{"id":43370,"date":"2021-07-21T14:30:00","date_gmt":"2021-07-21T14:30:00","guid":{"rendered":"https:\/\/discover.restek.com\/uncategorized\/simplify-alcohol-analyses-with-application-designed-chromatography-columns\/"},"modified":"2026-01-03T15:44:26","modified_gmt":"2026-01-03T15:44:26","slug":"simplify-alcohol-analyses-with-application-designed-chromatography-columns","status":"publish","type":"post","link":"https:\/\/discover.restek.com\/ja\/articles-fr\/cfar3698\/simplify-alcohol-analyses-with-application-designed-chromatography-columns","title":{"rendered":"Simplify Alcohol Analyses with Application-Designed Chromatography Columns"},"content":{"rendered":"\n<p class=\"wp-block-paragraph\">On the surface, alcohol analyses appear relatively straightforward. They involve few target analytes, are not extremely sensitive, and have quick sample preparation. However, they are not as simple as they seem. Ethanol analyses present distinct chromatography challenges when assaying in biological samples. Compounds found in biological matrices can interfere with the identification and quantitation of ethanol and its associated analytes.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Restek has developed tailored columns to address the common analyte interferences that can hinder productive chromatography. Dual capillary column sets for GC are available for common blood alcohol analyses, and high-throughput analytical LC column and guard cartridges are available for urinary alcohol assays. Starting method development with tailored columns designed specifically to lessen interferences can help labs develop methods faster and keep instruments running.<\/p>\n\n\n\n<h2 class=\"wp-block-heading\">Blood Alcohol Content (BAC)<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">Gas chromatography coupled with headspace sampling and a flame ionization detector (HS-GC-FID) is the most commonly used technique for the analysis of blood alcohol in biological specimens from both living and dead persons. HS-GC-FID is remarkedly simple, reliable, and linear over the analytical range needed for the quantitative detection of alcohol. The troublesome part of the chromatographic analysis is ensuring analyte specificity from endogenous compounds. This is extremely important when analyzing postmortem samples due to the risk that fermentation or putrefaction may produce volatile chemical byproducts that can interfere with the analysis. For this reason, it is always recommended to use a dual column method. By splitting the sample after the injection port and analyzing it simultaneously on two different stationary phases, we can both quantify and confirm the analyte&#8217;s identity in a single run (Figure 1). The target analytes will elute in the calibrated retention time window on both columns, thus confirming a positive identification. To simplify column selection, Restek has paired two distinct sets of column phases: Rtx-BAC1 and Rtx-BAC2 columns and Rtx-BAC Plus 1 and Rtx-BAC Plus 2 columns for in vivo and postmortem analysis.<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><strong>Figure 1:<\/strong>\u00a0HS-GC-FID dual column setup is common for BAC testing.<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n<style>.kb-image43370_187050-1b .kb-image-has-overlay:after{opacity:0.3;}<\/style>\n<div class=\"wp-block-kadence-image kb-image43370_187050-1b\"><figure class=\"aligncenter size-medium_large\"><img loading=\"lazy\" decoding=\"async\" width=\"768\" height=\"829\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-CFAR3698-01-768x829.jpg\" alt=\"\" class=\"kb-img wp-image-19252\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-CFAR3698-01-768x829.jpg 768w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-CFAR3698-01-278x300.jpg 278w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-CFAR3698-01-948x1024.jpg 948w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-CFAR3698-01-1422x1536.jpg 1422w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-CFAR3698-01.jpg 1800w\" sizes=\"auto, (max-width: 768px) 100vw, 768px\" \/><\/figure><\/div>\n\n<\/div><\/div><\/div>\n\n\n<p class=\"wp-block-paragraph\">Rtx-BAC1 and Rtx-BAC2 columns are designed for routine in vivo blood alcohol analyses (Figure 2). This column pair has been in use for decades and has proven to provide stable, defensible results for routine BAC testing. When using&nbsp;<em>n<\/em>-propanol as the internal standard, this is the industry standard.<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><strong>Figure 2:<\/strong>\u00a0The industry standard for in vivo BAC analysis is the Rtx-BAC1 and Rtx-BAC2 column set.<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n\n<p class=\"has-medium-font-size wp-block-paragraph\">BAC Resolution Control Standard with&nbsp;<em>n<\/em>-Propanol on Rtx-BAC1 Column<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article\"><div class=\"wp-block-custom-chromatogram-article\"><div class=\"chromatogram-image regular-image\"><img decoding=\"async\" src=\"https:\/\/ez.restek.com\/images\/cgram\/gc_cf1172.png\" alt=\"BAC Resolution Control Standard n-Propanol on Rtx-BAC1\" title=\"-\"><\/div><p class=\"article-id\" style=\"text-align:center\"> GC_CF1172<\/p><div class=\"chromatogram-peaks\"><h4>Peaks<\/h4><table class=\"peaks col-lg-6 col-12\">\n<thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 75px\">t<sub>R<\/sub> (min)<\/th><th style=\"text-align: center;width: 75px\">Conc.<br \/>(\u00b5g\/mL)<\/th><\/tr><\/thead>\n<tbody><tr><td class=\"num\">1.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Methanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/67-56-1\/Methanol\" target=\"_blank\" rel=\"noopener\">Methanol<\/a><\/td><td class=\"oth\">0.886<\/td><td class=\"oth\">200<\/td><\/tr>\n<tr><td class=\"num\">2.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Acetaldehyde\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/75-07-0\/Acetaldehyde\" target=\"_blank\" rel=\"noopener\">Acetaldehyde<\/a><\/td><td class=\"oth\">0.993<\/td><td class=\"oth\">200<\/td><\/tr>\n<tr><td class=\"num\">3.