{"id":43653,"date":"2020-10-20T14:30:00","date_gmt":"2020-10-20T14:30:00","guid":{"rendered":"https:\/\/discover.restek.com\/uncategorized\/phospholipid-and-protein-removal-in-one-simple-spe-process-prevents-matrix-based-signal-suppression\/"},"modified":"2026-01-03T17:22:15","modified_gmt":"2026-01-03T17:22:15","slug":"phospholipid-and-protein-removal-in-one-simple-spe-process-prevents-matrix-based-signal-suppression","status":"publish","type":"post","link":"https:\/\/discover.restek.com\/ja\/articles\/gnar3124\/phospholipid-and-protein-removal-in-one-simple-spe-process-prevents-matrix-based-signal-suppression","title":{"rendered":"Phospholipid and Protein Removal in One Simple SPE Process Prevents Matrix-Based Signal Suppression"},"content":{"rendered":"\n<ul class=\"wp-block-list\">\n<li>Resprep PLR SPE removes both proteins and phospholipids in one easy, high-efficiency procedure.<\/li>\n\n\n\n<li>Avoid signal suppression by removing interfering phospholipids from the sample matrix.<\/li>\n\n\n\n<li>No method development\u2014straightforward and effective sample preparation for acids, bases, and neutral compounds.<\/li>\n<\/ul>\n\n\n\n<p class=\"wp-block-paragraph\">Whether you are analyzing whole blood, plasma, or serum, sample matrix components can interfere with analysis, decreasing both accuracy and sensitivity. Removing proteins through precipitation and filtration is a very common strategy for preventing problems, but with this procedure the sample still contains a high concentration of phospholipids. While LC-MS\/MS analytical methods will not detect phospholipids if the ions they produce are not monitored, their presence can still affect the analytical results and the overall cleanliness of the instrument.<\/p>\n\n\n\n<h2 class=\"wp-block-heading\">How to Avoid Phospholipid Signal Suppression<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">If phospholipids in the sample coelute with analytes of interest, they can affect target analyte ionization, which can cause suppression and lower overall analytical sensitivity. To illustrate signal suppression, we established a post-column infusion of nortriptyline and monitored the steady signal that the constant flow of analyte created. Then, we injected a sample of protein-precipitated plasma onto the column and monitored the presence of phospholipids as well as their effect on the nortriptyline signal. Figure 1A shows how certain phospholipids significantly reduced the nortriptyline response (by 38% in this case). In situations where a target analyte coelutes with phospholipids, use of internal standards can help account for phospholipid signal suppression and further method development can often eliminate the coelution. However, when phospholipid and protein removal can be done concurrently in the same process normally used to remove proteins alone, the question becomes \u201cwhy try to design a method around something I can easily remove from my sample?\u201d<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Resprep PLR SPE products allow you to do just that: one simple SPE workflow removes proteins through precipitation and filtration, while phospholipids are selectively retained in an innovative sintered composite material. Figure 1B illustrates the effectiveness of removing phospholipids along with proteins using a Resprep PLR 96-well plate (cartridges are also available). Comparing again to a continuous post-column infusion of nortriptyline, we can see that the signal is suppressed with protein precipitation alone, but when the sample phospholipids are removed along with the proteins, the signal is no longer compromised. Higher accuracy and sensitivity can be obtained by avoiding phospholipid signal suppression.<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><strong>Figure 1<\/strong>: Protein precipitation alone does not prevent phospholipid signal suppression (A), but simultaneous phospholipid and protein removal with Resprep PLR SPE eliminates this problem (B).