Method 505: Analysis of Organohalide Pesticides and Commercial PCBs in Water by Microextraction and GC

by Microextraction and Gas Chromatography

27 Oct 2020

feature EVSS1013

With the advent of modern agriculture and its vast selection of chemical pest control measures, the farming community has made significant increases in productivity and efficiency. Crop yield per acre is extremely high, due in part to the role of pesticides and herbicides in mitigating the devastating effects of many plant and insect pests. However, the use of these chemicals can have drawbacks, including surface and ground water contamination. EPA Method 505 is used for organohalide pesticide analysis in drinking and ground water. The Rtx-CLPesticides and Rtx-CLPesticides2 column set works well for the analysis of organohalide compounds because the columns are highly selective for analytes that contain electronegative substituents. The optimized parallel dual-column method shown here provides fast analysis times, which improves sample throughput.

VendorColumn PairAnalysis Time (min)CoelutionsRestek Advantage
RestekRtx-CLPesticides /
Rtx-CLPesticides2
18 / 18.51 / 1All compounds resolved on at least one column.
Competitor Acolumn 1 / column 214 / 140 / 1
Competitor Bcolumn 1 / column 235 / 361 / 2

Comparison based on published competitor data. All columns tested were 0.32 mm ID.

Organohalide pesticide analysis (Method 505) on Rtx-CLPesticides and Rtx-CLPesticides2 columns.
Pesticides of U.S. EPA Method 505 on Rtx-CLPesticides and Rtx-CLPesticides2

GC_EV01004

Conditions

ColumnsRtx-CLPesticides 30 m, 0.32 mm ID, 0.32 µm (cat.# 11141)
and Rtx-CLPesticides2 30 m, 0.32 mm ID, 0.25 µm (cat.# 11324)
using Rxi guard column 5 m, 0.32 mm ID (cat.# 10039)
with deactivated universal “Y” Press-Tight connector (cat.# 20405-261)
Standard/Sample
200 ng/mL 505 organohalide pesticide mix (cat.# 32024)
4.2 µg/mL simazine (cat.# 32236)
4.2 µg/mL atrazine (cat.# 32208)
Diluent:Methanol
Injection
Inj. Vol.:2 µL splitless (hold 0.75 min)
Liner:Cyclo double taper (4 mm) (cat.# 20896)
Inj. Temp.:250 °C
Oven
Oven Temp.:90 °C (hold 1 min) to 310 °C at 10 °C/min (hold 5 min)
Carrier GasHe, constant flow
Linear Velocity:40 cm/sec
DetectorMicro-ECD @ 325 °C
InstrumentAgilent/HP6890 GC
NotesThis chromatogram was obtained using an Agilent micro-ECD. To obtain comparable results, you will need to employ a micro-ECD in addition to confirmational dual columns connected to a 5-meter guard column using a “Y” Press-Tight connector.

この記事で紹介した製品


Rtx-CLPesticides GC Capillary Column Kit, 0.32 mm ID, 0.32 µm
Rtx-CLPesticides GC Capillary Column, 15 m, 0.25 mm ID, 0.25 µm
Rtx-CLPesticides2 GC Capillary Column, 15 m, 0.32 mm ID, 0.25 µm

著者 / 執筆者

  • Restek Corporation

    Restek is a leading provider of chromatography columns, accessories, and certified reference materials. Trust Restek for reliable, high-quality analytical solutions.

    View Articles
EVAR4504