技術資料

Raptor FluoroPhenyl: Get the Power of HILIC and RP Modes in One LC Column

22 Dec 2020

feature GNBR2368

With Raptor LC columns, Restek chemists became the first to combine the speed of 2.7 and 5 μm superficially porous particles (also known as SPP or “core-shell” particles) with the resolution of highly selective USLC technology, improving separations and speeding up analysis times with standard HPLC instruments. Raptor then evolved to bring that same improved speed, efficiency, and selectivity to UHPLC analyses by offering 1.8 μm particle columns. Learn more about Raptor LC columns at www.restek.com/raptor

With the addition of Raptor FluoroPhenyl columns, Restek further expanded the speed and reliability of Raptor column technology into the HILIC realm. Restek’s Raptor FluoroPhenyl phase offers chromatographers the ability to run in reversed-phase or HILIC mode for a variety of compounds. The Restek Raptor FluoroPhenyl column is also amenable to LC-MS because it is extremely reliable and efficient with acidic mobile phases.

Switch to a Raptor FluoroPhenyl LC column for reliable performance in both reversed-phase and HILIC modes.

Column Description:

article ligand gnbr2368b 01 1

Stationary Phase Category:

Pentafluorophenyl propyl (L43)

Ligand Type:

Fluorophenyl

Particle:

1.8 μm, 2.7 μm, or 5 μm superficially porous particle (SPP or “core-shell” particle) silica

Pore Size:

90 Å

Surface Area:

125 m2/g (1.8 μm),
130 m2/g (2.7 μm),
or 100 m2/g (5 μm)

Recommended Usage:

pH Range: 2.0–8.0
Maximum Temperature: 80 °C
Maximum Pressure: 1034 bar/15,000 psi* (1.8 μm),
600 bar/8700 psi (2.7 μm); 400 bar/5800 psi (5 μm)
* For maximum lifetime, recommended maximum pressure for 1.8 μm particles is 830 bar/12,000 psi.

Properties:

  • Capable of both reversed-phase and HILIC separations.
  • Ideal for increasing sensitivity and selectivity in LC-MS analyses.
  • Offers increased retention for charged bases.

Switch to a Raptor FluoroPhenyl LC column when:

  • You observe limited retention and selectivity on a C18 for basic compounds
  • You need increased retention of hydrophilic compounds.

Column Interaction Profile:

article profile gnbr1891c 01

Defining Solute Interactions:

  • Cation exchange

Complementary Solute Interaction:

  • Polarizability
  • Dispersion

Solute Retention Profile:

article graph gnbr2368b 01 1

Target Analyte Structures:

  • Nitrogen-containing

Target Analyte Functionalities:

  • Protonated amines
  • Quaternary ammonium compounds
  • Positively charged moieties
  • Lewis bases

Raptor FluoroPhenyl LC Columns: Rugged, Check—Reproducible, Double Check.

Of course, Raptor FluoroPhenyl columns are rugged; that is to be expected. And, they are exceptionally reproducible as well. Reproducibility can be an issue for fluorinated phenyl phases, which is why we engineered all our columns for dependable performance. Lot to lot, column to column, and injection to injection, every Raptor FluoroPhenyl column gives a consistent performance that you can count on: consider it done.

Figure 1: Raptor FluoroPhenyl columns maintain efficiency in any dimension or particle size—even at their maximum recommended operating pressures—so you can run at high linear velocities with confidence.
article figure gnbr2368b 01
Figure 2: Strict quality control ensures Raptor FluoroPhenyl columns are exceptionally reproducible, so you get predictable performance from every column.
Taxanes Lot to Lot Comparison on Raptor™ FluoroPhenyl

LC_BA0352

Peaks

PeakstR (min)Precursor IonProduct Ion 1Product Ion 2
1.Baccatin III0.97587.0405.1105.0
2.Docetaxel1.25808.1527.3226.1
3.Paclitaxel1.33854.1569.3286.2

Conditions

ColumnRaptor™ FluoroPhenyl (cat.# 931955E)
Dimensions:50 mm x 3 mm ID
Particle Size:5 µm
Temp.:35 °C
Standard/Sample
Diluent:Water
Conc.:100 ng/mL
Inj. Vol.:5 µL
Mobile Phase
A:0.1% Formic acid in water
B:0.1% Formic acid in acetonitrile
Time (min)Flow (mL/min)%A%B
0.000.87525
2.000.8595
2.010.87525
3.500.87525
DetectorMS/MS
Ion Mode:ESI+
Mode:MRM
InstrumentUHPLC
Figure 3: From start to finish, Raptor FluoroPhenyl columns provide accurate, reproducible results.
4-Methylimidazole Lifetime Over 500 Injection on Raptor FluoroPhenyl