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Ethanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/64-17-5\/Ethanol\" target=\"_blank\" rel=\"noopener\">Ethanol<\/a><\/td><td class=\"oth\">1.123<\/td><td class=\"oth\">200<\/td><\/tr>\n<tr><td class=\"num\">4.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2-Propanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/67-63-0\/2-Propanol\" target=\"_blank\" rel=\"noopener\">2-Propanol<\/a><\/td><td class=\"oth\">1.381<\/td><td class=\"oth\">200<\/td><\/tr>\n<tr><td class=\"num\">5.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Acetone\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/67-64-1\/Acetone\" target=\"_blank\" rel=\"noopener\">Acetone<\/a><\/td><td class=\"oth\">1.709<\/td><td class=\"oth\">200<\/td><\/tr>\n<tr><td class=\"num\">6.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for n-Propanol (IS)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/71-23-8\/n-Propanol\" target=\"_blank\" rel=\"noopener\">n-Propanol (IS)<\/a><\/td><td class=\"oth\">1.873<\/td><td class=\"oth\">200<\/td><\/tr>\n<\/tbody><\/table><\/div><div class=\"chromatogram-conditions\"><h4>Conditions<\/h4><div class=\"conditions-container container-fluid\"><div class=\"row\"><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Column<\/th><td>Rtx-BAC1, 30 m, 0.32 mm ID, 1.8 \u00b5m (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/18003\" rel=\"noopener\">cat.# 18003<\/a>)<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Standard\/Sample<\/th><td>BAC resolution control standard n-P (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/36010\" rel=\"noopener\">cat.# 36010<\/a>)<\/td><\/tr>\n<tr><th class=\"sub conditions_header\" scope=\"row\">Diluent:<\/th><td>Water<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Conc.:<\/th><td>200 \u00b5g\/mL<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Injection<\/th><td>headspace-loop split (split ratio 50:1)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Liner:<\/th><td>Topaz 1.0 mm ID straight inlet liner (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/23333\" rel=\"noopener\">cat.# 23333<\/a>)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Port Temp.:<\/th><td>220 \u00b0C<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Headspace-Loop<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Instrument:<\/th><td>Tekmar HT3 <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Time:<\/th><td>0.5  min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Transfer Line Temp.:<\/th><td>200 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Valve Oven Temp.:<\/th><td>200 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Standby flow rate:<\/th><td>100  mL\/min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Sample Temp.:<\/th><td>60 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Platen temp equil. time:<\/th><td>60  min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Sample Equil. Time:<\/th><td>5  min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Vial Pressure:<\/th><td>30 psi<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Pressurize Time:<\/th><td>2  min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Pressure Equilibration Time:<\/th><td>0.2  min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Loop Pressure:<\/th><td>20 psi<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Loop Fill Time:<\/th><td>1  min<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\" colspan=\"2\">Oven<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Oven Temp.:<\/th><td>40 &#176;C (hold 4 min)<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Carrier Gas<\/th><td>He, constant flow<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Linear Velocity:<\/th><td>80 cm\/sec<\/td><\/tr><\/table><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Detector<\/th><td>FID @ 240 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Make-up Gas Flow Rate:<\/th><td>45 mL\/min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Make-up Gas Type:<\/th><td>N<sub>2<\/sub> <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Hydrogen flow:<\/th><td>40 mL\/min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Air flow:<\/th><td>400 mL\/min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Data Rate:<\/th><td>50 Hz<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Instrument<\/th><td>Agilent 7890A GC<\/td><\/tr><\/table><\/div><\/div><\/div><div class=\"chromatogram-related-skus\"><h4>Related SKUs<\/h4><ul><li>36010<\/li><li>18003<\/li><li>23333<\/li><\/ul><\/div><div class=\"chromatogram-related-searches\"><h4>Related Searches<\/h4><ul><li>blood alcohol<\/li><\/ul><\/div><div class=\"chromatogram-pdf-link\"><a href=\"https:\/\/ez.restek.com\/images\/cgram\/gc_cf1172.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"><svg xmlns=\"http:\/\/www.w3.org\/2000\/svg\" width=\"18\" height=\"18\" viewBox=\"0 0 18 18\"><g data-name=\"Group 2996\"><path data-name=\"Rectangle 1246\" d=\"M0 0h18v18H0z\" style=\"fill: none;\"><\/path><\/g><g data-name=\"Group 2997\"><path data-name=\"Path 729\" d=\"M13.412 11.4v2.017H5.345V11.4H4v2.017a1.349 1.349 0 0 0 1.345 1.345h8.068a1.349 1.349 0 0 0 1.345-1.345V11.4zm-.672-2.694-.948-.948-1.741 1.735V4H8.706v5.493L6.965 7.758l-.948.948 3.361 3.361z\" transform=\"translate(-.437 -.414)\" style=\"fill: rgb(13, 123, 196);\"><\/path><\/g><\/svg>Download PDF<\/a><\/div><\/div><\/div>\n\n\n<p class=\"has-medium-font-size wp-block-paragraph\">BAC Resolution Control Standard with&nbsp;<em>n<\/em>-Propanol on Rtx-BAC2 Column<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article\"><div class=\"wp-block-custom-chromatogram-article\"><div class=\"chromatogram-image regular-image\"><img decoding=\"async\" src=\"https:\/\/ez.restek.com\/images\/cgram\/gc_cf1174.png\" alt=\"BAC Resolution Control Standard n-Propanol on Rtx-BAC2\" title=\"-\"><\/div><p class=\"article-id\" style=\"text-align:center\"> GC_CF1174<\/p><div class=\"chromatogram-peaks\"><h4>Peaks<\/h4><table class=\"peaks col-lg-6 col-12\">\n<thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 75px\">t<sub>R<\/sub> (min)<\/th><th style=\"text-align: center;width: 75px\">Conc.