{"id":48522,"date":"2022-11-16T00:00:00","date_gmt":"2022-11-16T00:00:00","guid":{"rendered":"https:\/\/discover.restek.com\/uncategorized\/troubleshooting-gc-columns-a-methodological-approach-part-1\/"},"modified":"2026-09-08T13:55:08","modified_gmt":"2026-09-08T13:55:08","slug":"troubleshooting-gc-columns-a-methodological-approach-part-1","status":"publish","type":"post","link":"https:\/\/discover.restek.com\/it\/blogs\/gnbl4842\/troubleshooting-gc-columns-a-methodological-approach-part-1","title":{"rendered":"Troubleshooting GC Columns \u2013 a Methodological Approach \u2013 Part 1"},"content":{"rendered":"\n<p class=\"wp-block-paragraph\">When it comes to troubleshooting GC columns, often a methodological approach is best, unless you know exactly what has caused the issue. &nbsp;If you are struggling to find the cause for an issue\/problem, perhaps the outlines below will help.<\/p>\n\n\n\n<div style=\"height:20px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<h2 class=\"wp-block-heading\">New column (no samples\/matrix injected into the column)<\/h2>\n\n\n\n<div style=\"height:20px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<h3 class=\"wp-block-heading\">1. How old is the column?<\/h3>\n\n\n\n<ol class=\"wp-block-list\">\n<li>When exactly was this column purchased?<\/li>\n\n\n\n<li>How was this column stored?<\/li>\n<\/ol>\n\n\n\n<h3 class=\"wp-block-heading\">2. Was the column installation verified to be leak free?<\/h3>\n\n\n\n<ol style=\"list-style-type:lower-alpha\" class=\"wp-block-list\">\n<li>How was leak checking performed?<\/li>\n\n\n\n<li>Was the electronic leak detector working properly?<\/li>\n<\/ol>\n\n\n\n<h3 class=\"wp-block-heading\">3. Have the carrier gas traps\/filters been inspected and confirmed to be in good working order?<\/h3>\n\n\n\n<ol style=\"list-style-type:lower-alpha\" class=\"wp-block-list\">\n<li>Are the carrier gas filters \u201cindicating\u201d or \u201cnon-indicating\u201d?<\/li>\n\n\n\n<li>If non-indicating, how long ago were they replaced?<\/li>\n<\/ol>\n\n\n\n<h3 class=\"wp-block-heading\">4. Was the column purged long enough to remove all traces of air (oxygen) before heat was applied?<\/h3>\n\n\n\n<ol style=\"list-style-type:lower-alpha\" class=\"wp-block-list\">\n<li>What was the purity of the carrier gas?<\/li>\n\n\n\n<li>What was the flow rate used to purge the system and how long was the system purged?<\/li>\n<\/ol>\n\n\n\n<h3 class=\"wp-block-heading\">5. How exactly was the column conditioned?<\/h3>\n\n\n\n<ol style=\"list-style-type:lower-alpha\" class=\"wp-block-list\">\n<li>Was the column conditioned according to the manufacturer\u2019s instructions?<\/li>\n\n\n\n<li>Has a similar column been successfully conditioned in the past using the same instructions?<\/li>\n<\/ol>\n\n\n\n<h3 class=\"wp-block-heading\">6. When a column test mix was injected\/analyzed, how did the peak shapes, responses, retention times and baseline compare to the original QA column test chromatogram that was supplied by the manufacturer?<\/h3>\n\n\n\n<ol style=\"list-style-type:lower-alpha\" class=\"wp-block-list\">\n<li>Did your newly generated chromatogram closely match the one obtained from the manufacturer?<\/li>\n\n\n\n<li>If not, does the system\/instrument need to be optimized or serviced?<\/li>\n<\/ol>\n\n\n\n<div style=\"height:20px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<figure class=\"wp-block-image\"><img decoding=\"async\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-troubleshooting-gc-teaser-product-gc-columns-fused-silica-capillary.jpg\" alt=\"\" title=\"-\"><\/figure>\n\n\n\n<div style=\"height:50px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<h2 class=\"wp-block-heading\">Used column (samples\/matrix have been injected into the column)<\/h2>\n\n\n\n<h3 class=\"wp-block-heading\">1. How old is the column?