<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n<div class=\"wp-block-custom-chromatogram-article\"><div class=\"wp-block-custom-chromatogram-article\"><div class=\"chromatogram-image regular-image\"><img decoding=\"async\" src=\"https:\/\/ez.restek.com\/images\/cgram\/lc_gn0594.png\" alt=\"Signal Suppression from Phospholipids\" title=\"-\"><\/div><p class=\"article-id\" style=\"text-align:center\"> LC_GN0594<\/p><div class=\"chromatogram-peaks\"><h4>Peaks<\/h4><table class=\"peaks col-lg-6 col-12\">\n<thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 75px\">t<sub>R<\/sub> (min)<\/th><th style=\"text-align: center;width: 75px\">Precursor Ion<\/th><th style=\"text-align: center;width: 75px\">Product Ion<\/th><\/tr><\/thead>\n<tbody><tr><td class=\"num\">1.<\/td><td class=\"cmpd\">Phosphatidylcholine 34:2<\/td><td class=\"oth\">0.831<\/td><td class=\"oth\">758.5<\/td><td class=\"oth\">184.1<\/td><\/tr>\n<tr><td class=\"num\">2.<\/td><td class=\"cmpd\">Phosphatidylcholine 36:2<\/td><td class=\"oth\">1.119<\/td><td class=\"oth\">786.5<\/td><td class=\"oth\">184.1<\/td><\/tr>\n<\/tbody><\/table><\/div><div class=\"chromatogram-conditions\"><h4>Conditions<\/h4><div class=\"conditions-container container-fluid\"><div class=\"row\"><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Column<\/th><td>Raptor Biphenyl  (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/9309232\" rel=\"noopener\">cat.# 9309232<\/a>)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Dimensions:<\/th><td>30 mm x 2.1 mm ID<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Particle Size:<\/th><td>1.8 \u00b5m<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Pore Size:<\/th><td>90 \u00c5<\/td><\/tr><tr><td><\/td><tr><th class=\"sub conditions_header\" scope=\"row\">Temp.:<\/th><td>40 \u00b0C<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Standard\/Sample<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Diluent:<\/th><td>Methanol<\/td><\/tr><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Vol.:<\/th><td>1 \u00b5L <\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Mobile Phase<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">A:<\/th><td>0.1% Formic acid in water <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">B:<\/th><td>0.1% Formic acid in methanol <\/td><\/tr><tr><td><\/td><td><table class=\"cgram_ramp\"><thead><tr><th>Time (min)<\/th><th>Flow (mL\/min)<\/th><th>%A<\/th><th>%B<\/th><\/tr><\/thead><tbody><tr><td>0.00<\/td><td>0.75<\/td><td>13.3<\/td><td>86.7<\/td><\/tr><tr><td>2.50<\/td><td>0.75<\/td><td>13.3<\/td><td>86.7<\/td><\/tr><\/tbody><\/table><\/td><\/tr><\/table><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Detector<\/th><td>MS\/MS<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Ion Mode:<\/th><td>ESI+ <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Mode:<\/th><td>MRM <\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Instrument<\/th><td>UHPLC<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Sample Preparation<\/th><td>Proteins were precipitated out of plasma using a 3:1 1% formic acid in acetonitrile:human plasma mixture that was then centrifuged. The supernatant was then divided and cleanup was performed on half using a Resprep PLR SPE 96-well plate (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/28301\" rel=\"noopener\">cat.# 28301<\/a>).<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Notes<\/th><td>Nortriptyline (15 ng\/mL) was infused post-column at 60 \u03bcL\/min in order to demonstrate signal suppression from coeluting phospholipids. To demonstrate the effect of cleanup on signal strength, 1 \u03bcL injections were made of protein crashed plasma samples both with and without Resprep PLR SPE cleanup.<\/td><\/tr><\/table><\/div><\/div><\/div><div class=\"chromatogram-related-skus\"><h4>Related SKUs<\/h4><ul><li>9309232<\/li><li>28301<\/li><\/ul><\/div><div class=\"chromatogram-pdf-link\"><a href=\"https:\/\/ez.restek.com\/images\/cgram\/lc_gn0594.