{"id":51607,"date":"2018-07-17T14:30:00","date_gmt":"2018-07-17T14:30:00","guid":{"rendered":"https:\/\/discover.restek.com\/uncategorized\/annual-replacement-of-cartridge-gas-filters\/"},"modified":"2026-05-26T19:35:58","modified_gmt":"2026-05-26T19:35:58","slug":"annual-replacement-of-cartridge-gas-filters","status":"publish","type":"post","link":"https:\/\/discover.restek.com\/it\/blogs\/gnbl4580\/annual-replacement-of-cartridge-gas-filters","title":{"rendered":"Annual Replacement of Cartridge Gas Filters"},"content":{"rendered":"\n<p class=\"wp-block-paragraph\">A significant number of instrument and column complaints are carrier gas related, which could be caused by breakthrough from filters that are not replaced in time.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">The possible consequences of a filter breakthrough are:<\/p>\n\n\n\n<ul class=\"wp-block-list\">\n<li>Gas distribution system behind the filter will be contaminated (fast cleaning nearly impossible, bleeds for months)<\/li>\n\n\n\n<li>Instrument gets contaminated, expensive maintenance required<\/li>\n\n\n\n<li>Column lifetime reduced, bad analytical results, high cost of ownership, unnecessary changing of column brands<\/li>\n\n\n\n<li>MS source gets contaminated, expensive maintenance required, long system shutdown.<\/li>\n<\/ul>\n\n\n\n<p class=\"wp-block-paragraph\">This article explains why it is important to replace your <a href=\"https:\/\/www.restek.com\/c\/1166\" target=\"_blank\" rel=\"noopener\">gas filters<\/a> annually instead of waiting for the indicators to change color.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\"><strong>Filter Media Types<\/strong><\/p>\n\n\n\n<p class=\"wp-block-paragraph\">A typical GC-MS laboratory gas filter system contains three types of filtering media:<\/p>\n\n\n\n<ul class=\"wp-block-list\">\n<li>Oxygen Catalyst for absorbing traces of&nbsp;<strong>oxygen<\/strong><\/li>\n\n\n\n<li>Activated Carbon for adsorbing&nbsp;<strong>hydrocarbons<\/strong><\/li>\n\n\n\n<li>Molecular Sieve for adsorbing&nbsp;<strong>moisture<\/strong><\/li>\n<\/ul>\n\n\n\n<p class=\"wp-block-paragraph\">These filter media types can be divided into two groups:<\/p>\n\n\n\n<ul class=\"wp-block-list\">\n<li><strong>Chemically absorbent media<\/strong><\/li>\n<\/ul>\n\n\n\n<figure class=\"wp-block-image size-full\"><img loading=\"lazy\" decoding=\"async\" width=\"160\" height=\"139\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-annual-replacement-of-cartridge-gas-filters-01.png\" alt=\"\" class=\"wp-image-12724\" title=\"-\"><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\">Like a Venus fly trap, when it comes in contact with a contaminate it absorbs it and does not let it go. Adsorption differs from absorption, which also removes things, but the result is swelling of the media. The media size increase equals&nbsp;the amount of material removed<\/p>\n\n\n\n<ul class=\"wp-block-list\">\n<li><strong>Surface adsorbent media<\/strong><\/li>\n<\/ul>\n\n\n\n<figure class=\"wp-block-image\"><img decoding=\"async\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-annual-replacement-of-cartridge-gas-filters-02.png\" alt=\"decorative\" class=\"wp-image-48432\" title=\"-\"><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\">Like a floor mop, where the contaminant is trapped onto the surface of the media and retained. The material doing the adsorption does not change in size.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\"><strong>Filter Media Breakthrough Indicators<\/strong><\/p>\n\n\n<div class=\"wp-block-image\">\n<figure class=\"alignright\"><img decoding=\"async\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-annual-replacement-of-cartridge-gas-filters-03.png\" alt=\"decorative\" class=\"wp-image-48441\" title=\"-\"><\/figure>\n<\/div>\n\n\n<p class=\"wp-block-paragraph\">The visual indicators are mainly for urgent situations such as a leak or high amount of impurities breaking through.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Indicators are typically placed behind the filter media bed. When they change color shortly after installation of the filter it usually means there is a leak or that the gas distribution system including the manifold to which the filter is connected was not flushed properly prior to &#8211; or during &#8211; installation of the filter.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">When the indicator changes color during normal operation of the filter it indicates that the filtration media has reached its capacity, and the filter should be replaced&nbsp;<strong>immediately&nbsp;<\/strong>to avoid contaminant breakthrough.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Other components such as branched hydrocarbons which are also trapped by the molecular sieve are not shown by the moisture indicator.