{"id":76425,"date":"2015-10-15T00:00:00","date_gmt":"2015-10-15T00:00:00","guid":{"rendered":"https:\/\/discover.restek.com\/?p=76425"},"modified":"2025-12-31T13:32:07","modified_gmt":"2025-12-31T13:32:07","slug":"high-source-temperatures-and-meeting-epa-tuning-criteria","status":"publish","type":"post","link":"https:\/\/discover.restek.com\/it\/blogs\/gnbl5500\/high-source-temperatures-and-meeting-epa-tuning-criteria","title":{"rendered":"High Source Temperatures and Meeting EPA Tuning Criteria"},"content":{"rendered":"\n<p class=\"wp-block-paragraph\">My first blog here at Restek was a demonstration of how increasing the source temperature could improve the raw response of 2,4-dinitrophenol while maintaining the relative response. Four years later I&#8217;m here with the promised follow-up on the elevated source temperature effects on tune evaluations (and making adjustments via target tuning). This blog is going to deal exclusively with meeting the EPA Method 8270D DFTPP tune evaluation criteria because fear of failing tunes is the primary feedback we receive when we ask customers why they don&#8217;t try acquiring data at elevated source temperatures when performing semi-volatiles analysis.<\/p>\n\n\n<style>.kb-image76425_3d9186-e1:not(.kb-image-is-ratio-size) .kb-img, .kb-image76425_3d9186-e1.kb-image-is-ratio-size{padding-bottom:0px;}.kb-image76425_3d9186-e1 .kb-image-has-overlay:after{opacity:0.3;}<\/style>\n<figure class=\"wp-block-kadence-image kb-image76425_3d9186-e1 size-full\"><img loading=\"lazy\" decoding=\"async\" width=\"479\" height=\"569\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-high-source-temperatures-and-meeting-epa-tuning-criteria-01.png\" alt=\"\" class=\"kb-img wp-image-77964\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-high-source-temperatures-and-meeting-epa-tuning-criteria-01.png 479w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-high-source-temperatures-and-meeting-epa-tuning-criteria-01-253x300.png 253w\" sizes=\"auto, (max-width: 479px) 100vw, 479px\" \/><figcaption>Figure 1 &#8211; Table 3: DFTPP Ion Abundance Criteria, taken from EPA Method 8270D Rev. 4, Feb. 2007<br><\/figcaption><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\"><strong>Figure 1<\/strong>&nbsp;lists the DFTPP ion abundance criteria for quadrupole instrumentation. It is important to note that it allows for m\/z = 198 or m\/z = 442 to be the base peak. The first step is to make sure that your software is configured to automatically switch between alternate base peaks (<strong>Figure 2<\/strong>&nbsp;&#8211; Edit DFTPP tune evaluation criteria interface).<\/p>\n\n\n<style>.kb-image76425_888f9f-e0:not(.kb-image-is-ratio-size) .kb-img, .kb-image76425_888f9f-e0.kb-image-is-ratio-size{padding-top:var(--global-kb-spacing-xs, 1rem);padding-bottom:0px;}.kb-image76425_888f9f-e0 .kb-image-has-overlay:after{opacity:0.3;}<\/style>\n<figure class=\"wp-block-kadence-image kb-image76425_888f9f-e0 size-full\"><img loading=\"lazy\" decoding=\"async\" width=\"606\" height=\"355\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-high-source-temperatures-and-meeting-epa-tuning-criteria-02.png\" alt=\"\" class=\"kb-img wp-image-77958\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-high-source-temperatures-and-meeting-epa-tuning-criteria-02.png 606w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-high-source-temperatures-and-meeting-epa-tuning-criteria-02-300x176.png 300w\" sizes=\"auto, (max-width: 606px) 100vw, 606px\" \/><figcaption>Figure 2 &#8211; EnviroQuant Tune Evaluation Criteria Edit Screen<\/figcaption><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\"><strong>N.B.<\/strong>&nbsp;The 8270D tune evaluation criteria are NOT&nbsp;the default DFTPP tune evaluation criteria in EnviroQuant. If you haven&#8217;t updated them, be sure that you do.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">The goal here is to kill two birds with one stone. If you&#8217;ve read last week&#8217;s blog,&nbsp;<em>Which Source Configuration is Right for your Application<\/em>,&nbsp;you know that my conclusion was that a 6 mm ID or 9 mm ID extractor lens or drawout plate should be used for semivolatile analysis to minimize source tailing of low volatility compounds (while increasing signal to noise). Anticipating tune related hesitation, I performed all the tuning experiments detailed below with a 9 mm ID extractor lens installed and source and quadrupole temperatures set at 350\u2070C and 200\u2070C respectively.