{"id":43557,"date":"2023-10-24T00:00:00","date_gmt":"2023-10-24T00:00:00","guid":{"rendered":"https:\/\/discover.restek.com\/uncategorized\/best-practices-for-handling-and-using-volatile-analytical-standards\/"},"modified":"2026-01-03T18:05:48","modified_gmt":"2026-01-03T18:05:48","slug":"best-practices-for-handling-and-using-volatile-analytical-standards","status":"publish","type":"post","link":"https:\/\/discover.restek.com\/ja\/articles\/ffar4121\/best-practices-for-handling-and-using-volatile-analytical-standards","title":{"rendered":"Best Practices for Handling and Using Volatile Analytical Standards"},"content":{"rendered":"\n<p class=\"wp-block-paragraph\">This guide is appropriate for standards containing volatile compounds dissolved in solution, such as VOCs and residual solvent gases.&nbsp;<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Ensuring the integrity of your volatile gas reference standards starts with having the proper tools and skillset for best practices in handling and use. Because these standards are very volatile and are more sensitive to temperature and agitation than nonvolatile standards, it is especially important to observe proper care and handling in constructing your calibration solutions. Here, we describe best practice recommendations for use of your volatile gas standards as well as a simplified workflow to ensure you are set up for success in beginning your analyses.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\"><\/p>\n\n\n\n<div class=\"wp-block-columns is-layout-flex wp-container-core-columns-is-layout-5dc627e1 wp-block-columns-is-layout-flex\" style=\"margin-top:0;margin-bottom:0\">\n<div class=\"wp-block-column has-background is-layout-flow wp-block-column-is-layout-flow\" style=\"border-radius:0px;background-color:#e6f0fa;padding-top:20px;padding-right:20px;padding-bottom:20px;padding-left:20px\">\n<p class=\"has-text-align-center has-medium-font-size wp-block-paragraph\"><strong>General Best Practice Recommendations<\/strong><\/p>\n<\/div>\n<\/div>\n\n\n\n<div class=\"wp-block-columns has-background is-layout-flex wp-container-core-columns-is-layout-5b6e2339 wp-block-columns-is-layout-flex\" style=\"border-radius:0px;background-color:#e6f0fa;margin-top:0\">\n<div class=\"wp-block-column is-layout-flow wp-block-column-is-layout-flow\"><style>.kb-image43557_dc03a6-4e .kb-image-has-overlay:after{opacity:0.3;}<\/style>\n<div class=\"wp-block-kadence-image kb-image43557_dc03a6-4e\"><figure class=\"aligncenter size-medium\"><img loading=\"lazy\" decoding=\"async\" width=\"300\" height=\"300\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-FFAR4121-01-300x300.png\" alt=\"\" class=\"kb-img wp-image-19930\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-FFAR4121-01-300x300.png 300w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-FFAR4121-01-150x150.png 150w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-FFAR4121-01.png 500w\" sizes=\"auto, (max-width: 300px) 100vw, 300px\" \/><\/figure><\/div>\n\n\n\n<p class=\"has-text-align-center wp-block-paragraph\">Avoid vigorous mixing<br>and agitation.<\/p>\n<\/div>\n\n\n\n<div class=\"wp-block-column is-layout-flow wp-block-column-is-layout-flow\"><style>.kb-image43557_c014d5-5e .kb-image-has-overlay:after{opacity:0.3;}<\/style>\n<div class=\"wp-block-kadence-image kb-image43557_c014d5-5e\"><figure class=\"aligncenter size-medium\"><img loading=\"lazy\" decoding=\"async\" width=\"300\" height=\"300\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-FFAR4121-02-300x300.png\" alt=\"\" class=\"kb-img wp-image-19936\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-FFAR4121-02-300x300.png 300w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-FFAR4121-02-150x150.png 150w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-FFAR4121-02.png 500w\" sizes=\"auto, (max-width: 300px) 100vw, 300px\" \/><\/figure><\/div>\n\n\n\n<p class=\"has-text-align-center wp-block-paragraph\">Do not heat.<\/p>\n<\/div>\n\n\n\n<div class=\"wp-block-column is-layout-flow wp-block-column-is-layout-flow\"><style>.kb-image43557_007eec-71 .kb-image-has-overlay:after{opacity:0.3;}<\/style>\n<div class=\"wp-block-kadence-image kb-image43557_007eec-71\"><figure class=\"aligncenter size-medium\"><img loading=\"lazy\" decoding=\"async\" width=\"300\" height=\"300\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-FFAR4121-03-300x300.png\" alt=\"\" class=\"kb-img wp-image-19942\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-FFAR4121-03-300x300.png 300w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-FFAR4121-03-150x150.png 150w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-FFAR4121-03.png 500w\" sizes=\"auto, (max-width: 300px) 100vw, 300px\" \/><\/figure><\/div>\n\n\n\n<p class=\"has-text-align-center wp-block-paragraph\">Work efficiently to<br>minimize analyte loss.