{"id":45256,"date":"2020-11-01T13:30:00","date_gmt":"2020-11-01T13:30:00","guid":{"rendered":"https:\/\/discover.restek.com\/uncategorized\/analysis-of-nicotine-and-impurities-in-electronic-cigarette-solutions-and-vapor\/"},"modified":"2025-12-10T20:57:37","modified_gmt":"2025-12-10T20:57:37","slug":"analysis-of-nicotine-and-impurities-in-electronic-cigarette-solutions-and-vapor","status":"publish","type":"post","link":"https:\/\/discover.restek.com\/ja\/notes-dapplication\/ffan2127\/analysis-of-nicotine-and-impurities-in-electronic-cigarette-solutions-and-vapor","title":{"rendered":"Analysis of Nicotine and Impurities in Electronic Cigarette Solutions and Vapor"},"content":{"rendered":"\n<h2 class=\"wp-block-heading\">Abstract<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">Electronic cigarettes (e-cigarettes) are growing in popularity exponentially. Despite their ever-growing acceptance, relatively little work has been done to characterize their vapor. To date, the majority of e-cigarette research has focused on characterizing the solutions, which are ultimately vaporized for the end user to inhale. The current study focused on developing a complete analytical package for the quick and simple analysis of electronic cigarette solutions and vapor to determine nicotine content and impurity profiles. Rapid (&lt;5 min) gas chromatography\u2013flame ionization detector (GC-FID) methods (using both helium and hydrogen carrier gas) were developed for the determination of nicotine content in e-cigarette solutions. In addition, a straightforward GC mass spectrometry (GC-MS) method was developed for the determination of impurities in e-cigarette liquids. Lastly, a simple sampling device was developed to draw e-cigarette vapor into a thermal desorption (TD) tube, which was then thermally extracted and analyzed via the same GC-MS method. This novel approach was able to provide detectable levels of volatile organic compounds (VOCs) and semivolatile organic compounds (SVOCs), which were not detected in the liquids, from a single 40 mL puff. All three of the methods may be done with one GC, two detectors, and one analytical column (Rtx-VMS), thereby reducing required resources and affording easy comparison of results.<\/p>\n\n\n\n<h2 class=\"wp-block-heading\">Introduction<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">Electronic cigarettes (e-cigarettes) do not burn tobacco, rather they produce an aerosol (without flame or smoke) from a battery-powered, metal heating element and liquid-containing cartridge [1]. The liquid typically consists of humectants (propylene glycol [1,2-propanediol] and\/or glycerin), flavorants, and nicotine [2]. When an e-cigarette\u2019s power source is activated, the heating element vaporizes the liquid to form a mist, which the end user then may inhale (often referred to as \u201cvape\u201d) [3]. The smoke-like vapor imitates tobacco smoke visually and replicates the burning sensation in the throat and lungs (often referred to as \u201cthroat hit\u201d). These similarities to tobacco smoke, combined with the same hand-to-mouth behaviors, have contributed to the rapid adaptation of electronic cigarettes [4-6]. Despite their increasing use on a global scale [3], relatively little is known about the e-cigarette chemical components. The majority of studies have focused primarily on the nicotine content and impurities (e.g., nitrosamines) of e-cigarette liquid (e-juice) [7]. More important, relatively little is known about the chemical composition of the vapor, which is ultimately what end users are exposed to [7, 8].<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Only a few researchers (e.g., Goniewicz et al. [7], Kosmider [9], and Schober et al. [8]) have attempted to characterize e-cigarette vapor by analyzing it for the presence of volatile organic compounds (VOCs), nitrosamines, heavy metals, and polycyclic aromatic hydrocarbons (PAHs); however, their study designs have been relatively complex and\/or required the use of a specialized smoking machine and\/or an array of specialized analytical instruments. Such requirements are often not practical for routine contract laboratory testing. The current study evaluates the nicotine content and impurities of several commercially available e-cigarettes and their respective solutions via simple and rapid GC-FID and GC-MS methods. In addition, the primary e-cigarette emissions were analyzed for VOCs and semivolatile organic compounds (SVOCs) via a simple and novel technique that pairs thermal desorption (TD) with GC-MS. Results, analytical techniques, obstacles, and solutions are discussed.<\/p>\n\n\n\n<h2 class=\"wp-block-heading\">Experimental<\/h2>\n\n\n\n<h3 class=\"wp-block-heading\">Electronic Cigarettes and Liquids<\/h3>\n\n\n\n<p class=\"wp-block-paragraph\">Four commercially available electronic cigarettes (Table I) were chosen from the \u201cBest E-Cigarettes of 2014,\u201d which is a top 10 list of e-cigarettes as viewed by \u201cexperts and users\u201d [10]. It is important to note that these four chosen e-cigarettes also routinely appeared on other web-based review sites as \u201ctop 10\u201d performers. In addition, these four brands were readily obtained from local stores. All four e-cigarettes were \u201c1st generation\u201d cigarettes (i.e., generally mimicking the size and look of regular cigarettes) [11] and, with the exception of vendor D, were disposable. In addition to the e-cigarettes, their respective e-liquids (i.e., same brand, flavor, and nicotine content) were obtained.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Vendors A, B, and C indicated their claimed nicotine percentage was based on wt\/wt analysis. Vendor D indicated their labeled value was based on vol\/vol analysis; however, one side of the D refill solution bottle denoted 1,000 mg of nicotine, which is in keeping with a wt\/wt analysis (which appears to be the industry standard) or a wt\/vol analysis. Therefore, it was not entirely clear how vendor D determined their nicotine concentrations. Upon receipt, 1 mL of each e-cigarette solution was pipetted with a calibrated syringe onto a calibrated scale to determine the density of each solution. Measured densities were later used to convert wt\/wt label claims to wt\/vol values for direct comparison to the analytically determined wt\/vol values using the following equation:<\/p>\n\n\n\n<figure class=\"wp-block-image\"><img decoding=\"async\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/equation-article-FFAN2127-01.jpg\" alt=\"decorative\" title=\"-\"><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\"><strong>Table I:<\/strong>&nbsp;Characteristics of Electronic Cigarettes and Liquids<\/p>\n\n\n<figure class=\"wp-block-table\"><table class=\"has-fixed-layout\"><thead><tr><th><strong>Vendor<\/strong><\/th><th><strong>Claimed Nicotine % (wt \/wt)<\/strong><\/th><th><strong>Style<\/strong><\/th><th><strong>Measured Density (g\/mL)<\/strong><\/th><\/tr><\/thead><tbody><tr><td>A<\/td><td>1.8 (18 mg\/1,000 mg)<\/td><td>Classic Tobacco<\/td><td>1.1179<\/td><\/tr><tr><td>B<\/td><td>1.2 (12 mg\/1,000 mg)<\/td><td>Classic Tobacco<\/td><td>1.1843<\/td><\/tr><tr><td>C<\/td><td>1.2 (12 mg\/1,000 mg)<\/td><td>Menthol<\/td><td>1.2006<\/td><\/tr><tr><td>D<\/td><td>&nbsp;1.8 (18 mL\/1,000 mL)*<\/td><td>Classic Tobacco<\/td><td>1.1271<\/td><\/tr><tr><td colspan=\"4\">*One label on the solution refill bottle indicated the % nicotine was based on % vol\/vol; however, the other side of the bottle denoted 1,000 mg, which is in keeping with wt\/wt analysis.<\/td><\/tr><\/tbody><\/table><\/figure>\n\n\n<h3 class=\"wp-block-heading\">Nicotine<\/h3>\n\n\n\n<p class=\"wp-block-paragraph\">The following system was used to analyze electronic cigarette e-liquid nicotine concentrations: an Agilent 7890A GC equipped with an Agilent FID. An Rtx-VMS column was chosen as the analytical column based on its unique ability to separate volatile compounds. The GC-FID parameters for both helium and hydrogen carrier gases are presented in Table II. The nicotine levels of the e-cigarette solutions were determined by calibrating the GC-FID with a National Institute of Standards and Technology (NIST) traceable nicotine standard (cat.# 34085). The 1,000 \u00b5g\/mL nicotine standard was serially diluted with methylene chloride to generate a 7-point external calibration curve (Table III). Although not shown, a United States (U.S.) Environmental Protection Agency (EPA) Method 8260 internal standard (cat.# 30074) was found to be suitable for the current work.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">All electronic cigarette solutions were diluted with methylene chloride by one hundred fold. This dilution was carried out for the following reasons: 1) Initial work with the e-cigarette solutions indicated the liquids were relatively viscous in nature. This viscosity resulted in the formation of air bubbles in the autosampler syringe. A 100:1 dilution remedied any viscosity issues. 2) The e-cigarette solutions chosen for this study appeared to have nicotine concentrations of ~15\u201325 mg\/mL, which was outside the concentration range of the calibration curve (Table III). A 100:1 dilution resulted in nicotine levels that fell between the upper and lower limits of the calibration curve. It is important to note that methylene chloride was chosen as the diluent instead of methanol because the methanol solvent peak coeluted with the ethanol (one of the major constituents of e-cigarette solutions) peak.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\"><strong>Table II:<\/strong>&nbsp;Analytical system and parameters utilized for quantifying the nicotine content of electronic cigarette liquids.<\/p>\n\n\n<figure class=\"wp-block-table\">\n<table class=\"has-fixed-layout\">\n<tbody>\n<tr>\n<td colspan=\"4\"><strong>Agilent 7890A GC-FID<\/strong><\/td>\n<\/tr>\n<tr>\n<td><strong>Column<\/strong><\/td>\n<td colspan=\"3\">Rtx-VMS, 30 m, 0.25 mm ID, 1.40 \u00b5m (cat.# 19915)<\/td>\n<\/tr>\n<tr>\n<td><strong>Injection<\/strong><\/td>\n<td colspan=\"3\">Diluted (100:1) electronic cigarette liquid<\/td>\n<\/tr>\n<tr>\n<td>Inj. Vol.<\/td>\n<td colspan=\"3\">1.0 \u00b5L split (200:1)<\/td>\n<\/tr>\n<tr>\n<td>Liner<\/td>\n<td colspan=\"3\">Restek Premium Precision liner w\/wool (cat.# 23305)<\/td>\n<\/tr>\n<tr>\n<td>Inj. Temp.<\/td>\n<td colspan=\"3\">250 \u00b0C<\/td>\n<\/tr>\n<tr>\n<td>Septum Purge Flow<\/td>\n<td colspan=\"3\">3 mL\/min<\/td>\n<\/tr>\n<tr>\n<td><strong>Detector<\/strong><\/td>\n<td colspan=\"3\">FID @ 250 \u00b0C<\/td>\n<\/tr>\n<tr>\n<td><strong>Carrier Gas<\/strong><\/td>\n<td>He, constant flow<\/td>\n<td>H<sub>2<\/sub>, constant flow*<\/td>\n<td>H<sub>2<\/sub>, constant flow*<\/td>\n<\/tr>\n<tr>\n<td>Flow Rate<\/td>\n<td>2.0 mL\/min<\/td>\n<td>2.50 mL\/min<\/td>\n<td>2.50 mL\/min<\/td>\n<\/tr>\n<tr>\n<td>Linear Velocity<\/td>\n<td>44.4 cm\/sec<\/td>\n<td>67.2 cm\/sec<\/td>\n<td>67.2 cm\/sec<\/td>\n<\/tr>\n<tr>\n<td><strong>Oven<\/strong><\/td>\n<td>100 \u00b0C to 260 \u00b0C at 35 \u00b0C\/min<br \/>(hold 0.25 min)<\/td>\n<td>100 \u00b0C to 260 \u00b0C at 54 \u00b0C\/min<br \/>(hold 0.15 min)<\/td>\n<td>100 \u00b0C to 240 \u00b0C at 35 \u00b0C\/min<\/td>\n<\/tr>\n<tr>\n<td colspan=\"4\">*Requires a fast ramping oven\u00a0<\/td>\n<\/tr>\n<\/tbody>\n<\/table>\n<\/figure>\n\n\n<p class=\"wp-block-paragraph\"><strong>Table III:<\/strong>&nbsp;External nicotine calibration curve for quantifying the nicotine content of electronic cigarette liquids.<\/p>\n\n\n\n<figure class=\"wp-block-table\"><table class=\"has-fixed-layout\"><tbody><tr><td colspan=\"5\"><strong>1.00 mg\/mL Nicotine Standard (cat.# 34085)<\/strong><\/td><\/tr><tr><td><strong>Level<\/strong><\/td><td><strong>\u00b5L of Previous Level<\/strong><\/td><td><strong>\u00b5L of Methylene Chloride<\/strong><\/td><td><strong>Total Volume (\u00b5L)<\/strong><\/td><td><strong>Concentration (mg\/mL)<\/strong><\/td><\/tr><tr><td>1<\/td><td>NA<\/td><td>NA<\/td><td>NA<\/td><td>1.00<\/td><\/tr><tr><td>2<\/td><td>100<\/td><td>100<\/td><td>200<\/td><td>0.500<\/td><\/tr><tr><td>3<\/td><td>100<\/td><td>100<\/td><td>200<\/td><td>0.250<\/td><\/tr><tr><td>4<\/td><td>100<\/td><td>100<\/td><td>200<\/td><td>0.125<\/td><\/tr><tr><td>5<\/td><td>100<\/td><td>100<\/td><td>200<\/td><td>0.063<\/td><\/tr><tr><td>6<\/td><td>100<\/td><td>100<\/td><td>200<\/td><td>0.031<\/td><\/tr><tr><td>7<\/td><td>100<\/td><td>100<\/td><td>200<\/td><td>0.016<\/td><\/tr><\/tbody><\/table><\/figure>\n\n\n\n<h3 class=\"wp-block-heading\">Impurities<\/h3>\n\n\n\n<p class=\"wp-block-paragraph\">The following analytical system was used for the qualitative determination of any impurities found in the electronic cigarette solutions: an Agilent 7890B GC coupled with an Agilent 5977A MS detector. The GC-MS parameters are presented in Table IV. This analysis also utilized the Rtx-VMS column based on its proven performance for volatile compounds.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\"><strong>Table IV:<\/strong>&nbsp;Analytical system and parameters utilized for determination of electronic cigarette solution impurities.<\/p>\n\n\n\n<figure class=\"wp-block-table\"><table class=\"has-fixed-layout\"><tbody><tr><td colspan=\"2\"><strong>Agilent 7890<\/strong><strong>B\/5977A GC-MS Parameters<\/strong><\/td><\/tr><tr><td><strong>Column<\/strong><\/td><td>Rtx-VMS, 30 m, 0.25 mm ID, 1.40 \u00b5m (cat.# 19915)<\/td><\/tr><tr><td><strong>Injection<\/strong><\/td><td>Diluted (2:1) electronic cigarette liquid<\/td><\/tr><tr><td>Inj. Vol.<\/td><td>1.0 \u00b5L split (10:1)<\/td><\/tr><tr><td>Liner<\/td><td>Restek Premium 4 mm Precision liner w\/wool (cat.# 23305)<\/td><\/tr><tr><td>Inj. Temp.<\/td><td>250 \u00b0C<\/td><\/tr><tr><td>Purge Flow<\/td><td>3 mL\/min<\/td><\/tr><tr><td><strong>Oven<\/strong><\/td><td>35 \u00b0C (hold 1 min) to 250 \u00b0C at 11 \u00b0C\/min (hold 4 min)<\/td><\/tr><tr><td><strong>Carrier Gas<\/strong><\/td><td>He, constant flow<\/td><\/tr><tr><td>Flow Rate<\/td><td>2.0 mL\/min<\/td><\/tr><tr><td>Linear Velocity<\/td><td>51.15 cm\/sec<\/td><\/tr><tr><td><strong>Detector<\/strong><\/td><td>MS<\/td><\/tr><tr><td>Mode<\/td><td>Scan<\/td><\/tr><tr><td>Transfer Line Temp.<\/td><td>250 \u00b0C<\/td><\/tr><tr><td>Analyzer Type<\/td><td>Single quadrupole<\/td><\/tr><tr><td>Source Temp.<\/td><td>230 \u00b0C<\/td><\/tr><tr><td>Quad Temp.<\/td><td>150 \u00b0C<\/td><\/tr><tr><td>Electron Energy<\/td><td>70 eV<\/td><\/tr><tr><td>Tune Type<\/td><td>BFB<\/td><\/tr><tr><td>Ionization Mode<\/td><td>EI<\/td><\/tr><tr><td>Acquisition Range<\/td><td>15 \u2013 550 amu<\/td><\/tr><tr><td>Rate<\/td><td>5.2 scans\/sec<\/td><\/tr><\/tbody><\/table><\/figure>\n\n\n\n<h3 class=\"wp-block-heading\">Vapor<\/h3>\n\n\n\n<p class=\"wp-block-paragraph\">Electronic cigarette vapor was analyzed for nicotine and impurities by trapping the vapor on thermal desorption tubes. Goniewicz et al. and other researchers have used smoking machines (e.g., Teague TE-2, Borgwaldt RM20S) to generate and collect e-cigarette aerosols; however, access to such an apparatus was not available for this study [7]. Therefore, in order to provide reproducible and quantitative results, a simple sampling device (Figure 1) was adapted from a 50 mL gas-tight syringe (cat.# 24761). The syringe was used to draw 40 mL of vapor in ~4 seconds from the e-cigarettes across a stainless steel thermal desorption tube packed with Tenax TA, Carbograph TD, and Carboxen 1003 (unconditioned [cat.# 26469] or conditioned [cat.# 26470]). This tube was chosen based on the optimized combination of three sorbents to screen for VOCs in the C<sub>2-3<\/sub>\u00a0range up to SVOCs in the C<sub>30-32<\/sub>\u00a0range. Although this method was manual, a ~4-second puff was utilized, as suggested based on Farsalinos et al.