{"id":43476,"date":"2020-10-27T00:00:00","date_gmt":"2020-10-27T00:00:00","guid":{"rendered":"https:\/\/discover.restek.com\/uncategorized\/method-8151a-chlorinated-herbicides-analysis-by-gc-using-methylation-or-pentafluorobenzylation-derivatization\/"},"modified":"2026-01-02T18:31:51","modified_gmt":"2026-01-02T18:31:51","slug":"method-8151a-chlorinated-herbicides-analysis-by-gc-using-methylation-or-pentafluorobenzylation-derivatization","status":"publish","type":"post","link":"https:\/\/discover.restek.com\/ja\/articles\/evar4501\/method-8151a-chlorinated-herbicides-analysis-by-gc-using-methylation-or-pentafluorobenzylation-derivatization","title":{"rendered":"Method 8151A: Chlorinated Herbicides Analysis by GC Using Methylation or Pentafluorobenzylation Derivatization"},"content":{"rendered":"\n<p class=\"wp-block-paragraph\">Today, many modern herbicides are designed to selectively kill specific types of vegetation. Even though these herbicides are not environmentally persistent and degrade in less than one year, EPA Method 8151A was developed for chlorinated herbicide analysis in wastewater and soil matrices. The most commonly used compounds targeted in this method are the chlorophenoxy herbicides. Of these herbicides, (2,4-dicholorophenoxy)acetic acid, also known as 2,4-D, is the most widely used and widely monitored herbicide worldwide.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">According to Method 8151A, samples containing the target herbicides are analyzed using a parallel dual-column confirmation setup and an ECD, the recommended method of detection. Because of this, many laboratories acquire chlorinated herbicides on the same instrument setup used for chlorinated pesticide and PCB analyses. The&nbsp;Rtx-CLPesticides and Rtx-CLPesticides2 column pair&nbsp;also performs exceptionally well for chlorinated herbicide analysis; in fact, all 17 target compounds in EPA Method 8151A can be determined in just 13 minutes using these columns. One pair of compounds (bentazon and picloram methyl esters) coelutes on the Rtx-CLPesticides column, but these compounds are fully resolved on the Rtx-CLPesticides2 column. And, due to the difference in selectivity of the columns, several elution order changes occur filling the confirmational requirement of the method.<\/p>\n\n\n\n<figure class=\"wp-block-table\"><table class=\"has-fixed-layout\"><tbody><tr><th class=\"has-text-align-left\" data-align=\"left\">Vendor<\/th><th class=\"has-text-align-left\" data-align=\"left\">Column Pair<\/th><th class=\"has-text-align-left\" data-align=\"left\">Analysis Time (min)<\/th><th class=\"has-text-align-left\" data-align=\"left\">Coelutions<\/th><th class=\"has-text-align-left\" data-align=\"left\">Restek Advantage<\/th><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Restek<\/td><td class=\"has-text-align-left\" data-align=\"left\"><a href=\"https:\/\/www.restek.com\/p\/11120\" target=\"_blank\" rel=\"noopener\">Rtx-CLPesticides<\/a>&nbsp;\/<br><a href=\"https:\/\/www.restek.com\/p\/11321\" target=\"_blank\" rel=\"noopener\">Rtx-CLPesticides2<\/a><\/td><td class=\"has-text-align-left\" data-align=\"left\">13 \/ 13<\/td><td class=\"has-text-align-left\" data-align=\"left\">1 \/ 0<\/td><td class=\"has-text-align-left\" data-align=\"left\" rowspan=\"3\">More elution order changes improve confidence in confirmational results.<\/td><\/tr><tr><th class=\"has-text-align-left\" data-align=\"left\">Competitor A<\/th><td class=\"has-text-align-left\" data-align=\"left\">column 1 \/ column 2<\/td><td class=\"has-text-align-left\" data-align=\"left\">13 \/ 13<\/td><td class=\"has-text-align-left\" data-align=\"left\">0 \/ 0<\/td><\/tr><tr><th class=\"has-text-align-left\" data-align=\"left\">Competitor B<\/th><td class=\"has-text-align-left\" data-align=\"left\">column 1 \/ column 2<\/td><td class=\"has-text-align-left\" data-align=\"left\">16 \/ 15<\/td><td class=\"has-text-align-left\" data-align=\"left\">1 \/ 1<\/td><\/tr><\/tbody><\/table><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\">Comparison based on published competitor data. All columns tested were 0.32 mm ID.