<\/p>\n<\/div>\n<\/div>\n\n\n\n<div style=\"height:20px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<h2 class=\"wp-block-heading\">Simplified Example Workflow<\/h2>\n\n\n<figure class=\"wp-block-table\">\n<table>\n<tbody>\n<tr>\n<td style=\"background-color: #0083c6; color: #ffffff; border-color: #ffffff;\"><strong>1<\/strong><\/td>\n<td><strong>Store unopened ampuls at the recommended storage conditions until ready for use.<\/strong> Proper storage of volatile standards is especially critical to ensure the integrity of the analytes in solution<\/td>\n<\/tr>\n<tr>\n<td style=\"background-color: #0083c6; color: #ffffff; border-color: #ffffff;\"><strong>2<\/strong><\/td>\n<td><strong>Use the analytical reference standard cold (0 \u00b0C or colder).<\/strong> Do not heat the ampul to accelerate equilibration, as it can lead to volatilization of the compounds.<\/td>\n<\/tr>\n<tr>\n<td style=\"background-color: #0083c6; color: #ffffff; border-color: #ffffff;\"><strong>3<\/strong><\/td>\n<td><strong>Prior to opening the ampul, gather your diluent, glassware, and any other supplies that are needed to prepare your calibration solutions.<\/strong> Once the ampul is open, it\u2019s important to work quickly in transferring and diluting your standards. Having supplies ready to go can help to ensure an efficient workflow.<\/td>\n<\/tr>\n<tr>\n<td style=\"background-color: #0083c6; color: #ffffff; border-color: #ffffff;\"><strong>4<\/strong><\/td>\n<td><strong>Gently invert the unopened ampul several times to mix the contents prior to opening.<\/strong> It\u2019s important to ensure the contents of the ampul are homogenous prior to use, but agitation increases the surface area of the liquid and thus the volatile rate. We recommend gently inverting the ampul to minimize any volatilization of the analytes and then allowing the ampul to sit undisturbed for a couple of minutes before opening.<\/td>\n<\/tr>\n<tr>\n<td style=\"background-color: #0083c6; color: #ffffff; border-color: #ffffff;\"><strong>5<\/strong><\/td>\n<td><strong>Carefully open the ampul with the Restek safe cracker.<\/strong><\/td>\n<\/tr>\n<tr>\n<td style=\"background-color: #0083c6; color: #ffffff; border-color: #ffffff;\"><strong>6<\/strong><\/td>\n<td><strong>Transfer and dilute standards as quickly as possible without compromising technique.<\/strong> We suggest using gas-tight syringes for transfer, which minimizes any potential volatilization.<\/td>\n<\/tr>\n<tr>\n<td style=\"background-color: #0083c6; color: #ffffff; border-color: #ffffff;\"><strong>7<\/strong><\/td>\n<td><strong>After dilution, immediately place caps securely on the vials.<\/strong> Gently invert several times to mix.<\/td>\n<\/tr>\n<\/tbody>\n<\/table>\n<\/figure>\n\n\n<p class=\"wp-block-paragraph\"><strong>Note:<\/strong> To maintain long-term stability, store in micro-vials with Mininert Precision sampling valves at 0 \u00b0C or colder to minimize analyte loss. Refer to our <a href=\"https:\/\/www.restek.com\/articles\/storing-and-handling-volatile-residual-solvent-gases-in-secondary-containers\" data-type=\"link\" data-id=\"https:\/\/www.restek.com\/articles\/storing-and-handling-volatile-residual-solvent-gases-in-secondary-containers\" target=\"_blank\" rel=\"noopener\">&#8220;Storing and Handling Volatile Residual Solvent Gases in Secondary Containers&#8221;<\/a> article to learn more.<\/p>\n\n\n\n<div style=\"height:20px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<h3 class=\"wp-block-heading\">Recommended Tools:<\/h3>\n\n\n\n<ul class=\"wp-block-list\">\n<li>Gas-tight syringes<\/li>\n\n\n\n<li>Restek safe cracker<\/li>\n<\/ul>\n\n\n\n<div style=\"height:30px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n        <div class=\"cpb\">\n            <h3 class=\"cpb-heading\">\u3053\u306e\u8a18\u4e8b\u3067\u7d39\u4ecb\u3057\u305f\u88fd\u54c1<\/h3>\n            <hr class=\"cpb-heading-underline\" \/>\n            <div class=\"cpb-list\">\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/36024\">                                \u30ab\u30bf\u30ed\u30b0No. 36024                