文章

Molecular Sieve 5A and 13X Packed GC Columns for Permanent Gas Separations

28 Oct 2020

feature PCAR2969

Molecular sieves, synthetic forms of Zeolite packing, have been used since the 1950s for separating light gases (oxygen, nitrogen, methane, and carbon monoxide) and inert gases (helium, argon, neon, krypton, and xenon), yet there have been few changes or improvements in their basic composition or performance. Restek has designed the two most common molecular sieves—5A and 13X, as high-performance packings that feature batch-to-batch inertness and consistency. These packings are used in a wide range of packed and PLOT column analyses.

Molecular sieve 5A and 13X packings differ in pore size and composition, causing differences in retention and selectivity for many gases. The 5A packing provides greater retention, which improves the separation of argon, oxygen, and nitrogen, and is a better choice for analyzing the trace impurities in inert gases typically used in the semiconductor industry. The 13X packing often is preferred for analysis of carbon monoxide, particularly at trace concentrations, because lower retention results in sharper chromatographic peaks and improved detection limits.

Molecular Sieve 5A or 13X Packing in Silcosteel Treated Tubing Exhibits Excellent Chromatography for Permanent Gas Analysis

Figure 1 compares a gas mixture analyzed using a 1-meter, high-performance molecular sieve 5A column and a 2-meter, high-performance molecular sieve 13X column. Notice the excellent peak symmetry for carbon monoxide on each column, indicating that the packing and column tubing are extremely inert. Combining high-quality molecular sieves with Silcosteel-treated tubing produces a packed column optimized for trace analysis of the difficult components oxygen and carbon monoxide.

Figure 1: Permanent gases on Molecular Sieve 5A and Molecular Sieve 13X packed columns.
Permanent Gases on Molecular Sieve 5A and Molecular Sieve 13X

GC_PC00166

Peaks

Peaks
1.Oxygen
2.Nitrogen
3.Methane
4.Carbon monoxide

Conditions

ColumnsMolesieve 5A 1 m, 2 mm ID (cat.# 80440-800)
and Molesieve 13X 2 m, 2 mm ID (cat.# 80439-800)
Standard/Sample5-10% each component, balance helium
Injection
Inj. Vol.:10 µL Direct
Inj. Temp.:150 °C
Oven
Oven Temp.:50 °C
Carrier GasHe, constant flow
Flow Rate:30 mL/min
DetectorTCD @ 200 °C

Quality Testing Ensures Reproducibility

Each lot of Zeolite packing is purified to remove metals and other trace contaminants that can adsorb trace gases, such as carbon monoxide. Then, every batch is carefully classified to precise mesh ranges, ensuring lot-to-lot reproducibility. Our thermal conditioning process is critical for establishing the relative retention of methane/carbon monoxide and for producing excellent peak symmetry for active compounds like carbon monoxide. As a result, each lot of molecular sieve is precisely produced, conditioned, and QA tested for reproducible efficiency, column back pressure, peak symmetry, and relative retention time, using a permanent gas performance mixture.

Author

  • Restek Corporation

    Restek is a leading provider of chromatography columns, accessories, and certified reference materials. Trust Restek for reliable, high-quality analytical solutions.

    View all posts
PCAR2969