<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><strong>Excellent resolution of chlorinated herbicides on the Rtx-CLPesticides\/Rtx-CLPesticides2 column pair<\/strong><\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n<div class=\"wp-block-custom-chromatogram-article\"><div class=\"wp-block-custom-chromatogram-article\"><div class=\"chromatogram-image regular-image\"><img decoding=\"async\" src=\"https:\/\/ez.restek.com\/images\/cgram\/gc_ev00971.png\" alt=\"Herbicides on Rtx-CLPesticides &amp; Rtx-CLPesticides2\" title=\"-\"><\/div><p class=\"article-id\" style=\"text-align:center\"> GC_EV00971<\/p><div class=\"chromatogram-peaks\"><h4>Peaks<\/h4><table class=\"peaks col-lg-6 col-12 peak-50\">\n<thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><\/tr><\/thead>\n<tbody><tr><td class=\"num\">1.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Dalapon methyl ester\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/17640-02-7\/Dalapon methyl ester\" target=\"_blank\" rel=\"noopener\">Dalapon methyl ester<\/a><\/td><\/tr>\n<tr><td class=\"num\">2.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 3,5-Dichlorobenzoic acid methyl ester (SS)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/2905-67-1\/3,5-Dichlorobenzoic acid methyl ester\" target=\"_blank\" rel=\"noopener\">3,5-Dichlorobenzoic acid methyl ester (SS)<\/a><\/td><\/tr>\n<tr><td class=\"num\">3.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 4-Nitroanisole\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/100-17-4\/4-Nitroanisole\" target=\"_blank\" rel=\"noopener\">4-Nitroanisole<\/a><\/td><\/tr>\n<tr><td class=\"num\">4.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for DCAA methyl ester (SS)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/79-43-6\/DCAA methyl ester\" target=\"_blank\" rel=\"noopener\">DCAA methyl ester (SS)<\/a><\/td><\/tr>\n<tr><td class=\"num\">5.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Dicamba methyl ester\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/1918-00-9\/Dicamba methyl ester\" target=\"_blank\" rel=\"noopener\">Dicamba methyl ester<\/a><\/td><\/tr>\n<tr><td class=\"num\">6.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for MCPP methyl ester\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/93-65-2\/MCPP methyl ester\" target=\"_blank\" rel=\"noopener\">MCPP methyl ester<\/a><\/td><\/tr>\n<tr><td class=\"num\">7.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for MCPA Methyl ester\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/94-74-6\/MCPA Methyl ester\" target=\"_blank\" rel=\"noopener\">MCPA Methyl ester<\/a><\/td><\/tr>\n<tr><td class=\"num\">8.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Dichlorprop, methyl ester\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/120-36-5\/Dichlorprop, methyl ester\" target=\"_blank\" rel=\"noopener\">Dichlorprop, methyl ester<\/a><\/td><\/tr>\n<tr><td class=\"num\">9.<\/td><td class=\"cmpd\">4,4&#8242;-DBOB (IS)<\/td><\/tr>\n<tr><td class=\"num\">10.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2,4-D methyl ester\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/94-75-7\/2,4-D methyl ester\" target=\"_blank\" rel=\"noopener\">2,4-D methyl ester<\/a><\/td><\/tr>\n<\/tbody><\/table>\n<table class=\"peaks col-lg-6 col-12 peak-50\"><thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><\/tr><\/thead><tbody>\n<tr><td class=\"num\">11.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Pentachloroanisole\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/1825-21-4\/Pentachloroanisole\" target=\"_blank\" rel=\"noopener\">Pentachloroanisole<\/a><\/td><\/tr>\n<tr><td class=\"num\">12.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2,4,5-TP, methyl ester\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/93-72-1\/2,4,5-TP, methyl ester\" target=\"_blank\" rel=\"noopener\">2,4,5-TP, methyl ester<\/a><\/td><\/tr>\n<tr><td class=\"num\">13.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2,4,5-T methyl ester\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/93-76-5\/2,4,5-T methyl ester\" target=\"_blank\" rel=\"noopener\">2,4,5-T methyl ester<\/a><\/td><\/tr>\n<tr><td class=\"num\">14.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Chloramben, methyl ester\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/133-90-4\/Chloramben, methyl ester\" target=\"_blank\" rel=\"noopener\">Chloramben, methyl ester<\/a><\/td><\/tr>\n<tr><td class=\"num\">15.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2,4-DB methyl ester\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/94-82-6\/2,4-DB methyl ester\" target=\"_blank\" rel=\"noopener\">2,4-DB methyl ester<\/a><\/td><\/tr>\n<tr><td class=\"num\">16.