{"id":97348,"date":"2026-06-23T01:12:32","date_gmt":"2026-06-23T01:12:32","guid":{"rendered":"https:\/\/discover.restek.com\/?p=97348"},"modified":"2026-10-16T20:12:02","modified_gmt":"2026-10-16T20:12:02","slug":"gc-column-installation-guide","status":"publish","type":"post","link":"https:\/\/discover.restek.com\/it\/blogs\/gnbl5738\/gc-column-installation-guide","title":{"rendered":"GC Column Installation: Step-by-Step Guide for Capillary Columns"},"content":{"rendered":"\n<p class=\"wp-block-paragraph\">Any analyst facing GC column installation understands the underlying tension. The gas chromatography setup looks straightforward on paper, but getting it right is critical.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Rushing the step-by-step process can lead to immediate analytical problems such as ghost peaks, split peaks, fitting leaks, column bleed, and wasted runs. However, with the right preparation, GC column installation becomes a repeatable, stress-free skill rather than a guessing game.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">This guide covers every detail from the pre-installation checklist through final conditioning. You will learn to secure every connection point with a clear, repeatable technique, reducing the risk of leaks, poor peak shape, and unstable baselines.<\/p>\n\n\n\n<h2 class=\"wp-block-heading\"><strong>What You Need Before You Start<\/strong><\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">Successful lab GC column installation always begins with <strong>proper preparation<\/strong>. The most common GC best practices dictate that gathering the right tools prevents errors before they happen.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Catching gaps before starting saves time, extends column life, and conserves carrier gas. Ensure you have the following ready before you begin:<\/p>\n\n\n\n<h3 class=\"wp-block-heading\">The GC column<\/h3>\n\n\n\n<p class=\"wp-block-paragraph\">Check that the box documentation is intact. Review the lot certificate and quality test chromatogram to verify lot-to-lot consistency and confirm the <a href=\"https:\/\/www.restek.com\/c\/1109\" target=\"_blank\" rel=\"noreferrer noopener\">GC column<\/a> meets expected performance specifications.&nbsp;<\/p>\n\n\n\n<h3 class=\"wp-block-heading\">Correct ferrules<\/h3>\n\n\n\n<p class=\"wp-block-paragraph\">Match the ferrule inner diameter to the <strong>column outer diameter <\/strong>and choose ferrule formats and materials that are compatible with your instrument and application. Restek <a href=\"https:\/\/www.restek.com\/global\/en\/c\/1559\" target=\"_blank\" rel=\"noreferrer noopener\">Vespel\/graphite ferrules<\/a> provide reliable sealing without over-compression compared to pure graphite alternatives. Make sure your ferrule choice is appropriate for your instrument, temperature, and application.&nbsp;<\/p>\n\n\n\n<figure class=\"wp-block-embed is-type-video is-provider-youtube wp-block-embed-youtube wp-embed-aspect-16-9 wp-has-aspect-ratio\"><div class=\"wp-block-embed__wrapper\">\n<iframe loading=\"lazy\" title=\"What Ferrule Material and ID Can I Use in my GC?\" width=\"720\" height=\"405\" src=\"https:\/\/www.youtube.com\/embed\/Q7yoaM4CiFA?feature=oembed\" frameborder=\"0\" allow=\"accelerometer; autoplay; clipboard-write; encrypted-media; gyroscope; picture-in-picture; web-share\" referrerpolicy=\"strict-origin-when-cross-origin\" allowfullscreen><\/iframe>\n<\/div><\/figure>\n\n\n\n<div style=\"height:0px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<p class=\"wp-block-paragraph\">Using matched ferrule materials at both ends ensures consistent thermal expansion behavior across your full temperature range. Mismatched ferrule materials can expand differently during temperature cycling and may increase the risk of leaks.<\/p>\n\n\n\n<h3 class=\"wp-block-heading\">Column nuts<\/h3>\n\n\n\n<p class=\"wp-block-paragraph\">Verify compatibility with your specific inlet and detector fittings to properly connect components.<\/p>\n\n\n\n<h3 class=\"wp-block-heading\">Leak detection tools<\/h3>\n\n\n\n<p class=\"wp-block-paragraph\">Use an <a href=\"https:\/\/www.restek.com\/p\/28500\" target=\"_blank\" rel=\"noreferrer noopener\"><strong>electronic leak detector<\/strong><\/a> whenever possible. If permitted by your lab&#8217;s standard operating procedures, use chromatography-grade leak detection solution only as a backup. Restek does not recommend ever using liquid leak detector solutions because they can be drawn into the capillary column and cause irreparable damage.