup>\u00a5<\/sup><\/td><td class=\"has-text-align-center\" data-align=\"center\">2.9<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.9993<\/td><td class=\"has-text-align-center\" data-align=\"center\">106<\/td><td class=\"has-text-align-center\" data-align=\"center\">8<\/td><td class=\"has-text-align-center\" data-align=\"center\">101<\/td><td class=\"has-text-align-center\" data-align=\"center\">2<\/td><td class=\"has-text-align-center\" data-align=\"center\">103<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><td class=\"has-text-align-center\" data-align=\"center\">104<\/td><td class=\"has-text-align-center\" data-align=\"center\">4<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Spinetoram &#8211; spinosyn J (80%)<sup>c<\/sup><\/td><td class=\"has-text-align-center\" data-align=\"center\">3000<sup>\u00a7<\/sup><\/td><td class=\"has-text-align-center\" data-align=\"center\">4<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.9995<\/td><td class=\"has-text-align-center\" data-align=\"center\">104<\/td><td class=\"has-text-align-center\" data-align=\"center\">7<\/td><td class=\"has-text-align-center\" data-align=\"center\">102<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><td class=\"has-text-align-center\" data-align=\"center\">108<\/td><td class=\"has-text-align-center\" data-align=\"center\">4<\/td><td class=\"has-text-align-center\" data-align=\"center\">102<\/td><td class=\"has-text-align-center\" data-align=\"center\">2<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Trifloxystrobin<\/td><td class=\"has-text-align-center\" data-align=\"center\">30,000<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.9996<\/td><td class=\"has-text-align-center\" data-align=\"center\">107<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><td class=\"has-text-align-center\" data-align=\"center\">106<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><td class=\"has-text-align-center\" data-align=\"center\">106<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><td class=\"has-text-align-center\" data-align=\"center\">104<\/td><td class=\"has-text-align-center\" data-align=\"center\">2<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Prallethrin<\/td><td class=\"has-text-align-center\" data-align=\"center\">400<\/td><td class=\"has-text-align-center\" data-align=\"center\">10<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.9967<\/td><td class=\"has-text-align-center\" data-align=\"center\">107<\/td><td class=\"has-text-align-center\" data-align=\"center\">21<\/td><td class=\"has-text-align-center\" data-align=\"center\">99<\/td><td class=\"has-text-align-center\" data-align=\"center\">16<\/td><td class=\"has-text-align-center\" data-align=\"center\">111<\/td><td class=\"has-text-align-center\" data-align=\"center\">6<\/td><td class=\"has-text-align-center\" data-align=\"center\">101<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Pyrethrin II (34%)<sup>f<\/sup><\/td><td class=\"has-text-align-center\" data-align=\"center\">1000<sup>\u00a3<\/sup><\/td><td class=\"has-text-align-center\" data-align=\"center\">17<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.9977<\/td><td class=\"has-text-align-center\" data-align=\"center\">&#8211;<\/td><td class=\"has-text-align-center\" data-align=\"center\">&#8211;<\/td><td class=\"has-text-align-center\" data-align=\"center\">94<\/td><td class=\"has-text-align-center\" data-align=\"center\">11<\/td><td class=\"has-text-align-center\" data-align=\"center\">112<\/td><td class=\"has-text-align-center\" data-align=\"center\">4<\/td><td class=\"has-text-align-center\" data-align=\"center\">106<\/td><td class=\"has-text-align-center\" data-align=\"center\">6<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Spinetoram &#8211; spinosyn L (20%)<sup>d<\/sup><\/td><td class=\"has-text-align-center\" data-align=\"center\">3000<sup>\u00a7<\/sup><\/td><td class=\"has-text-align-center\" data-align=\"center\">2<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.9993<\/td><td class=\"has-text-align-center\" data-align=\"center\">107<\/td><td class=\"has-text-align-center\" data-align=\"center\">10<\/td><td class=\"has-text-align-center\" data-align=\"center\">109<\/td><td class=\"has-text-align-center\" data-align=\"center\">4<\/td><td class=\"has-text-align-center\" data-align=\"center\">107<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><td class=\"has-text-align-center\" data-align=\"center\">102<\/td><td class=\"has-text-align-center\" data-align=\"center\">1<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Piperonyl Butoxide<\/td><td class=\"has-text-align-center\" data-align=\"center\">8000<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.9998<\/td><td class=\"has-text-align-center\" data-align=\"center\">110<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><td class=\"has-text-align-center\" data-align=\"center\">99<\/td><td class=\"has-text-align-center\" data-align=\"center\">2<\/td><td class=\"has-text-align-center\" data-align=\"center\">94<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><td class=\"has-text-align-center\" data-align=\"center\">95<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Chlorpyrifos*<\/td><td class=\"has-text-align-center\" data-align=\"center\">&lt;LOD<\/td><td class=\"has-text-align-center\" data-align=\"center\">20<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.9995<\/td><td class=\"has-text-align-center\" data-align=\"center\">&#8211;<\/td><td class=\"has-text-align-center\" data-align=\"center\">&#8211;<\/td><td class=\"has-text-align-center\" data-align=\"center\">103<\/td><td class=\"has-text-align-center\" data-align=\"center\">9<\/td><td class=\"has-text-align-center\" data-align=\"center\">106<\/td><td class=\"has-text-align-center\" data-align=\"center\">4<\/td><td class=\"has-text-align-center\" data-align=\"center\">104<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Hexythiazox<\/td><td class=\"has-text-align-center\" data-align=\"center\">2000<\/td><td class=\"has-text-align-center\" data-align=\"center\">10<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.9975<\/td><td class=\"has-text-align-center\" data-align=\"center\">104<\/td><td class=\"has-text-align-center\" data-align=\"center\">15<\/td><td class=\"has-text-align-center\" data-align=\"center\">102<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><td class=\"has-text-align-center\" data-align=\"center\">107<\/td><td class=\"has-text-align-center\" data-align=\"center\">4<\/td><td class=\"has-text-align-center\" data-align=\"center\">106<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Etoxazole<\/td><td class=\"has-text-align-center\" data-align=\"center\">1500<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.9995<\/td><td class=\"has-text-align-center\" data-align=\"center\">103<\/td><td class=\"has-text-align-center\" data-align=\"center\">4<\/td><td class=\"has-text-align-center\" data-align=\"center\">104<\/td><td class=\"has-text-align-center\" data-align=\"center\">2<\/td><td class=\"has-text-align-center\" data-align=\"center\">103<\/td><td class=\"has-text-align-center\" data-align=\"center\">1<\/td><td class=\"has-text-align-center\" data-align=\"center\">100<\/td><td class=\"has-text-align-center\" data-align=\"center\">1<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Spiromesifen<\/td><td class=\"has-text-align-center\" data-align=\"center\">12,000<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.9992<\/td><td class=\"has-text-align-center\" data-align=\"center\">98<\/td><td class=\"has-text-align-center\" data-align=\"center\">8<\/td><td class=\"has-text-align-center\" data-align=\"center\">109<\/td><td class=\"has-text-align-center\" data-align=\"center\">7<\/td><td class=\"has-text-align-center\" data-align=\"center\">111<\/td><td class=\"has-text-align-center\" data-align=\"center\">2<\/td><td class=\"has-text-align-center\" data-align=\"center\">102<\/td><td class=\"has-text-align-center\" data-align=\"center\">2<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Pyrethrin I (54%)<sup>e<\/sup><\/td><td class=\"has-text-align-center\" data-align=\"center\">1000<sup>\u00a3<\/sup><\/td><td class=\"has-text-align-center\" data-align=\"center\">11<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.9977<\/td><td class=\"has-text-align-center\" data-align=\"center\">&#8211;<\/td><td class=\"has-text-align-center\" data-align=\"center\">&#8211;<\/td><td class=\"has-text-align-center\" data-align=\"center\">98<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><td class=\"has-text-align-center\" data-align=\"center\">105<\/td><td class=\"has-text-align-center\" data-align=\"center\">12<\/td><td class=\"has-text-align-center\" data-align=\"center\">104<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Cyfluthrin<\/td><td class=\"has-text-align-center\" data-align=\"center\">1000<\/td><td class=\"has-text-align-center\" data-align=\"center\">50<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.999<\/td><td class=\"has-text-align-center\" data-align=\"center\">&#8211;<\/td><td class=\"has-text-align-center\" data-align=\"center\">&#8211;<\/td><td class=\"has-text-align-center\" data-align=\"center\">93<\/td><td class=\"has-text-align-center\" data-align=\"center\">11<\/td><td class=\"has-text-align-center\" data-align=\"center\">102<\/td><td class=\"has-text-align-center\" data-align=\"center\">23<\/td><td class=\"has-text-align-center\" data-align=\"center\">115<\/td><td class=\"has-text-align-center\" data-align=\"center\">10<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Cypermethrin<\/td><td class=\"has-text-align-center\" data-align=\"center\">1000<\/td><td class=\"has-text-align-center\" data-align=\"center\">50<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.9961<\/td><td class=\"has-text-align-center\" data-align=\"center\">&#8211;<\/td><td class=\"has-text-align-center\" data-align=\"center\">&#8211;<\/td><td class=\"has-text-align-center\" data-align=\"center\">115<\/td><td class=\"has-text-align-center\" data-align=\"center\">11<\/td><td class=\"has-text-align-center\" data-align=\"center\">98<\/td><td class=\"has-text-align-center\" data-align=\"center\">18<\/td><td class=\"has-text-align-center\" data-align=\"center\">104<\/td><td class=\"has-text-align-center\" data-align=\"center\">6<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">(E)-Fenpyroximate<\/td><td class=\"has-text-align-center\" data-align=\"center\">2000<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.9995<\/td><td class=\"has-text-align-center\" data-align=\"center\">102<\/td><td class=\"has-text-align-center\" data-align=\"center\">9<\/td><td class=\"has-text-align-center\" data-align=\"center\">109<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><td class=\"has-text-align-center\" data-align=\"center\">109<\/td><td class=\"has-text-align-center\" data-align=\"center\">2<\/td><td class=\"has-text-align-center\" data-align=\"center\">105<\/td><td class=\"has-text-align-center\" data-align=\"center\">2<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Permethrin-<em>trans<\/em>&nbsp;(59%)<sup>h<\/sup><\/td><td class=\"has-text-align-center\" data-align=\"center\">20,000<sup>\u00a2<\/sup><\/td><td class=\"has-text-align-center\" data-align=\"center\">12<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.9994<\/td><td class=\"has-text-align-center\" data-align=\"center\">&#8211;<\/td><td class=\"has-text-align-center\" data-align=\"center\">&#8211;<\/td><td class=\"has-text-align-center\" data-align=\"center\">99<\/td><td class=\"has-text-align-center\" data-align=\"center\">8<\/td><td class=\"has-text-align-center\" data-align=\"center\">104<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><td class=\"has-text-align-center\" data-align=\"center\">101<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Permethrin-<em>cis<\/em>&nbsp;(41%)<sup>g<\/sup><\/td><td class=\"has-text-align-center\" data-align=\"center\">20,000<sup>\u00a2<\/sup><\/td><td class=\"has-text-align-center\" data-align=\"center\">8<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.9996<\/td><td class=\"has-text-align-center\" data-align=\"center\">95<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><td class=\"has-text-align-center\" data-align=\"center\">104<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><td class=\"has-text-align-center\" data-align=\"center\">103<\/td><td class=\"has-text-align-center\" data-align=\"center\">4<\/td><td class=\"has-text-align-center\" data-align=\"center\">101<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Avermectin B1a<\/td><td class=\"has-text-align-center\" data-align=\"center\">300<\/td><td class=\"has-text-align-center\" data-align=\"center\">50<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.9988<\/td><td class=\"has-text-align-center\" data-align=\"center\">&#8211;<\/td><td class=\"has-text-align-center\" data-align=\"center\">&#8211;<\/td><td class=\"has-text-align-center\" data-align=\"center\">114<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><td class=\"has-text-align-center\" data-align=\"center\">108<\/td><td class=\"has-text-align-center\" data-align=\"center\">2<\/td><td class=\"has-text-align-center\" data-align=\"center\">105<\/td><td class=\"has-text-align-center\" data-align=\"center\">2<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Etofenprox*<\/td><td