{"id":51659,"date":"2022-06-22T14:30:00","date_gmt":"2022-06-22T14:30:00","guid":{"rendered":"https:\/\/discover.restek.com\/uncategorized\/derivatization-of-sugars-for-gc-ms-part-3-maintenance\/"},"modified":"2025-10-29T13:12:34","modified_gmt":"2025-10-29T13:12:34","slug":"derivatization-of-sugars-for-gc-ms-part-3-maintenance","status":"publish","type":"post","link":"https:\/\/discover.restek.com\/it\/blogs\/gnbl4642\/derivatization-of-sugars-for-gc-ms-part-3-maintenance","title":{"rendered":"Derivatization of sugars for GC-MS (Part 3): Maintenance"},"content":{"rendered":"\n<p class=\"wp-block-paragraph\">If you have tried out any of the derivatization procedures that I outlined back in&nbsp;Part 1, you also saw the beautiful colors these derivatization reactions produce (Figure 1). While the colors may initially be a nice change of pace from the clear solvents and solutions you\u2019re used to, color can be a \u2018red flag\u2019 that you will need to stay ahead of instrument maintenance. Judging by the deep red\/brown of that fructose TFA-oxime (Figure 1), I just know I will need to change the liner soon.<\/p>\n\n\n\n<figure class=\"wp-block-image size-large\"><img loading=\"lazy\" decoding=\"async\" width=\"1024\" height=\"632\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-derivatization-of-sugars-for-gc-ms-part-3-maintenance-01-1024x632.png\" alt=\"\" class=\"wp-image-13684\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-derivatization-of-sugars-for-gc-ms-part-3-maintenance-01-1024x632.png 1024w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-derivatization-of-sugars-for-gc-ms-part-3-maintenance-01-300x185.png 300w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-derivatization-of-sugars-for-gc-ms-part-3-maintenance-01-768x474.png 768w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-derivatization-of-sugars-for-gc-ms-part-3-maintenance-01.png 1445w\" sizes=\"auto, (max-width: 1024px) 100vw, 1024px\" \/><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\"><strong>Figure 1<\/strong>: Vial inserts containing TMS oximes (in ethyl acetate), TFA oximes (in ethyl acetate), and alditol acetates (in chloroform) of sugars, analyzed by GC-MS.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Starting with the syringe, make sure you are using solvents that can adequately clean the syringe and keep it from&nbsp;getting stuck. I like to rinse with my sample solvent (TFA\/TMS-oxime = ethyl acetate; alditol acetate = chloroform) immediately before and after injection. Ethyl acetate and chloroform are polar, so it may be a good idea to wash with something non-polar as well, like hexane or isooctane. Washing with a polar and non-polar solvent after injection will help keep the syringe clean from a wider range of potential-contaminants in the sample. It is much easier to wash with two solvents after every injection, than to waste time by only using the second solvent when the syringe gets stuck.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Next up is the liner. The damage wasn\u2019t as bad as I expected from that fructose TFA-oxime, but a liner change was certainly warranted. I cannot emphasize enough how important a liner with wool is when working with sugar derivatives. Not only will the wool help volatilize the sample, but it will also help protect your column from non-volatile contaminants. After &lt;50 1 uL injections (from admittedly concentrated samples), a darkened ring was visible at the top of the wool of my inlet liner, and a concentrated dark spot was visible on the wool directly below the syringe needle (Figure 2). If you are like me, and don\u2019t like having to pull out your liner to check if it needs to be changed, there are also some indirect signs the liner may&nbsp;need to be replaced. Are peaks heights decreasing? Is recovery of your internal standard starting to drop? Are peaks getting wider? If yes, then it\u2019s time to check the liner.<\/p>\n\n\n\n<figure class=\"wp-block-image\"><img decoding=\"async\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-derivatization-of-sugars-for-gc-ms-part-3-maintenance-02.png\" alt=\"blog-derivatization-of-sugars-for-GC-MS-part-3-maintenance-02.png\" title=\"-\"><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\"><strong>Figure 2:<\/strong>&nbsp;Side (left) and top-down (right) views of a liner after &lt;50 injections of concentrated derivatized sugar solutions.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Finally, the column. Excess derivatizing agents will strip away phase from the head of the column, which creates excess bleed and impedes analyte retention. I like to trim 10-20 cm (4-8 inches) from the column head when I see retention times start to shift. After trimming you can use the EZGC method translator to&nbsp;adjust your methods&nbsp;and match your previously used retention times, that way you don\u2019t have to worry about updating your analysis software.