{"id":51733,"date":"2023-09-28T14:30:00","date_gmt":"2023-09-28T14:30:00","guid":{"rendered":"https:\/\/discover.restek.com\/uncategorized\/just-dont-do-it-dont-overtighten-your-lc-direct-connect-guard\/"},"modified":"2026-09-08T14:04:21","modified_gmt":"2026-09-08T14:04:21","slug":"just-dont-do-it-dont-overtighten-your-lc-direct-connect-guard","status":"publish","type":"post","link":"https:\/\/discover.restek.com\/it\/blog-fr\/gnbl4726\/just-dont-do-it-dont-overtighten-your-lc-direct-connect-guard","title":{"rendered":"Just don\u2019t do it!! Don\u2019t overtighten your LC direct connect guard."},"content":{"rendered":"\n<p class=\"wp-block-paragraph\">Overtightening High-Performance Liquid Chromatography <a href=\"https:\/\/www.restek.com\/c\/1277\" target=\"_blank\" rel=\"noopener\">(HPLC) guard columns<\/a> to HPLC columns can cause several issues, leading to suboptimal chromatographic performance and potential damage to the equipment.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">We offer three distinct direct-connect guard systems: EXP Holders, ROC Holders, and Trident Holders. It depends on the HPLC phase you are using. For more details, please refer here.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\"><a href=\"https:\/\/discover.restek.com\/GNSS2975\/how-to-choose-the-right-lc-guard-column\/\">How to Choose the Right LC Guard Column<\/a><\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Guard columns are small, protective columns placed before the main analytical column to protect it from sample debris and HPLC system contaminants. Overtightening the guard column can cause physical damage to the PEEK ferrule. Excessive tightening can cause the PEEK ferrule to partially block the inlet frit of the HPLC column causing elevated backpressure. Overtightening can make it extremely challenging to remove the guard column or the HPLC column during maintenance or when switching columns. This difficulty can result in accidental damage to the column threads or the column itself (unable to remove ferrule).<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">What are the consequences of overtightening on the PEEK ferrule and guard cartridge? Below are some images illustrating the effects of excessive tightening on the components of the Trident guard holder and cartridge.<\/p>\n\n\n\n<div class=\"wp-block-columns is-layout-flex wp-container-core-columns-is-layout-8f761849 wp-block-columns-is-layout-flex\">\n<div class=\"wp-block-column is-layout-flow wp-block-column-is-layout-flow\">\n<div class=\"wp-block-columns is-layout-flex wp-container-core-columns-is-layout-8f761849 wp-block-columns-is-layout-flex\">\n<div class=\"wp-block-column is-layout-flow wp-block-column-is-layout-flow\">\n<p class=\"wp-block-paragraph\">Figure 1<\/p>\n\n\n\n<figure class=\"wp-block-image size-full\"><img loading=\"lazy\" decoding=\"async\" width=\"251\" height=\"200\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-just-dont-do-it-dont-overtighten-your-lc-direct-connect-guard-01.jpg\" alt=\"PEEK ferrule immovably embedded within an HPLC column\" class=\"wp-image-15568\" title=\"-\"><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\">PEEK ferrule stuck in the HPLC column<br><a href=\"https:\/\/www.restek.com\/p\/27476\" target=\"_blank\" rel=\"noopener\">27476<\/a><\/p>\n<\/div>\n\n\n\n<div class=\"wp-block-column is-layout-flow wp-block-column-is-layout-flow\">\n<p class=\"wp-block-paragraph\">Figure 2<\/p>\n\n\n\n<figure class=\"wp-block-image size-full\"><img loading=\"lazy\" decoding=\"async\" width=\"250\" height=\"200\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-just-dont-do-it-dont-overtighten-your-lc-direct-connect-guard-02.jpg\" alt=\"impression on the guard cartridge resulting from excessive tightening\" class=\"wp-image-15574\" title=\"-\"><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\">Guard Cartridge inlet side.