{"id":76404,"date":"2014-08-01T00:00:00","date_gmt":"2014-08-01T00:00:00","guid":{"rendered":"https:\/\/discover.restek.com\/?p=76404"},"modified":"2026-01-23T23:58:13","modified_gmt":"2026-01-23T23:58:13","slug":"calculating-splitless-valve-time-for-splitless-injection-gc-with-the-new-ezgc-method-translator-and-flow-calculator-pesticides","status":"publish","type":"post","link":"https:\/\/discover.restek.com\/it\/blogs\/gnbl5473\/calculating-splitless-valve-time-for-splitless-injection-gc-with-the-new-ezgc-method-translator-and-flow-calculator-pesticides","title":{"rendered":"Calculating Splitless Valve Time for Splitless Injection GC with the New EZGC Method Translator and Flow Calculator &#8211; Pesticides"},"content":{"rendered":"\n<p class=\"wp-block-paragraph\">Properly setting the splitless valve time (SVT) for splitless injection in gas chromatography (GC) is probably one of the most overlooked optimization parameters for pesticide analysis today. &nbsp;Having too short of a SVT leads to poor transfer of active and involatile pesticides, especially as the liner becomes contaminated with \u201cdirt\u201d from injected sample extracts.&nbsp; This leads to poor quantitative results and possibly the premature need to recalibrate the GC.&nbsp; The reason a proper SVT is so important has to do with the long sample transfer time from the inlet to the GC column inherent in splitless injection.&nbsp; In general, the flow through the inlet is in the same order-of-magnitude as the capillary GC column outlet flow, which means it\u2019s slow.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">A rule-of-thumb for setting the SVT (sometimes known too as purge valve time) is to sweep the inlet liner completely with carrier gas 1.5 to 2 times before opening the split valve.\u00a0 This range provides a buffer for quantitative transfer as the liner\/seal gets dirty and analytes have to \u201cchromatograph\u201d out of the dirt to get to the column, while keeping the transfer time short enough to avoid a long solvent tail that might obscure earlier eluting analytes.\u00a0 We provide an easy way to calculate splitless valve time through our new\u00a0<a href=\"https:\/\/ez.restek.com\/ezgc-mtfc\/en\" target=\"_blank\" rel=\"noopener\"><em>EZ<\/em>GC<sup>\u00a0<\/sup>Method Translator and Flow Calculator<\/a>\u00a0(MTFC).<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Using the Flow Calculator:<\/p>\n\n\n\n<ul class=\"wp-block-list\">\n<li>Enter your Column length, inner diameter, film thickness, and starting oven temperature.<\/li>\n\n\n\n<li>Enter your Column outlet flow, and click \u201cAtm\u201d or \u201cVacuum\u201d depending on whether you\u2019re using an ambient pressure (Atm) detector like ECD, or a vacuum-outlet detector like a mass spectrometer (MS).<\/li>\n\n\n\n<li>Enter your inlet temperature and your inlet liner volume (this can be calculated or if you\u2019re using an HP or Agilent GC inlet liner, you can find liner volumes in the MTFC Glossary).<\/li>\n\n\n\n<li>The Splitless Valve Time range that should be used is calculated automatically.<\/li>\n<\/ul>\n\n\n\n<p class=\"wp-block-paragraph\">I\u2019ve outlined the steps graphically below.&nbsp; &nbsp;<\/p>\n\n\n<style>.kb-image76404_f38c89-a1 .kb-image-has-overlay:after{opacity:0.3;}<\/style>\n<figure class=\"wp-block-kadence-image kb-image76404_f38c89-a1 size-medium_large\"><img loading=\"lazy\" decoding=\"async\" width=\"768\" height=\"585\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-calculating-splitless-valve-time-for-splitless-injection-gc-with-the-new-ezgc-method-translator-and-flow-calculator-pesticides-01-768x585.jpg\" alt=\"\" class=\"kb-img wp-image-77538\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-calculating-splitless-valve-time-for-splitless-injection-gc-with-the-new-ezgc-method-translator-and-flow-calculator-pesticides-01-768x585.jpg 768w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-calculating-splitless-valve-time-for-splitless-injection-gc-with-the-new-ezgc-method-translator-and-flow-calculator-pesticides-01-300x229.jpg 