\u2019s observations on e-cigarette topography [12]. In addition to the single puff sample, a 10-puff sample was also taken in order to mimic a smoking regime. This sample was taken by manually drawing ten 4-second puffs separated by 10-second intervals between puffs. The desorption tube was then transferred to the following analytical system for determining the VOCs and SVOCs directly emitted from an e-cigarette: a Markes UNITY thermal desorption system paired with an Agilent 7890B GC coupled to an Agilent 5977A MS detector. The UNITY system and GC-MS parameters are presented in Table V and Table IV, respectively.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">The vapor concentrations of selected VOCs were calculated from a 5-point calibration curve generated by analyzing a series of volumes of a 10.0 ppbv primary standard (Table VI). The 10.0 ppbv primary standard was generated by injecting 180 mL of a 1.00 ppmv 75 component TO-15 + NJ mix (cat.# 34396) and 180 mL of a 1.00 ppmv ozone precursor mixture\/PAMS (cat.# 34420) into an evacuated 6-liter SilcoCan air monitoring canister (cat.# 27309) and pressurizing the canister to 30 psig with 50% RH nitrogen. Ochiai et al. [13] determined 50% RH to be optimal for stability. The standard was allowed to age for 7 days.<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><strong>Figure 1:<\/strong>\u00a0Gas-tight syringe sampling apparatus for quantitatively drawing electronic cigarette vapor into a thermal desorption tube.<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n<style>.kb-image45256_e26877-77 .kb-image-has-overlay:after{opacity:0.3;}<\/style>\n<div class=\"wp-block-kadence-image kb-image45256_e26877-77\"><figure class=\"aligncenter size-large\"><img loading=\"lazy\" decoding=\"async\" width=\"1024\" height=\"171\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-FFAN2127-01-1024x171.jpg\" alt=\"\" class=\"kb-img wp-image-19786\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-FFAN2127-01-1024x171.jpg 1024w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-FFAN2127-01-300x50.jpg 300w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-FFAN2127-01-768x128.jpg 768w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-FFAN2127-01-1536x256.jpg 1536w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-FFAN2127-01.jpg 1800w\" sizes=\"auto, (max-width: 1024px) 100vw, 1024px\" \/><\/figure><\/div>\n\n<\/div><\/div><\/div>\n\n\n<div style=\"height:20px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<p class=\"wp-block-paragraph\"><strong>Table V:<\/strong>&nbsp;Markes UNITY thermal desorption system and parameters utilized for thermally extracting electronic cigarette aerosols for the qualitative and quantitative determination of VOCs and SVOCs emitted from electronic cigarettes.<\/p>\n\n\n\n<figure class=\"wp-block-table\"><table class=\"has-fixed-layout\"><thead><tr><th colspan=\"4\"><strong>Markes UNITY Parameters<\/strong><\/th><\/tr><\/thead><tbody><tr><td colspan=\"2\"><strong>General Settings<\/strong><\/td><td colspan=\"2\"><strong>Trap Settings<\/strong><\/td><\/tr><tr><td>Operating Mode<\/td><td>Standard two stage<\/td><td>Pre-Trap Fire Purge<\/td><td>1.0 min<\/td><\/tr><tr><td>Standby Split<\/td><td>True<\/td><td>Flow<\/td><td>20.0 mL\/min<\/td><\/tr><tr><td>Standby Flow<\/td><td>5 mL\/min<\/td><td>Trap Low<\/td><td>0 \u00b0C<\/td><\/tr><tr><td>Flow Path Temperature<\/td><td>210 \u00b0C<\/td><td>Heating Rate<\/td><td>Max<\/td><\/tr><tr><td>Minimum Carrier Pressure<\/td><td>5.0 psi<\/td><td>Trap High<\/td><td>320 \u00b0C<\/td><\/tr><tr><td>GC Cycle Time<\/td><td>0.0<\/td><td>Trap Hold<\/td><td>5 min<\/td><\/tr><tr><td>&nbsp;<\/td><td>&nbsp;<\/td><td>Split On<\/td><td>True<\/td><\/tr><tr><td colspan=\"2\"><strong>Pre-Desorption<\/strong><\/td><td>Split On<\/td><td>20 mL\/min<\/td><\/tr><tr><td>Prepurge Time<\/td><td colspan=\"3\">1.0 min<\/td><\/tr><tr><td>Trap in Line<\/td><td colspan=\"3\">False<\/td><\/tr><tr><td>Split On<\/td><td colspan=\"3\">True<\/td><\/tr><tr><td>Flow<\/td><td colspan=\"3\">20 mL\/min<\/td><\/tr><tr><td colspan=\"4\"><strong>Tube\/Sample Desorption<\/strong><\/td><\/tr><tr><td>Time 1<\/td><td colspan=\"3\">10.0 min<\/td><\/tr><tr><td>Temperature 1<\/td><td colspan=\"3\">320 \u00b0C<\/td><\/tr><tr><td>Trap in Line<\/td><td colspan=\"3\">True<\/td><\/tr><tr><td>Split On<\/td><td colspan=\"3\">False<\/td><\/tr><\/tbody><\/table><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\"><\/p>\n\n\n\n<p class=\"wp-block-paragraph\"><strong>Table VI:<\/strong>&nbsp;Calibration curve for calculating vapor concentrations determined on a Markes UNITY thermal desorption system.<\/p>\n\n\n\n<figure class=\"wp-block-table\"><table class=\"has-fixed-layout\"><thead><tr><th><strong>Standard (ppbv)<\/strong><\/th><th><strong>Injection Volume (mL)<\/strong><\/th><th><strong>Calibration Concentration (ppbv)<\/strong><\/th><\/tr><\/thead><tbody><tr><td>10.0<\/td><td>720<\/td><td>180<\/td><\/tr><tr><td>10.0<\/td><td>360<\/td><td>90<\/td><\/tr><tr><td>10.0<\/td><td>120<\/td><td>30<\/td><\/tr><tr><td>10.0<\/td><td>40<\/td><td>10<\/td><\/tr><tr><td>10.0<\/td><td>4<\/td><td>1.00<\/td><\/tr><\/tbody><\/table><\/figure>\n\n\n\n<h3 class=\"wp-block-heading\">Blanks<\/h3>\n\n\n\n<p class=\"wp-block-paragraph\">The Markes UNITY system was operated with helium carrier gas for desorbing the thermal desorption tubes and the cryogenic trap during ballistic heating for analyte focusing on the head of the analytical column. The combination of helium gas (devoid of oxygen) and elevated temperatures may have established conditions that were ideal for pyrolysis of propylene glycol and\/or glycerin. The pyrolysis of propylene glycol and glycerin has been demonstrated to produce formaldehyde, acetaldehyde, and acrolein. Therefore, the following experiments were conducted to evaluate any compound contributions from the TD-GC-MS process itself: empty stainless steel tubes (i.e., no sorbents) and packed thermal desorption tubes (i.e., multi-bed sorbents) were injected with 1 \u00b5L aliquots of the electronic cigarette solutions and run through the TD-GC-MS analysis. In addition, the air drawn through the electronic cigarettes during sampling came from the laboratory. Due to the ubiquitous nature of VOCs such as formaldehyde and benzene, it was imperative to determine the background contributions of VOCs to the vapor analysis. Therefore, 40 mL samples of the laboratory air were periodically collected with thermal desorption tubes and analyzed with the same TD-GC-MS method.<\/p>\n\n\n\n<h2 class=\"wp-block-heading\">Results and Discussion<\/h2>\n\n\n\n<h3 class=\"wp-block-heading\">Nicotine<\/h3>\n\n\n\n<p class=\"wp-block-paragraph\">Analyses of electronic cigarette solutions, as shown in Figure 2 (helium), Figure 3 (hydrogen, fast ramp), and Figure 4 (hydrogen, standard ramp), using the GC-FID conditions in Table II afforded the rapid (i.e., &lt;5 minute GC run time) determination of the major chemical components. All four vendors\u2019 e-cigarette liquids appeared to contain ethanol, propylene glycol, glycerin, and nicotine. It is important to note that all four vendors listed propylene glycol, glycerin, and nicotine; however, none of the vendors listed ethanol as an ingredient. Blank analyses indicated that ethanol was not from laboratory contamination. Methylene chloride was used as the diluent to solve viscosity and concentrations issues, hence the abundant presence of methylene chloride. As shown in Figure 5, the rapid GC-FID method produced an acceptable external calibration of nicotine from 0.016 to 1.00 mg\/mL (r &gt; 0.995).<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">As shown in Table VII, the vendor claimed nicotine concentrations were lower than the actual measured nicotine concentrations by 4 to 28%. Recall the wt\/wt label claims were converted to wt\/vol values using the measured density of each solution in order to allow direct comparison to the actual values determined analytically using the calibration curve. The observation of increased nicotine content was consistent with what Schober et al. [8] and others have observed as well.<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><strong>Figure 2:<\/strong>\u00a0Analysis of major electronic cigarette solution components via GC-FID (helium).<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n<div class=\"wp-block-custom-chromatogram-article\"><div class=\"wp-block-custom-chromatogram-article\"><div class=\"chromatogram-image wide-image\"><img decoding=\"async\" src=\"https:\/\/ez.restek.com\/images\/cgram\/gc_ff1257.png\" alt=\"Electronic Cigarette Liquid on Rtx-VMS (Hydrogen\/Fast Ramp)\" title=\"-\"><\/div><p class=\"article-id\" style=\"text-align:center\"> GC_FF1257<\/p><div class=\"chromatogram-peaks\"><h4>Peaks<\/h4><table class=\"peaks col-lg-4 col-12\">\n<thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 50px\">t<sub>R<\/sub> (min)<\/th><\/tr><\/thead>\n<tbody><tr><td class=\"num\">1.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Methanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/67-56-1\/Methanol\" target=\"_blank\" rel=\"noopener\">Methanol<\/a><\/td><td class=\"oth\">0.861<\/td><\/tr>\n<tr><td class=\"num\">2.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Ethanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/64-17-5\/Ethanol\" target=\"_blank\" rel=\"noopener\">Ethanol<\/a><\/td><td class=\"oth\">0.905<\/td><\/tr>\n<tr><td class=\"num\">3.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Methylene chloride\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/75-09-2\/Methylene chloride\" target=\"_blank\" rel=\"noopener\">Methylene chloride<\/a><\/td><td class=\"oth\">0.957<\/td><\/tr>\n<tr><td class=\"num\">4.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Propylene glycol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/57-55-6\/Propylene glycol\" target=\"_blank\" rel=\"noopener\">Propylene glycol<\/a><\/td><td class=\"oth\">1.433<\/td><\/tr>\n<tr><td class=\"num\">5.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Glycerin\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/56-81-5\/Glycerin\" target=\"_blank\" rel=\"noopener\">Glycerin<\/a><\/td><td class=\"oth\">2.256<\/td><\/tr>\n<tr><td class=\"num\">6.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Nicotine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/54-11-5\/Nicotine\" target=\"_blank\" rel=\"noopener\">Nicotine<\/a><\/td><td class=\"oth\">3.030<\/td><\/tr>\n<\/tbody><\/table><\/div><div class=\"chromatogram-conditions\"><h4>Conditions<\/h4><div class=\"conditions-container container-fluid\"><div class=\"row\"><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Column<\/th><td>Rtx-VMS, 30 m, 0.25 mm ID, 1.40 \u00b5m (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/19915\" rel=\"noopener\">cat.# 19915<\/a>)<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Standard\/Sample<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Diluent:<\/th><td>Methylene chloride<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Conc.:<\/th><td>  Electronic cigarette liquid diluted 100:1<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Injection<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Vol.:<\/th><td>1.0 \u00b5L split (split ratio 200:1)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Liner:<\/th><td>Premium 4 mm Precision liner w\/wool (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/23305\" rel=\"noopener\">cat.# 23305<\/a>)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Temp.:<\/th><td>250 \u00b0C<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\" colspan=\"2\">Oven<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Oven Temp.:<\/th><td>100 &#176;C to 260 &#176;C at 54 &#176;C\/min (hold 0.15 min)<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Carrier Gas<\/th><td>H<sub>2<\/sub>, constant flow<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Flow Rate:<\/th><td>2.5 mL\/min<tr><th class=\"sub conditions_header\" scope=\"row\">Linear Velocity:<\/th><td>67.2 cm\/sec @ 100 \u00b0C<\/td><\/tr><\/table><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Detector<\/th><td>FID @ 250 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Make-up Gas Flow Rate:<\/th><td>50 mL\/min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Make-up Gas Type:<\/th><td>H<sub>2<\/sub> <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Hydrogen flow:<\/th><td>40 mL\/min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Air flow:<\/th><td>400 mL\/min<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Instrument<\/th><td>Agilent 7890A GC<\/td><\/tr><\/table><\/div><\/div><\/div><div class=\"chromatogram-related-skus\"><h4>Related SKUs<\/h4><ul><li>19915<\/li><li>23305<\/li><\/ul><\/div><div class=\"chromatogram-pdf-link\"><a href=\"https:\/\/ez.restek.com\/images\/cgram\/gc_ff1257.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"><svg xmlns=\"http:\/\/www.w3.org\/2000\/svg\" width=\"18\" height=\"18\" viewBox=\"0 0 18 18\"><g data-name=\"Group 2996\"><path data-name=\"Rectangle 1246\" d=\"M0 0h18v18H0z\" style=\"fill: none;\"><\/path><\/g><g data-name=\"Group 2997\"><path data-name=\"Path 729\" d=\"M13.412 11.4v2.017H5.345V11.4H4v2.017a1.349 1.349 0 0 0 1.345 1.345h8.068a1.349 1.349 0 0 0 1.345-1.345V11.4zm-.672-2.694-.948-.948-1.741 1.735V4H8.706v5.493L6.965 7.758l-.948.948 3.361 3.361z\" transform=\"translate(-.437 -.414)\" style=\"fill: rgb(13, 123, 196);\"><\/path><\/g><\/svg>Download PDF<\/a><\/div><\/div><\/div>\n<\/div><\/div><\/div>\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><strong>Figure 3:<\/strong>\u00a0Analysis of major electronic cigarette solution components via GC-FID (hydrogen, fast ramp).<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n<div class=\"wp-block-custom-chromatogram-article\"><div class=\"wp-block-custom-chromatogram-article\"><div class=\"chromatogram-image wide-image\"><img decoding=\"async\" src=\"https:\/\/ez.restek.com\/images\/cgram\/gc_ff1256.png\" alt=\"Electronic Cigarette Liquid on Rtx-VMS (Helium)\" title=\"-\"><\/div><p class=\"article-id\" style=\"text-align:center\"> GC_FF1256<\/p><div class=\"chromatogram-peaks\"><h4>Peaks<\/h4><table class=\"peaks col-lg-4 col-12\">\n<thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 50px\">t<sub>R<\/sub> (min)<\/th><\/tr><\/thead>\n<tbody><tr><td class=\"num\">1.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Methanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/67-56-1\/Methanol\" target=\"_blank\" rel=\"noopener\">Methanol<\/a><\/td><td class=\"oth\">1.285<\/td><\/tr>\n<tr><td class=\"num\">2.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Ethanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/64-17-5\/Ethanol\" target=\"_blank\" rel=\"noopener\">Ethanol<\/a><\/td><td class=\"oth\">1.355<\/td><\/tr>\n<tr><td class=\"num\">3.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Methylene chloride\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/75-09-2\/Methylene chloride\" target=\"_blank\" rel=\"noopener\">Methylene chloride<\/a><\/td><td class=\"oth\">1.430<\/td><\/tr>\n<tr><td class=\"num\">4.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Propylene glycol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/57-55-6\/Propylene glycol\" target=\"_blank\" rel=\"noopener\">Propylene glycol<\/a><\/td><td class=\"oth\">2.174<\/td><\/tr>\n<tr><td class=\"num\">5.<\/td><td class=\"cmpd\">Unknown<\/td><td class=\"oth\">3.371<\/td><\/tr>\n<tr><td class=\"num\">6.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Glycerin\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/56-81-5\/Glycerin\" target=\"_blank\" rel=\"noopener\">Glycerin<\/a><\/td><td class=\"oth\">3.446<\/td><\/tr>\n<tr><td class=\"num\">7.