<\/p>\n<\/div>\n<\/div>\n<\/div>\n\n\n\n<div class=\"wp-block-column is-layout-flow wp-block-column-is-layout-flow\">\n<div class=\"wp-block-columns is-layout-flex wp-container-core-columns-is-layout-8f761849 wp-block-columns-is-layout-flex\">\n<div class=\"wp-block-column is-layout-flow wp-block-column-is-layout-flow\">\n<p class=\"wp-block-paragraph\">Figure 3<\/p>\n\n\n\n<figure class=\"wp-block-image size-full\"><img loading=\"lazy\" decoding=\"async\" width=\"251\" height=\"200\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-just-dont-do-it-dont-overtighten-your-lc-direct-connect-guard-03.jpg\" alt=\"PEEK ferrule damaged due to overtightening\" class=\"wp-image-15580\" title=\"-\"><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\">Ferrule that has been overtightened<\/p>\n<\/div>\n\n\n\n<div class=\"wp-block-column is-layout-flow wp-block-column-is-layout-flow\">\n<p class=\"wp-block-paragraph\">Figure 4<\/p>\n\n\n\n<figure class=\"wp-block-image size-full\"><img loading=\"lazy\" decoding=\"async\" width=\"250\" height=\"200\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-just-dont-do-it-dont-overtighten-your-lc-direct-connect-guard-04.jpg\" alt=\"unused PEEK ferrule (catalog # 27476)\" class=\"wp-image-15586\" title=\"-\"><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\">&nbsp;<\/p>\n<\/div>\n<\/div>\n<\/div>\n<\/div>\n\n\n\n<p class=\"wp-block-paragraph\">Figure 1: displays the PEEK ferrule immovably embedded within an HPLC column (column will need to be discarded).<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Figure 2: exhibits the impression on the guard cartridge resulting from excessive tightening.&nbsp;<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Figure 3: showcases a photograph of a PEEK ferrule damaged due to overtightening.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Figure 4: illustrates an unused PEEK ferrule (<a href=\"https:\/\/www.restek.com\/p\/27476\" target=\"_blank\" rel=\"noopener\">catalog 27476<\/a>).<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">To prevent these problems, it is essential to adhere to our guidelines for attaching the guard column to the HPLC column. Usually, hand-tightening should suffice. A wrench may be necessary to ensure a leak-free connection, but it should only be turned an additional 1\/8 to \u00bc of a rotation beyond hand-tight. Please see the instruction sheets for the three different holders for more details.&nbsp;<\/p>\n\n\n\n<div data-wp-interactive=\"core\/file\" class=\"wp-block-file\"><object data-wp-bind--hidden=\"!state.hasPdfPreview\" hidden class=\"wp-block-file__embed\" data=\"https:\/\/discover.restek.com\/wp-content\/uploads\/730-40-001.pdf\" type=\"application\/pdf\" style=\"width:100%;height:600px\" aria-label=\"Embed of 730-40-001.\"><\/object><a id=\"wp-block-file--media-56768488-e6b6-414e-9cad-a064d7e5a008\" href=\"https:\/\/discover.restek.com\/wp-content\/uploads\/730-40-001.pdf\">730-40-001<\/a><a href=\"https:\/\/discover.restek.com\/wp-content\/uploads\/730-40-001.pdf\" class=\"wp-block-file__button wp-element-button\" aria-describedby=\"wp-block-file--media-56768488-e6b6-414e-9cad-a064d7e5a008\" download>Download<\/a><\/div>\n\n\n\n<div data-wp-interactive=\"core\/file\" class=\"wp-block-file\"><object data-wp-bind--hidden=\"!state.hasPdfPreview\" hidden class=\"wp-block-file__embed\" data=\"https:\/\/discover.restek.com\/wp-content\/uploads\/730-40-002.pdf\" type=\"application\/pdf\" style=\"width:100%;height:600px\" aria-label=\"Embed of 730-40-002.\"><\/object><a id=\"wp-block-file--media-d851d1f0-95a5-44a0-bac9-f71a2bf9c3a7\" href=\"https:\/\/discover.restek.com\/wp-content\/uploads\/730-40-002.pdf\">730-40-002<\/a><a href=\"https:\/\/discover.restek.com\/wp-content\/uploads\/730-40-002.pdf\" class=\"wp-block-file__button wp-element-button\" aria-describedby=\"wp-block-file--media-d851d1f0-95a5-44a0-bac9-f71a2bf9c3a7\" download>Download<\/a><\/div>\n\n\n\n<div data-wp-interactive=\"core\/file\" class=\"wp-block-file\"><object data-wp-bind--hidden=\"!state.hasPdfPreview\" hidden class=\"wp-block-file__embed\" data=\"https:\/\/discover.restek.com\/wp-content\/uploads\/730-10-006.pdf\" type=\"application\/pdf\" style=\"width:100%;height:600px\" aria-label=\"Embed of 730-10-006.