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"><svg xmlns=\"http:\/\/www.w3.org\/2000\/svg\" width=\"18\" height=\"18\" viewBox=\"0 0 18 18\"><g data-name=\"Group 2996\"><path data-name=\"Rectangle 1246\" d=\"M0 0h18v18H0z\" style=\"fill: none;\"><\/path><\/g><g data-name=\"Group 2997\"><path data-name=\"Path 729\" d=\"M13.412 11.4v2.017H5.345V11.4H4v2.017a1.349 1.349 0 0 0 1.345 1.345h8.068a1.349 1.349 0 0 0 1.345-1.345V11.4zm-.672-2.694-.948-.948-1.741 1.735V4H8.706v5.493L6.965 7.758l-.948.948 3.361 3.361z\" transform=\"translate(-.437 -.414)\" style=\"fill: rgb(13, 123, 196);\"><\/path><\/g><\/svg>Download PDF<\/a><\/div><\/div><\/div>\n<\/div><\/div><\/div>\n\n\n<p class=\"wp-block-paragraph\"><\/p>\n\n\n\n<h2 class=\"wp-block-heading\">Excellent Recoveries for Wide Range of Compounds with No Method Development<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">Simultaneous phospholipid and protein removal using Resprep PLR SPE cartridges or 96-well plates can improve sensitivity by reducing phospholipid signal suppression, but its broad applicability offers an additional benefit: less time spent on method development. Simple to use and effective for a wide variety of compound chemistries, Resprep PLR SPE provides reliable cleanup without the need for compound-specific method development. As shown in the competitor comparison in Figure 2, good recoveries were obtained in plasma with a high degree of analytical precision for acidic, basic, and neutral compounds. Tight manufacturing controls ensure consistently strong performance, as demonstrated by the accurate and precise recovery results obtained in plasma for all three target compound categories across three independent lots (Table I).<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><strong>Figure 2<\/strong>: Evaluation of Analyte Recovery for a Single 96-Well Plate by Manufacturer*<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n<style>.kb-image43653_da7103-69 .kb-image-has-overlay:after{opacity:0.3;}<\/style>\n<div class=\"wp-block-kadence-image kb-image43653_da7103-69\"><figure class=\"aligncenter size-large\"><img loading=\"lazy\" decoding=\"async\" width=\"820\" height=\"1024\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR3124-02-820x1024.jpg\" alt=\"\" class=\"kb-img wp-image-20812\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR3124-02-820x1024.jpg 820w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR3124-02-240x300.jpg 240w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR3124-02-768x959.jpg 768w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR3124-02-1230x1536.jpg 1230w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR3124-02-1640x2048.jpg 1640w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR3124-02.jpg 1800w\" sizes=\"auto, (max-width: 820px) 100vw, 820px\" \/><\/figure><\/div>\n\n<\/div><\/div><\/div>\n\n\n<div style=\"height:20px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<p class=\"wp-block-paragraph\"><strong>Table I:<\/strong>&nbsp;Grand average percent recoveries for acids, bases, and neutral compounds across three different lots of Resprep PLR 96-well plates (3 plates per lot).<\/p>\n\n\n\n<figure class=\"wp-block-table\"><table class=\"has-fixed-layout\"><thead><tr><th><strong>Compound<br>Class<\/strong><\/th><th><strong>Compound<\/strong><\/th><th><strong>Fortification Level<br>(ng\/mL)<\/strong><\/th><th><strong>Grand Average of<br>% Recoveries<br>(n = 787 wells)<\/strong><\/th><th><strong>Standard<br>Deviation<\/strong><\/th><th><strong>%RSD<\/strong><\/th><\/tr><\/thead><tbody><tr><td rowspan=\"5\"><strong>Acids<\/strong><\/td><td>CBDA<\/td><td>150<\/td><td>94.1<\/td><td>9.3<\/td><td>9.9<\/td><\/tr><tr><td>Furosemide<\/td><td>250<\/td><td>105.3<\/td><td>17.2<\/td><td>16.4<\/td><\/tr><tr><td>Ketoprofen<\/td><td>125<\/td><td>105.4<\/td><td>7.3<\/td><td>6.9<\/td><\/tr><tr><td>Sulfadiazine<\/td><td>25<\/td><td>108.2<\/td><td>5.7<\/td><td>5.3<\/td><\/tr><tr><td>THCA<\/td><td>100<\/td><td>68.9<\/td><td>9.6<\/td><td>14.0<\/td><\/tr><tr><td rowspan=\"5\"><strong>Bases<\/strong><\/td><td>Amphetamine<\/td><td>15<\/td><td>106.8<\/td><td>5.6<\/td><td>5.3<\/td><\/tr><tr><td>Fluvoxamine<\/td><td>50<\/td><td>104.3<\/td><td>6.4<\/td><td>6.1<\/td><\/tr><tr><td>Methadone<\/td><td>5<\/td><td>99.4<\/td><td>6.0<\/td><td>6.1<\/td><\/tr><tr><td>Nortriptyline<\/td><td>15<\/td><td>101.2<\/td><td>6.0<\/td><td>6.0<\/td><\/tr><tr><td>Propranolol<\/td><td>25<\/td><td>105.0<\/td><td>6.2<\/td><td>5.9<\/td><\/tr><tr><td