<\/p>\n\n\n\n<h3 class=\"wp-block-heading\">Clean, powder-free gloves<\/h3>\n\n\n\n<p class=\"wp-block-paragraph\">Skin oils and debris can contaminate the column end and contribute to poor peak shape, ghost peaks, or unstable early runs. Always wear clean, powder-free gloves for GC column installation.<\/p>\n\n\n\n<h3 class=\"wp-block-heading\">Column cutter<\/h3>\n\n\n\n<p class=\"wp-block-paragraph\">Use a ceramic scoring wafer or sapphire scribe to avoid scratching and damaging the tubing, plus a magnifying loupe to inspect the cut.<\/p>\n\n\n\n<h3 class=\"wp-block-heading\">Installation guides<\/h3>\n\n\n\n<p class=\"wp-block-paragraph\">Have your instrument manual and Restek&#8217;s <a href=\"https:\/\/discover.restek.com\/en\/instruction-sheet\/gnot1098\/restek-capillary-column-installation-guide-section-i-brief-installation-procedure\">GC column installation quick-start guide<\/a> open on the bench. <\/p>\n\n\n\n<p class=\"wp-block-paragraph\">A clean <a href=\"https:\/\/www.restek.com\/global\/en\/c\/1484\" target=\"_blank\" rel=\"noreferrer noopener\">Topaz inlet liner<\/a> is also crucial to install while you are installing the column. A contaminated liner can compromise an otherwise correct installation. A new inlet seal and septum should also be installed along with the GC column. Restek carries a full line of GC consumables to complement column installation and ensure a clean flow path during sample analysis.<\/p>\n\n\n\n<h2 class=\"wp-block-heading\"><strong>Step 1: Prepare the Instrument<\/strong><\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">The instrument must be stabilized before introducing the column. Installing into an uncontrolled environment increases the likelihood of leaks, contamination, and component damage.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Begin by cooling the GC oven, inlet, and detector to <strong>below 40 \u00b0C<\/strong>. While some instruments specify different limits, this threshold is a reliable general standard for safe handling.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Reduce or shut off carrier gas flow according to the instrument procedure. Attempting installation under pressure can disrupt ferrule seating and introduce contaminants at the column end.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Set detectors to off or standby. Connecting a room-temperature column to a heated detector or source complicates alignment and increases the risk of improper sealing.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Inspect the inlet liner carefully. It should be clean, undamaged, and correctly positioned. Cracks, contamination, or misalignment at this stage will directly affect peak shape and reproducibility.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\"><strong>Remove any residual ferrule material <\/strong>from previous GC column installations at both inlet and detector fittings. Leftover material or moisture prevents proper seating of new ferrules.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Skipping the cool-down step is one of the most common and costly installation errors. Waiting until the inlet, detector, and oven have cooled to the recommended temperature protects the column, ferrules, and inlet components.<\/p>\n\n\n\n<figure class=\"wp-block-table\"><table class=\"has-fixed-layout\"><tbody><tr><td><strong>Warning\/Important:<\/strong> Never install a column into a hot inlet or injection port. Cool the system first, then manage carrier gas flow according to the instrument procedure.<\/td><\/tr><\/tbody><\/table><\/figure>\n\n\n\n<h2 class=\"wp-block-heading\"><strong>Step 2: Prepare the Column End<\/strong><\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">The quality of the column cut determines the integrity of the seal. A<strong> poor cut cannot be corrected by tightening <\/strong>and will almost always result in a leak.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Carefully remove about one loop of each column end from its cage, avoiding sharp bends that may cause breakage or unnecessary tension. Identify the inlet and detector ends if specified by the manufacturer.