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Also note that most filters do not provide an indicator for hydrocarbons, so you would never know whether the activated carbon has reached its adsorption capacity.<\/p>\n\n\n\n<figure class=\"wp-block-image\"><img decoding=\"async\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-annual-replacement-of-cartridge-gas-filters-04.png\" alt=\"decorative\" class=\"wp-image-48439\" title=\"-\"><\/figure>\n\n\n\n<figure class=\"wp-block-image size-full\"><img loading=\"lazy\" decoding=\"async\" width=\"260\" height=\"251\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-annual-replacement-of-cartridge-gas-filters-05.png\" alt=\"\" class=\"wp-image-12748\" title=\"-\"><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\"><strong>Filter Media Breakthrough<\/strong><\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Both adsorption and absorption media have fixed capacities, meaning they hold just so much, since they are storing the material removed from the gas, not destroying it.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Normal situation<\/p>\n\n\n<div class=\"wp-block-image\">\n<figure class=\"aligncenter\"><img decoding=\"async\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-annual-replacement-of-cartridge-gas-filters-06.png\" alt=\"decorative\" class=\"wp-image-48443\" title=\"-\"><\/figure>\n<\/div>\n\n\n<p class=\"wp-block-paragraph\">Hydrocarbon breakthrough<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Over time, the pores of the activated carbon fill up. The molecules that are adsorbed with higher energy (larger mass) can displace the lower-energy molecules that are less tightly held. This phenomenon, called displacement, may knock the smaller particles off the media, straight to the column and into the instrument.<\/p>\n\n\n<div class=\"wp-block-image\">\n<figure class=\"aligncenter\"><img decoding=\"async\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-annual-replacement-of-cartridge-gas-filters-07.png\" alt=\"decorative\" class=\"wp-image-48444\" title=\"-\"><\/figure>\n<\/div>\n\n\n<p class=\"wp-block-paragraph\">Moisture breakthrough<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">The molecular sieve adsorbs moisture until it cannot adsorb anymore. If the humidity level of the gas is lower than in the molecular sieve, it will de-adsorb its moisture until it is in &#8220;balance&#8221; with the lower humidity level of the gas, which means the filter could&nbsp;<strong>increase&nbsp;<\/strong>the amount of moisture in your gas.<\/p>\n\n\n<div class=\"wp-block-image\">\n<figure class=\"aligncenter size-full\"><img loading=\"lazy\" decoding=\"async\" width=\"520\" height=\"108\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-annual-replacement-of-cartridge-gas-filters-08.png\" alt=\"\" class=\"wp-image-12766\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-annual-replacement-of-cartridge-gas-filters-08.png 520w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-annual-replacement-of-cartridge-gas-filters-08-300x62.png 300w\" sizes=\"auto, (max-width: 520px) 100vw, 520px\" \/><\/figure>\n<\/div>\n\n\n<p class=\"wp-block-paragraph\">Oxygen breakthrough<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">The media size increase equals the amount of material removed. When the media has reached its absorption capacity it will not release the already trapped contaminants but it will also not absorb any new ones.<\/p>\n\n\n<div class=\"wp-block-image\">\n<figure class=\"aligncenter size-full\"><img loading=\"lazy\" decoding=\"async\" width=\"512\" height=\"112\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-annual-replacement-of-cartridge-gas-filters-09.png\" alt=\"\" class=\"wp-image-12772\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-annual-replacement-of-cartridge-gas-filters-09.png 512w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-annual-replacement-of-cartridge-gas-filters-09-300x66.png 300w\" sizes=\"auto, (max-width: 512px) 100vw, 512px\" \/><\/figure>\n<\/div>\n\n\n<p class=\"wp-block-paragraph\"><strong>Replace your filter before the indicators start changing color to prevent breakthrough and to avoid high maintenance and repair cost of your instrument<\/strong><\/p>\n\n\n\n<p class=\"wp-block-paragraph\">As explained previously, filters which are not maintained on a regular interval can cause the outgoing gas to become more contaminated than the original source gas.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">The color indicators used in gas filters are so called \u2018last minute\u2019 indicators and require quick action.