300w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-calculating-splitless-valve-time-for-splitless-injection-gc-with-the-new-ezgc-method-translator-and-flow-calculator-pesticides-01-1024x781.jpg 1024w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-calculating-splitless-valve-time-for-splitless-injection-gc-with-the-new-ezgc-method-translator-and-flow-calculator-pesticides-01.jpg 1228w\" sizes=\"auto, (max-width: 768px) 100vw, 768px\" \/><\/figure>\n\n\n<style>.kb-image76404_953e62-c4 .kb-image-has-overlay:after{opacity:0.3;}<\/style>\n<figure class=\"wp-block-kadence-image kb-image76404_953e62-c4 size-medium_large\"><img loading=\"lazy\" decoding=\"async\" width=\"768\" height=\"588\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-calculating-splitless-valve-time-for-splitless-injection-gc-with-the-new-ezgc-method-translator-and-flow-calculator-pesticides-02-768x588.jpg\" alt=\"\" class=\"kb-img wp-image-77532\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-calculating-splitless-valve-time-for-splitless-injection-gc-with-the-new-ezgc-method-translator-and-flow-calculator-pesticides-02-768x588.jpg 768w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-calculating-splitless-valve-time-for-splitless-injection-gc-with-the-new-ezgc-method-translator-and-flow-calculator-pesticides-02-300x230.jpg 300w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-calculating-splitless-valve-time-for-splitless-injection-gc-with-the-new-ezgc-method-translator-and-flow-calculator-pesticides-02-1024x785.jpg 1024w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-calculating-splitless-valve-time-for-splitless-injection-gc-with-the-new-ezgc-method-translator-and-flow-calculator-pesticides-02.jpg 1476w\" sizes=\"auto, (max-width: 768px) 100vw, 768px\" \/><\/figure>\n\n\n<style>.kb-image76404_20ff4d-d5 .kb-image-has-overlay:after{opacity:0.3;}<\/style>\n<figure class=\"wp-block-kadence-image kb-image76404_20ff4d-d5 size-medium_large\"><img loading=\"lazy\" decoding=\"async\" width=\"768\" height=\"577\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-calculating-splitless-valve-time-for-splitless-injection-gc-with-the-new-ezgc-method-translator-and-flow-calculator-pesticides-03-768x577.jpg\" alt=\"\" class=\"kb-img wp-image-77526\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-calculating-splitless-valve-time-for-splitless-injection-gc-with-the-new-ezgc-method-translator-and-flow-calculator-pesticides-03-768x577.jpg 768w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-calculating-splitless-valve-time-for-splitless-injection-gc-with-the-new-ezgc-method-translator-and-flow-calculator-pesticides-03-300x225.jpg 300w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-calculating-splitless-valve-time-for-splitless-injection-gc-with-the-new-ezgc-method-translator-and-flow-calculator-pesticides-03-1024x769.jpg 1024w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-calculating-splitless-valve-time-for-splitless-injection-gc-with-the-new-ezgc-method-translator-and-flow-calculator-pesticides-03.jpg 1506w\" sizes=\"auto, (max-width: 768px) 100vw, 768px\" \/><\/figure>\n\n\n<style>.kb-image76404_99f443-15 .kb-image-has-overlay:after{opacity:0.3;}<\/style>\n<figure class=\"wp-block-kadence-image kb-image76404_99f443-15 size-medium_large\"><img loading=\"lazy\" decoding=\"async\" width=\"768\" height=\"544\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-calculating-splitless-valve-time-for-splitless-injection-gc-with-the-new-ezgc-method-translator-and-flow-calculator-pesticides-04-768x544.jpg\" alt=\"\" class=\"kb-img wp-image-77520\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-calculating-splitless-valve-time-for-splitless-injection-gc-with-the-new-ezgc-method-translator-and-flow-calculator-pesticides-04-768x544.jpg 768w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-calculating-splitless-valve-time-for-splitless-injection-gc-with-the-new-ezgc-method-translator-and-flow-calculator-pesticides-04-300x213.jpg 