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Nicotine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/54-11-5\/Nicotine\" target=\"_blank\" rel=\"noopener\">Nicotine<\/a><\/td><td class=\"oth\">4.632<\/td><\/tr>\n<\/tbody><\/table><\/div><div class=\"chromatogram-conditions\"><h4>Conditions<\/h4><div class=\"conditions-container container-fluid\"><div class=\"row\"><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Column<\/th><td>Rtx-VMS, 30 m, 0.25 mm ID, 1.40 \u00b5m (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/19915\" rel=\"noopener\">cat.# 19915<\/a>)<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Standard\/Sample<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Diluent:<\/th><td>Methylene chloride<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Conc.:<\/th><td>  Electronic cigarette liquid diluted 100:1<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Injection<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Vol.:<\/th><td>1.0 \u00b5L split (split ratio 200:1)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Liner:<\/th><td>Premium 4 mm Precision liner w\/wool (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/23305\" rel=\"noopener\">cat.# 23305<\/a>)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Temp.:<\/th><td>250 \u00b0C<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\" colspan=\"2\">Oven<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Oven Temp.:<\/th><td>100 &#176;C to 260 &#176;C at 35 &#176;C\/min (hold 0.25 min)<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Carrier Gas<\/th><td>He, constant flow<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Flow Rate:<\/th><td>2.0 mL\/min<tr><th class=\"sub conditions_header\" scope=\"row\">Linear Velocity:<\/th><td>44.4 cm\/sec @ 100 \u00b0C<\/td><\/tr><\/table><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Detector<\/th><td>FID @ 250 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Make-up Gas Flow Rate:<\/th><td>50 mL\/min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Make-up Gas Type:<\/th><td>H<sub>2<\/sub> <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Hydrogen flow:<\/th><td>40 mL\/min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Air flow:<\/th><td>400 mL\/min<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Instrument<\/th><td>Agilent 7890A GC<\/td><\/tr><\/table><\/div><\/div><\/div><div class=\"chromatogram-related-skus\"><h4>Related SKUs<\/h4><ul><li>19915<\/li><li>23305<\/li><\/ul><\/div><div class=\"chromatogram-pdf-link\"><a href=\"https:\/\/ez.restek.com\/images\/cgram\/gc_ff1256.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"><svg xmlns=\"http:\/\/www.w3.org\/2000\/svg\" width=\"18\" height=\"18\" viewBox=\"0 0 18 18\"><g data-name=\"Group 2996\"><path data-name=\"Rectangle 1246\" d=\"M0 0h18v18H0z\" style=\"fill: none;\"><\/path><\/g><g data-name=\"Group 2997\"><path data-name=\"Path 729\" d=\"M13.412 11.4v2.017H5.345V11.4H4v2.017a1.349 1.349 0 0 0 1.345 1.345h8.068a1.349 1.349 0 0 0 1.345-1.345V11.4zm-.672-2.694-.948-.948-1.741 1.735V4H8.706v5.493L6.965 7.758l-.948.948 3.361 3.361z\" transform=\"translate(-.437 -.414)\" style=\"fill: rgb(13, 123, 196);\"><\/path><\/g><\/svg>Download PDF<\/a><\/div><\/div><\/div>\n<\/div><\/div><\/div>\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><strong>Figure 4:<\/strong>\u00a0Analysis of major electronic cigarette solution components via GC-FID (hydrogen, standard ramp).<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n<div class=\"wp-block-custom-chromatogram-article\"><div class=\"wp-block-custom-chromatogram-article\"><div class=\"chromatogram-image wide-image\"><img decoding=\"async\" src=\"https:\/\/ez.restek.com\/images\/cgram\/gc_ff1258.png\" alt=\"Electronic Cigarette Liquid on Rtx-VMS (Hydrogen)\" title=\"-\"><\/div><p class=\"article-id\" style=\"text-align:center\"> GC_FF1258<\/p><div class=\"chromatogram-peaks\"><h4>Peaks<\/h4><table class=\"peaks col-lg-4 col-12\">\n<thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 50px\">t<sub>R<\/sub> (min)<\/th><\/tr><\/thead>\n<tbody><tr><td class=\"num\">1.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Methanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/67-56-1\/Methanol\" target=\"_blank\" rel=\"noopener\">Methanol<\/a><\/td><td class=\"oth\">0.875<\/td><\/tr>\n<tr><td class=\"num\">2.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Ethanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/64-17-5\/Ethanol\" target=\"_blank\" rel=\"noopener\">Ethanol<\/a><\/td><td class=\"oth\">0.926<\/td><\/tr>\n<tr><td class=\"num\">3.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Methylene chloride\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/75-09-2\/Methylene chloride\" target=\"_blank\" rel=\"noopener\">Methylene chloride<\/a><\/td><td class=\"oth\">0.987<\/td><\/tr>\n<tr><td class=\"num\">4.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Propylene glycol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/57-55-6\/Propylene glycol\" target=\"_blank\" rel=\"noopener\">Propylene glycol<\/a><\/td><td class=\"oth\">1.619<\/td><\/tr>\n<tr><td class=\"num\">5.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Glycerin\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/56-81-5\/Glycerin\" target=\"_blank\" rel=\"noopener\">Glycerin<\/a><\/td><td class=\"oth\">2.805<\/td><\/tr>\n<tr><td class=\"num\">6.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Nicotine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/54-11-5\/Nicotine\" target=\"_blank\" rel=\"noopener\">Nicotine<\/a><\/td><td class=\"oth\">3.927<\/td><\/tr>\n<\/tbody><\/table><\/div><div class=\"chromatogram-conditions\"><h4>Conditions<\/h4><div class=\"conditions-container container-fluid\"><div class=\"row\"><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Column<\/th><td>Rtx-VMS, 30 m, 0.25 mm ID, 1.40 \u00b5m (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/19915\" rel=\"noopener\">cat.# 19915<\/a>)<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Standard\/Sample<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Diluent:<\/th><td>Methylene chloride<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Conc.:<\/th><td>  Electronic cigarette liquid diluted 100:1<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Injection<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Vol.:<\/th><td>1.0 \u00b5L split (split ratio 200:1)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Liner:<\/th><td>Premium 4 mm Precision liner w\/wool (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/23305\" rel=\"noopener\">cat.# 23305<\/a>)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Temp.:<\/th><td>250 \u00b0C<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\" colspan=\"2\">Oven<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Oven Temp.:<\/th><td>100 &#176;C to 240 &#176;C at 35 &#176;C\/min<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Carrier Gas<\/th><td>H<sub>2<\/sub>, constant flow<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Flow Rate:<\/th><td>2.5 mL\/min<tr><th class=\"sub conditions_header\" scope=\"row\">Linear Velocity:<\/th><td>67.2 cm\/sec @ 100 \u00b0C<\/td><\/tr><\/table><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Detector<\/th><td>FID @ 250 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Make-up Gas Flow Rate:<\/th><td>50 mL\/min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Make-up Gas Type:<\/th><td>H<sub>2<\/sub> <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Hydrogen flow:<\/th><td>40 mL\/min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Air flow:<\/th><td>400 mL\/min<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Instrument<\/th><td>Agilent 7890A GC<\/td><\/tr><\/table><\/div><\/div><\/div><div class=\"chromatogram-related-skus\"><h4>Related SKUs<\/h4><ul><li>19915<\/li><li>23305<\/li><\/ul><\/div><div class=\"chromatogram-pdf-link\"><a href=\"https:\/\/ez.restek.com\/images\/cgram\/gc_ff1258.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"><svg xmlns=\"http:\/\/www.w3.org\/2000\/svg\" width=\"18\" height=\"18\" viewBox=\"0 0 18 18\"><g data-name=\"Group 2996\"><path data-name=\"Rectangle 1246\" d=\"M0 0h18v18H0z\" style=\"fill: none;\"><\/path><\/g><g data-name=\"Group 2997\"><path data-name=\"Path 729\" d=\"M13.412 11.4v2.017H5.345V11.4H4v2.017a1.349 1.349 0 0 0 1.345 1.345h8.068a1.349 1.349 0 0 0 1.345-1.345V11.4zm-.672-2.694-.948-.948-1.741 1.735V4H8.706v5.493L6.965 7.758l-.948.948 3.361 3.361z\" transform=\"translate(-.437 -.414)\" style=\"fill: rgb(13, 123, 196);\"><\/path><\/g><\/svg>Download PDF<\/a><\/div><\/div><\/div>\n<\/div><\/div><\/div>\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><strong>Figure 5:<\/strong>\u00a0A linear response was obtained for nicotine over a concentration range of 0.06\u20131.00 mg\/mL using the GC-FID method as demonstrated by the external calibration curve (r &gt; 0.995).<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n<style>.kb-image45256_f0c603-44 .kb-image-has-overlay:after{opacity:0.3;}<\/style>\n<div class=\"wp-block-kadence-image kb-image45256_f0c603-44\"><figure class=\"aligncenter size-large\"><img loading=\"lazy\" decoding=\"async\" width=\"1024\" height=\"498\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-FFAN2127-05-1024x498.jpg\" alt=\"\" class=\"kb-img wp-image-19792\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-FFAN2127-05-1024x498.jpg 1024w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-FFAN2127-05-300x146.jpg 300w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-FFAN2127-05-768x374.jpg 768w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-FFAN2127-05-1536x748.jpg 1536w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-FFAN2127-05.jpg 1800w\" sizes=\"auto, (max-width: 1024px) 100vw, 1024px\" \/><\/figure><\/div>\n\n<\/div><\/div><\/div>\n\n\n<div style=\"height:30px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<p class=\"wp-block-paragraph\"><strong>Table VII:<\/strong>\u00a0Vendor nicotine concentrations as claimed and as determined in the current study by direct comparison with pure nicotine standards via GC-FID.<\/p>\n\n\n<figure class=\"wp-block-table\">\n<table class=\"has-fixed-layout\">\n<thead>\n<tr>\n<th><strong>Vendor<\/strong><\/th>\n<th><strong>Vendor Claimed Nicotine (mg\/mL)^<\/strong><\/th>\n<th><strong>Nicotine (mg\/mL) Determined*<\/strong><\/th>\n<th><strong>% Difference<\/strong><\/th>\n<\/tr>\n<\/thead>\n<tbody>\n<tr>\n<td>A<\/td>\n<td>20.1<\/td>\n<td>23.4<\/td>\n<td>17%<\/td>\n<\/tr>\n<tr>\n<td>B<\/td>\n<td>14.2<\/td>\n<td>14.8<\/td>\n<td>4%<\/td>\n<\/tr>\n<tr>\n<td>C<\/td>\n<td>14.4<\/td>\n<td>17.4<\/td>\n<td>21%<\/td>\n<\/tr>\n<tr>\n<td>D<\/td>\n<td>20.3<\/td>\n<td>26.0<\/td>\n<td>28%<\/td>\n<\/tr>\n<tr>\n<td colspan=\"4\">^ Calculated based on determined density.<br \/>* Average of 3 analyses.\u00a0<\/td>\n<\/tr>\n<\/tbody>\n<\/table>\n<\/figure>\n\n\n<div style=\"height:30px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<h3 class=\"wp-block-heading\">Impurities in E-Cigarette Solutions<\/h3>\n\n\n\n<p class=\"wp-block-paragraph\">As shown in Figure 6, the analysis of electronic cigarette solutions revealed that they contained numerous compounds besides the vendor-listed propylene glycol, glycerin, and nicotine. For the solution shown in Figure 6 (vendor A) there were 64 unidentified and identified (some only tentatively) compounds found in the e-cigarette solution. Compounds were deemed \u201cidentified\u201d when verified with a subsequent run of an external standard with matching retention times and mass spectral data. Compounds were deemed \u201ctentatively identified\u201d when the mass spectral quality was 80% or greater according to the NIST 2011 database [14]. Several pyrazines were tentatively identified, which is consistent with manufacturer-added flavorings. For example, acetylpyrazine, which was tentatively identified, is a flavorant well known for producing \u201cnutty\u201d flavors\/aromas. In addition, several pyridines were identified, which is consistent with tobacco-derived nicotine. For example, 3-(3,4-dihydro-2H-pyrrol-5-yl)- pyridine(myosmine) was also tentatively identified and this compound is an alkaloid found in tobacco [15]. It is important to note that almost half (36) of the compounds were unidentified; future work should focus on identifying these compounds.<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><strong>Figure 6:<\/strong>\u00a0Analysis of electronic cigarette solution (e-juice) by GC-MS revealed the presence of numerous components in addition to the compounds listed on the product labels.<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n<div class=\"wp-block-custom-chromatogram-article\"><div class=\"wp-block-custom-chromatogram-article\"><div class=\"chromatogram-image wide-image\"><img decoding=\"async\" src=\"https:\/\/ez.restek.com\/images\/cgram\/gc_ff1260.png\" alt=\"Electronic Cigarette Liquid on Rtx-VMS (GC-MS)\" title=\"-\"><\/div><p class=\"article-id\" style=\"text-align:center\"> GC_FF1260<\/p><div class=\"chromatogram-peaks\"><h4>Peaks<\/h4><table class=\"peaks col-12\">\n<thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 75px\">t<sub>R<\/sub> (min)<\/th><th style=\"text-align: center;width: 75px\">Match Quality<\/th><th style=\"text-align: center;width: 75px\">EC Liquid<\/th><th style=\"text-align: center;width: 75px\">Blank<\/th><th style=\"text-align: center;width: 75px\">Region<\/th><\/tr><\/thead>\n<tbody><tr><td class=\"num\">1.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Nitrogen\/oxygen\/carbon dioxide\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/7727-37-9\/Nitrogen_oxygen_carbon dioxide\" target=\"_blank\" rel=\"noopener\">Nitrogen\/oxygen\/carbon dioxide<\/a><\/td><td class=\"oth\">1.051<\/td><td class=\"oth\">100<\/td><td class=\"oth\">x<\/td><td class=\"oth\">x<\/td><td class=\"oth\">Red<\/td><\/tr>\n<tr><td class=\"num\">2.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Water\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/7732-18-5\/Water\" target=\"_blank\" rel=\"noopener\">Water<\/a><\/td><td class=\"oth\">1.441<\/td><td class=\"oth\">100<\/td><td class=\"oth\">x<\/td><td class=\"oth\">x<\/td><td class=\"oth\">Red<\/td><\/tr>\n<tr><td class=\"num\">3.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Methanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/67-56-1\/Methanol\" target=\"_blank\" rel=\"noopener\">Methanol<\/a><\/td><td class=\"oth\">1.709<\/td><td class=\"oth\">100<\/td><td class=\"oth\">x<\/td><td class=\"oth\">x<\/td><td class=\"oth\">Red<\/td><\/tr>\n<tr><td class=\"num\">4.<\/td><td class=\"cmpd\">Unidentified<\/td><td class=\"oth\">1.934<\/td><td class=\"oth\"><\/td><td class=\"oth\">x<\/td><td class=\"oth\">x<\/td><td class=\"oth\">Red<\/td><\/tr>\n<tr><td class=\"num\">5.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for cis-1,2-Dimethylcyclopropane\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/930-18-7\/cis-1,2-Dimethylcyclopropane\" target=\"_blank\" rel=\"noopener\">cis-1,2-Dimethylcyclopropane<\/a><\/td><td class=\"oth\">2.117<\/td><td class=\"oth\">94<\/td><td class=\"oth\">x<\/td><td class=\"oth\">x<\/td><td class=\"oth\">Red<\/td><\/tr>\n<tr><td class=\"num\">6.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Ethanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/64-17-5\/Ethanol\" target=\"_blank\" rel=\"noopener\">Ethanol<\/a><\/td><td class=\"oth\">2.239<\/td><td class=\"oth\">100<\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Red<\/td><\/tr>\n<tr><td class=\"num\">7.