<br \/>(\u00b5g\/mL)<\/th><\/tr><\/thead>\n<tbody><tr><td class=\"num\">1.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Acetaldehyde\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/75-07-0\/Acetaldehyde\" target=\"_blank\" rel=\"noopener\">Acetaldehyde<\/a><\/td><td class=\"oth\">1.011<\/td><td class=\"oth\">200<\/td><\/tr>\n<tr><td class=\"num\">2.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Methanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/67-56-1\/Methanol\" target=\"_blank\" rel=\"noopener\">Methanol<\/a><\/td><td class=\"oth\">1.199<\/td><td class=\"oth\">200<\/td><\/tr>\n<tr><td class=\"num\">3.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Ethanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/64-17-5\/Ethanol\" target=\"_blank\" rel=\"noopener\">Ethanol<\/a><\/td><td class=\"oth\">1.623<\/td><td class=\"oth\">200<\/td><\/tr>\n<tr><td class=\"num\">4.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Acetone\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/67-64-1\/Acetone\" target=\"_blank\" rel=\"noopener\">Acetone<\/a><\/td><td class=\"oth\">1.758<\/td><td class=\"oth\">200<\/td><\/tr>\n<tr><td class=\"num\">5.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2-Propanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/67-63-0\/2-Propanol\" target=\"_blank\" rel=\"noopener\">2-Propanol<\/a><\/td><td class=\"oth\">1.929<\/td><td class=\"oth\">200<\/td><\/tr>\n<tr><td class=\"num\">6.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for n-Propanol (IS)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/71-23-8\/n-Propanol\" target=\"_blank\" rel=\"noopener\">n-Propanol (IS)<\/a><\/td><td class=\"oth\">3.165<\/td><td class=\"oth\">200<\/td><\/tr>\n<\/tbody><\/table><\/div><div class=\"chromatogram-conditions\"><h4>Conditions<\/h4><div class=\"conditions-container container-fluid\"><div class=\"row\"><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Column<\/th><td>Rtx-BAC2, 30 m, 0.32 mm ID, 1.2 \u00b5m (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/18002\" rel=\"noopener\">cat.# 18002<\/a>)<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Standard\/Sample<\/th><td>BAC resolution control standard n-P (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/36010\" rel=\"noopener\">cat.# 36010<\/a>)<\/td><\/tr>\n<tr><th class=\"sub conditions_header\" scope=\"row\">Diluent:<\/th><td>Water<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Conc.:<\/th><td>200 \u00b5g\/mL<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Injection<\/th><td>headspace-loop split (split ratio 50:1)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Liner:<\/th><td>Topaz 1.0 mm ID straight inlet liner (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/23333\" rel=\"noopener\">cat.# 23333<\/a>)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Port Temp.:<\/th><td>220 \u00b0C<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Headspace-Loop<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Instrument:<\/th><td>Tekmar HT3 <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Time:<\/th><td>0.5  min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Transfer Line Temp.:<\/th><td>200 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Valve Oven Temp.:<\/th><td>200 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Standby flow rate:<\/th><td>100  mL\/min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Sample Temp.:<\/th><td>60 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Platen temp equil. time:<\/th><td>1  min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Sample Equil. Time:<\/th><td>5  min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Vial Pressure:<\/th><td>30 psi<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Pressurize Time:<\/th><td>2  min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Pressure Equilibration Time:<\/th><td>0.2  min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Loop Pressure:<\/th><td>20 psi<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Loop Fill Time:<\/th><td>1  min<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\" colspan=\"2\">Oven<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Oven Temp.:<\/th><td>40 &#176;C (hold 4 min)<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Carrier Gas<\/th><td>He, constant flow<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Linear Velocity:<\/th><td>80 cm\/sec<\/td><\/tr><\/table><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Detector<\/th><td>FID @ 240 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Make-up Gas Flow Rate:<\/th><td>45 mL\/min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Make-up Gas Type:<\/th><td>N<sub>2<\/sub> <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Hydrogen flow:<\/th><td>40 mL\/min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Air flow:<\/th><td>400 mL\/min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Data Rate:<\/th><td>50 Hz<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Instrument<\/th><td>Agilent 7890A GC<\/td><\/tr><\/table><\/div><\/div><\/div><div class=\"chromatogram-related-skus\"><h4>Related SKUs<\/h4><ul><li>36010<\/li><li>18002<\/li><li>23333<\/li><\/ul><\/div><div class=\"chromatogram-related-searches\"><h4>Related Searches<\/h4><ul><li>blood alcohol<\/li><\/ul><\/div><div class=\"chromatogram-pdf-link\"><a href=\"https:\/\/ez.restek.com\/images\/cgram\/gc_cf1174.