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Because I&#8217;m using an extractor source, I ran the&nbsp;etune.u with the default targets from the dftpp.u tune file. The tune wizard screens are&nbsp;<strong>F<\/strong><strong>igures 3, 4, 5, and 6<\/strong>&nbsp;below.&nbsp;<strong>Figure 7<\/strong>&nbsp;is the detailed tune profile report.<\/p>\n\n\n<style>.kb-image76425_2d9992-c0:not(.kb-image-is-ratio-size) .kb-img, .kb-image76425_2d9992-c0.kb-image-is-ratio-size{padding-top:var(--global-kb-spacing-xs, 1rem);padding-bottom:0px;}.kb-image76425_2d9992-c0 .kb-image-has-overlay:after{opacity:0.3;}<\/style>\n<figure class=\"wp-block-kadence-image kb-image76425_2d9992-c0 size-full\"><img loading=\"lazy\" decoding=\"async\" width=\"532\" height=\"280\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-high-source-temperatures-and-meeting-epa-tuning-criteria-03.png\" alt=\"\" class=\"kb-img wp-image-77952\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-high-source-temperatures-and-meeting-epa-tuning-criteria-03.png 532w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-high-source-temperatures-and-meeting-epa-tuning-criteria-03-300x158.png 300w\" sizes=\"auto, (max-width: 532px) 100vw, 532px\" \/><figcaption>Figure 3 &#8211; Tune Wizard Screen 1<\/figcaption><\/figure>\n\n\n<style>.kb-image76425_c52cf3-04:not(.kb-image-is-ratio-size) .kb-img, .kb-image76425_c52cf3-04.kb-image-is-ratio-size{padding-top:var(--global-kb-spacing-xs, 1rem);padding-bottom:0px;}.kb-image76425_c52cf3-04 .kb-image-has-overlay:after{opacity:0.3;}<\/style>\n<figure class=\"wp-block-kadence-image kb-image76425_c52cf3-04 size-full\"><img loading=\"lazy\" decoding=\"async\" width=\"522\" height=\"402\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-high-source-temperatures-and-meeting-epa-tuning-criteria-04.png\" alt=\"\" class=\"kb-img wp-image-77946\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-high-source-temperatures-and-meeting-epa-tuning-criteria-04.png 522w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-high-source-temperatures-and-meeting-epa-tuning-criteria-04-300x231.png 300w\" sizes=\"auto, (max-width: 522px) 100vw, 522px\" \/><figcaption>Figure 4 &#8211; Tune Wizard Screen 2<\/figcaption><\/figure>\n\n\n<style>.kb-image76425_b59251-cd:not(.kb-image-is-ratio-size) .kb-img, .kb-image76425_b59251-cd.kb-image-is-ratio-size{padding-top:var(--global-kb-spacing-xs, 1rem);padding-bottom:0px;}.kb-image76425_b59251-cd .kb-image-has-overlay:after{opacity:0.3;}<\/style>\n<figure class=\"wp-block-kadence-image kb-image76425_b59251-cd size-full\"><img loading=\"lazy\" decoding=\"async\" width=\"523\" height=\"362\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-high-source-temperatures-and-meeting-epa-tuning-criteria-05.png\" alt=\"\" class=\"kb-img wp-image-77940\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-high-source-temperatures-and-meeting-epa-tuning-criteria-05.png 523w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-high-source-temperatures-and-meeting-epa-tuning-criteria-05-300x208.png 300w\" sizes=\"auto, (max-width: 523px) 100vw, 523px\" \/><figcaption>Figure 5 &#8211; Tune Wizard Screen 3<\/figcaption><\/figure>\n\n\n<style>.kb-image76425_0464b0-38:not(.kb-image-is-ratio-size) .kb-img, .kb-image76425_0464b0-38.kb-image-is-ratio-size{padding-top:var(--global-kb-spacing-xs, 1rem);padding-bottom:0px;}.kb-image76425_0464b0-38 .kb-image-has-overlay:after{opacity:0.3;}<\/style>\n<figure class=\"wp-block-kadence-image kb-image76425_0464b0-38 size-full\"><img loading=\"lazy\" decoding=\"async\" width=\"528\" height=\"234\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-high-source-temperatures-and-meeting-epa-tuning-criteria-06.png\" alt=\"\" class=\"kb-img wp-image-77934\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-high-source-temperatures-and-meeting-epa-tuning-criteria-06.png 