class=\"has-text-align-center\" data-align=\"center\">&lt;LOD<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.9994<\/td><td class=\"has-text-align-center\" data-align=\"center\">104<\/td><td class=\"has-text-align-center\" data-align=\"center\">7<\/td><td class=\"has-text-align-center\" data-align=\"center\">107<\/td><td class=\"has-text-align-center\" data-align=\"center\">1<\/td><td class=\"has-text-align-center\" data-align=\"center\">106<\/td><td class=\"has-text-align-center\" data-align=\"center\">2<\/td><td class=\"has-text-align-center\" data-align=\"center\">103<\/td><td class=\"has-text-align-center\" data-align=\"center\">1<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Bifenthrin<\/td><td class=\"has-text-align-center\" data-align=\"center\">500<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.999<\/td><td class=\"has-text-align-center\" data-align=\"center\">99<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><td class=\"has-text-align-center\" data-align=\"center\">103<\/td><td class=\"has-text-align-center\" data-align=\"center\">2<\/td><td class=\"has-text-align-center\" data-align=\"center\">108<\/td><td class=\"has-text-align-center\" data-align=\"center\">6<\/td><td class=\"has-text-align-center\" data-align=\"center\">104<\/td><td class=\"has-text-align-center\" data-align=\"center\">2<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Acequinocyl<\/td><td class=\"has-text-align-center\" data-align=\"center\">4000<\/td><td class=\"has-text-align-center\" data-align=\"center\">10<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.9997<\/td><td class=\"has-text-align-center\" data-align=\"center\">104<\/td><td class=\"has-text-align-center\" data-align=\"center\">7<\/td><td class=\"has-text-align-center\" data-align=\"center\">109<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><td class=\"has-text-align-center\" data-align=\"center\">108<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><td class=\"has-text-align-center\" data-align=\"center\">106<\/td><td class=\"has-text-align-center\" data-align=\"center\">2<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Quintozene (PCNB) (GC)<\/td><td class=\"has-text-align-center\" data-align=\"center\">200<\/td><td class=\"has-text-align-center\" data-align=\"center\">10<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.9966<\/td><td class=\"has-text-align-center\" data-align=\"center\">110<\/td><td class=\"has-text-align-center\" data-align=\"center\">14<\/td><td class=\"has-text-align-center\" data-align=\"center\">102<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><td class=\"has-text-align-center\" data-align=\"center\">101<\/td><td class=\"has-text-align-center\" data-align=\"center\">1<\/td><td class=\"has-text-align-center\" data-align=\"center\">97<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Methyl parathion (GC)*<\/td><td class=\"has-text-align-center\" data-align=\"center\">&lt;LOD<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.9934<\/td><td class=\"has-text-align-center\" data-align=\"center\">89<\/td><td class=\"has-text-align-center\" data-align=\"center\">10<\/td><td class=\"has-text-align-center\" data-align=\"center\">89<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><td class=\"has-text-align-center\" data-align=\"center\">87<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><td class=\"has-text-align-center\" data-align=\"center\">89<\/td><td class=\"has-text-align-center\" data-align=\"center\">6<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Captan (GC)<\/td><td class=\"has-text-align-center\" data-align=\"center\">5000<\/td><td class=\"has-text-align-center\" data-align=\"center\">10<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.9924<\/td><td class=\"has-text-align-center\" data-align=\"center\">110<\/td><td class=\"has-text-align-center\" data-align=\"center\">14<\/td><td class=\"has-text-align-center\" data-align=\"center\">96<\/td><td class=\"has-text-align-center\" data-align=\"center\">10<\/td><td class=\"has-text-align-center\" data-align=\"center\">94<\/td><td class=\"has-text-align-center\" data-align=\"center\">18<\/td><td class=\"has-text-align-center\" data-align=\"center\">92<\/td><td class=\"has-text-align-center\" data-align=\"center\">7<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Chlordane (GC)*<\/td><td class=\"has-text-align-center\" data-align=\"center\">&lt;LOD<\/td><td class=\"has-text-align-center\" data-align=\"center\">20<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.9913<\/td><td class=\"has-text-align-center\" data-align=\"center\">&#8211;<\/td><td class=\"has-text-align-center\" data-align=\"center\">&#8211;<\/td><td class=\"has-text-align-center\" data-align=\"center\">105<\/td><td class=\"has-text-align-center\" data-align=\"center\">9<\/td><td class=\"has-text-align-center\" data-align=\"center\">93<\/td><td class=\"has-text-align-center\" data-align=\"center\">8<\/td><td class=\"has-text-align-center\" data-align=\"center\">85<\/td><td class=\"has-text-align-center\" data-align=\"center\">10<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Chlorfenapyr (GC)*<\/td><td class=\"has-text-align-center\" data-align=\"center\">&lt;LOD<\/td><td class=\"has-text-align-center\" data-align=\"center\">10<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.9924<\/td><td class=\"has-text-align-center\" data-align=\"center\">97<\/td><td class=\"has-text-align-center\" data-align=\"center\">9<\/td><td class=\"has-text-align-center\" data-align=\"center\">90<\/td><td class=\"has-text-align-center\" data-align=\"center\">19<\/td><td class=\"has-text-align-center\" data-align=\"center\">89<\/td><td class=\"has-text-align-center\" data-align=\"center\">6<\/td><td class=\"has-text-align-center\" data-align=\"center\">85<\/td><td class=\"has-text-align-center\" data-align=\"center\">13<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Cyfluthrin (GC)<\/td><td class=\"has-text-align-center\" data-align=\"center\">1000<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.9935<\/td><td class=\"has-text-align-center\" data-align=\"center\">107<\/td><td class=\"has-text-align-center\" data-align=\"center\">10<\/td><td class=\"has-text-align-center\" data-align=\"center\">92<\/td><td class=\"has-text-align-center\" data-align=\"center\">18<\/td><td class=\"has-text-align-center\" data-align=\"center\">91<\/td><td class=\"has-text-align-center\" data-align=\"center\">7<\/td><td class=\"has-text-align-center\" data-align=\"center\">89<\/td><td class=\"has-text-align-center\" data-align=\"center\">11<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Cypermethrin (GC)<\/td><td class=\"has-text-align-center\" data-align=\"center\">1000<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.9938<\/td><td class=\"has-text-align-center\" data-align=\"center\">95<\/td><td class=\"has-text-align-center\" data-align=\"center\">9<\/td><td class=\"has-text-align-center\" data-align=\"center\">83<\/td><td class=\"has-text-align-center\" data-align=\"center\">17<\/td><td class=\"has-text-align-center\" data-align=\"center\">97<\/td><td class=\"has-text-align-center\" data-align=\"center\">15<\/td><td class=\"has-text-align-center\" data-align=\"center\">93<\/td><td class=\"has-text-align-center\" data-align=\"center\">9<\/td><\/tr><\/tbody><\/table><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\">*Category I pesticides, LOQ \u2264100 ng\/g<br><sup>a<\/sup>Spinosad- spinosyn A: Conc. 1: 7 ng\/g; Conc. 2: 35.5 ng\/g; Conc. 3: 71 ng\/g; Conc. 4: 355 ng\/g<br><sup>b<\/sup>Spinosad &#8211; spinosyn D: Conc. 1: 3 ng\/g; conc. 2: 14.5 ng\/g; Conc. 3: 29 ng\/g; Conc. 4: 145 ng\/g<br><sup>c<\/sup>Spinetoram &#8211; spinosyn J: Conc. 1: 8 ng\/g; Conc. 2: 40 ng\/g; Conc. 3: 80 ng\/g; Conc. 4: 400 ng\/g<br><sup>d<\/sup>Spinetoram &#8211; spinosyn L: Conc. 1: 2 ng\/g; Conc. 2: 10 ng\/g; Conc. 3: 20 ng\/g; Conc. 4: 100 ng\/g<br><sup>e<\/sup>Pyrethrin I: Conc. 1: 5 ng\/g; Conc. 2: 27 ng\/g; Conc. 3: 54 ng\/g; Conc. 4: 270 ng\/g<br><sup>f<\/sup>Pyrethrin II: Conc. 1: 3 ng\/g; Conc. 2: 17 ng\/g; Conc. 3: 34 ng\/g; Conc. 4: 170 ng\/g<br><sup>g<\/sup>Permethrin-<em>cis<\/em>: Conc. 1: 4 ng\/g; Conc. 2: 20.5 ng\/g; Conc. 3: 41 ng\/g; Conc. 4: 205 ng\/g<br><sup>h<\/sup>Permethrin-<em>trans<\/em>: Conc. 1: 6 ng\/g; Conc. 2: 29.5 ng\/g; Conc. 3: 59 ng\/g; Conc. 4: 295 ng\/g<br><sup>#<\/sup>Total of aflatoxin B1, B2, G1, and G2 should not exceed 20 ng\/g.<br><sup>\u00a5<\/sup>Total spinosad should not exceed 3000 ng\/g.<br><sup>\u00a7<\/sup>Total spinetoram should not exceed 3000 ng\/g.<br><sup>\u00a3<\/sup>&nbsp;Total pyrethrins should not exceed 1000 ng\/g.<br><sup>\u00a2<\/sup>Total permethrins should not exceed 20,000 ng\/g.<\/p>\n\n\n\n<div style=\"height:20px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<p class=\"wp-block-paragraph\"><strong>Tabella XI:<\/strong>&nbsp;Linearit\u00e0 per i cannabinoidi nelle caramelle gommose alla cannabis.<\/p>\n\n\n\n<figure class=\"wp-block-table\"><table class=\"has-fixed-layout\"><tbody><tr><td><strong>Cannabinoidi<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>Tempo di ritenzione<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>R<sup>2<\/sup><\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>Equazione<\/strong><\/td><\/tr><tr><td>Cannabidiolic acid (CBDA)<\/td><td class=\"has-text-align-center\" data-align=\"center\">2.142<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.9993<\/td><td class=\"has-text-align-center\" data-align=\"center\">y = 2.08e+004x + 1.03e+003<\/td><\/tr><tr><td>Cannabigerol (CBG)<\/td><td class=\"has-text-align-center\" data-align=\"center\">2.405<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.9981<\/td><td class=\"has-text-align-center\" data-align=\"center\">y = 1.16e+004x + 1.85e+003<\/td><\/tr><tr><td>Cannabidiol (CBD)<\/td><td class=\"has-text-align-center\" data-align=\"center\">2.535<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.9972<\/td><td class=\"has-text-align-center\" data-align=\"center\">y = 1.17e+004x + 1.42e+003<\/td><\/tr><tr><td>Cannabinol (CBN)<\/td><td class=\"has-text-align-center\" data-align=\"center\">3.776<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.9980<\/td><td class=\"has-text-align-center\" data-align=\"center\">y = 2.70e+004x + 6.14e+004<\/td><\/tr><tr><td>Delta-9-tetrahydrocannabinol (Delta-9-THC)<\/td><td class=\"has-text-align-center\" data-align=\"center\">4.753<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.9970<\/td><td class=\"has-text-align-center\" data-align=\"center\">y = 1.06e+004x + 7.29e+003<\/td><\/tr><tr><td>Tetrahydrocannabinolic acid (THCA)<\/td><td class=\"has-text-align-center\" data-align=\"center\">6.279<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.9986<\/td><td class=\"has-text-align-center\" data-align=\"center\">y = 1.78e+004x &#8211; 1.28e+003<\/td><\/tr><\/tbody><\/table><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\"><strong>Tabella XII:<\/strong>&nbsp;Accuratezza e precisione per i cannabinoidi nelle caramelle alla cannabis.<\/p>\n\n\n\n<figure class=\"wp-block-table\"><table class=\"has-fixed-layout\"><tbody><tr><td><strong>Cannabinoide\/Livello di&nbsp;fortificazione<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\" colspan=\"2\"><strong>Conc. estratto diluito (ppm)<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\" rowspan=\"2\"><strong>Media (ppm)<\/strong><\/td><td rowspan=\"2\"><strong>SD<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\" rowspan=\"2\"><strong>RSD (%)<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\" rowspan=\"2\"><strong>Conc. estratto non diluito (ppm)<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\" rowspan=\"2\"><strong>Conc. stimata del campione (mg\/g)<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\" rowspan=\"2\"><strong>Accuracy (%)<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\" rowspan=\"2\"><strong>Errore percentuale<\/strong><\/td><\/tr><tr><td><strong>Caramelle fortificate a 0,2 mg\/g<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>Replica 1<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>Replica 2<\/strong><\/td><\/tr><tr><td>Cannabidiolic acid (CBDA)<\/td><td class=\"has-text-align-center\" data-align=\"center\">4.2<\/td><td class=\"has-text-align-center\" data-align=\"center\">4.4<\/td><td>4.3<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><td class=\"has-text-align-center\" data-align=\"center\">43<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.2<\/td><td class=\"has-text-align-center\" data-align=\"center\">107<\/td><td>7<\/td><\/tr><tr><td>Cannabigerol (CBG)<\/td><td class=\"has-text-align-center\" data-align=\"center\">4.0<\/td><td class=\"has-text-align-center\" data-align=\"center\">3.9<\/td><td>4.0<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">2<\/td><td class=\"has-text-align-center\" data-align=\"center\">40<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.2<\/td><td class=\"has-text-align-center\" data-align=\"center\">99<\/td><td>1<\/td><\/tr><tr><td>Cannabidiol (CBD)<\/td><td class=\"has-text-align-center\" data-align=\"center\">4.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">4.2<\/td><td>4.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">2<\/td><td class=\"has-text-align-center\" data-align=\"center\">41<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.2<\/td><td class=\"has-text-align-center\" data-align=\"center\">103<\/td><td>3<\/td><\/tr><tr><td>Cannabinol (CBN)<\/td><td class=\"has-text-align-center\" data-align=\"center\">4.0<\/td><td class=\"has-text-align-center\" data-align=\"center\">4.1<\/td><td>4.