\"><\/object><a id=\"wp-block-file--media-43ee9327-cb38-4aa2-9a12-7f09527329ce\" href=\"https:\/\/discover.restek.com\/wp-content\/uploads\/730-10-006.pdf\">730-10-006<\/a><a href=\"https:\/\/discover.restek.com\/wp-content\/uploads\/730-10-006.pdf\" class=\"wp-block-file__button wp-element-button\" aria-describedby=\"wp-block-file--media-43ee9327-cb38-4aa2-9a12-7f09527329ce\" download>Download<\/a><\/div>\n\n\n\n<p class=\"wp-block-paragraph\">Furthermore, it is equally crucial to routinely inspect and replace the ferrule to maintain effective sealing without the risk of overtightening. You will notice a new ferrule (figure 4) to start to expand and look like figure 3 when it needs to be replaced. Replace ferrule every few times the guard holder is removed or shows signs of wear.&nbsp;<\/p>\n\n\n        <div class=\"cpb\">\n            <h3 class=\"cpb-heading\">Replacement ferrules:<\/h3>\n            <hr class=\"cpb-heading-underline\" \/>\n            <div class=\"cpb-list\">\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/25935\">                                N\u00b0 Catalogo 25935                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Ferrula in titanio ibrido EXP<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/25935\">Esplora Prodotto<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/26391\">                                N\u00b0 Catalogo 26391                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Ferrule di ricambio in PEEK per holder per precolonne LC Roc, 3 pz.<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/26391\">Esplora Prodotto<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/27476\">                                N\u00b0 Catalogo 27476                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Ferrule di ricambio per sistema Trident, PEEK, 10 pz.<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/27476\">Esplora Prodotto<\/a>\n                                                    <\/div>\n                    <\/div>\n                            <\/div>\n        <\/div>\n        \n\n\n<h2 class=\"wp-block-heading\">Further information:<\/h2>\n\n\n\n<figure class=\"wp-block-embed is-type-video is-provider-youtube wp-block-embed-youtube wp-embed-aspect-16-9 wp-has-aspect-ratio\"><div class=\"wp-block-embed__wrapper\">\n<iframe loading=\"lazy\" title=\"Do I Need an LC Guard Column?\" width=\"720\" height=\"405\" src=\"https:\/\/www.youtube.com\/embed\/vnMtEybkBUo?feature=oembed\" frameborder=\"0\" allow=\"accelerometer; autoplay; clipboard-write; encrypted-media; gyroscope; picture-in-picture; web-share\" referrerpolicy=\"strict-origin-when-cross-origin\" allowfullscreen><\/iframe>\n<\/div><\/figure>\n\n\n\n<div style=\"height:6px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<div style=\"height:100px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n","protected":false},"excerpt":{"rendered":"<p>Overtightening High-Performance Liquid Chromatography (HPLC) guard columns to HPLC columns can cause several issues, leading to suboptimal chromatographic performance and potential damage to the equipment. In this post, we look at the consequences of overtightening a PEEK ferrule and guard cartridge.