528w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-high-source-temperatures-and-meeting-epa-tuning-criteria-06-300x133.png 300w\" sizes=\"auto, (max-width: 528px) 100vw, 528px\" \/><figcaption>Figure 6 &#8211; Tune Wizard Screen 4<\/figcaption><\/figure>\n\n\n<style>.kb-image76425_491cbb-5d:not(.kb-image-is-ratio-size) .kb-img, .kb-image76425_491cbb-5d.kb-image-is-ratio-size{padding-top:var(--global-kb-spacing-xs, 1rem);padding-bottom:0px;}.kb-image76425_491cbb-5d .kb-image-has-overlay:after{opacity:0.3;}<\/style>\n<figure class=\"wp-block-kadence-image kb-image76425_491cbb-5d size-full\"><img loading=\"lazy\" decoding=\"async\" width=\"702\" height=\"897\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-high-source-temperatures-and-meeting-epa-tuning-criteria-07.png\" alt=\"\" class=\"kb-img wp-image-77928\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-high-source-temperatures-and-meeting-epa-tuning-criteria-07.png 702w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-high-source-temperatures-and-meeting-epa-tuning-criteria-07-235x300.png 235w\" sizes=\"auto, (max-width: 702px) 100vw, 702px\" \/><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\"><strong>Figure 8<\/strong>&nbsp;is the tune evaluation report&nbsp;for a 10:1 split injection of the 50 ng\/\u00b5L GC-MS tuning mix (cat# 31615) on a 30 m Rxi-5ms (cat# 13423), with an acquisition (scan) rate of 5.4 Hz and a gain factor of 3.0, using the etune.u file described above. The DFTPP tune evaluation met all criteria the first time it was evaluated for the peak apex scan (<strong>Figure 8<\/strong>), as well as the peak average (report not shown). Overall, the&nbsp;evaluation looks good, though the m\/z = 51 relative to the base peak check (m\/z = 198 in this case). 14.6% is a bit closer to the lower threshold of 10% relative response than I would like if I were setting this instrument up for routine work.<\/p>\n\n\n<style>.kb-image76425_d11ce6-4b:not(.kb-image-is-ratio-size) .kb-img, .kb-image76425_d11ce6-4b.kb-image-is-ratio-size{padding-top:var(--global-kb-spacing-xs, 1rem);padding-bottom:0px;}.kb-image76425_d11ce6-4b .kb-image-has-overlay:after{opacity:0.3;}<\/style>\n<figure class=\"wp-block-kadence-image kb-image76425_d11ce6-4b size-large\"><img loading=\"lazy\" decoding=\"async\" width=\"1024\" height=\"1003\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-high-source-temperatures-and-meeting-epa-tuning-criteria-08-1024x1003.png\" alt=\"\" class=\"kb-img wp-image-77916\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-high-source-temperatures-and-meeting-epa-tuning-criteria-08-1024x1003.png 1024w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-high-source-temperatures-and-meeting-epa-tuning-criteria-08-300x294.png 300w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-high-source-temperatures-and-meeting-epa-tuning-criteria-08-768x752.png 768w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-high-source-temperatures-and-meeting-epa-tuning-criteria-08.png 1037w\" sizes=\"auto, (max-width: 1024px) 100vw, 1024px\" \/><figcaption>Figure 8 &#8211; EnviroQuant DFTPP Tune Evaluation Report<\/figcaption><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\">In fact, when run under alternate acquisition conditions, with the gain set to 0.3, the tune evaluation fails low for m\/z = 51 relative to base peak response when evaluated on the peak apex scan (<strong>Figure 9<\/strong>) .<\/p>\n\n\n<style>.kb-image76425_ed2b1a-11:not(.kb-image-is-ratio-size) .kb-img, .kb-image76425_ed2b1a-11.kb-image-is-ratio-size{padding-top:var(--global-kb-spacing-xs, 1rem);padding-bottom:0px;}.kb-image76425_ed2b1a-11 .kb-image-has-overlay:after{opacity:0.3;}<\/style>\n<figure class=\"wp-block-kadence-image kb-image76425_ed2b1a-11 size-large\"><img loading=\"lazy\" decoding=\"async\" width=\"1024\" height=\"1003\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-high-source-temperatures-and-meeting-epa-tuning-criteria-09-1024x1003.png\" alt=\"\" class=\"kb-img wp-image-77910\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-high-source-temperatures-and-meeting-epa-tuning-criteria-09-1024x1003.png 1024w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-high-source-temperatures-and-meeting-epa-tuning-criteria-09-300x294.png 