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Dinoseb methyl ester\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/88-85-7\/Dinoseb methyl ester\" target=\"_blank\" rel=\"noopener\">Dinoseb methyl ester<\/a><\/td><\/tr>\n<tr><td class=\"num\">17.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Bentazon methyl ester\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/25057-89-0\/Bentazon methyl ester\" target=\"_blank\" rel=\"noopener\">Bentazon methyl ester<\/a><\/td><\/tr>\n<tr><td class=\"num\">18.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for DCPA methyl ester (Chlorthal-dimethyl)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/1861-32-1\/DCPA methyl ester\" target=\"_blank\" rel=\"noopener\">DCPA methyl ester (Chlorthal-dimethyl)<\/a><\/td><\/tr>\n<tr><td class=\"num\">19.<\/td><td class=\"cmpd\">Pichloram methyl ester<\/td><\/tr>\n<tr><td class=\"num\">20.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Acifluorfen methyl ester\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/50594-66-6\/Acifluorfen methyl ester\" target=\"_blank\" rel=\"noopener\">Acifluorfen methyl ester<\/a><\/td><\/tr>\n<\/tbody><\/table><div class=\"peaknotes\">C. contaminant<\/div><\/div><div class=\"chromatogram-conditions\"><h4>Conditions<\/h4><div class=\"conditions-container container-fluid\"><div class=\"row\"><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Columns<\/th><td>Rtx-CLPesticides2 30 m, 0.32 mm ID, 0.25 \u00b5m (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/11324\" rel=\"noopener\">cat.# 11324<\/a>)<tr><td><\/td><td>and Rtx-CLPesticides 30 m, 0.32 mm ID, 0.32 \u00b5m (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/11141\" rel=\"noopener\">cat.# 11141<\/a>)<tr><td><\/td><td>using Rxi deactivated guard column 5 m, 0.32 mm ID (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/10039\" rel=\"noopener\">cat.# 10039<\/a>)<tr><td><\/td><td>with universal \u201cY\u201d Press-Tight connector (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/20405-261\" rel=\"noopener\">cat.# 20405-261<\/a>)<tr><td><\/td><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Standard\/Sample<\/th><td><\/td><\/tr><tr><td><\/td><td>200 ng\/mL herbicide mix #1 (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/32055\" rel=\"noopener\">cat.# 32055<\/a>)<\/td><\/tr><tr><td><\/td><td>1,000 ng\/mL dalapon methyl ester (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/32057\" rel=\"noopener\">cat.# 32057<\/a>)<\/td><\/tr><tr><td><\/td><td>20,000 ng\/mL herbicide mix #3 (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/32059\" rel=\"noopener\">cat.# 32059<\/a>)<\/td><\/tr><tr><td><\/td><td>200 ng\/mL herbicide mix #4 (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/32062\" rel=\"noopener\">cat.# 32062<\/a>)<\/td><\/tr><tr><td><\/td><td>250 ng\/mL 4,4&#8242;-dibromooctafluorobiphenyl (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/32053\" rel=\"noopener\">cat.# 32053<\/a>)<\/td><\/tr><tr><td><\/td><td>400 ng\/mL 2,4-dichlorophenyl acetic acid methyl ester (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/32050\" rel=\"noopener\">cat.# 32050<\/a>)<\/td><\/tr>\n<tr><th class=\"sub conditions_header\" scope=\"row\">Diluent:<\/th><td>hexane<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Injection<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Vol.:<\/th><td>1.0 \u00b5L splitless (hold 0.75  min)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Liner:<\/th><td>Cyclo double taper (4 mm)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Temp.:<\/th><td>250 \u00b0C<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\" colspan=\"2\">Oven<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Oven Temp.:<\/th><td>70 &#176;C (hold 0.5 min) to 190 &#176;C at 25 &#176;C\/min (hold 1 min) to 300 &#176;C at 11 &#176;C\/min (hold 5 min)<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Carrier Gas<\/th><td>He, constant pressure<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Linear Velocity:<\/th><td>36 cm\/sec @ 70 \u00b0C<\/td><\/tr><\/table><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Detector<\/th><td>Micro_ECD @ 325 \u00b0C<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Instrument<\/th><td>Agilent\/HP6890  GC<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Notes<\/th><td>Liner cat.