<\/p>\n","protected":false},"author":33,"featured_media":0,"comment_status":"closed","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"_acf_changed":false,"_kad_blocks_custom_css":"","_kad_blocks_head_custom_js":"","_kad_blocks_body_custom_js":"","_kad_blocks_footer_custom_js":"","_kadence_starter_templates_imported_post":false,"_kad_post_transparent":"","_kad_post_title":"","_kad_post_layout":"","_kad_post_sidebar_id":"","_kad_post_content_style":"","_kad_post_vertical_padding":"","_kad_post_feature":"","_kad_post_feature_position":"","_kad_post_header":false,"_kad_post_footer":false,"footnotes":"","_ppma_block_editor_authors":""},"categories":[791],"tags":[],"industries-application":[],"post-badge":[],"resource-type":[],"product-library":[],"resource-technique":[2318],"ppma_author":[453,615],"class_list":["post-51733","post","type-post","status-publish","format-standard","hentry","category-blog-fr","resource-technique-chromatographie-en-phase-liquide"],"acf":[],"taxonomy_info":{"category":[{"value":791,"label":"Blog"}],"resource-technique":[{"value":2318,"label":"Chromatographie en phase liquide"}]},"featured_image_src_large":false,"author_info":{"display_name":"Gary Oden","author_link":"https:\/\/discover.restek.com\/it\/author\/gary-oden\/"},"comment_info":0,"category_info":[{"term_id":791,"name":"Blog","slug":"blog-fr","term_group":0,"term_taxonomy_id":791,"taxonomy":"category","description":"","parent":0,"count":457,"filter":"raw","cat_ID":791,"category_count":457,"category_description":"","cat_name":"Blog","category_nicename":"blog-fr","category_parent":0}],"tag_info":false,"authors":[{"term_id":453,"user_id":33,"is_guest":0,"slug":"gary-oden","display_name":"Gary Oden","avatar_url":{"url":"https:\/\/discover.restek.com\/wp-content\/uploads\/people-oden-gary.jpg","url2x":"https:\/\/discover.restek.com\/wp-content\/uploads\/people-oden-gary.jpg"},"author_category":"1","first_name":"Gary","last_name":"Oden","user_url":"https:\/\/www.linkedin.com\/in\/gary-oden-jr-a35ba74b\/","job_title":"Technical Service, Specialist I","description":"Gary has 20 years of hands-on experience with chromatography, including SPE, HPLC, and LC\/MS instrumentation. He began his chromatography career in the laboratory as a bonding chemist and has served as a study director for PFAS analysis, specializing in both bioanalytical and environmental fields. His work in developing HPLC and LC\/MS-MS applications has further broadened his expertise in chromatography. Since 2018, Gary has been a key member of Restek\u2019s European-based technical support team, where he offers expert assistance to customers in overcoming complex analytical challenges."},{"term_id":615,"user_id":59,"is_guest":0,"slug":"todd-shaffer","display_name":"Todd Shaffer","avatar_url":"https:\/\/secure.gravatar.com\/avatar\/?s=96&d=mm&r=g","author_category":"1","first_name":"Todd","last_name":"Shaffer","user_url":"","job_title":"","description":""}],"_links":{"self":[{"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/posts\/51733","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/users\/33"}],"replies":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/comments?post=51733"}],"version-history":[{"count":8,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/posts\/51733\/revisions"}],"predecessor-version":[{"id":102808,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/posts\/51733\/revisions\/102808"}],"wp:attachment":[{"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/media?parent=51733"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/categories?post=51733"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/tags?post=51733"},{"taxonomy":"industries-application","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/industries-application?post=51733"},{"taxonomy":"post-badge","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/post-badge?post=51733"},{"taxonomy":"resource-type","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/resource-type?post=51733"},{"taxonomy":"product-library","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/product-library?post=51733"},{"taxonomy":"resource-technique","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/resource-technique?post=51733"},{"taxonomy":"author","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/ppma_author?post=51733"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}