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Using a column cutter, score the column at a clean <strong>90-degree angle<\/strong>. Apply gentle pressure and snap the tubing at the score point. This quick <a href=\"https:\/\/discover.restek.com\/videos\/gnav3590\/gc-column-installation\">GC column installation video<\/a> includes a demonstration of how to cut a fused silica GC column. If you are installing a metal GC column, review <a href=\"https:\/\/discover.restek.com\/en\/videos-zh\/gnav3607\/how-to-cut-a-metal-capillary-gc-column\">this video<\/a> instead for a detailed procedure.&nbsp;<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Inspect the cut under magnification for chips or scratch marks. A proper cut should be flat, smooth, and perpendicular to the tubing axis, with no chips or jagged edges.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">If imperfections are visible, cut the column at least 3 to 5 mm further up.&nbsp;<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Even slight irregularities create micro-leaks that compromise system and column performance under operating conditions.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Once both column ends are properly prepared, carefully mount the column on the rack inside the GC oven, ensuring it is properly supported and positioned for connection to the inlet and detector.<\/p>\n\n\n\n<figure class=\"wp-block-table\"><table class=\"has-fixed-layout\"><tbody><tr><td><strong>Key Insight:<\/strong> A ragged or angled column cut cannot be fixed by tightening the ferrule further. Achieving a perfectly square, flat face is your only defense against system-compromising micro-leaks.<\/td><\/tr><\/tbody><\/table><\/figure>\n\n\n\n<h2 class=\"wp-block-heading\"><strong>Step 3: Thread the Ferrule and Column Nut<\/strong><\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">Ferrule installation errors are a leading cause of leaks because they are often not visible once assembled.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Slide the column nut onto the tubing first, ensuring correct orientation. Then slide the ferrule onto the capillary column with the tapered end facing the fitting. This orientation allows proper compression against the fitting seat. Cut approximately 2 cm from the end of the column to remove any residual ferrule material from the end of the column.&nbsp;<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Insert the column to the correct distance or depth based on instrument specifications. Because this varies by inlet type and injection mode, always confirm this requirement in the instrument manual.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">At this stage, hand-tighten only until the ferrule contacts the fitting. Avoid compressing the ferrule prematurely.<\/p>\n\n\n\n<figure class=\"wp-block-table\"><table class=\"has-fixed-layout\"><tbody><tr><td><strong>Pro Tip:<\/strong> Install to the correct column depth. Your instrument manufacturer will have specific information on the proper column insertion depth. Once you know the depth, you can use a <a href=\"https:\/\/www.restek.com\/c\/1201?Ns=restekProduct.x_productBadge%7C0\" target=\"_blank\" rel=\"noreferrer noopener\">capillary installation gauge<\/a> to measure and pre-swage your ferrule on the column. This simple step eliminates one of the most common causes of misaligned GC column installations.<\/td><\/tr><\/tbody><\/table><\/figure>\n\n\n\n<h2 class=\"wp-block-heading\"><strong>Step 4: Install at the Inlet<\/strong><\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">The inlet, or injection port, connection is one of the most important points in the GC column installation because it directly affects injection efficiency, peak shape, and leak risk. A ferrule compressed incorrectly at the inlet cannot be re-seated and must be replaced.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Carefully insert the prepared column end into the injection port fitting until the depth mark aligns with the face of the fitting. This confirms the column insertion depth is positioned correctly relative to the <a href=\"https:\/\/discover.restek.com\/articles\/gnar1991\/how-to-choose-a-gc-inlet-liner\">GC inlet liner<\/a>. It is essential to make sure the correct insertion depth is used for your GC system and injection type.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Align it so it enters straight and centered to avoid breakage at the connection point. Remember that fused silica cannot tolerate lateral bending force at the connection point.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Thread the column nut onto the fitting by hand until resistance is felt to secure the connection. It should be hand-tight and snug, but not to the point of strain.