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">\u6b8b\u7559\u6eb6\u5a92\u30ac\u30b9 Standard Mix, 5000 \u00b5g\/mL in 1,2,4-Trimethylbenzene, 2.2 mL\/\u30a2\u30f3\u30d7\u30eb<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/36024\">\u88fd\u54c1\u60c5\u5831\u3092\u898b\u308b<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/36023\">                                \u30ab\u30bf\u30ed\u30b0No. 36023                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">\u6b8b\u7559\u6eb6\u5a92\u30ac\u30b9 Neopentane Standard, 5000 \u00b5g\/mL in 1,2,4-Trimethylbenzene, 2.2 mL\/\u30a2\u30f3\u30d7\u30eb<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/36023\">\u88fd\u54c1\u60c5\u5831\u3092\u898b\u308b<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/36022\">                                \u30ab\u30bf\u30ed\u30b0No. 36022                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">\u6b8b\u7559\u6eb6\u5a92\u30ac\u30b9 Butane Standard, 5000 \u00b5g\/mL in 1,2,4-Trimethylbenzene, 2.2 mL\/\u30a2\u30f3\u30d7\u30eb<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/36022\">\u88fd\u54c1\u60c5\u5831\u3092\u898b\u308b<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/36021\">                                \u30ab\u30bf\u30ed\u30b0No. 36021                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">\u6b8b\u7559\u6eb6\u5a92\u30ac\u30b9 Isobutane Standard, 5000 \u00b5g\/mL in 1,2,4-Trimethylbenzene, 2.2 mL\/\u30a2\u30f3\u30d7\u30eb<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/36021\">\u88fd\u54c1\u60c5\u5831\u3092\u898b\u308b<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/36020\">                                \u30ab\u30bf\u30ed\u30b0No. 36020                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">\u6b8b\u7559\u6eb6\u5a92\u30ac\u30b9 Propane Standard, 5000 \u00b5g\/mL  in 1,2,4-Trimethylbenzene, 2.2 mL\/\u30a2\u30f3\u30d7\u30eb<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/36020\">\u88fd\u54c1\u60c5\u5831\u3092\u898b\u308b<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/24571\">                                \u30ab\u30bf\u30ed\u30b0No. 24571                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Syringe, Hamilton 1750 (500 \u00b5L\/LTN\/22\/2&quot;\/2pt), PTFE Tip, Gas-Tight<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/24571\">\u88fd\u54c1\u60c5\u5831\u3092\u898b\u308b<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/24742\">                                \u30ab\u30bf\u30ed\u30b0No. 24742                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Syringe, SGE (500 \u00b5L\/F\/25\/50 mm\/2pt), PTFE Tip, Gas-Tight<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/24742\">\u88fd\u54c1\u60c5\u5831\u3092\u898b\u308b<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/24559\">                                \u30ab\u30bf\u30ed\u30b0No. 24559                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Syringe, Hamilton 1702 (25 \u00b5L\/N\/22s\/2&quot;\/2pt), PTFE Tip, Gas-Tight<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/24559\">\u88fd\u54c1\u60c5\u5831\u3092\u898b\u308b<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/24727\">                                \u30ab\u30bf\u30ed\u30b0No. 24727                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Syringe, SGE (25 \u00b5L\/F\/25\/50 mm\/2pt), PTFE Tip, Gas-Tight<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/24727\">\u88fd\u54c1\u60c5\u5831\u3092\u898b\u308b<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/24563\">                                \u30ab\u30bf\u30ed\u30b0No. 24563                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Syringe, Hamilton 1710 (100 \u00b5L\/N\/22s\/2&quot;\/2pt), PTFE Tip, Gas-Tight<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/24563\">\u88fd\u54c1\u60c5\u5831\u3092\u898b\u308b<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/24734\">                                \u30ab\u30bf\u30ed\u30b0No. 24734                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Syringe, SGE (100 \u00b5L\/F\/25\/50 mm\/2pt), PTFE Tip, Gas-Tight<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/24734\">\u88fd\u54c1\u60c5\u5831\u3092\u898b\u308b<\/a>\n                                                    <\/div>\n                    <\/div>\n                            <\/div>\n        <\/div>\n        \n\n\n<div style=\"height:100px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n","protected":false},"excerpt":{"rendered":"<p>In this guide, we describe best practice recommendations for use of your volatile gas standards, as well as a simplified workflow to ensure you are set up for success in beginning your analyses.