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"><svg xmlns=\"http:\/\/www.w3.org\/2000\/svg\" width=\"18\" height=\"18\" viewBox=\"0 0 18 18\"><g data-name=\"Group 2996\"><path data-name=\"Rectangle 1246\" d=\"M0 0h18v18H0z\" style=\"fill: none;\"><\/path><\/g><g data-name=\"Group 2997\"><path data-name=\"Path 729\" d=\"M13.412 11.4v2.017H5.345V11.4H4v2.017a1.349 1.349 0 0 0 1.345 1.345h8.068a1.349 1.349 0 0 0 1.345-1.345V11.4zm-.672-2.694-.948-.948-1.741 1.735V4H8.706v5.493L6.965 7.758l-.948.948 3.361 3.361z\" transform=\"translate(-.437 -.414)\" style=\"fill: rgb(13, 123, 196);\"><\/path><\/g><\/svg>Download PDF<\/a><\/div><\/div><\/div>\n<\/div><\/div><\/div>\n\n\n<p class=\"wp-block-paragraph\"><\/p>\n\n\n\n<p class=\"wp-block-paragraph\">For labs looking to expand their HS-GC-FID platform beyond routine in vivo\u00a0BAC testing, the Rtx-BAC Plus 1\u00a0and Rtx-BAC Plus 2 column set is a more suitable option. When analyzing postmortem blood alcohol,\u00a0<em>t<\/em>-butanol is often used as the internal standard instead of\u00a0<em>n<\/em>-propanol.\u00a0<em>N<\/em>-propanol can be created by putrefaction, which can alter quantitation. When the internal standard, which exists to stabilize quantitation, is erroneously high, the results will lose accuracy. The Rtx-BAC Plus 1\u00a0and Rtx-BAC Plus 2 columns were designed to fully resolve the alternate internal standard,\u00a0<em>t<\/em>-butanol, making it amenable to a fast, two-minute\u00a0<a href=\"https:\/\/discover.restek.com\/wp-content\/uploads\/CFSS1483.pdf\" target=\"_blank\" data-type=\"link\" data-id=\"https:\/\/discover.restek.com\/wp-content\/uploads\/CFSS1483.pdf\" rel=\"noreferrer noopener\">postmortem BAC analysis<\/a>\u00a0(Figure 3). With this column set, GC-FID workflows can also be expanded to include\u00a0<a href=\"https:\/\/discover.restek.com\/cfar3506\/accelerate-inhalants-of-abuse-method-development-with-pro-ezgc-software\/\" data-type=\"post\" data-id=\"43369\">inhalants<\/a>,\u00a0<a href=\"chromatogram-detail\/GC_CF1163\">glycols<\/a>\u00a0and\u00a0<a href=\"chromatogram-detail\/GC_CF1164\">GHB<\/a>.<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><strong>Figure 3:<\/strong>\u00a0The Rtx-BAC Plus 1 and Rtx-BAC Plus 2 column set was designed to resolve the alternative internal standard,\u00a0<em>t<\/em>-butanol, for postmortem BAC analysis.<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n\n<p class=\"has-medium-font-size wp-block-paragraph\">BAC Resolution Control Standard with&nbsp;<em>t<\/em>-Butanol on Rtx-BAC Plus 1 Column<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article\"><div class=\"wp-block-custom-chromatogram-article\"><div class=\"chromatogram-image regular-image\"><img decoding=\"async\" src=\"https:\/\/ez.restek.com\/images\/cgram\/gc_cf1178.png\" alt=\"BAC Resolution Control Standard t-Butanol on Rtx-BAC Plus1\" title=\"-\"><\/div><p class=\"article-id\" style=\"text-align:center\"> GC_CF1178<\/p><div class=\"chromatogram-peaks\"><h4>Peaks<\/h4><table class=\"peaks col-lg-6 col-12\">\n<thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 75px\">t<sub>R<\/sub> (min)<\/th><th style=\"text-align: center;width: 75px\">Conc.<br \/>(\u00b5g\/mL)<\/th><\/tr><\/thead>\n<tbody><tr><td class=\"num\">1.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Methanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/67-56-1\/Methanol\" target=\"_blank\" rel=\"noopener\">Methanol<\/a><\/td><td class=\"oth\">0.859<\/td><td class=\"oth\">200<\/td><\/tr>\n<tr><td class=\"num\">2.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Acetaldehyde\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/75-07-0\/Acetaldehyde\" target=\"_blank\" rel=\"noopener\">Acetaldehyde<\/a><\/td><td class=\"oth\">0.910<\/td><td class=\"oth\">200<\/td><\/tr>\n<tr><td class=\"num\">3.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Ethanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/64-17-5\/Ethanol\" target=\"_blank\" rel=\"noopener\">Ethanol<\/a><\/td><td class=\"oth\">1.085<\/td><td class=\"oth\">200<\/td><\/tr>\n<tr><td class=\"num\">4.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2-Propanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/67-63-0\/2-Propanol\" target=\"_blank\" rel=\"noopener\">2-Propanol<\/a><\/td><td class=\"oth\">1.328<\/td><td class=\"oth\">200<\/td><\/tr>\n<tr><td class=\"num\">5.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Acetone\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/67-64-1\/Acetone\" target=\"_blank\" rel=\"noopener\">Acetone<\/a><\/td><td class=\"oth\">1.444<\/td><td class=\"oth\">200<\/td><\/tr>\n<tr><td class=\"num\">6.<\/td><td class=\"cmpd\">&lt;a class=&quot;cmpd_link&quot; title=&quot;View compound information for <i>t<\/i>-Butanol (IS)&#8221; title=&#8221;View compound information for <i>t<\/i>-Butanol (IS)&#8221; href=&#8221;https:\/\/ez.restek.com\/compound\/view\/en\/75-65-0\/<i>t-Butanol&#8221;&gt;<i>t<\/i>-Butanol (IS)<\/a><\/td><td class=\"oth\">1.574<\/td><td class=\"oth\">200<\/td><\/tr>\n<\/tbody><\/table><\/div><div class=\"chromatogram-conditions\"><h4>Conditions<\/h4><div class=\"conditions-container container-fluid\"><div class=\"row\"><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Column<\/th><td>Rtx-BAC Plus 1, 30 m, 0.32 mm ID, 1.8 \u00b5m (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/18004\" rel=\"noopener\">cat.# 18004<\/a>)<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Standard\/Sample<\/th><td>BAC resolution control standard t-B (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/36011\" rel=\"noopener\">cat.# 36011<\/a>)<\/td><\/tr>\n<tr><th class=\"sub conditions_header\" scope=\"row\">Diluent:<\/th><td>Water<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Conc.:<\/th><td>200 \u00b5g\/mL<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Injection<\/th><td>headspace-loop split (split ratio 50:1)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Liner:<\/th><td>Topaz 1.0 mm ID straight inlet liner (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/23333\" rel=\"noopener\">cat.