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">The syringe, liner, and column head should be your first stop for maintenance when analyzing derivatized sugars, but if your chromatograms still don\u2019t look right after maintenance, check out some of our other blogs to learn more about instrument maintenance and when to change and clean other parts of your GC system.<\/p>\n\n\n\n<div style=\"height:100px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n","protected":false},"excerpt":{"rendered":"<p>If you have tried out any of the derivatization procedures that I outlined back in Part 1, you also saw the beautiful colors these derivatization reactions produce. While the colors may initially be a nice change of pace from the clear solvents and solutions you&rsquo;re used to, color can be a &lsquo;red flag&rsquo; that you will need to stay ahead of instrument maintenance. Judging by the deep red\/brown of that fructose TFA-oxime, I just know I will need to change the liner soon.<\/p>\n","protected":false},"author":34,"featured_media":0,"comment_status":"closed","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"_acf_changed":false,"_kad_blocks_custom_css":"","_kad_blocks_head_custom_js":"","_kad_blocks_body_custom_js":"","_kad_blocks_footer_custom_js":"","_kadence_starter_templates_imported_post":false,"_kad_post_transparent":"","_kad_post_title":"","_kad_post_layout":"","_kad_post_sidebar_id":"","_kad_post_content_style":"","_kad_post_vertical_padding":"","_kad_post_feature":"","_kad_post_feature_position":"","_kad_post_header":false,"_kad_post_footer":false,"footnotes":"","_ppma_block_editor_authors":""},"categories":[9],"tags":[],"industries-application":[2153,2157,2167],"post-badge":[],"resource-type":[],"product-library":[],"resource-technique":[2294,2301],"ppma_author":[454],"class_list":["post-51659","post","type-post","status-publish","format-standard","hentry","category-blogs","industries-application-clinical","industries-application-energy","industries-application-food-beverage","resource-technique-gas-chromatography-gc","resource-technique-mass-spectrometry-ms"],"acf":[],"taxonomy_info":{"category":[{"value":9,"label":"Blogs"}],"industries-application":[{"value":2153,"label":"Clinical"},{"value":2157,"label":"Energy"},{"value":2167,"label":"Food &amp; Beverage"}],"resource-technique":[{"value":2294,"label":"Gas Chromatography (GC)"},{"value":2301,"label":"Mass Spectrometry (MS)"}]},"featured_image_src_large":false,"author_info":{"display_name":"Erica Pack, PhD","author_link":"https:\/\/discover.restek.com\/it\/author\/erica-pack-phd\/"},"comment_info":0,"category_info":[{"term_id":9,"name":"Blogs","slug":"blogs","term_group":0,"term_taxonomy_id":9,"taxonomy":"category","description":"","parent":0,"count":457,"filter":"raw","cat_ID":9,"category_count":457,"category_description":"","cat_name":"Blogs","category_nicename":"blogs","category_parent":0}],"tag_info":false,"authors":[{"term_id":454,"user_id":34,"is_guest":0,"slug":"erica-pack-phd","display_name":"Erica Pack, PhD","avatar_url":{"url":"https:\/\/discover.restek.com\/wp-content\/uploads\/people-pack-erica.jpg","url2x":"https:\/\/discover.restek.com\/wp-content\/uploads\/people-pack-erica.jpg"},"author_category":"1","first_name":"Erica","last_name":"Pack","user_url":"https:\/\/www.linkedin.com\/in\/ericapack\/","job_title":"Senior Scientist, GC Applied","description":"Erica is an interdisciplinary GC applications and technologies scientist at Restek. She obtained her bachelor's degree in forensic biology from The Pennsylvania State University, and her doctorate from Virginia Tech in plant pathology, physiology, and weed science. Since joining Restek in 2021, she has worked with a wide variety of GC columns, including fused silica, MXT, PLOT, and packed columns as well as accessories, such as liners, valves, and methanizers."}],"_links":{"self":[{"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/posts\/51659","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/users\/34"}],"replies":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/comments?post=51659"}],"version-history":[{"count":3,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/posts\/51659\/revisions"}],"predecessor-version":[{"id":61592,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/posts\/51659\/revisions\/61592"}],"wp:attachment":[{"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/media?parent=51659"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/categories?post=51659"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/tags?post=51659"},{"taxonomy":"industries-application","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/industries-application?post=51659"},{"taxonomy":"post-badge","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/post-badge?post=51659"},{"taxonomy":"resource-type","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/resource-type?post=51659"},{"taxonomy":"product-library","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/product-library?post=51659"},{"taxonomy":"resource-technique","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/resource-technique?post=51659"},{"taxonomy":"author","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/ppma_author?post=51659"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}