# 20895 was used to produce this chromatogram, but it has since been discontinued. We recommend cat.# 20896 [or cat.# 23310] as an alternative. For assistance choosing a replacement for this application, contact Restek Technical Service or your local Restek representative.<br \/>&#8211; &#8211; &#8211; &#8211; &#8211; &#8211;<br \/>This chromatogram was obtained using an Agilent micro-ECD. To obtain comparable results, you will need to employ a micro-ECD in addition to confirmational dual columns connected to a 5-meter guard column using a \u201cY\u201d Press-Tight connector.<\/td><\/tr><\/table><\/div><\/div><\/div><div class=\"chromatogram-related-skus\"><h4>Related SKUs<\/h4><ul><li>32055<\/li><li>32057<\/li><li>32059<\/li><li>32062<\/li><li>32053<\/li><li>32050<\/li><li>11324<\/li><li>11141<\/li><li>10039<\/li><li>20405-261<\/li><\/ul><\/div><div class=\"chromatogram-related-searches\"><h4>Related Searches<\/h4><ul><li>8151A<\/li><li>CLP<\/li><li>dalapon<\/li><li>dual column<\/li><li>herbicides<\/li><li>methyl esters<\/li><li>pesticides<\/li><\/ul><\/div><div class=\"chromatogram-pdf-link\"><a href=\"https:\/\/ez.restek.com\/images\/cgram\/gc_ev00971.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"><svg xmlns=\"http:\/\/www.w3.org\/2000\/svg\" width=\"18\" height=\"18\" viewBox=\"0 0 18 18\"><g data-name=\"Group 2996\"><path data-name=\"Rectangle 1246\" d=\"M0 0h18v18H0z\" style=\"fill: none;\"><\/path><\/g><g data-name=\"Group 2997\"><path data-name=\"Path 729\" d=\"M13.412 11.4v2.017H5.345V11.4H4v2.017a1.349 1.349 0 0 0 1.345 1.345h8.068a1.349 1.349 0 0 0 1.345-1.345V11.4zm-.672-2.694-.948-.948-1.741 1.735V4H8.706v5.493L6.965 7.758l-.948.948 3.361 3.361z\" transform=\"translate(-.437 -.414)\" style=\"fill: rgb(13, 123, 196);\"><\/path><\/g><\/svg>Download PDF<\/a><\/div><\/div><\/div>\n<\/div><\/div><\/div>\n\n\n<p class=\"wp-block-paragraph\"><\/p>\n\n\n\n<div style=\"height:50px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n        <div class=\"cpb\">\n            <h3 class=\"cpb-heading\">\u3053\u306e\u8a18\u4e8b\u3067\u7d39\u4ecb\u3057\u305f\u88fd\u54c1<\/h3>\n            <hr class=\"cpb-heading-underline\" \/>\n            <div class=\"cpb-list\">\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/11120\">                                \u30ab\u30bf\u30ed\u30b0No. 11120                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Rtx-CLPesticides GC Capillary Column, 15 m, 0.25 mm ID, 0.25 \u00b5m<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/11120\">\u88fd\u54c1\u60c5\u5831\u3092\u898b\u308b<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/11321\">                                \u30ab\u30bf\u30ed\u30b0No. 11321                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Rtx-CLPesticides2 GC Capillary Column, 15 m, 0.32 mm ID, 0.25 \u00b5m<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/11321\">\u88fd\u54c1\u60c5\u5831\u3092\u898b\u308b<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/11196\">                                \u30ab\u30bf\u30ed\u30b0No. 11196                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Rtx-CLPesticides GC Capillary Column Kit, 0.32 mm ID, 0.32 \u00b5m<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/11196\">\u88fd\u54c1\u60c5\u5831\u3092\u898b\u308b<\/a>\n                                                    <\/div>\n                    <\/div>\n                            <\/div>\n        <\/div>\n        \n\n\n<div style=\"height:100px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n","protected":false},"excerpt":{"rendered":"<p>Rtx-CLPesticides and Rtx-CLPesticides2 columns are ideal for multiple environmental ECD methods. Here we demonstrate the effectiveness of this column pair for chlorinated herbicides by EPA Method 8151A.