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">After hand-tightening, apply the manufacturer-recommended wrench increment. For many Vespel\/graphite ferrules, this is about 1\/4 turn for most applications, but always confirm the exact guidance for your system and avoid overtightening. Additional force will deform the ferrule, not improve the seal.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Critical Step: Over-tightening is a one-way mistake. If you feel the column nut stop suddenly before the 1\/4 turn is complete, stop. Check that the connection holds the column in place and does not leak.<\/p>\n\n\n\n<h2 class=\"wp-block-heading\"><strong>Step 5: Install at the Detector<\/strong><\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">The GC detector installation is often rushed, but the detector connection requires the same precision as the inlet. Getting this wrong affects data quality and shortens detector lifespan.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Repeat the nut and ferrule threading process at the detector end to properly connect the column using the same procedure. This ensures correct ferrule orientation. Insertion depth at the detector connection is not universal; it is governed entirely by the specific detector geometry. For a flame ionization detector (FID), the column end is typically positioned close to the jet according to the instrument manufacturer\u2019s specified distance.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Confirm this with your manual, as maintaining the correct distance directly affects flame stability. For a mass spectrometer (MS), insert the column into the transfer line or ion source interface. If it is too close, you risk thermal damage. If it is too far, you reduce sensitivity.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Check documentation for your TCD or ECD specifications as well.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Hand-tighten the column nut, then apply the same 1\/4 turn controlled wrench increment. After tightening, allow the column to form a gentle, natural loop inside the GC oven.<\/p>\n\n\n\n<figure class=\"wp-block-table\"><table class=\"has-fixed-layout\"><tbody><tr><td><strong>Key Insight:<\/strong> Always use matched ferrule materials at both the inlet and detector ends. This ensures consistent thermal expansion behavior across temperature cycles, preventing the formation of micro-leaks during analysis.<\/td><\/tr><\/tbody><\/table><\/figure>\n\n\n\n<h2 class=\"wp-block-heading\"><strong>Step 6: Verify the Installation<\/strong><\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">Verifying the GC column installation serves as the final quality gate before conditioning begins. An undetected leak at this stage will degrade data quality throughout every subsequent run.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Restore the carrier gas flow. Inspect both connections with an <a href=\"https:\/\/www.restek.com\/p\/28500\" target=\"_blank\" rel=\"noreferrer noopener\">electronic leak detector<\/a> to prevent oxygen ingress. If an electronic leak detector is unavailable, use only a manufacturer-approved leak detection solution and take care to ensure the liquid does not enter the analytical column or detector connections.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Confirm that the column head pressure and carrier gas flow rate match your method specifications. Use the instrument&#8217;s internal readout or an <a href=\"https:\/\/www.restek.com\/p\/22656\" target=\"_blank\" rel=\"noreferrer noopener\">external flow meter<\/a>.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">An unexpected pressure or flow can indicate a seating or ferrule compression issue that will compromise column performance. First, check the column in the oven and confirm the fused silica tubing is not kinked, compressed by the oven door, or coiled too tightly. If no problems with the column are observed, you may need to tighten the fitting (do not overtighten) or repeat the installation steps.&nbsp;<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Once pressure and flow rate specifications are met, record the installation in your log.&nbsp;<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Include the serial number, installation date, initial carrier gas conditions, and installer name. Reproducibility begins right here at the installation stage, so ensuring your GC column is properly installed will pay dividends during routine analysis.