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 1,1-Dichloroethene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/75-35-4\/1,1-Dichloroethene\" target=\"_blank\" rel=\"noopener\">1,1-Dichloroethene<\/a><\/td><td class=\"oth\">2.282<\/td><td class=\"oth\">94<\/td><td class=\"oth\">x<\/td><td class=\"oth\">x<\/td><td class=\"oth\">Red<\/td><\/tr>\n<tr><td class=\"num\">8.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Methylene chloride\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/75-09-2\/Methylene chloride\" target=\"_blank\" rel=\"noopener\">Methylene chloride<\/a><\/td><td class=\"oth\">2.757<\/td><td class=\"oth\">100<\/td><td class=\"oth\">x<\/td><td class=\"oth\">x<\/td><td class=\"oth\">Red<\/td><\/tr>\n<tr><td class=\"num\">9.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 1,2-Dichloroethene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/540-59-0\/1,2-Dichloroethene\" target=\"_blank\" rel=\"noopener\">1,2-Dichloroethene<\/a><\/td><td class=\"oth\">2.891<\/td><td class=\"oth\">94<\/td><td class=\"oth\">x<\/td><td class=\"oth\">x<\/td><td class=\"oth\">Red<\/td><\/tr>\n<tr><td class=\"num\">10.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Ethyl acetate\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/141-78-6\/Ethyl acetate\" target=\"_blank\" rel=\"noopener\">Ethyl acetate<\/a><\/td><td class=\"oth\">4.037<\/td><td class=\"oth\">91<\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Red<\/td><\/tr>\n<tr><td class=\"num\">11.<\/td><td class=\"cmpd\">Unidentified<\/td><td class=\"oth\">6.000<\/td><td class=\"oth\"><\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Red<\/td><\/tr>\n<tr><td class=\"num\">12.<\/td><td class=\"cmpd\">Unidentified<\/td><td class=\"oth\">6.085<\/td><td class=\"oth\"><\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Red<\/td><\/tr>\n<tr><td class=\"num\">13.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Toluene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/108-88-3\/Toluene\" target=\"_blank\" rel=\"noopener\">Toluene<\/a><\/td><td class=\"oth\">6.207<\/td><td class=\"oth\">100<\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Red<\/td><\/tr>\n<tr><td class=\"num\">14.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Propylene glycol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/57-55-6\/Propylene glycol\" target=\"_blank\" rel=\"noopener\">Propylene glycol<\/a><\/td><td class=\"oth\">7.853<\/td><td class=\"oth\">100<\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Orange<\/td><\/tr>\n<tr><td class=\"num\">15.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2,3-Dimethylpyrazine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/5910-89-4\/2,3-Dimethylpyrazine\" target=\"_blank\" rel=\"noopener\">2,3-Dimethylpyrazine<\/a><\/td><td class=\"oth\">9.243<\/td><td class=\"oth\">91<\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Orange<\/td><\/tr>\n<tr><td class=\"num\">16.<\/td><td class=\"cmpd\">Unidentified<\/td><td class=\"oth\">9.615<\/td><td class=\"oth\"><\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Orange<\/td><\/tr>\n<tr><td class=\"num\">17.<\/td><td class=\"cmpd\">Unidentified<\/td><td class=\"oth\">9.713<\/td><td class=\"oth\"><\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Orange<\/td><\/tr>\n<tr><td class=\"num\">18.<\/td><td class=\"cmpd\">Unidentified<\/td><td class=\"oth\">9.889<\/td><td class=\"oth\"><\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Orange<\/td><\/tr>\n<tr><td class=\"num\">19.<\/td><td class=\"cmpd\">Unidentified<\/td><td class=\"oth\">10.017<\/td><td class=\"oth\"><\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Orange<\/td><\/tr>\n<tr><td class=\"num\">20.<\/td><td class=\"cmpd\">Unidentified<\/td><td class=\"oth\">10.060<\/td><td class=\"oth\"><\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Orange<\/td><\/tr>\n<tr><td class=\"num\">21.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Trimethylpyrazine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/14667-55-1\/Trimethylpyrazine\" target=\"_blank\" rel=\"noopener\">Trimethylpyrazine<\/a><\/td><td class=\"oth\">10.383<\/td><td class=\"oth\">94<\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Orange<\/td><\/tr>\n<tr><td class=\"num\">22.<\/td><td class=\"cmpd\">Unidentified<\/td><td class=\"oth\">10.828<\/td><td class=\"oth\"><\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Orange<\/td><\/tr>\n<tr><td class=\"num\">23.<\/td><td class=\"cmpd\">Unidentified<\/td><td class=\"oth\">10.907<\/td><td class=\"oth\"><\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Orange<\/td><\/tr>\n<tr><td class=\"num\">24.<\/td><td class=\"cmpd\">Unidentified<\/td><td class=\"oth\">11.047<\/td><td class=\"oth\"><\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Orange<\/td><\/tr>\n<tr><td class=\"num\">25.<\/td><td class=\"cmpd\">Unidentified<\/td><td class=\"oth\">11.114<\/td><td class=\"oth\"><\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Orange<\/td><\/tr>\n<tr><td class=\"num\">26.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Acetylpyrazine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/22047-25-2\/Acetylpyrazine\" target=\"_blank\" rel=\"noopener\">Acetylpyrazine<\/a><\/td><td class=\"oth\">11.394<\/td><td class=\"oth\">95<\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Orange<\/td><\/tr>\n<tr><td class=\"num\">27.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for N-(1-Methylethyl)benzenamine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/768-52-5\/N-(1-Methylethyl)benzenamine\" target=\"_blank\" rel=\"noopener\">N-(1-Methylethyl)benzenamine<\/a><\/td><td class=\"oth\">11.864<\/td><td class=\"oth\">80<\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Orange<\/td><\/tr>\n<tr><td class=\"num\">28.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Dipropylene glycol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/110-98-5\/Dipropylene glycol\" target=\"_blank\" rel=\"noopener\">Dipropylene glycol<\/a><\/td><td class=\"oth\">12.071<\/td><td class=\"oth\">91<\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Orange<\/td><\/tr>\n<tr><td class=\"num\">29.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Glycerin\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/56-81-5\/Glycerin\" target=\"_blank\" rel=\"noopener\">Glycerin<\/a><\/td><td class=\"oth\">12.473<\/td><td class=\"oth\">100<\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Orange<\/td><\/tr>\n<tr><td class=\"num\">30.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Dipropylene glycol methyl ether\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/34590-94-8\/Dipropylene glycol methyl ether\" target=\"_blank\" rel=\"noopener\">Dipropylene glycol methyl ether<\/a><\/td><td class=\"oth\">13.040<\/td><td class=\"oth\">80<\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Green<\/td><\/tr>\n<tr><td class=\"num\">31.<\/td><td class=\"cmpd\">Unidentified<\/td><td class=\"oth\">13.107<\/td><td class=\"oth\"><\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Green<\/td><\/tr>\n<tr><td class=\"num\">32.<\/td><td class=\"cmpd\">Unidentified<\/td><td class=\"oth\">13.168<\/td><td class=\"oth\"><\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Green<\/td><\/tr>\n<tr><td class=\"num\">33.<\/td><td class=\"cmpd\">Unidentified<\/td><td class=\"oth\">13.229<\/td><td class=\"oth\"><\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Green<\/td><\/tr>\n<tr><td class=\"num\">34.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 1-(3-Pyridinyl)ethanone\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/350-03-8\/1-(3-Pyridinyl)ethanone\" target=\"_blank\" rel=\"noopener\">1-(3-Pyridinyl)ethanone<\/a><\/td><td class=\"oth\">13.321<\/td><td class=\"oth\">94<\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Green<\/td><\/tr>\n<tr><td class=\"num\">35.<\/td><td class=\"cmpd\">Unidentified<\/td><td class=\"oth\">13.412<\/td><td class=\"oth\"><\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Green<\/td><\/tr>\n<tr><td class=\"num\">36.<\/td><td class=\"cmpd\">Unidentified<\/td><td class=\"oth\">13.463<\/td><td class=\"oth\"><\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Green<\/td><\/tr>\n<tr><td class=\"num\">37.<\/td><td class=\"cmpd\">Unidentified<\/td><td class=\"oth\">14.479<\/td><td class=\"oth\"><\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Green<\/td><\/tr>\n<tr><td class=\"num\">38.<\/td><td class=\"cmpd\">Unidentified<\/td><td class=\"oth\">14.534<\/td><td class=\"oth\"><\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Green<\/td><\/tr>\n<tr><td class=\"num\">39.<\/td><td class=\"cmpd\">Unidentified<\/td><td class=\"oth\">14.643<\/td><td class=\"oth\"><\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Green<\/td><\/tr>\n<tr><td class=\"num\">40.<\/td><td class=\"cmpd\">Unidentified<\/td><td class=\"oth\">14.863<\/td><td class=\"oth\"><\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Green<\/td><\/tr>\n<tr><td class=\"num\">41.<\/td><td class=\"cmpd\">Unidentified<\/td><td class=\"oth\">15.003<\/td><td class=\"oth\"><\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Green<\/td><\/tr>\n<tr><td class=\"num\">42.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Nicotine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/54-11-5\/Nicotine\" target=\"_blank\" rel=\"noopener\">Nicotine<\/a><\/td><td class=\"oth\">15.800<\/td><td class=\"oth\">100<\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Green<\/td><\/tr>\n<tr><td class=\"num\">43.<\/td><td class=\"cmpd\">Unidentified<\/td><td class=\"oth\">16.161<\/td><td class=\"oth\"><\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Blue<\/td><\/tr>\n<tr><td class=\"num\">44.<\/td><td class=\"cmpd\">Unidentified<\/td><td class=\"oth\">16.222<\/td><td class=\"oth\"><\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Blue<\/td><\/tr>\n<tr><td class=\"num\">45.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for \u03b1-Damascone\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/31089-90-4\/\u03b1-Damascone\" target=\"_blank\" rel=\"noopener\">\u03b1-Damascone<\/a><\/td><td class=\"oth\">16.289<\/td><td class=\"oth\">95<\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Blue<\/td><\/tr>\n<tr><td class=\"num\">46.<\/td><td class=\"cmpd\">Unidentified<\/td><td class=\"oth\">16.374<\/td><td class=\"oth\"><\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Blue<\/td><\/tr>\n<tr><td class=\"num\">47.<\/td><td class=\"cmpd\">Unidentified<\/td><td class=\"oth\">16.417<\/td><td class=\"oth\"><\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Blue<\/td><\/tr>\n<tr><td class=\"num\">48.<\/td><td class=\"cmpd\">Unidentified<\/td><td class=\"oth\">16.478<\/td><td class=\"oth\"><\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Blue<\/td><\/tr>\n<tr><td class=\"num\">49.<\/td><td class=\"cmpd\">Unidentified<\/td><td class=\"oth\">16.643<\/td><td class=\"oth\"><\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Blue<\/td><\/tr>\n<tr><td class=\"num\">50.<\/td><td class=\"cmpd\">Unidentified<\/td><td class=\"oth\">16.984<\/td><td class=\"oth\"><\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Blue<\/td><\/tr>\n<tr><td class=\"num\">51.<\/td><td class=\"cmpd\">Unidentified<\/td><td class=\"oth\">17.033<\/td><td class=\"oth\"><\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Blue<\/td><\/tr>\n<tr><td class=\"num\">52.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Myosmine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/532-12-7\/Myosmine\" target=\"_blank\" rel=\"noopener\">Myosmine<\/a><\/td><td class=\"oth\">17.155<\/td><td class=\"oth\">95<\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Blue<\/td><\/tr>\n<tr><td class=\"num\">53.<\/td><td class=\"cmpd\">Unidentified<\/td><td class=\"oth\">17.276<\/td><td class=\"oth\"><\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Blue<\/td><\/tr>\n<tr><td class=\"num\">54.<\/td><td class=\"cmpd\">Unidentified<\/td><td class=\"oth\">17.380<\/td><td class=\"oth\"><\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Blue<\/td><\/tr>\n<tr><td class=\"num\">55.<\/td><td class=\"cmpd\">Unidentified<\/td><td class=\"oth\">17.441<\/td><td class=\"oth\"><\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Blue<\/td><\/tr>\n<tr><td class=\"num\">56.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Nicotine 1-N-oxide\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/2820-55-5\/Nicotine 1-N-oxide\" target=\"_blank\" rel=\"noopener\">Nicotine 1-N-oxide<\/a><\/td><td class=\"oth\">17.533<\/td><td class=\"oth\">93<\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Blue<\/td><\/tr>\n<tr><td class=\"num\">57.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Anabasine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/494-52-0\/Anabasine\" target=\"_blank\" rel=\"noopener\">Anabasine<\/a><\/td><td class=\"oth\">17.697<\/td><td class=\"oth\">98<\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Blue<\/td><\/tr>\n<tr><td class=\"num\">58.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Nicotyrine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/487-19-4\/Nicotyrine\" target=\"_blank\" rel=\"noopener\">Nicotyrine<\/a><\/td><td class=\"oth\">17.752<\/td><td class=\"oth\">91<\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Blue<\/td><\/tr>\n<tr><td class=\"num\">59.<\/td><td class=\"cmpd\">Unidentified<\/td><td class=\"oth\">18.105<\/td><td class=\"oth\"><\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Blue<\/td><\/tr>\n<tr><td class=\"num\">60.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2,3-Dipyridyl\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/581-50-0\/2,3-Dipyridyl\" target=\"_blank\" rel=\"noopener\">2,3-Dipyridyl<\/a><\/td><td class=\"oth\">18.550<\/td><td class=\"oth\">97<\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Blue<\/td><\/tr>\n<tr><td class=\"num\">61.<\/td><td class=\"cmpd\">Unidentified<\/td><td class=\"oth\">19.788<\/td><td class=\"oth\"><\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Blue<\/td><\/tr>\n<tr><td class=\"num\">62.<\/td><td class=\"cmpd\">Unidentified<\/td><td class=\"oth\">21.025<\/td><td class=\"oth\"><\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Blue<\/td><\/tr>\n<tr><td class=\"num\">63.<\/td><td class=\"cmpd\">Unidentified<\/td><td class=\"oth\">21.092<\/td><td class=\"oth\"><\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Blue<\/td><\/tr>\n<tr><td class=\"num\">64.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Cotinine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/486-56-6\/Cotinine\" target=\"_blank\" rel=\"noopener\">Cotinine<\/a><\/td><td class=\"oth\">21.635<\/td><td class=\"oth\">91<\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Blue<\/td><\/tr>\n<\/tbody><\/table><\/div><div class=\"chromatogram-conditions\"><h4>Conditions<\/h4><div class=\"conditions-container container-fluid\"><div class=\"row\"><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Column<\/th><td>Rtx-VMS, 30 m, 0.25 mm ID, 1.40 \u00b5m (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/19915\" rel=\"noopener\">cat.# 19915<\/a>)<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Standard\/Sample<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Diluent:<\/th><td>Methylene chloride<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Conc.:<\/th><td>  Electronic cigarette liquid diluted 2:1<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Injection<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Vol.:<\/th><td>1 \u00b5L split (split ratio 10:1)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Liner:<\/th><td>Premium 4 mm Precision liner w\/wool (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/23305\" rel=\"noopener\">cat.