300w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-high-source-temperatures-and-meeting-epa-tuning-criteria-09-768x752.png 768w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-high-source-temperatures-and-meeting-epa-tuning-criteria-09.png 1037w\" sizes=\"auto, (max-width: 1024px) 100vw, 1024px\" \/><figcaption>Figure 9 &#8211; EnviroQuant DFTPP Tune Evaluation Report<\/figcaption><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\">The mass selective detector can be tuned to elevate the relative response of low m\/z fragments while leaving the higher m\/z fragments largely unaffected. This is done by increasing the target response of PFTBA fragment m\/z = 50 relative to m\/z = 69 (the PFTBA base ion peak) while leaving the other target peaks the same. This is done on screen two of the tune wizard (<strong>Figure 10<\/strong>). For this exercise, I raised the m\/z = 50 target from 1.0% to 1.1% of m\/z = 69.<\/p>\n\n\n<style>.kb-image76425_1b65cc-73:not(.kb-image-is-ratio-size) .kb-img, .kb-image76425_1b65cc-73.kb-image-is-ratio-size{padding-top:var(--global-kb-spacing-xs, 1rem);padding-bottom:0px;}.kb-image76425_1b65cc-73 .kb-image-has-overlay:after{opacity:0.3;}<\/style>\n<figure class=\"wp-block-kadence-image kb-image76425_1b65cc-73 size-full\"><img loading=\"lazy\" decoding=\"async\" width=\"524\" height=\"401\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-high-source-temperatures-and-meeting-epa-tuning-criteria-10.png\" alt=\"\" class=\"kb-img wp-image-77904\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-high-source-temperatures-and-meeting-epa-tuning-criteria-10.png 524w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-high-source-temperatures-and-meeting-epa-tuning-criteria-10-300x230.png 300w\" sizes=\"auto, (max-width: 524px) 100vw, 524px\" \/><figcaption>Figure 10 &#8211; Tune Wizard Screen 2 Showing Change in M\/Z = 50 Tune Target<\/figcaption><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\">This 0.1% relative abundance target change for m\/z = 50 resulted in an approximately 30% increase of the m\/z = 51 response relative to the base peak in the DFTPP tune evaluation (<strong>Figure 11<\/strong>). This time the tune was evaluated for a 20:1 split injection of the 50 ng\/\u00b5L GC-MS tuning mix (cat# 31615) on a 20 m Rxi-5ms (cat# 13402), with an acquisition (scan) rate of&nbsp;9.7&nbsp;Hz and a gain factor of 3.0, using the etune 350 target I.u file generated from the change above.&nbsp;<strong>Figure 12<\/strong>&nbsp;is the detailed tune profile report.<\/p>\n\n\n<style>.kb-image76425_136101-7c:not(.kb-image-is-ratio-size) .kb-img, .kb-image76425_136101-7c.kb-image-is-ratio-size{padding-top:var(--global-kb-spacing-xs, 1rem);padding-bottom:0px;}.kb-image76425_136101-7c .kb-image-has-overlay:after{opacity:0.3;}<\/style>\n<figure class=\"wp-block-kadence-image kb-image76425_136101-7c size-large\"><img loading=\"lazy\" decoding=\"async\" width=\"1024\" height=\"1003\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-high-source-temperatures-and-meeting-epa-tuning-criteria-11-1024x1003.png\" alt=\"\" class=\"kb-img wp-image-77898\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-high-source-temperatures-and-meeting-epa-tuning-criteria-11-1024x1003.png 1024w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-high-source-temperatures-and-meeting-epa-tuning-criteria-11-300x294.png 300w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-high-source-temperatures-and-meeting-epa-tuning-criteria-11-768x752.png 768w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-high-source-temperatures-and-meeting-epa-tuning-criteria-11.png 1037w\" sizes=\"auto, (max-width: 1024px) 100vw, 1024px\" \/><figcaption>Figure 11 &#8211; EnviroQuant DFTPP Tune Evaluation<\/figcaption><\/figure>\n\n\n<style>.kb-image76425_a99808-ef:not(.kb-image-is-ratio-size) .kb-img, .kb-image76425_a99808-ef.kb-image-is-ratio-size{padding-top:var(--global-kb-spacing-xs, 1rem);padding-bottom:0px;}.kb-image76425_a99808-ef .kb-image-has-overlay:after{opacity:0.3;}<\/style>\n<figure class=\"wp-block-kadence-image kb-image76425_a99808-ef size-full\"><img