rowspan=\"2\"><strong>Neutrals<\/strong><\/td><td>CBN<\/td><td>100<\/td><td>87.7<\/td><td>9.1<\/td><td>10.4<\/td><\/tr><tr><td>Prednisolone<\/td><td>125<\/td><td>105.0<\/td><td>10.6<\/td><td>10.0<\/td><\/tr><\/tbody><\/table><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\">The reliable performance demonstrated here is due the robust design of Resprep PLR SPE products. A top frit is used to capture proteins, while the sorbent bed below is a specially designed porous polymer matrix that is co-sintered to silica particles that have bound ligands that retain phospholipids. The end result is a solid matrix for phospholipid and protein removal that\u2014unlike loose sorbent beds\u2014is not susceptible to channeling. Figure 3 shows the resulting performance: high levels of phospholipids are removed consistently across all tested wells in a representative 96-well plate. This means your target analytes can be effectively recovered and analytical accuracy will not be affected by phospholipid signal suppression. In addition, your instrument and overall productivity will benefit from not injecting high concentrations of phospholipids that will eventually reduce LC column lifetime and increase MS downtime as the ion source becomes contaminated.<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><strong>Figure 3<\/strong>: Representative phospholipid removal performance for a typical Resprep PLR 96-well plate.<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n<style>.kb-image43653_ced6a5-f3 .kb-image-has-overlay:after{opacity:0.3;}<\/style>\n<div class=\"wp-block-kadence-image kb-image43653_ced6a5-f3\"><figure class=\"aligncenter size-large\"><img loading=\"lazy\" decoding=\"async\" width=\"1024\" height=\"619\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR3124-03-1024x619.jpg\" alt=\"\" class=\"kb-img wp-image-20818\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR3124-03-1024x619.jpg 1024w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR3124-03-300x181.jpg 300w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR3124-03-768x464.jpg 768w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR3124-03-1536x928.jpg 1536w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR3124-03.jpg 1800w\" sizes=\"auto, (max-width: 1024px) 100vw, 1024px\" \/><\/figure><\/div>\n\n<\/div><\/div><\/div>\n\n\n<p class=\"wp-block-paragraph\">In conclusion, switching from protein precipitation alone to simultaneous phospholipid and protein removal with Resprep PLR SPE is an easy way to obtain more accurate, reliable results for clinically relevant substances in whole blood, plasma, or serum. The simple sample preparation process is effective for a wide range of analytes, so excellent results can be obtained for acids, bases, and neutral compounds without time and resources being spent on additional method development. In addition to removing interferences that could affect your data, Resprep PLR SPE ensures that less contamination is injected into your column and mass spectrometer, which results in less downtime for maintenance.<\/p>\n","protected":false},"excerpt":{"rendered":"<p>Biological samples contain proteins and phospholipids that can interfere with LC-MS\/MS analysis and increase the frequency of instrument maintenance. Here we used Resprep PLR SPE to simultaneously remove proteins and phospholipids from plasma in one high-efficiency sample preparation process. Concurrent phospholipid removal eliminated phospholipid signal suppression and excellent recoveries were obtained for a wide range of chemical classes without the need for additional method development.