# 23305<\/a>)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Temp.:<\/th><td>250 \u00b0C<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\" colspan=\"2\">Oven<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Oven Temp.:<\/th><td>35 &#176;C (hold 1 min) to 250 &#176;C at 11 &#176;C\/min (hold 4 min)<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Carrier Gas<\/th><td>He, constant flow<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Flow Rate:<\/th><td>2.0 mL\/min<tr><th class=\"sub conditions_header\" scope=\"row\">Linear Velocity:<\/th><td>51.15 cm\/sec @ 35 \u00b0C<\/td><\/tr><\/table><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Detector<\/th><td>MS<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Mode:<\/th><td>Scan<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Scan Program:<\/th><td><\/td><\/tr><tr><td style=\"text-align:left\" colspan=\"2\"><table class=\"conditions_sim\"><thead><tr><th style=\"padding: 0 5px\">Group<\/th><th style=\"padding: 0 5px\">Start Time<br \/>(min)<\/th><th style=\"padding: 0 5px\">Scan Range<br \/>(amu)<\/th><th style=\"padding: 0 5px\">Scan Rate<br \/>(scans\/sec)<\/th><\/tr><\/thead><tbody><tr><td>1<\/td><td>0<\/td><td>15-550<\/td><td>5.2<\/td><\/tr><\/tbody><\/table><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Transfer Line Temp.:<\/th><td>250 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Analyzer Type:<\/th><td>Quadrupole <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Source Type:<\/th><td>Extractor <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Extractor Lens:<\/th><td>6mm ID<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Source Temp.:<\/th><td>230 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Quad Temp.:<\/th><td>150 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Electron Energy:<\/th><td>70 eV<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Tune Type:<\/th><td>BFB <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Ionization Mode:<\/th><td>EI <\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Instrument<\/th><td>Agilent 7890B GC &amp; 5977A MSD<\/td><\/tr><\/table><\/div><\/div><\/div><div class=\"chromatogram-related-skus\"><h4>Related SKUs<\/h4><ul><li>19915<\/li><li>23305<\/li><\/ul><\/div><div class=\"chromatogram-pdf-link\"><a href=\"https:\/\/ez.restek.com\/images\/cgram\/gc_ff1260.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"><svg xmlns=\"http:\/\/www.w3.org\/2000\/svg\" width=\"18\" height=\"18\" viewBox=\"0 0 18 18\"><g data-name=\"Group 2996\"><path data-name=\"Rectangle 1246\" d=\"M0 0h18v18H0z\" style=\"fill: none;\"><\/path><\/g><g data-name=\"Group 2997\"><path data-name=\"Path 729\" d=\"M13.412 11.4v2.017H5.345V11.4H4v2.017a1.349 1.349 0 0 0 1.345 1.345h8.068a1.349 1.349 0 0 0 1.345-1.345V11.4zm-.672-2.694-.948-.948-1.741 1.735V4H8.706v5.493L6.965 7.758l-.948.948 3.361 3.361z\" transform=\"translate(-.437 -.414)\" style=\"fill: rgb(13, 123, 196);\"><\/path><\/g><\/svg>Download PDF<\/a><\/div><\/div><\/div>\n<\/div><\/div><\/div>\n\n\n<div style=\"height:30px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<h3 class=\"wp-block-heading\">Vapor<\/h3>\n\n\n\n<p class=\"wp-block-paragraph\">As shown in Figure 7, the simple sampling device (Figure 1) was able to successfully draw electronic cigarette vapor into a thermal desorption tube and provide detectable levels of VOCs and SVOCs from a single 40 mL puff. As observed in the impurities study, there clearly were numerous compounds (i.e., 82 unidentified and identified [some only tentatively]) in the e-cigarette vapor beyond propylene glycol, glycerin, and nicotine. However, the analysis of the vapor revealed the presence of 18 more compounds in addition to those found in the liquid analysis. Of particular interest was the presence of formaldehyde, acetaldehyde, acrolein, and xylenes, as well as several siloxanes. The current observation of these three carbonyls (formaldehyde, acetaldehyde, and acrolein) was consistent with Goniewicz et al.\u2019s [7] and Kosmider et al.\u2019s [9] observations.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">These observations are significant for the two following reasons: 1. All three of these carbonyls are acutely toxic; in addition, formaldehyde is a known human carcinogen [16] and acetaldehyde is a probable human carcinogen [17]. 2. These compounds were not present in the e-juice, which indicates they were generated during the vaporization process and\/or from the e-cigarette materials. This is consistent with the fact that pyrolysis of glycerin results in the formation of formaldehyde, acetaldehyde, and acrolein [18]. This is also consistent with the fact that polysiloxanes are often used as plastic additives and the majority of the first generation e-cigarettes, like those evaluated in this study, are made with plastic bodies. All of the aforementioned have profound implications for how e-cigarettes should be evaluated, especially when considering that end users are ultimately exposed to the e-cigarette vapor rather than the liquid.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">To expound upon this further, acrolein was not found in the electronic cigarette solutions. However, acrolein was found in the vapor from all four of the e-cigarettes evaluated in the current study. The acrolein concentrations ranged from 1.5 to 6.7 ppmv per 40 mL puff (0.003\u20130.015 \u00b5g\/mL), which is comparable to the 0.004 \u00b5g\/mL Goniewicz et al. reported [7]. To put these concentrations into perspective, these levels exceeded the National Institute of Occupational Safety and Health (NIOSH) short-term exposure limit (STEL) of 350 ppbv. Furthermore, assuming 40 mL per puff and 400 to 500 puffs per e-cigarette (values suggested by several e-cigarette manufacturers), each e-cigarette would generate ~20 to 230 \u00b5g of acrolein. From a human health perspective, the acrolein emissions observed in the current study appear to be on par with what has previously been reported for conventional tobacco cigarettes (3 to 220 \u00b5g of acrolein\/cigarette) [19]. Formaldehyde and acetaldehyde standards were not available at the time of publishing this application note. However, their peak areas were on the same order of magnitude as acrolein, thereby suggesting their concentrations were comparable, which is also consistent with what Goniewicz et al. reported [7].<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Currently, the U.S. Food and Drug Administration (FDA) does not have any regulatory authority over electronic cigarettes. However, the FDA does acknowledge that e-cigarettes, their associated risks, nicotine levels, and any potentially harmful chemicals inhaled are \u201cnot fully studied.\u201d Therefore, the FDA has issued a proposed rule to extend their authority to include e-cigarettes [20]. Regardless of the status of the FDA\u2019s authority over e-cigarettes, it is clear from the current research and the research of others that the e-cigarette landscape is not fully understood. However, it appears that e-cigarettes are not without human health risks. Most important, and as demonstrated by the current work, when designing future e-cigarette studies investigators should strongly consider the difference between analyzing electronic cigarette solutions and analyzing electronic cigarette vapor, as it very clear that their chemical profiles are different.<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><strong>Figure 7:<\/strong>\u00a0A single 40 mL puff of electronic cigarette vapor collected on a thermal desorption tube and analyzed via GC-MS.<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n<div class=\"wp-block-custom-chromatogram-article\"><div class=\"wp-block-custom-chromatogram-article\"><div class=\"chromatogram-image regular-image\"><img decoding=\"async\" src=\"https:\/\/ez.restek.com\/images\/cgram\/gc_ar1161.png\" alt=\"40 mL of Electronic Cigarette Vapor Collected on a Thermal Desorption Tube and Analyzed on Rtx-VMS\" title=\"-\"><\/div><p class=\"article-id\" style=\"text-align:center\"> GC_AR1161<\/p><div class=\"chromatogram-peaks\"><h4>Peaks<\/h4><table class=\"peaks col-12\">\n<thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 75px\">t<sub>R<\/sub> (min)<\/th><th style=\"text-align: center;width: 75px\">Match Quality<\/th><th style=\"text-align: center;width: 75px\">Vapor<\/th><th style=\"text-align: center;width: 75px\">Blank*<\/th><th style=\"text-align: center;width: 75px\">Region<\/th><\/tr><\/thead>\n<tbody><tr><td class=\"num\">1.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Nitrogen\/oxygen\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/7727-37-9\/Nitrogen_oxygen\" target=\"_blank\" rel=\"noopener\">Nitrogen\/oxygen<\/a><\/td><td class=\"oth\">0.685<\/td><td class=\"oth\">100<\/td><td class=\"oth\">x<\/td><td class=\"oth\">x<\/td><td class=\"oth\">Red<\/td><\/tr>\n<tr><td class=\"num\">2.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Carbon dioxide\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/124-38-9\/Carbon dioxide\" target=\"_blank\" rel=\"noopener\">Carbon dioxide<\/a><\/td><td class=\"oth\">1.063<\/td><td class=\"oth\">100<\/td><td class=\"oth\">x<\/td><td class=\"oth\">x<\/td><td class=\"oth\">Red<\/td><\/tr>\n<tr><td class=\"num\">3.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Propene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/115-07-1\/Propene\" target=\"_blank\" rel=\"noopener\">Propene<\/a><\/td><td class=\"oth\">1.200<\/td><td class=\"oth\">100<\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Red<\/td><\/tr>\n<tr><td class=\"num\">4.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Formaldehyde\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/50-00-0\/Formaldehyde\" target=\"_blank\" rel=\"noopener\">Formaldehyde<\/a><\/td><td class=\"oth\">1.227<\/td><td class=\"oth\">100<\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Red<\/td><\/tr>\n<tr><td class=\"num\">5.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Sulfur dioxide\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/7446-09-5\/Sulfur dioxide\" target=\"_blank\" rel=\"noopener\">Sulfur dioxide<\/a><\/td><td class=\"oth\">1.313<\/td><td class=\"oth\">90<\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Red<\/td><\/tr>\n<tr><td class=\"num\">6.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Chloromethane\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/74-87-3\/Chloromethane\" target=\"_blank\" rel=\"noopener\">Chloromethane<\/a><\/td><td class=\"oth\">1.380<\/td><td class=\"oth\">100<\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Red<\/td><\/tr>\n<tr><td class=\"num\">7.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Water\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/7732-18-5\/Water\" target=\"_blank\" rel=\"noopener\">Water<\/a><\/td><td class=\"oth\">1.453<\/td><td class=\"oth\">100<\/td><td class=\"oth\">x<\/td><td class=\"oth\">x<\/td><td class=\"oth\">Red<\/td><\/tr>\n<tr><td class=\"num\">8.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Acetaldehyde\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/75-07-0\/Acetaldehyde\" target=\"_blank\" rel=\"noopener\">Acetaldehyde<\/a><\/td><td class=\"oth\">1.672<\/td><td class=\"oth\">100<\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Red<\/td><\/tr>\n<tr><td class=\"num\">9.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Methanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/67-56-1\/Methanol\" target=\"_blank\" rel=\"noopener\">Methanol<\/a><\/td><td class=\"oth\">1.715<\/td><td class=\"oth\">100<\/td><td class=\"oth\">x<\/td><td class=\"oth\">x<\/td><td class=\"oth\">Red<\/td><\/tr>\n<tr><td class=\"num\">10.<\/td><td class=\"cmpd\">Unidentified<\/td><td class=\"oth\">1.885<\/td><td class=\"oth\"><\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Red<\/td><\/tr>\n<tr><td class=\"num\">11.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Ethanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/64-17-5\/Ethanol\" target=\"_blank\" rel=\"noopener\">Ethanol<\/a><\/td><td class=\"oth\">2.270<\/td><td class=\"oth\">100<\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Red<\/td><\/tr>\n<tr><td class=\"num\">12.<\/td><td class=\"cmpd\">Unidentified<\/td><td class=\"oth\">2.331<\/td><td class=\"oth\"><\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Red<\/td><\/tr>\n<tr><td class=\"num\">13.<\/td><td class=\"cmpd\">Unidentified<\/td><td class=\"oth\">2.410<\/td><td class=\"oth\"><\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Red<\/td><\/tr>\n<tr><td class=\"num\">14.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Acrolein\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/107-02-8\/Acrolein\" target=\"_blank\" rel=\"noopener\">Acrolein<\/a><\/td><td class=\"oth\">2.581<\/td><td class=\"oth\">100<\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Red<\/td><\/tr>\n<tr><td class=\"num\">15.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Propanal\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/123-38-6\/Propanal\" target=\"_blank\" rel=\"noopener\">Propanal<\/a><\/td><td class=\"oth\">2.629<\/td><td class=\"oth\">100<\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Red<\/td><\/tr>\n<tr><td class=\"num\">16.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Methylene chloride\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/75-09-2\/Methylene chloride\" target=\"_blank\" rel=\"noopener\">Methylene chloride<\/a><\/td><td class=\"oth\">2.770<\/td><td class=\"oth\">100<\/td><td class=\"oth\">x<\/td><td class=\"oth\">x<\/td><td class=\"oth\">Red<\/td><\/tr>\n<tr><td class=\"num\">17.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Acetone\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/67-64-1\/Acetone\" target=\"_blank\" rel=\"noopener\">Acetone<\/a><\/td><td class=\"oth\">2.843<\/td><td class=\"oth\">100<\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Red<\/td><\/tr>\n<tr><td class=\"num\">18.<\/td><td class=\"cmpd\">Unidentified<\/td><td class=\"oth\">2.892<\/td><td class=\"oth\"><\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Red<\/td><\/tr>\n<tr><td class=\"num\">19.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Hexane\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/110-54-3\/Hexane\" target=\"_blank\" rel=\"noopener\">Hexane<\/a><\/td><td class=\"oth\">2.928<\/td><td class=\"oth\">100<\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Red<\/td><\/tr>\n<tr><td class=\"num\">20.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Acetonitrile\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/75-05-8\/Acetonitrile\" target=\"_blank\" rel=\"noopener\">Acetonitrile<\/a><\/td><td class=\"oth\">3.160<\/td><td class=\"oth\">100<\/td><td class=\"oth\">x<\/td><td class=\"oth\">x<\/td><td class=\"oth\">Red<\/td><\/tr>\n<tr><td class=\"num\">21.<\/td><td class=\"cmpd\">Unidentified<\/td><td class=\"oth\">3.544<\/td><td class=\"oth\"><\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Orange<\/td><\/tr>\n<tr><td class=\"num\">22.<\/td><td class=\"cmpd\">Unidentified<\/td><td class=\"oth\">3.842<\/td><td class=\"oth\"><\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Orange<\/td><\/tr>\n<tr><td class=\"num\">23.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Trimethylsilanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/1066-40-6\/Trimethylsilanol\" target=\"_blank\" rel=\"noopener\">Trimethylsilanol<\/a><\/td><td class=\"oth\">3.928<\/td><td class=\"oth\">100<\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Orange<\/td><\/tr>\n<tr><td class=\"num\">24.<\/td><td class=\"cmpd\">Unidentified<\/td><td class=\"oth\">4.092<\/td><td class=\"oth\"><\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Orange<\/td><\/tr>\n<tr><td class=\"num\">25.<\/td><td class=\"cmpd\">Unidentified<\/td><td class=\"oth\">4.159<\/td><td class=\"oth\"><\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Orange<\/td><\/tr>\n<tr><td class=\"num\">26.