{"id":51802,"date":"2019-10-13T14:30:00","date_gmt":"2019-10-13T14:30:00","guid":{"rendered":"https:\/\/discover.restek.com\/uncategorized\/terpene-analysis-approaches-part-i\/"},"modified":"2026-06-17T19:45:43","modified_gmt":"2026-06-17T19:45:43","slug":"terpene-analysis-approaches-part-i","status":"publish","type":"post","link":"https:\/\/discover.restek.com\/it\/blog-fr\/gnbl4807\/terpene-analysis-approaches-part-i","title":{"rendered":"Terpene Analysis Approaches &#8211; Part I"},"content":{"rendered":"\n<p class=\"wp-block-paragraph\">Welcome to the newest blog series that totally stinks! In this series we\u2019re going to look at different approaches for analyzing terpenes in cannabis. To quickly brief those who may not be familiar with terpenes, this class of compounds is responsible for many of the flavors and aromas in cannabis. Not only are cannabis growers and manufacturers interested in terpene content though; researchers also have interest in cannabis terpenes because of their potential therapeutic benefits. It is important to classify terpenes in cannabis to better understand the Entourage Effect. This term refers to the synergistic interactions between cannabinoids and terpenes with respect to physical ailments.<sup>1<\/sup><\/p>\n\n\n\n<p class=\"wp-block-paragraph\">A common approach to analyzing these compounds is by static headspace (HS) \u2013 gas chromatography (GC) \u2013 mass spectrometry (MS). This can either be done with a HS Autosampler or a gas-tight syringe, but today, we are going to compare this technique to solid phase microextraction (SPME) with the SPME Arrow. For our initial comparison, we diluted our Terpene Standard Mix 1 &amp; 2 (cat#\u00a034095\u00a0&amp;\u00a034096) to 1 \u00b5g\/mL, added 1 mL of this to a 20 mL HS vial, then analyzed samples via HS-Syringe-GC-MS and HS-SPME-GC-MS. You\u2019ll notice that the parameters are very different for each approach, because we used the optimal parameters for each technique.<\/p>\n\n\n\n<div style=\"height:20px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n<div class=\"wp-block-image\">\n<figure class=\"alignleft size-full\"><img loading=\"lazy\" decoding=\"async\" width=\"530\" height=\"516\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-terpene-analysis-approaches-part-i-01.jpg\" alt=\"\" class=\"wp-image-16918\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-terpene-analysis-approaches-part-i-01.jpg 530w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-terpene-analysis-approaches-part-i-01-300x292.jpg 300w\" sizes=\"auto, (max-width: 530px) 100vw, 530px\" \/><\/figure>\n<\/div>\n\n\n<div style=\"height:20px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n<div class=\"wp-block-image\">\n<figure class=\"alignleft size-full\"><img loading=\"lazy\" decoding=\"async\" width=\"757\" height=\"526\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-terpene-analysis-approaches-part-i-02.jpg\" alt=\"\" class=\"wp-image-16924\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-terpene-analysis-approaches-part-i-02.jpg 757w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-terpene-analysis-approaches-part-i-02-300x208.jpg 300w\" sizes=\"auto, (max-width: 757px) 100vw, 757px\" \/><\/figure>\n<\/div>\n\n\n<div style=\"height:20px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<figure class=\"wp-block-image size-large\"><img loading=\"lazy\" decoding=\"async\" width=\"1024\" height=\"560\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-terpene-analysis-approaches-part-i-03-1024x560.jpg\" alt=\"\" class=\"wp-image-16930\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-terpene-analysis-approaches-part-i-03-1024x560.jpg 1024w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-terpene-analysis-approaches-part-i-03-300x164.jpg 300w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-terpene-analysis-approaches-part-i-03-768x420.jpg 768w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-terpene-analysis-approaches-part-i-03.jpg 1132w\" sizes=\"auto, (max-width: 1024px) 100vw, 1024px\" \/><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\">Figure. Analysis of 23 component terpene standard.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Results from each approach can be seen above. As you can tell from the data collected, both approaches were far from awesome. We were only able to identify 13 of the 23 terpenes in the standard. You\u2019ll notice that we were able to collect data for many of the monoterpenes, but not so much for the sesquiterpenes. For the compounds that were able to be identified; the HS-Syringe performed slightly better than the HS-SPME Arrow. However, the HS-SPME Arrow had better % RSDs than the HS-Syringe. So, why were we only able to identify 13 of the 23 terpenes and how can we improve this? Tune in next time!<\/p>\n\n\n\n<ol class=\"wp-block-list\">\n<li>Russo, E. B. (2011), Taming THC: potential cannabis synergy and phytocannabinoid\u2010terpenoid entourage effects. British Journal of Pharmacology, 163: 1344-1364.