{"id":43475,"date":"2020-10-26T14:30:00","date_gmt":"2020-10-26T14:30:00","guid":{"rendered":"https:\/\/discover.restek.com\/uncategorized\/method-8081b-organochlorine-pesticides-analysis-by-gc-extended\/"},"modified":"2026-01-02T18:30:10","modified_gmt":"2026-01-02T18:30:10","slug":"method-8081b-organochlorine-pesticides-analysis-by-gc-extended","status":"publish","type":"post","link":"https:\/\/discover.restek.com\/ja\/articles\/evar4500\/method-8081b-organochlorine-pesticides-analysis-by-gc-extended","title":{"rendered":"Method 8081B: Organochlorine Pesticides Analysis by GC (Extended)"},"content":{"rendered":"\n<p class=\"wp-block-paragraph\">Many labs extend the list of compounds tested using Method 8081B to include other organochlorine pesticides.&nbsp;Rtx-CLPesticides and Rtx-CLPesticides2 columns&nbsp;are ideal for these determinations and can be used in a parallel dual-column configuration to reliably quantify 38 commonly analyzed organochlorine pesticides, as well as many other compounds. Full separations are achieved between both columns and the unique, optimized selectivity allows additional compounds to be included. (Note that since extended compound lists vary among laboratories the lists compared here are not identical.)<\/p>\n\n\n\n<figure class=\"wp-block-table\"><table class=\"has-fixed-layout\"><tbody><tr><th class=\"has-text-align-center\" data-align=\"center\">Vendor<\/th><th class=\"has-text-align-left\" data-align=\"left\">Column Pair<\/th><th class=\"has-text-align-left\" data-align=\"left\">Analysis Time (min)<\/th><th class=\"has-text-align-left\" data-align=\"left\">Coelutions<\/th><th class=\"has-text-align-left\" data-align=\"left\">Restek Advantage<\/th><\/tr><tr><td class=\"has-text-align-center\" data-align=\"center\">Restek<\/td><td class=\"has-text-align-left\" data-align=\"left\"><a href=\"https:\/\/www.restek.com\/p\/11120\" target=\"_blank\" rel=\"noopener\">Rtx-CLPesticides<\/a>&nbsp;\/<br><a href=\"https:\/\/www.restek.com\/p\/11321\" target=\"_blank\" rel=\"noopener\">Rtx-CLPesticides2<\/a><\/td><td class=\"has-text-align-left\" data-align=\"left\">24 \/ 23<\/td><td class=\"has-text-align-left\" data-align=\"left\">1 \/ 2<\/td><td class=\"has-text-align-left\" data-align=\"left\" rowspan=\"3\">Best balance of speed and selectivity. All compounds are resolved on at least one column.<\/td><\/tr><tr><th class=\"has-text-align-center\" data-align=\"center\">Competitor A<\/th><td class=\"has-text-align-left\" data-align=\"left\">column 1 \/ column 2<\/td><td class=\"has-text-align-left\" data-align=\"left\">27 \/ 29<\/td><td class=\"has-text-align-left\" data-align=\"left\">0 \/ 3<\/td><\/tr><tr><th class=\"has-text-align-center\" data-align=\"center\">Competitor B<\/th><td class=\"has-text-align-left\" data-align=\"left\">column 1 \/ column 2<\/td><td class=\"has-text-align-left\" data-align=\"left\">NDP \/ 16<\/td><td class=\"has-text-align-left\" data-align=\"left\">NDP \/ 3<\/td><\/tr><\/tbody><\/table><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\">Comparison based on published competitor data. All columns tested were 0.32 mm ID. NDP = no data published<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><strong>The unique selectivity of the Rtx-CLPesticides column pair allows easy separation of extended lists of organochlorine pesticides.<\/strong><\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n<div class=\"wp-block-custom-chromatogram-article\"><div class=\"wp-block-custom-chromatogram-article\"><div class=\"chromatogram-image wide-image\"><img decoding=\"async\" src=\"https:\/\/ez.restek.com\/images\/cgram\/gc_ev1193.png\" alt=\"Organochlorine Pesticides by EPA Method 8081B on Rtx-CLPesticides\/Rtx-CLPesticides2 (extended list)\" title=\"-\"><\/div><p class=\"article-id\" style=\"text-align:center\"> GC_EV1193<\/p><div class=\"chromatogram-peaks\"><h4>Peaks<\/h4><table class=\"peaks col-lg-4 col-12 peak-33\">\n<thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 50px\">Conc.