<\/td><td class=\"cmpd\">Unidentified<\/td><td class=\"oth\">4.245<\/td><td class=\"oth\"><\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Orange<\/td><\/tr>\n<tr><td class=\"num\">27.<\/td><td class=\"cmpd\">Unidentified<\/td><td class=\"oth\">4.354<\/td><td class=\"oth\"><\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Orange<\/td><\/tr>\n<tr><td class=\"num\">28.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Benzene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/71-43-2\/Benzene\" target=\"_blank\" rel=\"noopener\">Benzene<\/a><\/td><td class=\"oth\">4.452<\/td><td class=\"oth\">100<\/td><td class=\"oth\">x<\/td><td class=\"oth\">x<\/td><td class=\"oth\">Orange<\/td><\/tr>\n<tr><td class=\"num\">29.<\/td><td class=\"cmpd\">Unidentified<\/td><td class=\"oth\">4.519<\/td><td class=\"oth\"><\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Orange<\/td><\/tr>\n<tr><td class=\"num\">30.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Acetic acid\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/64-19-7\/Acetic acid\" target=\"_blank\" rel=\"noopener\">Acetic acid<\/a><\/td><td class=\"oth\">5.055<\/td><td class=\"oth\">86<\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Orange<\/td><\/tr>\n<tr><td class=\"num\">31.<\/td><td class=\"cmpd\">Unidentified<\/td><td class=\"oth\">5.141<\/td><td class=\"oth\"><\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Orange<\/td><\/tr>\n<tr><td class=\"num\">32.<\/td><td class=\"cmpd\">Unidentified<\/td><td class=\"oth\">5.647<\/td><td class=\"oth\"><\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Orange<\/td><\/tr>\n<tr><td class=\"num\">33.<\/td><td class=\"cmpd\">Unidentified<\/td><td class=\"oth\">5.756<\/td><td class=\"oth\"><\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Orange<\/td><\/tr>\n<tr><td class=\"num\">34.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 1-Hydroxy-2-propanone\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/116-09-6\/1-Hydroxy-2-propanone\" target=\"_blank\" rel=\"noopener\">1-Hydroxy-2-propanone<\/a><\/td><td class=\"oth\">6.073<\/td><td class=\"oth\">80<\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Orange<\/td><\/tr>\n<tr><td class=\"num\">35.<\/td><td class=\"cmpd\">Unidentified<\/td><td class=\"oth\">6.165<\/td><td class=\"oth\"><\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Orange<\/td><\/tr>\n<tr><td class=\"num\">36.<\/td><td class=\"cmpd\">Unidentified<\/td><td class=\"oth\">6.220<\/td><td class=\"oth\"><\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Orange<\/td><\/tr>\n<tr><td class=\"num\">37.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Toluene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/108-88-3\/Toluene\" target=\"_blank\" rel=\"noopener\">Toluene<\/a><\/td><td class=\"oth\">6.280<\/td><td class=\"oth\">100<\/td><td class=\"oth\">x<\/td><td class=\"oth\">x<\/td><td class=\"oth\">Orange<\/td><\/tr>\n<tr><td class=\"num\">38.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Hexamethylcyclotrisiloxane\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/541-05-9\/Hexamethylcyclotrisiloxane\" target=\"_blank\" rel=\"noopener\">Hexamethylcyclotrisiloxane<\/a><\/td><td class=\"oth\">6.506<\/td><td class=\"oth\">91<\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Orange<\/td><\/tr>\n<tr><td class=\"num\">39.<\/td><td class=\"cmpd\">Unidentified<\/td><td class=\"oth\">7.231<\/td><td class=\"oth\"><\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Orange<\/td><\/tr>\n<tr><td class=\"num\">40.<\/td><td class=\"cmpd\">Unidentified<\/td><td class=\"oth\">7.530<\/td><td class=\"oth\"><\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Orange<\/td><\/tr>\n<tr><td class=\"num\">41.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Propylene glycol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/57-55-6\/Propylene glycol\" target=\"_blank\" rel=\"noopener\">Propylene glycol<\/a><\/td><td class=\"oth\">7.737<\/td><td class=\"oth\">100<\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Green<\/td><\/tr>\n<tr><td class=\"num\">42.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for m-Xylene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/108-38-3\/m-Xylene\" target=\"_blank\" rel=\"noopener\">m-Xylene<\/a><\/td><td class=\"oth\">8.048<\/td><td class=\"oth\">100<\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Green<\/td><\/tr>\n<tr><td class=\"num\">43.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for p-Xylene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/106-42-3\/p-Xylene\" target=\"_blank\" rel=\"noopener\">p-Xylene<\/a><\/td><td class=\"oth\">8.048<\/td><td class=\"oth\">100<\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Green<\/td><\/tr>\n<tr><td class=\"num\">44.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for o-Xylene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/95-47-6\/o-Xylene\" target=\"_blank\" rel=\"noopener\">o-Xylene<\/a><\/td><td class=\"oth\">8.530<\/td><td class=\"oth\">100<\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Green<\/td><\/tr>\n<tr><td class=\"num\">45.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Styrene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/100-42-5\/Styrene\" target=\"_blank\" rel=\"noopener\">Styrene<\/a><\/td><td class=\"oth\">8.597<\/td><td class=\"oth\">100<\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Green<\/td><\/tr>\n<tr><td class=\"num\">46.<\/td><td class=\"cmpd\">Unidentified<\/td><td class=\"oth\">9.158<\/td><td class=\"oth\"><\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Green<\/td><\/tr>\n<tr><td class=\"num\">47.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Octamethylcyclotetrasiloxane\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/556-67-2\/Octamethylcyclotetrasiloxane\" target=\"_blank\" rel=\"noopener\">Octamethylcyclotetrasiloxane<\/a><\/td><td class=\"oth\">9.218<\/td><td class=\"oth\">91<\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Green<\/td><\/tr>\n<tr><td class=\"num\">48.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 4-Methyl-1-(1-methylethyl)cyclohexene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/500-00-5\/4-Methyl-1-(1-methylethyl)cyclohexene\" target=\"_blank\" rel=\"noopener\">4-Methyl-1-(1-methylethyl)cyclohexene<\/a><\/td><td class=\"oth\">9.371<\/td><td class=\"oth\">95<\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Green<\/td><\/tr>\n<tr><td class=\"num\">49.<\/td><td class=\"cmpd\">Unidentified<\/td><td class=\"oth\">9.639<\/td><td class=\"oth\"><\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Green<\/td><\/tr>\n<tr><td class=\"num\">50.<\/td><td class=\"cmpd\">Unidentified<\/td><td class=\"oth\">9.852<\/td><td class=\"oth\"><\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Green<\/td><\/tr>\n<tr><td class=\"num\">51.<\/td><td class=\"cmpd\">Unidentified<\/td><td class=\"oth\">9.932<\/td><td class=\"oth\"><\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Green<\/td><\/tr>\n<tr><td class=\"num\">52.<\/td><td class=\"cmpd\">Unidentified<\/td><td class=\"oth\">10.121<\/td><td class=\"oth\"><\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Green<\/td><\/tr>\n<tr><td class=\"num\">53.<\/td><td class=\"cmpd\">Unidentified<\/td><td class=\"oth\">10.219<\/td><td class=\"oth\"><\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Green<\/td><\/tr>\n<tr><td class=\"num\">54.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Trimethylpyrazine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/14667-55-1\/Trimethylpyrazine\" target=\"_blank\" rel=\"noopener\">Trimethylpyrazine<\/a><\/td><td class=\"oth\">10.468<\/td><td class=\"oth\">80<\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Green<\/td><\/tr>\n<tr><td class=\"num\">55.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Benzaldehyde\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/100-52-7\/Benzaldehyde\" target=\"_blank\" rel=\"noopener\">Benzaldehyde<\/a><\/td><td class=\"oth\">10.657<\/td><td class=\"oth\">100<\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Green<\/td><\/tr>\n<tr><td class=\"num\">56.<\/td><td class=\"cmpd\">Unidentified<\/td><td class=\"oth\">10.858<\/td><td class=\"oth\"><\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Green<\/td><\/tr>\n<tr><td class=\"num\">57.<\/td><td class=\"cmpd\">Unidentified<\/td><td class=\"oth\">11.120<\/td><td class=\"oth\"><\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Green<\/td><\/tr>\n<tr><td class=\"num\">58.<\/td><td class=\"cmpd\">Unidentified<\/td><td class=\"oth\">11.187<\/td><td class=\"oth\"><\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Green<\/td><\/tr>\n<tr><td class=\"num\">59.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Acetylpyrazine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/22047-25-2\/Acetylpyrazine\" target=\"_blank\" rel=\"noopener\">Acetylpyrazine<\/a><\/td><td class=\"oth\">11.541<\/td><td class=\"oth\">93<\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Green<\/td><\/tr>\n<tr><td class=\"num\">60.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Decamethylcyclopentasiloxane\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/541-02-6\/Decamethylcyclopentasiloxane\" target=\"_blank\" rel=\"noopener\">Decamethylcyclopentasiloxane<\/a><\/td><td class=\"oth\">11.620<\/td><td class=\"oth\">91<\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Green<\/td><\/tr>\n<tr><td class=\"num\">61.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Phenol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/108-95-2\/Phenol\" target=\"_blank\" rel=\"noopener\">Phenol<\/a><\/td><td class=\"oth\">11.870<\/td><td class=\"oth\">94<\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Green<\/td><\/tr>\n<tr><td class=\"num\">62.<\/td><td class=\"cmpd\">Unidentified<\/td><td class=\"oth\">12.272<\/td><td class=\"oth\"><\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Green<\/td><\/tr>\n<tr><td class=\"num\">63.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 1,1&amp;apos;-Oxybis-2-propanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/110-98-5\/1,1&amp;apos;-Oxybis-2-propanol\" target=\"_blank\" rel=\"noopener\">1,1&#8242;-Oxybis-2-propanol<\/a><\/td><td class=\"oth\">12.333<\/td><td class=\"oth\">90<\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Green<\/td><\/tr>\n<tr><td class=\"num\">64.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Glycerin\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/56-81-5\/Glycerin\" target=\"_blank\" rel=\"noopener\">Glycerin<\/a><\/td><td class=\"oth\">12.748<\/td><td class=\"oth\">100<\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Blue<\/td><\/tr>\n<tr><td class=\"num\">65.<\/td><td class=\"cmpd\">Unidentified<\/td><td class=\"oth\">13.327<\/td><td class=\"oth\"><\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Blue<\/td><\/tr>\n<tr><td class=\"num\">66.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Dodecamethylcyclohexasiloxane\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/540-97-6\/Dodecamethylcyclohexasiloxane\" target=\"_blank\" rel=\"noopener\">Dodecamethylcyclohexasiloxane<\/a><\/td><td class=\"oth\">13.979<\/td><td class=\"oth\">94<\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Blue<\/td><\/tr>\n<tr><td class=\"num\">67.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Nicotine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/54-11-5\/Nicotine\" target=\"_blank\" rel=\"noopener\">Nicotine<\/a><\/td><td class=\"oth\">15.862<\/td><td class=\"oth\">100<\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Blue<\/td><\/tr>\n<tr><td class=\"num\">68.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Tetradecamethylhexasiloxane\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/107-52-8\/Tetradecamethylhexasiloxane\" target=\"_blank\" rel=\"noopener\">Tetradecamethylhexasiloxane<\/a><\/td><td class=\"oth\">16.082<\/td><td class=\"oth\">91<\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Blue<\/td><\/tr>\n<tr><td class=\"num\">69.<\/td><td class=\"cmpd\">Unidentified<\/td><td class=\"oth\">16.326<\/td><td class=\"oth\"><\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Blue<\/td><\/tr>\n<tr><td class=\"num\">70.<\/td><td class=\"cmpd\">Unidentified<\/td><td class=\"oth\">16.460<\/td><td class=\"oth\"><\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Blue<\/td><\/tr>\n<tr><td class=\"num\">71.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Myosmine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/532-12-7\/Myosmine\" target=\"_blank\" rel=\"noopener\">Myosmine<\/a><\/td><td class=\"oth\">17.216<\/td><td class=\"oth\">94<\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Blue<\/td><\/tr>\n<tr><td class=\"num\">72.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Nicotyrine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/487-19-4\/Nicotyrine\" target=\"_blank\" rel=\"noopener\">Nicotyrine<\/a><\/td><td class=\"oth\">17.807<\/td><td class=\"oth\">90<\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Blue<\/td><\/tr>\n<tr><td class=\"num\">73.<\/td><td class=\"cmpd\">Unidentified<\/td><td class=\"oth\">18.002<\/td><td class=\"oth\"><\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Blue<\/td><\/tr>\n<tr><td class=\"num\">74.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2,3&amp;apos;-Dipyridyl\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/581-50-0\/2,3&amp;apos;-Dipyridyl\" target=\"_blank\" rel=\"noopener\">2,3&#8242;-Dipyridyl<\/a><\/td><td class=\"oth\">18.618<\/td><td class=\"oth\">94<\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Blue<\/td><\/tr>\n<tr><td class=\"num\">75.<\/td><td class=\"cmpd\">Unidentified<\/td><td class=\"oth\">18.721<\/td><td class=\"oth\"><\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Blue<\/td><\/tr>\n<tr><td class=\"num\">76.<\/td><td class=\"cmpd\">Unidentified<\/td><td class=\"oth\">19.294<\/td><td class=\"oth\"><\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Blue<\/td><\/tr>\n<tr><td class=\"num\">77.<\/td><td class=\"cmpd\">Unidentified<\/td><td class=\"oth\">19.611<\/td><td class=\"oth\"><\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Blue<\/td><\/tr>\n<tr><td class=\"num\">78.<\/td><td class=\"cmpd\">Unidentified<\/td><td class=\"oth\">20.093<\/td><td class=\"oth\"><\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Blue<\/td><\/tr>\n<tr><td class=\"num\">79.<\/td><td class=\"cmpd\">Unidentified<\/td><td class=\"oth\">20.190<\/td><td class=\"oth\"><\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Blue<\/td><\/tr>\n<tr><td class=\"num\">80.<\/td><td class=\"cmpd\">Unidentified<\/td><td class=\"oth\">20.269<\/td><td class=\"oth\"><\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Blue<\/td><\/tr>\n<tr><td class=\"num\">81.<\/td><td class=\"cmpd\">Unidentified<\/td><td class=\"oth\">20.501<\/td><td class=\"oth\"><\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Blue<\/td><\/tr>\n<tr><td class=\"num\">82.<\/td><td class=\"cmpd\">Unidentified<\/td><td class=\"oth\">20.855<\/td><td class=\"oth\"><\/td><td class=\"oth\">x<\/td><td class=\"oth\"><\/td><td class=\"oth\">Blue<\/td><\/tr>\n<\/tbody><\/table><div class=\"peaknotes\">*The concentrations of these compounds in e-cigarette vapor were too close to blank and\/or laboratory air concentrations to definitively state they were emitted from the e-cigarettes.<\/div><\/div><div class=\"chromatogram-conditions\"><h4>Conditions<\/h4><div class=\"conditions-container container-fluid\"><div class=\"row\"><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Column<\/th><td>Rtx-VMS, 30 m, 0.25 mm ID, 1.40 \u00b5m (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/19915\" rel=\"noopener\">cat.