{"id":76416,"date":"2022-12-11T00:00:00","date_gmt":"2022-12-11T00:00:00","guid":{"rendered":"https:\/\/discover.restek.com\/?p=76416"},"modified":"2026-01-28T22:41:30","modified_gmt":"2026-01-28T22:41:30","slug":"dont-overlook-lc-column-temperature","status":"publish","type":"post","link":"https:\/\/discover.restek.com\/it\/blogs\/gnbl5488\/dont-overlook-lc-column-temperature","title":{"rendered":"Don\u2019t Overlook LC Column Temperature!"},"content":{"rendered":"\n<p class=\"wp-block-paragraph\">Many things can affect peak shape and retention times, such as column size, mobile phase additives, and analyte characteristics. Column temperature is an important and often overlooked factor when it comes to LC method development. This is because column temperature can impact peak shape, retention, and backpressure. Regarding retention time, as the temperature of the column is increased, the retention time is decreased. This is because, the higher the temperature of the column, the faster the analytes are exchanged between the stationary phase of the column and the mobile phase. As this process becomes faster, analytes are not retained as long, resulting in shorter retention times, and often more efficient peak shapes. However, keep in mind that increasing temperature does not affect all analytes equally, meaning that the impact of retention factor may be different for each analyte. For example, in Figure 1 and Figure 2 below the same method and same column were used but the methods were run using two different column temperatures, 30<sup>o<\/sup>C and 55<sup>o<\/sup>C. In this example, shorter retention times are noted for all of the analytes up to a difference of 0.19 minutes.<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\">Figure 1: Fatty acid ethyl esters on Force Biphenyl 50 x 3mm, 3 \u00b5m at 30<sup>o<\/sup>C.<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n\n<figure class=\"wp-block-image size-large\"><img loading=\"lazy\" decoding=\"async\" width=\"1024\" height=\"278\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-dont-overlook-lc-column-temperature-01-1024x278.png\" alt=\"\" class=\"wp-image-77686\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-dont-overlook-lc-column-temperature-01-1024x278.png 1024w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-dont-overlook-lc-column-temperature-01-300x81.png 300w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-dont-overlook-lc-column-temperature-01-768x209.png 768w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-dont-overlook-lc-column-temperature-01-1536x417.png 1536w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-dont-overlook-lc-column-temperature-01-2048x556.png 2048w\" sizes=\"auto, (max-width: 1024px) 100vw, 1024px\" \/><\/figure>\n\n<\/div><\/div><\/div>\n\n\n<div style=\"height:25px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\">Figure 2: Fatty acid ethyl esters on Force Biphenyl 50 x 3mm, 3 \u00b5m at 55<sup>o<\/sup>C.<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n\n<figure class=\"wp-block-image size-large\"><img loading=\"lazy\" decoding=\"async\" width=\"1024\" height=\"275\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-dont-overlook-lc-column-temperature-02-1024x275.png\" alt=\"\" class=\"wp-image-77680\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-dont-overlook-lc-column-temperature-02-1024x275.png 1024w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-dont-overlook-lc-column-temperature-02-300x81.png 300w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-dont-overlook-lc-column-temperature-02-768x207.png 768w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-dont-overlook-lc-column-temperature-02-1536x413.png 1536w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-dont-overlook-lc-column-temperature-02-2048x551.png 2048w\" sizes=\"auto, (max-width: 1024px) 100vw, 1024px\" \/><\/figure>\n\n<\/div><\/div><\/div>\n\n\n<div style=\"height:25px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<p class=\"wp-block-paragraph\">However, if you are having trouble separating peaks, or want to get better resolution between peaks, it could help to try a lower column temperature. Just as increasing the column temperature can cause the retention times to be shorter because the speed of interactions on the column has increased, lowering the temperature will do the opposite, which in turn can help with the resolution of analytes. There are also some analytes that show better separation with a lower column temperature. For example, some cannabinoids such as CBDA, CBGA, CBC, and THCA show best resolution at a lower temperature. In the example below, the first chromatogram the compounds were analyzed with a column temperature at 40<sup>o<\/sup>C and there is no baseline resolution between CBDA and CBGA or CBC and THCA however, when run with a column temperature of 30<sup>o<\/sup>C, the compounds are better resolved. This can be seen in Figure 3 and Figure 4 below.