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><td class=\"has-text-align-center\" data-align=\"center\">41<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.2<\/td><td class=\"has-text-align-center\" data-align=\"center\">101<\/td><td>1<\/td><\/tr><tr><td>Delta-9 tetrahydrocannabinol (Delta 9 THC)<\/td><td class=\"has-text-align-center\" data-align=\"center\">4.0<\/td><td class=\"has-text-align-center\" data-align=\"center\">4.2<\/td><td>4.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><td class=\"has-text-align-center\" data-align=\"center\">41<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.2<\/td><td class=\"has-text-align-center\" data-align=\"center\">103<\/td><td>3<\/td><\/tr><tr><td>Tetrahydrocannabinolic acid (THCA)<\/td><td class=\"has-text-align-center\" data-align=\"center\">4.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">4.3<\/td><td>4.2<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><td class=\"has-text-align-center\" data-align=\"center\">42<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.2<\/td><td class=\"has-text-align-center\" data-align=\"center\">105<\/td><td>5<\/td><\/tr><tr><td>&nbsp;<\/td><td class=\"has-text-align-center\" data-align=\"center\">&nbsp;<\/td><td class=\"has-text-align-center\" data-align=\"center\">&nbsp;<\/td><td>&nbsp;<\/td><td class=\"has-text-align-center\" data-align=\"center\">&nbsp;<\/td><td class=\"has-text-align-center\" data-align=\"center\">&nbsp;<\/td><td class=\"has-text-align-center\" data-align=\"center\">&nbsp;<\/td><td class=\"has-text-align-center\" data-align=\"center\">&nbsp;<\/td><td class=\"has-text-align-center\" data-align=\"center\">&nbsp;<\/td><td>&nbsp;<\/td><\/tr><tr><td><strong>Gummy spiked at 0.5 mg\/g<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\">&nbsp;<\/td><td class=\"has-text-align-center\" data-align=\"center\">&nbsp;<\/td><td>&nbsp;<\/td><td class=\"has-text-align-center\" data-align=\"center\">&nbsp;<\/td><td class=\"has-text-align-center\" data-align=\"center\">&nbsp;<\/td><td class=\"has-text-align-center\" data-align=\"center\">&nbsp;<\/td><td class=\"has-text-align-center\" data-align=\"center\">&nbsp;<\/td><td class=\"has-text-align-center\" data-align=\"center\">&nbsp;<\/td><td>&nbsp;<\/td><\/tr><tr><td>Cannabidiolic acid (CBDA)<\/td><td class=\"has-text-align-center\" data-align=\"center\">10.4<\/td><td class=\"has-text-align-center\" data-align=\"center\">10.7<\/td><td>10.6<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.2<\/td><td class=\"has-text-align-center\" data-align=\"center\">2<\/td><td class=\"has-text-align-center\" data-align=\"center\">106<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.5<\/td><td class=\"has-text-align-center\" data-align=\"center\">106<\/td><td>6<\/td><\/tr><tr><td>Cannabigerol (CBG)<\/td><td class=\"has-text-align-center\" data-align=\"center\">9.5<\/td><td class=\"has-text-align-center\" data-align=\"center\">10.3<\/td><td>9.9<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.5<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><td class=\"has-text-align-center\" data-align=\"center\">99<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.5<\/td><td class=\"has-text-align-center\" data-align=\"center\">99<\/td><td>1<\/td><\/tr><tr><td>Cannabidiol (CBD)<\/td><td class=\"has-text-align-center\" data-align=\"center\">9.8<\/td><td class=\"has-text-align-center\" data-align=\"center\">10.6<\/td><td>10.2<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.5<\/td><td class=\"has-text-align-center\" data-align=\"center\">5<\/td><td class=\"has-text-align-center\" data-align=\"center\">102<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.5<\/td><td class=\"has-text-align-center\" data-align=\"center\">102<\/td><td>2<\/td><\/tr><tr><td>Cannabinol (CBN)<\/td><td class=\"has-text-align-center\" data-align=\"center\">9.8<\/td><td class=\"has-text-align-center\" data-align=\"center\">10.3<\/td><td>10.0<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.4<\/td><td class=\"has-text-align-center\" data-align=\"center\">4<\/td><td class=\"has-text-align-center\" data-align=\"center\">100<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.5<\/td><td class=\"has-text-align-center\" data-align=\"center\">100<\/td><td>0<\/td><\/tr><tr><td>Delta-9-tetrahydrocannabinol (Delta-9-THC)<\/td><td class=\"has-text-align-center\" data-align=\"center\">9.9<\/td><td class=\"has-text-align-center\" data-align=\"center\">10.3<\/td><td>10.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.3<\/td><td class=\"has-text-align-center\" data-align=\"center\">3<\/td><td class=\"has-text-align-center\" data-align=\"center\">101<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.5<\/td><td class=\"has-text-align-center\" data-align=\"center\">101<\/td><td>1<\/td><\/tr><tr><td>Tetrahydrocannabinolic acid (THCA)<\/td><td class=\"has-text-align-center\" data-align=\"center\">10.3<\/td><td class=\"has-text-align-center\" data-align=\"center\">9.9<\/td><td>10.1<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.2<\/td><td class=\"has-text-align-center\" data-align=\"center\">2<\/td><td class=\"has-text-align-center\" data-align=\"center\">101<\/td><td class=\"has-text-align-center\" data-align=\"center\">0.5<\/td><td class=\"has-text-align-center\" data-align=\"center\">101<\/td><td>1<\/td><\/tr><\/tbody><\/table><\/figure>\n\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><strong>Figura 2:\u00a0<\/strong>Cromatogramma LC-MS\/MS di un estratto ottenuto da bianco di caramelle gommose fortificato con pesticidi e micotossine a 100 ng\/g.<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n<div class=\"wp-block-custom-chromatogram-article\"><div class=\"wp-block-custom-chromatogram-article\"><div class=\"chromatogram-image wide-image\"><img decoding=\"async\" src=\"https:\/\/ez.restek.com\/images\/cgram\/lc_gn0666.png\" alt=\"California Cannabis Pesticides and Mycotoxins in Gummy Edibles on Raptor ARC-18\" title=\"-\"><\/div><p class=\"article-id\" style=\"text-align:center\"> LC_GN0666<\/p><div class=\"chromatogram-peaks\"><h4>Peaks<\/h4><table class=\"peaks col-12\">\n<thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 75px\">t<sub>R<\/sub> (min)<\/th><th style=\"text-align: center;width: 75px\">Precursor Ion<\/th><th style=\"text-align: center;width: 75px\">Product Ion 1<\/th><th style=\"text-align: center;width: 75px\">Product Ion 2<\/th><th style=\"text-align: center;width: 75px\">Polarity<\/th><\/tr><\/thead>\n<tbody><tr><td class=\"num\">1.<\/td><td class=\"cmpd\">Daminozide-d6<\/td><td class=\"oth\">0.7<\/td><td class=\"oth\">167.0<\/td><td class=\"oth\">149.3<\/td><td class=\"oth\">49.3<\/td><td class=\"oth\">+<\/td><\/tr>\n<tr><td class=\"num\">2.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Daminozide\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/1596-84-5\/Daminozide\" target=\"_blank\" rel=\"noopener\">Daminozide<\/a><\/td><td class=\"oth\">0.7<\/td><td class=\"oth\">161.1<\/td><td class=\"oth\">44.1<\/td><td class=\"oth\">143.2<\/td><td class=\"oth\">+<\/td><\/tr>\n<tr><td class=\"num\">3.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Acephate\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/30560-19-1\/Acephate\" target=\"_blank\" rel=\"noopener\">Acephate<\/a><\/td><td class=\"oth\">1.7<\/td><td class=\"oth\">184.0<\/td><td class=\"oth\">143.1<\/td><td class=\"oth\">95.1<\/td><td class=\"oth\">+<\/td><\/tr>\n<tr><td class=\"num\">4.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Oxamyl\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/23135-22-0\/Oxamyl\" target=\"_blank\" rel=\"noopener\">Oxamyl<\/a><\/td><td class=\"oth\">2.0<\/td><td class=\"oth\">237.1<\/td><td class=\"oth\">72.1<\/td><td class=\"oth\">90.1<\/td><td class=\"oth\">+<\/td><\/tr>\n<tr><td class=\"num\">5.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Flonicamid\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/158062-67-0\/Flonicamid\" target=\"_blank\" rel=\"noopener\">Flonicamid<\/a><\/td><td class=\"oth\">2.1<\/td><td class=\"oth\">230.1<\/td><td class=\"oth\">203.1<\/td><td class=\"oth\">174.1<\/td><td class=\"oth\">+<\/td><\/tr>\n<tr><td class=\"num\">6.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Methomyl\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/16752-77-5\/Methomyl\" target=\"_blank\" rel=\"noopener\">Methomyl<\/a><\/td><td class=\"oth\">2.1<\/td><td class=\"oth\">163.1<\/td><td class=\"oth\">88.1<\/td><td class=\"oth\">106.1<\/td><td class=\"oth\">+<\/td><\/tr>\n<tr><td class=\"num\">7.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Thiamethoxam\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/153719-23-4\/Thiamethoxam\" target=\"_blank\" rel=\"noopener\">Thiamethoxam<\/a><\/td><td class=\"oth\">2.1<\/td><td class=\"oth\">292.0<\/td><td class=\"oth\">211.1<\/td><td class=\"oth\">181.1<\/td><td class=\"oth\">+<\/td><\/tr>\n<tr><td class=\"num\">8.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Imidacloprid\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/138261-41-3\/Imidacloprid\" target=\"_blank\" rel=\"noopener\">Imidacloprid<\/a><\/td><td class=\"oth\">2.3<\/td><td class=\"oth\">256.1<\/td><td class=\"oth\">209.1<\/td><td class=\"oth\">175.1<\/td><td class=\"oth\">+<\/td><\/tr>\n<tr><td class=\"num\">9.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Mevinphos\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/7786-34-7\/Mevinphos\" target=\"_blank\" rel=\"noopener\">Mevinphos<\/a><\/td><td class=\"oth\">2.4<\/td><td class=\"oth\">225.1<\/td><td class=\"oth\">127.1<\/td><td class=\"oth\">193.2<\/td><td class=\"oth\">+<\/td><\/tr>\n<tr><td class=\"num\">10.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Acetamiprid\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/135410-20-7\/Acetamiprid\" target=\"_blank\" rel=\"noopener\">Acetamiprid<\/a><\/td><td class=\"oth\">2.4<\/td><td class=\"oth\">223.0<\/td><td class=\"oth\">126.1<\/td><td class=\"oth\">56.1<\/td><td class=\"oth\">+<\/td><\/tr>\n<tr><td class=\"num\">11.<\/td><td class=\"cmpd\">Dimethoathe-d6<\/td><td class=\"oth\">2.4<\/td><td class=\"oth\">236.1<\/td><td class=\"oth\">205.1<\/td><td class=\"oth\">&#8211;<\/td><td class=\"oth\">+<\/td><\/tr>\n<tr><td class=\"num\">12.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Dimethoate\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/60-51-5\/Dimethoate\" target=\"_blank\" rel=\"noopener\">Dimethoate<\/a><\/td><td class=\"oth\">2.4<\/td><td class=\"oth\">230.0<\/td><td class=\"oth\">199.1<\/td><td class=\"oth\">125.1<\/td><td class=\"oth\">+<\/td><\/tr>\n<tr><td class=\"num\">13.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Thiacloprid\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/111988-49-9\/Thiacloprid\" target=\"_blank\" rel=\"noopener\">Thiacloprid<\/a><\/td><td class=\"oth\">2.5<\/td><td class=\"oth\">253.0<\/td><td class=\"oth\">126.0<\/td><td class=\"oth\">90.1<\/td><td class=\"oth\">+<\/td><\/tr>\n<tr><td class=\"num\">14.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Aflatoxin G2\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/7241-98-7\/Aflatoxin G2\" target=\"_blank\" rel=\"noopener\">Aflatoxin G2<\/a><\/td><td class=\"oth\">2.5<\/td><td class=\"oth\">331.2<\/td><td class=\"oth\">189.3<\/td><td class=\"oth\">115.2<\/td><td class=\"oth\">+<\/td><\/tr>\n<tr><td class=\"num\">15.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Aflatoxin G1\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/1165-39-5\/Aflatoxin G1\" target=\"_blank\" rel=\"noopener\">Aflatoxin G1<\/a><\/td><td class=\"oth\">2.5<\/td><td class=\"oth\">329.2<\/td><td class=\"oth\">243.2<\/td><td class=\"oth\">215.3<\/td><td class=\"oth\">+<\/td><\/tr>\n<tr><td class=\"num\">16.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Aldicarb\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/116-06-3\/Aldicarb\" target=\"_blank\" rel=\"noopener\">Aldicarb<\/a><\/td><td class=\"oth\">2.6<\/td><td class=\"oth\">116.0<\/td><td class=\"oth\">89.2<\/td><td class=\"oth\">70.2<\/td><td class=\"oth\">+<\/td><\/tr>\n<tr><td class=\"num\">17.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Aflatoxin B2\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/7220-81-7\/Aflatoxin B2\" target=\"_blank\" rel=\"noopener\">Aflatoxin B2<\/a><\/td><td class=\"oth\">2.6<\/td><td class=\"oth\">315.3<\/td><td class=\"oth\">287.2<\/td><td class=\"oth\">243.3<\/td><td class=\"oth\">+<\/td><\/tr>\n<tr><td class=\"num\">18.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Dichlorvos\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/62-73-7\/Dichlorvos\" target=\"_blank\" rel=\"noopener\">Dichlorvos<\/a><\/td><td class=\"oth\">2.7<\/td><td class=\"oth\">220.9<\/td><td class=\"oth\">109.1<\/td><td class=\"oth\">79.2<\/td><td class=\"oth\">+<\/td><\/tr>\n<tr><td class=\"num\">19.<\/td><td class=\"cmpd\">Dichlorvos-d6<\/td><td class=\"oth\">2.7<\/td><td class=\"oth\">227.0<\/td><td class=\"oth\">115.1<\/td><td class=\"oth\">&#8211;<\/td><td class=\"oth\">+<\/td><\/tr>\n<tr><td class=\"num\">20.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Aflatoxin B1\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/1162-65-8\/Aflatoxin B1\" target=\"_blank\" rel=\"noopener\">Aflatoxin B1<\/a><\/td><td class=\"oth\">2.7<\/td><td class=\"oth\">313.2<\/td><td class=\"oth\">241.2<\/td><td class=\"oth\">128.2<\/td><td class=\"oth\">+<\/td><\/tr>\n<tr><td class=\"num\">21.