# 19915<\/a>)<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Standard\/Sample<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Conc.:<\/th><td>  One 40 mL puff of electronic cigarette vapor drawn via a gas-tight syringe to replicate vaping<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Injection<\/th><td>Direct<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\" colspan=\"2\">Oven<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Oven Temp.:<\/th><td>35 &#176;C (hold 1 min) to 250 &#176;C at 11 &#176;C\/min (hold 4 min)<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Carrier Gas<\/th><td>He, constant flow<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Flow Rate:<\/th><td>2.0 mL\/min @ 35 \u00b0C<\/td><\/tr><\/table><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Detector<\/th><td>MS<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Mode:<\/th><td>Scan<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Scan Program:<\/th><td><\/td><\/tr><tr><td style=\"text-align:left\" colspan=\"2\"><table class=\"conditions_sim\"><thead><tr><th style=\"padding: 0 5px\">Group<\/th><th style=\"padding: 0 5px\">Start Time<br \/>(min)<\/th><th style=\"padding: 0 5px\">Scan Range<br \/>(amu)<\/th><th style=\"padding: 0 5px\">Scan Rate<br \/>(scans\/sec)<\/th><\/tr><\/thead><tbody><tr><td>1<\/td><td>0<\/td><td>15-550<\/td><td>5.2<\/td><\/tr><\/tbody><\/table><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Transfer Line Temp.:<\/th><td>250 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Analyzer Type:<\/th><td>Quadrupole <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Source Type:<\/th><td>Extractor <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Extractor Lens:<\/th><td>6mm ID<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Source Temp.:<\/th><td>230 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Quad Temp.:<\/th><td>150 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Electron Energy:<\/th><td>70 eV<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Tune Type:<\/th><td>BFB <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Ionization Mode:<\/th><td>EI <\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Preconcentrator<\/th><td>Markes UNITY\u2122<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Instrument<\/th><td>Agilent 7890B GC &amp; 5977A MSD<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Acknowledgement<\/th><td>Markes<\/td><\/tr><\/table><\/div><\/div><\/div><div class=\"chromatogram-related-skus\"><h4>Related SKUs<\/h4><ul><li>19915<\/li><\/ul><\/div><div class=\"chromatogram-pdf-link\"><a href=\"https:\/\/ez.restek.com\/images\/cgram\/gc_ar1161.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"><svg xmlns=\"http:\/\/www.w3.org\/2000\/svg\" width=\"18\" height=\"18\" viewBox=\"0 0 18 18\"><g data-name=\"Group 2996\"><path data-name=\"Rectangle 1246\" d=\"M0 0h18v18H0z\" style=\"fill: none;\"><\/path><\/g><g data-name=\"Group 2997\"><path data-name=\"Path 729\" d=\"M13.412 11.4v2.017H5.345V11.4H4v2.017a1.349 1.349 0 0 0 1.345 1.345h8.068a1.349 1.349 0 0 0 1.345-1.345V11.4zm-.672-2.694-.948-.948-1.741 1.735V4H8.706v5.493L6.965 7.758l-.948.948 3.361 3.361z\" transform=\"translate(-.437 -.414)\" style=\"fill: rgb(13, 123, 196);\"><\/path><\/g><\/svg>Download PDF<\/a><\/div><\/div><\/div>\n<\/div><\/div><\/div>\n\n\n<div style=\"height:30px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<h3 class=\"wp-block-heading\">Blanks<\/h3>\n\n\n\n<p class=\"wp-block-paragraph\">The 1 \u00b5L aliquots of electronic cigarette solutions injected into empty stainless steel tubes (i.e., no sorbents) and analyzed via the TD-GC-MS method resulted in the formation of formaldehyde, acetaldehyde, and acrolein. However, the concentrations of these three compounds did not increase when 1 \u00b5L aliquots of the e-cigarette solutions were injected into packed thermal desorption tubes (i.e., multi-bed sorbents) and analyzed via the TD-GC-MS method. The two aforementioned observations are consistent with the hypothesis that pyrolysis of propylene glycol and\/or glycerin was taking place within the TD-GC-MS system itself and not in the thermal desorption tube media (i.e., the multi-sorbent bed). However, it was unclear as to where the pyrolysis was taking place (i.e., on the cryogenic trap during ballistic heating versus in the heated transfer lines) within the TD-GC-MS system. Regardless, the pyrolysis was responsible for 14 to 23% of the vapor concentrations of formaldehyde, acetaldehyde, and acrolein observed in the current study. The aforesaid percent contributions were approximated by comparing the carbonyl\/nicotine ratios obtained from the empty stainless steel tubes and packed thermal desorption tubes to the 40 mL puff samples. In addition, the laboratory air was sometimes a source for certain VOCs; however, these levels (i.e., low ppbv) were often well below the e-cigarette levels (i.e., low to mid ppmv). Future investigators should be aware of their laboratory air concentrations and the potential pyrolysis within the TD-GC-MS system and make necessary adjustments in their reporting limits and\/or background corrections. It was outside the scope of the current work; however, future work should focus on reducing pyrolysis contribution by adjusting line temperatures, heating rates, flow rates, etc.<\/p>\n\n\n\n<h3 class=\"wp-block-heading\">Advantages\/Limitations\/Future Research<\/h3>\n\n\n\n<p class=\"wp-block-paragraph\">Researchers like Goniewicz et al. had access to specialized smoking machines, which enabled \u201crealistic\u201d smoking regimes (e.g., a 1.8 second puff with 10 second intervals between puffs). These smoking regimes may reveal more about e-cigarette vapor and\/or be more accurate than the simple sampling device (Figure 1) utilized in the current study. However, the current work is significant in that multiple puffs were not needed because the present analytical techniques demonstrated detectability from a single 40 mL puff. In fact, it is important to note that a smoking regime of a 4-second puff with 10-second intervals between 10 puffs was executed manually with the simple sampling device (Figure 1). The results of this 10-puff sample are shown in Figure 8. The 10-puff sample did reveal some early eluting compounds (i.e., identified, tentatively identified, and unidentified), which were not identified in the single-puff (Figure 7). However, the propylene glycol and glycerin peaks, which were already overloaded in the single-puff sample, became so large in the 10-puff sample that most of the peaks previously identified in the single-puff sample were lost due to interference with propylene glycol and glycerin. In addition, this overloading of propylene glycol and glycerin contaminated the Markes UNITY thermal desorption system, thereby requiring a time-consuming cleaning to avoid carryover.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">As previously mentioned in the discussion of the blanks results, future researchers should be aware of the potential pyrolysis conditions within the TD-GC-MS system and how that may affect their formaldehyde, acetaldehyde, and acrolein vapor concentrations. Alternative sampling\/analytical approaches (e.g., DNPH-coated solid sorbents) are available for these carbonyls, which would circumvent the pyrolysis issues; however, they come at the significant disadvantage of time-consuming solvent extractions and the inability to scan for a large number of compounds (e.g., the 82 VOCs\/SVOCs observed in the current study) in a single 40 mL puff. Future TD-GC-MS work on e-cigarette vapor should focus on optimizing the thermal desorption parameters in order to reduce pyrolysis contributions by adjusting line temperatures, heating rates, flow rates, etc. Overall, the current method may be well suited for the easy and rapid screening of e-cigarette vapor for a large number of VOCs and SVOCs.<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><strong>Figure 8:<\/strong>\u00a0Ten 40 mL puffs of electronic cigarette vapor collected on a thermal desorption tube and analyzed via GC-MS.<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n<div class=\"wp-block-custom-chromatogram-article\"><div class=\"wp-block-custom-chromatogram-article\"><div class=\"chromatogram-image regular-image\"><img decoding=\"async\" src=\"https:\/\/ez.restek.com\/images\/cgram\/gc_ar1162.png\" alt=\"400 mL of Electronic Cigarette Vapor Collected on a Thermal Desorption Tube and Analyzed on Rtx-VMS\" title=\"-\"><\/div><p class=\"article-id\" style=\"text-align:center\"> GC_AR1162<\/p><div class=\"chromatogram-conditions\"><h4>Conditions<\/h4><div class=\"conditions-container container-fluid\"><div class=\"row\"><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Column<\/th><td>Rtx-VMS, 30 m, 0.25 mm ID, 1.40 \u00b5m (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/19915\" rel=\"noopener\">cat.# 19915<\/a>)<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Standard\/Sample<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Conc.:<\/th><td>  Ten 40 mL puffs of electronic cigarette vapor drawn via a gas-tight syringe to replicate vaping<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Injection<\/th><td>Direct<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\" colspan=\"2\">Oven<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Oven Temp.:<\/th><td>35 &#176;C (hold 1 min) to 250 &#176;C at 11 &#176;C\/min (hold 4 min)<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Carrier Gas<\/th><td>He, constant flow<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Flow Rate:<\/th><td>2.0 mL\/min @ 35 \u00b0C<\/td><\/tr><\/table><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Detector<\/th><td>MS<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Mode:<\/th><td>Scan<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Scan Program:<\/th><td><\/td><\/tr><tr><td style=\"text-align:left\" colspan=\"2\"><table class=\"conditions_sim\"><thead><tr><th style=\"padding: 0 5px\">Group<\/th><th style=\"padding: 0 5px\">Start Time<br \/>(min)<\/th><th style=\"padding: 0 5px\">Scan Range<br \/>(amu)<\/th><th style=\"padding: 0 5px\">Scan Rate<br \/>(scans\/sec)<\/th><\/tr><\/thead><tbody><tr><td>1<\/td><td>0<\/td><td>15-550<\/td><td>5.2<\/td><\/tr><\/tbody><\/table><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Transfer Line Temp.:<\/th><td>250 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Analyzer Type:<\/th><td>Quadrupole <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Source Type:<\/th><td>Extractor <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Extractor Lens:<\/th><td>6mm ID<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Source Temp.:<\/th><td>230 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Quad Temp.:<\/th><td>150 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Electron Energy:<\/th><td>70 eV<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Tune Type:<\/th><td>BFB <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Ionization Mode:<\/th><td>EI <\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Preconcentrator<\/th><td>Markes UNITY\u2122<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Instrument<\/th><td>Agilent 7890B GC &amp; 5977A MSD<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Acknowledgement<\/th><td>Markes<\/td><\/tr><\/table><\/div><\/div><\/div><div class=\"chromatogram-related-skus\"><h4>Related SKUs<\/h4><ul><li>19915<\/li><\/ul><\/div><div class=\"chromatogram-pdf-link\"><a href=\"https:\/\/ez.restek.com\/images\/cgram\/gc_ar1162.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"><svg xmlns=\"http:\/\/www.w3.org\/2000\/svg\" width=\"18\" height=\"18\" viewBox=\"0 0 18 18\"><g data-name=\"Group 2996\"><path data-name=\"Rectangle 1246\" d=\"M0 0h18v18H0z\" style=\"fill: none;\"><\/path><\/g><g data-name=\"Group 2997\"><path data-name=\"Path 729\" d=\"M13.412 11.4v2.017H5.345V11.4H4v2.017a1.349 1.349 0 0 0 1.345 1.345h8.068a1.349 1.349 0 0 0 1.345-1.345V11.4zm-.672-2.694-.948-.948-1.741 1.735V4H8.706v5.493L6.965 7.758l-.948.948 3.361 3.361z\" transform=\"translate(-.437 -.414)\" style=\"fill: rgb(13, 123, 196);\"><\/path><\/g><\/svg>Download PDF<\/a><\/div><\/div><\/div>\n<\/div><\/div><\/div>\n\n\n<p class=\"wp-block-paragraph\"><\/p>\n\n\n\n<h3 class=\"wp-block-heading\">Conclusions<\/h3>\n\n\n\n<p class=\"wp-block-paragraph\">As electronic cigarettes explode in popularity, public attention is rapidly turning toward consumer safety. While research to date has focused primarily on the components of e-cigarette solutions, data presented here indicate a need for substantially more research into the chemical profile of vapor samples. To that end, this study included development of analytical methods for both solution and vapor samples. All three methods developed in the current study used an Rtx-VMS column\u2014a proprietary phase to Restek\u2014which was chosen to reduce required resources and afford easy comparison of results.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">For e-cigarette solutions, rapid GC-FID methods using helium or hydrogen carrier gas were established for the determination of nicotine content. These methods would be suitable for fast quality control testing of electronic cigarette solutions. In addition, a straightforward GC-MS method was developed for the determination of impurities in e-cigarette solutions. Results showed that electronic cigarette solutions contained numerous compounds in addition to the compounds listed on the label by the vendor (propylene glycol, glycerin, and nicotine). In this study, e-cigarette solution profiles revealed 64 identified (some only tentatively) and unidentified compounds, far more than the three that were listed on the product label.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">In order to analyze vapor samples, a simple yet novel sampling device was developed to draw electronic cigarette vapor into a thermal desorption tube, which was then thermally extracted and analyzed via a GC-MS method. This approach provided detectable levels of 82 VOCs and SVOCs from a single 40 mL puff and can be easily implemented by labs that do not have access to a smoking machine. Notably, some of compounds found are known to be detrimental to human health. These compounds were detected in the vapor, but not in the e-cigarette solution, which indicates they were produced during the vaporization process.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">It is unequivocal that electronic cigarette solutions, and more important\u2014vapor\u2014have numerous compounds beyond the ingredients listed on the product label. As these compounds have potential implications for human health, the scientific community needs to place more emphasis on vapor testing in order to definitively identify the chemicals present and to determine how typical usage patterns relate to human health exposure limits.<\/p>\n\n\n\n<h2 class=\"wp-block-heading\">Acknowledgements<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">Markes International Inc., 11126-D Kenwood Road, Cincinnati, OH 45242<\/p>\n\n\n\n<h2 class=\"wp-block-heading\">References<\/h2>\n\n\n\n<ol class=\"wp-block-list\">\n<li>A. Trtchounian, M. Williams, P. Talbot, Conventional and electronic cigarettes (e-cigarettes) have different smoking characteristics, Nicotine Tob Res 12 (2010) 905.<\/li>\n\n\n\n<li>C.J. Brown, J.M. Cheng, Electronic cigarettes: product characterisation and design considerations, Tob Control 23 Suppl 2 (2014) ii4.<\/li>\n\n\n\n<li>J.K. Pepper, T. Eissenberg, Waterpipes and Electronic Cigarettes: Increasing Prevalence and Expanding Science, Chemical Research in Toxicology 27 (2014) 1336.