<br \/>(mg\/mL)<\/th><\/tr><\/thead>\n<tbody><tr><td class=\"num\">1.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 1,2-Dibromo-3-chloropropane\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/96-12-8\/1,2-Dibromo-3-chloropropane\" target=\"_blank\" rel=\"noopener\">1,2-Dibromo-3-chloropropane<\/a><\/td><td class=\"oth\">100<\/td><\/tr>\n<tr><td class=\"num\">2.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Hexachloro-cyclopentadiene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/77-47-4\/Hexachloro-cyclopentadiene\" target=\"_blank\" rel=\"noopener\">Hexachloro-cyclopentadiene<\/a><\/td><td class=\"oth\">100<\/td><\/tr>\n<tr><td class=\"num\">3.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Tetrachloro-m-xylene (SS)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/877-09-8\/Tetrachloro-m-xylene\" target=\"_blank\" rel=\"noopener\">Tetrachloro-m-xylene (SS)<\/a><\/td><td class=\"oth\">100<\/td><\/tr>\n<tr><td class=\"num\">4.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for cis-Diallate\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/2303-16-4\/cis-Diallate\" target=\"_blank\" rel=\"noopener\">cis-Diallate<\/a><\/td><td class=\"oth\">500<\/td><\/tr>\n<tr><td class=\"num\">5.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Hexachlorobenzene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/118-74-1\/Hexachlorobenzene\" target=\"_blank\" rel=\"noopener\">Hexachlorobenzene<\/a><\/td><td class=\"oth\">100<\/td><\/tr>\n<tr><td class=\"num\">6.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for trans-Diallate\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/17708-58-6\/trans-Diallate\" target=\"_blank\" rel=\"noopener\">trans-Diallate<\/a><\/td><td class=\"oth\">500<\/td><\/tr>\n<tr><td class=\"num\">7.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for \u03b1-BHC\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/319-84-6\/\u03b1-BHC\" target=\"_blank\" rel=\"noopener\">\u03b1-BHC<\/a><\/td><td class=\"oth\">40<\/td><\/tr>\n<tr><td class=\"num\">8.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for \u03b3-BHC\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/58-89-9\/\u03b3-BHC\" target=\"_blank\" rel=\"noopener\">\u03b3-BHC<\/a><\/td><td class=\"oth\">40<\/td><\/tr>\n<tr><td class=\"num\">9.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for \u03b2-BHC\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/319-85-7\/\u03b2-BHC\" target=\"_blank\" rel=\"noopener\">\u03b2-BHC<\/a><\/td><td class=\"oth\">40<\/td><\/tr>\n<tr><td class=\"num\">10.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for \u03b4-BHC\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/319-86-8\/\u03b4-BHC\" target=\"_blank\" rel=\"noopener\">\u03b4-BHC<\/a><\/td><td class=\"oth\">40<\/td><\/tr>\n<tr><td class=\"num\">11.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Heptachlor\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/76-44-8\/Heptachlor\" target=\"_blank\" rel=\"noopener\">Heptachlor<\/a><\/td><td class=\"oth\">40<\/td><\/tr>\n<tr><td class=\"num\">12.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Aldrin\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/309-00-2\/Aldrin\" target=\"_blank\" rel=\"noopener\">Aldrin<\/a><\/td><td class=\"oth\">40<\/td><\/tr>\n<\/tbody><\/table>\n<table class=\"peaks col-lg-4 col-12 peak-33\"><thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 50px\">Conc.<br \/>(mg\/mL)<\/th><\/tr><\/thead><tbody>\n<tr><td class=\"num\">13.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Chlorpyrifos\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/2921-88-2\/Chlorpyrifos\" target=\"_blank\" rel=\"noopener\">Chlorpyrifos<\/a><\/td><td class=\"oth\">100<\/td><\/tr>\n<tr><td class=\"num\">14.