<\/p>\n","protected":false},"author":11,"featured_media":0,"comment_status":"closed","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"_acf_changed":false,"_kad_blocks_custom_css":"","_kad_blocks_head_custom_js":"","_kad_blocks_body_custom_js":"","_kad_blocks_footer_custom_js":"","_kadence_starter_templates_imported_post":false,"_kad_post_transparent":"","_kad_post_title":"","_kad_post_layout":"","_kad_post_sidebar_id":"","_kad_post_content_style":"","_kad_post_vertical_padding":"","_kad_post_feature":"","_kad_post_feature_position":"","_kad_post_header":false,"_kad_post_footer":false,"footnotes":"","_ppma_block_editor_authors":""},"categories":[7],"tags":[],"industries-application":[2153,2154,2156],"post-badge":[],"resource-type":[],"product-library":[2375,2402],"resource-technique":[2299,2304],"ppma_author":[414],"class_list":["post-43653","post","type-post","status-publish","format-standard","hentry","category-articles","industries-application-clinical","industries-application-clinical-diagnostics","industries-application-clinical-toxicology","product-library-sample-preparation-products","product-library-spe-sle","resource-technique-liquid-chromatography","resource-technique-solid-phase-extraction-spe"],"acf":[],"taxonomy_info":{"category":[{"value":7,"label":"Articles"}],"industries-application":[{"value":2153,"label":"Clinical"},{"value":2154,"label":"Clinical Diagnostics"},{"value":2156,"label":"Clinical Toxicology"}],"product-library":[{"value":2375,"label":"Sample Preparation Products"},{"value":2402,"label":"SPE &amp; SLE"}],"resource-technique":[{"value":2299,"label":"Liquid Chromatography"},{"value":2304,"label":"Solid-Phase Extraction (SPE)"}]},"featured_image_src_large":false,"author_info":{"display_name":"Restek Corporation","author_link":"https:\/\/discover.restek.com\/ja\/author\/restek-corporation\/"},"comment_info":0,"category_info":[{"term_id":7,"name":"Articles","slug":"articles","term_group":0,"term_taxonomy_id":7,"taxonomy":"category","description":"","parent":0,"count":511,"filter":"raw","cat_ID":7,"category_count":511,"category_description":"","cat_name":"Articles","category_nicename":"articles","category_parent":0}],"tag_info":false,"authors":[{"term_id":414,"user_id":11,"is_guest":0,"slug":"restek-corporation","display_name":"Restek Corporation","avatar_url":{"url":"https:\/\/discover.restek.com\/wp-content\/uploads\/Restek_Favicon_300x300.jpg","url2x":"https:\/\/discover.restek.com\/wp-content\/uploads\/Restek_Favicon_300x300.jpg"},"author_category":"1","first_name":"Restek","last_name":"Corporation","user_url":"https:\/\/www.linkedin.com\/company\/restek-corporation\/posts\/?feedView=all","job_title":"Chromatography Products and Solutions","description":"Restek is a leading provider of chromatography columns, accessories, and certified reference materials. Trust Restek for reliable, high-quality analytical solutions."}],"_links":{"self":[{"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/posts\/43653","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/users\/11"}],"replies":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/comments?post=43653"}],"version-history":[{"count":5,"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/posts\/43653\/revisions"}],"predecessor-version":[{"id":79430,"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/posts\/43653\/revisions\/79430"}],"wp:attachment":[{"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/media?parent=43653"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/categories?post=43653"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/tags?post=43653"},{"taxonomy":"industries-application","embeddable":true,"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/industries-application?post=43653"},{"taxonomy":"post-badge","embeddable":true,"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/post-badge?post=43653"},{"taxonomy":"resource-type","embeddable":true,"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/resource-type?post=43653"},{"taxonomy":"product-library","embeddable":true,"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/product-library?post=43653"},{"taxonomy":"resource-technique","embeddable":true,"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/resource-technique?post=43653"},{"taxonomy":"author","embeddable":true,"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/ppma_author?post=43653"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}