<\/li>\n\n\n\n<li>K.E. Farsalinos, G. Romagna, D. Tsiapras, S. Kyrzopoulos, V. Voudris, Evaluating nicotine levels selection and patterns of electronic cigarette use in a group of &#8220;vapers&#8221; who had achieved complete substitution of smoking, Subst Abuse 7 (2013) 139.<\/li>\n\n\n\n<li>C. Bullen, C. Howe, M. Laugesen, H. McRobbie, V. Parag, J. Williman, N. Walker, Electronic cigarettes for smoking cessation: a randomised controlled trial, Lancet 382 (2013) 1629.<\/li>\n\n\n\n<li>P. Caponnetto, D. Campagna, F. Cibella, J.B. Morjaria, M. Caruso, C. Russo, R. Polosa, EffiCiency and Safety of an eLectronic cigAreTte (ECLAT) as tobacco cigarettes substitute: a prospective 12-month randomized control design study, PLoS One 8 (2013) e66317.<\/li>\n\n\n\n<li>M.L. Goniewicz, J. Knysak, M. Gawron, L. Kosmider, A. Sobczak, J. Kurek, A. Prokopowicz, M. Jablonska-Czapla, C. Rosik-Dulewska, C. Havel, P. Jacob III, N. Benowitz, Levels of selected carcinogens and toxicants in vapour from electronic cigarettes, Tob Control 23 (2014) 133.<\/li>\n\n\n\n<li>W. Schober, K. Szendrei, W. Matzen, H. Osiander-Fuchs, D. Heitmann, T. Schettgen, R.A. Jorres, H. Fromme, Use of electronic cigarettes (e-cigarettes) impairs indoor air quality and increases FeNO levels of e-cigarette consumers, Int J Hyg Environ Health 217 (2014) 628.<\/li>\n\n\n\n<li>L. Kosmider, A. Sobczak, M. Fik, J. Knysak, M. Zaciera, J. Kurek, M.L., Goniewicz, Carbonyl compounds in electronic cigarette vapors: effects of nicotine solvent and battery output voltage, Nicotine Tob Res 16 (2014) 1319.<\/li>\n\n\n\n<li>E-Cig Reviews on the Best E-Cigarettes of 2014, Ecigarette Reviewed (2014).&nbsp;<a href=\"http:\/\/ecigarettereviewed.com\/\" target=\"_blank\" rel=\"noreferrer noopener\">http:\/\/ecigarettereviewed.com<\/a>&nbsp;(Accessed January 12, 2015).<\/li>\n\n\n\n<li>K.E. Farsalinos, R. Polosa, Safety evaluation and risk assessment of electronic cigarettes as tobacco cigarette substitutes: a systematic review, Ther Adv Drug Saf 5 (2014) 67.<\/li>\n\n\n\n<li>K.E. Farsalinos, G. Romagna, D. Tsiapras, S. Kyrzopoulos, V. Voudris, Evaluation of electronic cigarette use (vaping) topography and estimation of liquid consumption: implications for research protocol standards definition and for public health authorities&#8217; regulation, Int J Environ Res Public Health 10 (2013) 2500.<\/li>\n\n\n\n<li>N. Ochiai, A. Tsuji, N. Nakamura, S. Daishima, D.B., Cardin, Stabilities of 58 volatile organic compounds in fused-silica-lined and SUMMA polished canisters under various humidified conditions, J Environ Monit 4 (2002) 879.<\/li>\n\n\n\n<li>NIST Mass Spectrometry Data Center, U.S. Department of Commerce, 2014.<\/li>\n\n\n\n<li>A. Rodgman, T.A. Perfetti, The Chemical Components of Tobacco and Tobacco Smoke, CRC Press, 2nd ed., 2013.<\/li>\n\n\n\n<li>V.J. Cogliano, Y. Grosse, R.A. Baan, K. Straif, M.B. Secretan, F. El Ghissassi, Meeting report: summary of IARC monographs on formaldehyde, 2-butoxyethanol, and 1-tert-butoxy-2-propanol., Environ Health Perspect 113 (2005) 1205.<\/li>\n\n\n\n<li>IARC Working Group Lyon, 13-20 October 1987, Alcohol drinking, IARC Monogr Eval Carcinog Risks Hum 44 (1988) 1.<\/li>\n\n\n\n<li>Y.S. Stein, M.J. Antal, M. Jones, A study of the gas-phase pyrolysis of glycerol, Appl Pyrolysis, 4 (1983) 283.<\/li>\n\n\n\n<li>Toxicological Profile for Acrolein, Agency for Toxic Substances and Disease Registry, U.S. Department of Health and Human Services, Atlanta, 2007.<\/li>\n\n\n\n<li>Electronic Cigarettes (e-Cigarettes), U.S. Food and Drug Administration (2014)&nbsp;(Accessed January 13, 2015).<\/li>\n<\/ol>\n\n\n\n<div style=\"height:30px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n        <div class=\"cpb\">\n            <h3 class=\"cpb-heading\">\u3053\u306e\u8a18\u4e8b\u3067\u7d39\u4ecb\u3057\u305f\u88fd\u54c1<\/h3>\n            <hr class=\"cpb-heading-underline\" \/>\n            <div class=\"cpb-list\">\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/34085\">                                \u30ab\u30bf\u30ed\u30b0No. 34085                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Nicotine Standard, 1000 \u00b5g\/mL, Methanol, 1 mL\/ampul<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/34085\">\u88fd\u54c1\u60c5\u5831\u3092\u898b\u308b<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/30074\">                                \u30ab\u30bf\u30ed\u30b0No. 30074                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">8260 Internal Standard Mix, 2500 \u00b5g\/mL, P&amp;T Methanol, 1 mL\/ampul<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/30074\">\u88fd\u54c1\u60c5\u5831\u3092\u898b\u308b<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/19915\">                                \u30ab\u30bf\u30ed\u30b0No. 19915                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Rtx-VMS GC Capillary Column, 30 m, 0.25 mm ID, 1.40 \u00b5m<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/19915\">\u88fd\u54c1\u60c5\u5831\u3092\u898b\u308b<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/23305\">                                \u30ab\u30bf\u30ed\u30b0No. 23305                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Topaz, Precision Inlet Liner, 4.0 mm x 6.3 x 78.5, for Agilent GCs, w\/Quartz Wool, Premium Deactivation, 5-pk.<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/23305\">\u88fd\u54c1\u60c5\u5831\u3092\u898b\u308b<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/24761\">                                \u30ab\u30bf\u30ed\u30b0No. 24761                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">SGE \u30ac\u30b9\u30bf\u30a4\u30c8\u30b7\u30ea\u30f3\u30b8 50 mL\uff5cPTFE\u30c1\u30c3\u30d7\u4ed8\u304d\uff5c\u30eb\u30a2\u30fc\u30ed\u30c3\u30af, \u30cb\u30fc\u30c9\u30eb\u5225\u58f2<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/24761\">\u88fd\u54c1\u60c5\u5831\u3092\u898b\u308b<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/34396\">                                \u30ab\u30bf\u30ed\u30b0No. 34396                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Linde Air Standard, 75 Comp TO15 + NJ Mix, 1 ppm in N2, 98 L @ 1700 psig<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/34396\">\u88fd\u54c1\u60c5\u5831\u3092\u898b\u308b<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/34420\">                                \u30ab\u30bf\u30ed\u30b0No. 34420                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Linde Air Standard, Ozone Precursor Mix\/PAMS, 1 ppm in N2, 98 L @ 1700 psig<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/34420\">\u88fd\u54c1\u60c5\u5831\u3092\u898b\u308b<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/27309\">                                \u30ab\u30bf\u30ed\u30b0No. 27309                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">SilcoCan Air Sampling Canister, 6 L, with 3-Port Siltek-Treated RAVE+ Valve with Gauge<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/27309\">\u88fd\u54c1\u60c5\u5831\u3092\u898b\u308b<\/a>\n                                                    <\/div>\n                    <\/div>\n                            <\/div>\n        <\/div>\n        ","protected":false},"excerpt":{"rendered":"<p>Despite the growing popularity of electronic cigarettes, relatively little work has been done to characterize their vapor. In this study, we developed a quick, simple thermal desorption device for sampling vapor, as well as methods for analyzing vapor and solutions to determine nicotine content (GC-FID) and impurity profiles (GC-MS). This novel approach, which utilizes an Rtx-VMS column, provides detectable levels of volatile organic compounds (VOCs) and semivolatile organic compounds (SVOCs) from a single 40 mL puff.<\/p>\n","protected":false},"author":25,"featured_media":6829,"comment_status":"closed","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"_acf_changed":false,"_kad_blocks_custom_css":"","_kad_blocks_head_custom_js":"","_kad_blocks_body_custom_js":"","_kad_blocks_footer_custom_js":"","_kadence_starter_templates_imported_post":false,"_kad_post_transparent":"","_kad_post_title":"","_kad_post_layout":"","_kad_post_sidebar_id":"","_kad_post_content_style":"","_kad_post_vertical_padding":"","_kad_post_feature":"","_kad_post_feature_position":"","_kad_post_header":false,"_kad_post_footer":false,"footnotes":"","_ppma_block_editor_authors":""},"categories":[782],"tags":[],"industries-application":[2214,2216],"post-badge":[],"resource-type":[],"product-library":[2443,2444,2452,2456,2458,2446,2447],"resource-technique":[2316,2320],"ppma_author":[445,452,511],"class_list":["post-45256","post","type-post","status-publish","format-standard","has-post-thumbnail","hentry","category-notes-dapplication","industries-application-industrie-agroalimentaire","industries-application-securite-alimentaire","product-library-echantillonnage-dair-et-de-gaz","product-library-produits-pour-la-chromatographie-en-phase-gazeuse","product-library-etalons-gazeux-accessoires","product-library-colonnes-gc","product-library-pieces-de-rechange-pour-injecteur","product-library-etalons-de-reference","product-library-produits-pour-la-preparation-dechantillons","resource-technique-chromatographie-en-phase-gazeuse-gc","resource-technique-spectrometrie-de-masse-ms"],"acf":[],"taxonomy_info":{"category":[{"value":782,"label":"Notes d\u2019application"}],"industries-application":[{"value":2214,"label":"Industrie Agroalimentaire"},{"value":2216,"label":"S\u00e9curit\u00e9 alimentaire"}],"product-library":[{"value":2443,"label":"\u00c9chantillonnage d'air et de gaz"},{"value":2444,"label":"Produits pour la chromatographie en phase gazeuse"},{"value":2452,"label":"\u00c9talons gazeux et accessoires"},{"value":2456,"label":"Colonnes GC"},{"value":2458,"label":"Pi\u00e8ces de rechange pour injecteur"},{"value":2446,"label":"\u00c9talons de R\u00e9f\u00e9rence"},{"value":2447,"label":"Produits pour la pr\u00e9paration d'\u00e9chantillons"}],"resource-technique":[{"value":2316,"label":"Chromatographie en phase gazeuse (GC)"},{"value":2320,"label":"Spectrom\u00e9trie de masse (MS)"}]},"featured_image_src_large":["https:\/\/discover.restek.com\/wp-content\/uploads\/feature-FFAN2127-1024x536.jpg",1024,536,true],"author_info":{"display_name":"Jason S. Herrington, PhD","author_link":"https:\/\/discover.restek.com\/ja\/author\/jason-s-herrington-phd\/"},"comment_info":0,"category_info":[{"term_id":782,"name":"Notes d\u2019application","slug":"notes-dapplication","term_group":0,"term_taxonomy_id":782,"taxonomy":"category","description":"","parent":0,"count":88,"filter":"raw","cat_ID":782,"category_count":88,"category_description":"","cat_name":"Notes d\u2019application","category_nicename":"notes-dapplication","category_parent":0}],"tag_info":false,"authors":[{"term_id":445,"user_id":25,"is_guest":0,"slug":"jason-s-herrington-phd","display_name":"Jason S. Herrington, PhD","avatar_url":{"url":"https:\/\/discover.restek.com\/wp-content\/uploads\/people-herrington-jason.png","url2x":"https:\/\/discover.restek.com\/wp-content\/uploads\/people-herrington-jason.png"},"author_category":"1","first_name":"Jason","last_name":"Herrington","user_url":"https:\/\/www.linkedin.com\/in\/jason-herrington-97b0b620\/%20","job_title":"VP, Reference Standards Business","description":"Jason joined Restek in 2011 after spending over 10 years in environmental analysis, most notably as a postdoctoral research fellow with the U.S. EPA focusing on the development of techniques for organic speciation of ambient gas-phase and particulate air toxics. He has a BS in environmental sciences from Rutgers University and a PhD in exposure science from The University of Medicine and Dentistry of New Jersey. His wide-ranging work has been published in over a dozen peer-reviewed manuscripts and presented around the world. As a senior scientist with Restek, Jason is heavily involved with developing new air products and applications."},{"term_id":452,"user_id":32,"is_guest":0,"slug":"colton-myers","display_name":"Colton Myers","avatar_url":{"url":"https:\/\/discover.restek.com\/wp-content\/uploads\/people-myers-colton.jpg","url2x":"https:\/\/discover.restek.com\/wp-content\/uploads\/people-myers-colton.jpg"},"author_category":"1","first_name":"Colton","last_name":"Myers","user_url":"https:\/\/www.linkedin.com\/in\/myerscolton\/","job_title":"Product Manager - New Product Development","description":"Colton Myers is a Product Manager for GC columns at Restek Corporation with over 10 years of experience in research and development. He specializes in driving new GC column technologies from concept to market, combining deep technical expertise with a strong focus on innovation and commercialization. Colton holds a BS in chemistry from Juniata College and a Master of Management in Corporate Innovation and Entrepreneurship from Penn State University."},{"term_id":511,"user_id":85,"is_guest":0,"slug":"amanda-rigdon","display_name":"Amanda Rigdon","avatar_url":{"url":"https:\/\/discover.restek.com\/wp-content\/uploads\/people-rigdon-amanda.jpg","url2x":"https:\/\/discover.restek.com\/wp-content\/uploads\/people-rigdon-amanda.jpg"},"author_category":"1","first_name":"Amanda","last_name":"Rigdon","user_url":"https:\/\/www.linkedin.com\/in\/amanda-itnyre-rigdon-b06aa84\/","job_title":"Principal Scientist","description":"Amanda Rigdon is an analytical chemist with more than 21 years of experience in chromatography and mass spectrometry. Throughout her career, she has worked in a variety of analytical fields, including clinical forensics, food safety, environmental analysis, and cannabis testing. Her experience spans method development, validation, troubleshooting, and laboratory operations using both GC and LC methodologies. Amanda also served as Chief Scientific Officer of an ISO\/IEC 17025-accredited cannabis testing laboratory, where she gained extensive experience in quality systems, regulatory compliance, and laboratory management. She is currently a Principal Scientist in the GC Applications Group at Restek, where she enjoys helping customers solve analytical challenges and sharing practical chromatography knowledge through technical support, presentations, and educational content."}],"_links":{"self":[{"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/posts\/45256","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/users\/25"}],"replies":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/comments?post=45256"}],"version-history":[{"count":8,"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/posts\/45256\/revisions"}],"predecessor-version":[{"id":73218,"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/posts\/45256\/revisions\/73218"}],"wp:featuredmedia":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/media\/6829"}],"wp:attachment":[{"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/media?parent=45256"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/categories?post=45256"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/tags?post=45256"},{"taxonomy":"industries-application","embeddable":true,"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/industries-application?post=45256"},{"taxonomy":"post-badge","embeddable":true,"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/post-badge?post=45256"},{"taxonomy":"resource-type","embeddable":true,"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/resource-type?post=45256"},{"taxonomy":"product-library","embeddable":true,"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/product-library?post=45256"},{"taxonomy":"resource-technique","embeddable":true,"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/resource-technique?post=45256"},{"taxonomy":"author","embeddable":true,"href":"https:\/\/discover.restek.com\/ja\/wp-json\/wp\/v2\/ppma_author?post=45256"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}