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Isodrin\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/465-73-6\/Isodrin\" target=\"_blank\" rel=\"noopener\">Isodrin<\/a><\/td><td class=\"oth\">100<\/td><\/tr>\n<tr><td class=\"num\">15.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Oxychlordane\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/27304-13-8\/Oxychlordane\" target=\"_blank\" rel=\"noopener\">Oxychlordane<\/a><\/td><td class=\"oth\">50<\/td><\/tr>\n<tr><td class=\"num\">16.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Heptachlor epoxide\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/1024-57-3\/Heptachlor epoxide\" target=\"_blank\" rel=\"noopener\">Heptachlor epoxide<\/a><\/td><td class=\"oth\">40<\/td><\/tr>\n<tr><td class=\"num\">17.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for trans-Chlordane\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/5103-74-2\/trans-Chlordane\" target=\"_blank\" rel=\"noopener\">trans-Chlordane<\/a><\/td><td class=\"oth\">40<\/td><\/tr>\n<tr><td class=\"num\">18.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2,4&amp;apos;-DDE\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/3424-82-6\/2,4&amp;apos;-DDE\" target=\"_blank\" rel=\"noopener\">2,4&#8242;-DDE<\/a><\/td><td class=\"oth\">100<\/td><\/tr>\n<tr><td class=\"num\">19.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for trans-Nonachlor\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/39765-80-5\/trans-Nonachlor\" target=\"_blank\" rel=\"noopener\">trans-Nonachlor<\/a><\/td><td class=\"oth\">100<\/td><\/tr>\n<tr><td class=\"num\">20.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for cis-Chlordane\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/5103-71-9\/cis-Chlordane\" target=\"_blank\" rel=\"noopener\">cis-Chlordane<\/a><\/td><td class=\"oth\">40<\/td><\/tr>\n<tr><td class=\"num\">21.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Endosulfan I\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/959-98-8\/Endosulfan I\" target=\"_blank\" rel=\"noopener\">Endosulfan I<\/a><\/td><td class=\"oth\">40<\/td><\/tr>\n<tr><td class=\"num\">22.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Captan\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/133-06-2\/Captan\" target=\"_blank\" rel=\"noopener\">Captan<\/a><\/td><td class=\"oth\">100<\/td><\/tr>\n<tr><td class=\"num\">23.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 4,4&amp;apos;-DDE\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/72-55-9\/4,4&amp;apos;-DDE\" target=\"_blank\" rel=\"noopener\">4,4&#8242;-DDE<\/a><\/td><td class=\"oth\">100<\/td><\/tr>\n<tr><td class=\"num\">24.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Dieldrin\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/60-57-1\/Dieldrin\" target=\"_blank\" rel=\"noopener\">Dieldrin<\/a><\/td><td class=\"oth\">80<\/td><\/tr>\n<\/tbody><\/table>\n<table class=\"peaks col-lg-4 col-12 peak-33\"><thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 50px\">Conc.<br \/>(mg\/mL)<\/th><\/tr><\/thead><tbody>\n<tr><td class=\"num\">25.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2,4&amp;apos;-DDD\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/53-19-0\/2,4&amp;apos;-DDD\" target=\"_blank\" rel=\"noopener\">2,4&#8242;-DDD<\/a><\/td><td class=\"oth\">100<\/td><\/tr>\n<tr><td class=\"num\">26.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Endrin\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/72-20-8\/Endrin\" target=\"_blank\" rel=\"noopener\">Endrin<\/a><\/td><td class=\"oth\">80<\/td><\/tr>\n<tr><td class=\"num\">27.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Chlorobenzilate\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/510-15-6\/Chlorobenzilate\" target=\"_blank\" rel=\"noopener\">Chlorobenzilate<\/a><\/td><td class=\"oth\">500<\/td><\/tr>\n<tr><td class=\"num\">28.