300w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-calculating-splitless-valve-time-for-splitless-injection-gc-with-the-new-ezgc-method-translator-and-flow-calculator-pesticides-04-1024x726.jpg 1024w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-calculating-splitless-valve-time-for-splitless-injection-gc-with-the-new-ezgc-method-translator-and-flow-calculator-pesticides-04.jpg 1506w\" sizes=\"auto, (max-width: 768px) 100vw, 768px\" \/><\/figure>\n\n\n\n<div style=\"height:100px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n","protected":false},"excerpt":{"rendered":"<p>Properly setting the splitless valve time (SVT) for splitless injection in gas chromatography (GC) is probably one of the most overlooked optimization parameters for pesticide analysis today. Having too short of a SVT leads to poor transfer of active and&#8230;<\/p>\n","protected":false},"author":5,"featured_media":0,"comment_status":"closed","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"_acf_changed":false,"_kad_blocks_custom_css":"","_kad_blocks_head_custom_js":"","_kad_blocks_body_custom_js":"","_kad_blocks_footer_custom_js":"","_kadence_starter_templates_imported_post":false,"_kad_post_transparent":"","_kad_post_title":"","_kad_post_layout":"","_kad_post_sidebar_id":"","_kad_post_content_style":"","_kad_post_vertical_padding":"","_kad_post_feature":"","_kad_post_feature_position":"","_kad_post_header":false,"_kad_post_footer":false,"footnotes":"","_ppma_block_editor_authors":""},"categories":[2806],"tags":[],"industries-application":[],"post-badge":[],"resource-type":[],"product-library":[],"resource-technique":[],"ppma_author":[590],"class_list":["post-76404","post","type-post","status-publish","format-standard","hentry","category-blogs"],"acf":[],"taxonomy_info":{"category":[{"value":2806,"label":"Blogs"}]},"featured_image_src_large":false,"author_info":{"display_name":"Brenda Seprish","author_link":"https:\/\/discover.restek.com\/it\/author\/brenda-coganrestek-com\/"},"comment_info":0,"category_info":[{"term_id":2806,"name":"Blogs","slug":"blogs","term_group":0,"term_taxonomy_id":2806,"taxonomy":"category","description":"","parent":0,"count":457,"filter":"raw","cat_ID":2806,"category_count":457,"category_description":"","cat_name":"Blogs","category_nicename":"blogs","category_parent":0}],"tag_info":false,"authors":[{"term_id":590,"user_id":0,"is_guest":1,"slug":"jack-cochran","display_name":"Jack Cochran","avatar_url":"https:\/\/secure.gravatar.com\/avatar\/?s=96&d=mm&r=g","author_category":"","first_name":"","last_name":"","user_url":"","job_title":"","description":""}],"_links":{"self":[{"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/posts\/76404","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/users\/5"}],"replies":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/comments?post=76404"}],"version-history":[{"count":5,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/posts\/76404\/revisions"}],"predecessor-version":[{"id":84425,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/posts\/76404\/revisions\/84425"}],"wp:attachment":[{"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/media?parent=76404"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/categories?post=76404"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/tags?post=76404"},{"taxonomy":"industries-application","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/industries-application?post=76404"},{"taxonomy":"post-badge","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/post-badge?post=76404"},{"taxonomy":"resource-type","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/resource-type?post=76404"},{"taxonomy":"product-library","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/product-library?post=76404"},{"taxonomy":"resource-technique","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/resource-technique?post=76404"},{"taxonomy":"author","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/ppma_author?post=76404"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}