# 23333<\/a>)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Port Temp.:<\/th><td>220 \u00b0C<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Headspace-Loop<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Instrument:<\/th><td>Tekmar HT3 <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Time:<\/th><td>0.5  min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Transfer Line Temp.:<\/th><td>200 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Valve Oven Temp.:<\/th><td>200 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Standby flow rate:<\/th><td>100  mL\/min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Sample Temp.:<\/th><td>60 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Platen temp equil. time:<\/th><td>1  min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Sample Equil. Time:<\/th><td>5  min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Vial Pressure:<\/th><td>30 psi<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Pressurize Time:<\/th><td>2  min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Pressure Equilibration Time:<\/th><td>0.2  min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Loop Pressure:<\/th><td>20 psi<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Loop Fill Time:<\/th><td>1  min<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\" colspan=\"2\">Oven<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Oven Temp.:<\/th><td>40 &#176;C (hold 4 min)<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Carrier Gas<\/th><td>He, constant flow<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Linear Velocity:<\/th><td>80 cm\/sec<\/td><\/tr><\/table><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Detector<\/th><td>FID @ 240 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Make-up Gas Flow Rate:<\/th><td>45 mL\/min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Make-up Gas Type:<\/th><td>N<sub>2<\/sub> <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Hydrogen flow:<\/th><td>40 mL\/min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Air flow:<\/th><td>400 mL\/min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Data Rate:<\/th><td>50 Hz<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Instrument<\/th><td>Agilent 7890A GC<\/td><\/tr><\/table><\/div><\/div><\/div><div class=\"chromatogram-related-skus\"><h4>Related SKUs<\/h4><ul><li>36011<\/li><li>18004<\/li><li>23333<\/li><\/ul><\/div><div class=\"chromatogram-related-searches\"><h4>Related Searches<\/h4><ul><li>blood alcohol<\/li><\/ul><\/div><div class=\"chromatogram-pdf-link\"><a href=\"https:\/\/ez.restek.com\/images\/cgram\/gc_cf1178.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"><svg xmlns=\"http:\/\/www.w3.org\/2000\/svg\" width=\"18\" height=\"18\" viewBox=\"0 0 18 18\"><g data-name=\"Group 2996\"><path data-name=\"Rectangle 1246\" d=\"M0 0h18v18H0z\" style=\"fill: none;\"><\/path><\/g><g data-name=\"Group 2997\"><path data-name=\"Path 729\" d=\"M13.412 11.4v2.017H5.345V11.4H4v2.017a1.349 1.349 0 0 0 1.345 1.345h8.068a1.349 1.349 0 0 0 1.345-1.345V11.4zm-.672-2.694-.948-.948-1.741 1.735V4H8.706v5.493L6.965 7.758l-.948.948 3.361 3.361z\" transform=\"translate(-.437 -.414)\" style=\"fill: rgb(13, 123, 196);\"><\/path><\/g><\/svg>Download PDF<\/a><\/div><\/div><\/div>\n\n\n<p class=\"has-medium-font-size wp-block-paragraph\">BAC Resolution Control Standard with&nbsp;<em>t<\/em>-Butanol on Rtx-BAC Plus 2 Column<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article\"><div class=\"wp-block-custom-chromatogram-article\"><div class=\"chromatogram-image regular-image\"><img decoding=\"async\" src=\"https:\/\/ez.restek.com\/images\/cgram\/gc_cf1176.png\" alt=\"BAC Resolution Control Standard t-Butanol on Rtx-BAC Plus 2\" title=\"-\"><\/div><p class=\"article-id\" style=\"text-align:center\"> GC_CF1176<\/p><div class=\"chromatogram-peaks\"><h4>Peaks<\/h4><table class=\"peaks col-lg-6 col-12\">\n<thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 75px\">t<sub>R<\/sub> (min)<\/th><th style=\"text-align: center;width: 75px\">Conc.<br \/>(\u00b5g\/mL)<\/th><\/tr><\/thead>\n<tbody><tr><td class=\"num\">1.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Acetaldehyde\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/75-07-0\/Acetaldehyde\" target=\"_blank\" rel=\"noopener\">Acetaldehyde<\/a><\/td><td class=\"oth\">0.771<\/td><td class=\"oth\">200<\/td><\/tr>\n<tr><td class=\"num\">2.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Methanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/67-56-1\/Methanol\" target=\"_blank\" rel=\"noopener\">Methanol<\/a><\/td><td class=\"oth\">0.818<\/td><td class=\"oth\">200<\/td><\/tr>\n<tr><td class=\"num\">3.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Ethanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/64-17-5\/Ethanol\" target=\"_blank\" rel=\"noopener\">Ethanol<\/a><\/td><td class=\"oth\">0.979<\/td><td class=\"oth\">200<\/td><\/tr>\n<tr><td class=\"num\">4.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Acetone\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/67-64-1\/Acetone\" target=\"_blank\" rel=\"noopener\">Acetone<\/a><\/td><td class=\"oth\">1.060<\/td><td class=\"oth\">200<\/td><\/tr>\n<tr><td class=\"num\">5.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2-Propanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/67-63-0\/2-Propanol\" target=\"_blank\" rel=\"noopener\">2-Propanol<\/a><\/td><td class=\"oth\">1.116<\/td><td class=\"oth\">200<\/td><\/tr>\n<tr><td class=\"num\">6.<\/td><td class=\"cmpd\">&lt;a class=&quot;cmpd_link&quot; title=&quot;View compound information for <i>t<\/i>-Butanol (IS)&#8221; title=&#8221;View compound information for <i>t<\/i>-Butanol (IS)&#8221; href=&#8221;https:\/\/ez.restek.com\/compound\/view\/en\/75-65-0\/<i>t-Butanol&#8221;&gt;<i>t<\/i>-Butanol (IS)<\/a><\/td><td class=\"oth\">1.229<\/td><td class=\"oth\">200<\/td><\/tr>\n<\/tbody><\/table><\/div><div class=\"chromatogram-conditions\"><h4>Conditions<\/h4><div class=\"conditions-container container-fluid\"><div class=\"row\"><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Column<\/th><td>Rtx-BAC Plus 2, 30 m, 0.32 mm ID, 0.6 \u00b5m (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/18006\" rel=\"noopener\">cat.# 18006<\/a>)<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Standard\/Sample<\/th><td>BAC resolution control standard t-B (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/36011\" rel=\"noopener\">cat.