<\/h3>\n\n\n\n<ol style=\"list-style-type:lower-alpha\" class=\"wp-block-list\">\n<li>How long has this column been used?<\/li>\n\n\n\n<li>How was this column stored between analysis?<\/li>\n<\/ol>\n\n\n\n<h3 class=\"wp-block-heading\">2. Has this column ever worked well for the current analysis, or for any other analysis?<\/h3>\n\n\n\n<ol style=\"list-style-type:lower-alpha\" class=\"wp-block-list\">\n<li>If \u201cyes\u201d, when exactly did this column stop working well?<\/li>\n\n\n\n<li>If \u201cno\u201d, what exactly is the column doing or not doing that makes its performance unacceptable?<\/li>\n<\/ol>\n\n\n\n<h3 class=\"wp-block-heading\">3. If the column did work well in the past, is it installed into that exact same instrument?<\/h3>\n\n\n\n<ol style=\"list-style-type:lower-alpha\" class=\"wp-block-list\">\n<li>If the instrument is the same, has a different column been installed into this instrument to confirm the instrument is working properly?<\/li>\n\n\n\n<li>If the instrument is different, has the instrument been optimized for this method and verified that it is working properly?<\/li>\n<\/ol>\n\n\n\n<h3 class=\"wp-block-heading\">4. If the instrument has multiple injection ports and\/or detectors, have the column positions been switched?<\/h3>\n\n\n\n<ol style=\"list-style-type:lower-alpha\" class=\"wp-block-list\">\n<li>Has the column been installed into the second injection port and tested?<\/li>\n\n\n\n<li>Has the column been installed into the second detector and tested?<\/li>\n<\/ol>\n\n\n\n<h3 class=\"wp-block-heading\">5. If the column had been working well in a particular instrument and now is not, what has changed?<\/h3>\n\n\n\n<ol style=\"list-style-type:lower-alpha\" class=\"wp-block-list\">\n<li>Was it performing well until matrix\/samples were analyzed?<\/li>\n\n\n\n<li>Was it performing well until something else happened, like a power outage or loss of carrier gas?<\/li>\n<\/ol>\n\n\n\n<h3 class=\"wp-block-heading\">6. How quickly did the change take place from working well to working poorly?<\/h3>\n\n\n\n<ol style=\"list-style-type:lower-alpha\" class=\"wp-block-list\">\n<li>Was the change immediate?<\/li>\n\n\n\n<li>Did the change occur over time?<\/li>\n<\/ol>\n\n\n\n<h3 class=\"wp-block-heading\">7. Which GC maintenance steps were necessary to improve the column\u2019s performance?<\/h3>\n\n\n\n<ol style=\"list-style-type:lower-alpha\" class=\"wp-block-list\">\n<li>Did routine maintenance, such as trimming the column or replacing the injection port liner, bring back the performance?<\/li>\n\n\n\n<li>Did the column need replaced?<\/li>\n<\/ol>\n\n\n\n<p class=\"wp-block-paragraph\">I hope that the outlines above will assist you in determining the cause of the issue, or almost equally as important, what is not causing the issue.&nbsp; Once you have narrowed down possible reasons for the problem, see <a href=\"https:\/\/discover.restek.com\/blogs\/troubleshooting-gc-columns-a-methodological-approach-part-2\" data-type=\"link\" data-id=\"https:\/\/discover.restek.com\/blogs\/troubleshooting-gc-columns-a-methodological-approach-part-2\"><strong>Troubleshooting GC Columns \u2013 a Methodological Approach &#8211; Part 2<\/strong><\/a>&nbsp;for possible fixes\/solutions.