アプリケーション

A Rapid and Sensitive LC-MS/MS Method for the Analysis of Three Forms of Thyroid Hormones Using Raptor Biphenyl LC Columns

26 Oct 2020

feature BAAN2112

Thyroid hormones are essential for the regulation of development and growth in humans and animals. The thyroid gland produces thyroxin (T4) and tri-iodothyronine (T3) and quickly releases these compounds into the circulatory system. The concentration of circulating T4 is 50-60 times higher than T3 and the majority of these molecules are bound to blood proteins. The unbound or “free” T3 and T4 are the active forms of the hormone which only represent a small portion (less than 1%) of total thyroid hormones. Accurate and sensitive measurement of low pg/mL levels of free hormones is necessary to assess thyroid function for both veterinary and human diagnostics. Reverse tri-iodothyronine (rT3) is an inactive form that results from T4 biotransformation. Since rT3 functions as the feedback inhibitor of thyroid hormone production, the measurement of rT3 can be an important diagnostic marker with clinical implications.

The intent of this application was to develop an LC-MS/MS method for the analysis of thyroid hormones at the free form levels using the highly efficient and selective Raptor Biphenyl LC column. The clinical applicability of this method was demonstrated by analyzing fortified thyroid hormone in phosphate buffered saline (PBS) containing 4% human albumin.

Experimental

Instrument and Analytical Conditions

The instrument and analytical conditions are listed in Table I. The analyte MRMs are shown in Table II.

Table I:  Analytical Conditions for Waters Xevo TQ-S with Acquity UPLC

Analytical ColumnRaptor Biphenyl 2.7 µm, 100 mm x 2.1 mm (cat.# 9309A12)
Guard ColumnRaptor Biphenyl EXP guard column cartridge 2.7 µm, 5 mm x 2.1 mm (cat.# 9309A0252)
Mobile Phase A0.1% Formic acid in water
Mobile phase B0.1% Formic acid in methanol
GradientTime (min)%B
0.070
2.080
2.0170
3.570
Flow Rate0.4 mL/min
Column Temp.40 °C
Ion ModePositive ESI
Capillary Voltage2.0 kV
Gas Flow800 (L/Hr) desolvation
200 (L/Hr) cone
7.0 (bar) nebulizer
Desolvation Temp.600 °C

Table II:  Analyte MS/MS Transitions

AnalytePrecursor IonProduct Ion QuantifierProduct Ion Qualifier
T3652.07606.10508.10
rT3652.07606.10508.10
T4778.03731.97323.87
T4-13C6784.09738.04–

Sample/Calibration Standard Preparation

Human albumin was dissolved in PBS solution to a final concentration of 4%. This solution was used to prepare calibration standards ranging from 2 to 400 pg/mL. Standard solutions (0.5 mL) were fortified with 5 µL of internal standard (T4-13C6 [1 ng/mL]) and mixed with 1 mL of acetonitrile in a 4 mL glass vial. A 2 mL aliquot of ethyl acetate was added, stirred for 2 min, and then centrifuged for 10 min at 4,300 rpm. The organic phase was removed and placed into a 4 mL glass vial, then evaporated to dryness at 55 °C under a gentle stream of nitrogen. The dried extract was reconstituted with 80 µL of a 30:70 water:methanol solution and injected (10 µL) into the LC-MS/MS for analysis.

Results

Chromatographic Separation of Thyroid Hormones

Since the most sensitive mass transitions for T3 and rT3 are identical, it is necessary to chromatographically separate these two compounds for accurate quantitation. With independent injections of T3 and rT3, it was shown that T3 and rT3 were completely resolved with the Raptor Biphenyl column (Figure 1). An example chromatogram (Figure 2) shows the baseline resolution of 3 forms of thyroid hormone from extracted sample (20 pg/mL) within 2 minutes.