<\/p>\n","protected":false},"author":11,"featured_media":6751,"comment_status":"closed","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"_acf_changed":false,"_kad_blocks_custom_css":"","_kad_blocks_head_custom_js":"","_kad_blocks_body_custom_js":"","_kad_blocks_footer_custom_js":"","_kadence_starter_templates_imported_post":false,"_kad_post_transparent":"","_kad_post_title":"","_kad_post_layout":"","_kad_post_sidebar_id":"","_kad_post_content_style":"","_kad_post_vertical_padding":"","_kad_post_feature":"","_kad_post_feature_position":"","_kad_post_header":false,"_kad_post_footer":false,"footnotes":"","_ppma_block_editor_authors":""},"categories":[7],"tags":[],"industries-application":[2163,2161,2668,2166],"post-badge":[],"resource-type":[],"product-library":[2372,2384,2399,2374],"resource-technique":[2294],"ppma_author":[414],"class_list":["post-43476","post","type-post","status-publish","format-standard","has-post-thumbnail","hentry","category-articles","industries-application-drinking-water-wastewater-testing","industries-application-environmental","industries-application-semivolatiles","industries-application-soil-sediment-biota-testing","product-library-gas-chromatography-products","product-library-gc-columns","product-library-liquid-reference-standards","product-library-reference-standard-products","resource-technique-gas-chromatography-gc"],"acf":[],"taxonomy_info":{"category":[{"value":7,"label":"Articles"}],"industries-application":[{"value":2163,"label":"Drinking Water &amp; Wastewater Testing"},{"value":2161,"label":"Environmental"},{"value":2668,"label":"Semivolatiles"},{"value":2166,"label":"Soil, Sediment, Biota Testing"}],"product-library":[{"value":2372,"label":"Gas Chromatography Products"},{"value":2384,"label":"GC Columns"},{"value":2399,"label":"Liquid Reference Standards"},{"value":2374,"label":"Reference Standard Products"}],"resource-technique":[{"value":2294,"label":"Gas Chromatography (GC)"}]},"featured_image_src_large":["https:\/\/discover.restek.com\/wp-content\/uploads\/feature-EVSS1013-1024x536.jpg",1024,536,true],"author_info":{"display_name":"Restek Corporation","author_link":"https:\/\/discover.restek.com\/ja\/author\/restek-corporation\/"},"comment_info":0,"category_info":[{"term_id":7,"name":"Articles","slug":"articles","term_group":0,"term_taxonomy_id":7,"taxonomy":"category","description":"","parent":0,"count":511,"filter":"raw","cat_ID":7,"category_count":511,"category_description":"","cat_name":"Articles","category_nicename":"articles","category_parent":0}],"tag_info":false,"authors":[{"term_id":414,"user_id":11,"is_guest":0,"slug":"restek-corporation","display_name":"Restek Corporation","avatar_url":{"url":"https:\/\/discover.restek.com\/wp-content\/uploads\/Restek_Favicon_300x300.jpg","url2x":"https:\/\/discover.restek.com\/wp-content\/uploads\/Restek_Favicon_300x300.jpg"},"author_category":"1","first_name":"Restek","last_name":"Corporation","user_url":"https:\/\/www.linkedin.com\/company\/restek-corporation\/posts\/?feedView=all","job_title":"Chromatography Products and Solutions","description":"Restek is a leading provider of chromatography columns, accessories, and certified reference materials. Trust Restek for reliable, high-quality analytical solutions."}],"_links":{"self":[{"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/posts\/43476","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/users\/11"}],"replies":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/comments?post=43476"}],"version-history":[{"count":9,"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/posts\/43476\/revisions"}],"predecessor-version":[{"id":79174,"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/posts\/43476\/revisions\/79174"}],"wp:featuredmedia":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/media\/6751"}],"wp:attachment":[{"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/media?parent=43476"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/categories?post=43476"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/tags?post=43476"},{"taxonomy":"industries-application","embeddable":true,"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/industries-application?post=43476"},{"taxonomy":"post-badge","embeddable":true,"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/post-badge?post=43476"},{"taxonomy":"resource-type","embeddable":true,"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/resource-type?post=43476"},{"taxonomy":"product-library","embeddable":true,"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/product-library?post=43476"},{"taxonomy":"resource-technique","embeddable":true,"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/resource-technique?post=43476"},{"taxonomy":"author","embeddable":true,"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/ppma_author?post=43476"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}