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Imazalil\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/35554-44-0\/Imazalil\" target=\"_blank\" rel=\"noopener\">Imazalil<\/a><\/td><td class=\"oth\">2.7<\/td><td class=\"oth\">297.0<\/td><td class=\"oth\">159.0<\/td><td class=\"oth\">201.0<\/td><td class=\"oth\">+<\/td><\/tr>\n<tr><td class=\"num\">22.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Carbofuran\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/1563-66-2\/Carbofuran\" target=\"_blank\" rel=\"noopener\">Carbofuran<\/a><\/td><td class=\"oth\">2.7<\/td><td class=\"oth\">222.1<\/td><td class=\"oth\">123.1<\/td><td class=\"oth\">165.2<\/td><td class=\"oth\">+<\/td><\/tr>\n<tr><td class=\"num\">23.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Propoxur\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/114-26-1\/Propoxur\" target=\"_blank\" rel=\"noopener\">Propoxur<\/a><\/td><td class=\"oth\">2.7<\/td><td class=\"oth\">210.1<\/td><td class=\"oth\">111.1<\/td><td class=\"oth\">93.1<\/td><td class=\"oth\">+<\/td><\/tr>\n<tr><td class=\"num\">24.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Carbaryl-d7\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/362049-56-7\/Carbaryl-d7\" target=\"_blank\" rel=\"noopener\">Carbaryl-d7<\/a><\/td><td class=\"oth\">2.8<\/td><td class=\"oth\">209.2<\/td><td class=\"oth\">152.2<\/td><td class=\"oth\">&#8211;<\/td><td class=\"oth\">+<\/td><\/tr>\n<tr><td class=\"num\">25.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Carbaryl\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/63-25-2\/Carbaryl\" target=\"_blank\" rel=\"noopener\">Carbaryl<\/a><\/td><td class=\"oth\">2.8<\/td><td class=\"oth\">202.1<\/td><td class=\"oth\">145.1<\/td><td class=\"oth\">127.1<\/td><td class=\"oth\">+<\/td><\/tr>\n<tr><td class=\"num\">26.<\/td><td class=\"cmpd\">Diuron-d6<\/td><td class=\"oth\">3.0<\/td><td class=\"oth\">239.1<\/td><td class=\"oth\">78.2<\/td><td class=\"oth\">&#8211;<\/td><td class=\"oth\">+<\/td><\/tr>\n<tr><td class=\"num\">27.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Atrazine-d5\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/163165-75-1\/Atrazine-d5\" target=\"_blank\" rel=\"noopener\">Atrazine-d5<\/a><\/td><td class=\"oth\">3.0<\/td><td class=\"oth\">221.2<\/td><td class=\"oth\">179.1<\/td><td class=\"oth\">&#8211;<\/td><td class=\"oth\">+<\/td><\/tr>\n<tr><td class=\"num\">28.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Naled\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/300-76-5\/Naled\" target=\"_blank\" rel=\"noopener\">Naled<\/a><\/td><td class=\"oth\">3.1<\/td><td class=\"oth\">397.8<\/td><td class=\"oth\">127.1<\/td><td class=\"oth\">109.1<\/td><td class=\"oth\">+<\/td><\/tr>\n<tr><td class=\"num\">29.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Metalaxyl\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/57837-19-1\/Metalaxyl\" target=\"_blank\" rel=\"noopener\">Metalaxyl<\/a><\/td><td class=\"oth\">3.1<\/td><td class=\"oth\">280.2<\/td><td class=\"oth\">220.2<\/td><td class=\"oth\">192.2<\/td><td class=\"oth\">+<\/td><\/tr>\n<tr><td class=\"num\">30.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Spiroxamine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/118134-30-8\/Spiroxamine\" target=\"_blank\" rel=\"noopener\">Spiroxamine<\/a><\/td><td class=\"oth\">3.1<\/td><td class=\"oth\">298.3<\/td><td class=\"oth\">144.2<\/td><td class=\"oth\">100.2<\/td><td class=\"oth\">+<\/td><\/tr>\n<tr><td class=\"num\">31.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Chlorantraniliprole\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/500008-45-7\/Chlorantraniliprole\" target=\"_blank\" rel=\"noopener\">Chlorantraniliprole<\/a><\/td><td class=\"oth\">3.2<\/td><td class=\"oth\">483.9<\/td><td class=\"oth\">452.9<\/td><td class=\"oth\">285.9<\/td><td class=\"oth\">+<\/td><\/tr>\n<tr><td class=\"num\">32.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Phosmet\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/732-11-6\/Phosmet\" target=\"_blank\" rel=\"noopener\">Phosmet<\/a><\/td><td class=\"oth\">3.2<\/td><td class=\"oth\">318.0<\/td><td class=\"oth\">160.1<\/td><td class=\"oth\">77.2<\/td><td class=\"oth\">+<\/td><\/tr>\n<tr><td class=\"num\">33.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Azoxystrobin\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/131860-33-8\/Azoxystrobin\" target=\"_blank\" rel=\"noopener\">Azoxystrobin<\/a><\/td><td class=\"oth\">3.3<\/td><td class=\"oth\">404.0<\/td><td class=\"oth\">372.1<\/td><td class=\"oth\">344.1<\/td><td class=\"oth\">+<\/td><\/tr>\n<tr><td class=\"num\">34.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Linuron-d6\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/1219804-76-8\/Linuron-d6\" target=\"_blank\" rel=\"noopener\">Linuron-d6<\/a><\/td><td class=\"oth\">3.3<\/td><td class=\"oth\">255.1<\/td><td class=\"oth\">160.1<\/td><td class=\"oth\">&#8211;<\/td><td class=\"oth\">+<\/td><\/tr>\n<tr><td class=\"num\">35.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Fludioxonil\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/131341-86-1\/Fludioxonil\" target=\"_blank\" rel=\"noopener\">Fludioxonil<\/a><\/td><td class=\"oth\">3.4<\/td><td class=\"oth\">247.0<\/td><td class=\"oth\">180.0<\/td><td class=\"oth\">126.0<\/td><td class=\"oth\">&#8211;<\/td><\/tr>\n<tr><td class=\"num\">36.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Methiocarb\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/2032-65-7\/Methiocarb\" target=\"_blank\" rel=\"noopener\">Methiocarb<\/a><\/td><td class=\"oth\">3.4<\/td><td class=\"oth\">226.1<\/td><td class=\"oth\">169.1<\/td><td class=\"oth\">121.1<\/td><td class=\"oth\">+<\/td><\/tr>\n<tr><td class=\"num\">37.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Dimethomorph\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/110488-70-5\/Dimethomorph\" target=\"_blank\" rel=\"noopener\">Dimethomorph<\/a><\/td><td class=\"oth\">3.5<\/td><td class=\"oth\">388.2<\/td><td class=\"oth\">301.2<\/td><td class=\"oth\">165.3<\/td><td class=\"oth\">+<\/td><\/tr>\n<tr><td class=\"num\">38.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Boscalid\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/188425-85-6\/Boscalid\" target=\"_blank\" rel=\"noopener\">Boscalid<\/a><\/td><td class=\"oth\">3.5<\/td><td class=\"oth\">342.9<\/td><td class=\"oth\">307.1<\/td><td class=\"oth\">140.1<\/td><td class=\"oth\">+<\/td><\/tr>\n<tr><td class=\"num\">39.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Paclobutrazol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/76738-62-0\/Paclobutrazol\" target=\"_blank\" rel=\"noopener\">Paclobutrazol<\/a><\/td><td class=\"oth\">3.6<\/td><td class=\"oth\">294.3<\/td><td class=\"oth\">70.1<\/td><td class=\"oth\">125.1<\/td><td class=\"oth\">+<\/td><\/tr>\n<tr><td class=\"num\">40.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Malathion\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/121-75-5\/Malathion\" target=\"_blank\" rel=\"noopener\">Malathion<\/a><\/td><td class=\"oth\">3.6<\/td><td class=\"oth\">331.0<\/td><td class=\"oth\">127.2<\/td><td class=\"oth\">285.2<\/td><td class=\"oth\">+<\/td><\/tr>\n<tr><td class=\"num\">41.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Myclobutanil\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/88671-89-0\/Myclobutanil\" target=\"_blank\" rel=\"noopener\">Myclobutanil<\/a><\/td><td class=\"oth\">3.7<\/td><td class=\"oth\">289.1<\/td><td class=\"oth\">70.1<\/td><td class=\"oth\">125.1<\/td><td class=\"oth\">+<\/td><\/tr>\n<tr><td class=\"num\">42.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Bifenazate\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/149877-41-8\/Bifenazate\" target=\"_blank\" rel=\"noopener\">Bifenazate<\/a><\/td><td class=\"oth\">3.7<\/td><td class=\"oth\">301.0<\/td><td class=\"oth\">198.1<\/td><td class=\"oth\">170.2<\/td><td class=\"oth\">+<\/td><\/tr>\n<tr><td class=\"num\">43.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Ochratoxin A\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/303-47-9\/Ochratoxin A\" target=\"_blank\" rel=\"noopener\">Ochratoxin A<\/a><\/td><td class=\"oth\">3.8<\/td><td class=\"oth\">404.2<\/td><td class=\"oth\">239.1<\/td><td class=\"oth\">358.3<\/td><td class=\"oth\">+<\/td><\/tr>\n<tr><td class=\"num\">44.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Fenhexamid\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/126833-17-8\/Fenhexamid\" target=\"_blank\" rel=\"noopener\">Fenhexamid<\/a><\/td><td class=\"oth\">3.9<\/td><td class=\"oth\">302.1<\/td><td class=\"oth\">97.1<\/td><td class=\"oth\">55.2<\/td><td class=\"oth\">+<\/td><\/tr>\n<tr><td class=\"num\">45.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Spirotetramat\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/203313-25-1\/Spirotetramat\" target=\"_blank\" rel=\"noopener\">Spirotetramat<\/a><\/td><td class=\"oth\">4.0<\/td><td class=\"oth\">374.2<\/td><td class=\"oth\">302.1<\/td><td class=\"oth\">216.1<\/td><td class=\"oth\">+<\/td><\/tr>\n<tr><td class=\"num\">46.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Ethoprophos\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/13194-48-4\/Ethoprophos\" target=\"_blank\" rel=\"noopener\">Ethoprophos<\/a><\/td><td class=\"oth\">4.1<\/td><td class=\"oth\">243.1<\/td><td class=\"oth\">131.1<\/td><td class=\"oth\">97.1<\/td><td class=\"oth\">+<\/td><\/tr>\n<tr><td class=\"num\">47.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Fipronil\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/120068-37-3\/Fipronil\" target=\"_blank\" rel=\"noopener\">Fipronil<\/a><\/td><td class=\"oth\">4.1<\/td><td class=\"oth\">436.8<\/td><td class=\"oth\">331.8<\/td><td class=\"oth\">251.9<\/td><td class=\"oth\">&#8211;<\/td><\/tr>\n<tr><td class=\"num\">48.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Fenoxycarb\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/72490-01-8\/Fenoxycarb\" target=\"_blank\" rel=\"noopener\">Fenoxycarb<\/a><\/td><td class=\"oth\">4.2<\/td><td class=\"oth\">302.1<\/td><td class=\"oth\">88.1<\/td><td class=\"oth\">116.1<\/td><td class=\"oth\">+<\/td><\/tr>\n<tr><td class=\"num\">49.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Kresoxim-methyl\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/143390-89-0\/Kresoxim-methyl\" target=\"_blank\" rel=\"noopener\">Kresoxim-methyl<\/a><\/td><td class=\"oth\">4.4<\/td><td class=\"oth\">314.2<\/td><td class=\"oth\">267.2<\/td><td class=\"oth\">222.2<\/td><td class=\"oth\">+<\/td><\/tr>\n<tr><td class=\"num\">50.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Tebuconazole\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/107534-96-3\/Tebuconazole\" target=\"_blank\" rel=\"noopener\">Tebuconazole<\/a><\/td><td class=\"oth\">4.4<\/td><td class=\"oth\">308.1<\/td><td class=\"oth\">70.1<\/td><td class=\"oth\">125.1<\/td><td class=\"oth\">+<\/td><\/tr>\n<tr><td class=\"num\">51.<\/td><td class=\"cmpd\">Diazinon-d10 <\/td><td class=\"oth\">4.6<\/td><td class=\"oth\">315.2<\/td><td class=\"oth\">170.2<\/td><td class=\"oth\">&#8211;<\/td><td class=\"oth\">+<\/td><\/tr>\n<tr><td class=\"num\">52.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Spinosyn A (Spinosad)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/168316-95-8\/Spinosyn A\" target=\"_blank\" rel=\"noopener\">Spinosyn A (Spinosad)<\/a><\/td><td class=\"oth\">4.6<\/td><td class=\"oth\">732.4<\/td><td class=\"oth\">142.2<\/td><td class=\"oth\">98.1<\/td><td class=\"oth\">+<\/td><\/tr>\n<tr><td class=\"num\">53.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Diazinon\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/333-41-5\/Diazinon\" target=\"_blank\" rel=\"noopener\">Diazinon<\/a><\/td><td class=\"oth\">4.6<\/td><td class=\"oth\">305.1<\/td><td class=\"oth\">169.2<\/td><td class=\"oth\">153.2<\/td><td class=\"oth\">+<\/td><\/tr>\n<tr><td class=\"num\">54.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Coumaphos\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/56-72-4\/Coumaphos\" target=\"_blank\" rel=\"noopener\">Coumaphos<\/a><\/td><td class=\"oth\">4.7<\/td><td class=\"oth\">363.1<\/td><td class=\"oth\">227.1<\/td><td class=\"oth\">307.1<\/td><td class=\"oth\">+<\/td><\/tr>\n<tr><td class=\"num\">55.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Pyridaben\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/96489-71-3\/Pyridaben\" target=\"_blank\" rel=\"noopener\">Pyridaben<\/a><\/td><td class=\"oth\">4.7<\/td><td class=\"oth\">365.1<\/td><td class=\"oth\">309.2<\/td><td class=\"oth\">147.2<\/td><td class=\"oth\">+<\/td><\/tr>\n<tr><td class=\"num\">56.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Propiconazole\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/60207-90-1\/Propiconazole\" target=\"_blank\" rel=\"noopener\">Propiconazole<\/a><\/td><td class=\"oth\">4.7<\/td><td class=\"oth\">342.0<\/td><td class=\"oth\">159.0<\/td><td class=\"oth\">69.2<\/td><td class=\"oth\">+<\/td><\/tr>\n<tr><td class=\"num\">57.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Clofentezine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/74115-24-5\/Clofentezine\" target=\"_blank\" rel=\"noopener\">Clofentezine<\/a><\/td><td class=\"oth\">4.8<\/td><td class=\"oth\">303.0<\/td><td class=\"oth\">138.1<\/td><td class=\"oth\">102.1<\/td><td class=\"oth\">+<\/td><\/tr>\n<tr><td class=\"num\">58.