Figure 1: Chromatographic separation of T3 and rT3 on the Raptor Biphenyl column. Note: peak separation is critical for accurate analysis because these compounds have identical transitions and cannot be distinguished by MS/MS alone.
T3 and rT3 Analysis on Raptor™ Biphenyl by LC-MS/MS

LC_CF0603

Peaks

PeakstR (min)Precursor IonQuantifier IonQualifier Ion
1.Tri-iodothyronine (T3)1.03652.07606.10508.10
2.Reverse tri-iodothyronine (rT3)1.24652.07606.10508.10
Figure 2: Complete chromatographic separations of all three extracted thyroid hormones (20 pg/mL) are obtained in less than 2 minutes.
Thyroid Hormone Analysis in 4% Human Albumin in PBS on Raptor Biphenyl by LC-MS/MS

LC_CF0605

Peaks

PeakstR (min)Precursor IonProduct Ion
1.Tri-iodothyronine (T3)1.04652.07606.10
2.Reverse tri-iodothyronine (rT3)1.25652.07606.10
3.Thyroxin (T4)1.43778.03731.97
4.Thyroxin-13C61.43784.09738.04

Conditions

ColumnRaptor Biphenyl (cat.# 9309A12)
Dimensions:100 mm x 2.1 mm ID
Particle Size:2.7 µm
Pore Size:90 Å
Guard Column:Raptor Biphenyl EXP guard column cartridge 5 mm, 2.1 mm ID, 2.7 µm (cat.# 9309A0252)
Temp.:40 °C
Standard/Sample
Diluent:30:70 Water:methanol
Conc.:20 pg/mL
Inj. Vol.:10 µL
Mobile Phase
A:0.1% Formic acid in water
B:0.1% Formic acid in methanol
Time (min)Flow (mL/min)%A%B
0.000.43070
2.000.42080
2.010.43070
3.500.43070
DetectorMS/MS
Ion Mode:ESI+
Mode:MRM
InstrumentHPLC

Linearity

Good linearities (1/x weighted) were obtained for all 3 forms of thyroid hormones with coefficient of determination (R2) values > 0.990 from 2 to 400 pg/mL (for T3) or 5 to 400 pg/mL (for T4 and rT3). The %deviation was < 15%. Linearity plots are shown in Figure 3.

Figure 3: Good linear responses were obtained for thyroid hormones T4, T3, and rT3 at low pg/mL levels.

A) T4 Calibration Curve, R2 = 0.999

figure article BAAN2112 01

B) T3 Calibration Curve, R2 = 0.998

figure article BAAN2112 02

C) rT3 Calibration Curve, R2 = 0.991

figure article BAAN2112 03

Conclusions

The Raptor Biphenyl column is excellent for rapid and sensitive analysis of thyroid hormones. With the method described here, concentrations of thyroid hormones as low as 2 pg/mL (T3) or 5 pg/mL (T4 and rT3) can be determined with less than 3.5 minutes of total analysis time. The analytical method is thus applicable to the clinical analysis of free thyroid hormone at low pg/mL levels.

この記事で紹介した製品


Raptor Biphenyl, 2.7 µm, 100 x 2.1 mm HPLCカラム
Raptor Biphenyl EXP Guard Column Cartridge, 2.7 µm, 5 x 2.1 mm, 3-pk.

著者 / 執筆者

  • Shun-Hsin Liang, PhD

    Shun-Hsin is a senior principal scientist in LC Solutions at Restek. He received his bachelor’s degree from the National Taiwan University in 1988 and obtained his PhD from Michigan State University in 1996. He performed postdoctoral research at the University of Michigan from 1996 to 2000 for oncology studies. In 2001, he was appointed as research faculty at The Pennsylvania State University and focused on molecular toxicology research. In 2006, he joined MPI Research Inc. as a senior research scientist and was a study director for GLP analytical projects. In 2013, Dr. Liang joined the LC Solutions department at Restek and specialized in developing application methods across the fields of environmental, food safety, and life sciences.

    View Articles
  • Paul Connolly
  • Ty Kahler
BAAN2112