LC_FF0561

Peaks

PeakstR (min)Precursor IonProduct Ion
1.4-Methylimidazole (4-MEI)1.198356

Conditions

ColumnRaptor FluoroPhenyl (cat.# 9319A52)
Dimensions:50 mm x 2.1 mm ID
Particle Size:2.7 µm
Temp.:35 °C
Standard/Sample
Diluent:Acetonitrile
Conc.:100 ng/mL
Inj. Vol.:5 µL
Mobile Phase
A:0.1% Formic acid in water
B:0.1% Formic acid in acetonitrile
Time (min)Flow (mL/min)%A%B
0.000.6595
2.000.67030
2.010.6595
3.500.6595
DetectorMS/MS
Ion Mode:ESI+
Mode:MRM
InstrumentUHPLC

More Separating Power than a C18

C18 columns work well for many compounds, but they just don’t work for everything. Raptor FluoroPhenyl columns can provide greater selectivity and retention for analytes that are not easily separated by C18 phase chemistry. For example, interest in vitamin D status is on the rise in clinical diagnostics, but accurate analysis is only possible if the epimeric forms of both vitamin D2 and D3 25-hydroxy metabolites can be distinguished. Typical reversed-phase C18 columns cannot separate these isobaric epimers, which differ in bioactivity, but the new Raptor FluoroPhenyl column provides adequate chromatographic resolution so accurate results are generated and the proper diagnosis can be made.

Figure 4: Reversed-phase C18 columns do not have the right selectivity or retention mechanism to separate the epimers of vitamin D2 and D3 25-hydroxy metabolites.
25-Hydroxyvitamin D2/D3 and 3-Epi-25-Hydroxyvitamin D2/D3 on Raptor™ ARC-18 by LC-MS/MS

LC_CF0649

Conditions

ColumnRaptor ARC-18 (cat.# 9314A12)
Dimensions:100 mm x 2.1 mm ID
Particle Size:2.7 µm
Temp.:30 °C
Standard/Sample
Diluent:Water:methanol (50:50)
Conc.:25-50 ng/mL
Inj. Vol.:5 µL
Mobile Phase
A:0.1% Formic acid in water
B:Methanol
Time (min)Flow (mL/min)%A%B
0.000.52575
4.000.52080
12.000.52080
12.100.52575
14.000.52575
DetectorMS/MS
Ion Mode:ESI+
Mode:MRM
InstrumentHPLC
Figure 5: Raptor FluoroPhenyl columns have the selectivity and retention you need to quickly and easily separate compounds that coelute on a C18, such as the epimeric forms of vitamin D2 and D3 25-hydroxy metabolites.
25-Hydroxyvitamin D2/D3 and 3-Epi-25-Hydroxyvitamin D2/D3 on Raptor FluoroPhenyl by LC-MS/MS

LC_CF0644

Conditions

ColumnRaptor FluoroPhenyl (cat.# 9319A1E)
Dimensions:100 mm x 3 mm ID
Particle Size:2.7 µm
Temp.:30 °C
Standard/Sample
Diluent:Water:methanol (50:50)
Conc.:25-50 ng/mL
Inj. Vol.:5 µL
Mobile Phase
A:0.1% Formic acid in water
B:Methanol
Time (min)Flow (mL/min)%A%B
0.000.62575
4.000.61585
4.100.60100
5.000.60100
5.010.62575
7.000.62575
DetectorMS/MS
Ion Mode:ESI+
Mode:MRM
InstrumentHPLC
Lot Comparison of 25-Hydroxyvitamin D3 and 3-Epi-25-Hydroxyvitamin D3 on Raptor™ FluoroPhenyl by LC-MS/MS

LC_CF0646

Get the Power of HILIC and RP Modes in One LC Column

HILIC chromatography is becoming the go-to solution for compounds that are difficult to retain on a C18. The Raptor FluoroPhenyl column gives chromatographers the flexibility to evaluate compound retention in both reversed-phase and HILIC modes. The analysis of 4-methylimidazole (4-MEI), which is a byproduct of caramel coloring in foods and beverages, can be problematic by reversed-phase chromatography due to very limited retention. However, 4-MEI is well retained on a Raptor FluoroPhenyl column and can easily be analyzed using HILIC mode and simple LC and LC-MS/MS compatible mobile phases.

Figure 6: Sometimes, adequate retention cannot be obtained with a C18. The Raptor FluoroPhenyl column performs dependably in either HILIC or RP mode, so you can use the mode that is best for your analytes.
Analysis of 4-Methylimidazole on Raptor FluoroPhenyl by LC-MS/MS

LC_FF0559

Peaks

PeakstR (min)Precursor IonProduct Ion
1.4-Methylimidazole (4-MEI)1.198356

Conditions

ColumnRaptor FluoroPhenyl (cat.# 9319A52)
Dimensions:50 mm x 2.1 mm ID
Particle Size:2.7 µm
Temp.:35 °C
Standard/Sample
Diluent:Acetonitrile
Conc.:100 ng/mL
Inj. Vol.:5 µL
Mobile Phase
A:0.1% Formic acid in water
B:0.1% Formic acid in acetonitrile
Time (min)Flow (mL/min)%A%B
0.000.6595
2.000.67030
2.010.6595
3.500.6595
DetectorMS/MS
Ion Mode:ESI+
Mode:MRM
InstrumentUHPLC
4-Methylimidazole Lot to Lot Comparison on Raptor™ FluoroPhenyl