<\/p>\n\n\n\n<figure class=\"wp-block-embed is-type-video is-provider-youtube wp-block-embed-youtube wp-embed-aspect-16-9 wp-has-aspect-ratio\"><div class=\"wp-block-embed__wrapper\">\n<iframe loading=\"lazy\" title=\"Troubleshooting Principles\" width=\"720\" height=\"405\" src=\"https:\/\/www.youtube.com\/embed\/UU0H1KtccsA?feature=oembed\" frameborder=\"0\" allow=\"accelerometer; autoplay; clipboard-write; encrypted-media; gyroscope; picture-in-picture; web-share\" referrerpolicy=\"strict-origin-when-cross-origin\" allowfullscreen><\/iframe>\n<\/div><\/figure>\n\n\n\n<div style=\"height:100px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n","protected":false},"excerpt":{"rendered":"<p>When it comes to troubleshooting GC columns, often a methodological approach is best, unless you know exactly what has caused the issue. &nbsp;If you are struggling to find the cause for an issue\/problem, perhaps this information will help.<\/p>\n","protected":false},"author":89,"featured_media":0,"comment_status":"closed","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"_acf_changed":false,"_kad_blocks_custom_css":"","_kad_blocks_head_custom_js":"","_kad_blocks_body_custom_js":"","_kad_blocks_footer_custom_js":"","_kadence_starter_templates_imported_post":false,"_kad_post_transparent":"","_kad_post_title":"","_kad_post_layout":"","_kad_post_sidebar_id":"","_kad_post_content_style":"","_kad_post_vertical_padding":"","_kad_post_feature":"","_kad_post_feature_position":"","_kad_post_header":false,"_kad_post_footer":false,"footnotes":"","_ppma_block_editor_authors":""},"categories":[9],"tags":[],"industries-application":[],"post-badge":[],"resource-type":[],"product-library":[],"resource-technique":[2294],"ppma_author":[582],"class_list":["post-48522","post","type-post","status-publish","format-standard","hentry","category-blogs","resource-technique-gas-chromatography-gc"],"acf":[],"taxonomy_info":{"category":[{"value":9,"label":"Blogs"}],"resource-technique":[{"value":2294,"label":"Gas Chromatography (GC)"}]},"featured_image_src_large":false,"author_info":{"display_name":"Travis Prebble","author_link":"https:\/\/discover.restek.com\/it\/author\/travis-prebble\/"},"comment_info":0,"category_info":[{"term_id":9,"name":"Blogs","slug":"blogs","term_group":0,"term_taxonomy_id":9,"taxonomy":"category","description":"","parent":0,"count":457,"filter":"raw","cat_ID":9,"category_count":457,"category_description":"","cat_name":"Blogs","category_nicename":"blogs","category_parent":0}],"tag_info":false,"authors":[{"term_id":582,"user_id":0,"is_guest":1,"slug":"alan-sensue","display_name":"Alan Sensue","avatar_url":"https:\/\/secure.gravatar.com\/avatar\/?s=96&d=mm&r=g","author_category":"1","first_name":"Alan","last_name":"Sensue","user_url":"","job_title":"","description":""}],"_links":{"self":[{"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/posts\/48522","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/users\/89"}],"replies":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/comments?post=48522"}],"version-history":[{"count":8,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/posts\/48522\/revisions"}],"predecessor-version":[{"id":102801,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/posts\/48522\/revisions\/102801"}],"wp:attachment":[{"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/media?parent=48522"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/categories?post=48522"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/tags?post=48522"},{"taxonomy":"industries-application","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/industries-application?post=48522"},{"taxonomy":"post-badge","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/post-badge?post=48522"},{"taxonomy":"resource-type","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/resource-type?post=48522"},{"taxonomy":"product-library","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/product-library?post=48522"},{"taxonomy":"resource-technique","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/resource-technique?post=48522"},{"taxonomy":"author","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/ppma_author?post=48522"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}