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2,4&amp;apos;-DDT\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/789-02-6\/2,4&amp;apos;-DDT\" target=\"_blank\" rel=\"noopener\">2,4&#8242;-DDT<\/a><\/td><td class=\"oth\">100<\/td><\/tr>\n<tr><td class=\"num\">29.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for cis-Nonachlor\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/5103-73-1\/cis-Nonachlor\" target=\"_blank\" rel=\"noopener\">cis-Nonachlor<\/a><\/td><td class=\"oth\">100<\/td><\/tr>\n<tr><td class=\"num\">30.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 4,4&amp;apos;-DDD\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/72-54-8\/4,4&amp;apos;-DDD\" target=\"_blank\" rel=\"noopener\">4,4&#8242;-DDD<\/a><\/td><td class=\"oth\">80<\/td><\/tr>\n<tr><td class=\"num\">31.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Endosulfan II\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/33213-65-9\/Endosulfan II\" target=\"_blank\" rel=\"noopener\">Endosulfan II<\/a><\/td><td class=\"oth\">80<\/td><\/tr>\n<tr><td class=\"num\">32.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 4,4&amp;apos;-DDT\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/50-29-3\/4,4&amp;apos;-DDT\" target=\"_blank\" rel=\"noopener\">4,4&#8242;-DDT<\/a><\/td><td class=\"oth\">80<\/td><\/tr>\n<tr><td class=\"num\">33.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Endrin aldehyde\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/7421-93-4\/Endrin aldehyde\" target=\"_blank\" rel=\"noopener\">Endrin aldehyde<\/a><\/td><td class=\"oth\">80<\/td><\/tr>\n<tr><td class=\"num\">34.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Endosulfan sulfate\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/1031-07-8\/Endosulfan sulfate\" target=\"_blank\" rel=\"noopener\">Endosulfan sulfate<\/a><\/td><td class=\"oth\">80<\/td><\/tr>\n<tr><td class=\"num\">35.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Methoxychlor\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/72-43-5\/Methoxychlor\" target=\"_blank\" rel=\"noopener\">Methoxychlor<\/a><\/td><td class=\"oth\">400<\/td><\/tr>\n<tr><td class=\"num\">36.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Mirex\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/2385-85-5\/Mirex\" target=\"_blank\" rel=\"noopener\">Mirex<\/a><\/td><td class=\"oth\">100<\/td><\/tr>\n<tr><td class=\"num\">37.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Endrin ketone\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/53494-70-5\/Endrin ketone\" target=\"_blank\" rel=\"noopener\">Endrin ketone<\/a><\/td><td class=\"oth\">80<\/td><\/tr>\n<tr><td class=\"num\">38.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Decachloro-biphenyl\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/2051-24-3\/Decachloro-biphenyl\" target=\"_blank\" rel=\"noopener\">Decachloro-biphenyl<\/a><\/td><td class=\"oth\">100<\/td><\/tr>\n<\/tbody><\/table><\/div><div class=\"chromatogram-conditions\"><h4>Conditions<\/h4><div class=\"conditions-container container-fluid\"><div class=\"row\"><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Columns<\/th><td>Rtx-CLPesticides2 30 m, 0.32 mm ID, 0.25 \u00b5m (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/11324\" rel=\"noopener\">cat.# 11324<\/a>)<tr><td><\/td><td>and Rtx-CLPesticides 30 m, 0.32 mm ID, 0.32 \u00b5m (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/11141\" rel=\"noopener\">cat.# 11141<\/a>)<tr><td><\/td><td>using Rxi guard column 5 m, 0.32 mm ID (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/10039\" rel=\"noopener\">cat.# 10039<\/a>)<tr><td><\/td><td>with deactivated universal \u201cY\u201d Press-Tight connector (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/20405-261\" rel=\"noopener\">cat.