<\/p>\n\n\n\n<figure class=\"wp-block-image\"><img decoding=\"async\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-annual-replacement-of-cartridge-gas-filters-10.png\" alt=\"icon\" class=\"wp-image-48447\" title=\"-\"><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\">You can compare it to the engine oil indicator in your car. When the engine oil indicator is blinking on your dashboard, the car should not be driven and ignition should be switched off unless topped with engine oil &#8211; In event of taking risk to drive &#8211; high probability of engine getting seized causing high expenditure to repair or replace the engine.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">The same is valid for gas filters. When one of the indicators starts to change color, the filter should not be used and instrument should be switched off unless the filter is immediately being replaced with a new one &#8211; In event of taking risk to continue \u2013 there will be a high probability of contaminant breakthrough causing high expenditure on instrument maintenance as compared to planned annual preventive maintenance on your instrument and gas filters.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">So what you can do as an end-user is to always buy an additional set of filters to keep on standby in case the indicator starts changing color or to use a good preventive maintenance plan or tool (such as the electronic indicator) to replace your filters at least once a year.<\/p>\n","protected":false},"excerpt":{"rendered":"<p>A significant number of instrument and column complaints are carrier gas related, which could be caused by breakthrough from filters that are not replaced in time. The possible consequences of a filter breakthrough are: Gas distribution system behind the filter&#8230;<\/p>\n","protected":false},"author":65,"featured_media":0,"comment_status":"closed","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"_acf_changed":false,"_kad_blocks_custom_css":"","_kad_blocks_head_custom_js":"","_kad_blocks_body_custom_js":"","_kad_blocks_footer_custom_js":"","_kadence_starter_templates_imported_post":false,"_kad_post_transparent":"","_kad_post_title":"","_kad_post_layout":"","_kad_post_sidebar_id":"","_kad_post_content_style":"","_kad_post_vertical_padding":"","_kad_post_feature":"","_kad_post_feature_position":"","_kad_post_header":false,"_kad_post_footer":false,"footnotes":"","_ppma_block_editor_authors":""},"categories":[9],"tags":[],"industries-application":[],"post-badge":[],"resource-type":[565],"product-library":[],"resource-technique":[2294],"ppma_author":[599],"class_list":["post-51607","post","type-post","status-publish","format-standard","hentry","category-blogs","resource-type-maintenance","resource-technique-gas-chromatography-gc"],"acf":[],"taxonomy_info":{"category":[{"value":9,"label":"Blogs"}],"resource-type":[{"value":565,"label":"Maintenance"}],"resource-technique":[{"value":2294,"label":"Gas Chromatography (GC)"}]},"featured_image_src_large":false,"author_info":{"display_name":"SaVanna Cano","author_link":"https:\/\/discover.restek.com\/it\/author\/savanna-cano\/"},"comment_info":0,"category_info":[{"term_id":9,"name":"Blogs","slug":"blogs","term_group":0,"term_taxonomy_id":9,"taxonomy":"category","description":"","parent":0,"count":458,"filter":"raw","cat_ID":9,"category_count":458,"category_description":"","cat_name":"Blogs","category_nicename":"blogs","category_parent":0}],"tag_info":false,"authors":[{"term_id":599,"user_id":0,"is_guest":1,"slug":"mark-badger","display_name":"Mark Badger","avatar_url":"https:\/\/secure.gravatar.com\/avatar\/?s=96&d=mm&r=g","author_category":"1","first_name":"Mark","last_name":"Badger","user_url":"","job_title":"","description":""}],"_links":{"self":[{"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/posts\/51607","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/users\/65"}],"replies":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/comments?post=51607"}],"version-history":[{"count":3,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/posts\/51607\/revisions"}],"predecessor-version":[{"id":95583,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/posts\/51607\/revisions\/95583"}],"wp:attachment":[{"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/media?parent=51607"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/categories?post=51607"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/tags?post=51607"},{"taxonomy":"industries-application","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/industries-application?post=51607"},{"taxonomy":"post-badge","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/post-badge?post=51607"},{"taxonomy":"resource-type","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/resource-type?post=51607"},{"taxonomy":"product-library","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/product-library?post=51607"},{"taxonomy":"resource-technique","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/resource-technique?post=51607"},{"taxonomy":"author","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/ppma_author?post=51607"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}