# 36011<\/a>)<\/td><\/tr>\n<tr><th class=\"sub conditions_header\" scope=\"row\">Diluent:<\/th><td>Water<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Conc.:<\/th><td>200 \u00b5g\/mL<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Injection<\/th><td>headspace-loop split (split ratio 50:1)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Liner:<\/th><td>Topaz 1.0 mm ID straight inlet liner (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/23333\" rel=\"noopener\">cat.# 23333<\/a>)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Port Temp.:<\/th><td>220 \u00b0C<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Headspace-Loop<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Instrument:<\/th><td>Tekmar HT3 <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Time:<\/th><td>0.5  min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Transfer Line Temp.:<\/th><td>200 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Valve Oven Temp.:<\/th><td>200 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Standby flow rate:<\/th><td>100  mL\/min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Sample Temp.:<\/th><td>60 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Platen temp equil. time:<\/th><td>1  min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Sample Equil. Time:<\/th><td>5  min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Vial Pressure:<\/th><td>30 psi<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Pressurize Time:<\/th><td>2  min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Pressure Equilibration Time:<\/th><td>0.2  min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Loop Pressure:<\/th><td>20 psi<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Loop Fill Time:<\/th><td>1  min<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\" colspan=\"2\">Oven<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Oven Temp.:<\/th><td>40 &#176;C (hold 4 min)<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Carrier Gas<\/th><td>He, constant flow<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Linear Velocity:<\/th><td>80 cm\/sec<\/td><\/tr><\/table><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Detector<\/th><td>FID @ 240 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Make-up Gas Flow Rate:<\/th><td>45 mL\/min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Make-up Gas Type:<\/th><td>N<sub>2<\/sub> <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Hydrogen flow:<\/th><td>40 mL\/min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Air flow:<\/th><td>400 mL\/min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Data Rate:<\/th><td>50 Hz<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Instrument<\/th><td>Agilent 7890A GC<\/td><\/tr><\/table><\/div><\/div><\/div><div class=\"chromatogram-related-skus\"><h4>Related SKUs<\/h4><ul><li>36011<\/li><li>18006<\/li><li>23333<\/li><\/ul><\/div><div class=\"chromatogram-related-searches\"><h4>Related Searches<\/h4><ul><li>blood alcohol<\/li><\/ul><\/div><div class=\"chromatogram-pdf-link\"><a href=\"https:\/\/ez.restek.com\/images\/cgram\/gc_cf1176.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"><svg xmlns=\"http:\/\/www.w3.org\/2000\/svg\" width=\"18\" height=\"18\" viewBox=\"0 0 18 18\"><g data-name=\"Group 2996\"><path data-name=\"Rectangle 1246\" d=\"M0 0h18v18H0z\" style=\"fill: none;\"><\/path><\/g><g data-name=\"Group 2997\"><path data-name=\"Path 729\" d=\"M13.412 11.4v2.017H5.345V11.4H4v2.017a1.349 1.349 0 0 0 1.345 1.345h8.068a1.349 1.349 0 0 0 1.345-1.345V11.4zm-.672-2.694-.948-.948-1.741 1.735V4H8.706v5.493L6.965 7.758l-.948.948 3.361 3.361z\" transform=\"translate(-.437 -.414)\" style=\"fill: rgb(13, 123, 196);\"><\/path><\/g><\/svg>Download PDF<\/a><\/div><\/div><\/div>\n<\/div><\/div><\/div>\n\n\n<div style=\"height:20px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<h2 class=\"wp-block-heading\">Urinary Alcohol Metabolites<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">When ethanol metabolites are being monitored in urine, which is commonly performed in treatment programs, high-throughput LC-MS\/MS methods are employed. This technique is highly sensitive and fast. Since mass spectrometry is being used, analyte identification is not as significant a concern compared to when GC-FID is used.&nbsp;However, matrix interferences are still problematic during chromatographic assays. The presence of ethanol is determined using two urinary metabolites\u2014ethyl glucuronide and ethyl sulfate. These analytes are highly water-soluble and, therefore, can be difficult to retain using traditional reversed-phase columns. Because of the inherent variability of urine samples and the common presence of early eluting matrix components, not retaining the target analytes beyond the zone of matrix interferences can cause ionization suppression, resulting in low sensitivity and problematic quantitation. Restek has developed a novel stationary phase to retain and separate the polar analytes EtG and EtS from common urine matrix interferences. As part of our Raptor column line, the Raptor EtG\/EtS LC column features an SPP particle for improved cycle times and stability. It effectively separates EtG and EtS from potentially interfering matrix, ensuring sensitivity and accurate quantitation, in a&nbsp;<a href=\"https:\/\/discover.restek.com\/cfar3061\/successful-strategies-for-the-analysis-of-etg-and-ets-in-urine-01\/\" data-type=\"post\" data-id=\"43365\">fast, four-minute analysis<\/a>.<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><strong>Figure 4:<\/strong>\u00a0The Raptor EtG\/EtS column has the retention needed to separate ethanol metabolites from common matrix interference.