<\/li>\n<\/ol>\n\n\n\n<div style=\"height:50px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n        <div class=\"cpb\">\n            <h3 class=\"cpb-heading\">Prodotti Menzionati<\/h3>\n            <hr class=\"cpb-heading-underline\" \/>\n            <div class=\"cpb-list\">\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/34095\">                                N\u00b0 Catalogo 34095                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Standard 1 terpeni cannabis medica, 2500 \u00b5g\/mL ognuno in Isopropanolo, 1 ml\/ampolla<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/34095\">Esplora Prodotto<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/34096\">                                N\u00b0 Catalogo 34096                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Standard 2 terpeni cannabis medica, 2500 \u00b5g\/mL ognuno in Isopropanolo, 1 ml\/ampolla<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/34096\">Esplora Prodotto<\/a>\n                                                    <\/div>\n                    <\/div>\n                            <\/div>\n        <\/div>\n        \n\n\n<div style=\"height:100px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n","protected":false},"excerpt":{"rendered":"<p>Welcome to the newest blog series that totally stinks! In this series we&rsquo;re going to look at different approaches for analyzing terpenes in cannabis. To quickly brief those who may not be familiar with terpenes, this class of compounds is&#8230;<\/p>\n","protected":false},"author":32,"featured_media":0,"comment_status":"closed","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"_acf_changed":false,"_kad_blocks_custom_css":"","_kad_blocks_head_custom_js":"","_kad_blocks_body_custom_js":"","_kad_blocks_footer_custom_js":"","_kadence_starter_templates_imported_post":false,"_kad_post_transparent":"","_kad_post_title":"","_kad_post_layout":"","_kad_post_sidebar_id":"","_kad_post_content_style":"","_kad_post_vertical_padding":"","_kad_post_feature":"","_kad_post_feature_position":"","_kad_post_header":false,"_kad_post_footer":false,"footnotes":"","_ppma_block_editor_authors":""},"categories":[791],"tags":[],"industries-application":[2279,2198],"post-badge":[],"resource-type":[],"product-library":[],"resource-technique":[2316,2358,2320,2322],"ppma_author":[452],"class_list":["post-51802","post","type-post","status-publish","format-standard","hentry","category-blog-fr","industries-application-cannabis-fr","industries-application-cannabis-medical-recreatif","resource-technique-chromatographie-en-phase-gazeuse-gc","resource-technique-headspace-fr","resource-technique-spectrometrie-de-masse-ms","resource-technique-spme-solid-phase-microextraction"],"acf":[],"taxonomy_info":{"category":[{"value":791,"label":"Blog"}],"industries-application":[{"value":2279,"label":"Cannabis"},{"value":2198,"label":"Cannabis m\u00e9dical ou r\u00e9cr\u00e9atif"}],"resource-technique":[{"value":2316,"label":"Chromatographie en phase gazeuse (GC)"},{"value":2358,"label":"Headspace"},{"value":2320,"label":"Spectrom\u00e9trie de masse (MS)"},{"value":2322,"label":"SPME (Solid Phase Microextraction)"}]},"featured_image_src_large":false,"author_info":{"display_name":"Colton Myers","author_link":"https:\/\/discover.restek.com\/it\/author\/colton-myers\/"},"comment_info":0,"category_info":[{"term_id":791,"name":"Blog","slug":"blog-fr","term_group":0,"term_taxonomy_id":791,"taxonomy":"category","description":"","parent":0,"count":457,"filter":"raw","cat_ID":791,"category_count":457,"category_description":"","cat_name":"Blog","category_nicename":"blog-fr","category_parent":0}],"tag_info":false,"authors":[{"term_id":452,"user_id":32,"is_guest":0,"slug":"colton-myers","display_name":"Colton Myers","avatar_url":{"url":"https:\/\/discover.restek.com\/wp-content\/uploads\/people-myers-colton.jpg","url2x":"https:\/\/discover.restek.com\/wp-content\/uploads\/people-myers-colton.jpg"},"author_category":"1","first_name":"Colton","last_name":"Myers","user_url":"https:\/\/www.linkedin.com\/in\/myerscolton\/","job_title":"Product Manager - New Product Development","description":"Colton Myers is a Product Manager for GC columns at Restek Corporation with over 10 years of experience in research and development. He specializes in driving new GC column technologies from concept to market, combining deep technical expertise with a strong focus on innovation and commercialization. Colton holds a BS in chemistry from Juniata College and a Master of Management in Corporate Innovation and Entrepreneurship from Penn State University."}],"_links":{"self":[{"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/posts\/51802","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/users\/32"}],"replies":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/comments?post=51802"}],"version-history":[{"count":7,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/posts\/51802\/revisions"}],"predecessor-version":[{"id":73040,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/posts\/51802\/revisions\/73040"}],"wp:attachment":[{"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/media?parent=51802"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/categories?post=51802"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/tags?post=51802"},{"taxonomy":"industries-application","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/industries-application?post=51802"},{"taxonomy":"post-badge","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/post-badge?post=51802"},{"taxonomy":"resource-type","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/resource-type?post=51802"},{"taxonomy":"product-library","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/product-library?post=51802"},{"taxonomy":"resource-technique","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/resource-technique?post=51802"},{"taxonomy":"author","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/ppma_author?post=51802"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}