LC_FF0560

Fast, Accurate Analysis of Basic Compounds

Taxane drugs—such as paclitaxel, docetaxel, and their precursor, baccatin III—are early chemotherapy treatment options. Accurate analysis is critical because these drugs are used for many types of metastatic cancers. As shown here, the selectivity of the Raptor FluoroPhenyl column provides excellent retention and resolution of these structurally similar compounds. Reliable results for these basic compounds can be obtained in fast analysis times using LC-MS/MS friendly solvents.

Figure 7: Raptor FluoroPhenyl columns allow taxane drugs and other basic compounds to be separated quickly and effectively.
article structure gnbr2368b baccatiniii
article structure gnbr2368b
article structure gnbr2368b
Analysis of Taxanes on Raptor FluoroPhenyl by LC-MS/MS

LC_BA0351

Peaks

PeakstR (min)Precursor IonProduct Ion 1Product Ion 2
1.Baccatin III0.97587.0405.1105.0
2.Docetaxel1.25808.1527.3226.1
3.Paclitaxel1.33854.1569.3286.2

Conditions

ColumnRaptor FluoroPhenyl (cat.# 931955E)
Dimensions:50 mm x 3 mm ID
Particle Size:5 µm
Temp.:35 °C
Standard/Sample
Diluent:Water
Conc.:100 ng/mL
Inj. Vol.:5 µL
Mobile Phase
A:0.1% Formic acid in water
B:0.1% Formic acid in acetonitrile
Time (min)Flow (mL/min)%A%B
0.000.87525
2.000.8595
2.010.87525
3.500.87525
DetectorMS/MS
Ion Mode:ESI+
Mode:MRM
InstrumentUHPLC

Exceptional Selectivity for Clinical Analyses

The analysis of normetanephrine and metanephrine provides another example of the power of Raptor FluoroPhenyl columns for analyzing basic compounds. Accurately determining these metabolites of epinephrine/norepinephrine in plasma or urine is one of the best diagnostic tests for neuroendocrine tumors (pheochromocytomas). Normetanephrine and metanephrine are difficult to retain by typical C18 reversed-phase chromatography; however, the Raptor FluoroPhenyl column provides a simple, fast chromatographic solution to this challenging assay. The Raptor FluoroPhenyl column’s unique combination of aromatic retention and cation exchange mechanisms are not provided by a C18 column and result in reliable, high-quality separations.

Figure 8: Rapid analysis of metanephrine and normetanephrine on a Raptor FluoroPhenyl column.
Metanephrines on Raptor FluoroPhenyl by LC-MS/MS

LC_CF0642

Peaks

PeakstR (min)Precursor IonProduct Ion 1Product Ion 2
1.Normetanephrine1.15166.1121.1134.0
2.Metanephrine1.52180.1165.1148.3

Conditions

ColumnRaptor FluoroPhenyl (cat.# 9319A12)
Dimensions:100 mm x 2.1 mm ID
Particle Size:2.7 µm
Temp.:30 °C
Standard/Sample
Conc.: 20 ng/mL in water
Inj. Vol.:5 µL
Mobile Phase
A:Water + 0.2% formic acid
B:Methanol
Time (min)Flow (mL/min)%A%B
0.000.4982
2.000.46040
2.010.4982
6.000.4982
DetectorMS/MS
Ion Mode:ESI+
Mode:MRM
InstrumentUHPLC
Lot to Lot Comparison of Metanephrines on the Raptor™ FluoroPhenyl by LC-MS/MS

LC_CF0643

Experience Selectivity Accelerated. Order the Raptor FluoroPhenyl LC Column today.

Related Products


Raptor FluoroPhenyl, 5 µm, 50 x 3.0 mm HPLCカラム
Raptor FluoroPhenyl, 2.7 µm, 50 x 2.1 mm HPLCカラム
Raptor ARC-18, 2.7 µm, 100 x 2.1 mm HPLCカラム
Raptor FluoroPhenyl, 2.7 µm, 100 x 3.0 mm HPLCカラム
Raptor FluoroPhenyl, 2.7 µm, 100 x 2.1 mm HPLCカラム
EXP Hand-Tight Fitting (Nut w/Ferrule)
EXP Direct Connect Holder for EXP Guard Cartridges, Includes Fitting & Ferrules
UltraShield UHPLC PreColumn Filter, 0.2 µm Frit, ea.

著者 / 執筆者

  • Restek Corporation

    Restek is a leading provider of chromatography columns, accessories, and certified reference materials. Trust Restek for reliable, high-quality analytical solutions.

    View Articles
GNBR2368
バージョン B