FAQ

How long can I keep my samples in gas sampling bags?

31 Jul 2025

Single-use sample bags are typically intended for short-term storage (approximately 24–48 hours). Storage for >48 hours is usually not recommended unless validation studies verify longer stability of target analytes. We recommend the following guidelines when using gas sampling bags:

Before Sampling

  • Store unused bags in a clean environment, sealed in an outer bag to prevent adsorption of contaminants.
  • Preclean bags before use by flushing with high-purity nitrogen.
  • Leak rate should not exceed 0.1″ Hg/min.

During Sampling

  • Do not fill bags more than 80%.
  • Be sure the PTFE tubing used for bag connection is clean.
  • Use a vacuum box sampler for direct bag filling in order to avoid contamination from a sampling pump.
  • 3 L/min is a typical flow rate.

After Sampling

  • Bags are intended for a single use due to potential sample adsorption onto the bag film.
  • Hold times are typically 48 hours unless validation study demonstrates longer stability.
  • Protect samples from direct sunlight and store above 0 °C to prevent condensation.
  • Transport in rigid, opaque container to prevent bag puncture; do not ship by air unless samples will be kept in a pressurized area.

著者 / 執筆者

GNOT4989