# 20405-261<\/a>)<tr><td><\/td><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Standard\/Sample<\/th><td><\/td><\/tr><tr><td><\/td><td>Organochlorine pesticide mix AB #2 (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/32292\" rel=\"noopener\">cat.# 32292<\/a>)<\/td><\/tr><tr><td><\/td><td>Organochlorine pesticide mix C #1 (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/32296\" rel=\"noopener\">cat.# 32296<\/a>)<\/td><\/tr><tr><td><\/td><td>2,4,5,6-Tetrachloro-m-xylene (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/32027\" rel=\"noopener\">cat.# 32027<\/a>)<\/td><\/tr><tr><td><\/td><td>Decachlorobiphenyl (BZ #209) (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/32029\" rel=\"noopener\">cat.# 32029<\/a>)<\/td><\/tr><tr><td><\/td><td>Chlorpyrifos (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/32212\" rel=\"noopener\">cat.# 32212<\/a>)<\/td><\/tr><tr><td><\/td><td>2,4&#8242;-DDE (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/32099\" rel=\"noopener\">cat.# 32099<\/a>)<\/td><\/tr><tr><td><\/td><td>2,4&#8242;-DDD (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/32098\" rel=\"noopener\">cat.# 32098<\/a>)<\/td><\/tr><tr><td><\/td><td>Chlorobenzilate<\/td><\/tr><tr><td><\/td><td>Captan (custom)<\/td><\/tr><tr><td><\/td><td>cis-Nonachlor (custom)<\/td><\/tr><tr><td><\/td><td>Mirex (custom)<\/td><\/tr><tr><td><\/td><td>Oxychlordone (custom)<\/td><\/tr><tr><td><\/td><td>Diallate (custom)<\/td><\/tr><tr><td><\/td><\/td><\/tr>\n<tr><th class=\"sub conditions_header\" scope=\"row\">Diluent:<\/th><td>Hexane<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Injection<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Vol.:<\/th><td>1 \u00b5L splitless (hold 0.3  min)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Liner:<\/th><td>Cyclo double taper (4 mm) (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/20896\" rel=\"noopener\">cat.# 20896<\/a>)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Temp.:<\/th><td>250 \u00b0C<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\" colspan=\"2\">Oven<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Oven Temp.:<\/th><td>80 &#176;C to 175 &#176;C at 8 &#176;C\/min to 195 &#176;C at 11 &#176;C\/min to 235 &#176;C at 7 &#176;C\/min to 320 &#176;C at 25 &#176;C\/min (hold 3 min)<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Carrier Gas<\/th><td>He, constant flow<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Linear Velocity:<\/th><td>28 cm\/sec<\/td><\/tr><\/table><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Detector<\/th><td>Micro-ECD @ 330 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Make-up Gas Flow Rate:<\/th><td>60 mL\/min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Make-up Gas Type:<\/th><td>N<sub>2<\/sub> <\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Instrument<\/th><td>Agilent\/HP6890  GC<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Notes<\/th><td>This chromatogram was obtained using an Agilent micro-ECD. To obtain comparable results, you will need to employ a micro-ECD in addition to confirmational dual columns connected to a 5-meter guard column using a \u201cY\u201d Press-Tight connector.<br \/><br \/>Reference standard cat.# 32211 was used to produce this chromatogram, but has since been discontinued. For assistance choosing a replacement for this application, contact Restek Technical Service or your local Restek representative.<\/td><\/tr><\/table><\/div><\/div><\/div><div class=\"chromatogram-related-skus\"><h4>Related SKUs<\/h4><ul><li>32292<\/li><li>32296<\/li><li>32027<\/li><li>32029<\/li><li>32212<\/li><li>32099<\/li><li>32098<\/li><li>11324<\/li><li>11141<\/li><li>10039<\/li><li>20405-261<\/li><li>20896<\/li><\/ul><\/div><div class=\"chromatogram-related-searches\"><h4>Related Searches<\/h4><ul><li>8000<\/li><li>8081<\/li><li>BHC<\/li><li>Chlordane<\/li><li>EPA<\/li><\/ul><\/div><div class=\"chromatogram-pdf-link\"><a href=\"https:\/\/ez.restek.com\/images\/cgram\/gc_ev1193.