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Spinosyn D (Spinosad)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/168316-95-8\/Spinosyn D\" target=\"_blank\" rel=\"noopener\">Spinosyn D (Spinosad)<\/a><\/td><td class=\"oth\">5.0<\/td><td class=\"oth\">746.5<\/td><td class=\"oth\">142.3<\/td><td class=\"oth\">98.4<\/td><td class=\"oth\">+<\/td><\/tr>\n<tr><td class=\"num\">59.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Spinosyn J (Spinetoram)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/935545-74-7\/Spinosyn J\" target=\"_blank\" rel=\"noopener\">Spinosyn J (Spinetoram)<\/a><\/td><td class=\"oth\">5.1<\/td><td class=\"oth\">748.5<\/td><td class=\"oth\">142.3<\/td><td class=\"oth\">98.3<\/td><td class=\"oth\">+<\/td><\/tr>\n<tr><td class=\"num\">60.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Trifloxystrobin\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/141517-21-7\/Trifloxystrobin\" target=\"_blank\" rel=\"noopener\">Trifloxystrobin<\/a><\/td><td class=\"oth\">5.3<\/td><td class=\"oth\">409.2<\/td><td class=\"oth\">186.1<\/td><td class=\"oth\">145.1<\/td><td class=\"oth\">+<\/td><\/tr>\n<tr><td class=\"num\">61.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Prallethrin\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/23031-36-9\/Prallethrin\" target=\"_blank\" rel=\"noopener\">Prallethrin<\/a><\/td><td class=\"oth\">5.3<\/td><td class=\"oth\">301.2<\/td><td class=\"oth\">123.2<\/td><td class=\"oth\">105.2<\/td><td class=\"oth\">+<\/td><\/tr>\n<tr><td class=\"num\">62.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Pyrethrin II\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/121-29-9\/Pyrethrin II\" target=\"_blank\" rel=\"noopener\">Pyrethrin II<\/a><\/td><td class=\"oth\">5.5<\/td><td class=\"oth\">373.1<\/td><td class=\"oth\">161.1<\/td><td class=\"oth\">133.2<\/td><td class=\"oth\">+<\/td><\/tr>\n<tr><td class=\"num\">63.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Spinosyn L (Spinetoram)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/935545-74-7\/Spinosyn L\" target=\"_blank\" rel=\"noopener\">Spinosyn L (Spinetoram)<\/a><\/td><td class=\"oth\">5.6<\/td><td class=\"oth\">760.5<\/td><td class=\"oth\">142.2<\/td><td class=\"oth\">98.1<\/td><td class=\"oth\">+<\/td><\/tr>\n<tr><td class=\"num\">64.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Piperonyl butoxide\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/51-03-6\/Piperonyl butoxide\" target=\"_blank\" rel=\"noopener\">Piperonyl butoxide<\/a><\/td><td class=\"oth\">6.0<\/td><td class=\"oth\">356.3<\/td><td class=\"oth\">177.2<\/td><td class=\"oth\">119.2<\/td><td class=\"oth\">+<\/td><\/tr>\n<tr><td class=\"num\">65.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Chlorpyrifos\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/2921-88-2\/Chlorpyrifos\" target=\"_blank\" rel=\"noopener\">Chlorpyrifos<\/a><\/td><td class=\"oth\">6.1<\/td><td class=\"oth\">349.9<\/td><td class=\"oth\">198.0<\/td><td class=\"oth\">97.1<\/td><td class=\"oth\">+<\/td><\/tr>\n<tr><td class=\"num\">66.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Hexythiazox\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/78587-05-0\/Hexythiazox\" target=\"_blank\" rel=\"noopener\">Hexythiazox<\/a><\/td><td class=\"oth\">6.2<\/td><td class=\"oth\">353.1<\/td><td class=\"oth\">228.1<\/td><td class=\"oth\">168.1<\/td><td class=\"oth\">+<\/td><\/tr>\n<tr><td class=\"num\">67.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Etoxazole\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/153233-91-1\/Etoxazole\" target=\"_blank\" rel=\"noopener\">Etoxazole<\/a><\/td><td class=\"oth\">6.6<\/td><td class=\"oth\">360.2<\/td><td class=\"oth\">141.1<\/td><td class=\"oth\">304.2<\/td><td class=\"oth\">+<\/td><\/tr>\n<tr><td class=\"num\">68.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Spiromesifen\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/283594-90-1\/Spiromesifen\" target=\"_blank\" rel=\"noopener\">Spiromesifen<\/a><\/td><td class=\"oth\">6.7<\/td><td class=\"oth\">273.2<\/td><td class=\"oth\">255.2<\/td><td class=\"oth\">187.2<\/td><td class=\"oth\">+<\/td><\/tr>\n<tr><td class=\"num\">69.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Pyrethrin I\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/121-21-1\/Pyrethrin I\" target=\"_blank\" rel=\"noopener\">Pyrethrin I<\/a><\/td><td class=\"oth\">6.9<\/td><td class=\"oth\">329.2<\/td><td class=\"oth\">161.2<\/td><td class=\"oth\">105.2<\/td><td class=\"oth\">+<\/td><\/tr>\n<tr><td class=\"num\">70.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Cyfluthrin (qualifier)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/68359-37-5\/Cyfluthrin\" target=\"_blank\" rel=\"noopener\">Cyfluthrin (qualifier)<\/a><\/td><td class=\"oth\">6.9<\/td><td class=\"oth\">453.1<\/td><td class=\"oth\">193.2<\/td><td class=\"oth\">&#8211;<\/td><td class=\"oth\">+<\/td><\/tr>\n<tr><td class=\"num\">71.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Cyfluthrin\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/68359-37-5\/Cyfluthrin\" target=\"_blank\" rel=\"noopener\">Cyfluthrin<\/a><\/td><td class=\"oth\">6.9<\/td><td class=\"oth\">451.1<\/td><td class=\"oth\">191.2<\/td><td class=\"oth\">&#8211;<\/td><td class=\"oth\">+<\/td><\/tr>\n<tr><td class=\"num\">72.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Cypermethrin\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/52315-07-8\/Cypermethrin\" target=\"_blank\" rel=\"noopener\">Cypermethrin<\/a><\/td><td class=\"oth\">7.1<\/td><td class=\"oth\">433.1<\/td><td class=\"oth\">191.0<\/td><td class=\"oth\">416.0<\/td><td class=\"oth\">+<\/td><\/tr>\n<tr><td class=\"num\">73.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for (E)-Fenpyroximate\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/111812-58-9\/(E)-Fenpyroximate\" target=\"_blank\" rel=\"noopener\">(E)-Fenpyroximate<\/a><\/td><td class=\"oth\">7.1<\/td><td class=\"oth\">422.2<\/td><td class=\"oth\">366.1<\/td><td class=\"oth\">138.1<\/td><td class=\"oth\">+<\/td><\/tr>\n<tr><td class=\"num\">74.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for trans-Permethrin\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/52645-53-1\/trans-Permethrin\" target=\"_blank\" rel=\"noopener\">trans-Permethrin<\/a><\/td><td class=\"oth\">7.6<\/td><td class=\"oth\">408.3<\/td><td class=\"oth\">183.2<\/td><td class=\"oth\">355.1<\/td><td class=\"oth\">+<\/td><\/tr>\n<tr><td class=\"num\">75.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for cis-Permethrin\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/52645-53-1\/cis-Permethrin\" target=\"_blank\" rel=\"noopener\">cis-Permethrin<\/a><\/td><td class=\"oth\">7.9<\/td><td class=\"oth\">408.3<\/td><td class=\"oth\">183.2<\/td><td class=\"oth\">355.1<\/td><td class=\"oth\">+<\/td><\/tr>\n<tr><td class=\"num\">76.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Avermectin B1a\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/71751-41-2\/Avermectin B1a\" target=\"_blank\" rel=\"noopener\">Avermectin B1a<\/a><\/td><td class=\"oth\">7.9<\/td><td class=\"oth\">890.5<\/td><td class=\"oth\">305.4<\/td><td class=\"oth\">567.4<\/td><td class=\"oth\">+<\/td><\/tr>\n<tr><td class=\"num\">77.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Etofenprox\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/80844-07-1\/Etofenprox\" target=\"_blank\" rel=\"noopener\">Etofenprox<\/a><\/td><td class=\"oth\">8.0<\/td><td class=\"oth\">394.3<\/td><td class=\"oth\">177.2<\/td><td class=\"oth\">359.3<\/td><td class=\"oth\">+<\/td><\/tr>\n<tr><td class=\"num\">78.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Bifenthrin\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/82657-04-3\/Bifenthrin\" target=\"_blank\" rel=\"noopener\">Bifenthrin<\/a><\/td><td class=\"oth\">8.2<\/td><td class=\"oth\">440.0<\/td><td class=\"oth\">181.2<\/td><td class=\"oth\">166.2<\/td><td class=\"oth\">+<\/td><\/tr>\n<tr><td class=\"num\">79.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Acequinocyl (precursor ion 1)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/57960-19-7\/Acequinocyl\" target=\"_blank\" rel=\"noopener\">Acequinocyl (precursor ion 1)<\/a><\/td><td class=\"oth\">9.4<\/td><td class=\"oth\">402.3<\/td><td class=\"oth\">343.2<\/td><td class=\"oth\">189.0<\/td><td class=\"oth\">+<\/td><\/tr>\n<tr><td class=\"num\">80.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Acequinocyl (precursor ion 2)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/57960-19-7\/Acequinocyl\" target=\"_blank\" rel=\"noopener\">Acequinocyl (precursor ion 2)<\/a><\/td><td class=\"oth\">9.4<\/td><td class=\"oth\">386.0<\/td><td class=\"oth\">344.2<\/td><td class=\"oth\">189.1<\/td><td class=\"oth\">+<\/td><\/tr>\n<\/tbody><\/table><\/div><div class=\"chromatogram-conditions\"><h4>Conditions<\/h4><div class=\"conditions-container container-fluid\"><div class=\"row\"><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Column<\/th><td>Raptor ARC-18  (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/9314A12\" rel=\"noopener\">cat.# 9314A12<\/a>)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Dimensions:<\/th><td>100 mm x 2.1 mm ID<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Particle Size:<\/th><td>2.7 \u00b5m<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Pore Size:<\/th><td>90 \u00c5<\/td><\/tr><tr><td><\/td><tr><th class=\"sub conditions_header\" scope=\"row\">Guard Column:<\/th><td>Raptor ARC-18 EXP guard column cartridge 5 mm, 2.1 mm ID, 2.7 \u00b5m (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/9314A0252\" rel=\"noopener\">cat.# 9314A0252<\/a>)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Temp.:<\/th><td>40 \u00b0C<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Standard\/Sample<\/th><td><\/td><\/tr><tr><td><\/td><td>California pesticide standard #1 (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/34124\" rel=\"noopener\">cat.# 34124<\/a>)<\/td><\/tr><tr><td><\/td><td>California pesticide standard #2 (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/34125\" rel=\"noopener\">cat.# 34125<\/a>)<\/td><\/tr><tr><td><\/td><td>California pesticide standard #3 (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/34126\" rel=\"noopener\">cat.# 34126<\/a>)<\/td><\/tr><tr><td><\/td><td>California pesticide standard #4 (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/34127\" rel=\"noopener\">cat.# 34127<\/a>)<\/td><\/tr><tr><td><\/td><td>California pesticide standard #5 (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/34128\" rel=\"noopener\">cat.# 34128<\/a>)<\/td><\/tr><tr><td><\/td><td>California pesticide standard #6 (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/34129\" rel=\"noopener\">cat.# 34129<\/a>)<\/td><\/tr><tr><td><\/td><td>Dimethoate-d6 (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/31988\" rel=\"noopener\">cat.# 31988<\/a>)<\/td><\/tr><tr><td><\/td><td>Dichlorvos-d6 (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/31987\" rel=\"noopener\">cat.# 31987<\/a>)<\/td><\/tr><tr><td><\/td><td>Carbaryl-d7 (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/31985\" rel=\"noopener\">cat.# 31985<\/a>)<\/td><\/tr><tr><td><\/td><td>Diazinon-d10 (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/31986\" rel=\"noopener\">cat.# 31986<\/a>)<\/td><\/tr><tr><td><\/td><td>Atrazine-d5 (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/31984\" rel=\"noopener\">cat.# 31984<\/a>)<\/td><\/tr><tr><td><\/td><td>Diuron-d6 (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/31989\" rel=\"noopener\">cat.# 31989<\/a>)<\/td><\/tr><tr><td><\/td><td>Liuron-d6 (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/31990\" rel=\"noopener\">cat.# 31990<\/a>)<\/td><\/tr><tr><td><\/td><td>Aflatoxins standard (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/34121\" rel=\"noopener\">cat.# 34121<\/a>)<\/td><\/tr><tr><td><\/td><td>Ochratoxin A (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/34122\" rel=\"noopener\">cat.# 34122<\/a>)<\/td><\/tr><tr><td><\/td><td>Compounds not present in these mixes were obtained separately.<\/td><\/tr>\n<tr><th class=\"sub conditions_header\" scope=\"row\">Diluent:<\/th><td>75:25 Acetonitrile:water<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Conc.:<\/th><td>3.75-15 ng\/mL (Expected concentration range in extract of gummy initially spiked at 100 ng\/g.)