<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n<div class=\"wp-block-custom-chromatogram-article\"><div class=\"wp-block-custom-chromatogram-article\"><div class=\"chromatogram-image wide-image\"><img decoding=\"async\" src=\"https:\/\/ez.restek.com\/images\/cgram\/lc_cf0739.png\" alt=\"EtG\/EtS on Raptor EtG\/EtS: Updated Method Conditions (Mobile Phase and Gradient)\" title=\"-\"><\/div><p class=\"article-id\" style=\"text-align:center\"> LC_CF0739<\/p><div class=\"chromatogram-peaks\"><h4>Peaks<\/h4><table class=\"peaks col-12\">\n<thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 75px\">t<sub>R<\/sub> (min)<\/th><th style=\"text-align: center;width: 75px\">Conc.<br \/>(ng\/mL)<\/th><th style=\"text-align: center;width: 75px\">Precursor Ion<\/th><th style=\"text-align: center;width: 75px\">Product Ion<\/th><th style=\"text-align: center;width: 75px\">Product Ion<\/th><\/tr><\/thead>\n<tbody><tr><td class=\"num\">1.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Ethyl-\u03b2-D-glucuronide-d5 (EtG-d5)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/1135070-98-2\/Ethyl-\u03b2-D-glucuronide-d5\" target=\"_blank\" rel=\"noopener\">Ethyl-\u03b2-D-glucuronide-d5 (EtG-d5)<\/a><\/td><td class=\"oth\">1.21<\/td><td class=\"oth\">200<\/td><td class=\"oth\">226.2<\/td><td class=\"oth\">85.0<\/td><td class=\"oth\">&#8211;<\/td><\/tr>\n<tr><td class=\"num\">2.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Ethyl-\u03b2-D-glucuronide (EtG)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/17685-04-0\/Ethyl-\u03b2-D-glucuronide\" target=\"_blank\" rel=\"noopener\">Ethyl-\u03b2-D-glucuronide (EtG)<\/a><\/td><td class=\"oth\">1.23<\/td><td class=\"oth\">500<\/td><td class=\"oth\">221.2<\/td><td class=\"oth\">75.1<\/td><td class=\"oth\">85.1<\/td><\/tr>\n<tr><td class=\"num\">3.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Ethyl sulfate-d5 (EtS-d5)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/1329611-05-3\/Ethyl sulfate-d5\" target=\"_blank\" rel=\"noopener\">Ethyl sulfate-d5 (EtS-d5)<\/a><\/td><td class=\"oth\">2.36<\/td><td class=\"oth\">50<\/td><td class=\"oth\">130.1<\/td><td class=\"oth\">98.0<\/td><td class=\"oth\">&#8211;<\/td><\/tr>\n<tr><td class=\"num\">4.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Ethyl sulfate (EtS)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/64-67-5\/Ethyl sulfate\" target=\"_blank\" rel=\"noopener\">Ethyl sulfate (EtS)<\/a><\/td><td class=\"oth\">2.38<\/td><td class=\"oth\">500<\/td><td class=\"oth\">125.1<\/td><td class=\"oth\">97.1<\/td><td class=\"oth\">80.0<\/td><\/tr>\n<\/tbody><\/table><\/div><div class=\"chromatogram-conditions\"><h4>Conditions<\/h4><div class=\"conditions-container container-fluid\"><div class=\"row\"><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Column<\/th><td>Raptor EtG\/EtS  (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/9325A12\" rel=\"noopener\">cat.# 9325A12<\/a>)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Dimensions:<\/th><td>100 mm x 2.1 mm ID<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Particle Size:<\/th><td>2.7 \u00b5m<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Pore Size:<\/th><td>90 \u00c5<\/td><\/tr><tr><td><\/td><tr><th class=\"sub conditions_header\" scope=\"row\">Guard Column:<\/th><td>UltraShield UHPLC precolumn filter, 0.2 \u03bcm frit (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/25809\" rel=\"noopener\">cat.# 25809<\/a>)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Temp.:<\/th><td>35 \u00b0C<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Standard\/Sample<\/th><td><\/td><\/tr><tr><td><\/td><td>Ethyl-\u03b2-D-glucuronide-d5 (EtG-d5) (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/34102\" rel=\"noopener\">cat.# 34102<\/a>)<\/td><\/tr><tr><td><\/td><td>Ethyl-\u03b2-D-glucuronide (EtG) (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/34101\" rel=\"noopener\">cat.# 34101<\/a>)<\/td><\/tr><tr><td><\/td><td>Ethyl sulfate-d5 sodium salt (EtS-d5) (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/34104\" rel=\"noopener\">cat.# 34104<\/a>)<\/td><\/tr><tr><td><\/td><td>Ethyl sulfate sodium salt (EtS) (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/34103\" rel=\"noopener\">cat.# 34103<\/a>)<\/td><\/tr>\n<tr><th class=\"sub conditions_header\" scope=\"row\">Diluent:<\/th><td>0.01% Formic acid in water<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Conc.:<\/th><td>500 ng\/mL (40x acetonitrile precipitation sample prep)<\/td><\/tr><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Vol.:<\/th><td>10 \u00b5L <\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Mobile Phase<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">A:<\/th><td>0.01% Formic acid in water <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">B:<\/th><td>0.1% Formic acid in acetonitrile <\/td><\/tr><tr><td><\/td><td><table class=\"cgram_ramp\"><thead><tr><th>Time (min)<\/th><th>Flow (mL\/min)<\/th><th>%A<\/th><th>%B<\/th><\/tr><\/thead><tbody><tr><td>0.00<\/td><td>0.5<\/td><td>95<\/td><td>5<\/td><\/tr><tr><td>3.00<\/td><td>0.5<\/td><td>65<\/td><td>35<\/td><\/tr><tr><td>3.01<\/td><td>0.5<\/td><td>95<\/td><td>5<\/td><\/tr><tr><td>4.50<\/td><td>0.5<\/td><td>95<\/td><td>5<\/td><\/tr><\/tbody><\/table><\/td><\/tr><\/table><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Detector<\/th><td>MS\/MS<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Ion Source:<\/th><td>Electrospray <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Ion Mode:<\/th><td>ESI- <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Mode:<\/th><td>MRM <\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Instrument<\/th><td>HPLC<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Sample Preparation<\/th><td>A 500 ng\/mL standard was prepared in urine. A 50 \u03bcL aliquot was mixed with 10 \u03bcL of internal standard (20 \u03bcg\/mL EtG-d5 and 5 \u03bcg\/mL EtS-d5 in water) and 150 \u00b5L of acetonitrile by vortexing at 3000 rpm for 10 seconds and centrifuged at 4300 rpm for 10 minutes at 10 \u00b0C. After centrifugation, 100 \u00b5L of the supernatant was diluted with 900 \u00b5L (40x dilution) of 0.01% formic acid in water. The sample was then vortexed at 3000 rpm for 10 seconds and injected for LC-MS\/MS analysis.