<\/p>\n\n\n\n<figure class=\"wp-block-embed is-type-video is-provider-youtube wp-block-embed-youtube wp-embed-aspect-16-9 wp-has-aspect-ratio\"><div class=\"wp-block-embed__wrapper\">\n<iframe loading=\"lazy\" title=\"GC Column Installation\" width=\"720\" height=\"405\" src=\"https:\/\/www.youtube.com\/embed\/GGLDnWd4U40?feature=oembed\" frameborder=\"0\" allow=\"accelerometer; autoplay; clipboard-write; encrypted-media; gyroscope; picture-in-picture; web-share\" referrerpolicy=\"strict-origin-when-cross-origin\" allowfullscreen><\/iframe>\n<\/div><\/figure>\n\n\n\n<h2 class=\"wp-block-heading\"><strong>Common GC Column Installation Mistakes<\/strong><\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">Even experienced analysts make these typical errors when GC column installations are rushed. Catching them early prevents avoidable chromatographic issues.<\/p>\n\n\n\n<figure class=\"wp-block-table\"><table class=\"has-fixed-layout\"><tbody><tr><td><strong>Mistake<\/strong><\/td><td><strong>What Happens<\/strong><\/td><td><strong>Why It Matters<\/strong><br><\/td><td><strong>How to Prevent It<\/strong><\/td><\/tr><tr><td>Incorrect Ferrule Orientation<\/td><td>Ferrule installed backwards<\/td><td>Cannot seal properly, leading to persistent leaks<\/td><td>Ensure the tapered end faces the fitting before threading<\/td><\/tr><tr><td>Wrong Insertion Depth<\/td><td>Column seated too shallow or too deep<\/td><td>Causes peak broadening or contact with the liner\/detector components<\/td><td>Use a GC column installation gauge<\/td><\/tr><tr><td>Over-Tightening the Column Nut<\/td><td>Ferrule over-compressed<\/td><td>Damages ferrule, fitting, or fused-silica column<\/td><td>Tighten only to specification (typically ~1\/4 turn after hand-tight)<\/td><\/tr><tr><td>Ragged or Angled Column Cut<\/td><td>Non-square or chipped column end<\/td><td>Creates micro-leaks that cannot be sealed<\/td><td>Inspect under magnification before installation and re-cut if needed<\/td><\/tr><tr><td>Installing Into a Hot Instrument<\/td><td>Column installed above safe temperature<\/td><td>Leads to poor seating, leaks, and possible damage<\/td><td>Cool GC oven, inlet, and detector below ~40 \u00b0C before installing<\/td><\/tr><tr><td>Skipping the leak check (risk of oxygen ingress)<\/td><td>No verification before conditioning<\/td><td>Oxygen ingress causes baseline instability and column degradation<\/td><td>Use an electronic leak detector at both fittings<\/td><\/tr><\/tbody><\/table><\/figure>\n\n\n\n<h2 class=\"wp-block-heading\"><strong>Next Step: Column Conditioning<\/strong><\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">A correctly installed column is not yet ready for use. Conditioning removes residual solvents, moisture, and volatile manufacturing residues that interfere with early runs.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Skipping this step leads to baseline drift, noise, ghost peaks, and elevated column bleed. For detailed conditioning parameters and procedures, consult Restek&#8217;s <a href=\"https:\/\/discover.restek.com\/articles\/gnan1533\/how-to-condition-a-new-capillary-gc-column\">column conditioning guide<\/a>.<\/p>\n\n\n\n<figure class=\"wp-block-table\"><table class=\"has-fixed-layout\"><tbody><tr><td><strong>Warning\/Important:<\/strong> Never exceed your column&#8217;s maximum temperature during conditioning. Crossing this threshold causes rapid stationary phase decomposition, resulting in permanent column bleed and ruined analytical runs.<\/td><\/tr><\/tbody><\/table><\/figure>\n\n\n\n<h3 class=\"wp-block-heading\"><strong>Ensure Accurate GC Column Installation<\/strong><\/h3>\n\n\n\n<p class=\"wp-block-paragraph\">GC column installation is most reliable when each step is handled with care, from preparing the instrument and cutting the column end to seating ferrules correctly, checking for leaks, and conditioning before the first run.&nbsp;<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">A clean, leak-free setup protects peak shape, baseline stability, and column lifetime from the start.&nbsp;<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">If you need help ensuring optimal column performance, confirming installation details, or optimizing method conditions, <a href=\"https:\/\/www.restek.com\/global\/en\/contact-us\" target=\"_blank\" rel=\"noreferrer noopener\">contact Restek<\/a> to schedule a consultation or request a product demonstration.