<\/td><\/tr><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Vol.:<\/th><td>2 \u00b5L <\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Mobile Phase<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">A:<\/th><td>Water, 2 mM ammonium formate, 0.1% formic acid  <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">B:<\/th><td>Methanol, 2 mM ammonium formate, 0.1% formic acid <\/td><\/tr><tr><td><\/td><td><table class=\"cgram_ramp\"><thead><tr><th>Time (min)<\/th><th>Flow (mL\/min)<\/th><th>%A<\/th><th>%B<\/th><\/tr><\/thead><tbody><tr><td>0.00<\/td><td>0.5<\/td><td>95<\/td><td>5<\/td><\/tr><tr><td>1.5<\/td><td>0.5<\/td><td>35<\/td><td>65<\/td><\/tr><tr><td>8.5<\/td><td>0.5<\/td><td>5<\/td><td>95<\/td><\/tr><tr><td>9.5<\/td><td>0.5<\/td><td>0<\/td><td>100<\/td><\/tr><tr><td>10.5<\/td><td>0.5<\/td><td>0<\/td><td>100<\/td><\/tr><tr><td>10.6<\/td><td>0.5<\/td><td>95<\/td><td>5<\/td><\/tr><tr><td>12.0<\/td><td>0.5<\/td><td>95<\/td><td>5<\/td><\/tr><\/tbody><\/table><\/td><\/tr><\/table><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Detector<\/th><td>MS\/MS<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Ion Mode:<\/th><td>ESI+\/ESI- <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Mode:<\/th><td>MRM <\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Instrument<\/th><td>UHPLC<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Sample Preparation<\/th><td>Gummies were manually chopped into small pieces, and 1 g of sample was weighed in a 50 mL polypropylene tube. The sample was mixed with 5 mL of water and then vigorously vortexed until all gummy pieces were fully solubilized. The sample was fortified with pesticides and mycotoxins at 100 ng\/g. A mix of internal standards was added at 200 ng\/g. The spiked sample was further vortexed for 30 sec. 5 mL of acetonitrile acidified with 1% acetic acid was added to the sample, and this was followed by a 30 sec vortex agitation. Then, a pouch of European EN 15662 QuEChERS extraction salts (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/25849\" rel=\"noopener\">cat.# 25849<\/a>) was added to the sample. The sample was vortexed for 30 sec and then centrifuged for 5 min. 750 \u00b5L of organic extract was mixed with 250 \u00b5L of water. 2 \u03bcL of final extract was injected into the LC-MS\/MS system.<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Notes<\/th><td>Want even better performance when analyzing metal-sensitive compounds? Check out Inert LC columns at <a href=\"https:\/\/www.restek.com\/inert\" target=\"_blank\" rel=\"noopener\">www.restek.com\/inert<\/a>.<\/td><\/tr><\/table><\/div><\/div><\/div><div class=\"chromatogram-related-skus\"><h4>Related SKUs<\/h4><ul><li>34124<\/li><li>34125<\/li><li>34126<\/li><li>34127<\/li><li>34128<\/li><li>34129<\/li><li>31988<\/li><li>31987<\/li><li>31985<\/li><li>31986<\/li><li>31984<\/li><li>31989<\/li><li>31990<\/li><li>34121<\/li><li>34122<\/li><li>9314A0252<\/li><li>9314A12<\/li><li>25849<\/li><\/ul><\/div><div class=\"chromatogram-related-searches\"><h4>Related Searches<\/h4><ul><li>California<\/li><li>Cannabis<\/li><li>Edibles<\/li><li>Mycotoxins<\/li><li>Pesticides<\/li><\/ul><\/div><div class=\"chromatogram-pdf-link\"><a href=\"https:\/\/ez.restek.com\/images\/cgram\/lc_gn0666.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"><svg xmlns=\"http:\/\/www.w3.org\/2000\/svg\" width=\"18\" height=\"18\" viewBox=\"0 0 18 18\"><g data-name=\"Group 2996\"><path data-name=\"Rectangle 1246\" d=\"M0 0h18v18H0z\" style=\"fill: none;\"><\/path><\/g><g data-name=\"Group 2997\"><path data-name=\"Path 729\" d=\"M13.412 11.4v2.017H5.345V11.4H4v2.017a1.349 1.349 0 0 0 1.345 1.345h8.068a1.349 1.349 0 0 0 1.345-1.345V11.4zm-.672-2.694-.948-.948-1.741 1.735V4H8.706v5.493L6.965 7.758l-.948.948 3.361 3.361z\" transform=\"translate(-.437 -.414)\" style=\"fill: rgb(13, 123, 196);\"><\/path><\/g><\/svg>Download PDF<\/a><\/div><\/div><\/div>\n<\/div><\/div><\/div>\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><strong>Figura 3:<\/strong>\u00a0GC-MS\/MS chromatogram of an extract obtained from blank gummy spiked with pesticides and mycotoxins at 100 ng\/g.<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n<div class=\"wp-block-custom-chromatogram-article\"><div class=\"wp-block-custom-chromatogram-article\"><div class=\"chromatogram-image wide-image\"><img decoding=\"async\" src=\"https:\/\/ez.restek.com\/images\/cgram\/gc_gn1207.png\" alt=\"GC Amenable Pesticides Regulated by California in Gummy Edibles on Rxi-5ms\" title=\"-\"><\/div><p class=\"article-id\" style=\"text-align:center\"> GC_GN1207<\/p><div class=\"chromatogram-peaks\"><h4>Peaks<\/h4><table class=\"peaks col-12\">\n<thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 75px\">t<sub>R<\/sub> (min)<\/th><th style=\"text-align: center;width: 75px\">Polarity<\/th><th style=\"text-align: center;width: 75px\">Precursor Ion<\/th><th style=\"text-align: center;width: 75px\">Product Ion<\/th><th style=\"text-align: center;width: 75px\">Transition Type<\/th><\/tr><\/thead>\n<tbody><tr><td class=\"num\">1.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Atrazine-d5\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/163165-75-1\/Atrazine-d5\" target=\"_blank\" rel=\"noopener\">Atrazine-d5<\/a><\/td><td class=\"oth\">6.82<\/td><td class=\"oth\">Positive<\/td><td class=\"oth\">220.0<\/td><td class=\"oth\">58.0<\/td><td class=\"oth\">Quantifier<\/td><\/tr>\n<tr><td class=\"num\">2.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Atrazine-d5\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/163165-75-1\/Atrazine-d5\" target=\"_blank\" rel=\"noopener\">Atrazine-d5<\/a><\/td><td class=\"oth\">6.82<\/td><td class=\"oth\">Positive<\/td><td class=\"oth\">205.0<\/td><td class=\"oth\">127.0<\/td><td class=\"oth\">Qualifier<\/td><\/tr>\n<tr><td class=\"num\">3.<\/td><td class=\"cmpd\">Diazinon-d10 (diethyl-d10)<\/td><td class=\"oth\">7.01<\/td><td class=\"oth\">Positive<\/td><td class=\"oth\">183.0<\/td><td class=\"oth\">139.0<\/td><td class=\"oth\">Quantifier<\/td><\/tr>\n<tr><td class=\"num\">4.<\/td><td class=\"cmpd\">Diazinon-d10 (diethyl-d10)<\/td><td class=\"oth\">7.01<\/td><td class=\"oth\">Positive<\/td><td class=\"oth\">183.0<\/td><td class=\"oth\">168.0<\/td><td class=\"oth\">Qualifier<\/td><\/tr>\n<tr><td class=\"num\">5.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Quintozene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/82-68-8\/Quintozene\" target=\"_blank\" rel=\"noopener\">Quintozene<\/a><\/td><td class=\"oth\">7.03<\/td><td class=\"oth\">Positive<\/td><td class=\"oth\">294.9<\/td><td class=\"oth\">236.9<\/td><td class=\"oth\">Quantifier<\/td><\/tr>\n<tr><td class=\"num\">6.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Quintozene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/82-68-8\/Quintozene\" target=\"_blank\" rel=\"noopener\">Quintozene<\/a><\/td><td class=\"oth\">7.03<\/td><td class=\"oth\">Positive<\/td><td class=\"oth\">236.8<\/td><td class=\"oth\">118.9<\/td><td class=\"oth\">Qualifier<\/td><\/tr>\n<tr><td class=\"num\">7.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Methyl parathion\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/298-00-0\/Methyl parathion\" target=\"_blank\" rel=\"noopener\">Methyl parathion<\/a><\/td><td class=\"oth\">7.50<\/td><td class=\"oth\">Positive<\/td><td class=\"oth\">263.0<\/td><td class=\"oth\">109.0<\/td><td class=\"oth\">Quantifier<\/td><\/tr>\n<tr><td class=\"num\">8.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Methyl parathion\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/298-00-0\/Methyl parathion\" target=\"_blank\" rel=\"noopener\">Methyl parathion<\/a><\/td><td class=\"oth\">7.50<\/td><td class=\"oth\">Positive<\/td><td class=\"oth\">263.0<\/td><td class=\"oth\">79.0<\/td><td class=\"oth\">Qualifier<\/td><\/tr>\n<tr><td class=\"num\">9.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Captan\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/133-06-2\/Captan\" target=\"_blank\" rel=\"noopener\">Captan<\/a><\/td><td class=\"oth\">8.37<\/td><td class=\"oth\">Positive<\/td><td class=\"oth\">184.0<\/td><td class=\"oth\">149.1<\/td><td class=\"oth\">Quantifier<\/td><\/tr>\n<tr><td class=\"num\">10.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Captan\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/133-06-2\/Captan\" target=\"_blank\" rel=\"noopener\">Captan<\/a><\/td><td class=\"oth\">8.37<\/td><td class=\"oth\">Positive<\/td><td class=\"oth\">184.0<\/td><td class=\"oth\">134.1<\/td><td class=\"oth\">Qualifier<\/td><\/tr>\n<tr><td class=\"num\">11.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for trans-Chlordane\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/5103-74-2\/trans-Chlordane\" target=\"_blank\" rel=\"noopener\">trans-Chlordane<\/a><\/td><td class=\"oth\">8.41<\/td><td class=\"oth\">Positive<\/td><td class=\"oth\">271.9<\/td><td class=\"oth\">237.0<\/td><td class=\"oth\">Quantifier<\/td><\/tr>\n<tr><td class=\"num\">12.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for trans-Chlordane\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/5103-74-2\/trans-Chlordane\" target=\"_blank\" rel=\"noopener\">trans-Chlordane<\/a><\/td><td class=\"oth\">8.41<\/td><td class=\"oth\">Positive<\/td><td class=\"oth\">372.9<\/td><td class=\"oth\">265.9<\/td><td class=\"oth\">Qualifier<\/td><\/tr>\n<tr><td class=\"num\">13.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for cis-Chlordane\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/5103-71-9\/cis-Chlordane\" target=\"_blank\" rel=\"noopener\">cis-Chlordane<\/a><\/td><td class=\"oth\">8.53<\/td><td class=\"oth\">Positive<\/td><td class=\"oth\">372.9<\/td><td class=\"oth\">265.9<\/td><td class=\"oth\">Quantifier<\/td><\/tr>\n<tr><td class=\"num\">14.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for cis-Chlordane\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/5103-71-9\/cis-Chlordane\" target=\"_blank\" rel=\"noopener\">cis-Chlordane<\/a><\/td><td class=\"oth\">8.53<\/td><td class=\"oth\">Positive<\/td><td class=\"oth\">271.9<\/td><td class=\"oth\">237.0<\/td><td class=\"oth\">Qualifier<\/td><\/tr>\n<tr><td class=\"num\">15.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Chlorfenapyr\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/122453-73-0\/Chlorfenapyr\" target=\"_blank\" rel=\"noopener\">Chlorfenapyr<\/a><\/td><td class=\"oth\">8.80<\/td><td class=\"oth\">Positive<\/td><td class=\"oth\">247.1<\/td><td class=\"oth\">227.1<\/td><td class=\"oth\">Quantifier<\/td><\/tr>\n<tr><td class=\"num\">16.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Chlorfenapyr\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/122453-73-0\/Chlorfenapyr\" target=\"_blank\" rel=\"noopener\">Chlorfenapyr<\/a><\/td><td class=\"oth\">8.80<\/td><td class=\"oth\">Positive<\/td><td class=\"oth\">59.1<\/td><td class=\"oth\">31.1<\/td><td class=\"oth\">Qualifier<\/td><\/tr>\n<tr><td class=\"num\">17.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Cyfluthrin\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/68359-37-5\/Cyfluthrin\" target=\"_blank\" rel=\"noopener\">Cyfluthrin<\/a><\/td><td class=\"oth\">10.61<\/td><td class=\"oth\">Positive<\/td><td class=\"oth\">226.0<\/td><td class=\"oth\">206.0<\/td><td class=\"oth\">Quantifier<\/td><\/tr>\n<tr><td class=\"num\">18.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Cyfluthrin\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/68359-37-5\/Cyfluthrin\" target=\"_blank\" rel=\"noopener\">Cyfluthrin<\/a><\/td><td class=\"oth\">10.61<\/td><td class=\"oth\">Positive<\/td><td class=\"oth\">163.0<\/td><td class=\"oth\">127.0<\/td><td class=\"oth\">Qualifier<\/td><\/tr>\n<tr><td class=\"num\">19.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Cypermethrin\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/52315-07-8\/Cypermethrin\" target=\"_blank\" rel=\"noopener\">Cypermethrin<\/a><\/td><td class=\"oth\">10.87<\/td><td class=\"oth\">Positive<\/td><td class=\"oth\">163.0<\/td><td class=\"oth\">127.1<\/td><td class=\"oth\">Quantifier<\/td><\/tr>\n<tr><td class=\"num\">20.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Cypermethrin\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/52315-07-8\/Cypermethrin\" target=\"_blank\" rel=\"noopener\">Cypermethrin<\/a><\/td><td class=\"oth\">10.87<\/td><td class=\"oth\">Positive<\/td><td class=\"oth\">181.1<\/td><td class=\"oth\">152.1<\/td><td class=\"oth\">Qualifier<\/td><\/tr>\n<\/tbody><\/table><\/div><div class=\"chromatogram-conditions\"><h4>Conditions<\/h4><div class=\"conditions-container container-fluid\"><div class=\"row\"><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Column<\/th><td>Rxi-5ms, 30 m, 0.25 mm ID, 0.25 \u00b5m (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/13423\" rel=\"noopener\">cat.# 13423<\/a>)<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Standard\/Sample<\/th><td><\/td><\/tr><tr><td><\/td><td>California pesticide standard #1 (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/34124\" rel=\"noopener\">cat.# 34124<\/a>)<\/td><\/tr><tr><td><\/td><td>California pesticide standard #2 (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/34125\" rel=\"noopener\">cat.# 34125<\/a>)<\/td><\/tr><tr><td><\/td><td>California pesticide standard #3 (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/34126\" rel=\"noopener\">cat.# 34126<\/a>)<\/td><\/tr><tr><td><\/td><td>California pesticide standard #4 (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/34127\" rel=\"noopener\">cat.# 34127<\/a>)<\/td><\/tr><tr><td><\/td><td>California pesticide standard #5 (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/34128\" rel=\"noopener\">cat.# 34128<\/a>)<\/td><\/tr><tr><td><\/td><td>California pesticide standard #6 (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/34129\" rel=\"noopener\">cat.# 34129<\/a>)<\/td><\/tr><tr><td><\/td><td>Atrazine-d5 (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/31984\" rel=\"noopener\">cat.# 31984<\/a>)<\/td><\/tr><tr><td><\/td><td>Diazinon-d10 (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/31986\" rel=\"noopener\">cat.