4-Minute Bisphenol A (BPA) Analysis Increases Sample Throughput

Featured Application: Fast Analysis of Bisphenol A on a Raptor Biphenyl LC Column

14 Oct 2020

feature EVSS2395
  • Bisphenol A elutes in less than 2 minutes, increasing lab productivity.
  • Narrow, symmetrical peak shape means easy, accurate quantitation.
  • Alternate selectivity compared to a C18 allows easier identification in difficult matrices.
  • Fast separation regardless of instrument platform—compatible with both HPLC and UHPLC systems.

Bisphenol A (BPA) is a high production volume industrial chemical used to make polycarbonate plastics and epoxy resins that are used in many industries. BPA is released into the environment through industrial effluents and can also leach from packaging materials into food and beverage products. Although BPA concentrations are often low, repeated daily exposure can result in high cumulative doses. Interest in BPA analysis is growing as bisphenol A is an endocrine disruptor that can mimic estrogen, thus disrupting normal hormone levels and causing a significant impact on health. It is also potentially linked to obesity, diabetes, ADHD, and other health conditions. BPA HPLC analysis can be done using a C18 LC column, but analysis times can be long and peak shape often is not ideal.

Restek’s Raptor Biphenyl column is a better alternative than a C18 column for BPA analysis. As shown in the chromatogram below, a Raptor Biphenyl column provides a fast and selective means to isolate and quantitate bisphenol A by LC-MS/MS. It generates narrow, symmetrical peaks in a fast 4-minute analysis time, with bisphenol A eluting in less than 2 minutes. In addition, Raptor Biphenyl columns can be used with both conventional HPLC and UHPLC instruments. For BPA analysis, Raptor Biphenyl columns are recommended because they provide better peak shapes with short analysis times that allow labs to increase sample throughput and overall productivity.

Analysis of Bisphenol A on Raptor Biphenyl by LC-MS/MS

LC_GN0559

Peaks

PeakstR (min)Precursor IonQuantifer Product IonQualifier Product Ion
1.Bisphenol A (BPA)1.27227.1212.1133.1

Conditions

ColumnRaptor Biphenyl (cat.# 9309552)
Dimensions:50 mm x 2.1 mm ID
Particle Size:5 µm
Temp.:30 °C
Standard/Sample
Diluent:50:50 Mobile phase A:mobile phase B
Conc.:100 ng/mL
Inj. Vol.:10 µL
Mobile Phase
A:5 mM Ammonium acetate in water
B:Acetonitrile
Time (min)Flow (mL/min)%A%B
0.000.57525
2.000.50100
2.010.57525
4.000.57525
DetectorMS/MS
Ion Mode:ESI-
Mode:MRM
InstrumentHPLC

Products Mentioned


Raptor Biphenyl, 5 µm, 50 x 2.1 mm HPLC Column

Author

  • Restek Corporation

    Restek is a leading provider of chromatography columns, accessories, and certified reference materials. Trust Restek for reliable, high-quality analytical solutions.

    View Articles
EVSS2395
Version: A