loading=\"lazy\" decoding=\"async\" width=\"702\" height=\"897\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-high-source-temperatures-and-meeting-epa-tuning-criteria-12.png\" alt=\"\" class=\"kb-img wp-image-77892\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-high-source-temperatures-and-meeting-epa-tuning-criteria-12.png 702w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-high-source-temperatures-and-meeting-epa-tuning-criteria-12-235x300.png 235w\" sizes=\"auto, (max-width: 702px) 100vw, 702px\" \/><figcaption>Figure 12 &#8211; etune 350 target I.u Detailed Profile Report<\/figcaption><\/figure>\n\n\n\n<div style=\"height:100px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n","protected":false},"excerpt":{"rendered":"<p>My first blog here at Restek was a demonstration of how increasing the source temperature could improve the raw response of 2,4-dinitrophenol while maintaining the relative response. Four years later I&#8217;m here with the promised follow-up on the elevated source&#8230;<\/p>\n","protected":false},"author":3,"featured_media":0,"comment_status":"closed","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"_acf_changed":false,"_kad_blocks_custom_css":"","_kad_blocks_head_custom_js":"","_kad_blocks_body_custom_js":"","_kad_blocks_footer_custom_js":"","_kadence_starter_templates_imported_post":false,"_kad_post_transparent":"","_kad_post_title":"","_kad_post_layout":"","_kad_post_sidebar_id":"","_kad_post_content_style":"","_kad_post_vertical_padding":"","_kad_post_feature":"","_kad_post_feature_position":"","_kad_post_header":false,"_kad_post_footer":false,"footnotes":"","_ppma_block_editor_authors":""},"categories":[9],"tags":[],"industries-application":[],"post-badge":[],"resource-type":[],"product-library":[],"resource-technique":[],"ppma_author":[584],"class_list":["post-76425","post","type-post","status-publish","format-standard","hentry","category-blogs"],"acf":[],"taxonomy_info":{"category":[{"value":9,"label":"Blogs"}]},"featured_image_src_large":false,"author_info":{"display_name":"Audrey Crater","author_link":"https:\/\/discover.restek.com\/it\/author\/audrey-craterrestek-com\/"},"comment_info":0,"category_info":[{"term_id":9,"name":"Blogs","slug":"blogs","term_group":0,"term_taxonomy_id":9,"taxonomy":"category","description":"","parent":0,"count":457,"filter":"raw","cat_ID":9,"category_count":457,"category_description":"","cat_name":"Blogs","category_nicename":"blogs","category_parent":0}],"tag_info":false,"authors":[{"term_id":584,"user_id":0,"is_guest":1,"slug":"chris-rattray","display_name":"Chris Rattray","avatar_url":"https:\/\/secure.gravatar.com\/avatar\/?s=96&d=mm&r=g","author_category":"1","first_name":"Chris","last_name":"Rattray","user_url":"","job_title":"","description":""}],"_links":{"self":[{"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/posts\/76425","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/users\/3"}],"replies":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/comments?post=76425"}],"version-history":[{"count":4,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/posts\/76425\/revisions"}],"predecessor-version":[{"id":78962,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/posts\/76425\/revisions\/78962"}],"wp:attachment":[{"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/media?parent=76425"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/categories?post=76425"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/tags?post=76425"},{"taxonomy":"industries-application","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/industries-application?post=76425"},{"taxonomy":"post-badge","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/post-badge?post=76425"},{"taxonomy":"resource-type","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/resource-type?post=76425"},{"taxonomy":"product-library","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/product-library?post=76425"},{"taxonomy":"resource-technique","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/resource-technique?post=76425"},{"taxonomy":"author","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/ppma_author?post=76425"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}