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"><svg xmlns=\"http:\/\/www.w3.org\/2000\/svg\" width=\"18\" height=\"18\" viewBox=\"0 0 18 18\"><g data-name=\"Group 2996\"><path data-name=\"Rectangle 1246\" d=\"M0 0h18v18H0z\" style=\"fill: none;\"><\/path><\/g><g data-name=\"Group 2997\"><path data-name=\"Path 729\" d=\"M13.412 11.4v2.017H5.345V11.4H4v2.017a1.349 1.349 0 0 0 1.345 1.345h8.068a1.349 1.349 0 0 0 1.345-1.345V11.4zm-.672-2.694-.948-.948-1.741 1.735V4H8.706v5.493L6.965 7.758l-.948.948 3.361 3.361z\" transform=\"translate(-.437 -.414)\" style=\"fill: rgb(13, 123, 196);\"><\/path><\/g><\/svg>Download PDF<\/a><\/div><\/div><\/div>\n<\/div><\/div><\/div>\n\n\n<p class=\"wp-block-paragraph\"><\/p>\n\n\n\n<div style=\"height:50px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n        <div class=\"cpb\">\n            <h3 class=\"cpb-heading\">\u3053\u306e\u8a18\u4e8b\u3067\u7d39\u4ecb\u3057\u305f\u88fd\u54c1<\/h3>\n            <hr class=\"cpb-heading-underline\" \/>\n            <div class=\"cpb-list\">\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/11120\">                                \u30ab\u30bf\u30ed\u30b0No. 11120                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Rtx-CLPesticides GC Capillary Column, 15 m, 0.25 mm ID, 0.25 \u00b5m<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/11120\">\u88fd\u54c1\u60c5\u5831\u3092\u898b\u308b<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/11321\">                                \u30ab\u30bf\u30ed\u30b0No. 11321                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Rtx-CLPesticides2 GC Capillary Column, 15 m, 0.32 mm ID, 0.25 \u00b5m<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/11321\">\u88fd\u54c1\u60c5\u5831\u3092\u898b\u308b<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/11196\">                                \u30ab\u30bf\u30ed\u30b0No. 11196                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Rtx-CLPesticides GC Capillary Column Kit, 0.32 mm ID, 0.32 \u00b5m<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/11196\">\u88fd\u54c1\u60c5\u5831\u3092\u898b\u308b<\/a>\n                                                    <\/div>\n                    <\/div>\n                            <\/div>\n        <\/div>\n        \n\n\n<div style=\"height:100px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n","protected":false},"excerpt":{"rendered":"<p>Rtx-CLPesticides and Rtx-CLPesticides2 columns are ideal for multiple environmental ECD methods. Here we demonstrate the effectiveness of this column pair for chlorinated pesticides by EPA Method 8081B (extended).<\/p>\n","protected":false},"author":11,"featured_media":6751,"comment_status":"closed","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"_acf_changed":false,"_kad_blocks_custom_css":"","_kad_blocks_head_custom_js":"","_kad_blocks_body_custom_js":"","_kad_blocks_footer_custom_js":"","_kadence_starter_templates_imported_post":false,"_kad_post_transparent":"","_kad_post_title":"","_kad_post_layout":"","_kad_post_sidebar_id":"","_kad_post_content_style":"","_kad_post_vertical_padding":"","_kad_post_feature":"","_kad_post_feature_position":"","_kad_post_header":false,"_kad_post_footer":false,"footnotes":"","_ppma_block_editor_authors":""},"categories":[7],"tags":[],"industries-application":[2187,2185,2668,2190],"post-badge":[],"resource-type":[],"product-library":[2408,2420,2435,2410],"resource-technique":[2307],"ppma_author":[414],"class_list":["post-43475","post","type-post","status-publish","format-standard","has-post-thumbnail","hentry","category-articles","industries-application-drinking-water-wastewater-testing-zh","industries-application-environmental-zh","industries-application-semivolatiles","industries-application-soil-sediment-biota-testing-zh","product-library-gas-chromatography-products-zh","product-library-gc-columns-zh","product-library-liquid-reference-standards-zh","product-library-reference-standard-products-zh","resource-technique-gas-chromatography-gc-zh"],"acf":[],"taxonomy_info":{"category":[{"value":7,"label":"Articles"}],"industries-application":[{"value":2187,"label":"\u996e\u7528\u6c34&amp;\u5e9f\u6c34\u68c0\u6d4b"},{"value":2185,"label":"\u73af\u5883"},{"value":2668,"label":"Semivolatiles"},{"value":2190,"label":"\u571f\u58e4\u3001\u6c89\u79ef\u7269\u3001\u751f\u7269\u68c0\u6d4b"}],"product-library":[{"value":2408,"label":"\u6c14\u76f8\u8272\u8c31\u76f8\u5173\u4ea7\u54c1"},{"value":2420,"label":"GC\u8272\u8c31\u67f1"},{"value":2435,"label":"\u6db2\u4f53\u53c2\u8003\u6807\u51c6\u54c1"},{"value":2410,"label":"\u53c2\u8003\u6807\u51c6\u54c1"}],"resource-technique":[{"value":2307,"label":"\u6c14\u76f8\u8272\u8c31\uff08GC\uff09"}]},"featured_image_src_large":["https:\/\/discover.restek.com\/wp-content\/uploads\/feature-EVSS1013-1024x536.jpg",1024,536,true],"author_info":{"display_name":"Restek Corporation","author_link":"https:\/\/discover.restek.com\/ja\/author\/restek-corporation\/"},"comment_info":0,"category_info":[{"term_id":7,"name":"Articles","slug":"articles","term_group":0,"term_taxonomy_id":7,"taxonomy":"category","description":"","parent":0,"count":511,"filter":"raw","cat_ID":7,"category_count":511,"category_description":"","cat_name":"Articles","category_nicename":"articles","category_parent":0}],"tag_info":false,"authors":[{"term_id":414,"user_id":11,"is_guest":0,"slug":"restek-corporation","display_name":"Restek Corporation","avatar_url":{"url":"https:\/\/discover.restek.com\/wp-content\/uploads\/Restek_Favicon_300x300.jpg","url2x":"https:\/\/discover.restek.com\/wp-content\/uploads\/Restek_Favicon_300x300.jpg"},"author_category":"1","first_name":"Restek","last_name":"Corporation","user_url":"https:\/\/www.linkedin.com\/company\/restek-corporation\/posts\/?feedView=all","job_title":"Chromatography Products and Solutions","description":"Restek is a leading provider of chromatography columns, accessories, and certified reference materials. Trust Restek for reliable, high-quality analytical solutions."}],"_links":{"self":[{"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/posts\/43475","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/users\/11"}],"replies":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/comments?post=43475"}],"version-history":[{"count":7,"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/posts\/43475\/revisions"}],"predecessor-version":[{"id":79172,"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/posts\/43475\/revisions\/79172"}],"wp:featuredmedia":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/media\/6751"}],"wp:attachment":[{"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/media?parent=43475"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/categories?post=43475"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/tags?post=43475"},{"taxonomy":"industries-application","embeddable":true,"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/industries-application?post=43475"},{"taxonomy":"post-badge","embeddable":true,"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/post-badge?post=43475"},{"taxonomy":"resource-type","embeddable":true,"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/resource-type?post=43475"},{"taxonomy":"product-library","embeddable":true,"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/product-library?post=43475"},{"taxonomy":"resource-technique","embeddable":true,"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/resource-technique?post=43475"},{"taxonomy":"author","embeddable":true,"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/ppma_author?post=43475"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}