<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\">Figure 3: Chromatogram of cannabinoids at 40<sup>o<\/sup>C column temperature.<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n\n<figure class=\"wp-block-image size-large\"><img loading=\"lazy\" decoding=\"async\" width=\"1024\" height=\"544\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-dont-overlook-lc-column-temperature-03-1024x544.png\" alt=\"\" class=\"wp-image-77674\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-dont-overlook-lc-column-temperature-03-1024x544.png 1024w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-dont-overlook-lc-column-temperature-03-300x159.png 300w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-dont-overlook-lc-column-temperature-03-768x408.png 768w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-dont-overlook-lc-column-temperature-03.png 1129w\" sizes=\"auto, (max-width: 1024px) 100vw, 1024px\" \/><\/figure>\n\n<\/div><\/div><\/div>\n\n\n<div style=\"height:25px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\">Figure 4: Chromatogram of cannabinoids at 30<sup>o<\/sup>C column temperature.<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n\n<figure class=\"wp-block-image size-large\"><img loading=\"lazy\" decoding=\"async\" width=\"1024\" height=\"598\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-dont-overlook-lc-column-temperature-04-1024x598.png\" alt=\"\" class=\"wp-image-77668\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-dont-overlook-lc-column-temperature-04-1024x598.png 1024w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-dont-overlook-lc-column-temperature-04-300x175.png 300w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-dont-overlook-lc-column-temperature-04-768x448.png 768w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-dont-overlook-lc-column-temperature-04.png 1095w\" sizes=\"auto, (max-width: 1024px) 100vw, 1024px\" \/><\/figure>\n\n<\/div><\/div><\/div>\n\n\n<div style=\"height:25px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<p class=\"wp-block-paragraph\">While the above is true in most cases, there are certain instances where a higher temperature works better for both shorter retention times and analyte resolution. Ergot alkaloid epimers and PFAS compounds show best results with a higher column temperature. In the example below (Figure 5), ergosine and ergosinine were analyzed using a Raptor Biphenyl 100 x 2.1mm, 2.7\u00b5m column at 50<sup>o<\/sup>C and 60<sup>o<\/sup>C. At 50<sup>o<\/sup>C there is no separation of the two analytes, but at a higher column temperature, 60<sup>o<\/sup>C, the two analytes begin to separate.<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\">Figure 5: Effects of column temperature on ergot alkaloid epimers.<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n\n<figure class=\"wp-block-image size-full\"><img loading=\"lazy\" decoding=\"async\" width=\"960\" height=\"418\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-dont-overlook-lc-column-temperature-05.png\" alt=\"\" class=\"wp-image-77662\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-dont-overlook-lc-column-temperature-05.png 960w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-dont-overlook-lc-column-temperature-05-300x131.png 300w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-dont-overlook-lc-column-temperature-05-768x334.png 768w\" sizes=\"auto, (max-width: 960px) 100vw, 960px\" \/><\/figure>\n\n<\/div><\/div><\/div>\n\n\n<div style=\"height:25px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<p class=\"wp-block-paragraph\">Column temperature can also affect the back pressure of the column. Because high mobile phase viscosity can cause high back pressures, raising the temperature of the column can help to decrease the mobile phase\u2019s viscosity and therefore, lower the back pressure. For example, in Figure 1 and Figure 2 above, the starting conditions, mobile phases, and gradient conditions are the same for both. However, the column temperature was different for both. When the column temperature was set to 30<sup>o<\/sup>C, the pressure was 2100 psi. When the column temperature was set to 55<sup>o<\/sup>C, the pressure was 1500 psi, yielding a different in psi of approximately 600 psi.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">While higher column temperatures can help with peak shape, and shorter retention times, it is important to not go above the recommended temperature for the column. Using too high of a temperature on a column can lead to a drastically shortened column lifetime. You can usually find the maximum temperature for your column in the manufacturer\u2019s column usage guide.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">So, remember, though it is often overlooked, column temperature is an important parameter to optimize whenever you are setting up a method. It can take a bit of trial and error to find the right column temperature for the analytes on your method, but considering this parameter in early method development can lead to better resolution and more efficient methods.<\/p>\n\n\n\n<div style=\"height:100px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n","protected":false},"excerpt":{"rendered":"<p>Many things can affect peak shape and retention times, such as column size, mobile phase additives, and analyte characteristics. Column temperature is an important and often overlooked factor when it comes to LC method development.&nbsp;<\/p>\n","protected":false},"author":24,"featured_media":0,"comment_status":"closed","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"_acf_changed":false,"_kad_blocks_custom_css":"","_kad_blocks_head_custom_js":"","_kad_blocks_body_custom_js":"","_kad_blocks_footer_custom_js":"","_kadence_starter_templates_imported_post":false,"_kad_post_transparent":"","_kad_post_title":"","_kad_post_layout":"","_kad_post_sidebar_id":"","_kad_post_content_style":"","_kad_post_vertical_padding":"","_kad_post_feature":"","_kad_post_feature_position":"","_kad_post_header":false,"_kad_post_footer":false,"footnotes":"","_ppma_block_editor_authors":""},"categories":[9],"tags":[],"industries-application":[],"post-badge":[],"resource-type":[],"product-library":[],"resource-technique":[],"ppma_author":[444],"class_list":["post-76416","post","type-post","status-publish","format-standard","hentry","category-blogs"],"acf":[],"taxonomy_info":{"category":[{"value":9,"label":"Blogs"}]},"featured_image_src_large":false,"author_info":{"display_name":"Samantha Herbick","author_link":"https:\/\/discover.restek.com\/it\/author\/samantha-herbick\/"},"comment_info":0,"category_info":[{"term_id":9,"name":"Blogs","slug":"blogs","term_group":0,"term_taxonomy_id":9,"taxonomy":"category","description":"","parent":0,"count":457,"filter":"raw","cat_ID":9,"category_count":457,"category_description":"","cat_name":"Blogs","category_nicename":"blogs","category_parent":0}],"tag_info":false,"authors":[{"term_id":444,"user_id":24,"is_guest":0,"slug":"samantha-herbick","display_name":"Samantha Herbick","avatar_url":{"url":"https:\/\/discover.restek.com\/wp-content\/uploads\/people-herbick-samantha.jpg","url2x":"https:\/\/discover.restek.com\/wp-content\/uploads\/people-herbick-samantha.jpg"},"author_category":"1","first_name":"Samantha","last_name":"Herbick","user_url":"https:\/\/www.linkedin.com\/in\/samantha-herbick-656378136\/","job_title":"Advanced Scientist, LC Solutions","description":"Samantha Herbick is an applications scientist within the LC Solutions Department at Restek. Her primary focus is on the development of applications in the toxicology and life science markets. She attended Duquesne University where she earned a bachelor\u2019s degree in Biochemistry and a master's degree in forensic science and law. Prior to joining Restek, Samantha worked as a scientist and method developer in a forensic toxicology lab. In this role, she performed analysis on toxicology casework and worked on the development and validation of new and existing assays using LC-MS\/MS, GC-MS, and GC-FID."}],"_links":{"self":[{"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/posts\/76416","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/users\/24"}],"replies":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/comments?post=76416"}],"version-history":[{"count":3,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/posts\/76416\/revisions"}],"predecessor-version":[{"id":85815,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/posts\/76416\/revisions\/85815"}],"wp:attachment":[{"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/media?parent=76416"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/categories?post=76416"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/tags?post=76416"},{"taxonomy":"industries-application","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/industries-application?post=76416"},{"taxonomy":"post-badge","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/post-badge?post=76416"},{"taxonomy":"resource-type","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/resource-type?post=76416"},{"taxonomy":"product-library","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/product-library?post=76416"},{"taxonomy":"resource-technique","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/resource-technique?post=76416"},{"taxonomy":"author","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/ppma_author?post=76416"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}