<br \/><\/td><\/tr><\/table><\/div><\/div><\/div><div class=\"chromatogram-related-skus\"><h4>Related SKUs<\/h4><ul><li>34102<\/li><li>34101<\/li><li>34104<\/li><li>34103<\/li><li>25809<\/li><li>9325A12<\/li><\/ul><\/div><div class=\"chromatogram-related-searches\"><h4>Related Searches<\/h4><ul><li>EtG<\/li><li>EtS<\/li><li>Sensitivity<\/li><\/ul><\/div><div class=\"chromatogram-pdf-link\"><a href=\"https:\/\/ez.restek.com\/images\/cgram\/lc_cf0739.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"><svg xmlns=\"http:\/\/www.w3.org\/2000\/svg\" width=\"18\" height=\"18\" viewBox=\"0 0 18 18\"><g data-name=\"Group 2996\"><path data-name=\"Rectangle 1246\" d=\"M0 0h18v18H0z\" style=\"fill: none;\"><\/path><\/g><g data-name=\"Group 2997\"><path data-name=\"Path 729\" d=\"M13.412 11.4v2.017H5.345V11.4H4v2.017a1.349 1.349 0 0 0 1.345 1.345h8.068a1.349 1.349 0 0 0 1.345-1.345V11.4zm-.672-2.694-.948-.948-1.741 1.735V4H8.706v5.493L6.965 7.758l-.948.948 3.361 3.361z\" transform=\"translate(-.437 -.414)\" style=\"fill: rgb(13, 123, 196);\"><\/path><\/g><\/svg>Download PDF<\/a><\/div><\/div><\/div>\n<\/div><\/div><\/div>\n\n\n<p class=\"wp-block-paragraph\"><\/p>\n\n\n\n<div style=\"height:100px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n","protected":false},"excerpt":{"rendered":"<p>Compounds found in biological matrices can interfere with the identification and quantitation of ethanol and its associated analytes. However, tailored columns that are designed specifically to lessen interferences can produce productive chromatography with these matrices.<\/p>\n","protected":false},"author":11,"featured_media":0,"comment_status":"closed","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"_acf_changed":false,"_kad_blocks_custom_css":"","_kad_blocks_head_custom_js":"","_kad_blocks_body_custom_js":"","_kad_blocks_footer_custom_js":"","_kadence_starter_templates_imported_post":false,"_kad_post_transparent":"","_kad_post_title":"","_kad_post_layout":"","_kad_post_sidebar_id":"","_kad_post_content_style":"","_kad_post_vertical_padding":"","_kad_post_feature":"","_kad_post_feature_position":"","_kad_post_header":false,"_kad_post_footer":false,"footnotes":"","_ppma_block_editor_authors":""},"categories":[462],"tags":[],"industries-application":[2177,2194],"post-badge":[],"resource-type":[],"product-library":[2408,2420],"resource-technique":[2307,2308,2310,2361],"ppma_author":[414],"class_list":["post-43370","post","type-post","status-publish","format-standard","hentry","category-articles-fr","industries-application-clinical-zh","industries-application-forensics-zh","product-library-gas-chromatography-products-zh","product-library-gc-columns-zh","resource-technique-gas-chromatography-gc-zh","resource-technique-headspace-zh","resource-technique-liquid-chromatography-zh","resource-technique-ms-ms-zh"],"acf":[],"taxonomy_info":{"category":[{"value":462,"label":"Articles"}],"industries-application":[{"value":2177,"label":"\u4e34\u5e8a"},{"value":2194,"label":"\u53f8\u6cd5\u9274\u5b9a"}],"product-library":[{"value":2408,"label":"\u6c14\u76f8\u8272\u8c31\u76f8\u5173\u4ea7\u54c1"},{"value":2420,"label":"GC\u8272\u8c31\u67f1"}],"resource-technique":[{"value":2307,"label":"\u6c14\u76f8\u8272\u8c31\uff08GC\uff09"},{"value":2308,"label":"\u9876\u7a7a"},{"value":2310,"label":"\u6db2\u76f8\u8272\u8c31"},{"value":2361,"label":"MS\/MS"}]},"featured_image_src_large":false,"author_info":{"display_name":"Restek Corporation","author_link":"https:\/\/discover.restek.com\/ja\/author\/restek-corporation\/"},"comment_info":0,"category_info":[{"term_id":462,"name":"Articles","slug":"articles-fr","term_group":0,"term_taxonomy_id":462,"taxonomy":"category","description":"","parent":0,"count":510,"filter":"raw","cat_ID":462,"category_count":510,"category_description":"","cat_name":"Articles","category_nicename":"articles-fr","category_parent":0}],"tag_info":false,"authors":[{"term_id":414,"user_id":11,"is_guest":0,"slug":"restek-corporation","display_name":"Restek Corporation","avatar_url":{"url":"https:\/\/discover.restek.com\/wp-content\/uploads\/Restek_Favicon_300x300.jpg","url2x":"https:\/\/discover.restek.com\/wp-content\/uploads\/Restek_Favicon_300x300.jpg"},"author_category":"1","first_name":"Restek","last_name":"Corporation","user_url":"https:\/\/www.linkedin.com\/company\/restek-corporation\/posts\/?feedView=all","job_title":"Chromatography Products and Solutions","description":"Restek is a leading provider of chromatography columns, accessories, and certified reference materials. Trust Restek for reliable, high-quality analytical solutions."}],"_links":{"self":[{"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/posts\/43370","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/users\/11"}],"replies":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/comments?post=43370"}],"version-history":[{"count":7,"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/posts\/43370\/revisions"}],"predecessor-version":[{"id":79372,"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/posts\/43370\/revisions\/79372"}],"wp:attachment":[{"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/media?parent=43370"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/categories?post=43370"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/tags?post=43370"},{"taxonomy":"industries-application","embeddable":true,"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/industries-application?post=43370"},{"taxonomy":"post-badge","embeddable":true,"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/post-badge?post=43370"},{"taxonomy":"resource-type","embeddable":true,"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/resource-type?post=43370"},{"taxonomy":"product-library","embeddable":true,"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/product-library?post=43370"},{"taxonomy":"resource-technique","embeddable":true,"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/resource-technique?post=43370"},{"taxonomy":"author","embeddable":true,"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/ppma_author?post=43370"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}