# 31986<\/a>)<\/td><\/tr>\n<tr><th class=\"sub conditions_header\" scope=\"row\">Diluent:<\/th><td>Acetonitrile <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Conc.:<\/th><td>2.5-10 ng\/mL Expected concentration range in extract after extracting from gummy fortified at 100 ng\/g (final extract was diluted in half with acetonitrile).<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Injection<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Vol.:<\/th><td>1 \u00b5L splitless<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Liner:<\/th><td>Topaz 4.0 mm ID single taper inlet liner w\/wool (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/23447\" rel=\"noopener\">cat.# 23447<\/a>)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Temp.:<\/th><td>250 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Purge Flow:<\/th><td>5 mL\/min<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\" colspan=\"2\">Oven<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Oven Temp.:<\/th><td>90 &#176;C (hold 1 min) to 310 &#176;C at 25 &#176;C\/min (hold 10 min)<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Carrier Gas<\/th><td>He, constant flow<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Flow Rate:<\/th><td>1.4 mL\/min<\/td><\/tr><\/table><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Detector<\/th><td>MS\/MS<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Transfer Line Temp.:<\/th><td>290 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Analyzer Type:<\/th><td>Quadrupole <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Source Temp.:<\/th><td>330 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Electron Energy:<\/th><td>70 eV<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Tune Type:<\/th><td>PFTBA <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Ionization Mode:<\/th><td>EI <\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Instrument<\/th><td>Thermo Scientific TSQ 8000 Triple Quadrupole GC-MS<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Sample Preparation<\/th><td>Gummies were manually chopped into small pieces, and 1 g of sample was weighed in a 50 mL polypropylene tube. The sample was mixed with 5 mL of water and then vigorously vortexed until all gummy pieces were fully solubilized. The sample was fortified with pesticides and mycotoxins at 100 ng\/g. A mix of internal standards was added at 200 ng\/g. The spiked sample was further vortexed for 30 sec. 5 mL of acetonitrile acidified with 1% acetic acid was added to the sample, and this was followed by 30 sec vortex agitation. Then, a pouch of European EN 15662 QuEChERS extraction salts (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/25849\" rel=\"noopener\">cat.# 25849<\/a>) was added to the sample. The sample was vortexed for 30 sec and then centrifuged for 5 min. 1.9 mL of supernatant was transferred to a Q-sep QuEChERS dSPE tube containing pre-weighed magnesium sulfate, PSA, and GCB (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/26217\" rel=\"noopener\">cat.# 26217<\/a>). After vortexing and centrifuging, 500 \u03bcL of extract was mixed with 500 \u03bcL of acidified acetonitrile. 1 \u03bcL of final extract was injected into the GC-MS\/MS system.<\/td><\/tr><\/table><\/div><\/div><\/div><div class=\"chromatogram-related-skus\"><h4>Related SKUs<\/h4><ul><li>34124<\/li><li>34125<\/li><li>34126<\/li><li>34127<\/li><li>34128<\/li><li>34129<\/li><li>31984<\/li><li>31986<\/li><li>13423<\/li><li>23447<\/li><li>25849<\/li><li>26217<\/li><\/ul><\/div><div class=\"chromatogram-related-searches\"><h4>Related Searches<\/h4><ul><li>California<\/li><li>Cannabis<\/li><li>Pesticides<\/li><li>QuEChERS<\/li><\/ul><\/div><div class=\"chromatogram-pdf-link\"><a href=\"https:\/\/ez.restek.com\/images\/cgram\/gc_gn1207.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"><svg xmlns=\"http:\/\/www.w3.org\/2000\/svg\" width=\"18\" height=\"18\" viewBox=\"0 0 18 18\"><g data-name=\"Group 2996\"><path data-name=\"Rectangle 1246\" d=\"M0 0h18v18H0z\" style=\"fill: none;\"><\/path><\/g><g data-name=\"Group 2997\"><path data-name=\"Path 729\" d=\"M13.412 11.4v2.017H5.345V11.4H4v2.017a1.349 1.349 0 0 0 1.345 1.345h8.068a1.349 1.349 0 0 0 1.345-1.345V11.4zm-.672-2.694-.948-.948-1.741 1.735V4H8.706v5.493L6.965 7.758l-.948.948 3.361 3.361z\" transform=\"translate(-.437 -.414)\" style=\"fill: rgb(13, 123, 196);\"><\/path><\/g><\/svg>Download PDF<\/a><\/div><\/div><\/div>\n<\/div><\/div><\/div>\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\">F<strong>Figura 4:<\/strong>\u00a0Cromatogramma HPLC-UV di un estratto ottenuto da bianco di caramelle gommose fortificato con sei cannabinoidi a 0,2 mg\/g.<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n<div class=\"wp-block-custom-chromatogram-article\"><div class=\"wp-block-custom-chromatogram-article\"><div class=\"chromatogram-image wide-image\"><img decoding=\"async\" src=\"https:\/\/ez.restek.com\/images\/cgram\/lc_gn0667.png\" alt=\"Cannabinoids in Gummy Edibles on Raptor ARC-18\" title=\"-\"><\/div><p class=\"article-id\" style=\"text-align:center\"> LC_GN0667<\/p><div class=\"chromatogram-peaks\"><h4>Peaks<\/h4><table class=\"peaks col-lg-6 col-12\">\n<thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 75px\">t<sub>R<\/sub> (min)<\/th><th style=\"text-align: center;width: 75px\">mg\/g*<\/th><\/tr><\/thead>\n<tbody><tr><td class=\"num\">1.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Cannabidiolic acid (CBDA)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/1244-58-2\/Cannabidiolic acid\" target=\"_blank\" rel=\"noopener\">Cannabidiolic acid (CBDA)<\/a><\/td><td class=\"oth\">2.142<\/td><td class=\"oth\">0.2<\/td><\/tr>\n<tr><td class=\"num\">2.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Cannabigerol (CBG)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/25654-31-3\/Cannabigerol\" target=\"_blank\" rel=\"noopener\">Cannabigerol (CBG)<\/a><\/td><td class=\"oth\">2.405<\/td><td class=\"oth\">0.2<\/td><\/tr>\n<tr><td class=\"num\">3.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Cannabidiol (CBD)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/13956-29-1\/Cannabidiol\" target=\"_blank\" rel=\"noopener\">Cannabidiol (CBD)<\/a><\/td><td class=\"oth\">2.535<\/td><td class=\"oth\">0.2<\/td><\/tr>\n<tr><td class=\"num\">4.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Cannabinol (CBN)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/521-35-7\/Cannabinol\" target=\"_blank\" rel=\"noopener\">Cannabinol (CBN)<\/a><\/td><td class=\"oth\">3.776<\/td><td class=\"oth\">0.2<\/td><\/tr>\n<tr><td class=\"num\">5.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for \u03949-Tetrahydrocannabinol (\u03949-THC)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/1972-08-3\/\u03949-Tetrahydrocannabinol\" target=\"_blank\" rel=\"noopener\">\u03949-Tetrahydrocannabinol (\u03949-THC)<\/a><\/td><td class=\"oth\">4.753<\/td><td class=\"oth\">0.2<\/td><\/tr>\n<tr><td class=\"num\">6.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Tetrahydrocannabinolic acid A (THCA-A)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/23978-85-0\/Tetrahydrocannabinolic acid A\" target=\"_blank\" rel=\"noopener\">Tetrahydrocannabinolic acid A (THCA-A)<\/a><\/td><td class=\"oth\">6.279<\/td><td class=\"oth\">0.2<\/td><\/tr>\n<\/tbody><\/table><div class=\"peaknotes\">*Extract from a gummy sample initially spiked at 0.2 mg\/g.<\/div><\/div><div class=\"chromatogram-conditions\"><h4>Conditions<\/h4><div class=\"conditions-container container-fluid\"><div class=\"row\"><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Column<\/th><td>Raptor ARC-18  (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/9314A62\" rel=\"noopener\">cat.# 9314A62<\/a>)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Dimensions:<\/th><td>150 mm x 2.1 mm ID<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Particle Size:<\/th><td>2.7 \u00b5m<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Pore Size:<\/th><td>90 \u00c5<\/td><\/tr><tr><td><\/td><tr><th class=\"sub conditions_header\" scope=\"row\">Guard Column:<\/th><td>Raptor ARC-18 EXP guard column cartridge 5 mm, 2.1 mm ID, 2.7 \u00b5m (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/9314A0252\" rel=\"noopener\">cat.# 9314A0252<\/a>)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Temp.:<\/th><td>30 \u00b0C<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Standard\/Sample<\/th><td><\/td><\/tr><tr><td><\/td><td>Cannabinoids standard  (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/34014\" rel=\"noopener\">cat.# 34014<\/a>)<\/td><\/tr><tr><td><\/td><td>Cannabigerol (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/34091\" rel=\"noopener\">cat.# 34091<\/a>)<\/td><\/tr><tr><td><\/td><td>d9-Tetrahydrocannabinol  (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/34067\" rel=\"noopener\">cat.# 34067<\/a>)<\/td><\/tr><tr><td><\/td><td>d9-Tetrahydrocannabinolic acid A  (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/34111\" rel=\"noopener\">cat.# 34111<\/a>)<\/td><\/tr>\n<tr><th class=\"sub conditions_header\" scope=\"row\">Diluent:<\/th><td>75:25 Acetonitrile:water<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Conc.:<\/th><td>  Expected concentration of 4 ppm in final extract from gummy initially spiked at 0.2 mg\/g.<\/td><\/tr><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Vol.:<\/th><td>2 \u00b5L <\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Mobile Phase<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">A:<\/th><td>Water, 5 mM ammonium formate, 0.1% formic acid  <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">B:<\/th><td>Acetonitrile, 0.1% formic acid <\/td><\/tr><tr><td><\/td><td><table class=\"cgram_ramp\"><thead><tr><th>Time (min)<\/th><th>Flow (mL\/min)<\/th><th>%A<\/th><th>%B<\/th><\/tr><\/thead><tbody><tr><td>0.00<\/td><td>0.4<\/td><td>25<\/td><td>75<\/td><\/tr><tr><td>10.00<\/td><td>0.4<\/td><td>25<\/td><td>75<\/td><\/tr><\/tbody><\/table><\/td><\/tr><\/table><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Detector<\/th><td>UV\/Vis @ 228 nm<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Instrument<\/th><td>UHPLC<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Sample Preparation<\/th><td>Gummies were manually chopped into small pieces, and 1 g of sample was weighed in a 50 mL polypropylene tube. The sample was mixed with 5 mL of water and then vigorously vortexed until all gummy pieces were fully solubilized. The sample was fortified with cannabinoids at 0.2 mg\/g. The spiked sample was further vortexed for 30 sec. 5 mL of acetonitrile acidified with 1% acetic acid was added to the sample, and this was followed by 30 sec vortex agitation. Then, a pouch of European EN 15662 QuEChERS extraction salts (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/25849\" rel=\"noopener\">cat.# 25849<\/a>) was added to the sample. The sample was vortexed for 30 sec and then centrifuged for 5 min. 100 \u00b5L of organic extract was mixed with 900 \u00b5L of 75:25 acetonitrile:water. 2 \u00b5L of final extract was injected into the HPLC-UV system.<\/td><\/tr><\/table><\/div><\/div><\/div><div class=\"chromatogram-related-skus\"><h4>Related SKUs<\/h4><ul><li>34014<\/li><li>34091<\/li><li>34067<\/li><li>34111<\/li><li>9314A0252<\/li><li>9314A62<\/li><li>25849<\/li><\/ul><\/div><div class=\"chromatogram-related-searches\"><h4>Related Searches<\/h4><ul><li>Edibles<\/li><li>Gummy<\/li><li>Potency<\/li><li>QuEChERs<\/li><\/ul><\/div><div class=\"chromatogram-pdf-link\"><a href=\"https:\/\/ez.restek.com\/images\/cgram\/lc_gn0667.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"><svg xmlns=\"http:\/\/www.w3.org\/2000\/svg\" width=\"18\" height=\"18\" viewBox=\"0 0 18 18\"><g data-name=\"Group 2996\"><path data-name=\"Rectangle 1246\" d=\"M0 0h18v18H0z\" style=\"fill: none;\"><\/path><\/g><g data-name=\"Group 2997\"><path data-name=\"Path 729\" d=\"M13.412 11.4v2.017H5.345V11.4H4v2.017a1.349 1.349 0 0 0 1.345 1.345h8.068a1.349 1.349 0 0 0 1.345-1.345V11.4zm-.672-2.694-.948-.948-1.741 1.735V4H8.706v5.493L6.965 7.758l-.948.948 3.361 3.361z\" transform=\"translate(-.437 -.414)\" style=\"fill: rgb(13, 123, 196);\"><\/path><\/g><\/svg>Download PDF<\/a><\/div><\/div><\/div>\n<\/div><\/div><\/div>\n\n\n<div style=\"height:20px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<h2 class=\"wp-block-heading\">Conclusioni<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">\u00c8 stato sviluppato un flusso di lavoro semplice ed efficace per l\u2019analisi di pesticidi, micotossine e cannabinoidi nelle caramelle gommose alla cannabis. Le condizioni per la preparazione del campione includono l\u2019omogenizzazione della matrice di pezzi di caramelle in acqua, l\u2019estrazione degli analiti con acetonitrile acidificato seguita da una fase di salting-out con i sali di estrazione Q-sep QuEChERS, la diluizione dell\u2019estratto (per i contaminanti e cannabinoidi analizzabili in LC-MS\/MS) e infine la purificazione dSPE, per la quale si \u00e8 utilizzato solfato di magnesio, PSA e GCB (per i pesticidi compatibili con GC-MS\/MS). Per tutti i contaminanti target sono stati ottenuti risultati eccellenti in termini di LOQ, linearit\u00e0, accuratezza e precisione. Inoltre, i dati hanno mostrato che la metodologia proposta \u00e8 idonea per l\u2019analisi di potenza, offrendo valori di accuratezza fra il 99% e il 107% per i sei cannabinoidi elencati dalle normative del governo californiano. In generale, il flusso di lavoro presentato snellisce il lavoro di laboratorio per l\u2019analisi della cannabis in quanto permette una quantificazione eccellente di diverse classi di analiti in campioni di caramelle gommose utilizzando un solo estratto.