<\/p>\n","protected":false},"author":11,"featured_media":0,"comment_status":"closed","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"_acf_changed":false,"_kad_blocks_custom_css":"","_kad_blocks_head_custom_js":"","_kad_blocks_body_custom_js":"","_kad_blocks_footer_custom_js":"","_kadence_starter_templates_imported_post":false,"_kad_post_transparent":"","_kad_post_title":"","_kad_post_layout":"","_kad_post_sidebar_id":"","_kad_post_content_style":"","_kad_post_vertical_padding":"","_kad_post_feature":"","_kad_post_feature_position":"","_kad_post_header":false,"_kad_post_footer":false,"footnotes":""},"categories":[7],"tags":[],"industries-application":[2149],"post-badge":[],"resource-type":[],"product-library":[2371,2380,2399,2374],"resource-technique":[2296],"ppma_author":[414],"class_list":["post-43557","post","type-post","status-publish","format-standard","hentry","category-articles","industries-application-cannabis","product-library-air-gas-sampling-products","product-library-gas-standards-accessories","product-library-liquid-reference-standards","product-library-reference-standard-products","resource-technique-headspace"],"acf":[],"taxonomy_info":{"category":[{"value":7,"label":"Articles"}],"industries-application":[{"value":2149,"label":"Cannabis"}],"product-library":[{"value":2371,"label":"Air &amp; Gas Sampling Products"},{"value":2380,"label":"Gas Standards &amp; Accessories"},{"value":2399,"label":"Liquid Reference Standards"},{"value":2374,"label":"Reference Standard Products"}],"resource-technique":[{"value":2296,"label":"Headspace"}]},"featured_image_src_large":false,"author_info":{"display_name":"Restek Corporation","author_link":"https:\/\/discover.restek.com\/ja\/author\/restek-corporation\/"},"comment_info":0,"category_info":[{"term_id":7,"name":"Articles","slug":"articles","term_group":0,"term_taxonomy_id":7,"taxonomy":"category","description":"","parent":0,"count":460,"filter":"raw","cat_ID":7,"category_count":460,"category_description":"","cat_name":"Articles","category_nicename":"articles","category_parent":0}],"tag_info":false,"authors":[{"term_id":414,"user_id":11,"is_guest":0,"slug":"restek-corporation","display_name":"Restek Corporation","avatar_url":{"url":"https:\/\/discover.restek.com\/wp-content\/uploads\/Restek_Favicon_300x300.jpg","url2x":"https:\/\/discover.restek.com\/wp-content\/uploads\/Restek_Favicon_300x300.jpg"},"author_category":"1","first_name":"Restek","last_name":"Corporation","user_url":"","job_title":"Chromatography Products and Solutions","description":"Restek is a leading provider of chromatography columns, accessories, and certified reference materials. Trust Restek for reliable, high-quality analytical solutions."}],"_links":{"self":[{"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/posts\/43557","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/users\/11"}],"replies":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/comments?post=43557"}],"version-history":[{"count":16,"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/posts\/43557\/revisions"}],"predecessor-version":[{"id":79452,"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/posts\/43557\/revisions\/79452"}],"wp:attachment":[{"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/media?parent=43557"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/categories?post=43557"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/tags?post=43557"},{"taxonomy":"industries-application","embeddable":true,"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/industries-application?post=43557"},{"taxonomy":"post-badge","embeddable":true,"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/post-badge?post=43557"},{"taxonomy":"resource-type","embeddable":true,"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/resource-type?post=43557"},{"taxonomy":"product-library","embeddable":true,"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/product-library?post=43557"},{"taxonomy":"resource-technique","embeddable":true,"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/resource-technique?post=43557"},{"taxonomy":"author","embeddable":true,"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/ppma_author?post=43557"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}