Blogs

Have there been any new applications developed for Diquat and Paraquat?

06 Sep 2018

I am happy to answer “yes” to this question. Those of you who have been doing this analysis might have been using our Ultra Quat column that I talked about in an earlier blog post.

This updated method uses the Raptor HILIC-Si, which is a newer product we are excited to have.

Paraquat and Diquat on Raptor HILIC-Si

Paraquat and Diquat on Raptor HILIC-Si

LC_FS0512

Peaks

PeakstR (min)Precursor Ion 1Product Ion 1Precursor Ion 2Product Ion 2
1.Diquat3.25183.3157.2183.3130.1
2.Paraquat3.46185.3170.3171.377.2

Conditions

ColumnRaptor HILIC-Si (cat.# 9310A52)
Dimensions:50 mm x 2.1 mm ID
Particle Size:2.7 µm
Pore Size:90 Å
Temp.:45 °C
Standard/SampleParaquat & diquat calibration mix (cat.# 32437)
Diluent:Mobile phase B
Conc.:50 ng/mL
Inj. Vol.:5 µL
Mobile Phase
A:Water, 50 mM ammonium formate, 0.5% formic acid
B:25:75 Water:acetonitrile, 50 mM ammonium formate, 0.5% formic acid
Time (min)Flow (mL/min)%A%B
0.000.60100
4.000.63565
4.010.60100
7.000.60100
DetectorMS/MS
Ion Mode:ESI+
Mode:MRM
InstrumentHPLC

We are pleased with this application, particularly since it does not require any ion pairing agents or chaotropic salts. The mobile phase modifiers are very common ones that most labs will have on hand and extended equilibration times are not needed (as compared to NaPF6 or Heptafluorobutyric acid). The run time is short and sweet.  🙂

I hope you find this post helpful and thanks for reading.

この記事で紹介した製品


Raptor HILIC-Si, 2.7 µm, 50 x 2.1 mm HPLCカラム
Paraquat & Diquat Calibration Mix, 1000 µg/mL, Water, 1 mL/ampul
GNBL4687