<\/p>\n\n\n\n<h2 class=\"wp-block-heading\">Bibliografia<\/h2>\n\n\n\n<ol class=\"wp-block-list\">\n<li>Text of Regulations, Bureau of Cannabis Control, California Code of Regulations,&nbsp;<a href=\"https:\/\/cannabis.ca.gov\/wp-content\/uploads\/sites\/13\/2019\/01\/Order-of-Adoption-Clean-Version-of-Text.pdf\" target=\"_blank\" rel=\"noreferrer noopener\">https:\/\/cannabis.ca.gov\/wp-content\/uploads\/sites\/13\/2019\/01\/Order-of-Adoption-Clean-Version-of-Text.pdf<\/a>, (accessed 8 November 2019).<\/li>\n\n\n\n<li>X. Wang, D. Mackowsky, J. Searfoss, M. J. Telepchak, Determination of cannabinoid content and pesticide residues in cannabis edibles and beverages, Cannabis Sci. and Tech. (2018).&nbsp;<a href=\"https:\/\/www.cannabissciencetech.com\/view\/determination-cannabinoid-content-and-pesticide-residues-cannabis-edibles-and-beverages\" target=\"_blank\" rel=\"noreferrer noopener\">https:\/\/www.cannabissciencetech.com\/view\/determination-cannabinoid-content-and-pesticide-residues-cannabis-edibles-and-beverages<\/a>.<\/li>\n\n\n\n<li>Fast, low-solvent analysis of cannabinoids increases lab productivity and decreases solvent costs, Restek Corporation.\u00a0<a href=\"https:\/\/discover.restek.com\/articles\/fffa3123\/fast-low-solvent-analysis-of-cannabinoids-increases-lab-productivity-and-decreases-solvent-costs\" data-type=\"post\" data-id=\"43558\">https:\/\/www.restek.com\/Technical-Resources\/Technical-Library\/Foods-Flavors-Fragrances\/fff_FFFA3123-UNV<\/a>\u00a0(accessed 22 January 2021).<\/li>\n\n\n\n<li>N. Reyes-Garces, C. Myers, Analysis of pesticides and mycotoxins cannabis brownies, Restek Corporation.\u00a0<a href=\"https:\/\/discover.restek.com\/ffan3149-it\/lanalisi-di-pesticidi-e-micotossine-nei-brownie-alla-cannabis\/\" data-type=\"post\" data-id=\"45313\">https:\/\/www.restek.com\/Technical-Resources\/Technical-Library\/Foods-Flavors-Fragrances\/fff_FFAN3149-UNV<\/a>\u00a0(accessed 3 December 2020).<\/li>\n\n\n\n<li>B. K. Matuszewski, M. L. Constanzer, C. M. Chavez-Eng, Strategies for the assessment of matrix effect in quantitative bioanalytical methods based on HPLC\u2212MS\/MS, Anal. Chem. 75 (2003) 3019\u20133030.&nbsp;<a href=\"https:\/\/pubs.acs.org\/doi\/10.1021\/ac020361s\" target=\"_blank\" rel=\"noreferrer noopener\">https:\/\/pubs.acs.org\/doi\/10.1021\/ac020361s<\/a><\/li>\n<\/ol>\n\n\n\n<div style=\"height:50px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n        <div class=\"cpb\">\n            <h3 class=\"cpb-heading\">Prodotti Menzionati<\/h3>\n            <hr class=\"cpb-heading-underline\" \/>\n            <div class=\"cpb-list\">\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/25846\">                                N\u00b0 Catalogo 25846                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Provette da centrifuga vuote da 50 ml, Polipropilene con tappo, 50 provette, 50-conf.<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/25846\">Esplora Prodotto<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/25849\">                                N\u00b0 Catalogo 25849                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Solo pacchetti di sali di estrazione Q-sep QuEChERS (EN), 4 g MgSO4, 1 g NaCI, 1 g TSCD, 0,5 g DHS, 50 pacchetti<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/25849\">Esplora Prodotto<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/26217\">                                N\u00b0 Catalogo 26217                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Provetta da centrifuga da 2 mL dSPE Q-sep QuEChERS, contiene 150 mg MgSO4, 25 mg PSA, 2,5 mg GCB, 100 provette<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/26217\">Esplora Prodotto<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/9314A12\">                                N\u00b0 Catalogo 9314A12                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Colonna HPLC Raptor ARC-18, 2,7 \u00b5m, 100 x 2,1 mm<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/9314A12\">Esplora Prodotto<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/9314A0252\">                                N\u00b0 Catalogo 9314A0252                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Cartucce Precolonna EXP Raptor ARC-18, 2,7 \u00b5m, 5 x 2,1 mm, 3 pz.<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/9314A0252\">Esplora Prodotto<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/13423\">                                N\u00b0 Catalogo 13423                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Colonna capillare GC Rxi-5ms, 30 m, ID 0,25 mm, 0,25 \u00b5m<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/13423\">Esplora Prodotto<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/23447\">                                N\u00b0 Catalogo 23447                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Topaz Inlet Liner a Cono Singolo, 4,0 mm x 6,5 x 78,5, per GC Thermo TRACE 1300\/1310, 1600\/1610 con Iniettori SSL, con Lana di Quarzo, Deattivazione Premium, 5 pz.<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/23447\">Esplora Prodotto<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/9314A62\">                                N\u00b0 Catalogo 9314A62                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Colonna HPLC Raptor ARC-18, 2,7 \u00b5m, 150 x 2,1 mm<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/9314A62\">Esplora Prodotto<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/25851\">                                N\u00b0 Catalogo 25851                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Solo pacchetti di sali di estrazione Q-sep QuEChERS (AOAC), 6 g MgSO4, 1,5 g NaOAc, 50 pacchetti<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/25851\">Esplora Prodotto<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/25847\">                                N\u00b0 Catalogo 25847                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Solo pacchetti di sali di estrazione Q-sep QuEChERS (originale), 4 g MgSO4, 1 g NaCI, 50 pacchetti<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/25847\">Esplora Prodotto<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/26215\">                                N\u00b0 Catalogo 26215                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Provetta da centrifuga da 2 mL dSPE Q-sep QuEChERS, contiene 150 mg MgSO4, 25 mg PSA, 100 provette<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/26215\">Esplora Prodotto<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/26216\">                                N\u00b0 Catalogo 26216                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Provetta da centrifuga da 2 mL dSPE Q-sep QuEChERS, contiene 150 mg MgSO4, 25 mg PSA, 25 mg C18-EC, 100 provette<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/26216\">Esplora Prodotto<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/26242\">                                N\u00b0 Catalogo 26242                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Provetta da centrifuga da 2 mL dSPE Q-sep QuEChERS, contiene 150 mg MgSO4, 50 mg C18-EC, 100 provette<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/26242\">Esplora Prodotto<\/a>\n                                                    <\/div>\n                    <\/div>\n                            <\/div>\n        <\/div>\n        \n\n\n<div style=\"height:100px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n","protected":false},"excerpt":{"rendered":"<p>In questa nota applicativa, presentiamo un flusso di lavoro semplice ed efficace per l&#8217;analisi di pesticidi, micotossine e cannabinoidi nelle caramelle alla cannabis. Questo metodo semplifica i test consentendo la quantificazione di pi\u00f9 classi di analiti da un singolo estratto di campione.<\/p>\n","protected":false},"author":5,"featured_media":6841,"comment_status":"closed","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"_acf_changed":false,"_kad_blocks_custom_css":"","_kad_blocks_head_custom_js":"","_kad_blocks_body_custom_js":"","_kad_blocks_footer_custom_js":"","_kadence_starter_templates_imported_post":false,"_kad_post_transparent":"","_kad_post_title":"","_kad_post_layout":"","_kad_post_sidebar_id":"","_kad_post_content_style":"","_kad_post_vertical_padding":"","_kad_post_feature":"","_kad_post_feature_position":"","_kad_post_header":false,"_kad_post_footer":false,"footnotes":"","_ppma_block_editor_authors":""},"categories":[784],"tags":[],"industries-application":[2281,2243,2290,2292],"post-badge":[],"resource-type":[],"product-library":[2528,2535,2515,2517,2516,2519,2544,2530,2518,2543,2524],"resource-technique":[2334,2332,2364,2369],"ppma_author":[603],"class_list":["post-45314","post","type-post","status-publish","format-standard","has-post-thumbnail","hentry","category-application-notes-it","industries-application-cannabis-it","industries-application-cannabis-terapeutica-ricreativa","industries-application-food-beverage-it","industries-application-sicurezza-alimentare-it","product-library-colonne-gc","product-library-colonne-lc","product-library-prodotti-per-campionamento-dellaria-e-dei-gas","product-library-prodotti-per-cromatografia-liquida","product-library-prodotti-per-gascromatografia","product-library-prodotti-per-la-preparazione-del-campione","product-library-prodotti-quechers","product-library-ricambi-per-iniettore-gc","product-library-standard-di-riferimento","product-library-standard-di-riferimento-liquidi","product-library-standard-gassosi-accessori","resource-technique-cromatografia-liquida-lc","resource-technique-gascromatografia-gc","resource-technique-ms-ms-it","resource-technique-quechers-it"],"acf":[],"taxonomy_info":{"category":[{"value":784,"label":"Application Notes"}],"industries-application":[{"value":2281,"label":"Cannabis"},{"value":2243,"label":"Cannabis Terapeutica o Ricreativa"},{"value":2290,"label":"Food &amp; Beverage"},{"value":2292,"label":"Sicurezza alimentare"}],"product-library":[{"value":2528,"label":"Colonne GC"},{"value":2535,"label":"Colonne LC"},{"value":2515,"label":"Prodotti per Campionamento dell'Aria e dei Gas"},{"value":2517,"label":"Prodotti per Cromatografia Liquida"},{"value":2516,"label":"Prodotti per Gascromatografia"},{"value":2519,"label":"Prodotti per la Preparazione del Campione"},{"value":2544,"label":"Prodotti QuEChERS"},{"value":2530,"label":"Ricambi per Iniettore GC"},{"value":2518,"label":"Standard di Riferimento"},{"value":2543,"label":"Standard di Riferimento Liquidi"},{"value":2524,"label":"Standard Gassosi e Accessori"}],"resource-technique":[{"value":2334,"label":"Cromatografia Liquida (LC)"},{"value":2332,"label":"Gascromatografia (GC)"},{"value":2364,"label":"MS\/MS"},{"value":2369,"label":"QuEChERS"}]},"featured_image_src_large":["https:\/\/discover.restek.com\/wp-content\/uploads\/feature-FFAN3481-1024x536.jpg",1024,536,true],"author_info":{"display_name":"Brenda Seprish","author_link":"https:\/\/discover.restek.com\/it\/author\/brenda-coganrestek-com\/"},"comment_info":0,"category_info":[{"term_id":784,"name":"Application Notes","slug":"application-notes-it","term_group":0,"term_taxonomy_id":784,"taxonomy":"category","description":"","parent":0,"count":88,"filter":"raw","cat_ID":784,"category_count":88,"category_description":"","cat_name":"Application Notes","category_nicename":"application-notes-it","category_parent":0}],"tag_info":false,"authors":[{"term_id":603,"user_id":0,"is_guest":1,"slug":"nathaly-reyes-garces-phd","display_name":"Nathaly Reyes-Garc\u00e9s, PhD","avatar_url":"https:\/\/secure.gravatar.com\/avatar\/?s=96&d=mm&r=g","author_category":"1","first_name":"Nathaly","last_name":"Reyes-Garc\u00e9s","user_url":"","job_title":"","description":""}],"_links":{"self":[{"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/posts\/45314","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/users\/5"}],"replies":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/comments?post=45314"}],"version-history":[{"count":21,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/posts\/45314\/revisions"}],"predecessor-version":[{"id":85757,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/posts\/45314\/revisions\/85757"}],"wp:featuredmedia":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/media\/6841"}],"wp:attachment":[{"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/media?parent=45314"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/categories?post=45314"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/tags?post=45314"},{"taxonomy":"industries-application","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/industries-application?post=45314"},{"taxonomy":"post-badge","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/post-badge?post=45314"},{"taxonomy":"resource-type","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/resource-type?post=45314"},{"taxonomy":"product-library","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/product-library?post=45314"},{"taxonomy":"resource-technique","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/resource-technique?post=45314"},{"taxonomy":"author","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/ppma_author?post=45314"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}