<\/p>\n","protected":false},"excerpt":{"rendered":"<p>Any analyst facing GC column installation understands the underlying tension. The gas chromatography setup looks straightforward on paper, but getting it right is critical. Rushing the step-by-step process can lead to immediate analytical problems such as ghost peaks, split peaks, fitting leaks, column bleed, and wasted runs. However, with the right preparation, GC column installation&#8230;<\/p>\n","protected":false},"author":29,"featured_media":0,"comment_status":"closed","ping_status":"closed","sticky":false,"template":"","format":"standard","meta":{"_acf_changed":false,"_kad_blocks_custom_css":"","_kad_blocks_head_custom_js":"","_kad_blocks_body_custom_js":"","_kad_blocks_footer_custom_js":"","_kadence_starter_templates_imported_post":false,"_kad_post_transparent":"","_kad_post_title":"","_kad_post_layout":"","_kad_post_sidebar_id":"","_kad_post_content_style":"","_kad_post_vertical_padding":"","_kad_post_feature":"","_kad_post_feature_position":"","_kad_post_header":false,"_kad_post_footer":false,"footnotes":"","_ppma_block_editor_authors":""},"categories":[2806],"tags":[],"industries-application":[],"post-badge":[],"resource-type":[],"product-library":[],"resource-technique":[],"ppma_author":[449],"class_list":["post-97348","post","type-post","status-publish","format-standard","hentry","category-blogs"],"acf":[],"taxonomy_info":{"category":[{"value":2806,"label":"Blogs"}]},"featured_image_src_large":false,"author_info":{"display_name":"Jonathan Keim","author_link":"https:\/\/discover.restek.com\/it\/author\/jonathan-keim\/"},"comment_info":0,"category_info":[{"term_id":2806,"name":"Blogs","slug":"blogs","term_group":0,"term_taxonomy_id":2806,"taxonomy":"category","description":"","parent":0,"count":456,"filter":"raw","cat_ID":2806,"category_count":456,"category_description":"","cat_name":"Blogs","category_nicename":"blogs","category_parent":0}],"tag_info":false,"authors":[{"term_id":449,"user_id":29,"is_guest":0,"slug":"jonathan-keim","display_name":"Jonathan Keim","avatar_url":{"url":"https:\/\/discover.restek.com\/wp-content\/uploads\/people-keim-jonathan.png","url2x":"https:\/\/discover.restek.com\/wp-content\/uploads\/people-keim-jonathan.png"},"author_category":"1","first_name":"Jonathan","last_name":"Keim","user_url":"https:\/\/www.linkedin.com\/in\/jonathan-keim-24a7342\/","job_title":"Content Marketing Manager","description":"Jonathan \"Munch\" Keim joined Restek in 2001 with a BS in chemistry from Juniata College and an MS in analytical chemistry from the University of Pittsburgh. With over twenty years at Restek, he has served in a variety of roles, including managing the technical service group and as the product manager for Sample Prep. Jonathan is passionate about chromatography education and now serves as the Content Marketing Manager leading an outstanding team that helps people all over the world understand and optimize their chromatography."}],"_links":{"self":[{"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/posts\/97348","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/users\/29"}],"replies":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/comments?post=97348"}],"version-history":[{"count":6,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/posts\/97348\/revisions"}],"predecessor-version":[{"id":105436,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/posts\/97348\/revisions\/105436"}],"wp:attachment":[{"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/media?parent=97348"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/categories?post=97348"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/tags?post=97348"},{"taxonomy":"industries-application","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/industries-application?post=97348"},{"taxonomy":"post-badge","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/post-badge?post=97348"},{"taxonomy":"resource-type","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/resource-type?post=97348"},{"taxonomy":"product-library","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/product-library?post=97348"},{"taxonomy":"resource-technique","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/resource-technique?post=97348"},{"taxonomy":"author","embeddable":true,"href":"https:\/\/discover.restek.com\/it\/wp-json\/wp\/v2\/ppma_author?post=97348"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}